  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Plainsboro",
      "address_1": "311 Enterprise Dr",
      "reason_for_recall": "A single lot of left and a single lot of right Reverse Shoulder System cutting templates were manufactured incorrectly.  Specifically, the threaded handle has been welded backwards on the template resulting in the cutting angle being the reverse of what it should be.",
      "address_2": "",
      "product_quantity": "12 units",
      "code_info": "Lot Numbers PM0278 and PM0277",
      "center_classification_date": "20141107",
      "distribution_pattern": "Nationwide Distribution including the states of TX, CA, OH, MS, VA, IL, TN, GA, CA, FL, and AZ.",
      "state": "NJ",
      "product_description": "Integra Titan Reverse Shoulder System right and left head cutting templates, 142.5 degrees.    The Reverse Shoulder System Cutting Template, 142.5 degrees is an accessory for use with the Integra Titan Reverse Shoulder System. It is used to project the osteotomy angle on the humeral head.",
      "report_date": "20141119",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Integra LifeSciences Corp.",
      "recall_number": "Z-0197-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "69596",
      "termination_date": "20150721",
      "more_code_info": "",
      "recall_initiation_date": "20141024",
      "postal_code": "08536-3344",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}