      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69437",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the countries of   Algeria, Argentina, Australia, Austria, Bangladesh, Belarus, Bosnia-Herzegovina, Botswana, Brunei Darussalam, Bulgaria, Canada, China, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, El Salvador, Ethiopia, France, Germany, Gibraltar, Greece, Guam, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Korea, Lebanon, Libyan Arab Jamahiriya, Malaysia, Mexico, Morocco, Namibia, Pakistan, Palestinian Territory, Panama, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Kingdom, United States including Puerto Rico, Venezuela, and Vietnam.",
      "recall_number": "Z-0176-2015",
      "product_description": "UniCel DxC 600i SYNCHRON Access Clinical Analyzer, Catalog No. A25640, Instructions for Use Part No. A93659AB.    Product Usage:  The DxI 600i systems are in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analytes concentrations found in human body fluids.",
      "product_quantity": "1,519 units total (403 units total)",
      "reason_for_recall": "Beckman Coulter is recalling the Small Form Factor Console PC (SFF PC) (B23083) for use with the DxC 600i analyzer because it may experience a \"MFC Exception\" error during normal operation of the analyzer system.",
      "recall_initiation_date": "20141006",
      "center_classification_date": "20141105",
      "termination_date": "20150629",
      "report_date": "20141112",
      "code_info": "Serial No.    504489  502713  503114  503824  504400  504373  501946  507570; 503285, 800337, 700292  505091  509407  509349  509793  509801  509764  509768  509671  509628  509670  509690, 509705  509724  509800  509771  509744  509746, 509735  509668  509799  509626  509797  509669  509765  509790  509798  509739  509736  509769  509287  504104  509063  509256  509511  503600  505722  502105  504597  503503  504557  505476  504095  504060  503183  505154  506131  509267  509524  500646  509532  501760  509369  509514  509177  800135  509313  504619  800074  504591, 508910, 508912, 508907  509373, 509304  506566  508920  505811  508914, 508917  509625  502796  508933, 508934  509291  509634, 509635  509530, 509588  509787, 509788, 509783  509433, 509437, 509662, 509663, 509660, 509659, 509658  509715  509657  508916, 508566, 509344, 509526, 509424, 509461, 509459, 509466, 509062, 509197, 509200, 509195, 509196, 509201, 509199, 509605, 509604, 509571, 509585, 509567, 509569, 509568, 509566, 509579, 509570  508887, 509489, 509542, 509538, 509541, 509645, 509647, 509652, 509649, 509648, 509642, 509644  509341, 509346  508885, 508891, 509781, 509782, 509770  509026, 509020, 509025, 509027, 509028, 509021  509651, 509784  509194, 509190  509713, 509710  509175  509560, 509557, 509556, 509558  509553, 509554  509219, 509237, 509236, 509233, 509231, 509234, 509238, 509212, 509240, 509230, 509208, 509216, 509232, 509214  509535  509072, 508994, 508981, 508990  509531, 509417, 509398, 509759  509089  509050, 508991  508884, 508888, 508870, 509075, 509074, 509076, 509069, 509147, 509145, 509155  509442, 509443  509622  509107  508937, 509697, 509698  509821, 509810  509512, 509169, 509174, 509170, 509695  509610, 509620, 509613  509355, 509471, 509468, 509699, 509700, 509702, 509703  509517, 509520, 509664, 509665, 509666   509637  508954, 508996, 509203, 509210, 509205, 509209  509401, 509202, 509780, 509779   509423 ,509426, 509806, 509804, 509812, 509807, 509805  509561  508940  509564  509141  509142  509156  509507  509598  509023  509045  509090  509342  509217  509098  509104  509359  509415  509586  509539  509005  509002  509487  509345  508872  509404  509405  508961  509456  509061  508995  509220  509211  509052  509540  509469  509154  508992  509362  508976  508883  508875  509022  509448  509410  509436  508882  508971  509485  509483  509159  508998  508922  508874  509043  509017  509446  509227  509215  509397  509066  509091  509711  509106  509631  508957  509223  508986  509440  509431  509418  509229  508974  509445  509225  509425  509422  508879  508972  509048  509137  509133  509478  509482  509488  509082  509081  509619  509218  509000  509470  509108  509343  508876  508997  509457  509109  508894  509235  509084  509595  509008  509492  509044  509079  509412  509708  509709  509224  508869  509226  509064  509420  509013  508983  509411  509379  509399  508985  509712  509480  508953  509476  509228  509559  509083  509653  508987  509434  508864  509058  508902  508893  508975  