      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72892",
      "recalling_firm": "GE Medical Systems, LLC",
      "address_1": "3200 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1693",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "Worldwide Distribution- US Distribution including the states of AZ, CA, CO, IL IN, MI, MN, NY, OH, OK, SD, TX, WI, VA, DE, FL, MD, ME, MO, NC, NH, OR, PA, TN, WI, DC, PR  and the countries of CANADA, CHINA, CZECH REPUBLIC, ITALY, JAPAN, NETHERLANDS, NEW ZEALAND, PORTUGAL, ROMANIA, SOUTH AFRICA, SWITZERLAND, TURKEY, SWEDEN, AUSTRALIA, BRAZIL, CZECH REPUBLIC, EGYPT, FRANCE, GERMANY, ITALY, MEXICO, RUSSIA, SAUDI ARABIA, TAIWAN, UNITED KINGDOM,  INDONESIA, ESTONIA, REPUBLIC OF KOREA,  POLAND, KUWAIT, SPAIN.",
      "recall_number": "Z-0201-2019",
      "product_description": "GE Healthcare, Discovery 3.0T MR750, 5148810-2, 5373011-2, 5499460-2.",
      "product_quantity": "356 (98 US; 258 OUS)",
      "reason_for_recall": "GE Healthcare has recently become aware of a potential safety issue with the 3.0T 6 Channel Flex Coil used with the 3T MR750W Surgical Suite Scanners.    Coil overheating can occur when the device is used in Mode 2 setup. This could lead to a serious patient thermal injury. There have been no injuries reported as a result of this issue.    UPDATE 5/27/2016:  Recall has been updated to include an additional product which includes the  Discovery 3.0T MR750 systems running Application Software DV24 R01, DV25 R02, or DV25.1 M3.",
      "recall_initiation_date": "20151207",
      "center_classification_date": "20181022",
      "termination_date": "20181108",
      "report_date": "20181031",
      "code_info": "Mfg Lot or Serial # System ID  000000HDMR6188 301496MR3T4  000873824-0007 82427140106  00001105XF503F ASLW750  00001026XF518E 82427180022  00001048XF500F 82427210091  00001106XF5125 82427120092  00001106XF512A 82427120097  00001106XF5105 82427260003  00001105XF5030 82427040143  0000000UA0199A 82427040146  00001307FMM04C 82427200058  00000005271M22 82427210118  00000000395M24 EM0003  00000000WB0792 82427200076  641 434UVAMR  00000950BD62C2 G001MR01  00000914BD5A87 910272064  00001007BD6466 910272071  00001011BD663A 910272072  00001012BD66AE MRWB0558  00001009BD65AC 82427230048  00001009BD65A1 82427210068  00001012BD6696 82427230051  00001010BD6607 600110MR01  00000952BD62F0 EG1793MR01  00000903BD55F9 M4215566  00001002BD639D M40180117  00000914BD5A86 615010MR01  00000934BD5E44 J5274809  00000905BD567C A5279314  00000926BD5D00 2806034DVR750  00001007BD6471 RU3508MR01  00000919BD5B81 SA2180MR02  00001011BD6689 E001MR02  00000848BD5326 K002MR02  00000931BD5D88 853270103  00000944BD6163 853270104  00001009BD65E7 853270109  00001011BD6639 70065MRS03  00000931BD5D89 70328MRS05  00000836BD4DB5 475290MR01  00000291244MR2 MSK65MR2  00000293015MR4 910907WAMR1  00000296461MR7 MR129037  00000297637MR1 650723PSYMR  00000293065MR9 918491MR1  00000291464MR6 217528MR1  00000289961MR5 SINAI750MR  00000290050MR4 281276MR750  00000290659MR2 610740MR3  00000291363MR0 616391DVMR  00000291980MR1 704384DVMR  00000292367MR0 INSIGHTMR3228  00000292471MR0 804675VA3T  00000292634MR3 2486613TMR  00000293874MR4 HSSGMR750  00000292431MR4 541CMRIMR  00000290773MR1 415L3TMR2  00000291877MR9 415L3TMR1  00000298591MR9 414805MR2  00000292843MR0 217788MR3  00000294556MR6 540665MR3T  00000293261MR4 415OI3TMR  00000292678MR0 919684MR5  00000293561MR7 405752I3T  00000294018MR7 903525MR750  00000294057MR5 281274MR1  00000294576MR4 602830MR3  00000294698MR6 812485MR3  00000294665MR5 