{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83813",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.      The countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BERMUDA, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, GUAM, GUATEMALA, HONDURAS, HONG KONG, HUNGARY, INDIA, Indonesia, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, Korea (Republic of), KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, MACEDONIA, Malaysia, MEXICO, MOROCCO, Myanmar, NETHERLANDS, New Zealand, NORWAY, PANAMA, PHILIPPINES, POLAND, PORTUGAL, PUERTO RICO, QATAR, ROMANIA, Russia, SAUDI ARABIA, Singapore, SLOVAKIA, SOUTH AFRICA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, TAJIKISTAN, Thailand, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, Viet Nam, YUGOSLAVIA.",
      "recall_number": "Z-0175-2020",
      "product_description": "ApexPro Telemetry Server System.  Also identified as Modification To: ApexPro Telemetry System and ApexPro FH Telemetry System - Product Usage: These systems are typically deployed in sub acute care areas in hospitals or clinical sites where patient mobility can enhance the effectiveness of the medical procedures administered.",
      "product_quantity": "22,438 (17,971 US; 4,467 OUS) devices total",
      "reason_for_recall": "May not provide visual and/or audible alarms at the CARESCAPE Central Station or Clinical Information Center monitor for ECG arrhythmias, ECG LEADS FAIL or Pulse Oximetry (SpO2) under certain conditions.",
      "recall_initiation_date": "20190830",
      "center_classification_date": "20191023",
      "report_date": "20191030",
      "code_info": "1. Product Code: 3F, No UDI Number, Serial Numbers: M13F1043G, B23F1531G, B23F1532G, B23F2216G, A23F1290G, B23F1324G, M13F1152G, M13F1153G, M13F1154G, M13F1155G, M13F1164G, M13F1165G, M13F1166G, M13F1168G, M13F1169G, L13F0687G, D23F1925G, D23F1968G, E23F2074G, E23F2075G, B23F1402G, M13F1059G, M13F1173G, M13F1109G, C23F1815G, A23F1260G, J13F0272G, C23F1607G, D23F1981G, M13F0968G, L13F0683G, L13F0576G, J13F0274G, B23F1302G, K13F0437G, C23F1793G, D23F1943G, M13F1094G, B23F1495G, K13F0481G, L13F0571G, B23F1304G, C23F1808G, D23F1899G, M13F1151G, J13F0098G, J13F0099G, J13F0100G, C23F1594G, M13F1099G, B23F1439G, J13F0229G, L13F0600G, K13F0499G, K13F0500G, K13F0501G, L13F0623G, L13F0624G, L13F0625G, B23F13886, B23F1388G, B23F1390G, D23F1851G, D23F1861G, J13F0369G, M13F1075G, M13F1084G, M13F0891G, M13F0892G, A23F1229G, E23F2036G, L13F0776G, A23F1258G, B23F1315G, B23F1496G, B23F1501G, B23F1502G, B23F1506G, B23F1508G, C23F1784G, F13F0006G, J13F0006G, L13F0864G, L13F0865G, L13F0866G, L13F0668G, J13F0291G, J13F0292G, J13F0300G, M13F0950G, M13F1150G, B23F1313G, M13F0995G, M13F1044G, C23F1699G, J13F0293G, J13F0294G, J13F0295G, J13F0374G, J13F0375G, J13F0378G, J13F0379G, L13F0721G, L13F0729G, L13F0767G, L13F0768G, L13F0769G, M13F1062G, M13F1063G, E23F2068G, M13F1069G, D23F1852G, K13F0415G, K13F0426G, B23F1454G, D23F1971G, D23F1987G, K13F0411G, K13F0387G, L13F0755G, M13F1141G, M13F1142G, J13F0333G, J13F0203G, D23F1936G, M13F0991G, B23F1452G, B23F1509G, D23F1980G, D23F1982G, E23F2064G, J13F0308G, M13F1032G, M13F1034G, M13F1046G, M13F1047G, L13F0669G, A23F1266G, A23F1267G, A23F1297G, C23F1589G, B23F1357G, B23F1358G, M13F1074G, D23F1859G, D23F1933G, F23F2081G, F23F2090G, F23F2091G, F23F2092G, F23F2111G, A23F1263G, B23F1362G, C23F1690G, D23F1877G, D23F1888G, D23F1997G, D23F2014G, J13F0250G, J13F0252G, J13F0255G, K13F0462G, K13F0464G, K13F0518G, K13F0533G, L13F0691G, L13F0692G, L13F0741G, M13F0917G, M13F0927G, M13F0939G, M13F0972G, M13F0973G, M13F0993G, M13F0994G, M13F1033G, M13F1045G, M13F1048G, M13F1051G, M13F1070G, M13F1106G, M13F1107G, M13F1108G, M13F1114G, M13F1146G, D23F1867G, L13F0631G, C23F1809G, D23F1835G, L13F0607G, C23F1777G, J13F0222G, J13F0219G, C23F1800G, C23F1801G, C23F1802G, J13F0072G, B23F1346G, B23F1347G, B23F1348G, D23F1853G, M13F1073G, D23F1854G, D23F1855G, D23F1856G, B23F1516G, B23F1526G, B23F1527G, C23F1649G, C23F1779G, K13F0566G, J13F0245G, M13F0940G, D23F1958G, E23F2079G, E23F2072G, K13F0431G, K13F0432G, K13F0435G, K13F0563G, K13F0564G, K13F0565G, A23F1250G, A23F1257G, B23F1316G, J13F0313G, J13F0314G, J13F0315G, A23F1246G, A23F1249G, C23F1728G, C23F1757G, D23F1994G, M13F1040G, L13F0616G, L13F0861G, J13F0273G, A23F1282G, K13F0475G, C23F1767G, C23F1769G, C23F1770G, C23F1772G, D23F1846G, J13F0367G, J13F0368G, D23F1946G, B23F1510G, B23F1512G, B23F1513G, B23F1517G, B23F1521G, B23F1529G, B23F1530G, B23F1534G, B23F1539G, B23F1544G, C23F1725G, C23F1726G, K13F0524G, L13F0724G, L13F0730G, A23F1196G, A23F1198G, B23F1473G, L13F0809G, L13F0810G, L13F0811G, L13F0813G, L13F0814G, C23F1817G, C23F1818G, D23F1839G, M13F0949G, J13F0048G, J13F0189G, J13F0280G, J13F0281G, J13F0348G, M13F1049G, M13F1050G, C23F1723G, D23F1967G, K13F0402G, K13F0406G, K13F0407G, M13F0969G, M13F0975G, M13F0941G, M13F0948G, J13F0324G, B23F1518G, B23F1538G, M13F1052G, M13F1053G, M13F1066G, M13F1067G, K13F0506G, E23F2030G, B23F1423G, B23F1498G, C23F1750G, C23F1756G, D23F1864G, D23F1879G, D23F1960G, D23F2614G, M13F1130G, M13F1131G, D23F2001G, E23F2069G, M13F0976G, M13F1054G, M13F1136G, M13F1137G, M13F1148G, M13F1149G, A23F1286G, M13F1030G, M13F1031G, B23F1309G, J13F0323G, J13F0271G, L13F0614G, L13F0615G, L13F0663G, M13F0945G, M13F0946G, A23F1294G, B23F1528G, M13F1098G, C23F1786G, A23F1224G, A23F1240G, A23F1269G, B23F1329G, B23F1546G, B23F1547G, D23F1841G, D23F1845G, D23F1848G, D23F1947G, D23F1953G, D23F1954G, D23F1956G, D23F2013G, D23F2016G, D23F2017G, J13F0242G, J43F0481G, L13F0612G, M13F1122G, C23F1585G, L13F0637G, A23F1291G, A23F1293G, C23F1791G, L13F0713G, L13F0714G, L13F0685G, B23F1363G, C23F1789G, C23F1790G, C23F1813G, D23F1926G, D23F1966G, D23F1969G, D23F2000G, D23F2018G, D23F2019G, J13F0026G, J13F0247G, J13F0316G, J13F0317G, J13F0382G, J13F0383G, J13F0384G, K13F0390G, K13F0463G, L13F0753G, L13F0840G, L13F0841G, L13F0842G, M13F1056G, M13F1160G, B23F1372G, D23F1908G, D23F2012G, E23F2031G, L13F0594G, L13F0606G, M13F1110G, M13F1111G, M13F1112G, M13F1113G, M13F1115G, M13F1116G, M13F1117G, M13F1118G, M13F1119G, K13F0465G, K13F0482G, K13F0483G, L13F0596G, L13F0602G, D23F1998G, L13F0772G, M13F1089G, M13F1143G, M13F0928G, F23F2080G, C23F1785G, C23F1795G, D23F1913G, A23F1247G, A23F1248G, A23F1251G, A23F1252G, A23F1253G, B23F1305G, B23F1306G, B23F1314G, B23F1317G, B23F1318G, B23F1320G, B23F1376G, B23F1407G, B23F1408G, B23F1409G, B23F1410G, B23F1457G, B23F1458G, B23F1459G, B23F1460G, B23F1461G, B23F1462G, B23F1463G, B23F1464G, B23F1472G, B23F1474G, B23F1481G, B23F1482G, B23F1491G, B23F1492G, B23F1493G, B23F1494G, C23F1695G, C23F1696G, C23F1697G, C23F1698G, C23F1700G, C23F1701G, C23F1735G, C23F1749G, L13F0599G, D23F1871G, F23F2101G, J13F0343G, J13F0344G, L13F0695G, L13F0696G, L13F0760G, L13F0761G, M13F1082G, C23F1740G, C23F1747G, B23F1505G, D23F1860G, B23F1497G, C23F1774G, K13F0396G, K13F0398G, K13F0399G, K13F0400G, K13F0401G, L13F0646G, M13F0894G, B23F1387G, J13F0084G, J13F0102G, J13F0342F, J13F0342G, K13F0395G, K13F0494G, L13F0588G, K13F0517G, K13F0425G, J13F0004G, J13F0005G, J13F0077G, J13F0078G, J13F0206G, J13F0207G, D23F1865G, E23F2025G, D23F1988G, D23F1989G, D23F1993G, M13F0996G, A23F1296G, C23F1587G, C23F1604G, D23F1945G, L13F0699G, L13F0700G, L13F0701G, L13F0702G, L13F0703G, L13F0704G, L13F0705G, L13F0706G, J13F0223G, J13F0232G, J13F0233G, J13F0234G, L13F0790G, L13F0791G, M13F1128G, M13F1129G, M13F1057G, M13F1058G, D23F2009G, E23F2073G, E23F2076G, E23F2077G, E23F2078G, K13F0561G, J13F0126G, B23F1351G, L13F0860G, F23F2084G, F23F2085G, F23F2086G, F23F2089G, D23F1847G, D23F1880G, D23F1881G, D23F1928G, D23F1944G, C23F1755G, B23F1361G, B23F1364G, L13F0773G, M13F0965G, M13F0966G, M13F0967G, L13F0650G, L13F0651G, L13F0652G, J13F0211G, B23F1338G, A23F1287G, A23F1288G, B23F1322G, B23F1426G, B23F1427G, B23F1430G, B23F1431G, B23F1432G, B23F1433G, B23F1438G, B23F1441G, K13F0491G, L13F0632G, D23F2007G, L13F0694G, J13F0340G, M13F1140G, L13F0778G, L13F0779G, L13F0781G, C23F1729G, J13F0246G, L13F0666G, A23F1285G, B23F1379G, B23F1380G, B23F1381G, B23F1382G, B23F1392G, B23F1401G, B23F1429G, B23F1434G, B23F1436G, B23F1422G, C23F1753G, J13F0096G, B23F1524G, D23F2004G, D23F2021G, J13F0090G, J13F0205G, A23F1272G, L13F0722G, A23F1292G, B23F1319G, B23F1418G, B23F1420G, B23F1475G, K13F0474G, L13F0821G, L13F0822G, L13F0823G, M13F1095G, M13F1096G, M13F1100G, M13F1101G, M13F1102G, M13F1127G, M13F1097G, A23F1234G, J13F0231G, M13F1175G, J13F0239G, J13F0329G, J13F0365G, J13F0377G, B23F1368G, B23F1523G, M13F1158G, C23F1751G, C23F1692G, D23F1894G, D23F1895G, D23F1897G, A23F1182G, A23F1289G, B23F1389G, B23F1503G, D23F1842G, M13F1083G, J13F0243G, J13F0341G, B23F1469G, B23F1476G, C23F1724G, C23F1731G, J13F0248G, J13F0249G, B23F1310G, B23F1344G, K13F0490G, K13F0495G, K13F0496G, K13F0502G, M13F1156G, K13F0514G, L13F0728G, J13F0269G, A23F1237G, J13F0251G, J13F0279G, J13F0289G, J13F0290G, L13F0848G, M13F0947G, J13F0235G, B23F1399G, B23F1400G, K13F0478G, K13F0479G, K13F0480G, A23F1298G, B23F1349G, M13F1055G, M13F1065G, M13F1086G, B23F1425G, C23F1797G, C23F1598G, J13F0093G, J13F0094C, J13F0094G, K13F0428G, K13F0436G, L13F0580G, M13F0992G, M13F1076G, B23F1411G, C23F1657G, B23F1340G, B23F1359G, B23F1417G, B23F1468G, B23F1480G, B23F1543G, C23F1703G, C23F1754G, C23F1792G, C23F1794G, C23F1798G, C23F1820G, D23F1872G, B23F1542G, F23F2097G, F23F2099G, F23F2100G, J13F0350G, J13F0351G, C23F1804G, L13F0592G, L13F0638G, A23F1223G, A23F1277G, A23F1278G, A23F1279G, A23F1280G, B23F1378G, J13F0349G, L13F0708G, L13F0709G, L13F0710G, L13F0711G, L13F0723G, L13F0836G, B23F1448G, M13F0926G, J13F0228G, L13F0593G, B23F1391G, M13F0944G, K13F0545G, L13F0754G, L13F0757G, L13F0758G, L13F0759G, M13F0986G, J13F0240G, K13F0552G, K13F0553G, K13F0554G, L13F0610G, C23F1733G, C23F1775G, C23F1776G, C23F1778G, B23F1500G, L13F0667G, L13F0677G, L13F0689G, L13F0690G, J13F0204G, J13F0210G, J13F0270G, J13F0339G, B23F1446G, D23F1907G, D23F1909G, D23F1972G, D23F1973G, D23F1984G, D23F1986G, D23F1990G, E23F2023G, L13F0777G, L13F0820G, L13F0827G, L13F0850G, L13F0851G, L13F0852G, L13F0853G, L13F0854G, L13F0855G, M13F1139G, K13F0413G, K13F0520G, L13F0587G, L13F0589G, L13F0603G, L13F0604G, K13F0537G, B23F1323G, L13F0796G, L13F0797G, L13F0800G, L13F0801G, L13F0803G, J13F0241G, A23F1228G, M13F0974G, M13F0979G, F23F2114G, F23F2115G, F23F2116G, F23F2117G, F23F2118G, F23F2119G, J13F0253G, J13F0254G, J13F0262G, J13F0263G, J13F0264G, J13F0275G, J13F0276G, J13F0277G, J13F0302G, J13F0303G, J13F0304G, J13F0305G, C23F1592G, D23F1911G, D23F1912G, D23F1914G, D23F1927G, M13F1132G, M13F1133G, J13F0196G, D23F2006G, B23F1339G, B23F1341G, B23F1442G, B23F1449G, B23F1451G, B23F1535G, C23F1732G, C23F1734G, C23F1745G, C23F1746G, D23F1991G, L13F0693G, M13F1068G, A23F1230G, A23F1233G, A23F1235G, D23F1974G, J13F0038G, J13F0286G, J13F0370G, J13F0372G, J13F0373G, K13F0388G, K13F0393G, K13F0394G, K13F0416G, K13F0417G, K13F0418G, K13F0420G, K13F0421G, K13F0422G, K13F0423G, K13F0424G, L13F0654G, A23F1190G, L13F0597G, A23F1284G, K13F0515G, M13F0962G, A23F1226G, B23F1331G, B23F1334G, B23F1335G, B23F1336G, B23F1337G, B23F1350G, B23F1352G, B23F1353G, B23F1354G, B23F1367G, B23F1369G, B23F1370G, B23F1444G, B23F1445G, B23F1450G, B23F1453G, D23F1970G, E23F2026G, J13F0097G, J13F0318G, J13F0327G, J13F0357G, A23F1281G, A23F1283G, B23F1332G, A23F1225G, A23F1227G, B23F1326G, B23F1343G, B23F1345G, B23F1356G, B23F1360G, B23F1375G, B23F1383G, B23F1384G, B23F1385F, B23F1385G, B23F1455G, B23F1456G, B23F1467G, B23F1470G, B23F1471G, B23F1875G, C23F1691G, C23F1806G, C23F1810G, C23F1816G, D23F1843G, D23F1866G, D23F1885G, D23F1904G, D23F1905G, D23F1937G, D23F1948G, D23F1949G, D23F1950G, D23F1951G, D23F1952G, D23F1957G, D23F1959G, D23F2015G, J13F0168G, J13F0366G, L13F0817G, L13F0818G, L13F0819G, L13F0825G, L13F0826G, L13F0857G, M13F1157G, A23F1206G, J13F0216G, J13F0217G, K13F0438G, K13F0458G, K13F0459G, M13F1170G, D23F1964G, J13F0236G, J13F0237G, A23F1212G, B23F1540G, D23F1849G, J13F0042G, J13F0069G, J13F0085G, J13F0091G, J13F0092G, J13F0106G, J13F0107G, J13F0118G, J13F0119G, J13F0175G, J13F0177G, J13F0182G, J13F0198G, J13F0199G, J13F0208G, J13F0209G, J13F0212G, J13F0213G, J13F0214G, J13F0215G, J13F0296G, J13F0299G, J13F0301G, J13F0307G, J13F0309G, J13F0310G, J13F0311G, J13F0312G, J13F0360G, K13F0486G, K13F0487G, K13F0488G, L13F0815G, L13F0824G, A23F1202G, B23F1393G, J13F0170G, J13F0171G, J13F0265G, J13F0285G, K13F0403G, K13F0538G, L13F0774G, L13F0775G, M13F1159G, M13F1171G, F23F2082G, F23F2083G, D23F1999G, B23F1465G, D23F2011G, L13F0611G, K13F0454G, B23F1507G, D23F1836G, J13F0079G, J13F0080G, J13F0202G, J13F0244G, J13F0345G, J13F0346G, J13F0347G, J13F0354G, J13F0355G, J13F0356G, K13F0460G, K13F0508G, M13F1134G, M13F1135G, B23F1435G, B23F1478G, B23F1479G, L13F0756G, B23F1419G, D23F2008G, A23F1195G, B23F1321G, B23F1504G, J13F0001G, J13F0002G, J13F0331G, J13F0332G, J13F0361G, J13F0362G, J13F0363G, L13F0843G, L13F0844G, M13F0884G, M13F0885G, M13F0886G, M13F0887G, M13F0987G, B23F1327G, B23F1325G, J13F0278G, L13F0640G, D23F1870G, A23F1188G, L13F0712G, M13F0997G, M13F1120G, M13F1121G, M13F1123G, M13F1124G, M13F1125G, M13F1126G, J13F0328G, J13F0376G, L13F0635G, L13F0636G, J13F0201G, J13F0221G, J13F0230G, M13F1178G, L13F0572G, J13F0218G, K13F0441G, K13F0511G, L13F0780G, M13F0935G, M13F1072G, J13F0322G, C23F1780G, B23F1533G, B23F1386G, B23F1377G, J13F0191G, J13F0192G, D23F1890G, D23F1891G, C23F1742G, C23F1743G, C23F1744G, C23F1748G, C23F1758G, D23F1955G, D23F1992G, L13F0737G, L13F0750G, L13F0751G, L13F0752G, D23F2020G, J13F0238G, A23F1265G, B23F1447G, B23F1519G, A23F1264G, B23F1514G, B23F1548G, B23F1549G, B23F1553G, B23F1554G, B23F1555G, C23F1586G, C23F1593G, C23F1595G, C23F1596G, C23F1597G, C23F1605G, C23F1606G, C23F1819G, D23F1863G, D23F1869G, D23F1910G, D23F2002G, D23F2005G, E23F2024G, E23F2061G, J13F00081G, J13F0065G, J13F0066G, J13F0081G, J13F0082G, J13F0083G, J13F0101G, J13F0103G, J13F0256G, J13F0257G, J13F0259G, J13F0261G, K13F0392G, K13F0405G, K13F0419G, L13F0003G, L13F0573G, L13F0574G, L13F0575G, L13F0688G, L13F0733G, L13F0738G, L13F0740G, M13F0882G, M13F1041G, M13F1060G, M13F1061G, C23F1603G, D23F1965G, E23F2065G, B23F1311G, B23F1330G, A23F1295G, M13F1138G, M13F1144G, M13F1145G, L13F0725G, M13F0888G, J13F0095G, J13F0184G, J13F0224G, J13F0225G, J13F0226G, A23F1191G, A23F1199G, A23F1208G, A23F1209G, A23F1210G, A23F1211G, A23F1218G, A23F1219G, A23F1220G, D23F1850G, D23F1858G, D23F1862G, K13F0489G, M13F1161G, M13F0961G, B23F1301G, J13F0297G, D23F1934G, F23F2109G, B23F1333G, D23F1939G, D23F1940G, K13F0449G, D23F1962G, E23F2027G, C23F1713G, C23F1787G, D23F1827G, D23F1828G, J13F0044G, J13F0185G, J13F0193G, J13F0200G, L13F0581G, L13F0681G, M13F0898G, M13F0899G, D23F1832G, J13F0064G, J13F0186G, J13F0195G, L13F0598G, L13F0605G, L13F0648G, L13F0673G, L13F0674G, L13F0676G, M13F0936G, M13F0959G, K13F0569G, J13F0128G, J13F0141G, J13F0282G, J13F0283G, J13F0330G, K13F0410G, K13F0461G, M13F0953G, M13F0956G, M13F0963G, M13F0964G, M13F0970G, M13F0971G, J13F0029G, L13F0707G, L13F0655G, C23F1727G, C23F1805G, C23F1807G, C23F1811G, K13F0386G, J13F0023G, J13F0122G, B23F1525G, M13F0960G, M13F0977G, J13F0114G, J13F0123G, J13F0129G, J13F0194G, J13F0227G, K13F0468G, K13F0469G, K13F0476G, L13F0577G, L13F0578G, M13F0985G, M13F0895G, M13F1104G, M13F1105G, J13F0160G, A23F1241G, A23F1242G, A23F1245G, B23F1421G, D23F1941G, L13F0579G, L13F0670G, M13F0931G, B23F1397G, C23F1668G, J13F0138G, M13F0981G, C23F1717G, J13F0121G, C23F1722G, J13F0137G, L13F0837G, C23F1721G, J13F0139G, J13F0140G, J13F0054G, J13F0154G, L13F0839G, J13F0144G, A23F1201G, C23F1645G, J13F0125G, J13F0055G, J13F0150G, J13F0143G, J13F0149G, C23F1718G, J13F0052G, M13F0952G, J13F0104G, J13F0130G, J13F0131G, J13F0146G, J13F0147G, L13F0834G, L13F0835G, M13F0877G, C23F1591G, J13F0153G, J13F0157G, C23F1648G, B23F1303G, A23F1238G, A23F1244G, C23F1644G, M13F0930G, J13F0142G, K13F0412G, D23F1838G, C23F1658G, J13F0053G, J13F0117G, J13F0151G, J13F0156G, J13F0158G, J13F0159G, L13F0867G, L13F0868G, L13F0869G, L13F0871G, M13F0873G, M13F0874G, M13F0875G, M13F0879G, M13F0880G, M13F0881G, J13F0046G, J13F0113G, J13F0148G, J13F0162G, J13F0163G, L13F0831G, L13F0833G, L13F0830G, A23F1243G, J13F0152G, M13F0929G, J13F0105G, L13F0832G, C23F1643G, C23F1623G, C23F1642G, C23F1647G, L13F0872G, J13F0112G, L13F0870G, J13F0155G, J13F0145G, J13F0161G, M13F0876G, J13F0180G, J13F0181G, J13F0187G, J13F0188G, C23F1781G, J13F0178G, C23F1655G, C23F1656G, C23F1652G, C23F1803G, D23F1868G, D23F1900G, D23F2003G, M13F1176G, F23F2102G, F23F2103G, B23F1415G, J13F0041G, J13F0050G, L13F0672G, B23F1373G, E23F2034G, M13F0942G, J13F0043G, J13F0040G, J13F0057G, J13F0059G, L13F0675G, B23F1487G, M13F0914G, M13F0915G, A23F1262G, D23F1831F, D23F1831G, D23F1935G, B23F1374G, M13F1036G, L13F0828G, L13F0829G, L13F0802G, L13F0838F, L13F0838G, C23F1680G, C23F16826, C23F1682G, B23F1978G, D23F1978G, B23F1371G, C23F1584G, C23F1662G, C23F1663G, C23F1665G, C23F1681G, C23F1683G, C23F1684G, C23F1689G, C23F1708G, C23F1709G, C23F1738G, J13F0050GA, A23F1216G, J13F0051G, J13F0056G, J13F0058G, M13F0058G, M13F1093G, J13F0039G, M13F0955G, M13F1147G, D23F1932G, J13F0024G, J13F0025G, J13F0047G, L13F0629G, L13F0630G, A23F1271G, H13F0934G, B23F1488G, E23F2037G, B23F1489G, C23F1660G, J13F0027G, J13F0031G, J13F0035G, L13F0671G, B23F1365G, L13F0763G, D23F1929G, K13F0455G, M13F0902G, D23F1930G, C23F1615G, B23F1395G, B23F1398G, D23F1975G, L13F0583G, C23F1737G, C23F1788G, E23F2060G, L13F0858G, L13F0859G, M13F0896G, M13F0909G, J13F0019G, J13F0020G, K13F0457G, L13F0582G, M13F0920G, M13F1037G, B23F1485G, B23F1486G, C23F1617G, L13F0585G, B23F1405G, D23F1875G, B23F1403G, B23F1483G, B23F1490G, E23F2029G, M13F0901G, L13F8192G, J13F0033G, J13F0037G, J13F0061G, J13F0062G, M13F0900G, E23F2062G, D23F1976G, D23F1977G, C23F1616G, D23F1996G, J13F0032G, F23F2106G, L13F0584G, K13F0568G, J13F0063G, L13F0764G, H13F0567G, L13F06716, B23F1404G, M13F0897G, C23F1714G, M13F0918G, C23F1711G, C23F1712G, E23F2038G, E23F2039G, G23F1711G, F23F2095G, E23F2032G, E23F2033G, E23F2035G, F23F2107G, B13F3997G, A23F1217G, B23F1406G, J13F0021G, L13F0653G, J13F0022G, F23F2105G, C23F1685G, F23F2094G, F23F2093G, J13F0028G, M13F0921G, B23F1414G, K13F0492G, L13F0697G, L13F0726G, L13F0727G, C23F1716G, M13F1088G, M13F1087G, D23F1837G, L13F0731G, L13F0659G, L13F0660G, D23F1844G, M13F0910G, M13F0911G, M13F0912G, D23F1902G, D23F1903G, C23F1608G, D23F1979G, F23F2104G, J13F0076G, J13F0011G, L13F0845G, A23F1222G, L13F0783G, K13F0472G, K13F0473G, L13F0863G, J13F0045G, J13F0012G, M13F0958G, K13F0470G, K13F0471G, K13F0485G, K13F0484G, C23F1599G, C23F1600G, C23F1601G, C23F1602G, C23F1670G, C23F1671G, L13F0732G, J13F0013G, D23F1857G, A23F1183G, A23F1184G, L13F0856G, M13F1028G, C23F1760G, C23F1761G, C23F1762G, C23F1609G, C23F1613G, M13F1017G, D23F1915G, M13F1023G, M13F1004G, C23F1636G, D23F1919G, C23F1759G, D23F1840G, M13F1020G, M13F1021G, M13F1022G, M13F1024G, M13F1027G, C23F1625G, C23F1632G, C23F1633G, C23F1635G, E23F2058G, E23F2059G, C23F1763G, C23F1766G, C23F1627G, D23F2022G, E23F2063G, M13F0988G, M13F1025G, M13F1026G, D23F1916G, D23F1917G, D23F1918G, D23F1920G, D23F1921G, D23F1923G, D23F1924G, M13F0999G, M13F1000G, M13F1001G, M13F1002G, M13F1003G, E23F2053G, E23F2054G, C23F1628G, M13F1019G, E23F2051G, M13F1007G, M13F1011G, M13F1012G, C23F1629G, C23F1672G, C23F1673G, C23F1674G, C23F1676G, C23F1677G, C23F1679G, C23F1688G, C23F1705G, E23F2043G, C23F1764G, C23F1765G, E23F2070G, C23F1614G, E23F2052G, C23F1687G, C23F1704G, E23F2050G, C23F1612G, C23F1634G, F23F2044G, C23F1610G, C23F1626G, C23F1630G, B23F1511G, L13F0684G, L13F0686G, M13F0937G, D23F1834G, A23F1254G, A23F1255G, A23F1259G, A23F1261G, J13F0088G, J13F0115G, D23F1833G, C23F1621G, J13F0016G, M13F1064G, J13F0086G, D23F1983G, K13F0439G, E23F2040G, D23F1822G, D23F1823G, D23F1824G, D23F1825G, J13F0018G, L13F0619G, L13F0620G, L13F0622G, L13F0736G, L13F0784G, J13F0164G, J03F0164G, J13F0338G, L13F0847G, L13F0849G, M13F0890G, L13F0627G, L13F0643G, L13F0644G, L13F0645G, L13F0656G, L13F0657G, L13F0658G, J13F0306G, C23F1637G, C23F1638G, M13F1103G, C23F1782G, M13F0904G, M13F0905G, M13F0906G, M13F1077G, M13F1078G, M13F1079G, M13F1080G, M13F1081G, M13F1090G, M13F1091G, M13F1092G, D23F1874G, E23F2028G, K13F0467G, L13F0789G, E23F2071G, K13F0555G, J13F0172G, J13F0173G, J13F0176G, J13F0353G, J13F0068G, J13F0120G, J13F0127G, J13F0165G, J13F0371G, K13F0445G, K13F0446G, K13F0447G, C13F1620G, C23F1619G, C23F1620G, K13F0452G, K13F0453G, M13F1177G, E23F2041G, C23F1646G, C23F1730G, D23F1821G, D23F1961G, L13F0716G, B23F1416G, B23F1522G, B23F1556G, C23F1567G, D23F1906G, C23F1571G, K13F0442G, K13F0443G, K13F0444G, B23F1557G, B23F1552G, C23F1578G, C23F1579G, C23F1580G, C23F1562G, C23F1560G, C23F1569G, C23F1572G, C23F1573G, L13F0590G, L13F0591G, L13F0595G, B23F1558G, C23F1564G, C23F1566G, C23F1561G, C23F1568G, C23F1574G, B23F1550G, B23F1559G, C23F1563G, C23F1565G, C23F1570G, C23F1577G, C23F1581G, C23F1582G, C23F1583G, C23F1618G, F23F2110G, K13F0546G, K13F0547G, L13F0626G, C23F1575G, B23F1394G, C23F1702G, D23F1901G, J13F0015G, L13F0662G, C23F1669G, C23F1736G, A23F1194G, K13F0466G, M13F0913G, M13F0932G, M13F0916G, B23F1299G, A23F1193G, F23F2108G, L13F0762G, B23F1515G, B23F1520G, J13F0135G, J13F0136G, B23F1484G, C23F1661G, C23F1710G, C23F1715G, D23F1829G, D23F1830G, D23F1931G, L13F0586G, A23F1179G, L13F0601G, M13F0938G, J13F0110G, C23F1667G, A23F1204G, M13F0883, M13F0883G, M13F0883S, C23F1659G, A23F1203G, L13F0771G, A23F1268G, E23F2042G, M13F0923G, C23F1640G, C23F1641G, D23F1826G, K13F0427G, K13F0429G, K13F0434G, K13F0440G, B23F1443G, K13F0448G, B23F1413G, C23F1654G, K13F0430G, C23F1653G, L13F0765G, L13F0766G, A23F1192G, A23F1200G, J13F0111G, L13F0661G, M13F0957G, B23F1412G, L13F0816G, C23F16549, D23F1963G, A23F1205G, J13F0174G, C23F1639G;     2. Product Code: 4T, UDI Number (Serial Numbers): Not Applicable (F24T0254G), Not Applicable (F24T0259G), Not Applicable (F24T0264G), Not Applicable (F24T0279G), Not Applicable (A44T4691G), Not Applicable (A44T4713G), Not Applicable (C44T4931G), Not Applicable (D34T2574G), Not Applicable (H34T3263G), Not Applicable (H34T3264G), Not Applicable (H34T3266G), Not Applicable (H34T3267G), Not Applicable (H34T3268G), Not Applicable (H34T3270G), Not Applicable (H34T3272G), Not Applicable (H34T3274G), Not Applicable (M24T1660G), Not Applicable (M24T1661G), Not Applicable (M24T1662G), Not Applicable (M24T1672G), Not Applicable (M24T1710G), Not Applicable (M24T1716G), Not Applicable (M24T1718G), Not Applicable (M24T1721G), Not Applicable (M34T4314G), Not Applicable (M34T4315G), Not Applicable (M34T4316G), Not Applicable (M34T4321G), Not Applicable (M34T4360G), Not Applicable (M34T4368G), Not Applicable (M34T4482G), Not Applicable (M34T4215G), Not Applicable (M34T4216G), Not Applicable (J34T3725G), Not Applicable (J34T3752G), Not Applicable (J34T3755G), Not Applicable (M34T4340G), Not Applicable (G24T0384G), Not Applicable (K24T1151G), Not Applicable (L24T1274G), Not Applicable (L24T1284G), Not Applicable (L24T1377G), Not Applicable (L24T1401G), Not Applicable (L34T4001G), Not Applicable (A44T4739G), Not Applicable (G24T0475G), Not Applicable (A44T4573G), Not Applicable (C34T2261G), Not Applicable (B44T4778G), Not Applicable (F24T0081G), Not Applicable (F24T0292G), Not Applicable (F24T0294G), Not Applicable (F24T0296G), Not Applicable (F24T0297G), Not Applicable (F24T0318G), Not Applicable (F34T2928G), Not Applicable (K24T1003G), Not Applicable (L34T4174G), Not Applicable (F34T3021G), Not Applicable (C34T2336G), Not Applicable (F24T0242G), Not Applicable (L34T4041G), Not Applicable (M34T4479G), Not Applicable (F24T0084G), Not Applicable (G24T0564G), Not Applicable (M24T1736G), Not Applicable (M34T4449G), Not Applicable (M34T4450G), Not Applicable (B34T1952G), Not Applicable (A34T1762G), Not Applicable (F24T0124G), Not Applicable (F24T0241G), Not Applicable (F34T2959G), Not Applicable (J24T0609G), Not Applicable (J24T0826G), Not Applicable (K24T1143G), Not Applicable (K34T3974G), Not Applicable (M34T4381G), Not