{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Twinsburg",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88840",
      "recalling_firm": "FujiFilm Healthcare Americas Corporation",
      "address_1": "1959 Summit Commerce Park",
      "address_2": "N/A",
      "postal_code": "44087-2371",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-0191-2022",
      "product_description": "Lisendo 880 Ultrasound Software Version: V1.0 through V4.1.3",
      "product_quantity": "N/A",
      "reason_for_recall": "Diagnostic ultrasound system with the specified software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display the measurement results correctly with two cardiology measurement functions: Mitral Regurgitation Flow (MR Flow) Measurement; Mitral Valve (MV) Measurement. In addition there is an error in the instruction manual.",
      "recall_initiation_date": "20211014",
      "center_classification_date": "20211102",
      "report_date": "20211110",
      "code_info": "Serial Numbers/UDI:  20514657  (01)04562122958868(11)161220(21)20514657;     20659147  (01)04562122958868(11)20181124(21)20659147;    20659148  (01)04562122958868(11)20181124(21)20659148;     20659149  (01)04562122958868(11)20181124(21)20659149;     20659150  (01)04562122958868(11)20181124(21)20659150;     20659153  (01)04562122958868(11)20181126(21)20659153;     20659154  (01)04562122958868(11)20181126(21)20659154;     20659155  (01)04562122958868(11)20181126(21)20659155;     20659156  (01)04562122958868(11)20181126(21)20659156;    205G0225  (01)04562122958868(11)20170418(21)205G0225;     205G0226  (01)04562122958868(11)20170424(21)205G0226;     205P2146  (01)04562122958868(11)20170615(21)205P2146;     205P2147  (01)04562122958868(11)20170615(21)205P2147;     205P2148  (01)04562122958868(11)20170615(21)205P2148;     205P2149  (01)04562122958868(11)20170615(21)205P2149;     205Q0809  (01)04562122958868(11)20170707(21)205Q0809;     205Q0810  (01)04562122958868(11)20170707(21)205Q0810;     205Q0812  (01)04562122958868(11)20170707(21)205Q0812;     205Q0815  (01)04562122958868(11)20171006(21)205Q0815;     205T7139  (01)04562122958868(11)20171006(21)205T7139;     205T7143  (01)04562122958868(11)20171026(21)205T7143;     205V5544  (01)04562122958868(11)20171026(21)205V5544;    205V5545  (01)04562122958868(11)20171030(21)205V5545;     205V5547  (01)04562122958868(11)20171031(21)205V5547;     205V5548  (01)04562122958868(11)20171031(21)205V5548;    G3005239  (01)04562122958868(11)20190308(21)G3005239;     G3005240  (01)04562122958868(11)20190308(21)G3005240;     G3025041  (01)04562122958868(11)20190603(21)G3025041;     G3025042  (01)04562122958868(11)20190604(21)G3025042;     G3025044  (01)04562122958868(11)20190605(21)G3025044;     G3025045  (01)04562122958868(11)20190605(21)G3025045;    G3025048  (01)04562122958868(11)20190606(21)G3025048;     G3025049  (01)04562122958868(11)20190606(21)G3025049;     G3037183  (01)04562122958868(11)20190624(21)G3037183;     G3037184  (01)04562122958868(11)20190624(21)G3037184;     G3037192  (01)04562122958868(11)20190709(21)G3037192;     G3043580  (01)04562122958868(11)20190820(21)G3043580;    G3043581  (01)04562122958868(11)20190820(21)G3043581;     G3043582  (01)04562122958868(11)20190820(21)G3043582;     G3052217  (01)04562122958868(11)20191016(21)G3052217;     G3052218  (01)04562122958868(11)20191016(21)G3052218;     G3052225  (01)04562122958868(11)20191202(21)G3052225;     G3052234  (01)04562122958868(11)20200128(21)G3052234;     G3052236  (01)04562122958868(11)20200109(21)G3052236;     G3054831  (01)04562122958868(11)20200115(21)G3054831;    G3054833  (01)04562122958868(11)20191212(21)G3054833;     G3054840  (01)04562122958868(11)20200115(21)G3054840;     G3080616  (01)04562122958868(11)20201215(21)G3080616;     G3080617  (01)04562122958868(11)20201215(21)G3080617;    G3106256  (01)04562122958868(11)20210413(21)G3106256;     G3106258  (01)04562122958868(11)20210413(21)G3106258;     G3111572  (01)04562122958868(11)20210413(21)G3111572;     G3111573  (01)04562122958868(11)20210421(21)G3111573"
    }
  ]
}