      "country": "United States",
      "city": "Warsaw",
      "address_1": "700 Orthopaedic Dr",
      "reason_for_recall": "Issuing a device correction because if the pin bushing is over loaded or inverted, there is the potential for partial or full displacement and fracture of the plastic around the pin bushing.",
      "address_2": "",
      "product_quantity": "3,082",
      "code_info": "Catalog NOs. 2544-00-521 and 2544-00-520; All lots are affected for the Device Correction.      Lot numbers below should be inspected and swapped out if the bushing is incorrectly assembled:  Part No.                       DePuy Synthes Lot No:                  MFG Lot No:  254400520                          ABC22844                                  C22844                                                 ABC22845                                  C22845                                                 ABC34279                                  C34279                                                 ABC34280                                  C34280                                                 ABC4107                                     C4107                                                 ABC4108                                     C4108                                                 ABC4112                                     C4112                                                 ABC4113                                     C4113                                                 ABC50741                                  C50471  254400521                          ABB13662                                   B13662                                                  ABB19116                                  B19116                                                 ABB19117                                   B19117                                                 ABB19118                                   B19118                                                 ABB35392                                   B35392                                                 ABB35393                                   B35393                                                 ABB39835                                   B39835                                                 ABB41384                                   B41384                                                 ABB46872                                   B46872                                                 ABB49200                                   B49200                                                 ABB50724                                   B50724                                                 ABB55749                                   B55749                                                 ABB63598                                   B63598                                                 ABB66346                                   B66346                                                 ABB72107                                   B72107                                                 ABB77183                                   B77183                                                 ABB79390                                   B79390                                                 ABB79391                                   B79391                                                 ABB80712                                   N80712                                                 ABC17165                                  C17165                                                 ABC5023                                     C5023",
      "center_classification_date": "20141110",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) except AK, including Puerto Rico and the countries of  Canada, Australia, Austria, Belgium, China, France, Germany, India, Ireland, Israel, Italy, Japan, Malaysia, Netherlands, New Zealand, Singapore, South Africa, South Korea, Spain, Switzerland, Thailand, UAE, United Kingdom and Chile.",
      "state": "IN",
      "product_description": "ATTUNE Intuition Distal Femoral Jig; sold non-sterile; product is wrapped in bubble wrap and then sealed inside a polyethylene bag",
      "report_date": "20141119",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "DePuy Orthopaedics, Inc.",
      "recall_number": "Z-0199-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "69271",
      "termination_date": "20151119",
      "more_code_info": "",
      "recall_initiation_date": "20141015",
      "postal_code": "46582-3994",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}