      "status": "Ongoing",
      "city": "Tustin",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81026",
      "recalling_firm": "Canon Medical System, USA, INC.",
      "address_1": "2441 Michelle Dr",
      "address_2": "N/A",
      "postal_code": "92780-7047",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic: AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, WA, WI, WV, and WY.    International: Algeria, Australia, Bangladesh, Brazil, Canada, China, Colombia, Congo, Dominican Republic, Egypt, Guinea, Honduras, India, Indonesia,  Iran, Japan, Kenya, Korea, Lebanon, Malaysia, Mexico, Morocco, Netherlands, Nigeria, Oman,  Pakistan, Paraguay, Qatar, Saudi Arabia, Singapore, South Africa, Sudan, Taiwan, Thailand, Turkey, UAE, and Vietnam",
      "recall_number": "Z-0199-2019",
      "product_description": "Nuclear Magnetic Resonance Imaging Systems that are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body.",
      "product_quantity": "252",
      "reason_for_recall": "When using respiratory gating with Steady Mode ON in Spin Echo sequence scanning, the SAR (Specific Absorption Rate) value displayed on the console may be lower than the actual SAR value.  The actual applied SAR values may exceed the FIRST LEVEL CONTROLLED OPERATING MODE.",
      "recall_initiation_date": "20180530",
      "center_classification_date": "20181022",
      "report_date": "20181031",
      "code_info": "MODEL                   SERIAL NUMBER SID  MEXL-1504/U4 U4A12Z2020 30005914  MEXL-1504/T4 S4D1132069 339325  MEXL-1504/S4 S4A1032017 326165  MEXL-1504/U4 U4A1312023 30000547  MEXL-1504/S4 S4A1032019 327601  MEXL-1504/U4 U4A1312024 30002325  MEXL-1504/T4 S4D1152082 330035  MEXL-1504/S4 S4A1082040 330134  MEXL-1504/A4 A4A11Z2011 346866  MEXL-1504/U4 U4A1282003 356105  MEXL-1504/U4 U4A1322025 30009003  MEXL-1504/U4 U4B1442070 30012516  MEXL-1504/U4 U4A1352035 389636  MEXL-1504/U4 U4B1552097 30010318  MEXL-1504/U4 U4A12X2010 351817  MEXL-1504/S4 S4C1112065 330381  MEXL-1504/S4 S4C1112064 334599  MEXL-1504/A4 A4A11Z2009 352039  MEXL-1504/S4.004 S4A1072033 30025756  MEXL-1504/S4 S4A09Y2002 345231  MEXL-1504/U4.004 U4A1282002 30027784  MEXL-1504/U4.004 U4A1382047 30036224  MEXL-1504/S3 S3A0972030 312421  MEXL-1504/U4 U4A12X2011 345074  MEXL-1504/S4 S4C1112061 337212  MEXL-1504/U4 U4A13X2054 30011038  MEXL-1504/A4 A4A1222015 351023  MEXL-1504/U4 U4A1322026 337071  MEXL-1504/U4 U4B1642106 30008395  MEXL-1504/A4 