      "status": "Terminated",
      "city": "Concord",
      "state": "N/A",
      "country": "Canada",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88748",
      "recalling_firm": "AB SCIEX",
      "address_1": "71 Four Valley Dr",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of District of Columbia, Florida, Illinois, Louisiana, Maryland, Michigan, Missouri, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Tennessee, Utah, Wisconsin and the countries of Austria, Finland, Germany, Hong Kong, Italy, Japan, Norway, Spain, Switzerland, United Kingdom, Italy, France, and Japan.",
      "recall_number": "Z-0175-2022",
      "product_description": "Cliquid MD version 3.4 software used in conjunction with Analyst MD software on SCIEX Citrine & 4500MD series liquid chromatography-tandem mass spectrometry (LC-MS/MS) systems. Model No. 5088288",
      "product_quantity": "30",
      "reason_for_recall": "The values of the Internal Standard (IS) concentrations are incorrectly derived when the user builds a customized test, then set the IS column to hidden and left the column empty. This may lead to the user making incorrect conclusions using incorrect results.",
      "recall_initiation_date": "20211004",
      "center_classification_date": "20211026",
      "termination_date": "20240123",
      "report_date": "20211103",
      "code_info": "Software version 3.4"
    }
  ]
}