480768MR2  00000294718MR2 715838MR3T  00000295112MR7 801377MR3  00000295024MR4 207945MR750  00000295111MR9 703391MR2  00000294953MR5 7579533T  00000295228MR1 6036503T  00000296046MR6 309683MNMR750  00000295904MR7 281477MR2  00000295506MR0 858822HDMR1  00000296014MR4 816932DVMR  00000296512MR7 916LST3TMR  00000295215MR8 301295MR4  00000296602MR6 7168453T  00000296821MR2 DUPONT3T  00000294398MR3 713441MR9  00000297469MR9 954772MR1  00000296517MR6 615MTMR1  00000297974MR8 810235DVMR  00000297624MR9 415493VMR2  00000296908MR7 386FMR3T  00000290502MR4 MR129030  00000298420MR1 6162673T750  00000298537MR2 832522MR1  00000298765MR9 9155693TMR  00000298981MR2 480425MR6  00000297374MR1 650PH3T  00000298738MR6 RADNETMR1593  00000299181MR8 703858MR3T  00000299213MR9 303699MR2  00000299250MR1 847618MR1  00000298580MR2 570271MR6  00000288907MR9 MR129035  00000299411MR9 920288MR750  00000290605MR5 661KRMR750  00000291154MR3 519685MR3T  00000291008MR1 416486MR3T  00000298951MR5 514345BMR1  00000296265MR2 416595MR1  00000288909MR5 835270017  00000295757MR9 F5119155  00000291429MR9 910274032  00000297898MR9 2875893  00000295060MR8 MR13905YR5  00000296291MR8 EG1000MR03  00000294713MR3 EE4019MR02  00000290985MR1 M4484966  00000295131MR7 GOLD0DEM01  00000288156MR3 PC7479MR08  00000293710MR0 H10645MR17  00000294545MR9 A51152177  00000291345MR7 A5108257  00000299147MR9 EM0004  00000298566MR1 EM0007  00000298331MR0 850270341  00000292216MR9 RTD0125  00000288409MR6 920279028  00000296622MR4 PL1803MR01  00000290456MR3 70383MRS05  00000291274MR9 10307MRS03  00000298501MR8 00269MRS04  00000309742MR5 82427040165  00001106XF511E 202476MR3  00001106XF5055 82427040116  00001106XF5042 82427150038  00001018XF5019 82427040119  00001101XF5019 82427020036  00001106XF502D 82427090148  00001106XF50C3 82427140076  00001106XF5073 82427040120  00001101XF5016 82427040123  00001044XF5030 82427140074  00001106XF50B0 82427030076  00001106XF5043 82427040127  00001101XF5027 82427100147  00001106XF506B 82427230055  00001106XF50D7 82427010056  00001106XF507A 82427100149  00001106XF504D 82427090169  00001106XF508A 82427150047  00001106XF50C5 82427020042  00001106XF50DF 82427040132  00001106XF50D0 82427140087  00001106XF50DD 82427210081  00001106XF507B 82427060042  00001106XF50B6 82427270037  00001101XF5033 82427130057  00001106XF508B 82427040128  00001106XF5089 82427050029  00001101XF5035 82427070075  00001106XF50B4 82427050028  00001105XF502F 82427160088  00001106XF5090 82427020039  00001106XF508E 82427070079  00001106XF5104 82427100184  00001105XF502B C4200438  00001105XF5031 B4498422  00001106XF5064 M4155416  00001106XF5114 M4152248  000001106XF039 RMUE01MR03  00001044XF5033 8.3027E+13  00001026XF5151 IL1020MR04  00001106XF5094 EM0002  00001106XF5068 EM0022  00001026XF515D EM0005  00001027XF5017 EM0006  00001026XF5176 EM0009  00001106XF502E EM0014  00001106XF506C EM0018  00001027XF501B EM0017  00001026XF516F EM0016  00001101XF5020 EM0019  00001106XF507C EM0020  00001106XF503E EM0023  00001106XF5134 EM0030  00001106XF510E EM0013  00001044XF5031 EM0010  00001106XF503D 850270351  00001101XF5039 850270370  00001044XF5035 KW1102MR01  00001026XF5174 MRWB0632  00001106XF507E RTD0156  00001105XF5047 PL2853MR01  00001106XF502F RU1857MR03  00001101XF5022 RU1086MR03  00001101XF501C SA1012MR02  00001101XF5018 CPZ43623  00001106XF505A AC001MR01  00001106XF5076 541010MR04  00001101XF5037 853270111  00001106XF5035 853270118  00001106XF5078 853270119  00001106XF50C6 853270125  00001101XF503B 00040MRS01  00000302690MR3 HCIMR1805  00000299291MR5 205592750  00000299321MR0 509473DMR3T  