Applicable (A34T1853G), Not Applicable (B44T4800G), Not Applicable (D44T5014G), Not Applicable (K34T3833G), Not Applicable (K34T3834G), Not Applicable (K34T3837G), Not Applicable (K34T3843G), Not Applicable (K34T3858G), Not Applicable (K34T3859G), Not Applicable (K34T3966G), Not Applicable (M24T1560G), Not Applicable (M24T1665G), Not Applicable (M24T1673G), Not Applicable (M24T1692G), Not Applicable (M34T4413G), Not Applicable (M34T4415G), Not Applicable (M34T4442G), Not Applicable (M34T4443G), Not Applicable (M34T4444G), Not Applicable (M34T4445G), Not Applicable (L34T4036G), Not Applicable (C34T2256G), Not Applicable (K24T1165G), Not Applicable (L24T1309G), Not Applicable (M24T1528G), Not Applicable (C34T2324G), Not Applicable (C34T2326G), Not Applicable (C34T2328G), Not Applicable (C34T2341G), Not Applicable (A34T1825G), Not Applicable (A34T1826G), Not Applicable (A34T1827G), Not Applicable (A34T1829G), Not Applicable (A34T1830G), Not Applicable (A34T1831G), Not Applicable (A34T1832G), Not Applicable (A34T1848G), Not Applicable (A34T1896G), Not Applicable (A34T1899G), Not Applicable (A34T1900G), Not Applicable (A34T1903G), Not Applicable (B34T2033G), Not Applicable (C34T2298G), Not Applicable (C44T4984G), Not Applicable (G24T0512G), Not Applicable (J34T3789G), Not Applicable (J34T3791G), Not Applicable (C34T2275G), Not Applicable (C34T2361G), Not Applicable (C34T2365G), Not Applicable (C34T2378G), Not Applicable (F24T0247G), Not Applicable (J24T0867G), Not Applicable (J24T0873G), Not Applicable (J34T3597G), Not Applicable (J34T3598G), Not Applicable (J34T3604G), Not Applicable (J34T3605G), Not Applicable (J34T3606G), Not Applicable (J34T3628G), Not Applicable (J34T3631G), Not Applicable (J34T3669G), Not Applicable (K34T3871G), Not Applicable (F34T2927G), Not Applicable (M24T1608G), Not Applicable (M24T1647G), Not Applicable (D34T2530G), Not Applicable (D34T2534G), Not Applicable (D34T2538G), Not Applicable (J34T3517G), Not Applicable (J24T0831G), Not Applicable (C34T2265G), Not Applicable (C34T2267G), Not Applicable (F24T0269G), Not Applicable (J34T3816G), Not Applicable (B34T1947G), Not Applicable (B34T1951G), Not Applicable (J24T0846G), Not Applicable (J24T0862G), Not Applicable (J24T0876G), Not Applicable (H34T3437G), Not Applicable (A44T4581G), Not Applicable (B34T1984G), Not Applicable (B34T1989G), Not Applicable (B44T4905G), Not Applicable (B44T4913G), Not Applicable (C44T4930G), Not Applicable (F24T0054G), Not Applicable (F24T0239G), Not Applicable (G24T0414G), Not Applicable (G24T0506G), Not Applicable (G24T0520G), Not Applicable (G24T0521G), Not Applicable (G24T0524G), Not Applicable (H34T3284G), Not Applicable (J24T0738G), Not Applicable (J24T0833G), Not Applicable (J24T0836G), Not Applicable (M24T1668G), Not Applicable (M24T1671G), Not Applicable (D34T2454G), Not Applicable (D34T2535G), Not Applicable (D34T2545G), Not Applicable (F24T0053G), Not Applicable (D44T5027G), Not Applicable (K24T1117G), Not Applicable (K24T1118G), Not Applicable (K24T1190G), Not Applicable (K24T1191G), Not Applicable (K24T1196G), Not Applicable (K24T1197G), Not Applicable (K24T1200G), Not Applicable (B44T4880G), Not Applicable (B44T4882G), Not Applicable (J24T0642G), Not Applicable (J24T0813G), Not Applicable (M24T1569G), Not Applicable (B34T2021G), Not Applicable (M24T1540G), Not Applicable (J34T3601G), Not Applicable (M24T1603G), Not Applicable (M24T1749G), Not Applicable (M34T4339G), Not Applicable (M34T4493G), Not Applicable (M34T4494G), Not Applicable (K34T3968G), Not Applicable (D34T2629G), Not Applicable (B34T2006G), Not Applicable (M24T1609G), Not Applicable (M24T1663G), Not Applicable (A44T4676G), Not Applicable (B34T1991G), Not Applicable (B34T1998G), Not Applicable (B34T1999G), Not Applicable (B34T2105G), Not Applicable (B34T2107G), Not Applicable (B34T2108G), Not Applicable (B34T2109G), Not Applicable (D34T2440G), Not Applicable (H34T3419G), Not Applicable (J24T0704G), Not Applicable (J24T0790G), Not Applicable (J24T0792G), Not Applicable (J24T0799G), Not Applicable (J24T0854G), Not Applicable (J34T3807G), Not Applicable (K34T3839G), Not Applicable (K34T3840G), Not Applicable (K34T3841G), Not Applicable (K34T3844G), Not Applicable (L24T1304G), Not Applicable (F34T3027G), Not Applicable (K34T3920G), Not Applicable (L34T4102G), Not Applicable (L34T4103G), Not Applicable (L34T4104G), Not Applicable (L34T4105G), Not Applicable (L34T4106G), Not Applicable (K24T1214G), Not Applicable (K24T1215G), Not Applicable (K24T1225G), Not Applicable (L24T1362G), Not Applicable (H34T3477G), Not Applicable (K24T1266G), Not Applicable (L34T4005G), Not Applicable (M24T1455G), Not Applicable (F34T2841G), Not Applicable (M34T4432G), Not Applicable (F24T0308G), Not Applicable (F24T0309G), Not Applicable (G24T0417G), Not Applicable (G24T0420G), Not Applicable (J34T3521G), Not Applicable (G24T0462G), Not Applicable (G24T0470G), Not Applicable (B34T1979G), Not Applicable (B34T1986G), Not Applicable (B34T1987G), Not Applicable (L24T1294G), Not Applicable (L24T1303G), Not Applicable (L24T1315G), Not Applicable (H34T3361G), Not Applicable (K24T1193G), Not Applicable (K24T1198G), Not Applicable (B34T2043G), Not Applicable (B34T2047G), Not Applicable (B34T2055G), Not Applicable (B34T2202G), Not Applicable (L24T1391G), Not Applicable (F24T0278G), Not Applicable (F24T0149G), Not Applicable (F24T0211G), Not Applicable (G24T0324G), Not Applicable (H34T3435G), Not Applicable (M24T1512G), Not Applicable (J24T0900G), Not Applicable (L34T4109G), Not Applicable (M34T4446G), Not Applicable (M34T4481G), Not Applicable (M34T4539G), Not Applicable (H34T3441G), Not Applicable (K34T3978G), Not Applicable (B44T4781G), Not Applicable (F34T2978G), Not Applicable (B44T4745G), Not Applicable (B44T4751G), Not Applicable (D34T2470G), Not Applicable (K24T1184G), Not Applicable (A34T1781G), Not Applicable (A34T1784G), Not Applicable (A34T1785G), Not Applicable (A34T1789G), Not Applicable (A34T1791G), Not Applicable (C34T2407G), Not Applicable (J24T0823G), Not Applicable (D34T2509G), Not Applicable (K34T3886G), Not Applicable (K34T3928G), Not Applicable (K34T3960G), Not Applicable (F24T0276G), Not Applicable (B34T2163G), Not Applicable (B44T4783G), Not Applicable (C34T2235G), Not Applicable (F24T0256G), Not Applicable (J24T0671G), Not Applicable (J24T0700G), Not Applicable (J24T0879G), Not Applicable (J24T0884G), Not Applicable (J34T3686G), Not Applicable (J34T3688G), Not Applicable (J34T3689G), Not Applicable (J34T3693G), Not Applicable (K34T3900G), Not Applicable (K34T3902G), Not Applicable (A44T4704G), Not Applicable (A44T4707G), Not Applicable (K24T1033G), Not Applicable (K24T1039G), Not Applicable (A44T4637G), Not Applicable (A44T4639G), Not Applicable (B44T4747G), Not Applicable (F24T0203G), Not Applicable (G34T3207G), Not Applicable (B34T2019G), Not Applicable (D34T2473G), Not Applicable (D34T2475G), Not Applicable (D34T2491G), Not Applicable (D34T2493G), Not Applicable (D34T2495G), Not Applicable (D34T2518G), Not Applicable (D34T2527G), Not Applicable (D34T2541G), Not Applicable (D34T2546G), Not Applicable (D34T2570G), Not Applicable (D34T2602G), Not Applicable (D34T2604G), Not Applicable (D34T2605G), Not Applicable (D34T2606G), Not Applicable (M34T4318G), Not Applicable (M24T1600G), Not Applicable (M24T1700G), Not Applicable (D34T2482G), Not Applicable (D34T2516G), Not Applicable (D34T2573G), Not Applicable (F34T2854G), Not Applicable (F34T2858G), Not Applicable (F34T2863G), Not Applicable (F34T2864G), Not Applicable (F34T2865G), Not Applicable (F34T2867G), Not Applicable (G24T0448G), Not Applicable (K24T1231G), Not Applicable (K24T1239G), Not Applicable (M24T1677G), Not Applicable (M24T1687G), Not Applicable (M24T1706G), Not Applicable (M24T1712G), Not Applicable (C34T2351G), Not Applicable (C34T2372G), Not Applicable (J34T3653G), Not Applicable (H34T3486G), Not Applicable (J34T3681G), Not Applicable (E34T2726G), Not Applicable (E34T2727G), Not Applicable (E34T2729G), Not Applicable (E34T2732G), Not Applicable (E34T2737G), Not Applicable (F24T0288G), Not Applicable (F24T0289G), Not Applicable (F24T0290G), Not Applicable (F24T0291G), Not Applicable (M24T1504G), Not Applicable (M24T1511G), Not Applicable (M24T1555G), Not Applicable (M24T1561G), Not Applicable (M34T4390G), Not Applicable (B44T4909G), Not Applicable (G24T0543G), Not Applicable (G24T0545G), Not Applicable (H24T0594G), Not Applicable (H34T3265G), Not Applicable (A34T1803G), Not Applicable (F24T0271G), Not Applicable (F24T0284G), Not Applicable (F24T0285G), Not Applicable (F24T0298G), Not Applicable (F24T0299G), Not Applicable (F24T0300G), Not Applicable (F24T0301G), Not Applicable (F24T0303G), Not Applicable (F24T0316G), Not Applicable (G24T0484G), Not Applicable (G24T0488G), Not Applicable (G24T0490G), Not Applicable (G24T0499G), Not Applicable (G24T0542G), Not Applicable (J24T0692G), Not Applicable (J24T0709G), Not Applicable (J24T0710G), Not Applicable (J24T0711G), Not Applicable (J24T0712G), Not Applicable (J24T0716G), Not Applicable (J24T0717G), Not Applicable (J24T0718G), Not Applicable (J24T0722G), Not Applicable (J24T0830G), Not Applicable (K24T1004G), Not Applicable (K24T1014G), Not Applicable (K24T1080G), Not Applicable (F24T0236G), Not Applicable (G24T0411G), Not Applicable (B44T4904G), Not Applicable (F24T0130G), Not Applicable (F24T0272G), Not Applicable (F24T0273G), Not Applicable (G24T0529G), Not Applicable (G24T0538G), Not Applicable (J24T0630G), Not Applicable (J24T0651G), Not Applicable (J24T0691G), Not Applicable (J24T0721G), Not Applicable (J24T0723G), Not Applicable (K24T1007G), Not Applicable (K24T1116G), Not Applicable (K24T1119G), Not Applicable (K24T1120G), Not Applicable (K24T1122G), Not Applicable (K24T1123G), Not Applicable (L24TE1JG), Not Applicable (H34T3277G), Not Applicable (B34T2029G), Not Applicable (B34T2203G), Not Applicable (C34T2268G), Not Applicable (C34T2269G), Not Applicable (C34T2281G), Not Applicable (C34T2397G), Not Applicable (F24T0074G), Not Applicable (F24T0077G), Not Applicable (F24T0078G), Not Applicable (F24T0079G), Not Applicable (G24T0468G), Not Applicable (M24T1442G), Not Applicable (M24T1513G), Not Applicable (M34T4452G), Not Applicable (M34T4453G), Not Applicable (A34T1870G), Not Applicable (B34T1942G), Not Applicable (B34T2111G",
      "more_code_info": "), Not Applicable (D34T2542G), Not Applicable (E34T2722G), Not Applicable (E34T2746G), Not Applicable (E34T2765G), Not Applicable (F24T0002G), Not Applicable (F24T0003G), Not Applicable (F24T0004G), Not Applicable (F24T0006G), Not Applicable (F24T0007G), Not Applicable (F24T0008G), Not Applicable (F24T0009G), Not Applicable (F24T0010G), Not Applicable (F24T0011G), Not Applicable (F24T0012G), Not Applicable (F24T0014G), Not Applicable (F24T0015G), Not Applicable (F24T0017G), Not Applicable (F24T0026G), Not Applicable (F24T0028G), Not Applicable (F24T0033G), Not Applicable (F24T0035G), Not Applicable (F24T0036G), Not Applicable (F24T0042G), Not Applicable (F24T0043G), Not Applicable (F24T0068G), Not Applicable (F24T0069G), Not Applicable (F24T0133G), Not Applicable (F24T0134G), Not Applicable (F34T2818G), Not Applicable (F34T2862G), Not Applicable (F34T3003G), Not Applicable (G24T0343G), Not Applicable (G24T0367G), Not Applicable (G24T0502G), Not Applicable (G24T0504G), Not Applicable (H34T3373G), Not Applicable (J24T0678G), Not Applicable (K24T1160G), Not Applicable (K24T1188G), Not Applicable (K34T3838G), Not Applicable (K34T3865G), Not Applicable (M24T1481G), Not Applicable (M24T1482G), Not Applicable (M24T1573G), Not Applicable (M24T1682G), Not Applicable (M24T1686G), Not Applicable (M34T4533G), Not Applicable (A44T4598G), Not Applicable (A44T4627G), Not Applicable (A44T4714G), Not Applicable (A44T4716G), Not Applicable (B44T4758G), Not Applicable (B44T4769G), Not Applicable (B44T4809G), Not Applicable (H34T3329G), Not Applicable (F34T2951G), Not Applicable (F24T0020G), Not Applicable (F24T0233G), Not Applicable (J24T0617G), Not Applicable (C34T2334G), Not Applicable (G24T0395G), Not Applicable (J24T0637G), Not Applicable (J24T0638G), Not Applicable (J24T0639G), Not Applicable (J34T3662G), Not Applicable (J34T3763G), Not Applicable (M34T4406G), Not Applicable (M34T4407G), Not Applicable (M34T4408G), Not Applicable (M34T4428G), Not Applicable (M34T4429G), Not Applicable (A44T4732G), Not Applicable (C34T2363G), Not Applicable (E34T2731G), Not Applicable (L34T2731G), Not Applicable (L34T4086G), Not Applicable (L34T4125G), Not Applicable (B34T2191G), Not Applicable (L34T4151G), Not Applicable (M34T4531G), Not Applicable (A34T1874G), Not Applicable (A34T1883G), Not Applicable (A34T1884G), Not Applicable (B34T1944G), Not Applicable (F34T2904G), Not Applicable (G24T0349G), Not Applicable (H34T3305G), Not Applicable (J24T0753G), Not Applicable (K24T1015G), Not Applicable (K24T1016G), Not Applicable (K24T1008G), Not Applicable (K24T1034G), Not Applicable (K24T1035G), Not Applicable (E34T2739G), Not Applicable (E34T2740G), Not Applicable (E34T2742G), Not Applicable (E34T2744G), Not Applicable (C34T2290G), Not Applicable (J24T0941G), Not Applicable (L34T4026G), Not Applicable (L34T4079G), Not Applicable (M24T1503G), Not Applicable (G24T0463G), Not Applicable (C34T2299G), Not Applicable (C34T2301G), Not Applicable (J24T0765G), Not Applicable (J24T0767G), Not Applicable (J24T0832G), Not Applicable (J24T0834G), Not Applicable (J24T0863G), Not Applicable (J24T0887G), Not Applicable (K34T3927G), Not Applicable (A44T4547G), Not Applicable (A44T4564G), Not Applicable (A44T4565G), Not Applicable (B34T2048G), Not Applicable (C34T2228G), Not Applicable (C34T2283G), Not Applicable (C34T2325G), Not Applicable (E34T2724G), Not Applicable (F24T0249G), Not Applicable (G34T3077G), Not Applicable (M34T4334G), Not Applicable (M34T4356G), Not Applicable (M34T4434G), Not Applicable (G34T3099G), Not Applicable (G34T3102G), Not Applicable (G34T3109G), Not Applicable (E34T2653G), Not Applicable (C44T4938G), Not Applicable (G24T0509G), Not Applicable (F24T0126G), Not Applicable (H34T3369G), Not Applicable (H34T3370G), Not Applicable (H34T3372G), Not Applicable (H34T3374G), Not Applicable (H34T3376G), Not Applicable (H34T3377G), Not Applicable (H34T3380G), Not Applicable (J34T3800G), Not Applicable (L34T4054G), Not Applicable (H34T3301G), Not Applicable (H34T3302G), Not Applicable (H34T3346G), Not Applicable (J24T0707G), Not Applicable (J24T0759G), Not Applicable (J24T0840G), Not Applicable (J24T0880G), Not Applicable (J24T0905G), Not Applicable (J24T0929G), Not Applicable (J24T0931G), Not Applicable (F34T3012G), Not Applicable (F34T3019G), Not Applicable (M34T4221G), Not Applicable (A34T1765G), Not Applicable (A34T1808G), Not Applicable (A34T1809G), Not Applicable (A34T1815G), Not Applicable (A34T1816G), Not Applicable (A34T1822G), Not Applicable (F34T3004G), Not Applicable (F34T3007G), Not Applicable (M24T1739G), Not Applicable (G24T0368G), Not Applicable (M24T1745G), Not Applicable (M24T1750G), Not Applicable (J24T0610G), Not Applicable (J34T3754G), Not Applicable (J34T3806G), Not Applicable (K34T3823G), Not Applicable (K34T3870G), Not Applicable (K34T3899G), Not Applicable (B44T4801G), Not Applicable (C34T2354G), Not Applicable (C34T2355G), Not Applicable (F24T0235G), Not Applicable (G24T0338G), Not Applicable (J24T0679G), Not Applicable (K24T1220G), Not Applicable (M24T1483G), Not Applicable (M34T4504G), Not Applicable (M34T4505G), Not Applicable (F24T0051G), Not Applicable (H34T3395G), Not Applicable (H34T3396G), Not Applicable (H34T3420G), Not Applicable (J34T3600G), Not Applicable (F24T0244G), Not Applicable (F24T0245G), Not Applicable (F24T0253G), Not Applicable (F34T2835G), Not Applicable (F24T0125G), Not Applicable (F24T0128G), Not Applicable (K24T1148G), Not Applicable (D34T2522G), Not Applicable (D34T2559G), Not Applicable (D34T2576G), Not Applicable (B34T1960G), Not Applicable (C34T2241G), Not Applicable (G34T3062G), Not Applicable (G34T3202G), Not Applicable (G34T3144G), Not Applicable (F24T0270G), Not Applicable (A44T4687G), Not Applicable (K24T0999G), Not Applicable (K24T1042G), Not Applicable (K24T1043G), Not Applicable (M34T4378G), Not Applicable (M34T4486G), Not Applicable (B44T4899G), Not Applicable (D34T2438G), Not Applicable (A34T1879G), Not Applicable (G34T3211G), Not Applicable (L34T4021G), Not Applicable (B44T4744G), Not Applicable (B44T4750G), Not Applicable (B44T4752G), Not Applicable (B44T4754G), Not Applicable (B44T4799G), Not Applicable (B44T4804G), Not Applicable (F24T0047G), Not Applicable (F24T0048G), Not Applicable (L34T4082G), Not Applicable (L34T4084G), Not Applicable (L34T4085G), Not Applicable (L34T4087G), Not Applicable (L34T4088G), Not Applicable (L34T4090G), Not Applicable (B44T4782G), Not Applicable (E34T2668G), Not Applicable (E34T2669G), Not Applicable (L34T4081G), Not Applicable (L34T4089G), Not Applicable (B34T2005G), Not Applicable (K24T1183G), Not Applicable (B34T2065G), Not Applicable (B34T2066G), Not Applicable (B34T2067G), Not Applicable (B34T2069G), Not Applicable (B34T2070G), Not Applicable (B34T2071G), Not Applicable (B34T2072G), Not Applicable (B34T2154G), Not Applicable (B34T2155G), Not Applicable (D34T2524G), Not Applicable (E34T2747G), Not Applicable (E34T2752G), Not Applicable (F34T2888G), Not Applicable (F34T2990G), Not Applicable (M24T1584G), Not Applicable (M24T1585G), Not Applicable (J34T3674G), Not Applicable (B34T1930G), Not Applicable (B34T1938G), Not Applicable (B34T2135G), Not Applicable (C34T2231G), Not Applicable (C34T2237G), Not Applicable (F34T2824G), Not Applicable (K24T1101G), Not Applicable (K24T1126G), Not Applicable (K24T1127G), Not Applicable (K24T1186G), Not Applicable (K24T1224G), Not Applicable (K24T1236G), Not Applicable (K24T1238G), Not Applicable (L24T1327G), Not Applicable (L34T4025G), Not Applicable (M34T4435G), Not Applicable (B34T1977G), Not Applicable (D34T2480G), Not Applicable (D34T2555G), Not Applicable (D34T2560G), Not Applicable (F24T0001G), Not Applicable (F24T0002), Not Applicable (F24T0005G), Not Applicable (F24T0013G), Not Applicable (F24T0016G), Not Applicable (F24T0025G), Not Applicable (F24T0027G), Not Applicable (F24T0028), Not Applicable (F24T0029G), Not Applicable (F24T0030G), Not Applicable (F24T0031G), Not Applicable (F24T0032), Not Applicable (F24T0032G), Not Applicable (F24T0034G), Not Applicable (F24T0037G), Not Applicable (F24T0038G), Not Applicable (F24T0039G), Not Applicable (F24T0040G), Not Applicable (F24T0059G), Not Applicable (F24T0067G), Not Applicable (F24T0072G), Not Applicable (F34T2834G), Not Applicable (F34T2870G), Not Applicable (F34T2872G), Not Applicable (F34T3035G), Not Applicable (H34T3285G), Not Applicable (H34T3292G), Not Applicable (H34T3334G), Not Applicable (L24T1322G), Not Applicable (M34T4394G), Not Applicable (C44T4925G), Not Applicable (C44T4947G), Not Applicable (F24T0221G), Not Applicable (F24T0222G), Not Applicable (J34T3775G), Not Applicable (L34T4057G), Not Applicable (L34T4058G), Not Applicable (L34T4059G), Not Applicable (L34T4060G), Not Applicable (L34T4061G), Not Applicable (L34T4063G), Not Applicable (L34T4065G), Not Applicable (L34T4066G), Not Applicable (F24T0260G), Not Applicable (F24T0261G), Not Applicable (F24T0262G), Not Applicable (F24T0263G), Not Applicable (F24T0265G), Not Applicable (F24T0266G), Not Applicable (F24T0267G), Not Applicable (G34T3119G), Not Applicable (M34T4471G), Not Applicable (C34T2396G), Not Applicable (L24T1310G), Not Applicable (J34T3770G), Not Applicable (J24T0859G), Not Applicable (M34T4470G), Not Applicable (E34T2748G), Not Applicable (E34T2750G), Not Applicable (E34T2757G), Not Applicable (E34T2763G), Not Applicable (E34T2764G), Not Applicable (E34T2773G), Not Applicable (E34T2781G), Not Applicable (E34T2783G), Not Applicable (F34T2840G), Not Applicable (F34T2934G), Not Applicable (F34T2936G), Not Applicable (F34T2937G), Not Applicable (F34T2939G), Not Applicable (F34T2952G), Not Applicable (F34T2953G), Not Applicable (F34T2954G), Not Applicable (F34T2955G), Not Applicable (F34T2956G), Not Applicable (G34T3138G), Not Applicable (G34T3145G), Not Applicable (M34T4469G), Not Applicable (A34T1806G), Not Applicable (A34T1807G), Not Applicable (A34T1810G), Not Applicable (C44T4924G), Not Applicable (D34T2436G), Not Applicable (L24T1351G), Not Applicable (L24T1358G), Not Applicable (L24T1365G), Not Applicable (M34T4483G), Not Applicable (J34T3785G), Not Applicable (J24T0892G), Not Applicable (J34T3772G), Not Applicable (J34T3776G), Not Applicable (J34T3777G), Not Applicable (F34T3040G), Not Applicable (G34T3127G), Not Applicable (G34T3128G), Not Applicable (G34T3132G), Not Applicable (K24T1044G), Not Applicable (M24T1606G), Not Applicable (M24T1611G), Not Applicable (M34T4235G), Not Applicable (M34T4409G), Not Applicable (M34T4410G), Not Applicable (M34T4411G), Not Applicable (M34T4456G), Not Applicable (B44T4841G), Not Applicable (E34T2778G), Not Applicable (E34T2779G), Not Applicable (F24T0189G), Not Applicable (F24T0196G), Not Applicable (F34T2873G), Not Applicable (F34T2895G), Not Applicable (J24T0695G), Not Applicable (J24T0881G), Not Applicable (G24T0536G), Not Applicable (K34T3877G), Not Applicable (K34T3882G), Not Applicable (L34T4111G), Not Applicable (M24T1533G), Not Applicable (M34T4322G), Not Applicable (M34T4524G), Not Applicable (F24T0071G), Not Applicable (G34T3139G), Not Applicable (E34T2685G), Not Applicable (E34T2688G), Not Applicable (E34T2689G), Not Applicable (E34T2691G), Not Applicable (E34T2693G), Not Applicable (E34T2715G), Not Applicable (F34T2086G), Not Applicable (F34T2976G), Not Applicable (F34T2986G), Not Applicable (F34T2987G), Not Applicable (F34T2992G), Not Applicable (F34T2993G), Not Applicable (F34T2994G), Not Applicable (F34T2995G), Not Applicable (F34T3022G), Not Applicable (M24T1693G), Not Applicable (A34T1768G), Not Applicable (C34T2352G), Not Applicable (C44T4983G), Not Applicable (D34T2547G), Not Applicable (F24T0287G), Not Applicable (J24T0860G), Not Applicable (K24T1246G), Not Applicable (L24T1281G), Not Applicable (M24T1456G), Not Applicable (M24T1471G), Not Applicable (M24T1506G), Not Applicable (M24T1525G), Not Applicable (M24T1526G), Not Applicable (M24T1618G), Not Applicable (M24T1636G), Not Applicable (M24T1674G), Not Applicable (M24T1675G), Not Applicable (F24T0131G), Not Applicable (F24T0135G), Not Applicable (F24T0218G), Not Applicable (F24T0219G), Not Applicable (F24T0220G), Not Applicable (J34T3796G), Not Applicable (B34T2097G), Not Applicable (B44T4851G), Not Applicable (G24T0391G), Not Applicable (G24T0392G), Not Applicable (D34T2628G), Not Applicable (B34T2201G), Not Applicable (B34T2056G), Not Applicable (B34T2204G), Not Applicable (C34T2216G), Not Applicable (F34T3032G), Not Applicable (J34T3520G), Not Applicable (J34T3546G), Not Applicable (J34T3558G), Not Applicable (J34T3559G), Not Applicable (J34T3560G), Not Applicable (J34T3561G), Not Applicable (M34T4420G), Not Applicable (K24T1210G), Not Applicable (K24T1211G), Not Applicable (K24T1222G), Not Applicable (C34T2409G), Not Applicable (J34T3585G), Not Applicable (J34T3595G), Not Applicable (J34T3783G), Not Applicable (M34T4260G), Not Applicable (M34T4269G), Not Applicable (J34T3764G), Not Applicable (J34T3765G), Not Applicable (F34T2857G), Not Applicable (K24T1029G), Not Applicable (K24T1273G), Not Applicable (B34T2004G), Not Applicable (J34T3575G), Not Applicable (J34T3787G), Not Applicable (J34T3811G), Not Applicable (J34T3813G), Not Applicable (H24T0598G), Not Applicable (H24T0599G), Not Applicable (L34T4100G), Not Applicable (M24T1423G), Not Applicable (M24T1451G), Not Applicable (M24T1458G), Not Applicable (B44T4807G), Not Applicable (C34T2358G), Not Applicable (F24T0304G), Not Applicable (F24T0305G), Not Applicable (F24T0306G), Not Applicable (F24T0307G), Not Applicable (D34T2434G), Not Applicable (D34T2439G), Not Applicable (D34T2565G), Not Applicable (K24T1099G), Not Applicable (K34T3937G), Not Applicable (M34T4461G), Not Applicable (M34T4535G), Not Applicable (M34T4537G), Not Applicable (B34T2058G), Not Applicable (B34T2060G), Not Applicable (B34T2062G), Not Applicable (J24T0801G), Not Applicable (L34T4013G), Not Applicable (L34T4046G), Not Applicable (A44T4685G), Not Applicable (C34T2305G), Not Applicable (D34T2455G), Not Applicable (F24T0019G), Not Applicable (M24T1598G), Not Applicable (M24T1644G), Not Applicable (M24T1645G), Not Applicable (G24T0528G), Not Applicable (G24T0531G), Not Applicable (J24T0714G), Not Applicable (J24T0715G), Not Applicable (J24T0741G), Not Applicable (J24T0742G), Not Applicable (J34T3803G), Not Applicable (H34T3423G), Not Applicable (M24T1713G), Not Applicable (F34T2980G), Not Applicable (F34T3026G), Not Applicable (F34T3042G), Not Applicable (F24T0055G), Not Applicable (F34T3025G), Not Applicable (F34T3047G), Not Applicable (H34T3442G), Not Applicable (B34T1980G), Not Applicable (J24T0732G), Not Applicable (A34T1925G), Not Applicable (K24T1229G), Not Applicable (L24T1399G), Not Applicable (B34T2089G), Not Applicable (M24T1439G), Not Applicable (M24T1446G), Not Applicable (M24T1448G), Not Applicable (M24T1449G), Not Applicable (M24T1450G), Not Applicable (M24T1452G), Not Applicable (M24T1453G), Not Applicable (M24T1454G), Not Applicable (M24T1464G), Not Applicable (F34T3009G), Not Applicable (F34T3011G), Not Applicable (F34T3013G), Not Applicable (F34T3015G), Not Applicable (F34T3016G), Not Applicable (J34T3690G), Not Applicable (J34T3691G), Not Applicable (J34T3692G), Not Applicable (L24T1366G), Not Applicable (L24T1367G), Not Applicable (L24T1368G), Not Applicable (L24T1369G), Not Applicable (C34T2383G), Not Applicable (C34T2386G), Not Applicable (C34T2387G), Not Applicable (C34T2388G), Not Applicable (C34T2393G), Not Applicable (C34T2394G), Not Applicable (F24T0199G), Not Applicable (F24T0200G), Not Applicable (F24T0202G), Not Applicable (K24T1149G), Not Applicable (K24T1253G), Not Applicable (B34T1963G), Not Applicable (F24T0238G), Not Applicable (J24T0800G), Not Applicable (C34T2315G), Not Applicable (C34T2318G), Not Applicable (J24T0693G), Not Applicable (J24T0694G), Not Applicable (J24T0699G), Not Applicable (J24T0802G), Not Applicable (J24T0803G), Not Applicable (J24T0845G), Not Applicable (B34T1982G), Not Applicable (L24T1297G), Not Applicable (A44T4595G), Not Applicable (B34T2007G), Not Applicable (B34T2010G), Not Applicable (B34T2011G), Not Applicable (B34T2012G), Not Applicable (B34T2013G), Not Applicable (B34T2015G), Not Applicable (B34T2017G), Not Applicable (M24T1411G), Not Applicable (M24T1414G), Not Applicable (M24T1415G), Not Applicable (M24T1416G), Not Applicable (M24T1420G), Not Applicable (M24T1431G), Not Applicable (M34T4491G), Not Applicable (B34T1966G), Not Applicable (B34T1971G), Not Applicable (B34T2150G), Not Applicable (B44T4840G), Not Applicable (B44T4849G), Not Applicable (B44T5023G), Not Applicable (C34T2302G), Not Applicable (D44T5023G), Not Applicable (F24T0215G), Not Applicable (B34T2199G), Not Applicable (B34T2200G), Not Applicable (A44T4710G), Not Applicable (G34T3180G), Not Applicable (G34T3181G), Not Applicable (G34T3182G), Not Applicable (A44T4638G), Not Applicable (A44T4640G), Not Applicable (B44T4773G), Not Applicable (D34T2579G), Not Applicable (D34T2582G), Not Applicable (D34T2635G), Not Applicable (D44T5037G), Not Applicable (F34T3024G), Not Applicable (F34T3029G), Not Applicable (F34T3033G), Not Applicable (F34T3044G), Not Applicable (G24T0340G), Not Applicable (G24T0341G), Not Applicable (G24T0342G), Not Applicable (G24T0344G), Not Applicable (G24T0345G), Not Applicable (G24T0346G), Not Applicable (G24T0350G), Not Applicable (G24T0351G), Not Applicable (G24T0352G), Not Applicable (G24T0353G), Not Applicable (G24T0354G), Not Applicable (G24T0355G), Not Applicable (G24T0357G), Not Applicable (G24T0358G), Not Applicable (G24T0359G), Not Applicable (G24T0381G), Not Applicable (G24T0382G), Not Applicable (G24T0454G), Not Applicable (G24T0461G), Not Applicable (G24T0466G), Not Applicable (G24T0467G), Not Applicable (G24T0469G), Not Applicable (G24T0471G), Not Applicable (G24T0472G), Not Applicable (G24T0473G), Not Applicable (G24T0496G), Not Applicable (G24T0497G), Not Applicable (G24T0498G), Not Applicable (G34T3068G), Not Applicable (G34T3071G), Not Applicable (G34T3072G), Not Applicable (G34T3073G), Not Applicable (G34T3074G), Not Applicable (G34T3075G), Not Applicable (G34T3076G), Not Applicable (G34T3078G), Not Applicable (G34T3079G), Not Applicable (G34T3080G), Not Applicable (G34T3081G), Not Applicable (G34T3082G), Not Applicable (G34T3083G), Not Applicable (G34T3085G), Not Applicable (G34T3087G), Not Applicable (G34T3088G), Not Applicable (G34T3089G), Not Applicable (G34T3091G), Not Applicable (G34T3117G), Not Applicable (H34T3454G), Not Applicable (J24T0733G), Not Applicable (J34T3562G), Not Applicable (K24T1208G), Not Applicable (K24T1219G), Not Applicable (B34T1992G), Not Applicable (B34T1994G), Not Applicable (B34T2053G), Not Applicable (B34T2054G), Not Applicable (B34T2059G), Not Applicable (G34T3092G), Not Applicable (G34T3174G), Not Applicable (G24T0515G), Not Applicable (K24T1223G), Not Applicable (L34T4012G), Not Applicable (K34T3919G), Not Applicable (D44T5029G), Not Applicable (J24T0697G), Not Applicable (J24T0897G), Not Applicable (J24T0901G), Not Applicable (K24T1227G), Not Applicable (M34T4512G), Not Applicable (H34T3318G), Not Applicable (C44T4957G), Not Applicable (D34T2497G), Not Applicable (G34T3172G), Not Applicable (G34T3177G), Not Applicable (J24T0804G), Not Applicable (J24T0930G), Not Applicable (J24T0773G), Not Applicable (J24T0794G), Not Applicable (J24T0653G), Not Applicable (G24T0373G), Not Applicable (J34T3702G), Not Applicable (J34T3706G), Not Applicable (J24T0727G), Not Applicable (A44T4567G), Not Applicable (B44T4788G), Not Applicable (B44T4808G), Not Applicable (B44T4813G), Not Applicable (C34T2243G), Not Applicable (C44T4943G), Not Applicable (D34T2504G), Not Applicable (D34T2508G), Not Applicable (D34T2512G), Not Applicable (D44T5028G), Not Applicable (E04TMLJ1B3), Not Applicable (F34T2896G), Not Applicable (G24T0552G), Not Applicable (G24T0562G), Not Applicable (G34T3155G), Not Applicable (H34T3347G), Not Applicable (H34T3368G), Not Applicable (J24T0757G), Not Applicable (J24T0853G), Not Applicable (J24T0856G), Not Applicable (J24T0891G), Not Applicable (J24T0894G), Not Applicable (J34T3805G), Not Applicable (J34T3808G), Not Applicable (K24T1009G), Not Applicable (K24T1154G), Not Applicable (K34T3884G), Not Applicable (L34T4139G), Not Applicable (L34T4140G), Not Applicable (L34T4141G), Not Applicable (L34T4143G), Not Applicable (M24T1607G), Not Applicable (M24T1610G), Not Applicable (M34T4229G), Not Applicable (M34T4511G), Not Applicable (K34T3932G), Not Applicable (D44T5019G), Not Applicable (D44T5020G), Not Applicable (J34T3515G), Not Applicable (J34T3518G), Not Applicable (J34T3526G), Not Applicable (J34T3528G), Not Applicable (J34T3529G), Not Applicable (J34T3534G), Not Applicable (A34T1906G), Not Applicable (A34T1912G), Not Applicable (B34T2132G), Not Applicable (J34T3661G), Not Applicable (J34T3663G), Not Applicable (B34T2044G), Not Applicable (B34T2112G), Not Applicable (B34T2113G), Not Applicable (C34T2255G), Not Applicable (C34T2257G), Not Applicable (C34T2260G), Not Applicable (C34T2264G), Not Applicable (C34T2270G), Not Applicable (C34T2278G), Not Applicable (D34T2462G), Not Applicable (D34T2584G), Not Applicable (D34T2586G), Not Applicable (D34T2612G), Not Applicable (D34T2617G), Not Applicable (K34T3879G), Not Applicable (C44T4959G), Not Applicable (F34T2922G), Not Applicable (M34T4495G), Not Applicable (M34T4496G), Not Applicable (M34T4375G), Not Applicable (M34T4457G), Not Applicable (K34T3922G), Not Applicable (D34T2469G), Not Applicable (D34T2478G), Not Applicable (F34T2912G), Not Applicable (F34T2916G), Not Applicable (F34T2917G), Not Applicable (F34T2918G), Not Applicable (M34T4440G), Not Applicable (M34T4441G), Not Applicable (M34T4447G), Not Applicable (M34T4464G), Not Applicable (M34T4465G), Not Applicable (J34T3536G), Not Applicable (J34T3537G), Not Applicable (J34T3538G), Not Applicable (J34T3539G), Not Applicable (J34T3541G), Not Applicable (J34T3543G), Not Applicable (J34T3545G), Not Applicable (J34T3694G), Not Applicable (J34T3695G), Not Applicable (J34T3696G), Not Applicable (J34T3768G), Not Applicable (L24T9KWG), Not Applicable (L24TAJYG), Not Applicable (L24TDZJG), Not Applicable (J34T3564G), Not Applicable (J34T3566G), Not Applicable (K24T1153G), Not Applicable (K24T1083G), Not Applicable (H34T3326G), Not Applicable (G24T0464G), Not Applicable (G24T0575G), Not Applicable (G34T3095G), Not Applicable (A34T1779G), Not Applicable (A34T1864G), Not Applicable (B34T2080G), Not Applicable (D34T2499G), Not Applicable (F24T0121G), Not Applicable (L34T4038G), Not Applicable (L34T4145G), Not Applicable (A44T4566G), Not Applicable (J34T3678G), Not Applicable (J34T3709G), Not Applicable (J34T3710G), Not Applicable (J34T3711G), Not Applicable (J34T3712G), Not Applicable (J34T3713G), Not Applicable (J34T3714G), Not Applicable (J34T3715G), Not Applicable (J34T3716G), Not Applicable (J34T3719G), Not Applicable (L34T4147G), Not Applicable (M34T4202G), Not Applicable (M34T4205G), Not Applicable (M34T4207G), Not Applicable (M34T4209G), Not Applicable (M34T4210G), Not Applicable (M34T4211G), Not Applicable (M34T4212G), Not Applicable (M34T4213G), Not Applicable (M34T4214G), Not Applicable (F34T2962G), Not Applicable (J34T3649G), Not Applicable (J34T3650G), Not Applicable (J34T3652G), Not Applicable (J34T3657G), Not Applicable (J34T3659G), Not Applicable (J34T3682G), Not Applicable (J34T3683G), Not Applicable (J34T3700G), Not Applicable (J34T3721G), Not Applicable (J34T3722G), Not Applicable (J34T3732G), Not Applicable (J34T3749G), Not Applicable (J34T3750G), Not Applicable (J34T3756G), Not Applicable (M34T4193G), Not Applicable (M34T4195G), Not Applicable (M34T4196G), Not Applicable (M34T4197G), Not Applicable (M34T4198G), Not Applicable (M34T4199G), Not Applicable (M34T4200G), Not Applicable (M34T4201G), Not Applicable (M34T4203G), Not Applicable (M34T4204G), Not Applicable (M34T4208G), Not Applicable (M34T4251G), Not Applicable (M34T4313G), Not Applicable (M34T4317G), Not Applicable (M34T4431G), Not Applicable (M34T4433G), Not Applicable (E34T2705G), Not Applicable (J24T0665G), Not Applicable (J24T0677G), Not Applicable (J24T0740G), Not Applicable (F24T0317G), Not Applicable (G24T0474G), Not Applicable (G24T0480G), Not Applicable (B44T4850G), Not Applicable (B44T4852G), Not Applicable (B44T4857G), Not Applicable (B44T4859G), Not Applicable (B44T4861G), Not Applicable (B44T4864G), Not Applicable (B44T4867G), Not Applicable (B44T4870G), Not Applicable (B44T4871G), Not Applicable (B44T4873G), Not Applicable (G24T0567G), Not Applicable (G24T0573G), Not Applicable (J24T0963G), Not Applicable (K24T1056G), Not Applicable (K24T1057G), Not Applicable (K24T1144G), Not Applicable (B34T2037G), Not Applicable (B34T2039G), Not Applicable (A44T4587G), Not Applicable (B44T4814G), Not Applicable (F34T2946G), Not Applicable (F34T2975G), Not Applicable (J34T3658G), Not Applicable (J34T3701G), Not Applicable (J34T3703G), Not Applicable (J34T3704G), Not Applicable (J34T3708G), Not Applicable (J34T3744G), Not Applicable (J34T3746G), Not Applicable (M34T4248G), Not Applicable (M34T4278G), Not Applicable (M34T4279G), Not Applicable (M34T4280G), Not Applicable (M34T4283G), Not Applicable (M34T4284G), Not Applicable (M34T4285G), Not Applicable (M34T4286G), Not Applicable (M34T4287G), Not Applicable (J34T3675G), Not Applicable (J34T3677G), Not Applicable (L34T4146G), Not Applicable (L34T4167G), Not Applicable (L34T4173G), Not Applicable (E34T2741G), Not Applicable (J34T3625G), Not Applicable (J34T3626G), Not Applicable (J34T3627G), Not Applicable (J34T3629G), Not Applicable (M34T4254G), Not Applicable (B44T4756G), Not Applicable (B34T2115G), Not Applicable (C34T2284G), Not Applicable (L24T1280G), Not Applicable (D34T2468G), Not Applicable (M34T4239G), Not Applicable (K24T1097G), Not Applicable (H34T3480G), Not Applicable (H34T3300G), Not Applicable (H34T3306G), Not Applicable (H34T3310G), Not Applicable (B44T4823G), Not Applicable (G24T0386G), Not Applicable (G24T0459G), Not Applicable (G24T0460G), Not Applicable (H34T3350G), Not Applicable (H34T3351G), Not Applicable (H34T3399G), Not Applicable (H34T3438G), Not Applicable (H34T3439G), Not Applicable (H34T3440G), Not Applicable (H34T3443G), Not Applicable (H34T3444G), Not Applicable (H34T3445G), Not Applicable (H34T3446G), Not Applicable (H34T3447G), Not Applicable (H34T3448G), Not Applicable (H34T3449G), Not Applicable (H34T3476G), Not Applicable (H34T3478G), Not Applicable (H34T3484G), Not Applicable (M24T1680G), Not Applicable (M24T1717G), Not Applicable (M24T1727G), Not Applicable (B34T2035G), Not Applicable (G34T3229G), Not Applicable (G34T3230G), Not Applicable (H34T3299G), Not Applicable (H34T3307G), Not Applicable (H34T3366G), Not Applicable (G24T0519G), Not Applicable (G24T0577G), Not Applicable (G24T0580G), Not Applicable (B34T1962G), Not Applicable (A44T4690G), Not Applicable (A44T4692G), Not Applicable (A44T4693G), Not Applicable (A44T4702G), Not Applicable (A44T4703G), Not Applicable (A44T4706G), Not Applicable (B44T4853G), Not Applicable (B44T4854G), Not Applicable (B44T4855G), Not Applicable (B44T4856G), Not Applicable (B44T4858G), Not Applicable (B44T4860G), Not Applicable (D44T5015G), Not Applicable (H34T3463G), Not Applicable (M24T1746G), Not Applicable (D34T2614G), Not Applicable (J34T3563G), Not Applicable (J34T3565G), Not Applicable (M24T1580G), Not Applicable (J34T3753G), Not Applicable (A44T4561G), Not Applicable (C34T2330G), Not Applicable (C34T2364G), Not Applicable (C34T2367G), Not Applicable (H34T3261G), Not Applicable (M34T4301G), Not Applicable (M34T4302G), Not Applicable (M34T4303G), Not Applicable (M34T4304G), Not Applicable (M34T4305G), Not Applicable (M34T4306G), Not Applicable (M34T4307G), Not Applicable (M34T4309G), Not Applicable (M34T4310G), Not Applicable (M34T4311G), Not Applicable (M34T4349G), Not Applicable (M34T4351G), Not Applicable (M34T4352G), Not Applicable (B34T2206G), Not Applicable (C34T2382G), Not Applicable (C34T2401G), Not Applicable (D34T2597G), Not Applicable (B34T2031G), Not Applicable (B44T4907G), Not Applicable (L34T4009G), Not Applicable (L34T4014G), Not Applicable (L34T4015G), Not Applicable (L34T4018G), Not Applicable (L34T4019G), Not Applicable (K24T1261G), Not Applicable (L24T1296G), Not Applicable (L24T1317G), Not Applicable (L24T1318G), Not Applicable (L24T1320G), Not Applicable (L24T1321G), Not Applicable (M24T1520G), Not Applicable (M24T1524G), Not Applicable (B34T1941G), Not Applicable (B34T1949G), Not Applicable (B34T2104G), Not Applicable (E34T2684G), Not Applicable (J24T0749G), Not Applicable (J24T0754G), Not Applicable (J24T0755G), Not Applicable (J24T0756G), Not Applicable (J24T0785G), Not Applicable (K34T3855G), Not Applicable (M24T1696G), Not Applicable (M24T1697G), Not Applicable (M24T1698G), Not Applicable (M24T1699G), Not Applicable (M34T4522G), Not Applicable (A44T4544G), Not Applicable (M34T4536G), Not Applicable (A34T1804G), Not Applicable (A34T1846G), Not Applicable (B34T2024G), Not Applicable (B34T2026G), Not Applicable (F24T0283G), Not Applicable (F24T0312G), Not Applicable (F24T0313G), Not Applicable (F34T2947G), Not Applicable (F34T2981G), Not Applicable (F34T3051G), Not Applicable (H34T3483G), Not Applicable (J34T3707G), Not Applicable (D34T2441G), Not Applicable (G24T0481G), Not Applicable (J34T3581G), Not Applicable (C34T2366G), Not Applicable (J24T0779G), Not Applicable (L24T1359G), Not Applicable (G24T0576G), Not Applicable (G24T0402G), Not Applicable (B44T4760G), Not Applicable (B44T4763G), Not Applicable (B44T4770G), Not Applicable (G24T0559G), Not Applicable (G24T0508G), Not Applicable (C44T4945G), Not Applicable (M24T1536G), Not Applicable (M24T1664G), Not Applicable (C34T2344G), Not Applicable (C34T2379G), Not Applicable (C34T2392G), Not Applicable (G24T0574G), Not Applicable (H34T3280G), Not Applicable (L34T4101G), Not Applicable (M34T4194G), Not Applicable (M34T4466G), Not Applicable (M34T4517G), Not Applicable (M34T4518G), Not Applicable (M34T4530G), Not Applicable (M34T4532G), Not Applicable (B34T1953G), Not Applicable (B34T1978G), Not Applicable (B34T1990G), Not Applicable (B34T2001G), Not Applicable (D34T2556G), Not Applicable (J34T3603G), Not Applicable (A34T1843G), Not Applicable (F24T0280G), Not Applicable (F24T0281G), Not Applicable (F24T0282G), Not Applicable (F24T0319G), Not Applicable (F24T0320G), Not Applicable (F24T0321G), Not Applicable (K24T1017G), Not Applicable (M34T4538G), Not Applicable (J24T0701G), Not Applicable (B44T4862G), Not Applicable (B44T4872G), Not Applicable (E34T2758G), Not Applicable (E34T2780G), Not Applicable (E34T2793G), Not Applicable (G34T3186G), Not Applicable (M34T4534G), Not Applicable (J24T0837G), Not Applicable (K24T1010G), Not Applicable (D34T2476G), Not Applicable (D34T2477G), Not Applicable (A44T4635G), Not Applicable (H34T3357G), Not Applicable (H34T3332G), Not Applicable (K24T1257G), Not Applicable (J24T0906G), Not Applicable (J24T0925G), Not Applicable (M24T1436G), Not Applicable (C44T4932G), Not Applicable (C44T4934G), Not Applicable (A34T1770G), Not Applicable (C34T2362G), Not Applicable (A44T4542G), Not Applicable (D34T2484G), Not Applicable (F24T0246G), Not Applicable (G34T3090G), Not Applicable (J24T0899G), Not Applicable (K34T3832G), Not Applicable (L34T4010G), Not Applicable (L34T4024G), Not Applicable (L34T4030G), Not Applicable (L34T4042G), Not Applicable (L34T4043G), Not Applicable (L34T4045G), Not Applicable (L34T4052G), Not Applicable (M34T4336G), Not Applicable (M34T4341G), Not Applicable (D34T2496G), Not Applicable (M34T4423G), Not Applicable (K34T3921G), Not Applicable (K34T3924G), Not Applicable (K34T3950G), Not Applicable (M24T1582G), Not Applicable (M24T1588G), Not Applicable (M24T1593G), Not Applicable (M24T1597G), Not Applicable (M24T1601G), Not Applicable (M24T1641G), Not Applicable (M24T1642G), Not Applicable (A34T1872G), Not Applicable (B44T4797G), Not Applicable (D34T2562G), Not Applicable (F24T0277G), Not Applicable (F34T2855G), Not Applicable (G24T0449G), Not Applicable (G34T3141G), Not Applicable (G34T3232G), Not Applicable (G34T3233G), Not Applicable (G34T3234G), Not Applicable (G34T3237G), Not Applicable (G34T3238G), Not Applicable (G34T3245G), Not Applicable (G34T3246G), Not Applicable (G34T3248G), Not Applicable (G34T3250G), Not Applicable (H34T3283G), Not Applicable (H34T3312G), Not Applicable (H34T3313G), Not Applicable (H34T3314G), Not Applicable (H34T3321G), Not Applicable (H34T3322G), Not Applicable (H34T3331G), Not Applicable (H34T3342G), Not Applicable (H34T3344G), Not Applicable (H34T3360G), Not Applicable (H34T3364G), Not Applicable (H34T3367G), Not Applicable (H34T3411G), Not Applic able (H34T3413G), Not Applicable (H34T3416G), Not Applicable (H34T3421G), Not Applicable (H34T3427G), Not Applicable (H34T3428G), Not Applicable (H34T3429G), Not Applicable (H34T3430G), Not Applicable (H34T3431G), Not Applicable (H34T3432G), Not Applicable (H34T3433G), Not Applicable (H34T3436G), Not Applicable (H34T3450G), Not Applicable (H34T3451G), Not Applicable (H34T3452G), Not Applicable (H34T3453G), Not Applicable (H34T3455G), Not Applicable (H34T3456G), Not Applicable (H34T3457G), Not Applicable (H34T3458G), Not Applicable (H34T3461G), Not Applicable (H34T3489G), Not Applicable (J24T0613G), Not Applicable (J24T0614G), Not Applicable (J24T0615G), Not Applicable (J24T0616G), Not Applicable (J24T0848G), Not Applicable (J24T0850G), Not Applicable (J24T0868G), Not Applicable (J24T0869G), Not Applicable (J34T3586G), Not Applicable (M34T4256G), Not Applicable (M34T4484G), Not Applicable (M34T4485G), Not Applicable (K24T1212G), Not Applicable (L24TDWSG), Not Applicable (G34T3135G), Not Applicable (K24T1040G), Not Applicable (L34T4003G), Not Applicable (M24T1460G), Not Applicable (M24T1491G), Not Applicable (M24T1499G), Not Applicable (M24T1501G), Not Applicable (M24T1502G), Not Applicable (M24T1605G), Not Applicable (M24T1634G), Not Applicable (M24T1637G), Not Applicable (M24T1638G), Not Applicable (F24T0120G), Not Applicable (K24T1121G), Not Applicable (J24T0626G), Not Applicable (J24T0628G), Not Applicable (J24T0629G), Not Applicable (K24T1262G), Not Applicable (K34T3835G), Not Applicable (K34T3836G), Not Applicable (K34T3842G), Not Applicable (L24TE71G), Not Applicable (M24T1554G), Not Applicable (M34T4395G), Not Applicable (K24T1053G), Not Applicable (M24T1535G), Not Applicable (M24T1576G), Not Applicable (B34T1943G), Not Applicable (F24T0223G), Not Applicable (G24T0516G), Not Applicable (A44T4596G), Not Applicable (B34T2022G), Not Applicable (B34T2028G), Not Applicable (F34T2832G), Not Applicable (F34T2909G), Not Applicable (F34T2982G), Not Applicable (K34T3987G), Not Applicable (K34T3988G), Not Applicable (K34T3992G), Not Applicable (K34T3994G), Not Applicable (M24T1438G), Not Applicable (E34T2772G), Not Applicable (M24T1655G), Not Applicable (K34T3822G), Not Applicable (K34T3830G), Not Applicable (B44T4811G), Not Applicable (B44T4812G), Not Applicable (B44T4832G), Not Applicable (B44T4833G), Not Applicable (L34T4132G), Not Applicable (L34T4133G), Not Applicable (K24T1159G), Not Applicable (A34T1873G), Not Applicable (A34T1880G), Not Applicable (E34T2655G), Not Applicable (F24T0216G), Not Applicable (J24T0664G), Not Applicable (J24T0825G), Not Applicable (J34T3655G), Not Applicable (J34T3665G), Not Applicable (K34T3916G), Not Applicable (L24T1355G), Not Applicable (L24T1364G), Not Applicable (L34T4073G), Not Applicable (L34T4074G), Not Applicable (A44T4715G), Not Applicable (K34T3964G), Not Applicable (K34T3969G), Not Applicable (L34T4053G), Not Applicable (M34T4520G), Not Applicable (M34T4527G), Not Applicable (M34T4528G), Not Applicable (B44T4779G), Not Applicable (B44T4839G), Not Applicable (B44T4843G), Not Applicable (B44T4844G), Not Applicable (B44T4845G), Not Applicable (B44T4846G), Not Applicable (K24T1038G), Not Applicable (E34T2784G), Not Applicable (G24T0339G), Not Applicable (G24T0348G), Not Applicable (M24T1679G), Not Applicable (F24T41255G), Not Applicable (G24T0332G), Not Applicable (G24T0393G), Not Applicable (K24T1155G), Not Applicable (M24T1737G), Not Applicable (M24T1738G), Not Applicable (L34T4129G), Not Applicable (L34T4134G), Not Applicable (L34T4136G), Not Applicable (D44T5013G), Not Applicable (C34T2398G), Not Applicable (D34T2430G), Not Applicable (D34T2431G), Not Applicable (D34T2536G), Not Applicable (D34T2539G), Not Applicable (E34T2660G), Not Applicable (F34T2970G), Not Applicable (F34T2971G), Not Applicable (F34T2972G), Not Applicable (F34T2973G), Not Applicable (F34T2977G), Not Applicable (F34T2979G), Not Applicable (G24T0397G), Not Applicable (G24T0438G), Not Applicable (G34T3094G), Not Applicable (G34T3111G), Not Applicable (G34T3112G), Not Applicable (G34T3113G), Not Applicable (G34T3114G), Not Applicable (G34T3115G), Not Applicable (G34T3116G), Not Applicable (G34T3121G), Not Applicable (G34T3122G), Not Applicable (G34T3149G), Not Applicable (J24T0645G), Not Applicable (J24T0646G), Not Applicable (J24T0656G), Not Applicable (L24TALKG), Not Applicable (L34T4047G), Not Applicable (B34T2051G), Not Applicable (B34T2052G), Not Applicable (B34T2079G), Not Applicable (J24T0864G), Not Applicable (J34T3508G), Not Applicable (A34T1904G), Not Applicable (A34T1907G), Not Applicable (B34T1928G), Not Applicable (B34T1932G), Not Applicable (B34T1934G), Not Applicable (B34T1935G), Not Applicable (B34T1939G), Not Applicable (B44T4863G), Not Applicable (B44T4869G), Not Applicable (C34T2239G), Not Applicable (C34T2244G), Not Applicable (F34T2893G), Not Applicable (G24T0478G), Not Applicable (H34T3341G), Not Applicable (J34T3687G), Not Applicable (J34T3739G), Not Applicable (J34T3790G), Not Applicable (L34T4099G), Not Applicable (L34T4177G), Not Applicable (M24T1577G), Not Applicable (D34T2472G), Not Applicable (H34T3496G), Not Applicable (B44T4753G), Not Applicable (B44T4755G), Not Applicable (D34T2591G), Not Applicable (D34T2600G), Not Applicable (M24T1468G), Not Applicable (M24T1565G), Not Applicable (M24T1681G), Not Applicable (M24T1690G), Not Applicable (M34T4270G), Not Applicable (J24T0815G), Not Applicable (J24T0824G), Not Applicable (J34T3795G), Not Applicable (B34T2036G), Not Applicable (B44T4865G), Not Applicable (B44T4866G), Not Applicable (B44T4868G), Not Applicable (B44T4906G), Not Applicable (F34T3036G), Not Applicable (F34T3048G), Not Applicable (F34T3053G), Not Applicable (G34T3105G), Not Applicable (M24T1751G), Not Applicable (M24T1753G), Not Applicable (F34T3020G), Not Applicable (A34T1927G), Not Applicable (M24T1741G), Not Applicable (M24T1742G), Not Applicable (M24T1748G), Not Applicable (M24T1756G), Not Applicable (M24T1757G), Not Applicable (M24T1758G), Not Applicable (F24T0274G), Not Applicable (F24T0275G), Not Applicable (J24T0620G), Not Applicable (J34T3736G), Not Applicable (K24T1141G), Not Applicable (K24T1145G), Not Applicable (K24T1146G), Not Applicable (L34T4169G), Not Applicable (K34T3901G), Not Applicable (K34T3905G), Not Applicable (A34T1839G), Not Applicable (A34T1842G), Not Applicable (J24T0621G), Not Applicable (J24T0736G), Not Applicable (E34T2776G), Not Applicable (F34T2871G), Not Applicable (F34T3054G), Not Applicable (F34T3055G), Not Applicable (G34T3101G), Not Applicable (G34T3106G), Not Applicable (G34T3120G), Not Applicable (G34T3151G), Not Applicable (G34T3153G), Not Applicable (H34T3464G), Not Applicable (H34T3466G), Not Applicable (H34T3479G), Not Applicable (H34T3482G), Not Applicable (H34T3488G), Not Applicable (J34T3651G), Not Applicable (K24T1073G), Not Applicable (K24T1074G), Not Applicable (K24T1075G), Not Applicable (K24T1076G), Not Applicable (K24T1077G), Not Applicable (K24T1078G), Not Applicable (K24T1079G), Not Applicable (K24T1081G), Not Applicable (K24T1082G), Not Applicable (K24T1084G), Not Applicable (K24T1086G), Not Applicable (K34T3917G), Not Applicable (M34T4261G), Not Applicable (B34T2131G), Not Applicable (A44T4686G), Not Applicable (B34T1931G), Not Applicable (L24T1384G), Not Applicable (L24T1387G), Not Applicable (M34T4510G), Not Applicable (M34T4513G), Not Applicable (M34T4514G), Not Applicable (M34T4515G), Not Applicable (L24T1290G), Not Applicable (A34T1861G), Not Applicable (D34T2506G), Not Applicable (D34T2507G), Not Applicable (D34T2554G), Not Applicable (D34T2558G), Not Applicable (D34T2583G), Not Applicable (D34T2585G), Not Applicable (D34T2588G), Not Applicable (D34T2636G), Not Applicable (J24T0661G), Not Applicable (J24T0687G), Not Applicable (B34T2138G), Not Applicable (A44T4662G), Not Applicable (B44T4749G), Not Applicable (H34T3492G), Not Applicable (H34T3497G), Not Applicable (J34T3525G), Not Applicable (L24T1308G), Not Applicable (M24T1466G), Not Applicable (A44T4552G), Not Applicable (A44T4553G), Not Applicable (A34T1787G), Not Applicable (A34T1788G), Not Applicable (A34T1790G), Not Applicable (F24T0255G), Not Applicable (F24T0258G), Not Applicable (G24T0369G), Not Applicable (G24T0401G), Not Applicable (G24T0404G), Not Applicable (G24T0410G), Not Applicable (G24T0421G), Not Applicable (G24T0500G), Not Applicable (G24T0501G), Not Applicable (G24T0588G), Not Applicable (G24T0589G), Not Applicable (G24T0590G), Not Applicable (H34T3358G), Not Applicable (H34T3359G), Not Applicable (H34T3362G), Not Applicable (J24T0622G), Not Applicable (J24T0623G), Not Applicable (J24T0632G), Not Applicable (J24T0763G), Not Applicable (J24T0764G), Not Applicable (J24T0841G), Not Applicable (J24T0844G), Not Applicable (J24T0849G), Not Applicable (J24T0851G), Not Applicable (J24T0861G), Not Applicable (J24T0866G), Not Applicable (J24T0871G), Not Applicable (J24T0882G), Not Applicable (J24T0885G), Not Applicable (J24T0886G), Not Applicable (J34T3698G), Not Applicable (J34T3784G), Not Applicable (J34T3786G), Not Applicable (J34T3797G), Not Applicable (J34T3798G), Not Applicable (J34T3799G), Not Applicable (J34T3801G), Not Applicable (K24T1063G), Not Applicable (K24T1090G), Not Applicable (K24T1091G), Not Applicable (K24T1103G), Not Applicable (K24T1173G), Not Applicable (L24T1300G), Not Applicable (L24T1311G), Not Applicable (J24T0828G), Not Applicable (C34T2376G), Not Applicable (D34T2567G), Not Applicable (D34T2572G), Not Applicable (H34T3417G), Not Applicable (J24T0680G), Not Applicable (J24T0683G), Not Applicable (J24T0684G), Not Applicable (J24T0685G), Not Applicable (J24T0686G), Not Applicable (J24T0690G), Not Applicable (M24T1676G), Not Applicable (A44T4545G), Not Applicable (A44T4550G), Not Applicable (A44T4551G), Not Applicable (B34T2025G), Not Applicable (B44T4757G), Not Applicable (B44T4889G), Not Applicable (B44T4891G), Not Applicable (D34T2427G), Not Applicable (D34T2531G), Not Applicable (D34T2590G), Not Applicable (D34T2592G), Not Applicable (F24T0080G), Not Applicable (F24T0085G), Not Applicable (H34T3408G), Not Applicable (K24T1234G), Not Applicable (K34T3898G), Not Applicable (L24T1328G), Not Applicable (M34T4265G), Not Applicable (M34T4324G), Not Applicable (M34T4330G), Not Applicable (M34T4345G), Not Applicable (M34T4373G), Not Applicable (M34T4380G), Not Applicable (M34T4389G), Not Applicable (M34T4393G), Not Applicable (M34T4418G), Not Applicable (M34T4419G), Not Applicable (A34T1913G), Not Applicable (A44T4634G), Not Applicable (B44T4824G), Not Applicable (F24T0248G), Not Applicable (J24T0865G), Not Applicable (J24T0870G), Not Applicable (J24T0893G), Not Applicable (A34T1840G), Not Applicable (A34T1841G), Not Applicable (J34T3757G), Not Applicable (J34T3758G), Not Applicable (J34T3760G), Not Applicable (J34T3767G), Not Applicable (C44T4979G), Not Applicable (J24T0675G), Not Applicable (J24T0708G), Not Applicable (G34T3108G), Not Applicable (A44T4548G), Not Applicable (L34T4135G), Not Applicable (L34T4137G), Not Applicable (L34T4144G), Not Applicable (L34T4148G), Not Applicable (L34T4150G), Not Applicable (L34T4154G), Not Applicable (L34T4166G), Not Applicable (M34T4252G), Not Applicable (E34T2794G), Not Applicable (E34T2795G), Not Applicable (F34T2894G), Not Applicable (J24t0922G), Not Applicable (J24T0923G), Not Applicable (J24T0926G), Not Applicable (J24T0928G), Not Applicable (J34T3506G), Not Applicable (J34T3532G), Not Applicable (A44T4624G), Not Applicable (A44T4625G), Not Applicable (J34T3815G), Not Applicable (M24T1429G), Not Applicable (M24T1476G), Not Applicable (G34T3179G), Not Applicable (G34T3188G), Not Applicable (G34T3189G), Not Applicable (C34T2331G), Not Applicable (C34T2342G), Not Applicable (C34T2369G), Not Applicable (C34T2377G), Not Applicable (C34T2404G), Not Applicable (J34T3584G), Not Applicable (G24T0412G), Not Applicable (G24T0554G), Not Applicable (G24T0568G), Not Applicable (M24T1510G), Not Applicable (M24T1599G), Not Applicable (M24T1602G), Not Applicable (J34T3608G), Not Applicable (J34T3726G), Not Applicable (J34T3728G), Not Applicable (G24T0544G), Not Applicable (G24T0551G), Not Applicable (G24T0553G), Not Applicable (H34T3426G), Not Applicable (A44T4557G), Not Applicable (B34T2129G), Not Applicable (B34T2133G), Not Applicable (C34T2306G), Not Applicable (D34T2521G), Not Applicable (D34T2630G), Not Applicable (D34T2532G), Not Applicable (F24T0024G), Not Applicable (F24T0041G), Not Applicable (G24T0388G), Not Applicable (G24T0565G), Not Applicable (G24T0566G), Not Applicable (H34T3393G), Not Applicable (J34T3613G), Not Applicable (M34T4401G), Not Applicable (M34T4404G), Not Applicable (H34T3390G), Not Applicable (H34T3391G), Not Applicable (K34T3965G), Not Applicable (K34T3971G), Not Applicable (B34T2152G), Not Applicable (B34T2185G), Not Applicable (F24T0217G), Not Applicable (G24T0525G), Not Applicable (M24T1484G), Not Applicable (M24T1486G), Not Applicable (M24T1487G), Not Applicable (M24T1550G), Not Applicable (G24T0507G), Not Applicable (F24T0250G), Not Applicable (F24T0251G), Not Applicable (F24T0252G), Not Applicable (F34T2890G), Not Applicable (F34T2892G), Not Applicable (K34T3915G), Not Applicable (L34T4039G), Not Applicable (L34T4048G), Not Applicable (L34T4050G), Not Applicable (L34T4062G), Not Applicable (L34T4070G), Not Applicable (L34T4071G), Not Applicable (L34T4072G), Not Applicable (L34T4076G), Not Applicable (L34T4078G), Not Applicable (F24T0142G), Not Applicable (F24T0143G), Not Applicable (F24T0144G), Not Applicable (C34T2329G), Not Applicable (C34T2337G), Not Applicable (C34T2339G), Not Applicable (C34T2340G), Not Applicable (J24T0625G), Not Applicable (J24T0644G), Not Applicable (M34T4348G), Not Applicable (M34T4353G), Not Applicable (C34T2321G), Not Applicable (H34T3348G), Not Applicable (H34T3392G), Not Applicable (H34T3398G), Not Applicable (H34T3400G), Not Applicable (H34T3422G), Not Applicable (J34T3592G), Not Applicable (L24T1286G), Not Applicable (L24T1301G), Not Applicable (G24T0563G), Not Applicable (G24T0569G), Not Applicable (G24T0570G), Not Applicable (G24T0572G), Not Applicable (K34T3977G), Not Applicable (L34T3997G), Not Applicable (L34T3999G), Not Applicable (L34T4016G), Not Applicable (F24T0293G), Not Applicable (A44T4629G), Not Applicable (A44T4631G), Not Applicable (B44T4836G), Not Applicable (G24T0491G), Not Applicable (G24T0547G), Not Applicable (K24T1128G), Not Applicable (M24T1479G), Not Applicable (M24T1485G), Not Applicable (M34T4506G), Not Applicable (M34T4507G), Not Applicable (L34T4075G), Not Applicable (K24T1245G), Not Applicable (A44T4563G), Not Applicable (A44T4720G), Not Applicable (M24T1649G), Not Applicable (M24T1653G), Not Applicable (M24T1654G), Not Applicable (M24T1656G), Not Applicable (M24T1657G), Not Applicable (M24T1658G), Not Applicable (C34T2384G), Not Applicable (K24T1037G), Not Applicable (K24T1104G), Not Applicable (K24T1107G), Not Applicable (K24T1109G), Not Applicable (K24T1110G), Not Applicable (K24T1111G), Not Applicable (D34T2457G), Not Applicable (D34T2464G), Not Applicable (D34T2529G), Not Applicable (D34T2540G), Not Applicable (H24T0602G), Not Applicable (M24T1413G), Not Applicable (M24T1459G), Not Applicable (J34T3574G), Not Applicable (A44T4588G), Not Applicable (A44T4592G), Not Applicable (A44T4626G), Not Applicable (A44T4666G), Not Applicable (A44T4668G), Not Applicable (A44T4678G), Not Applicable (A44T4679G), Not Applicable (A44T4681G), Not Applicable (A44T4682G), Not Applicable (A44T4683G), Not Applicable (A44T4688G), Not Applicable (A44T4689G), Not Applicable (A44T4695G), Not Applicable (B34T2093G), Not Applicable (B44T4903G), Not Applicable (C34T2312G), Not Applicable (C34T2345G), Not Applicable (E34T2782G), Not Applicable (F24T0123G), Not Applicable (F24T0257G), Not Applicable (F34T2838G), Not Applicable (F34T2886G), Not Applicable (F34T2887G), Not Applicable (F34T2943G), Not Applicable (F34T2957G), Not Applicable (F34T2958G), Not Applicable (F34T2960G), Not Applicable (F34T2963G), Not Applicable (G34T3064G), Not Applicable (G34T3067G), Not Applicable (G34T3069G), Not Applicable (G34T3070G), Not Applicable (G34T3093G), Not Applicable (G34T3118G), Not Applicable (G34T3142G), Not Applicable (G34T3146G), Not Applicable (G34T3147G), Not Applicable (G34T3150G), Not Applicable (G34T3152G), Not Applicable (G34T3156G), Not Applicable (G34T3183G), Not Applicable (G34T3187G), Not Applicable (G34T3201G), Not Applicable (G34T3203G), Not Applicable (G34T3210G), Not Applicable (G34T3239G), Not Applicable (G34T3240G), Not Applicable (G34T3241G), Not Applicable (G34T3242G), Not Applicable (G34T3243G), Not Applicable (G34T3244G), Not Applicable (G34T3247G), Not Applicable (G34T3249G), Not Applicable (J34T3654G), Not Applicable (J34T3656G), Not Applicable (J34T3660G), Not Applicable (J34T3817G), Not Applicable (K34T3821G), Not Applicable (K34T3824G), Not Applicable (K34T3827G), Not Applicable (K34T3861G), Not Applicable (K34T3869G), Not Applicable (K34T3887G), Not Applicable (K34T3891G), Not Applicable (K34T3918G), Not Applicable (K34T3961G), Not Applicable (L34T4156G), Not Applicable (L34T4107G), Not Applicable (F34T2843G), Not Applicable (E34T2666G), Not Applicable (E34T2667G), Not Applicable (E34T2775G), Not Applicable (F34T2889G), Not Applicable (F24T0150G), Not Applicable (F24T0151G), Not Applicable (F24T0152G), Not Applicable (F24T0204G), Not Applicable (G24T0330G), Not Applicable (G24T0371G), Not Applicable (G24T0511G), Not Applicable (G24T0513G), Not Applicable (E34T2665G), Not Applicable (G24T0534G), Not Applicable (G24T0537G), Not Applicable (A34T1817G), Not Applicable (F24T0075G), Not Applicable (F24T0076G), Not Applicable (F24T0224G), Not Applicable (F24T0225G), Not Applicable (F24T0237G), Not Applicable (G24T0457G), Not Applicable (K24T1267G), Not Applicable (K24T1269G), Not Applicable (K24T1270G), Not Applicable (M24T1477G), Not Applicable (M24T1539G), Not Applicable (M24T1557G), Not Applicable (F24T0073G), Not Applicable (F24T0086G), Not Applicable (F34T3034G), Not Applicable (F34T3037G), Not Applicable (F34T3039G), Not Applicable (G34T3173G), Not Applicable (G34T3175G), Not Applicable (H34T3319G), Not Applicable (L24T1361G), Not Applicable (M24T1478G), Not Applicable (M24T1684G), Not Applicable (M24T1714G), Not Applicable (M34T4476G), Not Applicable (C44T4939G), Not Applicable (C44T4946G), Not Applicable (C44T4951G), Not Applicable (C44T4954G), Not Applicable (C44T4978G), Not Applicable (G24T0485G), Not Applicable (G24T0487G), Not Applicable (H34T3275G), Not Applicable (H34T3325G), Not Applicable (G34T3143G), Not Applicable (H34T3269G), Not Applicable (H34T3271G), Not Applicable (H34T3273G), Not Applicable (H34T3308G), Not Applicable (K24T1135G), Not Applicable (K24T1216G), Not Applicable (K24T1218G), Not Applicable (H34T3425G), Not Applicable (L34T4168G), Not Applicable (L34T4171G), Not Applicable (L34T4172G), Not Applicable (L34T4176G), Not Applicable (F24T0023G), Not Applicable (G34T3148G), Not Applicable (H34T3371G), Not Applicable (D34T2608G), Not Applicable (K24T1181G), Not Applicable (M34T4455G), Not Applicable (M34T4477G), Not Applicable (D34T2458G), Not Applicable (G34T3134G), Not Applicable (A44T4623G), Not Applicable (B44T4831G), Not Applicable (D44T5038G), Not Applicable (D44T5039G), Not Applicable (F24T0315G), Not Applicable (F34T2967G), Not Applicable (F34T3008G), Not Applicable (F34T3014G), Not Applicable (G24T0479G), Not Applicable (H34T3296G), Not Applicable (M24T1571G), Not Applicable (M34T4238G), Not Applicable (M34T4282G), Not Applicable (M34T4403G), Not Applicable (M34T4414G), Not Applicable (M34T4416G), Not Applicable (M34T4427G), Not Applicable (M34T4480G), Not Applicable (M34T4488G), Not Applicable (M34T4490G), Not Applicable (M34T4492G), Not Applicable (M34T4500G), Not Applicable (M34T4502G), Not Applicable (M34T4508G), Not Applicable (M34T4519G), Not Applicable (J24T0939G), Not Applicable (J24T0951G), Not Applicable (D44T5007G), Not Applicable (D44T5008G), Not Applicable (D44T5009G), Not Applicable (D44T5010G), Not Applicable (L24T1374G), Not Applicable (L24T1375G), Not Applicable (L24T1378G), Not Applicable (L24T1381G), Not Applicable (L24T1382G), Not Applicable (L24T1383G), Not Applicable (L24T1392G), Not Applicable (L24T1396G), Not Applicable (L24T1397G), Not Applicable (L24T1402G), Not Applicable (L24T1403G), Not Applicable (F24T0311G), Not Applicable (J24T0702G), Not Applicable (J24T0703G), Not Applicable (K24T1134G), Not Applicable (M34T4281G), Not Applicable (K34T3897G), Not Applicable (A44T4651G), Not Applicable (A44T4657G), Not Applicable (H34T3471G), Not Applicable (H34T3473G), Not Applicable (H34T3495G), Not Applicable (J34T3519G), Not Applicable (J34T3590G), Not Applicable (M34T3495G), Not Applicable (F24T0295G), Not Applicable (G24T0465G), Not Applicable (G34T3154G), Not Applicable (F34T2964G), Not Applicable (C34T2233G), Not Applicable (C34T2411G), Not Applicable (D34T2443G), Not Applicable (D34T2445G), Not Applicable (D34T2544G), Not Applicable (D34T2599G), Not Applicable (E34T2649G), Not Applicable (E34T2650G), Not Applicable (E34T2652G), Not Applicable (K24T1254G), Not Applicable (K24T1268G), Not Applicable (M24T1517G), Not Applicable (M24T1521G), Not Applicable (E34T2738G), Not Applicable (J34T3510G), Not Applicable (A44T4554G), Not Applicable (M24T1515G), Not Applicable (G24T0447G), Not Applicable (B34T1945G), Not Applicable (B34T2100G), Not Applicable (B34T2101G), Not Applicable (C44T4937G), Not Applicable (D34T2596G), Not Applicable (E34T2791G), Not Applicable (F24T0243G), Not Applicable (J34T3727G), Not Applicable (K34T3881G), Not Applicable (K34T3954G), Not Applicable (M24T1562G), Not Applicable (M24T1563G), Not Applicable (M24T1564G), Not Applicable (M24T1581G), Not Applicable (M24T1614G), Not Applicable (M24T1615G), Not Applicable (M24T1651G), Not Applicable (M24T1755G), Not Applicable (C44T4941G), Not Applicable (K24T1232G), Not Applicable (K34T3955G), Not Applicable (L24T1388G), Not Applicable (A44T4628G), Not Applicable (K24T1115G), Not Applicable (L34T4069G), Not Applicable (L34T4077G), Not Applicable (M34T4240G), Not Applicable (M34T4241G), Not Applicable (M34T4246G), Not Applicable (M34T4247G), Not Applicable (M34T4250G), Not Applicable (C34T2346G), Not Applicable (C34T2347G), Not Applicable (C34T2348G), Not Applicable (C34T2349G), Not Applicable (C34T2373G), Not Applicable (C34T2374G), Not Applicable (C34T2375G), Not Applicable (J24T0752G), Not Applicable (J24T0761G), Not Applicable (J24T0847G), Not Applicable (J24T0875G), Not Applicable (J34T3792G), Not Applicable (J34T3793G), Not Applicable (J34T3794G), Not Applicable (D34T2561G), Not Applicable (J34T3567G), Not Applicable (J34T3568G), Not Applicable (K24T1069G), Not Applicable (K24T1147G), Not Applicable (K24T1242G), Not Applicable (K24T1247G), Not Applicable (M24T1505G), Not Applicable (M24T1715G), Not Applicable (M24T1719G), Not Applicable (M24T1720G), Not Applicable (M24T1723G), Not Applicable (A44T4664G), Not Applicable (L34T4004G), Not Applicable (L34T4115G), Not Applicable (M34T4245G), Not Applicable (B34T2106G), Not Applicable (B34T2110G), Not Applicable (C34T2242G), Not Applicable (C34T2247G), Not Applicable (C34T2282G), Not Applicable (G24T0523G), Not Applicable (K24T1032G), Not Applicable (K24T1209G), Not Applicable (A44T4696G), Not Applicable (A44T4701G), Not Applicable (B34T1956G), Not Applicable (B34T2194G), Not Applicable (C34T2240G), Not Applicable (C44T4956G), Not Applicable (C44T4962G), Not Applicable (F24T0231G), Not Applicable (F24T0232G), Not Applicable (G24T0390G), Not Applicable (G24T0476G), Not Applicable (G24T0494G), Not Applicable (H34T3251G), Not Applicable (H34T3253G), Not Applicable (H34T3256G), Not Applicable (H34T3257G), Not Applicable (H34T3343G), Not Applicable (H34T3345G), Not Applicable (J24T0647G), Not Applicable (J24T0648G), Not Applicable (J24T0669G), Not Applicable (J24T0696G), Not Applicable (J24T0698G), Not Applicable (J24T0706G), Not Applicable (J34T3633G), Not Applicable (J34T3636G), Not Applicable (K24T0998G), Not Applicable (K24T1031G), Not Applicable (K24T1052G), Not Applicable (K24T1054G), Not Applicable (K24T1055G), Not Applicable (K24T1062G), Not Applicable (K24T1064G), Not Applicable (K24T1067G), Not Applicable (K34T3885G), Not Applicable (K34T3888G), Not Applicable (K34T3889G), Not Applicable (K34T3956G), Not Applicable (M24T1531G), Not Applicable (M24T1532G), Not Applicable (A34T1897G), Not Applicable (A34T1915G), Not Applicable (G24T0581G), Not Applicable (G24T0548G), Not Applicable (B34T2057G), Not Applicable (B34T2196G), Not Applicable (B34T2197G), Not Applicable (B44T4806G), Not Applicable (B44T4816G), Not Applicable (B44T4817G), Not Applicable (B44T4818G), Not Applicable (C34T2251G), Not Applicable (E34T2759G), Not Applicable (K34T3933G), Not Applicable (E34T2751G), Not Applicable (E34T2787G), Not Applicable (G34T3123G), Not Applicable (G34T3124G), Not Applicable (G34T3126G), Not Applicable (K34T3958G), Not Applicable (F34T2825G), Not Applicable (G34T3129G), Not Applicable (G34T3130G), Not Applicable (G34T3131G), Not Applicable (J34T3673G), Not Applicable (K24T1203G), Not Applicable (L34T4119G), Not Applicable (L34T4153G), Not Applicable (B44T4911G), Not Applicable (D44T5012G), Not Applicable (K34T3979G), Not Applicable (M34T4233G), Not Applicable (M34T4236G), Not Applicable (M34T4237G), Not Applicable (M34T4242G), Not Applicable (M34T4243G), Not Applicable (J24T0822G), Not Applicable (F24T0147G), Not Applicable (F34T2880G), Not Applicable (G34T3140G), Not Applicable (K34T3951G), Not Applicable (J24T0670G), Not Applicable (J24T0935G), Not Applicable (C44T4960G), Not Applicable (D34T2517G), Not Applicable (H34T3276G), Not Applicable (K24T1248G), Not Applicable (K24T1250G), Not Applicable (K24T1271G), Not Applicable (K34T3894G), Not Applicable (K34T3936G), Not Applicable (K34T3972G), Not Applicable (K34T3973G), Not Applicable (M24T1470G), Not Applicable (M24T1578G), Not Applicable (M24T1578GG), Not Applicable (M24T1579G), Not Applicable (F24T0175G), Not Applicable (F24T0182G), Not Applicable (F24T0183G), Not Applicable (K34T3826G), Not Applicable (K34T3831G), Not Applicable (M24T1461G), Not Applicable (F34T30506), Not Applicable (F34T3050G), Not Applicable (A44T4589G), Not Applicable (A44T4684G), Not Applicable (B34T2000G), Not Applicable (B44T4885G), Not Applicable (B44T4886G), Not Applicable (B44T4896G), Not Applicable (C34T2357G), Not Applicable (C34T2370G), Not Applicable (C34T2416G), Not Applicable (D34T2500G), Not Applicable (E34T2755G), Not Applicable (E34T2766G), Not Applicable (E34T2777G), Not Applicable (F24T0018G), Not Applicable (F24T0087G), Not Applicable (F24T0088G), Not Applicable (F24T0122G), Not Applicable (F34T2755G), Not Applicable (F34T2830G), Not Applicable (F34T2831G), Not Applicable (F34T2837G), Not Applicable (F34T2842G), Not Applicable (F34T2844G), Not Applicable (F34T2845G), Not Applicable (F34T2846G), Not Applicable (F34T2848G), Not Applicable (F34T2849G), Not Applicable (F34T2851G), Not Applicable (F34T2853G), Not Applicable (F34T2948G), Not Applicable (F34T2949G), Not Applicable (F34T2950G), Not Applicable (F34T2996G), Not Applicable (G24T0458G), Not Applicable (G24T0489G), Not Applicable (G24T0526G), Not Applicable (G24T0527G), Not Applicable (G34T3065G), Not Applicable (G34T3066G), Not Applicable (H34T3472G), Not Applicable (J24T0705G), Not Applicable (J24T0735G), Not Applicable (J24T0751G), Not Applicable (J24T0760G), Not Applicable (J24T0762G), Not Applicable (J24T0771G), Not Applicable (J24T0772G), Not Applicable (J24T0778G), Not Applicable (J24T0780G), Not Applicable (J24T0786G), Not Applicable (J24T0789G), Not Applicable (J24T0796G), Not Applicable (J24T0798G), Not Applicable (J24T0808G), Not Applicable (J24T0809G), Not Applicable (J24T0811G), Not Applicable (J24T0816G), Not Applicable (J24T0829G), Not Applicable (J24T0857G), Not Applicable (J24T0858G), Not Applicable (J24T0883G), Not Applicable (J24T0895G), Not Applicable (J24T0903G), Not Applicable (J24T0904G), Not Applicable (J24T0909G), Not Applicable (J24T0914G), Not Applicable (J24T0915G), Not Applicable (J24T0918G), Not Applicable (J24T0920G), Not Applicable (J24T0921G), Not Applicable (J24T0933G), Not Applicable (J34T3579G), Not Applicable (J34T3666G), Not Applicable (J34T3679G), Not Applicable (J34T3685G), Not Applicable (K24T1093G), Not Applicable (K24T1098G), Not Applicable (K24T1138G), Not Applicable (K24T1139G), Not Applicable (K24T1252G), Not Applicable (L24T1279G), Not Applicable (L24T1298G), Not Applicable (M24T1457G), Not Applicable (M24T1472G), Not Applicable (M24T1507G), Not Applicable (M24T1508G), Not Applicable (M24T1509G), Not Applicable (M24T1542G), Not Applicable (M24T1551G), Not Applicable (M24T1566G), Not Applicable (M24T1583G), Not Applicable (M24T1587G), Not Applicable (M24T1590G), Not Applicable (M24T1592G), Not Applicable (M24T1683G), Not Applicable (M24T1685G), Not Applicable (M24T1694G), Not