      "status": "Ongoing",
      "city": "Atlanta",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88860",
      "recalling_firm": "Elekta, Inc.",
      "address_1": "400 Perimeter Center Ter NE Ste 50",
      "address_2": "N/A",
      "postal_code": "30346-1227",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution throughout USA  OUS distribution including:Algeria, Argentina, Australia, Austria Bahamas, Bangladesh, Belarus, Belgium, Bolivia, Botswana, Brazil Brunei, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran (Islamic Republic of), Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Libyan Arab Jamahiriya, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Peru, Philippines, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Saudi Arabia, Senegal, Singapore, Slovakia, South Africa, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam",
      "recall_number": "Z-0182-2022",
      "product_description": "Oncentra Brachy users using version 4.0 and above with Applicator Modeling license 170283-00",
      "product_quantity": "1072 devices",
      "reason_for_recall": "Under certain circumstances, there can be a difference between the planned and delivered dose of the two ovoid channels.",
      "recall_initiation_date": "20211013",
      "center_classification_date": "20211028",
      "report_date": "20211103",
      "code_info": "Oncentra Brachy users using version 4.0 and above with Applicator Modeling license 170283-00  GTIN Numbers: 08717213052758, 08717213038660, 08717213020610, 08717213052321, 08717213052314, 08717213052307, 08717213052291, 08717213052246, 08717213052239, 08717213051881, 08717213051843, 08717213051782, 08717213051775, 08717213051294, 08717213020610, 08717213053717, 08717213053700"
    }
  ]
}