A4A1222017 347963  MEXL-1504/S4 S4A10X2047 315713  MEXL-1504/S3 S3A08Z2019 303586  MEXL-1504/U4 U4B1522092 358163  MEXL-1504/A4 A4A11X2004 30018147  MEXL-1504/T4 S4D1132074 347260  MEXL-1504/U4 U4B1422064 30021502  MEXL-1504/U4 U4A1312022 30004266  MEXL-1504/U4 U4A1352036 30001882  MEXL-1504/U4 U4A1372046 337083  MEXL-1504/T4 S4D1142076 341529  MEXL-1504/U4 U4A13Z2057 30005810  MEXL-1504/A4 A4A11Z2010 349167  MEXL-1504/U4 U4A1292007 349175  MEXL-1504/U4 U4B1632105 366015  MEXL-1504/U4 U4B1492081 376523  MEXL-1504/U4 U4A1352037 389696  MEXL-1504/U4 U4A1312021 353532  MEXL-1504/U4 U4B1482077 392326  MEXL-1504/A4 A4A11X2002 333633  MEXL-1504/U4 U4B14Y2082 30018555  MEXL-1504/U4 U4A13X2051 325434  MEXL-1504/A4 A4A11Y2006 350868  MEXL-1504/A4 A4A1252024 356295  MEXL-1504/U4 U4B1532094 30027785  MEXL-1504/U4 U4B1652107 30009792  MEXL-1504/S3 S3A0922023 299545  MEXL-1504/U4 U4A1382049 30007016  MEXL-1504/U4 U4A12Z2014 356808  MEXL-1504/S3 S3A0892009 301747  MEXL-1504/A4 A4A11X2003 324905  MEXL-1504/S3.004 S3A0882007 30024991  MEXL-1504/T4 S4D1132073 339838  MEXL-1504/U4 U4A1362039 393723  MEXL-1504/U4 U4B1462074 356287  MEXL-1504/U4 U4B1522091 30006245  MEXL-1504/U4 U4A1342033 30004737  MEXL-1504/U4 U4A1342034 30001465  MEXL-1504/U4 U4A12X2009 354860  MEXL-1504/U4 U4A1412059 383292  MEXL-1504/U4 U4A1412063 30003291  MEXL-1504/U4 U4A1282004 356881  MEXL-1504/U4 U4A1352038 30004544  MEXL-1504/U4 U4A12Y2013 359834  MEXL-1504/U4 U4B14Z2086 30023668  MEXL-1504/U4 U1A13Z2001 30007654  MEXL-1504/S3 S3A0942027 314575  MEXL-1504/U4 U4A1272001 357343  MEXL-1504/U4 U4A1292006 384206  MEXL-1504/S4 S4A1042023 308445  MEXL-1504/S3 S3A08Z2016 308437  MEXL-1504/A4 A4A11X2001 342766  MEXL-1504/U4 U4B1512089 30006322  MEXL-1504/A4 A4A1212012 352641  MEXL-1504/U4 U4A1342029 314591  MEXL-1504/T4 S4E1192095 340919  MEXL-1504/U4 U4A1362041 30004910  MEXL-1504/U4 U4A12Y2012 356816  MEXL-1504/U4 U4A1382050 30006997  MEXL-1504/A4 A4A1242021 345363  MEXL-1504/S4 S4A1092046 333104  MEXL-1504/A4 A4A1252023 348078  MEXL-1504/U4 U4B15Y2102 30002102  MEXL-1504/U5 U5B15Z2001 30002103  MEXL-1504/T4 S4D1132071 328690  MEXL-1504/U4 U4A13Y2056 30004937  MEXL-1504/S4 S4A1092043 310631  MEXL-1504/U4 U4A1322027 394526  MEXL-1504/U4 U4B1522093 30007522  MEXL-1504/U4 U4A1412061 389344  MEXL-1504/S3 S3A0872005 298422  MEXL-1504/U4 U4A12Z2017 357400  MEXL-1504/A4 A4A1212013 349118  MEXL-1504/U4 U4B1482076 30020797  MEXL-1504/U4 U4B1552098 30012104  