{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Marburg",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77515",
      "recalling_firm": "CSL Behring GmbH",
      "address_1": "Emil-von-Behring-Str. 76",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide including the countries of United States, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia-Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, Cyprus, Denmark, Egypt, Finland, France, Germany, Greece, Great Britain, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, japan, Kazakhstan, Kuwait, Latvia, Lebanon, Luxembourg, Macedonia, Mexico, Morocco, Netherlands, New Zealand, Norway, Poland, Portugal, Republic of Korea (S), Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, South Africa, Switzerland, Syria, Taiwan, Thailand, Trinidad-Tobago, Turkey, Turkmenistan, United Arab Emirates, Uruguay, Vatican, Venezuela,",
      "recall_number": "Z-0191-2018",
      "product_description": "SIEMENS N/T Protein Control SL for use on the BN II, BN ProSpec Systems: N/T Protein Control SL/L: REF/Product Code OQIN19, SMN 10446079 (Low); N/T Protein Control SL/M, REF/Product Code OQIO19, SMN 10446085 (Mid);  N/T Protein Control SL/H, REF/Product Code OQIP19, SMN 10446089 (High);",
      "product_quantity": "N/A",
      "reason_for_recall": "Siemens Healthcare Diagnostics has confirmed a negative bias of approximately -20% for results obtained with N Antiserum to Human Ceruloplasmin when calibrating Ceruloplasmin with various lots of N Protein Standard SL.  N/T Protein Controls SL for Ceruloplasmin are similarly biased. --- The affected standards and controls for the affected lots were incorrectly value assigned, which led to a lower result compared to correctly assigned standards.  Note: N Antiserum to Human Ceruloplasmin (REF OUIE) works as intended. --- Siemens investigation indicated that N Protein Standard SL and corresponding N/T Protein Control SL (Levels Low, Mid and High) are inaccurately assigned for Ceruloplasmin.",
      "recall_initiation_date": "20170605",
      "center_classification_date": "20171130",
      "termination_date": "20200407",
      "report_date": "20171206",
      "code_info": "SMN 10446079: Lot (UDI): 084647A (00842768016127084647A20170913),  084647C (00842768016127084647C20170913) exp. date  2017-09-13; 084648(0084276801612708464820180430),   084648C  (00842768016127084648C20180430),   084648H (00842768016127084648H20180430), exp. date 2018-04-30. ---  SMN 10446085: Lot (UDI): 084749D (00842768016134084749D20170917)  exp. date 2017-09-17; 084750 (03 0084276801613408475020180203),    084750B (00842768016134084750B20180203), 084750E (00842768016134084750E20180203) exp. date 2018-02-03. ---  SMN 10446089: Lot (UDI) 084847A (00842768016110084847A20170613),   084847B (00842768016110084847B20170613) exp. date 2017-06-13; 084848A (00842768016110084848A20171217),   084848D (00842768016110084848D20171217),   084848E (00842768016110084848E20171217),   084848H (00842768016110084848H20171217),   084848L (00842768016110084848L20171217) exp. date 2017-12-17;   084849 (0084276801611008484920180704),   084849C (00842768016110084849C20180704),  084849D (00842768016110084849D20180704)  084849E (00842768016110084849E20180704) exp. date 2018-07-04."
    }
  ]
}