509172  509655  508988  509105  509331, 509332  505348, 504964, 503973, 503803  509363  509347  509365  505980  509357  509484  509350  507659  509361  502352  504206  502973  507942  504121  501988  501981  504243  504320  508080, 507913, 508662  509704, 509707  509528  503193, 800313  501985  509167  504816  503459  509479  504132  509148  504423  504337  509018  504406  509533  504249  509335  504990  508909  503084  504415  508588  508589  502899  505130  509261  509271  508944  503700  508370  505951  509413  509036  509438  501685  509577  506229  505444  506246  800237  506188  504526  503954  507118  502219  509516  500172  509500  502246  700287  504019  800020  508968  505082  800085  501846  504717  501986, 502495  505178  503525  509326  501672  508555  506177, 506081  509059  500039  800022  501896, 500039  509789  505041  503212  501489  502545  501480  502836  509518  509029  504139  503467  501272  505281  501431  502309  509374, 503292, 503484  501135  504260  504536  506062  504611  508943  500061  504025  501953  505222  501372  506788  508694  509400  505144  503187  503496  505994  502412  502180  507584  504558  500403  506181  501610  503995  501746  504016  507063  500478  504757  503551  509101  504613  502019  502423  501348  509617  505898  508900  509116  503400  508895  508823  507795  501076  508945  505647  501375  501366  504167, 503888  500490  500533  505008  503921  501743  500436  508941  508928  503585  509375  501085  504108  504181  700119  500787  500117  500775  503792  505866  504923  506009  502266  500901  501770  505220  504665  503681  502918  501156  500579  505477  500300  503469  504035  504590  504270  503166  504384  504190  509416  508821  505465  503463  509033  505289  501739  508778  506813  503095  504670  502909  500589  501680  508599  502130  500357  502566  501095  509388, 509115  504803  508786  509386  500648  505294  509204, 509207, 509575, 509786, 509785  508862  501227  500836  504648  506101  501414, 501698, 502626  505107  800171  800553  506987  509314  509352  800220  508905  504885  509130  509274  509371. 509370  509103  509049  508970  504357  507755  509070, 509067  509047, 509046  509093, 509094  800387  506626  509686  500531  502548  505118  501916  509552  505110  502776  504203  503038  502706  506368  505519  502399  502962  502591  503521  502311  505494  509543  502658  502473  504006  508818  502670  500629  503180  502880  508939  503696  502976  503659  503176  502304  505419  505549  505287  501983  504503  500539  505941  502206  500494  508863  505004  504504  509795  508938  509279  505814  503814  509315  509316  507864  509092  508813  509603  508903, 508901  509288  509006  508849  502969  507741  508808  509596  509444  508602  509121  508899  506034, 509706  503333  501991  700075  501117  505061, 508807, 507733, 507476, 509041  501314  508568, 508134  500686  508140  503235  502410  504052  502239  508805, 507728, 506416, 509486  508963  509298, 509430, 509428  504430, 503270, 502846  501367, 501034, 500914  509073  501286  509584  509656, 509761  504781  506674, 505930, 507401, 506091, 505972  503371  501900  403732  500402  505889, 505912, 505340  506243  506012  502653  505071  506039  502108  506032  505284  505253  502261  502608  504153  504169  503867  503208  503107  503331, 502834  505229  503252  502177  502109, 502643, 502501  502117  500620  503495  508955  504832  504225  505573  504055  505996  504450  504562  504096, 504798, 503680, 503770, 503253  500603  509302, 509300, 509301, 509037, 509040, 509038, 509039, 509035, 509125, 509126, 509127, 509716  509153, 509701, 509714  508354, 509643, 509640, 509639, 509641, 700320  501479  506359  508861  504403  509338  505526  504246  505188  503708  800340  505335  503607  505928  503173  505503  800158  508401  508374  507319  800357  504359  506634  504870  504587, 506238  502464  501149  500954  508918  509012  509693  509136  509143  509024  509525  509527  509594  509323  509467  508775  507436, 506696, 509280 ,509281, 509439, 509435, 509589, 509688, 509755  509114  508915  508969  509158  509146  508921  508479  509161  509574  509010  508960, 508956, 508959, 508965, 508999, 509150, 509144  508926  509004  508932  508967  509334, 509333, 509149  508913  509152  508925  509330  508951  505784  508107  509309  508966  509592  509687  509685  509682, 509679  508935  508824  508919  509384  508896  509394  509380, 509385, 509391, 509389, 509387, 509390, 509393, 509408, 509395, 509751, 509752, 509750, 509749  509392  509011  