00000299515MR7 816698CP3T  00000299966MR2 540MIF3T750  00000299329MR3 209225VC750  00000299544MR7 703805MR3  00000299331MR9 720321MR  00000300713MR5 425688MR  00000301049MR3 707DG3T  00000301752MR2 858822HDMR2  00000300562MR6 202476MR4  00000300732MR5 805687MR1  00000301924MR7 312996MR3T  00000301926MR2 510537MR750  00000303805MR6 312355DVMR  00000302688MR7 607CORNELL3T  00000303439MR4 713566SMITH2  00000304177MR9 301295MR3T2  00000301597MR1 UM750MR  00000304536MR6 904202MR9  00000304836MR0 919684BIAC6  00000301599MR7 703573MR2DV  00000300711MR9 514848MR1  00000300563MR4 905522MR3T  00000301753MR0 82427040121  00000299454MR9 B51142110  00000301194MR7 EM0012  00000301925MR4 EM0015  00000300734MR1 EM0021  00000299419MR2 850270347  00000300733MR3 850270367  00000302922MR0 MR302922MR0  00000303872MR6 MR303872MR6  00000304910MR3 AMS0424  00000305118MR2 MR16061YR4  00000300792MR9 853270117  00000301296MR0 70043MRS05  00000307132MR1 7347633TMRFIX  00000309403MR4 918491MR2  00000306640MR4 305284MR1  00000030493MR0 602839MR1  00000306641MR2 212543MR750  00000308357MR3 650780MR750  00000310010MR4 82427140107  00000309404MR2 82427110049  309131MR1 82427160128  00000309560MR1 82427050034  00000307615MR5 82427150062  00000309158MR4 82427100213  00000309626MR0 82427040155  00000310274MR6 82427100229  00000310044MR3 82427120132  00000309405MR9 82427270050  00000310043MR5 82427130078  00000306734MR5 82427140094  00000306732MR9 82427040133  00000306731MR1 82427310057  00000307111MR5 82427040142  00000307131MR3 82427310062  00000308161MR9 82427090202  00000308162MR7 82427040137  00000308210MR4 82427160114  00000310285MR2 82427120122  00000310334MR8 82427210129  00000310608MR5 82427010080  00000310845MR3 82427200086  00001302FMM04W 82427210116  00001302FMM04A 82427040114  00001245FMM03L 82427010072  00001245FMM04P 82427210111  00001302FMM04Y 82427130072  00001245FMM03W 82427130071  00001245FMM04R 82427100202  00001302FMM053 82427010073  00001302FMM04C 82427140098  00001302FMM04M 82427040141  00001302FMM057 82427120101  00001302FMM04F 82427040148  00001201FMM089 82427040153  00001246FMM03C 82427120107  00001302FMM050 82427310059  00001245FMM04H 82427210121  00001307FMM08G 82427160118  00001307FMM07W 82427040154  00001302FMM04U 82427160101  00001307FMM07V 82427020056  00001302FMM056 82427100215  00001318FMM009 82427030092  00001406NMM02Y 82427120129  00001307FMM084 82427040163  00001318FMM006 82427040161  00001406NMM03H 82427030094  00001302FMM044 82427150058  00001302FMM04J 82427050033  00001302FMM052 82427180027    EM0024  00001215FMMOMD EM0025  00001245FMM044 10197MRS03  00001301FMM048 MRUA0184  00001308FMM047 82427060052  00001308FMM04N 82427120124  00001236FMM01N 82427210109  00001251FMM013 82427040135  00001301FMM04N 82427120110  00001251FMM01L 82427120116  00001308FMM059 82427100219  00001308FMM014 A4171551  00001250FMM02V MR102094WH0  00001308FMM00N EM0001  00001147FMM0EE EM0222  00001236FMM01R EM0214  00001227FMM00K EM0011  00001309FMM08Y EM0243  00001204FMM04P DPN26332  00001308FMM00A A010MR07  00001308FMM037 70001MRS11  N/A 82427060055  N/A 82427160149  N/A 82427120135  N/A 82427160127  N/A 82427010087  N/A 82427130076  N/A 82427010094  N/A 82427030100  N/A 82427040150  N/A 82427040151  N/A 82427040172  N/A 82427070137  N/A 82427090241  N/A 82427100223  N/A 82427100225  N/A 82427100235  N/A 82427100241  N/A 82427110051  N/A 82427110058  N/A 82427120131  N/A 82427120133  N/A 82427180028  N/A 82427130075  N/A 82427140113  N/A 82427020059  N/A 82427160136  N/A 82427210130  N/A 82427210136  N/A 82427220040"
    }
  ]
}