Applicable (M24T1695G), Not Applicable (M24T1701G), Not Applicable (M24T1703G), Not Applicable (M24T1707G), Not Applicable (M24T1708G), Not Applicable (M24T1709G), Not Applicable (M24T1728G), Not Applicable (M34T4228G), Not Applicable (M34T4525G), Not Applicable (C34T2414G), Not Applicable (J24T0818G), Not Applicable (J34T3766G), Not Applicable (K24T1132G), Not Applicable (K24T1133G), Not Applicable (K34T3864G), Not Applicable (K34T3866G), Not Applicable (L24T1295G), Not Applicable (M24T1430G), Not Applicable (M34T4255G), Not Applicable (M34T4472G), Not Applicable (M34T4473G), Not Applicable (B44T4768G), Not Applicable (A34T1792G), Not Applicable (M24T1519G), Not Applicable (C44T4955G), Not Applicable (B44T4765G), Not Applicable (B44T4774G), Not Applicable (D44T5011G), Not Applicable (E34T2728G), Not Applicable (E34T2745G), Not Applicable (E34T2760G), Not Applicable (E34T2789G), Not Applicable (E34T2797G), Not Applicable (F34T2868G), Not Applicable (F34T2940G), Not Applicable (H34T3282G), Not Applicable (A34T1819G), Not Applicable (F34T3010G), Not Applicable (F34T3017G), Not Applicable (K24T1161G), Not Applicable (L24T1287G), Not Applicable (L24T1289G), Not Applicable (L24T1293G), Not Applicable (L24T1312G), Not Applicable (L24T1313G), Not Applicable (L34T4008G), Not Applicable (M24T1434G), Not Applicable (C44T4921G), Not Applicable (C44T4922G), Not Applicable (C44T4933G), Not Applicable (C44T4952G), Not Applicable (C44T4953G), Not Applicable (C44T4958G), Not Applicable (C34T2234G), Not Applicable (F24T0022G), Not Applicable (F24T0145G), Not Applicable (G24T0518G), Not Applicable (H34T3465G), Not Applicable (H34T3467G), Not Applicable (J34T3502G), Not Applicable (J34T3504G), Not Applicable (J34T3505G), Not Applicable (J34T3512G), Not Applicable (J34T3569G), Not Applicable (G24T0333G), Not Applicable (C44T4923G), Not Applicable (C44T4949G), Not Applicable (D44T5041G), Not Applicable (J34T3578G), Not Applicable (J34T3580G), Not Applicable (M24T1538G), Not Applicable (K34T3981G), Not Applicable (L34T4049G), Not Applicable (L34T4083G), Not Applicable (L34T4023G), Not Applicable (D34T2624G), Not Applicable (F24T0240G), Not Applicable (B34T1959G), Not Applicable (B34T1993G), Not Applicable (K34T3849G), Not Applicable (M24T1514G), Not Applicable (M24T1541G), Not Applicable (M24T1553G), Not Applicable (M24T1556G), Not Applicable (M24T1558G), Not Applicable (M24T1589G), Not Applicable (M24T1643G), Not Applicable (M24T1648G), Not Applicable (M24T1650G), Not Applicable (M24T1704G), Not Applicable (M24T1705G), Not Applicable (M24T1740G), Not Applicable (M24T1747G), Not Applicable (F24T0146G), Not Applicable (L24T1282G), Not Applicable (L24T1283G), Not Applicable (L24T1285G), Not Applicable (L24T1288G), Not Applicable (L24T1379G), Not Applicable (L24T1380G), Not Applicable (L24T1394G), Not Applicable (L24T1395G), Not Applicable (M34T4222G), Not Applicable (M34T4232G), Not Applicable (K24T1174G), Not Applicable (A34T1911G), Not Applicable (F24T0129G), Not Applicable (L34T4027G), Not Applicable (L34T4037G), Not Applicable (M24T1426G), Not Applicable (M24T1702G), Not Applicable (D44T5040G), Not Applicable (H34T3493G), Not Applicable (J34T3514G), Not Applicable (J34T3516G), Not Applicable (M34T4223G), Not Applicable (D34T2603G), Not Applicable (D34T2625G), Not Applicable (D34T2626G), Not Applicable (E34T2690G), Not Applicable (E34T2692G), Not Applicable (K34T3895G), Not Applicable (K34T3923G), Not Applicable (L34T4011G), Not Applicable (L34T4029G), Not Applicable (F24T0268G), Not Applicable (E34T2743G), Not Applicable (J34T3596G), Not Applicable (J34T3788G), Not Applicable (L24T1350G), Not Applicable (L24 T1353G), Not Applicable (C34T2250G), Not Applicable (C34T2253G), Not Applicable (E34T2682G), Not Applicable (G24T0331G), Not Applicable (G34T3209G), Not Applicable (M34T4357G), Not Applicable (J24T0624G), Not Applicable (B34T1981G), Not Applicable (B44T4842G), Not Applicable (B44T4901G), Not Applicable (C34T2292G), Not Applicable (C34T2295G), Not Applicable (F34T2965G), Not Applicable (F34T2966G), Not Applicable (F34T2968G), Not Applicable (F34T2969G), Not Applicable (F34T2974G), Not Applicable (F34T2988G), Not Applicable (F34T2989G), Not Applicable (L34T4031G), Not Applicable (M24T1567G), Not Applicable (M24T1568G), Not Applicable (M24T1570G), Not Applicable (M24T1572G), Not Applicable (M24T1574G), Not Applicable (M24T1678G), Not Applicable (M24T1635G), Not Applicable (M24T1639G), Not Applicable (B34T2032G), Not Applicable (B34T2040G), Not Applicable (C34T2273G), Not Applicable (C34T2280G), Not Applicable (C34T2332G), Not Applicable (C34T2333G), Not Applicable (C34T2371G), Not Applicable (E34T2723G), Not Applicable (A44T4705G), Not Applicable (D44T5026G), Not Applicable (A44T4717G), Not Applicable (A44T4719G), Not Applicable (A44T4727G), Not Applicable (K24T1192G), Not Applicable (A44T4697G), Not Applicable (D44T5021G), Not Applicable (D44T5022G), Not Applicable (M34T4268G), Not Applicable (M34T4377G), Not Applicable (F34T2925G), Not Applicable (G34T3227G), Not Applicable (G34T3235G), Not Applicable (H34T3259G), Not Applicable (H34T3262G), Not Applicable (M34T4267G), Not Applicable (M34T4320G), Not Applicable (M34T4384G), Not Applicable (A44T4594G), Not Applicable (B44T4761G), Not Applicable (B44T4847G), Not Applicable (K34T3867G), Not Applicable (K34T3984G), Not Applicable (K34T3985G), Not Applicable (K34T3986G), Not Applicable (K34T3989G), Not Applicable (A44T4582G), Not Applicable (A44T4656G), Not Applicable (F34T3018G), Not Applicable (F34T3043G), Not Applicable (F34T3045G), Not Applicable (F34T3046G), Not Applicable (D34T2610G), Not Applicable (G24T0555G), Not Applicable (G24T0579G), Not Applicable (J24T0634G), Not Applicable (L24T1319G), Not Applicable (M24T1752G), Not Applicable (M24T1754G), Not Applicable (F34T3030G), Not Applicable (F34T3049G), Not Applicable (G34T3061G), Not Applicable (K24T1249G), Not Applicable (M24T1530G), Not Applicable (M24T1534G), Not Applicable (M34T4487G), Not Applicable (M34T4489G), Not Applicable (D34T2520G), Not Applicable (F34T2819G), Not Applicable (F34T2881G), Not Applicable (F34T2882G), Not Applicable (F34T2885G), Not Applicable (F34T2891G), Not Applicable (G24T0370G), Not Applicable (G24T0415G), Not Applicable (G24T0416G), Not Applicable (G24T0418G), Not Applicable (G24T0493G), Not Applicable (J24T0776G), Not Applicable (J24T0788G), Not Applicable (K24T1006G), Not Applicable (M24T1613G), Not Applicable (G34T3205G), Not Applicable (A34T1812G), Not Applicable (A34T1920G), Not Applicable (A34T1923G), Not Applicable (A34T1924G), Not Applicable (A34T1926G), Not Applicable (A44T4574G), Not Applicable (A44T4575G), Not Applicable (A44T4576G), Not Applicable (A44T4584G), Not Applicable (A44T4585G), Not Applicable (A44T4586G), Not Applicable (A44T4590G), Not Applicable (A44T4591G), Not Applicable (A44T4593G), Not Applicable (A44T4597G), Not Applicable (A44T4599G), Not Applicable (A44T4604G), Not Applicable (A44T4605G), Not Applicable (A44T4606G), Not Applicable (A44T4607G), Not Applicable (A44T4609G), Not Applicable (A44T4610G), Not Applicable (A44T4611G), Not Applicable (A44T4612G), Not Applicable (A44T4615G), Not Applicable (A44T4616G), Not Applicable (A44T4617G), Not Applicable (A44T4618G), Not Applicable (A44T4619G), Not Applicable (A44T4621G), Not Applicable (A44T4622G), Not Applicable (A44T4630G), Not Applicable (A44T4632G), Not Applicable (A44T4636G), Not Applicable (A44T4641G), Not Applicable (A44T4646G), Not Applicable (A44T4647G), Not Applicable (A44T4648G), Not Applicable (A44T4649G), Not Applicable (A44T4650G), Not Applicable (A44T4652G), Not Applicable (A44T4653G), Not Applicable (A44T4654G), Not Applicable (A44T4655G), Not Applicable (A44T4658G), Not Applicable (A44T4659G), Not Applicable (A44T4660G), Not Applicable (A44T4661G), Not Applicable (A44T4669G), Not Applicable (A44T4672G), Not Applicable (A44T4675G), Not Applicable (A44T4677G), Not Applicable (A44T4718G), Not Applicable (A44T4726G), Not Applicable (A44T4728G), Not Applicable (B34T1933G), Not Applicable (B44T4780G), Not Applicable (D34T2501G), Not Applicable (D34T2505G), Not Applicable (D34T2514G), Not Applicable (D34T2528G), Not Applicable (E34T2749G), Not Applicable (E34T2753G), Not Applicable (F34T2866G), Not Applicable (F34T2991G), Not Applicable (G24T0510G), Not Applicable (G34T3063G), Not Applicable (G34T3137G), Not Applicable (G34T3204G), Not Applicable (G34T3236G), Not Applicable (H34T3402G), Not Applicable (H34T3403G), Not Applicable (H34T3404G), Not Applicable (H34T3405G), Not Applicable (H34T3406G), Not Applicable (H34T3407G), Not Applicable (H34T3409G), Not Applicable (H34T3410G), Not Applicable (H34T3412G), Not Applicable (H34T3424G), Not Applicable (H34T3475G), Not Applicable (H34T3481G), Not Applicable (H34T3485G), Not Applicable (H34T3490G), Not Applicable (H34T3491G), Not Applicable (J24T0919G), Not Applicable (J24T0924G), Not Applicable (J34T3667G), Not Applicable (J34T3668G), Not Applicable (J34T3705G), Not Applicable (J34T3781G), Not Applicable (J34T3802G), Not Applicable (K24T1255G), Not Applicable (K34T3934G), Not Applicable (K34T3935G), Not Applicable (L34T4035G), Not Applicable (L34T4158G), Not Applicable (L34T4163G), Not Applicable (M34T4262G), Not Applicable (M34T4263G), Not Applicable (M34T4264G), Not Applicable (M34T4288G), Not Applicable (M34T4325G), Not Applicable (M34T4326G), Not Applicable (M34T4327G), Not Applicable (M34T4332G), Not Applicable (M34T4333G), Not Applicable (M34T4335G), Not Applicable (M34T4338G), Not Applicable (M34T4342G), Not Applicable (M34T4344G), Not Applicable (M34T4374G), Not Applicable (M34T4376G), Not Applicable (M34T4379G), Not Applicable (M34T4382G), Not Applicable (M34T4385G), Not Applicable (M34T4386G), Not Applicable (M34T4387G), Not Applicable (M34T4388G), Not Applicable (M34T4391G), Not Applicable (M34T4392G), Not Applicable (M34T4448G), Not Applicable (M34T4451G), Not Applicable (B44T4784G), Not Applicable (G34T3084G), Not Applicable (G34T3086G), Not Applicable (G34T3107G), Not Applicable (G34T3171G), Not Applicable (G34T3185G), Not Applicable (J34T3542G), Not Applicable (E34T2762G), Not Applicable (F34T2826G), Not Applicable (F34T2827G), Not Applicable (F34T2911G), Not Applicable (L24T1306G), Not Applicable (L24T1307G), Not Applicable (M24T1518G), Not Applicable (M34T4227G), Not Applicable (M34T4400G), Not Applicable (M34T4405G), Not Applicable (M34T4467G), Not Applicable (M34T4468G), Not Applicable (M34T4523G), Not Applicable (E34T2656G), Not Applicable (G24T0443G), Not Applicable (G24T0514G), Not Applicable (J34T3738G), Not Applicable (J34T3769G), Not Applicable (K24T1156G), Not Applicable (M34T4206G), Not Applicable (M34T4422G), Not Applicable (M34T4462G), Not Applicable (M34T4606G), Not Applicable (M34T4474G), Not Applicable (E34T2767G), Not Applicable (F34T3031G), Not Applicable (L24T1389G), Not Applicable (M24T1421G), Not Applicable (M24T1424G), Not Applicable (M24T1425G), Not Applicable (M24T1433G), Not Applicable (B34T2170G), Not Applicable (C34T2263G), Not Applicable (C34T2266G), Not Applicable (C34T2296G), Not Applicable (J24T0775G), Not Applicable (K24T1066G), Not Applicable (K24T1217G), Not Applicable (K34T3962G), Not Applicable (K34T3963G), Not Applicable (G24T0546G), Not Applicable (J24T0643G), Not Applicable (J24T0649G), Not Applicable (J24T0650G), Not Applicable (J24T0652G), Not Applicable (K24T1114G), Not Applicable (H34T3287G), Not Applicable (H34T3288G), Not Applicable (H34T3289G), Not Applicable (H34T3294G), Not Applicable (H34T3295G), Not Applicable (H34T3297G), Not Applicable (H34T3317G), Not Applicable (H34T3328G), Not Applicable (K34T3931G), Not Applicable (K34T3976G), Not Applicable (M34T4253G), Not Applicable (M34T4396G), Not Applicable (J34T3522G), Not Applicable (J34T3599G), Not Applicable (J34T3602G), Not Applicable (K24T1235G), Not Applicable (K24T1264G), Not Applicable (K24T1265G), Not Applicable (B44T4916G), Not Applicable (B44T4917G), Not Applicable (B44T4918G), Not Applicable (F34T2905G), Not Applicable (F34T2906G), Not Applicable (K34T3845G), Not Applicable (K34T3846G), Not Applicable (A44T4729G), Not Applicable (L34T4028G), Not Applicable (K34T3982G), Not Applicable (D34T2618G), Not Applicable (K24T1182G), Not Applicable (K34T3957G), Not Applicable (M24T1480G), Not Applicable (D34T2633G), Not Applicable (K34T3825G), Not Applicable (K34T3828G), Not Applicable (L34T4175G), Not Applicable (H34T3255G), Not Applicable (D24T2492G), Not Applicable (D34T2492G), Not Applicable (D34T2510G), Not Applicable (D34T2542J), Not Applicable (D34T2569G), Not Applicable (E34T2659G), Not Applicable (L24T1305G), Not Applicable (M24T1735G), Not Applicable (B44T4894G), Not Applicable (H24T1228G), Not Applicable (K24T1228G), Not Applicable (C44T4965G), Not Applicable (D44T5030G), Not Applicable (A44T4671G), Not Applicable (C34T2417G), Not Applicable (A44T4699G), Not Applicable (B44T4785G), Not Applicable (E34T2754G), Not Applicable (F34T2929G), Not Applicable (D34T2543G), Not Applicable (C34T2225G), Not Applicable (H34T3460G), Not Applicable (B34T2098G), Not Applicable (A34T1889G), Not Applicable (M34T4350G), Not Applicable (D34T2486G), Not Applicable (D34T2598G), Not Applicable (C44T4940G), Not Applicable (C44T4982G), Not Applicable (D44T5034G), Not Applicable (F34T2900G), Not Applicable (J24T0743G), Not Applicable (F24T0093G), Not Applicable (F24T0094G), Not Applicable (F24T0095G), Not Applicable (F24T0109G), Not Applicable (F24T0114G), Not Applicable (F24T0115G), Not Applicable (G24T0403G), Not Applicable (G24T0409G), Not Applicable (G24T0419G), Not Applicable (G24T0539G), Not Applicable (H24T0591G), Not Applicable (H24T0592G), Not Applicable (H24T0593G), Not Applicable (J24T0604G), Not Applicable (L24T1338G), Not Applicable (L24T1344G), Not Applicable (L34T4092G), Not Applicable (M34T4343G), Not Applicable (F34T3023G), Not Applicable (J34T3531G), Not Applicable (K24T1092G), Not Applicable (K34T3853G), Not Applicable (K34T3856G), Not Applicable (K34T3857G), Not Applicable (M34T4259G), Not Applicable (L24T1357G), Not Applicable (A34T1786G), Not Applicable (A34T1855G), Not Applicable (A34T1908G), Not Applicable (A34T1909G), Not Applicable (A44T4560G), Not Applicable (B34T1948G), Not Applicable (B34T1954G), Not Applicable (B34T1976G), Not Applicable (C34T2232G), Not Applicable (J24T0943G), Not Applicable (A44T4620G), Not Applicable (A44T4644G), Not Applicable (A34T1772G), Not Applicable (A34T1773G), Not Applicable (J34T3733G), Not Applicable (J34T3734G), Not Applicable (J34T3735G), Not Applicable (A44T4731G), Not Applicable (A44T4733G), Not Applicable (A44T4734G), Not Applicable (A44T4736G), Not Applicable (A44T4738G), Not Applicable (B44T4767G), Not Applicable (B44T4830G), Not Applicable (H34T3415G), Not Applicable (H34T3418G), Not Applicable (M34T4509G), Not Applicable (B44T4775G), Not Applicable (A44T4721G), Not Applicable (K24T1230G), Not Applicable (G24T0561G), Not Applicable (B34T1995G), Not Applicable (D34T2467G), Not Applicable (D34T2533G), Not Applicable (D34T2550G), Not Applicable (D34T2551G), Not Applicable (K34T3854G), Not Applicable (B44T4794G), Not Applicable (C34T2230G), Not Applicable (B44T4893G), Not Applicable (B44T4766G), Not Applicable (B44T4835G), Not Applicable (B44T4819G), Not Applicable (D34T2446G), Not Applicable (K34T3850G), Not Applicable (K34T3893G), Not Applicable (L34T4067G), Not Applicable (B44T4802G), Not Applicable (A34T1856G), Not Applicable (B34T1975G), Not Applicable (B44T4837G), Not Applicable (L34T4002G), Not Applicable (L34T4006G), Not Applicable (L34T4007G), Not Applicable (L34T4017G), Not Applicable (L34T4022G), Not Applicable (L34T4149G), Not Applicable (L34T4152G), Not Applicable (J24T0793G), Not Applicable (B34T2063G), Not Applicable (B34T2193G), Not Applicable (C44T4963G), Not Applicable (C44T4964G), Not Applicable (G24T0363G), Not Applicable (G24T0364G), Not Applicable (G24T0365G), Not Applicable (G24T0366G), Not Applicable (G24T0396G), Not Applicable (G24T0482G), Not Applicable (G24T0483G), Not Applicable (G24T0486G), Not Applicable (G24T0492G), Not Applicable (G24T0517G), Not Applicable (G24T0522G), Not Applicable (H24T0595G), Not Applicable (J24T0726G), Not Applicable (J24T0783G), Not Applicable (J24T0784G), Not Applicable (K24T1001G), Not Applicable (K24T1041G), Not Applicable (K24T1068G), Not Applicable (K24T1088G), Not Applicable (K24T1185G), Not Applicable (K34T3875G), Not Applicable (K34T3878G), Not Applicable (L34T4155G), Not Applicable (D34T2593G), Not Applicable (L34T3996G), Not Applicable (L34T3998G), Not Applicable (E44T5052G), Not Applicable (E44T5053G), Not Applicable (E44T5054G), Not Applicable (D44T5042G), Not Applicable (G34T3164G), Not Applicable (K24T1013G), Not Applicable (K24T1022G), Not Applicable (K24T1023G), Not Applicable (A34T1798G), Not Applicable (L24T1331G), Not Applicable (L24T1334G), Not Applicable (L24T1335G), Not Applicable (L24T1349G), Not Applicable (A34T1875G), Not Applicable (C34T2309G), Not Applicable (F34T2984G), Not Applicable (G24T0389G), Not Applicable (G24T0444G), Not Applicable (G24T0477G), Not Applicable (H34T3500G), Not Applicable (J34T3623G), Not Applicable (M24T1545G), Not Applicable (M24T1544G), Not Applicable (M24T1407G), Not Applicable (B34T2124G), Not Applicable (C44T4993G), Not Applicable (H34T3330G), Not Applicable (B44T4821G), Not Applicable (C34T2224G), Not Applicable (C34T2356G), Not Applicable (D34T2568G), Not Applicable (E34T2661G), Not Applicable (E44T5051G), Not Applicable (F24T0160G), Not Applicable (F24T0166G), Not Applicable (F24T0167G), Not Applicable (F24T0170G), Not Applicable (F24T0186G), Not Applicable (F24T0191G), Not Applicable (F24T0193G), Not Applicable (F24T0194G), Not Applicable (F24T0205G), Not Applicable (F24T0206G), Not Applicable (F44T5065G), Not Applicable (F44T5066G), Not Applicable (G24T0584G), Not Applicable (G34T3192G), Not Applicable (G34T3193G), Not Applicable (G34T3194G), Not Applicable (G34T3195G), Not Applicable (G34T3196G), Not Applicable (G34T3197G), Not Applicable (G34T3198G), Not Applicable (G34T3199G), Not Applicable (G34T3200G), Not Applicable (J24T0667G), Not Applicable (J34T3737G), Not Applicable (J34T3740G), Not Applicable (J34T3748G), Not Applicable (K24T1176G), Not Applicable (L24T1299G), Not Applicable (M24T1612G), Not Applicable (M24T1616G), Not Applicable (M24T1617G), Not Applicable (M24T1621G), Not Applicable (M24T1623G), Not Applicable (M24T1624G), Not Applicable (M24T1625G), Not Applicable (M24T1626G), Not Applicable (M24T1628G), Not Applicable (B34T1961G), Not Applicable (C34T2229G), Not Applicable (C44T5003G), Not Applicable (D44T5018G), Not Applicable (J34T3577G), Not Applicable (J34T3587G), Not Applicable (M34T4460G), Not Applicable (M34T4417G), Not Applicable (M34T4499G), Not Applicable (C34T2353G), Not Applicable (C34T2368G), Not Applicable (E34T2687G), Not Applicable (E34T2735G), Not Applicable (E34T2736G), Not Applicable (F44T5072G), Not Applicable (B34T2134G), Not Applicable (E34T2714G), Not Applicable (F24T0154G), Not Applicable (F24T0169G), Not Applicable (B34T2116G), Not Applicable (F24T0174G), Not Applicable (F24T0207G), Not Applicable (F24T0208G), Not Applicable (F24T0209G), Not Applicable (F24T0286G), Not Applicable (J34T3699G), Not Applicable (K24T1244G), Not Applicable (F24T0159G), Not Applicable (F24T0172G), Not Applicable (B34T2130G), Not Applicable (K24T1150G), Not Applicable (M24T1516G), Not Applicable (F24T0195G), Not Applicable (F24T0192G), Not Applicable (C34T2227G), Not Applicable (C44T4995G), Not Applicable (E34T2769G), Not Applicable (F24T0021G), Not Applicable (F44T5074G), Not Applicable (J24T0728G), Not Applicable (J24T0730G), Not Applicable (K24T1169G), Not Applicable (F24T0162G), Not Applicable (M34T4182G), Not Applicable (M24T1473G), Not Applicable (F24T0190G), Not Applicable (J24T0731G), Not Applicable (F24T0148G), Not Applicable (L24T1329G), Not Applicable (M24T1689G), Not Applicable (J24T0729G), Not Applicable (K24T1162G), Not Applicable (M24T1469G), Not Applicable (L24T1291G), Not Applicable (L24T1292G), Not Applicable (G24T0323G), Not Applicable (M24T1622G), Not Applicable (F44T5064G), Not Applicable (F44T5067G), Not Applicable (F44T5068G), Not Applicable (B34T2144G), Not Applicable (F44T5063G), Not Applicable (B44T4820G), Not Applicable (F34T2829G), Not Applicable (F34T3038G), Not Applicable (F44T5073G), Not Applicable (G24T0378G), Not Applicable (E34T2730G), Not Applicable (F34T2822G), Not Applicable (F34T2828G), Not Applicable (J34T3644G), Not Applicable (B34T2147G), Not Applicable (J34T3570G), Not Applicable (F44T5075G), Not Applicable (J34T3523G), Not Applicable (J34T3524G), Not Applicable (J34T3624G), Not Applicable (J34T3646G), Not Applicable (C34T2319G), Not Applicable (F44T5061G), Not Applicable (F44T5062G), Not Applicable (J34T3751G), Not Applicable (M34T4399G), Not Applicable (M34T4436G), Not Applicable (M34T4437G), Not Applicable (M34T4438G), Not Applicable (M34T4439G), Not Applicable (M34T4397G), Not Applicable (E34T2770G), Not Applicable (F24T0064G), Not Applicable (M24T1604G), Not Applicable (K24T1259G), Not Applicable (M34T4257G), Not Applicable (J24T0746G), Not Applicable (F24T0187G), Not Applicable (M24T1620G), Not Applicable (C34T2286G), Not Applicable (G24T0372G), Not Applicable (G24T0440G), Not Applicable (C34T2287G), Not Applicable (C34T2293G), Not Applicable (J34T3676G), Not Applicable (J34T3745G), Not Applicable (F24T0184G), Not Applicable (F24T0185G), Not Applicable (J34T3680G), Not Applicable (J24T0724G), Not Applicable (C34T2294G), Not Applicable (B34T2160G), Not Applicable (C44T4996G), Not Applicable (F24T0163G), Not Applicable (F24T0168G), Not Applicable (J34T3742G), Not Applicable (M24T1619G), Not Applicable (M24T1627G), Not Applicable (C34T2317G), Not Applicable (F24T0158G), Not Applicable (F24T0226G), Not Applicable (H34T3290G), Not Applicable (H34T3291G), Not Applicable (J34T3609G), Not Applicable (J34T3610G), Not Applicable (J34T3731G), Not Applicable (C44T4985G), Not Applicable (F24T0070G), Not Applicable (F24T0165G), Not Applicable (H34T3298G), Not Applicable (F24T0157G), Not Applicable (B34T2117G), Not Applicable (B34T2123G), Not Applicable (B34T2146G), Not Applicable (B34T2148G), Not Applicable (B34T2157G), Not Applicable (B44T4887G), Not Applicable (B44T4890G), Not Applicable (C34T2316G), Not Applicable (C44T4999G), Not Applicable (C44T5000G), Not Applicable (C44T5002G), Not Applicable (F24T0171G), Not Applicable (F24T0173G), Not Applicable (J34T3640G), Not Applicable (J34T3641G), Not Applicable (J34T3645G), Not Applicable (F44T5060G), Not Applicable (F44T5070G), Not Applicable (A34T1794G), Not Applicable (F44T5071G), Not Applicable (F24T0164G), Not Applicable (G44T5077G), Not Applicable (J24T0758G), Not Applicable (M24T1527G), Not Applicable (F24T0188G), Not Applicable (J24T0608G), Not Applicable (J24T0662G), Not Applicable (B44T4787G), Not Applicable (J34T3607G), Not Applicable (J34T3743G), Not Applicable (M24T1591G), Not Applicable (M34T4298G), Not Applicable (C34T2238G), Not Applicable (C44T4994G), Not Applicable (M34T4299G), Not Applicable (C44T4948G), Not Applicable (E44T5058G), Not Applicable (K24T1168G), Not Applicable (B34T2168G), Not Applicable (C34T2343G), Not Applicable (F24T0161G), Not Applicable (J34T3617G), Not Applicable (F44T5069G), Not Applicable (B34T2162G), Not Applicable (C34T2310G), Not Applicable (B34T2142G), Not Applicable (B34T2119G), Not Applicable (B34T2156G), Not Applicable (C34T2335G), Not Applicable (C44T5001G), Not Applicable (J34T3642G), Not Applicable (C34T2311G), Not Applicable (C34T2313G), Not Applicable (J34T3639G), Not Applicable (J34T3643G), Not Applicable (B44T4884G), Not Applicable (C44T5006G), Not Applicable (J34T3637G), Not Applicable (B34T2143G), Not Applicable (C34T2338G), Not Applicable (B34T2149G), Not Applicable (C34T2320G), Not Applicable (C44T4986G), Not Applicable (B34T2145G), Not Applicable (C34T2314G), Not Applicable (J34T3638G), Not Applicable (B44T4829G), Not Applicable (J24T0911G), Not Applicable (J24T0913G), Not Applicable (K24T1237G), Not Applicable (G24T0455G), Not Applicable (E34T2734G), Not Applicable (G24T0530G), Not Applicable (A34T1761G), Not Applicable (A34T1777G), Not Applicable (A34T1793G), Not Applicable (A34T1918G), Not Applicable (F34T2919G), Not Applicable (F34T2921G), Not Applicable (A34T1796G), Not Applicable (A34T1878G), Not Applicable (A44T4642G), Not Applicable (F34T2884G), Not Applicable (F34T3028G), Not Applicable (H34T3469G), Not Applicable (J34T3723G), Not Applicable (J34T3724G), Not Applicable (J34T3729G), Not Applicable (J34T3730G), Not Applicable (A34T1862G), Not Applicable (A34T1782G), Not Applicable (A34T1795G), Not Applicable (B34T2077G), Not Applicable (E34T2683G), Not Applicable (F34T2913G), Not Applicable (F34T2920G), Not Applicable (F34T2923G), Not Applicable (G24T0433G), Not Applicable (K24T1163G), Not Applicable (G34T3125G), Not Applicable (G24T0430G), Not Applicable (K34T3389G), Not Applicable (J24T0912G), Not Applicable (F24T0212G), Not Applicable (J24T0719G), Not Applicable (B44T4915G), Not Applicable (J24T0878G), Not Applicable (D44T5016G), Not Applicable (H34T3494G), Not Applicable (J34T3615G), Not Applicable (J34T3616G), Not Applicable (J34T3618G), Not Applicable (M34T4184G), Not Applicable (J24T0666G), Not Applicable (G24T0585G), Not Applicable (H34T3381G), Not Applicable (L24T1342G), Not Applicable (M24T1631G), Not Applicable (A34T1887G), Not Applicable (E34T2676G), Not Applicable (L24T1332G), Not Applicable (A24T1026G), Not Applicable (A34T1834G), Not Applicable (A34T1835G), Not Applicable (A34T1836G), Not Applicable (A34T1837G), Not Applicable (K24T1026G), Not Applicable (K24T1027G), Not Applicable (G24T0533G), Not Applicable (K24T0553G), Not Applicable (K24T0984G), Not Applicable (A34T1797G), Not Applicable (M34T4179G), Not Applicable (A34T1849G), Not Applicable (A34T1852G), Not Applicable (G34T3059G), Not Applicable (D34T2426G), Not Applicable (A34T1838G), Not Applicable (B34T2999G), Not Applicable (J34T4405G), Not Applicable (F34T2859G), Not Applicable (H34T3353G), Not Applicable (K34T3945G), Not Applicable (G24T0541G), Not Applicable (M34T4258G), Not Applicable (H34T3385G), Not Applicable (D44T0531G), Not Applicable (D44T5031G), Not Applicable (C44T4936G), Not Applicable (G44T4936G), Not Applicable (A34T1876G), Not Applicable (D44T5044G), Not Applicable (A34T1869G), Not Applicable (B34T1869G), Not Applicable (B34T1940G), Not Applicable (B34T194OG), Not Applicable (G34T3060G), Not Applicable (A34T850G), Not Applicable (F24T0105G), Not Applicable (D34T2425G), Not Applicable (G34T3213G), Not Applicable (K34T3944G), Not Applicable (C44T4988G), Not Applicable (A44T4602G), Not Applicable (A44T4603G), Not Applicable (C44T4928G), Not Applicable (A34T1885G), Not Applicable (G34T3220G), Not Applicable (A34T1771G), Not Applicable (A34T1851G), Not Applicable (A44T4578G), Not Applicable (C34T2420G), Not Applicable (E34T2681G), Not Applicable (F34T2800G), Not Applicable (F34T2801G), Not Applicable (H34T3336G), Not Applicable (K34T3947G), Not Applicable (L24T1336G), Not Applicable (L24T1339G), Not Applicable (M24T1465G), Not Applicable (B44T4878G), Not Applicable (C44T4975G), Not Applicable (G24T03756), Not Applicable (G24T0375G), Not Applicable (K24T1025), Not Applicable (K24T1025G), Not Applicable (H34T3352G), Not Applicable (K24T1018G), Not Applicable (A34T1799), Not Applicable (A34T1799G), Not Applicable (A34T1858G), Not Applicable (G34T3219G), Not Applicable (K34T3847G), Not Applicable (K34T8347G), Not Applicable (L24T1333G), Not Applicable (L24T1347G), Not Applicable (B44T4777G), Not Applicable (C34T2419G), Not Applicable (C34T2449G), Not Applicable (C44T4927G), Not Applicable (L24T1341), Not Applicable (L24T1341G), Not Applicable (G34T3214G), Not Applicable (G34T3212G), Not Applicable (A34T1847G), Not Applicable (B44T4776G), Not Applicable (G24T0337G), Not Applicable (L24T1337G), Not Applicable (L24T1346G), Not Applicable (M34T4398G), Not Applicable (E34T2686G), Not Applicable (G34T3221G), Not Applicable (G34T3222G), Not Applicable (H34T3386G), Not Applicable (L24T1343G), Not Applicable (M34T4503G), Not Applicable (K24T0995G), Not Applicable (A34T3354G), Not Applicable (A34T3355G), Not Applicable (H34T3354G), Not Applicable (H34T3355G), Not Applicable (F24T0098G), Not Applicable (A34T1850G), Not Applicable (A44T4600G), Not Applicable (L34T4094G), Not Applicable (A44T4577G), Not Applicable (C44T4991G), Not Applicable (C44T4992G), Not Applicable (C44T4976G), Not Applicable (G24T0452G), Not Applicable (H24T0597G), Not Applicable (F34T2907G), Not Applicable (K24T1172G), Not Applicable (J24T0672G), Not Applicable (J24T0673G), Not Applicable (L24T1372G), Not Applicable (J34T3557G), Not Applicable (C34T2285G), Not Applicable (G24T0432G), Not Applicable (F34T2803G), Not Applicable (C44T4950G), Not Applicable (G34T3218G), Not Applicable (E34T2679G), Not Applicable (H34T3337G), Not Applicable (D34T2453G), Not Applicable (G34T3216G), Not Applicable (E34T2670G), Not Applicable (J24T0720G), Not Applicable (C34T2276G), Not Applicable (C34T2277G), Not Applicable (B44T4827G), Not Applicable (H34T3338G), Not Applicable (G24T0450G), Not Applicable (G24T0451G), Not Applicable (J24T0745G), Not Applicable (D34T2526G), Not Applicable (J34T3622G), Not Applicable (J34T3647G), Not Applicable (D34T2423G), Not Applicable (K24T1205G), Not Applicable (M34T4192G), Not Applicable (J24T0744G), Not Applicable (K24T1002G), Not Applicable (C44T4926G), Not Applicable (C44T4935G), Not Applicable (C44T4967G), Not Applicable (C44T4968G), Not Applicable (C44T4969G), Not Applicable (C44T4970G), Not Applicable (C44T4971G), Not Applicable (C44T4972G), Not Applicable (E34T2671G), Not Applicable (G44T4968G), Not Applicable (F24T0089G), Not Applicable (F24T0090G), Not Applicable (F24T0091G), Not Applicable (H24T0601G), Not Applicable (A34T1801G), Not Applicable (M24T1463G), Not Applicable (M34T4294G), Not Applicable (G34T3057G), Not Applicable (A34T1868G), Not Applicable (D34T2452G), Not Applicable (F34T2804G), Not Applicable (K24T1070G), Not Applicable (A44T4570G), Not Applicable (D34T2627G), Not Applicable (A44T4572G), Not Applicable (K24T1020G), Not Applicable (G24T0436G), Not Applicable (G24T0437G), Not Applicable (A44T4569G), Not Applicable (B44T4798G), Not Applicable (G24T0329G), Not Applicable (D44T5035G), Not Applicable (D44T5043G), Not Applicable (G34T3217G), Not Applicable (C34T2222G), Not Applicable (F34T2901G), Not Applicable (G34T3096G), Not Applicable (A34T1890G), Not Applicable (H24T0596G), Not Applicable (C34T2323G), Not Applicable (D34T2502G), Not Applicable (A34T1775G), Not Applicable (A34T1871G), Not Applicable (D34T2646G), Not Applicable (M34T4290G), Not Applicable (A34T1802G), Not Applicable (C44T4990G), Not Applicable (B44T4828G), Not Applicable (B44T4874G), Not Applicable (B44T4875G), Not Applicable (B44T4877G), Not Applicable (G34T3165G), Not Applicable (L34T4093G), Not Applicable (J34T3761G), Not Applicable (F34T2897G), Not Applicable (F34T2898G), Not Applicable (F34T2899G), Not Applicable (F34T2908G), Not Applicable (K24T1071G), Not Applicable (A44T4722G), Not Applicable (C34T3388G), Not Applicable (C34T3609G), Not Applicable (C34T3622G), Not Applicable (D44T5047G), Not Applicable (D44T5048G), Not Applicable (D44T5049G), Not Applicable (D44T5050G), Not Applicable (M34T4180G), Not Applicable (M34T4292G), Not Applicable (M34T4293G), Not Applicable (J34T3501G), Not Applicable (A34T1821G), Not Applicable (B34T2986G), Not Applicable (F34T2802G), Not Applicable (K24T1170G), Not Applicable (K24T1171G), Not Applicable (K24T1206G), Not Applicable (L24T1371G), Not Applicable (M34T4291G), Not Applicable (M24T1546G), Not Applicable (E34T2678G), Not Applicable (M34T4277G), Not Applicable (G34T3166G), Not Applicable (G34T3191G), Not Applicable (G24T0327G), Not Applicable (G24T0532G), Not Applicable (K34T3943G), Not Applicable (E44T5055G), Not Applicable (E44T5056G), Not Applicable (B44T4876G), Not Applicable (M34T4181G), Not Applicable (B34T2182G), Not Applicable (A44T4571G), Not Applicable (H34T3387G), Not Applicable (D34T2432G), Not Applicable (B34T2181G), Not Applicable (A44T4549G), Not Applicable (A34T1800G), Not Applicable (D34T2548G), Not Applicable (H34T3474G), Not Applicable (F34T2805G), Not Applicable (F24T0092G), Not Applicable (M34T4188G), Not Applicable (M34T4189G), Not Applicable (M34T4190G), Not Applicable (M34T4217G), Not Applicable (G24T0431G), Not Applicable (M24T1408G), Not Applicable (M34T4218G), Not Applicable (G34T3097G), Not Applicable (C34T2410G), Not Applicable (M34T4187G), Not Applicable (M34T4191G), Not Applicable (B34T1958G), Not Applicable (B34T2102G), Not Applicable (B34T2103G), Not Applicable (B34T2140G), Not Applicable (F24T0101G), Not Applicable (F24T0102G), Not Applicable (G34T3168G), Not Applicable (G34T3170G), Not Applicable (H34T3260G), Not Applicable (H34T3278G), Not Applicable (H34T3281G), Not Applicable (H34T3311G), Not Applicable (J24T0795G), Not Applicable (J34T3591G), Not Applicable (K24T1058G), Not Applicable (M24T1652G), Not Applicable (M34T4421G), Not Applicable (D34T2448G), Not Applicable (D34T2456G), Not Applicable (G34T3215G), Not Applicable (L34T4113G), Not Applicable (L34T4118G), Not Applicable (K34T3925G), Not Applicable (A44T4700G), Not Applicable (A44T4708G), Not Applicable (A44T4709G), Not Applicable (F24T0213G), Not Applicable (K24T1050G), Not Applicable (G24T0540G), Not Applicable (H34T3383G), Not Applicable (H34T3384G), Not Applicable (H34T3462G), Not Applicable (J34T3773G), Not Applicable (M34T4271G), Not Applicable (M34T4272G), Not Applicable (C44T4980G), Not Applicable (F24T0136G), Not Applicable (F24T0137G), Not Applicable (F24T0138G), Not Applicable (F24T0139G), Not Applicable (F24T0140G), Not Applicable (F24T0176G), Not Applicable (F24T0177G), Not Applicable (F24T0179G), Not Applicable (F24T0180G), Not Applicable (F24T0181G), Not Applicable (J24T0688G), Not Applicable (F24T0153G), Not Applicable (A44T4712G), Not Applicable (G24T0427G), Not Applicable (G24T0439G), Not Applicable (J34T3573G), Not Applicable (E34T2712G), Not Applicable (F34T2914G), Not Applicable (B34T1996G), Not Applicable (F24T0052G), Not Applicable (M24T1522G), Not Applicable (F24T0127G), Not Applicable (F24T0201G), Not Applicable (M24T1529G), Not Applicable (F24T0107G), Not Applicable (K34T3930G), Not Applicable (B34T2164G), Not Applicable (B34T2165G), Not Applicable (B34T2166G), Not Applicable (B34T2167G), Not Applicable (B34T2169G), Not Applicable (C34T2236G), Not Applicable (F24T0197G), Not Applicable (C34T2308G), Not Applicable (M34T4185G), Not Applicable (M34T4273G), Not Applicable (J34T3619G), Not Applicable (M34T4183G), Not Applicable (L34T4091G), Not Applicable (M24T1659G), Not Applicable (M34T4274G), Not Applicable (M24T1629G), Not Applicable (C34T2322G), Not Applicable (G24T0383G), Not Applicable (G24T0550G), Not Applicable (C34T2418G), Not Applicable (D34T2422G), Not Applicable (A34T1886G), Not Applicable (M24T1467G), Not Applicable (C44T4961G), Not Applicable (A34T1863G), Not Applicable (M24T1406G), Not Applicable (E34T2672G), Not Applicable (E34T2673G), Not Applicable (E34T2674G), Not Applicable (E34T2675G), Not Applicable (E34T2677G), Not Applicable (G34T3056G), Not Applicable (C34T2258G), Not Applicable (M24T1547G), Not Applicable (G34T3223G), Not Applicable (M24T1409G), Not Applicable (M34T4219G), Not Applicable (G34T3225G), Not Applicable (G34T3098G), Not Applicable (M34T4220G), Not Applicable (M24T1548G), Not Applicable (D34T2421G), Not Applicable (G34T3224G), Not Applicable (K24T1005G), Not Applicable (B44T4789G), Not Applicable (B44T4790G), Not Applicable (B44T4791G), Not Applicable (B44T4792G), Not Applicable (B44T4793G), Not Applicable (J24T0674G), Not Applicable (D34T2424G), Not Applicable (A34T1844G), Not Applicable (C44T4989G), Not Applicable (M34T4275G), Not Applicable (D34T2607G), Not Applicable (M24T1462G), Not Applicable (M34T4186G), Not Applicable (L34T4095G), Not Applicable (C34T2212G), Not Applicable (M34T4178G), Not Applicable (A34T1881G), Not Applicable (J44T5093G), Not Applicable (F34T2850G), Not Applicable (C34T2248G), Not Applicable (C34T2249G), Not Applicable (E34T2657G), Not Applicable (L24T1385G), Not Applicable (L24T1405G), Not  Applicable (C34T2307G), Not Applicable (H44T5080G), Not Applicable (H44T5081G), Not Applicable (G34T3163G), Not Applicable (H44T5078G), Not Applicable (L44T5100G), Not Applicable (L44T5101G), Not Applicable (M24T1492G), Not Applicable (B34T2159G), Not Applicable (K34T3995G), Not Applicable (L44T5096G), Not Applicable (J44T5085G), Not Applicable (L24T1398G), Not Applicable (D34T2645G), Not Applicable (G34T3110G), Not Applicable (B34T2183G), Not Applicable (J44T5088G), Not Applicable (J44T5090G), Not Applicable (J44T5092G), Not Applicable (L44T5095G), Not Applicable (M24T1493G), Not Applicable (M24T1494G), Not Applicable (F34T2903G), Not Applicable (H44T5079G), Not Applicable (M34T4354G), Not Applicable (H34T3293G), Not Applicable (L34T4034G), Not Applicable (F34T2878G), Not Applicable (B34T1988G), Not Applicable (B44T4815G), Not Applicable (E34T2662G), Not Applicable (D34T2450G), Not Applicable (G24T0535G), Not Applicable (L24T1323G), Not Applicable (L24T1324G), Not Applicable (M34T4244G), Not Applicable (J34T3533G), Not Applicable (J34T3509G), Not Applicable (M34T4230G), Not Applicable (B34T1997G), Not Applicable (C34T2415G), Not Applicable (C44T4966G), Not Applicable (D34T2525G), Not Applicable (E34T2768G), Not Applicable (E34T2771G), Not Applicable (F24T0049G), Not Applicable (F24T0050G), Not Applicable (F24T0118G), Not Applicable (F24T0119G), Not Applicable (F24T0155G), Not Applicable (F34T2869G), Not Applicable (F34T2875G), Not Applicable (F34T2876G), Not Applicable (H34T3320G), Not Applicable (K24T0979G), Not Applicable (K34T3929G), Not Applicable (L24T1276G), Not Applicable (M34T4224G), Not Applicable (J24T0689G), Not Applicable (K34T3952G), Not Applicable (E34T2785G), Not Applicable (C34T2359G), Not Applicable (C34T2390G), Not Applicable (C34T2400G), Not Applicable (C34T2402G), Not Applicable (C34T2403G), Not Applicable (C34T2405G), Not Applicable (F34T2821G), Not Applicable (F34T2823G), Not Applicable (F34T3005G), Not Applicable (J24T0839G), Not Applicable (J24T0877G), Not Applicable (J24T0888G), Not Applicable (J24T0889G), Not Applicable (J24T0890G), Not Applicable (J34T3571G), Not Applicable (J34T3593G), Not Applicable (M34T4359G), Not Applicable (F34T2997G), Not Applicable (F34T3001G), Not Applicable (H34T3258G), 00840682109260M24T1559G (M24T1559G), Not Applicable (B34T2128G), Not Applicable (B44T4910G), Not Applicable (J34T3588G), Not Applicable (J34T3612G), Not Applicable (B34T2120G), Not Applicable (B44T4879G), Not Applicable (B44T4881G), Not Applicable (F24T0065G), Not Applicable (F34T2820G), Not Applicable (J34T3697G), Not Applicable (K34T3863G), Not Applicable (K34T3983G), Not Applicable (G24T0398G), Not Applicable (G24T0429G), Not Applicable (F24T0044G), Not Applicable (J24T0725G), Not Applicable (L34T4032G), Not Applicable (G24T0376G), Not Applicable (L34T4164G), Not Applicable (M34T4463G), Not Applicable (J34T3507G), Not Applicable (K24T1195G), Not Applicable (M24T1422G), Not Applicable (L34T4117G), Not Applicable (K34T3906G), Not Applicable (K34T3907G), Not Applicable (K34T3908G), Not Applicable (K34T3909G), Not Applicable (K34T3910G), Not Applicable (K34T3911G), Not Applicable (K34T3912G), Not Applicable (K34T3938G), Not Applicable (K34T3939G), Not Applicable (M34T4459G), Not Applicable (M34T4516G), Not Applicable (J24T0842G), Not Applicable (J24T0843G), Not Applicable (J24T0852G), Not Applicable (B34T2177G), Not Applicable (B34T2178G), Not Applicable (B34T2179G), Not Applicable (C44T4929G), Not Applicable (C44T4942G), Not Applicable (M34T4297G), Not Applicable (A34T1859G), Not Applicable (K24T1059G), Not Applicable (K24T1089G), Not Applicable (K24T1094G), Not Applicable (K24T1095G), Not Applicable (K24T1096G), Not Applicable (K24T1102G), Not Applicable (K24T1178G), Not Applicable (K24T1187G), Not Applicable (K34T3818G), Not Applicable (K34T3819G), Not Applicable (L24T1326G), Not Applicable (D44T5046G), Not Applicable (D34T2640G), Not Applicable (D34T2641G), Not Applicable (D34T2642G), Not Applicable (M24T1729G), Not Applicable (M24T1730G), Not Applicable (M24T1731G), Not Applicable (M24T1732G), Not Applicable (M24T1733G), Not Applicable (M24T1734G), Not Applicable (K24T1240G), Not Applicable (K24T1241G), Not Applicable (F24T0096G), Not Applicable (B34T2172G), Not Applicable (B34T2173G), Not Applicable (J34T3621G), Not Applicable (J34T3648G), Not Applicable (A34T1780G), Not Applicable (A34T1860G), Not Applicable (A34T1888G), Not Applicable (B34T2127G), Not Applicable (C34T2289G), Not Applicable (C34T2303G), Not Applicable (H34T3254G), Not Applicable (H34T3315G), Not Applicable (H34T3316G), Not Applicable (H34T3414G), Not Applicable (J24T0806G), Not Applicable (J24T0807G), Not Applicable (J34T3511G), Not Applicable (M34T4249G), Not Applicable (M34T4266G), Not Applicable (K24T1213G), Not Applicable (A44T4614G), Not Applicable (G34T3167G), Not Applicable (C34T2350G), Not Applicable (C44T4987G), Not Applicable (A34T1916G), Not Applicable (H34T3375G), Not Applicable (H34T3378G), Not Applicable (J34T3547G), Not Applicable (J34T3548G), Not Applicable (J34T3549G), Not Applicable (J34T3550G), Not Applicable (J34T3551G), Not Applicable (J34T3552G), Not Applicable (J34T3553G), Not Applicable (J34T3554G), Not Applicable (J34T3555G), Not Applicable (J34T3556G), Not Applicable (M34T4312G), Not Applicable (M34T4497G), Not Applicable (M34T4498G), Not Applicable (M34T4501G), Not Applicable (A34T1882G), Not Applicable (L34T4122G), Not Applicable (F34T2932G), Not Applicable (M34T4226G), Not Applicable (F34T2930G), Not Applicable (F34T2933G), Not Applicable (F34T2935G), Not Applicable (F34T2941G), Not Applicable (J34T3804G), Not Applicable (K34T3874G), Not Applicable (K34T3883G), Not Applicable (L34T4110G), Not Applicable (L34T4112G), Not Applicable (M24T1445G), Not Applicable (D34T2490G), Not Applicable (D34T2553G), Not Applicable (D34T2639G), Not Applicable (K34T3862G), Not Applicable (K34T3868G), Not Applicable (C34T2245G), Not Applicable (C34T2246G), Not Applicable (H34T3286G), Not Applicable (C44T4997G), Not Applicable (F24T0046G), Not Applicable (F24T0056G), Not Applicable (F24T0063G), Not Applicable (F24T0112G), Not Applicable (F34T2879G), Not Applicable (G24T0503G), Not Applicable (G24T0571G), Not Applicable (H34T3279G), Not Applicable (J24T0633G), Not Applicable (J24T0654G), Not Applicable (J24T0655G), Not Applicable (J24T0659G), Not Applicable (J24T0947G), Not Applicable (K24T1199G), Not Applicable (K34T3941G), Not Applicable (L24T1302G), Not Applicable (K24T1051G), Not Applicable (K24T1061G), Not Applicable (D34T2471G), Not Applicable (J24T0791G), Not Applicable (B44T4908G), Not Applicable (J24T0658G), Not Applicable (J24T0660G), Not Applicable (J34T3594G), Not Applicable (J34T3611G), Not Applicable (J34T3614G), Not Applicable (M24T1537G), Not Applicable (M24T1575G), Not Applicable (F34T3002G), Not Applicable (C34T2327G), Not Applicable (F34T2961G), Not Applicable (G24T0441G), Not Applicable (L24T1325G), Not Applicable (B44T4822G), Not Applicable (D34T2447G), Not Applicable (C34T2291G), Not Applicable (A34T1866G), Not Applicable (F34T3041G), Not Applicable (A44T4694G), Not Applicable (F34T2915G), Not Applicable (F34T2998G), Not Applicable (H34T3459G), Not Applicable (J34T3771G), Not Applicable (J34T3778G), Not Applicable (J34T3779G), Not Applicable (J34T3780G), Not Applicable (J34T3782G), Not Applicable (J24T0668G), Not Applicable (B44T4795G), Not Applicable (F24T0103G), Not Applicable (J24T0657G), Not Applicable (J24T0748G), Not Applicable (B34T2046G), Not Applicable (F24T0198G), Not Applicable (D34T2564G), Not Applicable (G34T3136G), Not Applicable (G24T0422G), Not Applicable (G24T0424G), Not Applicable (G24T0442G), Not Applicable (A34T1769G), Not Applicable (F34T2999G), Not Applicable (F34T3000G), Not Applicable (F34T3006G), Not Applicable (M24T1725G), Not Applicable (M24T1726G), Not Applicable (C44T5004G), Not Applicable (B34T2176G), Not Applicable (B34T2180G), Not Applicable (F24T0099G), Not Applicable (B34T1965G), Not Applicable (C34T2223G), Not Applicable (K24T1060G), Not Applicable (M34T4372G), Not Applicable (H34T3499G), Not Applicable (E34T2680G), Not Applicable (B34T2174G), Not Applicable (L24T1373G), Not Applicable (A44T4743G), Not Applicable (G24T0445G), Not Applicable (G24T0446G), Not Applicable (M24T1543G), Not Applicable (M24T1549G), Not Applicable (C44T5005G), Not Applicable (H34T3335G), Not Applicable (L24T1345G), Not Applicable (A34T1891G), Not Applicable (G24T0405G), Not Applicable (G24T0406G), Not Applicable (G24T0407G), Not Applicable (G24T0408G), Not Applicable (D44T5033G), Not Applicable (E34T2798G), Not Applicable (E34T2799G), Not Applicable (K34T3940G), Not Applicable (K34T3946G), Not Applicable (K34T3949G), Not Applicable (K24T1152G), Not Applicable (A34T1805G), Not Applicable (A34T1820G), Not Applicable (A34T1824G), Not Applicable (B34T2045G), Not Applicable (B44T4920G), Not Applicable (C44T4998G), Not Applicable (D34T2503G), Not Applicable (F34T2860G), Not Applicable (F34T2861G), Not Applicable (F34T2902G), Not Applicable (G24T0328G), Not Applicable (G24T0423G), Not Applicable (G24T0426G), Not Applicable (G24T0435G), Not Applicable (G24T0456G), Not Applicable (G34T3058G), Not Applicable (G34T3190G), Not Applicable (H34T3388G), Not Applicable (J34T3762G), Not Applicable (K34T3948G), Not Applicable (M34T4295G), Not Applicable (F24T0108G), Not Applicable (L34T4097G), Not Applicable (L34T4108G), Not Applicable (L34T4114G), Not Applicable (L34T4116G), Not Applicable (E34T2713G), Not Applicable (F34T2833G), Not Applicable (K24T1272G), Not Applicable (D34T2575G), Not Applicable (K24T1243G), Not Applicable (F34T2836G), Not Applicable (J24T0838G), Not Applicable (L24T1277G), Not Applicable (L24T1278G), Not Applicable (L24T1523G), Not Applicable (J24T0611G), Not Applicable (C34T2408G), Not Applicable (C34T2412G), Not Applicable (D34T2474G), Not Applicable (D34T2557G), Not Applicable (D34T2631G), Not Applicable (E34T2664G), Not Applicable (K34T3872G), Not Applicable (K34T3876G), Not Applicable (K34T3926G), Not Applicable (M34T4231G), Not Applicable (K34T3880G), Not Applicable (A34T1857G), Not Applicable (J24T0810G), Not Applicable (D34T2563G), Not Applicable (E34T2733G), Not Applicable (D34T2601G), Not Applicable (D34T2621G), Not Applicable (D34T2621G1), Not Applicable (E34T2658G), Not Applicable (D34T2498G), Not Applicable (D34T2566G), Not Applicable (D34T2523G), Not Applicable (G24T0326G), Not Applicable (L34T4055G), Not Applicable (G24T0413G), Not Applicable (B34T2068G), Not Applicable (C34T2406G), Not Applicable (D34T2449G), Not Applicable (D34T2571G), Not Applicable (D34T2622G), Not Applicable (J34T3576G), Not Applicable (J34T3583G), Not Applicable (L34T4040G), Not Applicable (D34T2619G), Not Applicable (D34T2620G), Not Applicable (J34T3759G), Not Applicable (B44T4892G), Not Applicable (A34T1774G), Not Applicable (B34T1936G), Not Applicable (G34T3103G), Not Applicable (A34T1823G), Not Applicable (B34T1955G), Not Applicable (K24T1030G), Not Applicable (H34T3468G), Not Applicable (B44T4838G), Not Applicable (B44T4883G), Not Applicable (A34T1818G), Not Applicable (A34T1865G), Not Applicable (A44T4673G), Not Applicable (A34T1767G), Not Applicable (L34T4000G), Not Applicable (M24T1428G), Not Applicable (C24T1046G), Not Applicable (A44T4698G), Not Applicable (B34T2122G), Not Applicable (H34T3333G), Not Applicable (K34T3975G), Not Applicable (M34T4454G), Not Applicable (B34T2083G), Not Applicable (J24T0797G), Not Applicable (F24T0062G), Not Applicable (B44T4897G), Not Applicable (K24T1177G), Not Applicable (A44T4663G), Not Applicable (M24T1594G), Not Applicable (M24T1595G), Not Applicable (A34T1783G), Not Applicable (D44T5024G), Not Applicable (A44T4740G), Not Applicable (B34T2118G), Not Applicable (K34T3953G), Not Applicable (G34T3184G), Not Applicable (K24T1046G), Not Applicable (K24T1157G), Not Applicable (K24T1194G), Not Applicable (K24T1202G), Not Applicable (K24T1204G), Not Applicable (M24T1444G), Not Applicable (F24T0106G), Not Applicable (G34T3133G), Not Applicable (K24T1137G), Not Applicable (C34T2226G), Not Applicable (K24T1164G), Not Applicable (B44T4803G), Not Applicable (A34T1845G), Not Applicable (J24T0940G), Not Applicable (J24T0944G), Not Applicable (B44T4759G), Not Applicable (B44T4772G), Not Applicable (K24T1047G), Not Applicable (K24T1065G), Not Applicable (A34T1867G), Not Applicable (A44T4633G), Not Applicable (A44T4674G), Not Applicable (G34T3169G), Not Applicable (K24T1049G), Not Applicable (K24T1045G), Not Applicable (M24T1474G), Not Applicable (G24T0425G), Not Applicable (K24T1140G), Not Applicable (B34T2125G), Not Applicable (B34T2002G), Not Applicable (B34T2003G), Not Applicable (B34T2942G), Not Applicable (F34T2942G), Not Applicable (K24T1179G), Not Applicable (K24T1201G), Not Applicable (K24T1263G), Not Applicable (L24T1400G), Not Applicable (A44T4680G), Not Applicable (F24T0061G), Not Applicable (B34T1957G), Not Applicable (B34T1969G), Not Applicable (B44T4825G), Not Applicable (A34T1776G), Not Applicable (B44T4786G), Not Applicable (C34T2413G), Not Applicable (B44T4796G), Not Applicable (J34T3630G), Not Applicable (J34T3632G), Not Applicable (B44T4762G), Not Applicable (B44T4771G), Not Applicable (J24T0603G), Not Applicable (M34T4234G), Not Applicable (C34T2297G), Not Applicable (G24T0399G), Not Applicable (B44T4805G), Not Applicable (G34T3100G), Not Applicable (J24T0605G), Not Applicable (J24T0606G);     3. Product Code: GU, No UDI Number, Serial Numbers: K0GU2966G, K0GU2976G, L0GU3003G, L0GU3004G, L0GU3007G, L0GU3008G, L0GU3009G, L0GU3010G, M0GU3113G, M0GU3148G, C1GU3291G, C1GU3292G, D9GU0690G, E9GU0800G, G9GU1036G, G9GU1043G, G9GU1049G, G9GU1050G, G9GU1051G, G9GU1052G, M9GU1713G, M9GU1723G, M9GU1730G, C9GU0649G, H9GU1240G, H9GU1241G, H9GU1242G, J0GU2875G, C0GU2061G, H9GU1096G, H9GU1098G, M0GU3094G, C1GU3331G, G0GU2522G, M0GU3118G, F0GU2452G, F0GU2453G, F0GU2454G, F0GU2456G, F0GU2475G, D9GU0715G, D9GU0716G, D9GU0750G, D9GU0751G, D9GU0752G, K8GU0028G, L0GU3052G, H9GU1102G, H9GU1137G, F9GU0969G, F9GU0970G, H9GU1127G, J9GU1352G, J9GU1353G, C9GU0583G, C9GU0584G, J0GU2839G, J0GU2840G, K8GU0019G, K8GU0020G, K9GU1390G, L8GU0137G, L8GU0180G, L8GU0182G, L8GU0183G, L8GU0184G, L8GU0185G, L8GU0186G, L8GU0196G, L8GU0197G, M8GU0328G, M8GU0329G, M8GU0338G, M8GU0345G, M8GU0346G, M8GU0362G, M9GU1679G, M9GU1680G, E9GU0869G, E9GU0870G, E9GU0875G, E9GU0876G, E9GU0877G, E9GU0878G, H9GU1196G, J9GU1339G, L9GU1539G, K8GU0045G, A9GU0413G, A9GU0414G, E0GU2227G, E0GU2239G, E0GU2242G, E0GU2264G, E0GU2271G, E0GU2272G, E0GU2273G, L0GU3036G, M0GU0095G, M0GU0096G, K0GU2908G, B0GU1918G, B0GU1919G, B1GU3224G, D9GU0719G, D9GU0720G, G9GU1024G, G9GU1025G, G9GU1026G, H9GU1205G, B9GU0529G, C9GU0621G, C9GU0622G, F9GU0963G, F9GU0964G, J0GU2880G, L0GU3026G, E0GU2246G, J0GU2720G, G0GU2485G, B9GU0499G, B9GU0500G, F0GU2362G, F0GU2363G, H9GU1101G, H9GU1151G, M9GU1151G, A1GU0230G, C0GU2003G, E0GU2318G, L9GU1547G, L9GU1548G, L9GU1549G, L9GU1550G, L9GU1551G, L9GU1569G, F9GU0929G, F9GU0930G, A0GU1873G, A0GU1874G, B0GU1981G, E0GU2206G, B1GU3220G, E0GU2285G, A1GU3189G, A9GU0365G, A9GU0366G, A9GU0367G, B1GU3270G, D0GU2128G, E0GU2207G, E0GU2208G, E0GU2209G, E0GU2210G, E0GU2211G, E0GU2212G, E0GU2213G, E0GU2217G, E0GU2218G, E0GU2322G, E0GU2323G, F0GU2401G, H9GU1174G, H9GU1175G, K0GU2925G, L9GU1545G, M9GU1728G, J0GU2742G, J0GU2743G, J0GU2744G, J0GU2745G, H9GU1227G, H9GU1228G, J9GU1328G, J9GU1329G, J9GU1363G, J9GU1369G, K9GU1429G, K9GU1435G, B9GU0451G, E0GU2255G, E0GU2345G, E0GU2346G, J0GU2850G, M8GU0351G, L8GU0133G, L8GU0134G, L8GU0135G, L8GU0136G, C9GU0581G, H9GU1139G, H9GU1157G, H9GU1158G, H9GU1159G, H9GU1160G, H9GU1161G, H9GU1162G, H9GU1163G, H9GU1164G, H9GU1165G, H9GU1166G, M8GU0228G, B0GU1976G, M8GU0359G, E9GU0859G, F9GU0911G, G9GU1047G, H0GU2717G, H0GU2719G, J0GU2721G, J0GU2722G, J0GU2723G, L8GU0170G, E0GU2288G, B0GU1927G, B0GU1936G, B0GU1937G, J0GU2836G, J0GU2837G, M8GU0243G, J0GU2789G, J9GU1350G, J9GU1351G, L9GU1600G, L9GU1601G, L9GU1602G, L9GU1603G, M0GU3155G, M0GU3156G, B9GU0501G, B9GU0502G, B9GU0503G, J9GU1356G, J9GU1362G, A0GU1813G, A0GU1823G, A0GU1824G, A0GU1825G, A0GU1826G, A0GU1827G, J0GU2854G, K0GU2927G, K0GU2928G, K0GU2929G, C9GU0620G, J9GU1315G, K8GU0116G, L8GU0160G, L8GU0161G, B9GU0455G, C9GU0633G, C9GU0634G, C9GU0638G, C9GU0639G, C9GU0646G, D9GU0709G, D9GU0710G, D9GU0711G, D9GU0717G, D9GU0746G, D9GU0747G, D9GU0748G, D9GU0749G, D9GU0758G, E9GU0855G, E9GU0856G, F9GU0899G, F9GU0900G, F9GU0901G, F9GU0902G, F9GU0903G, F9GU0904G, F9GU0905G, F9GU0906G, F9GU0965G, G9GU0646G, H0GU2619G, H0GU2665G, H9GU1090G, H9GU1100G, H9GU1203G, H9GU1204G, J9GU1259G, J9GU1334G, L8GU0227G, A0GU1894G, B9GU0485G, J9GU1371G, L8GU0201G, B9GU0443G, B9GU0527G, C9GU0655G, B9GU0431G, H0GU2695G, H0GU2696G, M8GU0342G, C9GU0663G, K9GU1420G, K9GU1421G, C9GU0566G, C9GU0568G, D9GU0760G, D9GU0761G, J9GU1316G, J9GU1335G, J9GU1336G, D9GU0725G, K9GU1494G, K9GU1495G, K9GU1496G, K9GU1497G, K9GU1498G, K9GU1499G, M8GU0298G, B1GU3242G, B9GU0531G, B9GU0534G, B9GU0535G, C1GU3321G, D9GU0703G, F0GU2402G, F0GU2403G, F0GU2479G, F9GU0944G, F9GU0988G, J0GU2892G, L9GU1511G, M0GU3092G, M0GU3093G, M0GU3103G, M9GU1677G, H9GU1193G, C1GU3313G, C1GU3314G, C9GU0585GR, D0GU2176G, F0GU2355G, F0GU2411G, F0GU2450G, H9GU1169G, H9GU1171G, C0GU2041G, D9GU0790G, K8GU0063G, A0GU1875G, A0GU1876G, A0GU1877G, G9GU1039G, A1GU3177G, A1GU3178G, B1GU3244G, B1GU3245G, H9GU1167G, H9GU1168G, H9GU1176G, H9GU1177G, H9GU1178G, J9GU1380G, L0GU3062G, L0GU3063G, M0GU3096G, M0GU3151G, M0GU3152G, K8GU0052G, M8GU0336G, C1GU3333G, E0GU2338G, L0GU3015G, L0GU3016G, L0GU3017G, L0GU3018G, M9GU1690G, D0GU2135G, G9GU1079G, D0GU2138G, E0GU2247G, E0GU2251G, H0GU2620G, A1GU3200G, K8GU0088G, K9GU1440G, K9GU1441G, L8GU0126G, L8GU0131G, A0GU1769G, A0GU1788G, A0GU1789G, A9GU0411G, B9GU0444G, H0GU2684G, K9GU1455G, M9GU1663G, M9GU1664G, M9GU1665G, K9GU1474G, K9GU1475G, M9GU1735G, M9GU1736G, M9GU1737G, M9GU1738G, M9GU1739G, M9GU1740G, M9GU1741G, J0GU2877G, M0GU3137G, B0GU1942G, L0GU3033G, L0GU3034G, A9GU0008G, A9GU0415G, A9GU0416G, A9GU0417G, B0GU1988G, B9GU0436G, B9GU0437G, B9GU0438G, C9GU0665G, D9GU0738G, H0GU2635G, H0GU2636G, C0GU2084G, C0GU2085G, G9GU1077G, K9GU1436G, D9GU0712G, D9GU0739G, E9GU0739G, A9GU0387G, A9GU0388G, A9GU0405G, A9GU0406G, E0GU2320G, C9GU0625G, C9GU0626G, C9GU0627G, C9GU0628G, H9GU1103G, H9GU1206G, K9GU1476G, K9GU1477G, L9GU1543G, M9GU1636G, J0GU2807G, J0GU2808G, J0GU2841G, J0GU2842G, M8GU0290G, M8GU0291G, M8GU0293G, M8GU0294G, A9GU0420G, B9GU0440G, F9GU0913G, E0GU2316G, B9GU0495G, E9GU0798G, F0GU2391G, F0GU2392G, F9GU0931G, G9GU1003G, G9GU1004G, G9GU1005G, G9GU1082G, H9GU1212G, E0GU2308G, J9GU1330G, K9GU1433G, M9GU1659G, M9GU1695G, A0GU1872G, H0GU2694G, H9GU1099G, K9GU1444G, K9GU1450G, H9GU1105G, F0GU2438G, F0GU2439G, F0GU2442G, F0GU2443G, F0GU2455G, F0GU2457G, A9GU0426G, B9GU0457G, B9GU0458G, B9GU0487G, C9GU0657G, L8GU0172G, L8GU0173G, M8GU0355G, G0GU2569G, G0GU2570G, J0GU2823G, J0GU2824G, A0GU1752G, A0GU1758G, A0GU1759G, A9GU0392G, A9GU0393G, A9GU0394G, A9GU0395G, A9GU0398G, A9GU0401G, B9GU0468G, B9GU0469G, B9GU0496G, B9GU0497G, B9GU0498G, B9GU0506G, B9GU0507G, B9GU0510G, B9GU0511G, B9GU0512G, B9GU0519G, B9GU0536G, B9GU0537G, C9GU0589G, D9GU0787G, D9GU788G, E0GU2286G, E9GU0821G, E9GU0848G, E9GU0849G, E9GU0850G, E9GU0851G, E9GU0852G, E9GU0853G, E9GU0854G, E9GU0857G, E9GU0865G, E9GU0881G, E9GU0882G, F9GU0908G, F9GU0920G, F9GU0921G, F9GU0922G, F9GU0923G, F9GU0925G, F9GU0926G, F9GU0996G, F9GU0997G, G9GU1013G, G9GU1014G, G9GU1015G, G9GU1016G, G9GU1017G, G9GU1019G, G9GU1020G, G9GU1037G, G9GU1061G, G9GU1919G, H9GU1121G, H9GU1122G, H9GU1123G, H9GU1124G, K8GU0048G, L8GU0147G, L8GU0165G, L8GU0211G, L8GU0212G, L8GU0217G, M8GU0244G, M8GU0245G, M8GU0246G, M8GU0247G, M8GU0248G, M8GU0249G, M8GU0250G, M8GU0264G, M8GU0265G, M8GU0266G, M8GU0267G, M8GU0268G, M8GU0269G, M8GU0270G, M8GU0271G, M8GU0272G, M8GU0273G, M8GU0274G, M8GU0276G, M8GU0277G, M8GU0278G, M8GU0279G, M8GU0280G, M8GU0281G, M8GU0285G, M8GU0286G, M8GU0292G, M8GU0295G, M8GU0296G, M8GU0297G, M8GU0299G, M8GU0300G, M8GU0332G, M8GU0347G, B1GU3274G, F0GU2373G, J9GU1368G, K0GU2965G, K9GU1432G, A0GU1878G, B9GU0523G, C1GU3318G, F0GU2357G, F0GU2358G, F0GU2395G, C9GU0015G, C9GU0661G, C9GU0662G, G9GU1028G, J0GU2814G, J0GU2815G, J0GU2817G, L9GU1552G, L9GU1553G, M8GU0321G, A0GU1848G, A0GU1849G, B0GU1940G, M8GU0275G, M0GU3121G, M0GU3122G, M0GU3124G, M0GU3125G, M0GU3126G, D0GU2107G, J0GU2857G, K9GU1502G, E0GU2205G, E9GU0840G, M9GU1714G, M9GU1715G, M9GU1716G, M9GU1717G, M9GU1718G, B9GU0517G, B0GU1950G, B0GU1951G, B0GU1952G, B0GU1953G, B0GU1954G, C0GU2089G, C0GU2090G, C0GU2091G, C0GU2092G, C0GU2093G, J0GU2811G, J0GU2812G, J9GU1377G, J9GU1378G, F0GU2432G, F0GU2433G, F0GU2434G, C0GU1999G, F9GU0095G, F9GU0955G, K8GU0024G, K8GU0025G, K8GU0057G, E9GU0842G, E9GU0847G, E9GU0956G, F9GU0956G, D9GU0707G, F9GU0973G, F9GU0974G, F9GU0975G, J9GU1264G, J9GU1265G, G9GU1027G, G9GU1029G, G9GU1030G, G9GU1031G, C1GU3329G, F9GU0907G, M8GU0323G, M8GU0324G, E0GU2276G, A0GU1865G, G0GU2524G, G0GU2542G, F9GU0966G, F9GU0967G, B0GU1959G, E0GU2233G, H9GU1189G, H9GU1190G, H9GU1191G, H9GU1192G, H9GU1992G, L9GU1624G, L9GU1625G, M9GU1652G, M9GU1691G, M9GU1694G, D9GU0701G, F0GU2426G, F0GU2427G, J0GU2835G, J0GU2838G, F0GU2353G, F0GU2354G, B9GU0538G, B9GU0539G, A0GU1835G, F0GU2474G, F9GU1000G, J0GU2816G, L8GU0190G, L8GU0191G, E1GU3373G, L8GU0194G, L8GU0195G, L8GU0220G, L8GU0221G, B9GU0447G, B0GU1973G, B0GU1974G, B0GU1975G, L8GU0188G, L8GU0189G, L8GU0198G, C9GU0669G, C9GU0670G, C9GU0671G, C9GU0672G, G9GU1075G, G9GU1076G, J0GU2891G, J9GU1252G, C9GU0551G, B0GU1967G, C9GU0610G, C9GU0611G, C9GU0612G, C9GU0615G, C9GU0616G, D0GU2174G, E0GU2244G, E0GU2248G, E0GU2282G, F0GU2430G, L8GU0193G, F0GU2451G, F0GU2367G, E9GU0803G, E9GU0808G, E9GU0814G, E9GU0871G, E9GU0872G, E9GU0873G, E9GU0874G, L9GU1577G, M0GU3135G, G9GU1044G, E1GU3372G, A0GU1866G, L8GU0204G, L8GU0205G, J9GU1305G, A9GU0377G, A9GU0378G, A9GU0379G, A9GU0380G, A9GU0381G, A9GU0382G, A9GU0385G, A9GU0386G, A9GU0396G, A9GU0397G, B9GU0528G, C9GU0673G, C9GU0674G, C9GU0675G, C9GU0676G, C9GU0680G, D9GU0741G, D9GU0754G, D9GU0759G, G0GU2486G, G0GU2487G, G0GU2513G, G9GU1089G, H9GU1093G, H9GU1201G, H9GU1202G, J9GU1359G, K9GU1427G, K9GU1428G, L8GU0200G, L9GU1517G, M8GU0333G, J9GU1256G, J9GU1257G, L9GU1542G, E9GU0816G, F9GU0982G, G9GU1058G, G9GU1059G, H9GU1210G, H9GU1211G, L9GU1525G, L9GU1558G, M9GU1710G, C0GU2050G, C0GU2051G, L0GU3041G, C0GU2029G, A0GU1841G, F0GU2470G, F0GU2473G, K8GU0023G, B0GU1906G, B0GU1907G, B0GU1908G, B0GU1909G, B0GU1910G, B0GU1911G, B0GU1912G, B0GU1913G, C0GU2031G, C0GU2086G, B0GU1930G, B0GU1932G, B0GU1933G, B0GU1966G, B0GU1968G, B1GU3273G, D0GU2121G, E9GU0797G, F0GU2429G, F0GU2477G, F9GU0950G, F9GU0951G, F9GU0958G, G0GU2604G, H0GU2650G, H0GU2651G, H0GU2667G, H0GU2668G, H0GU2669G, H9GU1138G, H9GU1215G, H9GU1216G, H9GU1217G, H9GU1218G, J9GU1307G, J9GU1308G, J9GU1309G, J9GU1310G, J9GU1327G, K0GU2955G, K9GU1423G, K9GU1425G, K9GU1430G, K9GU1431G, K9GU1467G, K9GU1468G, K9GU1469G, K9GU1470G, L0GU3066G, L0GU3067G, L9GU1554G, L9GU1606G, L9GU1607G, M8GU0357G, M8GU0358G, H0GU2677G, J0GU2728G, J0GU2787G, J0GU2788G, J0GU2821G, J0GU2822G, J0GU2827G, J0GU2828G, J0GU2829G, J0GU2830G, J0GU2832G, J0GU2833G, M0GU3147G, G0GU2595G, G0GU2611G, H0GU2685G, E0GU2200G, A0GU1750G, F0GU2393G, F0GU2394G, F0GU2404G, J0GU2810G, L0GU3012G, G0GU2582G, J9GU1343G, J9GU1344G, J9GU1345G, J9GU1347G, J9GU1348G, J9GU1349G, L9GU1618G, D9GU0691GR, H9GU1197G, J9GU1251G, L9GU1570G, F0GU2407G, F0GU2428G, J9GU1375G, K9GU1403G, K0GU2946G, A0GU1748G, A0GU1756G, A0GU1757G, A0GU1796G, A0GU1799G, A0GU1805G, A0GU1816G, A0GU1850G, A0GU1854G, A0GU1870G, H9GU1148G, H9GU1229G, H9GU1247G, H9GU1248G, H9GU1249G, H9GU1250G, J9GU1253G, J9GU1254G, J9GU1255G, J9GU1272G, J9GU1273G, J9GU1278G, J9GU1279G, J9GU1280G, K9GU1381G, K9GU1382G, K9GU1500G, L9GU1538G, L9GU1541G, L9GU1578G, L9GU1579G, L9GU1580G, L9GU1581G, M9GU1644G, M9GU1645G, M9GU1646G, M9GU1707G, M9GU1722G, M9GU1729G, C9GU0613G, L0GU2999G, K8GU0026G, B1GU3216G, B9GU0449G, D0GU2179G, D9GU0702G, E0GU2265G, F1GU3384G, L8GU0222G, L8GU0224G, L8GU0225G, M8GU0340G, M8GU0341G, M8GU0343G, H9GU1133G, L9GU1546G, C1GU3328G, G0GU2504G, G0GU2505G, C9GU0577G, C9GU0588G, E0GU2219G, E0GU2220G, J0GU2796G, J0GU2819G, E9GU0819G, E9GU0860G, E9GU0861G, E9GU0862G, E9GU0863G, E9GU0864G, C0GU2072G, E1GU3371G, G0GU2492G, G0GU2493G, G0GU2581G, H0GU2637G, J0GU2748G, J0GU2774G, J0GU2775G, J0GU2794G, J0GU2795G, J0GU2797G, J0GU2820G, C9GU0591G, C9GU0597G, E9GU0868G, G0GU2613G, H0GU2666G, G9GU1021G, G9GU1022G, D1GU3348G, D1GU3349G, E0GU2224G, E0GU2225G, E0GU2226G, B0GU1956G, H9GU1172G, H9GU1173G, H9GU1181G, H9GU1182G, H9GU1184G, H9GU1194G, H9GU1195G, K9GU1395G, K9GU1396G, K9GU1398G, K9GU1399G, K9GU1400G, K9GU1401G, K9GU1424G, C0GU2068G, J9GU1262G, J9GU1263G, K8GU0064G, K8GU0065G, K8GU0066G, K8GU0070G, K8GU0071G, K8GU0086G, K8GU0087G, K8GU0091G, K8GU0092G, K8GU0093G, K8GU0095G, L9GU1518G, L9GU1508G, L9GU1509G, E0GU2223G, E0GU2230G, E0GU2234G, E0GU2237G, E0GU2238G, E0GU2250G, F0GU2467G, M0GU3101G, M0GU3109G, M0GU3111G, M0GU3112G, L9GU1559G, L9GU1560G, L9GU1561G, L9GU1562G, L9GU1563G, L9GU1564G, A9GU0371G, A9GU0372G, A9GU0373G, A9GU0400G, A9GU0422G, A9GU0423G, A9GU0424G, B9GU0505G, C9GU0573G, D9GU0742G, D9GU0743G, M8GU0239G, M8GU0240G, M8GU0241G, M8GU0353G, M8GU0356G, C0GU2073G, G0GU2573G, B0GU1914G, B0GU1915G, C9GU0642G, C9GU0643G, D0GU2162G, E0GU2216G, H9GU1245G, K8GU0112GR, M9GU1661G, M9GU1662G, M9GU1671G, G0GU2500G, M8GU0301G, M8GU0302G, M8GU0304G, M8GU0322G, D0GU2178G, A0GU1775G, L9GU1520G, M0GU3134G, M0GU3143G, M8GU0242G, E0GU2328G, E0GU2330G, G0GU2503G, J9GU1370G, J9GU1374G, K9GU1422G, M8GU0229G, M8GU0232G, M8GU0233G, M8GU0234G, M8GU0235G, M8GU0236G, M8GU0237G, E9GU0815G, E9GU0992G, F9GU0992G, K8GU0049G, K8GU0050G, K8GU0082G, K8GU0084G, K8GU0085G, F9GU0971G, L9GU1594G, L9GU1595G, M0GU3145G, A0GU1885G, K8GU0079G, L9GU1582G, M9GU1649G, M9GU1650G, M9GU1672G, B0GU1969G, B0GU1970G, B0GU1971G, L0GU3005G, D9GU0791G, D9GU0792G, L0GU2987G, L0GU3035G, B0GU1916G, L9GU1587G, F0GU2374G, C9GU0555G, C9GU0556G, C9GU0560G, C9GU0561G, C9GU0562G, C9GU0563G, C9GU0565G, C9GU0598G, C9GU0677G, C9GU0678G, C9GU0679G, D9GU0696G, D9GU0697G, E9GU0884G, E9GU0885G, F0GU2435G, G9GU1034G, H9GU1112G, J0GU2890G, K0GU2938G, K0GU2983G, K9GU1397G, L9GU1557G, E0GU2342G, F0GU2416G, K9GU1451G, K9GU1452G, K9GU1453G, K9GU1456G, A9GU0390G, A9GU0391G, B0GU1920G, B0GU1921G, C1GU3319G, D1GU3347G, M8GU0363G, C9GU0607G, L0GU3072G, F0GU2408G, D9GU0722G, E0GU2198G, A0GU1751G, G0GU2601G, L9GU1585G, F9GU1001G, G0GU2589G, H0GU2699G, H0GU2700G, H0GU2701G, H0GU2702G, G0GU2574G, G0GU2575G, G0GU2576G, C0GU2083G, M9GU1675G, M9GU1676G, D0GU2108G, A0GU1801G, B0GU1986G, B0GU1987G, A9GU0418G, A9GU0425G, E9GU0846G, H9GU1136G, M8GU0348G, M9GU1670G, M8GU0330G, C9GU0599G, K0GU2918G, K9GU1411G, K9GU1412G, K9GU1454G, M0GU3153G, H9GU1183G, J9GU1318G, J0GU2740G, J0GU2741G, F9GU0961G, F9GU0962G, M0GU3157G, M0GU3158G, B0GU1960G, K8GU0099G, K8GU0106G, K8GU0107G, K9GU0106G, E0GU2253G, E0GU2254G, B1GU3250G, J9GU1372G, J9GU1373G, J9GU1376G, K9GU1413G, B9GU0522G, H0GU2692G, H0GU2693G, J9GU1266G, K8GU0047G, L8GU0192G, J0GU2780G, J0GU2806G, A0GU1809G, A0GU1810G, A0GU1830G, A0GU1831G, A0GU1840G, C0GU2040G, E0GU2278G, E0GU2317G, J9GU1298G, J9GU1299G, K9GU1442G, K9GU1443G, L9GU1528G, L9GU1529G, L9GU1530G, L9GU1531G, E9GUO890G, F0GU2469G, H9GU1239G, J9GU1286G, J9GU1287G, J9GU1288G, K8GU5049G, M9GU1638G, B9GU0453G, C9GU0592G, C9GU0593G, D9GU0744G, D9GU0745G, G9GU1041G, B1GU3248G, L9GU1626G, H9GU1222G, A0GU1814G, B9GU0486G, J9GU1303G, J9GU1304G, B9GU0518G, H9GU1115G, H9GU1116G, H9GU1117G, L8GU0226G, H9GU1223G, H9GU1243G, H9GU1244G, G0GU2520G, J0GU2738G, J0GU2739G, J0GU2879G, E0GU2275G, E0GU2295G, A0GU1845G, B0GU1902G, B0GU1903G, J9GU1364G, J9GU1365G, B9GU0442G, L8GU0202G, A9GU0383G, A9GU0384G, D0GU2131G, D0GU2132G, L9GU1535G, L9GU1536G, L9GU1537G, F9GU0984G, F9GU0985G, F9GU0916G, F9GU0960G, G0GU2515G, G0GU2516G, G0GU2517G, G9GU1007G, G9GU1008G, H0GU2676G, H0GU2683G, H9GU1220G, J0GU2777G, J0GU2798G, J0GU2799G, J0GU2800G, M0GU3106G, M0GU3107G, M0GU3108G, M0GU3127G, M9GU1704G, M9GU1705G, A9GU0404G, C0GU2046G, C0GU2047G, C0GU2048G, C0GU2049G, C0GU2059G, C0GU2060G, C1GU3330G, E0GU2231G, E0GU2232G, E9GU0828G, E9GU0829G, G9GU1060G, H9GU1125G, H9GU1126G, J0GU2858G, J0GU2876G, K0GU2954G, L9GU1604G, L9GU1605G, M9GU1706G, G9GU1054G, G9GU1055G, G9GU1056G, G9GU1057G, H0GU2655G, H0GU2656G, H0GU2657G, H0GU2658G, H0GU2715G, H0GU2716G, J0GU2844G, J0GU2845G, J9GU1274G, J9GU1300G, J9GU1317G, J9GU1366G, K8GU0072G, K8GU0073G, K8GU0075G, K8GU0076G, K9GU1383G, C0GU2080G, C0GU2081G, C0GU2082G, M8GU0334G, M8GU0335G, C1GU3315G, M9GU1732G, D0GU2110G, K8GU0046G, B9GU0543G, B9GU0544G, D0GU2119G, D0GU2120G, A0GU1855G, A0GU1856G, A0GU1857G, A0GU1858G, A0GU1859G, A0GU1860G, C0GU2020G, F9GU0947G, B9GU0470G, B9GU0471G, M8GU0230G, C9GU0550G, C9GU0554G, E9GU0886G, B0GU1955G, B0GU1972G, K8GU0053G, K8GU0054G, K8GU0055G, F9GU0954G, H9GU1107G, D0GU2143G, D9GU0718G, D9GU0762G, D9GU7186G, L9GU1627G, L0GU3040G, B0GU1943G, B0GU1944G, H0GU2671G, K0GU2934G, K0GU2935G, K0GU2978G, A9GU0369G, A9GU0374G, L9GU1544G, L9GU1574G, L9GU1575G, L9GU1576G, L9GU1615G, L9GU1616G, B1GU3251G, G0GU2612G, B9GU0520G, C9GU0558G, C9GU0559G, C9GU0567G, G0GU2596G, G0GU2597G, F9GU0928G, F9GU0933G, G9GU1040G, M9GU1724G, D0GU2163G, B0GU1904G, B0GU1905G, B9GU0462G, B9GU0463G, H0GU2688G, H0GU2689G, H0GU2690G, J1GU3406G, J9GU1281G, K8GU0034G, K8GU0035G, K8GU0036G, K8GU0037G, M9GU1731G, B9GU0448G, E9GU0827G, M9GU1702G, M9GU1703G, F9GU0968G, C9GU0645G, F9GU0978G, H9GU1226G, A9GU0421G, B9GU0509G, B9GU0513G, F0GU2436G, F0GU2437G, G0GU2544G, G0GU2586G, H9GU1104G, H9GU1132G, H9GU1208G, J9GU1277G, J9GU1301G, K9GU1503G, D9GU0704G, D9GU0740G, E9GU0820G, A0GU1766G, F0GU2406G, M9GU1648G, B0GU1917G, F0GU2349G, F0GU2365G, F0GU2366G, M9GU1681G, M9GU1682G, B9GU0480G, B9GU0484G, B9GU0508G, B9GU0526G, C9GU0549G, C9GU0553G, D9GU0776G, D9GU0777G, D9GU0778G, L0GU3014G, H0GU2649G, H9GU1106G, J0GU2873G, D9GU0706G, E9GU0841G, A0GU1745G, A0GU1746G, A0GU1747G, A0GU1851G, A0GU1852G, A0GU1853G, B1GU3221G, M9GU1692G, M9GU1693G, K8GU0027G, J0GU2852G, J0GU2831G, A0GU1895G, C9GU0552G, J0GU2715G, J0GU2783G, J0GU2785G, J0GU2T84G, M8GU0326G, M8GU0327G, J0GU2782G, J0GU2784G, M9GU1635G, C9GU0590G, D9GU0683G, D9GU0684G, D9GU0705G, G0GU2523G, G9GU1032G, G9GU1033G, J0GU2843G, H0GU2675G, L9GU1532G, E9GU0858G, L8GU0168G, L8GU0169G, J9GU1258G, C0GU2030G, M0GU3105G, A9GU0403G, A9GU0407G, A9GU0408G, A9GU0412G, B0GU1939G, A0GU1808G, A0GU1832G, A0GU1833G, A0GU1834G, C1GU3323G, C9GU0636G, C9GU0637G, G9GU1063G, J0GU2834G, J9GU1314G, L8GU0171G, L9GU1555G, L9GU1556G, M0GU3115G, M0GU3116G, M0GU3117G, M0GU3136G, M8GU0325G, K9GU1437G, K9GU1439G, A9GU0402G, C1GU3320G, E9GU0839G, H0GU2632G, J9GU1379G, B9GU0493G, B9GU0494G, J0GU2863G, J0GU2864G, J9GU1282G, J9GU1283G, J9GU1284G, J9GU1285G, J9GU1289G, J9GU1290G, J9GU1291G, J9GU1292G, J9GU1293G, J9GU1294G, J9GU1295G, J9GU1296G, J9GU1297G, K8GU0074G, K8GU0080G, K8GU0081G, K8GU0094G, K8GU0096G, K8GU0097G, K8GU0100G, K8GU0101G, K8GU0102G, K8GU0104G, K8GU0105G, K8GU0108G, K8GU0109G, K8GU0110G, K8GU0111G, K8GU0112G, K8GU0114G, K8GU0115G, K8GU0117G, K8GU0118G, L8GU0123G, L8GU0127G, L8GU0128G, L8GU0129G, L8GU0130G, L8GU0140G, L8GU0142G, L8GU0143G, L8GU0144G, L8GU0145G, L8GU0148G, L8GU0149G, L8GU0150G, L8GU0151G, L8GU0153G, L8GU0154G, L8GU0157G, L8GU0206G, L9GU1586G, M8GU0305G, M8GU0306G, M8GU0307G, M8GU0308G, M8GU0309G, M8GU0310G, M8GU0311G, M8GU0314G, M8GU0315G, M8GU0316G, M8GU0317G, M8GU0318G, M8GU0319G, M8GU0320G, M9GU1712G, A0GU1802G, C9GU0618G, G0GU2531G, G0GU2534G, G0GU2536G, G0GU2537G, G0GU2566G, G0GU2600G, H0GU2621G, H0GU2623G, H0GU2624G, H0GU2629G, H0GU2633G, J9GU1355G, L8GU0199G, L1GU0154G, L8GU0155G, B9GU0445G, B9GU0435G, B9GU0441G, G0GU2483G, H0GU2618G, C1GU3322G, M0GU3146G, E9GU0826G, G0GU2538G, G0GU2539G, G0GU2540G, G0GU2541G, K0GU2982G, M8GU0331G, G9GU1088G, F1GU3399G, J0GU2886G, A1GU3199G, C0GU2037G, C0GU2038G, C0GU2039G, C1GU3316G, C9GU0576G, D0GU2134G, E0GU2314G, M0GU3102G, M0GU3110G, M0GU3131G, M0GU3132G, M0GU3133G, A1GU3167G, A1GU3168G, B1GU3253G, C0GU2036G, C0GU2042G, C0GU2062G, C0GU2063G, C0GU2064G, C1GU3326G, C1GU3327G, C1GU3332G, D0GU2189G, D9GU0770G, D9GU0771G, D9GU0772G, D9GU0773G, D9GU0780G, D9GU0781G, D9GU0782G, D9GU0783G, D9GU0784G, D9GU0785G, E0GU2343G, F9GU0917G, G0GU2490G, G0GU2491G, K0GU2923G, K0GU2924G, K0GU2933G, K9GU1464G, K9GU1465G, K9GU1466G, L8GU0174G, L8GU0175G, L8GU0176G, L8GU0177G, L8GU0178G, L8GU0179G, L8GU0181G, M9GU1666G, M9GU1708G, G9GU1062G, G9GU1065G, G9GU1067G, G9GU1081G, G9GU1083G, G9GU1084G, G9GU1085G, G9GU1086G, C0GU2065G, G0GU2484G, A0GU1867G, A0GU1868G, D9GU0713G, G0GU2526G, G0GU2527G, G0GU2528G, G0GU2529G, G0GU2530G, H0GU2 