  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Waukesha",
      "address_1": "3000 N Grandview Blvd",
      "reason_for_recall": "GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems.  This check is being added to prevent a portion of the system falling onto the patient during a scan due to fasteners being loose that secure the camera to the gantry.",
      "address_2": "",
      "product_quantity": "147",
      "code_info": "22002 619474D630  22004 618443DISC1  22007 814623D630  22010 251639D630  22012 704355D630  22013 956523D630  22017 845454D630  22018 618842DISC1  22024 601354D630  22025 773282D630  22026 813558D630  22027 941LWRNM630  22029 PRINCE630  22030 281540D630  22032 417781D630  22033 601288D630  22034 913345D630  22036 318798D630  22039 614NCH630  22040 904321DSC1  22041 317887D630  22042 318448D630  22043 JHDSV630NM  22045 217544PH630  22046 301877D630  22047 601798D630  22048 716372NM630  22049 580531NM630  22050 651DSC630  22052 205333D630  22053 618942NM630  22054 813615D630  22055 719587D630  22056 903GSD630  22057 417533D630  22058 708684D630  22059 937SNHC630  22063 707961D630  22064 918502DIS1  22065 804675D630A  22066 469814NM630  22067 978354DSV630  22068 585922NM630  22069 305596D630A  22070 956523D630  22072 POIN630  22077 816234D630  22081 808485NM630  22082 989227D630  22085 305596D630B  22087 801714D630  22088 815564D630  22091 702388VNM2  22092 760778D630  22094 269945D630  22095 843347D630  22096 201418D630  22097 319364D630B  22098 207664NM630  22099 334287NM630  22101 609978D630  22103 843797TD630  22105 319364D630A  22112 OSC630  22117 330386D630  22119 909599D630  22122 904276DSC1  22123 270417D630  22124 901226D630  22125 208455NM630  22126 409212NM630  22127 770751D630  22131 DMP630A  22132 936NMD630  22134 856355D630  22135 DMP630B  22136 605322D630  22137 775STMANM630A  22138 951303D630  22139 850862D630  22141 765447D630  22143 203384D630  22144 WH231NM630  22145 603354NM630  22148 302430NUC630  22149 229259D630  22150 910FHHOKENM630  22151 479967DSC  22156 607763DIS630  22163 262798D630  00107BH3 DZ1060NM01  22090 NM22090  22003 030359NU07  22038 030029NU05  22080 030151NU15  22073 NM22073  22074 NM22074  22118 NM22118  22120 NM22120  22130 NM22130  22051 905684D630  22114 514843AD630  22102 NM22102  22116 NM22116  22129 NM22129  22161 NM22161  00000000083BH6 EG1040NM01  22014 M4216453  22016 M4476428  22079 V6414107  22084 M45052111  22121 M2148252  22001 TP0004NU01  22011 PMAB01NU04  22023 PC3003NU01  22078 407431NU01  22089 HC0863NU02  22108 870939NU02  22109 870939NU01  22115 600791NU01  22140 HG4437NU04  22147 HC0038NU01  22107 H3001NUC01  22005 083026001113611  22133 083026200011213  22021 IE1044NM01  22022 A5384208  22031 A5812287  22035 A5131865  22086 A5366307  22044 NT1008  22071 NT3010  22104 NT3002  22110 NT3003  22111 NT3020  CZC1321LZ6 NT3001  22037 0850260134  22093 0850260125  22009 NO1035MN06  22083 RU1136NM01  22106 ZA2029NM01  22075 H002NU03  22113 O009NU09  22128 34363NUC01  22015 00001NUC11  22028 00052NUC11  22100 00271NUC10",
      "center_classification_date": "20141107",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) including Guam, Puerto Rico and DC; and countries of: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BELRARUS, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, HUNGARY, GREECE, GUADELOUPE, HONG KONG, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAPAN, REPUBLIC OF KOREA, KUWAIT, LATVIA, LEBANON, LITHUANIA, MALAYSIA, MALTA,MARTINIQUE, MEXICO, MOROCCO, NAMIBIA, NETHERLANDS, NEW ZEALAND, NORWAY, PAKISTAN, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN,SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UNITED KINGDOM, UKRAINE, URUGUAY, VENEZUELA, VIET NAAM.",
      "state": "WI",
      "product_description": "Discovery NM 630, dual detector free-geometry integrated nuclear imaging system.  Models H3100RC, H3100RD, H2401LW.       Product Usage: The GE Discovery NM 630 system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and in assessment of organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM 630 system is a Nuclear Medicine (NM) system, which is intended to yield General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The scanning modes include planar mode (Static, Multi-gated, Dynamic and Whole body scanning) and tomographic mode (SPECT, Gated SPECT, Whole body SPECT). The acquisition types include single and multi-isotope/ multi peak frame/list mode single-photon imaging. The imaging-enhancement features include assortment of collimators, gating by physiological signals, and real-time automatic body contouring. The GE Discovery NM 630 system may include signal analysis and display equipment, patient and equipment supports, components and accessories. The system may be used for patients of all ages.",
      "report_date": "20141119",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "GE Healthcare, LLC",
      "recall_number": "Z-0187-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "69570",
      "termination_date": "20150227",
      "more_code_info": "",
      "recall_initiation_date": "20141014",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}