509009  509278  509277  507223, 509689, 509747  509402  509447  508878  509014  509007  509003  509001  506364  508877  509100, 509722, 509721, 509728, 509730, 509727, 509726, 509723, 509725, 509729, 509811, 509813, 509818  508906  509099, 508886  508911  509097  506176, 506141, 506027  508355  504032  501232  502186  502205  503610  509473  502164  504352  509065, 509068  509521  700132  500879  509376, 503571  501294  502729  501295  500880  500659  502368  501440  502035  509419  509515  509451  500231  500507  502079  509504  505351  504116  500335  504114  509582  509460  509583  800292  502427  506597  509618  509732  509163  506595  509414, 509409, 509030, 509031, 509032, 509222, 509221, 509621, 509623  504128  509290  505970  501446  504394, 506720, 506686, 509295  509138  504835  504323, 508892, 509243, 509317, 509183, 509239, 509242, 509165  505547, 509244, 509248, 509307, 509308, 509306, 503599  501405  504412  508897  504305  504729  506833  504353  504379, 504256, 505768, 505116, 506699  508814  508889, 509245, 509696  505439  502047  500484  505010  508950  502305  507757  501643  505087  500855  500480  509529  509720  509753  500973  500840  508973  800373  500994  506231, 506160  506258, 506199  700290, 700236  507380, 502480  504815, 508228  505927, 502362  700201, 700177  507379, 506515  508124, 700270  502116  509019  509441  507693  502512  509111  509129  508936  503091  800036  509760, 509762  509718  509283, 800809  800372  509312, 509310  509602  506252  502940  509537  509717  509263  509358  509270  509453  506766  509672  509016  509694  509803  509311  800650  505704  509377  508931  505298, 509513  509481  509499  509549  509508, 509576  509421  509284, 509454  509547, 509551  507848  509320  509354  509151  509015  506093  509122  508800  508881  509120  509339  509296  509188  500530  509173  509676  509429  502289  509615  509123  509646  509654  509472  509124  508029  509452  508964  509766  800804  505215  509464  509677  509135  509673  509282  509383  509113  500143, 509490  508930  509600  800774  509054  700190  509088  509368  509285  509609  506477  509817  509275  509182  509734  509458  508923  507846  509794  509292  509053  509550  800178  501309  509522  509572  509340  506236  509455  509773  505932  503495  503400  509611  509611, 509606  509573  506417  509096  506586  509757  505124  509266  506211  506155  507050  509250  502537  506536  508946  800205  509327  509692  509189  509071, 509077  509179  509192 509198  509591, 509590  509181, 509180  502844  505574  502839  504213  509777, 509774, 509776, 509775, 509778  508430  502841  502853, 509809  509477  700271  509495  509562  508349  508898  509102  507024  509178  504550  503369  509616  505557  508993  509497  509259  509738  509767  509493  509505  509276  509674  509128  509254  509624  509772  509057  509247  509299  509348  509607  509328  508982  508929  509251  509260  509255  509272  509548  509273  508904  508984  509661  509055  800878  800001  509494  800697  508908  505308  509132  505939  508942  509563  507594  508871  509353  505892  509187  505643  509206  509650  504826  507522  505995  509360  509286  507156  509257  509509  507695  509249  509581, 509796  508978  800694  509110  508859  505170  503542  509095  509519  508962  503214  509680  502008  509745  509164  509325  509820  504635  505843  509819, 501974  509808  509731, 509733  509268  504905  509056  508977  504232  509119  500921  700306  506663  506318  509465  509367  509378  509737  509246, 509241  509289  800515, 505991  509265  506694  509112  509051  509253  509293  509501  508890  509632  509523  503138  509503  509496  509502  509593  509060  509319  509681  800877  509087  509085  509086  509678  509318  509802  509297  509336  506678  508958  509193, 509191  504461  800674  509382  508949  506166  509366  509463  509822  509396  504472  509599  509171, 509166  501439  503511  509491  509763  800394  509140  504490  509118  509719  509351  509691  509042  509630  509252, 509258  509184  508356  509474  509185  509601  505342  509329  509356  502631  509269  509636  509534  507774  506072  509667  500350  509372  509578  500580  509162  504061, 509741  507872  509176  509186  508948  509262  509381  509303  509792  509337  500360  509546, 509555  501984  800270  509264  509758  500425  505890, 800238  506050  502358  509321, 509324, 509322, 509475, 509545, 509544, 509427, 509080, 509078, 508980, 509213, 509139, 509131, 509134, 509627, 509633, 509816, 509815, 509814  508989"
    }
  ]
}