MEXL-1504/A4 A4A1222019 352146  MEXL-1504/S4 S4A1052025 324327  MEXL-1504/U4 U4A1292008 356568  MEXL-1504/A4 A4A1222016 340950  MEXL-1504/U4 U4B1432067 30007279  MEXL-1504/U4 U4A1362042 357418  MEXL-1504/S4 S4A1092045 333047  MEXL-1504/U4 U4B1452071 30009112  MEXL-1504/A4 A4A1242022 351429  MEXL-1504/A4 A4A1262025 394677  MEXL-1504/S4 S4A1082036 331306  MEXL-1504/S4 S4C1112060 340539  MEXL-1504/T4 S4D1152081 342436  MEXL-1504/S4 S4A09X2001 317842  MEXL-1504/U4 U4A12Z2019 356063  MEXL-1504/T4.004 S4D1142079 372281  MEXL-1504/U4 U4B1512090 30011831  MEXL-1504/U4 U4A13X2055 372615  MEXL-1504/U4 U4A1362044 30008519  MEXL-1504/A4 A4A11X2005 343723  MEXL-1504/U4 U4B15X2101 30008684  MEXL-1504/S4 S4A1072032 316471  MEXL-1504/U4 U4B1482079 377483  MEXL-1504/U4 U4B1582100 30004842  MEXL-1504/S4 S4A1022013 313254  MEXL-1504/S4 S4A1022014 313262  MEXL-1504/U4 U4B1612103 30029397  MEXL-1504/T4 S4D1132070 337303  MEXL-1504/U4 U4A1382048 30003156  MEXL-1504/U4 U4B1612104 385256  MEXL-1504/T4 S4E1182093 342311  MEXL-1504/U4 U4A13X2052 380285  MEXL-1504/U4 U4A12Z2015 356840  MEXL-1504/U4 U4B1452072 369092  MEXL-1504/U4 U4B1452073 30007850  MEXL-1504/U4 U4B1582099 30006070  MEXL-1504/S4 S4A1022015 326512  MEXL-1504/U4 U4B14Y2084 365309  MEXL-1504/S3 S3A08Y2013 292490  MEXL-1504/S4 S4B10Z2058 336701  MEXL-1504/S3 S3A09X2039 351542  MEXL-1504/S4 S4A1092044 308189  MEXL-1504/A4 A4A1222018 345587  MEXL-1504/A4.007 A4A1222014 392554  MEXL-1504/T4 S4D1172089 345140  MEXL-1504/U4 U4B1432066 30008575  MEXL-1504/U4 U4B1442069 30007942  MEXL-1504/S4 S4A1082041 329656  MEXL-1504/S3 S3A09X2041 311514  MEXL-1504/A4 A4A11Y2008 348169  MEXL-1504/S3 S3A08Y2012 311506  MEXL-1504/S4 S4A1062030 311522  MEXL-1504/S3 S3A0992038 304907  MEXL-1504/U4 U4B1472075 348151  MEXL-1504/S3 S3A09X2040   MEXL-1504/S4 S4A1042022 326355  MEXL-1504/U4 U4B1512087 30009776  MEXL-1504/T4 S4D1162086 344267  MEXL-1504/U4 U4A1342031 351320  MEXL-1504/U4 U4A1342032 30005999  MEXL-1504/U4 U4A1362040 30007243  MEXL-1504/S4 S4B10Y2053 334979  MEXL-1504/S4 S4C1112066 337048  MEXL-1504/U4 U4A12Z2016 394744  MEXL-1504/S3 S3A0972031 299453  MEXL-1504/U4 U4A1282005 30003217  MEXL-1504/U4 U4B1512088 30027396  MEXL-1504/S4 S4A1052026 336677  MEXL-1504/S4 S4A1012011 299271  MEXL-1504/U4 U4B14Z2085 30020944  MEXL-1504/U4 U4A1332028 30002698  MEXL-1504/U4 U4A13X2053 30008346  MEXL-1504/U4 U4B1442068 30008347  MEXL-1504/U4 U4A1412060 30008348  MEXL-1504/U4 U4A1362045 