{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Le Noirmont",
      "state": "N/A",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86419",
      "recalling_firm": "Bien-Air Surgery Sa",
      "address_1": "rue de l' ouest 2b",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution including in the states of CA, MI and MN.",
      "recall_number": "Z-0176-2021",
      "product_description": "Bien-Air Surgery REF 1600407-001 SN **** FOOTCTRL OSSEODOC - Product Usage: intended to convert electrical energy to mechanically drive handpieces and microsaws used to cut and shape bones.",
      "product_quantity": "Total US = 1120 foot pedals",
      "reason_for_recall": "The magnet located inside the foot pedal may come off and may result in unattended start of unstoppable motor.",
      "recall_initiation_date": "20200911",
      "center_classification_date": "20201021",
      "report_date": "20201028",
      "code_info": "Model 1600407 / Catalog number 1600407-001  Serial Numbers:  16C0017  16F0013  16I0009  16K0017  17F0039  17I0015  17L0013"
    }
  ]
}