{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northridge",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90910",
      "recalling_firm": "Medtronic MiniMed",
      "address_1": "18000 Devonshire St",
      "address_2": "N/A",
      "postal_code": "91325-1219",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY.  OUS: Andorra, Australia, Austria, Barbados, Belgium, Bermuda, Bolivia, Botswana, Canada, Canary Islands, Cayman Islands, Chile, Czech Republic, Denmark, Dominican Rep., El Salvador, Faroe Islands, Finland, France, Germany, Gibraltar, Greece, Guatemala, Hungary, Ireland, Israel, Italy, Kenya, Liechtenstein, Luxembourg, Mauritius, Mexico, Namibia, Netherlands, New Zealand, Norway, Paraguay, Peru, Poland, Portugal, Romania, Slovakia, South Africa, Spain, Sweden, Switzerland, Trinidad and Tobago, Turkey, Uganda, United Kingdom, Venezuela",
      "recall_number": "Z-0194-2023",
      "product_description": "Insulin Pump/Model:  MiniMed 620G/ MMT-1750  MiniMed 640G/ MMT-1711, MMT-1712, MMT-1751, MMT-1752",
      "product_quantity": "316326",
      "reason_for_recall": "Medtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems associated with the communication protocol that could allow unauthorized access to the pump system. This unauthorized access could be used to deliver too much or too little insulin through delivery of an unintended insulin bolus or because insulin delivery is slowed or stopped, which could lead to hypoglycemia or hyperglycemia.  The Remote Bolus feature on insulin pumps should be turned off to prevent the unauthorized access.",
      "recall_initiation_date": "20220920",
      "center_classification_date": "20221107",
      "report_date": "20221116",
      "code_info": "Model/UDI-GTIN (All Serial Numbers): MMT-1750/00763000375461, 00643169559745; MMT-1711/643169554931, 763000367039, 00643169890039, 00643169554917, 00643169629813, 00643169554924, 00643169554931, 00643169742062, 00763000013066, 00643169554948, 00643169554955, 00643169621954; MMT-1712/643169662612, 643169577701, 00763000205409, 00643169890046, 00643169577664, 00643169629820, 00643169577671, 00643169577688, 00763000155346, 00643169577695, 00643169577701, 00643169621961; MMT-1751/643169672239, 00643169574410, 00643169574427, 00643169574434, 00763000253288, 00643169574441, 00643169574458, 00643169643512, 00643169521155, 00643169521704, 00763000015596, 00643169520882, 00643169521421, 00643169521971; MMT-1752/763000192181, 00643169596368, 00643169596382, 00643169596405, 00763000318291, 00763000318307, 00643169596443, 00643169522305, 00643169521292, 00643169522381, 00643169521841, 00643169522268, 00643169521025, 00643169522343, 00643169521575, 00643169522114"
    }
  ]
}