  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Waltham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91005",
      "recalling_firm": "Beaver Visitec International, Inc.",
      "address_1": "500 Totten Pond Rd",
      "address_2": "10 City Point",
      "postal_code": "02451-1916",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "FL, GA, IN, LA, MA, ME, VA Foreign: France, Japan, Netherlands, Switzerland, United Kingdom",
      "recall_number": "Z-0197-2023",
      "product_description": "Beaver Arthro-Lok Pointed Tip Curve L 4mm Blade-an orthopedic blade, which may be used in various types of orthopedic surgery (i.e. foot, ankle, hand, knee, hip or spine surgery).  Catalog#: 376780",
      "product_quantity": "150 units US; 1116 units OUS",
      "reason_for_recall": "Incorrect configuration of the blades in the package. The blades curve to the right instead of curving to the left. If used may cause cause unintended tissue damage during the procedure or if noticed prior to use may cause a delay in surgery",
      "recall_initiation_date": "20220930",
      "center_classification_date": "20221107",
      "report_date": "20221116",
      "code_info": "UDI# 00886158001249 LOT#'s: 6034831,6039837"
    }
  ]
}