      "status": "Ongoing",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88820",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-0200-2022",
      "product_description": "Artis zeego, Fluoroscopic X-Ray System  Models: 10280959,10502505, 10848283 in combination with the following workplace/system application software models:  Product/Model #:  syngo X Workplace, 10281061  syngo X Workplace, 10281299  syngo X Workplace, 10502522  Syngo Application Software, 10848815",
      "product_quantity": "43 US",
      "reason_for_recall": "Software error, the software can produce an incorrect interpretation of the table rotation if the table has been rotated by more than 4degrees in either direction. This can cause the 3D image reconstruction to be shown rotated by up to 5 degrees relative to the patient and to be used inappropriately by the system and result in incorrect treatment of the patient",
      "recall_initiation_date": "20210924",
      "center_classification_date": "20211104",
      "report_date": "20211110",
      "code_info": "Artis zeego systems with Artis patient table and AppSW VD20C/VD20N"
    }
  ]
}