  "results": [
    {
      "status": "Ongoing",
      "city": "Bronshoj",
      "state": "",
      "country": "Denmark",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99772",
      "recalling_firm": "Radiometer Medical ApS",
      "address_1": "Akandevej 21",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution including in the states of CA, OH, LA, MA, FL, NY, CO, IL, TX, GA, NC, PA, SC, MI, OK, WI, AR, MN, OR, UT, WA, SD, IN, TN, AZ, NE, WY, KY, AL, AK, ID, NJ, DE, MT, CT, MS, MO, VT, WV, IA, MD, KS, DC, NV, VA, NM, NH, ND, ME, HI, RI.",
      "recall_number": "Z-0190-2027",
      "product_description": "The ABL90 FLEX and ABL90 FLEX PLUS analyzers are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting. These tests are only performed under a physician's order.",
      "product_quantity": "5395",
      "reason_for_recall": "In vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer suitable for testing FHbF and ctBil, 2) Reportable ranges for some parameters, 3) Analyzer configuration, and 4) limitations and warnings.",
      "recall_initiation_date": "20260818",
      "center_classification_date": "20261015",
      "report_date": "20261021",
      "code_info": "UDI-DI: 05700693930923, 05700693930909 Software versions 3.6 and lower. ABL90 FLEX PLUS Instructions for Use versions prior to: 996-223, Version: 202606O.   ABL90 FLEX Instructions for Use versions prior to: 996-189, Version: 202606L",
      "more_code_info": ""
    }
  ]
}