  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-21",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99888",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of AZ, CA, IN, MD, MO, NJ, OH, and WI.",
      "recall_number": "Z-0183-2027",
      "product_description": "Convenience Kits containing CONMED AirSeal Access Ports    CHN GYN ONCOLOGY\t  \tDYNJ906731K          DYNJ906731L  \tDYNJ906731N  \t  DAVINCI PROSTATECTOMY\t  \tDYNJG901002A  \t  \t  \tDYNJG901002B          DYNJG901002C  \tDYNJG901002D    GENERAL DAVINCI\t  \tDYNJG901004B  \tDYNJG901004D    HERNANDEZ DAVINCI ACC BUNDLE M\t  \tDYNJ911258A  \t  KIT ROBOTICS UROLOGY PROSTATE\t  \tDYKMBNDL116D  \tDYKMBNDL116F  \t  ROBOTIC GYN\t  \tDYNJ907827F  \t  ROBOTIC PACK\t  \tDYNJ50474G    RR-DAVINCI PROSTATE ACCESSORY\t  \tDYNJ63392C          DYNJ63392F",
      "product_quantity": "3,551 kits",
      "reason_for_recall": "Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.",
      "recall_initiation_date": "20260831",
      "center_classification_date": "20261015",
      "report_date": "20261021",
      "code_info": "DYNJ906731K    UDI-DI 10195327436346    Lots 24ADB566   24CDB291   24CDB969   24GDB449   24JDA590   24KDB303   25BDB494   25CDA446   25EDB457   25FDA837   25FDB582   25GDA082  DYNJ906731L    UDI-DI 10198459403675    Lots 25HDB211   25JDA226  DYNJ906731N    UDI-DI 10198459758249    Lots 26EDA043   26FDB001   26FDB487    DYNJG901002A    UDI-DI 10195327401481    Lot 24BLA300  DYNJG901002B    UDI-DI 10195327677145    Lots 24GLA290   24ILA519   24JLA479   24LLA939   25ALA154   25ELA210   25FLA112   25GLA399   25HLA560  DYNJG901002C    UDI-DI 10198459545672    Lots 25JLA619   25KLA897   26ALA207   26BLA625   26CLA984  DYNJG901002D    UDI-DI 10198459723742    Lots 26ELA161   26ELA744   26GLA765   26HLA179    DYNJG901004B    UDI-DI 10195327677138    Lots 24FLA735   24GLA836   24KLA502   25ALA409   25ALA837   25DLA322   25ELA451   25FLA291   25GLB051   25HLA207   25ILA840   25JLA027   25JLA621  DYNJG901004D    UDI-DI 10198459723759    Lots 26DLA382   26ELA743   26FLB123    DYNJ911258A    UDI-DI 10198459582097    Lot 26AMF842    DYKMBNDL116D    UDI-DI 10195327608866    Lots 24CBF889   24DBA349  DYKMBNDL116F    UDI-DI 10198459000706    Lots 24DBS388   24FBC411   24GBK948   24IBV184   24JBB328    DYNJ907827F    UDI-DI 10195327345365    Lots 24ABV361   24CBM651   24DBD866   24GBH678    DYNJ50474G    UDI-DI 10198459140235    Lots 24KDA630   25ADB023   25BDB826   25CDB231   25DDA066   25HDA254   25JDA520   25LDA008   25LDC029   26DDA900   26FDA459   26FDB573    DYNJ63392C    UDI-DI 10195327634643    Lots 24BLA318   24CLA990   24ELA575   24FLA977   24HLA137   24ILA935  DYNJ63392F    UDI-DI 10198459315572    Lots 25DLA944   25ELA626   25FLA272   25GLA905   25HLA240   25HLA894   25JLA091   25KLA775   25LLA985   26ALA670   26DLB121   26FLA701",
      "more_code_info": ""
    }
  ]
}