  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Westlake Village",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78350",
      "recalling_firm": "Implant Direct Sybron Manufacturing, LLC",
      "address_1": "3050 E Hillcrest Dr",
      "address_2": "N/A",
      "postal_code": "91362-3171",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution: US (nationwide) including Puerto Rico and country of Japan.",
      "recall_number": "Z-0185-2018",
      "product_description": "ImplantDirect  Legacy(TM) 3 Implant, 13MM L, REF 864713    This Legacy implant is a two-piece implant for one-stage or two-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework.",
      "product_quantity": "75 units",
      "reason_for_recall": "One reported complaint and discovered that the outer vial was out of round and undersized.  Therefore a portion of the opening may not allow inner vial and components to slide freely out of the vial.  As a result, the cap could possibly not be screwed on properly and the sterilization compromised.",
      "recall_initiation_date": "20171010",
      "center_classification_date": "20171128",
      "termination_date": "20180927",
      "report_date": "20171206",
      "code_info": "Lot Number:  79607"
    }
  ]
}