  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69712",
      "recalling_firm": "Philips Medical Systems, Inc.",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide), Australia, Austria, Bahrain, Belgium, Brunei Darussalam, Canada, China, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Republic of Korea, Libyan Arab Jamahiriya, Lithuania, Mexico, Namibia, Netherlands, Palestine, Philippines, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates and United Kingdom.",
      "recall_number": "Z-0204-2016",
      "product_description": "Philips HeartStart MRx Monitor/Defibrillators  Models: M3535A and M3536A",
      "product_quantity": "1553 units",
      "reason_for_recall": "1. Device will perform the weekly automated tests hourly, which could cause the therapy capacitors to degrade sooner than intended and 2. While connected to AC or DC power and with no battery installed or the battery installed has a charge level of less than 10%, the Ready for Use (RFU) indicator will not provide the expected low battery indication",
      "recall_initiation_date": "20141119",
      "center_classification_date": "20151103",
      "termination_date": "20160815",
      "report_date": "20151111",
      "code_info": "Serial Numbers:  US00101159  US00322848  US00326834  US00328432  US00328439  US00328442  US00328443  US00328446  US00328450  US00328461  US00328464  US00328465  US00328468  US00328473  US00328478  US00330393  US00533518  US00533521 through US00535118  US00539526  US00540124  US00543102  US00543104  US00543138  US00543158  US00543161  US00543166  US00543167  US00543187  US00543204  US00543223  US00543239  US00546804"
    }
  ]
}