      "country": "United States",
      "city": "Ann Arbor",
      "address_1": "6200 Jackson Rd",
      "reason_for_recall": "During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications on certain product lots of Sarns\"  Soft-Flow¿ Aortic Cannulae and Sarns\" Venous Return Cannulae.",
      "address_2": "",
      "product_quantity": "40 units",
      "code_info": "Catalog Number 9473  Lot Numbers 0689812",
      "center_classification_date": "20131108",
      "distribution_pattern": "Worldwide Distribution-USA including DC and the states of MO,TX,OK, LA, MA, CA, WI, and MI and the country of Canada.",
      "state": "MI",
      "product_description": "Sarns\"  Venous Return Cannulae, 20 Fr with 1/4\" flare, 14.5\" (37 cm) long.    Indicated for venous drainage during cardiopulmonary bypass surgery for dial cannulation of the superior and inferior vena cava.",
      "report_date": "20131120",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "recall_number": "Z-0196-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "66624",
      "termination_date": "20140718",
      "more_code_info": "",
      "recall_initiation_date": "20131023",
      "postal_code": "48103-9586",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}