  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Le Noirmont",
      "state": "N/A",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86419",
      "recalling_firm": "Bien-Air Surgery Sa",
      "address_1": "rue de l' ouest 2b",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution including in the states of CA, MI and MN.",
      "recall_number": "Z-0177-2021",
      "product_description": "Bien-Air Surgery REF 1600517-001 SN ******* MULTIFUNCTION PEDAL OSSEODUO - Product Usage: Product Usage: intended to convert electrical energy to mechanically drive attachment, microsaws, shaver blades and burs, burs and perforator tools for cutting and shaping bone including spine and cranium.",
      "product_quantity": "Total US = 1120 foot pedals",
      "reason_for_recall": "The magnet located inside the foot pedal may come off and may result in unattended start of unstoppable motor.",
      "recall_initiation_date": "20200911",
      "center_classification_date": "20201021",
      "report_date": "20201028",
      "code_info": "Model 1600517 / Catalog number 1600517-001  Serial Numbers:  18F0017  18F0030  18G0003  18G0018  18G0019  18G0020  18G0021  18G0025  18H0016  18H0017  18H0018  18H0019  18H0026  18H0027  18I0003  18I0004  18I0005  18I0006  18I0018  18I0022  18I0023  18I0024  18I0025  18I0026  18I0027  18I0028  18I0029  18I0030  18J0001  18J0002  18J0003  18J0004  18J0005  18J0009  18J0010  18J0011  18J0012  18J0013  18J0014  18J0015  18J0016  18J0017  18J0018  18J0019  18J0020  18J0021  18J0022  18J0023  18J0024  18J0025  18J0026  18J0037  18J0038  18J0039  18J0040  18J0041  18L0012  18L0014  18L0017  18L0018  18L0019  18L0020  18L0021  18L0023  19B0021  19B0022  19B0023  19B0024  19B0025  19B0028  19B0029  19B0030  19B0038  19B0041  19B0042  19B0043  19B0044  19B0051  19B0052  19B0053  19B0054  19B0055  19B0056  19B0058  19B0059  19E0004  19E0005  19E0008  19E0009  19E0011  19E0012  19E0013  19E0014  19E0015  19E0016  19E0017  19E0028  19E0029  19E0030  19E0035  19E0037  19E0047  19E0048  19E0053  19E0054  19E0055  19E0056  19E0057  19E0058  19E0060  19E0063  19E0064  19E0068  19E0069  16A0015  16A0016  16G0024  16I0007  16I0026  16I0027  16I0028  16I0029  16K0027  17A0009  17B0008  17B0009  17B0010  17B0011  17B0012  17B0013  17B0039  17B0040  17B0041  17B0049  17B0050  17C0012  17C0013  17E0033  17G0004  17G0024  17G0025  19E0006  19E0007  19E0010"
    }
  ]
}