30008837  MEXL-1504/U4 U4B14Y2083 30008842  MEXL-1504/U4 U4B1482078 30010457  MEXL-1504/U4 U4B1422065 350207  MEXL-1504/U4 U4A1342030 362189  MEXL-1504/S4 S4A09Y2003 306985  MEXL-1504/U4 U4B1492080 30002625  MEXL-1504/S3 S3A0992036 307538  MEXL-1504/U4 U4B1542095 391411  MEXL-1504/U4 U4B1542096 380834  MEXL-1504/U4 U4A1412062 366506  MEXL-1504/S4 S4A1012009 308056  MEXL-1504/U4.004 U4A1282002 30027784  MEXL-1504/U4.004 U4A1382047 30036224  MRT-1504/S4 S4A1052027 330217  MEXL-1504/U4.004 U4A12Z2018   MEXL-1504/U4.004 U4A1362043   MEXL-1504/U4.004 U4A12Z2018   MEXL-1504/U4.004 U4A1362043    Model MEXL-1510  MEXL-1510/S2 S2B1642037  MEXL-1510/S2 S2A15Y2019  MEXL-1510/S2/E S2B16Y2055  MEXL-1510/S2 S2B1652041  MEXL-1510/S2 S2C1762072  MEXL-1510/S3 S3A16Z2007  MEXL-1510/S2/E S2C1772077  MEXL-1510/S2 S2A1592013  MEXL-1510/S3 S3B1782012  MEXL-1510/S2 S2A1612025  MEXL-1510/S2/E S2B1692050  MEXL-1510/S3 S3A1692003  MEXL-1510/S2/E S2C1772075  MEXL-1510/S3 S3B17Y2016  MEXL-1510/S2 S2A15Z2023  MEXL-1510/S2 S2A15Y2018  MEXL-1510/S2/E S2C1772074  MEXL-1510/S2 S2C1782081  MEXL-1510/S3 S3B1762010  MEXL-1510/S2 S2C1762071  MEXL-1510/S3 S3B1762009  MEXL-1510/S2 S2C1782082  MEXL-1510/S2 S2B1632035  MEXL-1510/S2/E S2D1812097  MEXL-1510/S2 S2A1582010  MEXL-1510/S2/E S2B16Z2061  MEXL-1510/S2/E S2B1732065  MEXL-1510/S2 S2B1632036  MEXL-1510/S2 S2B1672045  MEXL-1510/S3 S3B17X2015  MEXL-1510/S2 S2B1632030  MEXL-1510/S3 S3A16Z2006  MEXL-1510/S2/E S2B1642040  MEXL-1510/S2 S2A1612024  MEXL-1510/S2 S2C1752070  MEXL-1510/S2/E S2C1772076  MEXL-1510/S2 S2B1692048  MEXL-1510/S2 S2B1692049  MEXL-1510/S2 S2A1572008  MEXL-1510/S2 S2B1622029  MEXL-1510/S2 S2A1582012  MEXL-1510/S3 S3A16Y2004  MEXL-1510/S2/E S2B16X2054  MEXL-1510/S3 S3B1782013  MEXL-1510/S2 S2A15Y2016  MEXL-1510/S2 S2B1642039  MEXL-1510/S2 S2B16Y2058  MEXL-1510/S2 S2B1632031  MEXL-1510/S2/E S2C17Y2093  MEXL-1510/S2 S2C1772078  MEXL-1510/S3 S3B1772011  MEXL-1510/S3 S3A16Z2005  MEXL-1510/S3 S3B1782014  MEXL-1510/S2 S2B16Z2060  MEXL-1510/S2 S2A1562006  MEXL-1510/S2 S2A15Z2021  MEXL-1510/S2/E S2C1772079  MEXL-1510/S2 S2A1552005  MEXL-1510/S2 S2A1612026  MEXL-1510/S2/E S2C17Y2090  MEXL-1510/S2 S2D1822098  MEXL-1510/S2 S2D1832101  MEXL-1510/S3 S3C1812017  MEXL-1510/S3 S3C1832018    Model MEXL-1520  MEXL-1520/S1 S1C1672334 30032698  MEXL-1520/S1 S1C1682336 30035150  MEXL-1520/S1 S1C1722390 30040716  MEXL-1520/S1 S1D17Y2461 30036289  MEXL-1520/S1 S1B1492068 70000226"
    }
  ]
}