{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Austin",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80930",
      "recalling_firm": "OriGen Biomedical, Inc.",
      "address_1": "7000 Burleson Rd Bldg D",
      "address_2": "N/A",
      "postal_code": "78744-3202",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: U.S.(nationwide) to states of: IN, NY, CA, MA, TX, OH, KY, MI, MS, NM, NC, VA, WI, TN, CT, FL, AR, MO, GA, LA, DC, OR, MN, SC, OK, AZ, UT, RI, SD, WA, PA, AL, MD, IL, HI; and countries (OUS) of:  Argentina, Austria, Belgium, CANADA, Colombia, EU, Germany, India, Italy, Japan, Kingdom of Saudi Arabia, Netherlands, New Zealand, Poland, Qatar, Romania, South Africa, SWEDEN, Thailand, UK, United Arab Emirates, and VIETNAM.",
      "recall_number": "Z-0181-2019",
      "product_description": "Reinforced Dual Lumen Cannula, 19Fr., REF: VV19F, Sterile EO     Indications: The OriGen Reinforced Dual Lumen Cannula is indicated for use as a single cannula for both venous drainage and re-infusion of blood in the internal jugular vein during extracorporeal life support procedures.",
      "product_quantity": "435",
      "reason_for_recall": "These catheters may have the following deficiencies: 1) barium sulfate omission, and 2) actual catheter tip length is one centimeter shorter than printed on the tip",
      "recall_initiation_date": "20180821",
      "center_classification_date": "20181018",
      "termination_date": "20210715",
      "report_date": "20181024",
      "code_info": "UDI: 00816203022097, Lot: N18394, N18447, N18447-1, N18447-1A, P18909, P18971"
    }
  ]
}