  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Newark",
      "state": "DE",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88865",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "500 Gbc Dr",
      "address_2": "N/A",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.",
      "recall_number": "Z-0193-2022",
      "product_description": "Aptio Automation Interface Module to ADVIA Centaur XP/XPT. Aptio by Siemens Centaur Interface Firmware (FW) CENa_3-1-0-1 and lower. SMN 10703038",
      "product_quantity": "1142",
      "reason_for_recall": "The Aptio Automation Firmware for the ADVIA Centaur XP/XPT Interface Module may lead, in some cases, to an incorrect association of test results to sample ID. Incorrect association of test results could impact clinical interpretations and clinical decisions.",
      "recall_initiation_date": "20211006",
      "center_classification_date": "20211102",
      "report_date": "20211110",
      "code_info": "UDI 00630414596754"
    }
  ]
}