{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Andover",
      "address_1": "3000 Minuteman Rd",
      "reason_for_recall": "Faulty Automatic Motion Controller (AMC), a problem in the Power On Self Test (POST) error handling was detected, can result in a hazardous movement of the C-arc. system.",
      "address_2": "",
      "product_quantity": "22 systems",
      "code_info": "Allura Xper Software version: R8.2.0",
      "center_classification_date": "20141107",
      "distribution_pattern": "Distributed in the states of IL, and TN, and the countries of  Argentina , Australia, Belgium, Denmark, Egypt, France, Germany, Hong Kong, Italy, Japan, Korea, Republic of Netherlands, Norway, Saudi Arabia, Switzerland, United Kingdom, and Viet Nam.",
      "state": "MA",
      "product_description": "Philips Medical Systems Allura Xper FD20C Radiological Imaging with software version R8.2.O; System Code: 722028.    The Allura Xper FD series are intended for use on human patients to perform:  Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures.",
      "report_date": "20141119",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Medical Systems, Inc.",
      "recall_number": "Z-0181-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68723",
      "termination_date": "20170118",
      "more_code_info": "",
      "recall_initiation_date": "20140708",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}