  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Glendora",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63412",
      "recalling_firm": "Ormco/Sybronendo",
      "address_1": "1332 S Lone Hill Ave",
      "address_2": "N/A",
      "postal_code": "91740-5339",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution, including Nationwide (USA) and the countries of Canada, Austria, Belgium, Bulgaria, Switzerland, Germany, Denmark, France, Estonia, Spain, Finland, Iceland, United Kingdom, Greece, Croatia, Hungary, Ireland, Italy, Luxembourg, Netherlands, Monaco, Poland, Romania, Serbia, Slovenia, Turkey, Jersey, and Martinique.",
      "recall_number": "Z-0205-2013",
      "product_description": "Vector TAS Modular Driver, a manual driver for Vector TAS dental screws.",
      "product_quantity": "4,543",
      "reason_for_recall": "The Instructions For Use (IFUs) for Vector TAS Modular Driver for some foreign language sections incorrectly lists the sterilization temperature as 130¿C. The correct sterilization temperature should be listed as 132¿C and/or 270¿F.",
      "recall_initiation_date": "20111212",
      "center_classification_date": "20121105",
      "termination_date": "20121105",
      "report_date": "20121114",
      "code_info": "Part Numbers 601-0007 and 601-0010, all lots"
    }
  ]
}