  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Tustin",
      "address_1": "2441 Michelle Dr",
      "reason_for_recall": "Toshiba America Medical Systems, Inc. is recalling due to a potential problem with Specific Absorption Rate (SAR)  values when using the Toshiba Titan 3T System.   When the system is used in combination with either the QD Head Coil (MJQH-142A) or the Large Knee  SPEEDER (MJAJ-182A), SAR values lower than the actual SAR values may be displayed on the operating  monitor.",
      "address_2": "P.O. Box 2068",
      "product_quantity": "18 units",
      "code_info": "Lot numbers not utilized. Serial Numbers:A5B1192003  A5B1192004,  A5B11Y2005,  A5B11Z2007,  A5C1212008,  A5C1212009,  A5C1222010,  A5C1232011,  A5C1252012,  A5C1262013,  A5C1282014,  A5C1292015,  A5C12X2016,  A5C12Y2017,  A5C12Z2018,  A5D1312019,  A5D1322020,  A5D1362021,  XIB-A5A1082001,  XIB-A5B11Y2006,  A5D1372022,  A5B1182002.",
      "center_classification_date": "20131112",
      "distribution_pattern": "USA Nationwide Distribution in the states of: CA, FL, GA, IA, NV, NY, LA, IN, PA, OR, TX, and WI.",
      "state": "CA",
      "product_description": "Titan 3T MRI System, MRT-3010A/5.      Product Usage:  Usage:  MRI system.",
      "report_date": "20131120",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Toshiba American Medical Systems Inc",
      "recall_number": "Z-0201-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "66725",
      "termination_date": "20141224",
      "more_code_info": "",
      "recall_initiation_date": "20131028",
      "postal_code": "92780-7047",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}