686G, H0GU2687G, L0GU2989G, D0GU2140G, D0GU2141G, D0GU2142G, B0GU1949G, B1GU3210G, B1GU3211G, B1GU3212G, B1GU3213G, B1GU3214G, B1GU3215G, B1GU3229G, B1GU3231G, B1GU3233G, B1GU3238G, B1GU3239G, B1GU3240G, B1GU3241G, B1GU3243G, B1GU3246G, B1GU3247G, B9GU0483G, B9GU0514G, B9GU0515G, B9GU0516G, B9GU0521G, B9GU0540G, B9GU0541G, B9GU0542G, B9GU0547G, B9GU0548G, C9GU0575G, D9GU0763G, E0GU2260G, E0GU2347G, E1GU3352G, E1GU3353G, E1GU3354G, E1GU3355G, E1GU3356G, E1GU3357G, E1GU3358G, E9GU0834G, F0GU2468G, F9GU0939G, F9GU0940G, F9GU0942G, F9GU0943G, G0GU2603G, H0GU2638G, J0GU2847G, J0GU2848G, J0GU2849G, J0GU2855G, J0GU2856G, J0GU2859G, J0GU2860G, J0GU2865G, J0GU2866G, J0GU2868G, J0GU2869G, J0GU2870G, J0GU2871G, J0GU2872G, K8GU0032G, K8GU0033G, K8GU0040G, K8GU0041G, K8GU0042G, K8GU0043G, K8GU0044G, M8GU0364G, L8GU0162G, L8GU0164G, A0GU1842G, A0GU1846G, A0GU1869G, A0GU1888G, A0GU1887G, C0GU2043G, C0GU2044G, C0GU2077G, L0GU3011G, L9GU1510G, G0GU2511G, B9GU0450G, C9GU0569G, D1GU3343G, K8GU0038G, L0GU3049G, L0GU3053G, A0GU1770G, A0GU1771G, A0GU1871G, A9GU0409G, A9GU0410G, E9GU0887G, E9GU0888G, E9GU0889G, E9GU1001G, F9GU0957G, F9GU0980G, G0GU2585G, H9GU1149G, H9GU1150G, K0GU2977G, L0GU2998G, L9GU1612G, M0GU3089G, J9GU1332G, J9GU1333G, K9GU1472G, K9GU1473G, F0GU2461G, M8GU0251G, D9GU0753G, F9GU0945G, F9GU0946G, K8GU0056G, A0GU1829G, D0GU2137G, J0GU2753G, F9GU0948G, F9GU0977G, K9GU1449G, E9GU0817G, E9GU0818G, A9GU0389G, C9GU0653G, D9GU0726G, D9GU0727G, D9GU0728G, D9GU0729G, D9GU0730G, D9GU0731G, D9GU0732G, D9GU0733G, D9GU0734G, D9GU0735G, D9GU0736G, D9GU0737G, D9GU0789G, E9GU0880G, E9GU1880G, L8GU0158G, L8GU0159G, L8GU0163G, L8GU0213G, L8GU0214G, L8GU0215G, L8GU0216G, L8GU0218G, L8GU0219G, M8GU0253G, M8GU0254G, M8GU0255G, M8GU0256G, M8GU0257G, M8GU0258G, M8GU0259G, M8GU0260G, M8GU0261G, M8GU0262G, M8GU0263G, J9GU1302G, E9GU0867G, L9GU1617G, D9GU0689G, E0GU2197G, K8GU0060G, K8GU0061G, K8GU0062G, K8GU0068G, K8GU0069G, B0GU1983G, A9GU0399G, B0GU1900G, B1GU3252G, B9GU0533G, B9GU0545G, B9GU0546G, F0GU2359G, F0GU2360G, F9GU0959G, F9GU1002G, G0GU2501G, J0GU2776G, K8GU0077G, K8GU0078G, J0GU2825G, J0GU2888G, B9GU0446G, F0GU2431G, L8GU0124G, L8GU0125G, L8GU0152G, M8GU0337G, M8GU0349G, C1GU3334G, H0GU2697G, C9GU0640G, C9GU0641G, C9GU0647G, C9GU0650G, C9GU0651G, C9GU0652G, D9GU0688G, E0GU2240G, F9GU0999G, K8GU0098G, L8GU0207G, L8GU0208G, L8GU0209G, L8GU0210G, M8GU0252G, J0GU2853G, L9GU1547GR, E9GU0801G, J0GU2818G, J0GU2846G, J0GU2851G, L0GU3051G, B1GU3222G, K9GU1405G, H0GU2674G, D0GU2127G, G0GU2545G, G0GU2546G, G0GU2547G, G0GU2548G, G0GU2549G, G0GU2550G, G0GU2551G, G0GU2552G, G0GU2553G, G0GU2554G, G0GU2555G, G0GU2556G, G0GU557G, B1GU3223G, A0GU1749G, C9GU0595G, C9GU0596G, D9GU0692G, D9GU0693G, D9GU0694G, K8GU0058G, K8GU0059G, C0GU2054G, C0GU2056G, C0GU2057G, C0GU2058G, C1GU3281G, M0GU3104G, A0GU1806G, A0GU1807G, E9GU0844G, E0GU2201G, E0GU2202G, E0GU2203G, E0GU2305G, E0GU2306G, E0GU2307G, F0GU2458G, F0GU2459G, F0GU2460G, F0GU2462G, F0GU2463G, F0GU2464G, F0GU2465G, F9GU0989G, F9GU0990G, F9GU0991G, H9GU1091G, H9GU1092G, H9GU1094G, H9GU1095G, K9GU1384G, K9GU1407G, K9GU1408G, K9GU1409G, L9GU1094G, M9GU1721G, D0GU2111G, D0GU2112G, D0GU2113G, D0GU2114G, D0GU2118G, B9GU0472G, B9GU0473G, B9GU0474G, B9GU0475G, B9GU0476G, B9GU0477G, B9GU0478G, K0GU2980G, C0GU2074G, C9GU0574G, D0GU2184G, D0GU2185G, A0GU1881G, A0GU1882G, B9GU0530G, B9GU0532G, D8GU0295G, F9GU0927G, F9GU0934G, F9GU0935G, H9GU1152G, L9GU1596G, L9GU1597G, L9GU1598G, L9GU1599G, E0GU2204G, C1GU3335G, D0GU2129G, E8GU0806G, E9GU0809G, E9GU0822G, E9GU0823G, F1GU3387G, G0GU2499G, G0GU2567G, G0GU2568G, G0GU2579G, G9GU1035G, K9GU1404G, G9GU1073G, G9GU1074G, K8GU0089G, K8GU0090G, L0GU2986G, L0GU2988G, K9GU1471G, C0GU2021G, C0GU2088G, L0GU3000G, L0GU3001G, L0GU3002G, E0GU2269G, E0GU2270G, E0GU2235G, E0GU2236G, M9GU1699G, M9GU1700G, A9GU0368G, M0GU3074G, G9GU1009G, J0GU2746G, M8GU0282G, M8GU0283G, M8GU0288G, B1GU3276G, G0GU2506G, C0GU2055G, F0GU2466G, C0GU2052G, C0GU2070G, C0GU2071G, B0GU1961G, B0GU1964G, K9GU1416G, D9GU0779G, A0GU1768G, B0GU1948G, L9GU1588G, L9GU1589G, L9GU1590G, L9GU1591G, L9GU1592G, L9GU1593G, A0GU1862G, B0GU1938G, E9GU0793G, E9GU0824G, G0GU2602G, M9GU1673G, H9GU1134G, A0GU1779G, H9GU1246G, J9GU1324G, J9GU1342G, J9GU1325G, G0GU2521G, G0GU2563G, G0GU2564G, G0GU2565G, G0GU2584G, G0GU2598G, J0GU2726G, F0GU2441G, B0GU1957G, B1GU3230G, E0GU2341G, F0GU2405G, G0GU2482G, B1GU3236G, E9GU0813G, A0GU1822G, H0GU2713G, M0GU3099G, M9GU1687G, M9GU1688G, G0GU2592G, G0GU2593G, L9GU1629G, J9GU1361G, M0GU3119G, J0GU2805G, J9GU1360G, C0GU2076G, F0GU2471G, F0GU2472G, F0GU2422G, F0GU2423G, F0GU2444G, G0GU2509G, J0GU2813G, L0GU3060G, C9GU0594G, G0GU2518G, J0GU2804G, J0GU2764G, F0GU2400G, D0GU2105G, J0GU2803G, A0GU1787G, F0GU2419G, F0GU2415G, M0GU3130G, H9GU1200G, J0GU2763G, J0GU2765G, E1GU3360G, G0GU2594G, G0GU2599G, J0GU2762G, K0GU2942G, G0GU2591G, J0GU2766G, B0GU1935G, F0GU2410G, J0GU2737G, M0GU3129G, G0GU2519G, C0GU2005G, K0GU2974G, J0GU2767G, J0GU2769G, F0GU2413G, J0GU2768G, F0GU2414G, F0GU2412G, J0GU2786G, A9GU373G, C1GU3317G, G0GU2497G, J0GU2752G, F9GU0937G, H0GU2714G, H9GU1224G, F0GU2371G, H9GU1119G, K9GU1419G, G9GU1042G, G0GU2590G, M9GU1683G, G0GU2496G, G0GU2588G, A0GU1820G, H0GU2663G, H9GU1118G, K0GU29230G, F9GU0976G, H0GU2707G, H0GU2708G, D0GU2147G, B0GU1995G, H0GU2628G, E0GU2333G, E0GU2334G, B1GU3237G, H0GU2679, H0GU2679G, H0GU2680G, H9GU1120G, B0GU2193G, D0GU2193G, A1GU3191G, A9GU0423, C9GU0571G, D0GU2102G, D0GU2139G, E0GU2335G, H0GU1818G, J9GUI319G, D0GU2101G, A1GU3190G, K0GU2936G, F9GU0938G, G8GU1023G, G9GU1023G, J0GU2750G, F0GU2385G, B0GU1991G, B0GU1993G, K0GU2919G, B1GU3235G, B0GU1995, F0GU2386G, C9GU0570G, H9GU1236G, H9GU1237G, L9GU1516G, M9GU1685G, H9GU1144G, A0GU1765G, D0GU2151G, D0GU2152G, D0GU2166G, D0GU2167G, D0GU2168G, C0GU2014G, B0GU1996G, C1GU3297G, A0GU1776G, A0GU1828G, A0GU4128G, D9GU0695G, H0GU2626G, H0GU2627G, A0GU1777G, J9GU1337G, J9GU1338G, L0GU2985G, M8GU0238G, B0GU1931G, B1GU3217G, C1GU3295G, E0GU2214G, F0GU1486G, F9GU0124G, F9GU0912G, G0GU2558G, G0GU2559G, M0GU3080G, G9GU1012G, C0GU2045G, C0GU2027G, M8GU0352G, C9GU0557G, C9GU0564G, G0GU2562G, D0GU2148G, L8GU0119G, L8GU0120G, D9GU0687G, K0GU2807G, K0GU2907G, B9GU0465G, G0GU2606G, M0GU3077G, D0GU2115G, G0GU2508G, L0GU3054G, L9GU1521G, M8GU0344G, F9GU0993G, K9GU1479G, K9GU1483G, K9GU1484G, K9GU1485G, K9GU1486G, K9GU1488G, K9GU1490G, K9GU1491G, L9GU1512G, L9GU1513G, L9GU1515G, L8GU0138G, B0GU1998G, C0GU2001G, C0GU2004G, C0GU2008G, H0GU2642G, E0GU2279G, D9GU0723G, D9GU0724G, K9GU1493G, K9GU1489G, M9GU1696G, C1GU3296G, J0GU2861G, L9GU1504G, K9GU1463G, C9GU0600G, C9GU0601G, C9GU0602G, C9GU0603G, C9GU0604G, C9GU0605G, C9GU0606G, D9GU0708G, E9GU0795G, M0GU3085G, L0GU3044G, L0GU3045G, K8GU0009G, K8GU0010G, K8GU0011G, K8GU0012G, K8GU0013G, K8GU0014G, C0GU2099G, L9GU1572G, H0GU2643G, L8GU0139G, D9GU0765G, K0GU2964G, L0GU3042G, L0GU3043G, A1GU3161G, L8GU0132G, M8GU0361G, A1GU3164G, A1GU3195G, B1GU3226G, J9GU12366, A0GU2641G, M0GU3082G, B9GU0466G, L9GU1571G, M8GU0360G, K0GU2963G, K8GU0039G, F9GU0804G, L9GU1533G, F0GU2388G, B0GU1984G, B0GU1985G, C0GU2013G, J9GU1311G, J9GU1312G, J9GU1313G, B9GU0504G, D9GU0764G, K0GU2961G, G0GU2561G, J9GU1340G, K8GU0015G, A0GU1890G, M9GU1719G, E0GU2280G, L0GU2995G, J0GU2862G, L9GU1573G, M0GU3083G, D0GU2103G, D9GU0714G, C9GU0582G, F9GU0932G, K0GU2906G, D0GU2149G, C0GU2009G, A1GU3162G, G0GU2577G, G0GU2578G, H0GU2639G, H0GU2640G, H0GU2644G, H0GU2645G, G0GU2007G, L9GU1524G, E0GU2281G, C0GU2096G, D0GU2116G, F0GU2356G, A0GU1763G, G9GU1011G, H9GU1111G, A1GU3150G, A1GU3159G, D0GU2124G, M0GU3084G, D0GU2150G, H9GU1145G, B1GU3263G, A1GU3160G, A1GU3194G, L9GU1506G, L0GU3024G, A1GU3179G, A1GU3192G, B1GU3225G, A0GU1762G, D1GU3339G, A0GU1864G, A0GU1883G, J0GU2893G, L9GU1565G, M9GU1639G, M9GU1640G, M9GU1660G, M9GU1674G, K0GU2971G, L0GU3028G, J1GU3410G, J1GU3409G, F1GU3402G, E1GU3364G, F1GU3378G, J9GU1270G, H9GU1185G, J1GU3412G, G0GU2510G, A0GU1792G, H0GU2681G, B0GU1962G, K0GU2932G, A0GU1774G, E1GU3370G, E9GU0883G, L0GU3039G, B1GU3228G, L9GU1526G, B0GU1946G, J0GU2773G, E0GU2268G, J0GU2770G, J0GU2771G, J0GU2772G, J0GU2778G, J0GU2895G, K0GU2897G, K0GU2898G, K0GU2896G, J9GU1346G, H0GU2664G, L0GU3050G, G9GU1045G, L9GU1610G, B1GU3275G, H0GU2691G, L0GU3061G, A0GU1811G, A0GU1812G, A1GU3165G, D0GU2117G, A9GU0370G, F9GU0949G, K0GU2981G, C0GU2026G, B9GU0479G, B9GU0482G, A1GU3208G, A0GU17556, A0GU1755G, C0GU2094G, M0GU3091G, A1GU3203G, A1GU3176G, L8GU0121G, K9GU1457G, C1GU3311G, C1GU3285G, J9GU1275G, A9GU0429G, C0GU2025G, H9GU1180G, B0GU1958G, H9GU1154G, D9GU0700G, A0GU1880G, K0GU2948G, K0GU2949G, A1GU3207G, K0GU2947G, K0GU2950G, F0GU2390, F0GU2390G, A0GU1754G, H9GU1155G, D1GU3350G, D1GU3351G, A0GU1800G, F0GU2369G, M0GU3139G, M0GU3140G, M0GU3141G, M0GU3142G, J9GU1326G, L0GU3058G, L0GU3059G, E1GU3369G, C9GU0619G, E9GU0796G, A9GU0428G, A0GU1892G, A0GU1893G, K0GU2926G, D1GU3342G, K9GU1406G, J0GU2725G, M0GU1733G, B1GU3267G, J0GU2730G, J0GU2781G, L0GU3048G, F0GU2368G, F0GU2375G, F0GU2376G, D9GU0685G, D9GU685G, M8GU0303G, J0GU2826G, K9GU1410G, H0GU2711G, L8GU0203G, B9GU0492G, L0GU3071G, H0GU2659G, L0GU3164G, G0GU2535G, M0GU3154G;    4. Product Code: SAH, UDI Number, Serial Numbers: SAH06432207GA, SAH06442214GA, SAH06442222GA, SAH06181833GA, SAH06241925GA, SAH04350057GA, SAH08423888GA, SAH08423889GA, SAH08423891GA, SAH08423893GA, SAH08423894GA, SAH08423899GA, SAH08423904GA, SAH07052497GA, SAH05160826GA, SAH07052500GA, SAH05110701GA, SAH05140732GA, SAH05140738GA, SAH05140739GA, SAH05140744GA, SAH06392094GA, SAH06392095GA, SAH06392096GA, SAH06392097GA, SAH06392098GA, SAH06392099GA, SAH06392100GA, SAH06392101GA, SAH06392102GA, SAH06392103GA, SAH05491340GA, SAH05491341GA, SAH05491342GA, SAH05491345GA, SAH05491346GA, SAH05501387GA, SAH05501388GA, SAH06141763GA, SAH04420229GA, SAH04470318GA, SAH04470322GA, SAH05250895GA, SAH06432188GA, SAH06432191GA, SAH05431146GA, SAH05010520GA, SAH05010523GA, SAH05160823GA, SAH08023359GA, SAH05150795GA, SAH05150796GA, SAH05150798GA, SAH05150800GA, SAH05150801GA, SAH05150802GA, SAH05150803GA, SAH05160812GA, SAH05160819GA, SAH05431151GA, SAH05431152GA, SAH05431153GA, SAH06021460GA, SAH06452257GA, SAH06462294GA, SAH06462296GA, SAH06462300GA, SAH06462306GA, SAH06462307GA, SAH06462310GA, SAH06462312GA, SAH04460304GA, SAH07403012GA, SAH07403013GA, SAH07403014GA, SAH07403016GA, SAH07403040GA, SAH07403043GA, SAH07403059GA, SAH04420221GA, SAH04420233GA, SAH05220863GA, SAH05290966GA, SAH08383841GA, SAH08403876GA, SAH05300984GA, SAH07142623GA, SAH06141736GA, SAH06141738GA, SAH06141740GA, SAH06141742GA, SAH06141743GA, SAH08303681GA, SAH08313695GA, SAH05040568GA, SAH05511395GA, SAH08023361GA, SAH08443929GA, SAH05010515GA, SAH04360061GA, SAH04360065GA, SAH04360068GA, SAH04410185GA, SAH04410189GA, SAH04410190GA, SAH04410192GA, SAH04410194GA, SAH04410197GA, SAH04410198GA, SAH04410201GA, SAH04410202GA, SAH04470314GA, SAH04490379GA, SAH05250899GA, SAH05250900GA, SAH05250902GA, SAH06522422GA, SAH06522423GA, SAH06522424GA, SAH07212733GA, SAH07212740GA, SAH07212742GA, SAH07212749GA, SAH07212750GA, SAH07212751GA, SAH07372991GA, SAH07372992GA, SAH07403035GA, SAH07493255GA, SAH06051517GA, SAH06061525GA, SAH06071545GA, SAH06071546GA, SAH06071547GA, SAH07142624GA, SAH08203506GA, SAH08203528GA, SAH08203532GA, SAH08213559GA, SAH08253589GA, SAH08253594GA, SAH08253595GA, SAH08253596GA, SAH08253598GA, SAH08253599GA, SAH08253601GA, SAH08253602GA, SAH08253603GA, SAH08253604GA, SAH08373771GA, SAH08423892GA, SAH08423898GA, SAH04490380GA, SAH06191871GA, SAH04530498GA, SAH05010512GA, SAH05010521GA, SAH05260922GA, SAH05290956GA, SAH08373815GA, SAH05250897GA, SAH07423092GA, SAH07423093GA, SAH04500399GA, SAH04500404GA, SAH04500405GA, SAH04500408GA, SAH05080643GA, SAH05260914GA, SAH05260920GA, SAH05260921GA, SAH07372943GA, SAH07372946GA, SAH07372951GA, SAH07372954GA, SAH07372958GA, SAH07372960GA, SAH05040565GA, SAH07202702GA, SAH08283634GA, SAH08293663GA, SAH08313688GA, SAH08313694GA, SAH05140729GA, SAH07062545GA, SAH06021455GA, SAH08494058GA, SAH08514087GA, SAH04490361GA, SAH09024176GA, SAH05471247GA, SAH05471248GA, SAH05471250GA, SAH05471279GA, SAH05471282GA, SAH07292841GA, SAH07493281GA, SAH08373775GA, SAH08373780GA, SAH05321006GA, SAH05511397GA, SAH07172678GA, SAH07172679GA, SAH07172680GA, SAH07172681GA, SAH07202727GA, SAH07202728GA, SAH07202729GA, SAH08353729GA, SAH08383838GA, SAH08423923GA, SAH05070619GA, SAH05070624GA, SAH05491359GA, SAH08143430GA, SAH08143437GA, SAH06191864GA, SAH06191868GA, SAH06191869GA, SAH07312885GA, SAH08253617GA, SAH08283639GA, SAH08283641GA, SAH08283643GA, SAH08293659GA, SAH08313693GA, SAH04520471GA, SAH04530487GA, SAH05441184GA, SAH07172673GA, SAH07172674GA, SAH07202723GA, SAH07202724GA, SAH07202725GA, SAH07232763GA, SAH07232764GA, SAH07232765GA, SAH07232766GA, SAH07232767GA, SAH07232771GA, SAH07232772GA, SAH07232776GA, SAH07232777GA, SAH07232778GA, SAH07232779GA, SAH07232780GA, SAH07232781GA, SAH07232782GA, SAH07232783GA, SAH07232786GA, SAH07232787GA, SAH07232788GA, SAH07232791GA, SAH07292832GA, SAH04510440GA, SAH05280935GA, SAH07372953GA, SAH07372999GA, SAH04410204GA, SAH08443940GA, SAH06171782GA, SAH07142639GA, SAH07102585GA, SAH07102586GA, SAH06382064GA, SAH06382068GA, SAH06382069GA, SAH06382072GA, SAH05010530GA, SAH05290958GA, SAH06412135GA, SAH06412141GA, SAH04410186GA, SAH04410196GA, SAH04420231GA, SAH04430242GA, SAH04450265GA, SAH04450276GA, SAH04460301GA, SAH04500393SP, SAH07052505GA, SAH07142633GA, SAH07152651GA, SAH07202721GA, SAH06452250GA, SAH07072554GA, SAH07072561GA, SAH07172676GA, SAH05010517GA, SAH05010537GA, SAH05010555GA, SAH07403047GA, SAH07443143GA, SAH08343707GA, SAH08353736GA, SAH08353737GA, SAH08353739GA, SAH08353743GA, SAH08363745GA, SAH08363747GA, SAH08363753GA, SAH08524133GA, SAH06382065GA, SAH07372998GA, SAH07403027GA, SAH08023331GA, SAH08023333GA, SAH05160818GA, SAH05521405GA, SAH05521413GA, SAH08133423GA, SAH08173477GA, SAH06522393GA, SAH07212753GA, SAH07172687GA, SAH07172688GA, SAH07172689GA, SAH07172690GA, SAH07292836GA, SAH07292839GA, SAH05220843GA, SAH05010538GA, SAH05150776GA, SAH05150778GA, SAH05150799GA, SAH05160806GA, SAH06502356GA, SAH06502357GA, SAH07012432GA, SAH07012434GA, SAH07012439GA, SAH07012450GA, SAH07012451GA, SAH07302877GA, SAH07302879GA, SAH07302881GA, SAH07443145GA, SAH06332016GA, SAH06332022GA, SAH08143425GA, SAH08173482GA, SAH05381046GA, SAH05381065GA, SAH04410193GA, SAH04500385GA, SAH04500386GA, SAH04500387GA, SAH04500389GA, SAH04500390GA, SAH04500391GA, SAH06241958GA, SAH06241963GA, SAH06241906GA, SAH07342937GA, SAH07372945GA, SAH07372959GA, SAH07372994GA, SAH07372995GA, SAH07372996GA, SAH07372997GA, SAH04450260GA, SAH04450266GA, SAH04450275GA, SAH06291978GA, SAH06462297GA, SAH06462313GA, SAH07292823GA, SAH07292824GA, SAH07292825GA, SAH07292828GA, SAH07443129GA, SAH07443135GA, SAH08423903GA, SAH04510436GA, SAH04510437GA, SAH05110710GA, SAH05140733GA, SAH05160810GA, SAH06181828GA, SAH06181850GA, SAH06191854GA, SAH06382079GA, SAH06382090GA, SAH06382091GA, SAH06382092GA, SAH06402120GA, SAH07423090GA, SAH08423914GA, SAH06382085GA, SAH07052504GA, SAH07142625GA, SAH05321004GA, SAH06241937GA, SAH06241938GA, SAH06241939GA, SAH08143427GA, SAH08143428GA, SAH08143439GA, SAH05451205GA, SAH05451207GA, SAH05451211GA, SAH05451213GA, SAH05451220GA, SAH06412134GA, SAH07152656GA, SAH07102599GA, SAH08203525GA, SAH08203527GA, SAH08203530GA, SAH08203531GA, SAH08213549GA, SAH08494007GA, SAH08494008GA, SAH06091593GA, SAH07152663GA, SAH07152664GA, SAH05491351GA, SAH06241966GA, SAH07372948GA, SAH08033386GA, SAH08203521GA, SAH08353735GA, SAH08353738GA, SAH08353740GA, SAH08353741GA, SAH08353742GA, SAH08353744GA, SAH08363748GA, SAH08363757GA, SAH08363760GA, SAH08363763GA, SAH08373769GA, SAH08383827GA, SAH08383835GA, SAH08383836GA, SAH08383843GA, SAH08494057GA, SAH09024157GA, SAH09024158GA, SAH09024170GA, SAH09024186GA, SAH04360070GA, SAH05431160GA, SAH05521407GA, SAH08423901GA, SAH05150797GA, SAH08213555GA, SAH08213560GA, SAH06462302GA, SAH06462303GA, SAH04430250GA, SAH05381067GA, SAH05471276GA, SAH05481298GA, SAH05481299GA, SAH05481302GA, SAH05481303GA, SAH05481308GA, SAH05481313GA, SAH05491320GA, SAH05491321GA, SAH05491322GA, SAH05491323GA, SAH05491324GA, SAH05491325GA, SAH05491326GA, SAH05491327GA, SAH05491328GA, SAH05491329GA, SAH05491330GA, SAH05491332GA, SAH05491333GA, SAH05491334GA, SAH05491335GA, SAH05491336GA, SAH06021445GA, SAH06021446GA, SAH06181822GA, SAH07052522GA, SAH08213561GA, SAH08343720GA, SAH08343721GA, SAH07493273GA, SAH05290977GA, SAH05290981GA, SAH05290983GA, SAH05320998GA, SAH06442215GA, SAH06442218GA, SAH06442220GA, SAH06442223GA, SAH06502354GA, SAH06502355GA, SAH06502362GA, SAH06502363GA, SAH06502364GA, SAH06502365GA, SAH06502366GA, SAH06502367GA, SAH06502368GA, SAH06502369GA, SAH06502370GA, SAH06502371GA, SAH07052494GA, SAH07052495GA, SAH07052499GA, SAH07052502GA, SAH07052520GA, SAH07062525GA, SAH08373767GA, SAH08373796GA, SAH08373799GA, SAH05140735GA, SAH05220835GA, SAH05240893GA, SAH05260908GA, SAH05260913GA, SAH05280924GA, SAH05280926GA, SAH05280927GA, SAH05280928GA, SAH05280930GA, SAH05280933GA, SAH05280943GA, SAH05290959GA, SAH05320993GA, SAH06041482GA, SAH07372968GA, SAH07372970GA, SAH08473982GA, SAH06141737GA, SAH07483193GA, SAH07483222GA, SAH08023366GA, SAH08023367GA, SAH08023371GA, SAH04360069GA, SAH04400181GA, SAH04400183GA, SAH04410205GA, SAH04420210GA, SAH04420211GA, SAH04420212GA, SAH04420213GA, SAH04420214GA, SAH04420215GA, SAH04420216GA, SAH04420217GA, SAH04420218GA, SAH04420219GA, SAH04480337GA, SAH04480356GA, SAH04510434GA, SAH04510456GA, SAH04530495GA, SAH04530499GA, SAH04530508GA, SAH04530509GA, SAH05010510GA, SAH05010511GA, SAH05010514GA, SAH05010518GA, SAH05010526GA, SAH05010527GA, SAH05010531GA, SAH05010532GA, SAH05010534GA, SAH05010542GA, SAH05010544GA, SAH05010551GA, SAH05010562GA, SAH05040569GA, SAH05070597GA, SAH05070602GA, SAH05070603GA, SAH05070604GA, SAH05070605GA, SAH05070606GA, SAH05070617GA, SAH05080636GA, SAH05080640GA, SAH05080642GA, SAH05140717GA, SAH05140727GA, SAH05140730GA, SAH05140743GA, SAH05240873GA, SAH05391075GA, SAH05391089GA, SAH05391093GA, SAH05421120GA, SAH05431130GA, SAH05471237GA, SAH05471244GA, SAH06011433GA, SAH06081565GA, SAH06091592GA, SAH06091609GA, SAH06091611GA, SAH06091620GA, SAH06101638GA, SAH06101643GA, SAH06101650GA, SAH06121705GA, SAH06121708GA, SAH06141745GA, SAH06141751GA, SAH06171792GA, SAH06171798GA, SAH07052518GA, SAH07102583GA, SAH07282811GA, SAH08033375GA, SAH08133405GA, SAH08133409GA, SAH08143426GA, SAH08143431GA, SAH08143434GA, SAH08143435GA, SAH08143440GA, SAH08143441GA, SAH08143443GA, SAH08153456GA, SAH08153457GA, SAH08153458GA, SAH08153461GA, SAH08153463GA, SAH08213566GA, SAH08213578GA, SAH08373782GA, SAH08443933GA, SAH08443938GA, SAH08443957GA, SAH08443961GA, SAH08473968GA, SAH08473973GA, SAH08473979GA, SAH08473983GA, SAH08473985GA, SAH08473994GA, SAH08473999GA, SAH08524129GA, SAH08524132GA, SAH08524134GA, SAH08133414GA, SAH08253605GA, SAH08253607GA, SAH08253608GA, SAH08253612GA, SAH08473972GA, SAH08443943GA, SAH08443956GA, SAH05260923GA, SAH06041477GA, SAH07493277GA, SAH07493298GA, SAH07493302GA, SAH07493304GA, SAH04360064GA, SAH04400161GA, SAH04400166GA, SAH04400169GA, SAH04400182GA, SAH04400184GA, SAH06041465GA, SAH06422162GA, SAH06462285GA, SAH06462290GA, SAH06462291GA, SAH07282815GA, SAH07403021GA, SAH05290961GA, SAH06522425GA, SAH04510432GA, SAH05240880GA, SAH05290971GA, SAH05441167GA, SAH05441169GA, SAH05451196GA, SAH05451198GA, SAH05451202GA, SAH05451206GA, SAH05461223GA, SAH05481307GA, SAH06422178GA, SAH06482319GA, SAH06482321GA, SAH06482322GA, SAH05481305GA, SAH05481311GA, SAH05481312GA, SAH07102595GA, SAH07102596GA, SAH07493295GA, SAH07493296GA, SAH07493297GA, SAH07493300GA, SAH07493305GA, SAH07493306GA, SAH07493307GA, SAH07493308GA, SAH07493310GA, SAH07493311GA, SAH04470310GA, SAH05441172GA, SAH08213573GA, SAH08313691GA, SAH05320997GA, SAH05321000GA, SAH05321001GA, SAH05321002GA, SAH06291986GA, SAH06291993GA, SAH05441179GA, SAH06522379GA, SAH07493276GA, SAH05441170GA, SAH05441189GA, SAH05441190GA, SAH05441191GA, SAH05441192GA, SAH05441194GA, SAH06051515GA, SAH07282813GA, SAH07372986GA, SAH07372988GA, SAH04380107GA, SAH04450279GA, SAH04460297GA, SAH05391076GA, SAH06342058GA, SAH07152667GA, SAH07212754GA, SAH07212755GA, SAH08253622GA, SAH07403022GA, SAH07423071GA, SAH07493268GA, SAH08023335GA, SAH08023343GA, SAH06342049GA, SAH07022470GA, SAH04520466GA, SAH04520467GA, SAH07052498GA, SAH07052503GA, SAH07212737GA, SAH07212738GA, SAH07212747GA, SAH05070600GA, SAH05070612GA, SAH05391077GA, SAH05391078GA, SAH05391079GA, SAH05391080GA, SAH05481285GA, SAH05481286GA, SAH05481288GA, SAH05481289GA, SAH05481291GA, SAH05481293GA, SAH05481294GA, SAH08293657GA, SAH04430246GA, SAH04430247GA, SAH07342936GA, SAH06241879GA, SAH06241882GA, SAH06241884GA, SAH06241949GA, SAH07423108GA, SAH05351018GA, SAH05351021GA, SAH05351022GA, SAH05351029GA, SAH06081556GA, SAH07493275GA, SAH05133397GA, SAH08133395GA, SAH08133396GA, SAH08133397GA, SAH08133401GA, SAH08133402GA, SAH08133422GA, SAH05150793GA, SAH05220846GA, SAH05220851GA, SAH05240875GA, SAH05250894GA, SAH05250896GA, SAH05451197GA, SAH05451199GA, SAH05451201GA, SAH06041493GA, SAH06241911GA, SAH08373784GA, SAH06091628GA, SAH06091632GA, SAH06021450GA, SAH07212757GA, SAH05140740GA, SAH05431157GA, SAH06021438GA, SAH06021449GA, SAH06021452GA, SAH06021453GA, SAH06021459GA, SAH08203539GA, SAH08473990GA, SAH08473992GA, SAH08494053GA, SAH08494062GA, SAH06241885GA, SAH06241887GA, SAH06101642GA, SAH06151768GA, SAH06241897GA, SAH06292001GA, SAH06522378GA, SAH06522404GA, SAH06522405GA, SAH06522406GA, SAH08023338GA, SAH07052531GA, SAH06041487GA, SAH07142626GA, SAH04530505GA, SAH05040578GA, SAH07403005GA, SAH07483230GA, SAH07493269GA, SAH06011427GA, SAH06011432GA, SAH04520470GA, SAH04520472GA, SAH04520474GA, SAH06041492GA, SAH08173501GA, SAH06041471GA, SAH04500414GA, SAH04500418GA, SAH05080670GA, SAH07403052GA, SAH06291982GA, SAH06332018GA, SAH06382084GA, SAH08173496GA, SAH08363761GA, SAH07483218GA, SAH07483219GA, SAH07483229GA, SAH07483234GA, SAH08473964GA, SAH05471252GA, SAH06432184GA, SAH05220842GA, SAH05220844GA, SAH05220850GA, SAH07062530GA, SAH08253597GA, SAH08303684GA, SAH07282821GA, SAH07403057GA, SAH07403061GA, SAH07403064GA, SAH09024202GA, SAH05140731GA, SAH04470329GA, SAH04470333GA, SAH04480346GA, SAH04480347GA, SAH05080656GA, SAH05471241GA, SAH05471267GA, SAH05481310GA, SAH05491360GA, SAH05491363GA, SAH05501365GA, SAH05501366GA, SAH05501367GA, SAH05501368GA, SAH05501369GA, SAH05501370GA, SAH05501371GA, SAH05501372GA, SAH05501373GA, SAH05501374GA, SAH06051510GA, SAH0610141761GA, SAH06101633GA, SAH06101634GA, SAH06101636GA, SAH06101637GA, SAH06101639GA, SAH06101641GA, SAH06141749GA, SAH06141757GA, SAH06141761GA, SAH06191862GA, SAH06241930GA, SAH06241962GA, SAH06332027GA, SAH06332031GA, SAH06332032GA, SAH06332033GA, SAH06332034GA, SAH06332037GA, SAH06332038GA, SAH06332040GA, SAH06332041GA, SAH06332042GA, SAH06332043GA, SAH06342045GA, SAH06342048GA, SAH06342056GA, SAH06522387GA, SAH06522412GA, SAH06522421GA, SAH07202726GA, SAH07372944GA, SAH07372955GA, SAH07372956GA, SAH07403002GA, SAH07403017GA, SAH07403018GA, SAH07403032GA, SAH07403033GA, SAH08473996GA, SAH08494035GA, SAH08494036GA, SAH08494037GA, SAH08494038GA, SAH08494040GA, SAH08494041GA, SAH08494042GA, SAH08494043GA, SAH08524115GA, SAH08524116GA, SAH08524117GA, SAH08524118GA, SAH08524119GA, SAH08524120GA, SAH08524121GA, SAH08524122GA, SAH08524123GA, SAH09024161GA, SAH09024162GA, SAH09024167GA, SAH09024168GA, SAH09024171GA, SAH09024172GA, SAH04430252GA, SAH07212734GA, SAH07212739GA, SAH07212741GA, SAH06041473GA, SAH06051500GA, SAH06051505GA, SAH08373798GA, SAH05431140GA, SAH05431141GA, SAH05431142GA, SAH05431145GA, SAH05431147GA, SAH07012435GA, SAH06291988GA, SAH06402106GA, SAH06402109GA, SAH06402116GA, SAH06402117GA, SAH06402118GA, SAH06402119GA, SAH06402122GA, SAH06402123GA, SAH05110685GA, SAH05110688GA, SAH05110694GA, SAH05110695GA, SAH05140718GA, SAH05140722GA, SAH05381068GA, SAH06091584GA, SAH06101658GA, SAH06101681GA, SAH06191865GA, SAH06291976GA, SAH06342059GA, SAH05331015GA, SAH06412131GA, SAH06412138GA, SAH07172699GA, SAH07202731GA, SAH07403046GA, SAH07403049GA, SAH07403050GA, SAH07443140GA, SAH07443142GA, SAH05471235GA, SAH05471240GA, SAH05471272GA, SAH08023349GA, SAH08033390GA, SAH08153471GA, SAH08173485GA, SAH07423113GA, SAH04360063GA, SAH04360066GA, SAH04430235GA, SAH04430241GA, SAH05471262GA, SAH05471265GA, SAH08353731GA, SAH08423900GA, SAH08423902GA, SAH07202717GA, SAH06071550GA, SAH06051506GA, SAH06051507GA, SAH06051508GA, SAH06051509GA, SAH06051511GA, SAH06081568GA, SAH05491347GA, SAH05491348GA, SAH05491352GA, SAH05491353GA, SAH05491356GA, SAH05501376GA, SAH05501377GA, SAH06051514GA, SAH07493283GA, SAH05080675GA, SAH05110686GA, SAH05110687GA, SAH05110689GA, SAH05110692GA, SAH05080658GA, SAH05080662GA, SAH05080664GA, SAH05080680GA, SAH06141754GA, SAH06171790GA, SAH06171793GA, SAH06181814GA, SAH06181820GA, SAH08293654GA, SAH08524126GA, SAH08524127GA, SAH08524128GA, SAH08133412GA, SAH05080684GA, SAH07463176GA, SAH07463177GA, SAH07483188GA, SAH06241914GA, SAH06241917GA, SAH06241969GA, SAH04480349GA, SAH04480350GA, SAH04480355GA, SAH04490362GA, SAH05010545GA, SAH05451217GA, SAH06241926GA, SAH06241928GA, SAH06241932GA, SAH06292004GA, SAH06412148GA, SAH07062540GA, SAH07062548GA, SAH08443944GA, SAH05280941GA, SAH05280948GA, SAH06091617GA, SAH06091619GA, SAH06121729GA, SAH06151772GA, SAH07423085GA, SAH07423087GA, SAH07423095GA, SAH06021435GA, SAH05140716GA, SAH05140720GA, SAH06091614GA, SAH06091616GA, SAH06091621GA, SAH06091622GA, SAH06091627GA, SAH06121720GA, SAH06121726GA, SAH04480335GA, SAH05351020GA, SAH05481304GA, SAH07312892GA, SAH07312894GA, SAH07312908GA, SAH07372976GA, SAH06241886GA, SAH06241888GA, SAH06241935GA, SAH06241941GA, SAH07052543GA, SAH07102592GA, SAH07102593GA, SAH07102597GA, SAH07102600GA, SAH07102601GA, SAH07102608GA, SAH07423067GA, SAH07423077GA, SAH07493239GA, SAH07493244GA, SAH07493266GA, SAH08153472GA, SAH08173478GA, SAH08173479GA, SAH08203505GA, SAH08203515GA, SAH08203534GA, SAH08203538GA, SAH07463156GA, SAH07463159GA, SAH07463184GA, SAH06332014GA, SAH06332020GA, SAH06332021GA, SAH06332023GA, SAH06502346GA, SAH06502352GA, SAH06502353GA, SAH07312893GA, SAH06101690GA, SAH06181819GA, SAH06181830GA, SAH08494051GA, SAH07342914GA, SAH07342916GA, SAH07483226GA, SAH08494015GA, SAH08494016GA, SAH08494017GA, SAH08494018GA, SAH08494020GA, SAH08494021GA, SAH08494022GA, SAH08494029GA, SAH05351016GA, SAH05351017GA, SAH05351023GA, SAH05351026GA, SAH05351033GA, SAH05351034GA, SAH07012445GA, SAH07062552GA, SAH04520484GA, SAH04400172GA, SAH05431154GA, SAH05381072GA, SAH05381074GA, SAH05391081GA, SAH05391082GA, SAH07172701GA, SAH07202722GA, SAH04360073GA, SAH07022473GA, SAH07282806GA, SAH07282809GA, SAH07423069GA, SAH08023352GA, SAH05441171GA, SAH05441173GA, SAH05441188GA, SAH05441193GA, SAH06241895GA, SAH06241900GA, SAH06241956GA, SAH04480348GA, SAH06181823GA, SAH07493282GA, SAH06081573GA, SAH06181816GA, SAH06181827GA, SAH06181835GA, SAH06181840GA, SAH06181843GA, SAH06181848GA, SAH06482338GA, SAH07232770GA, SAH07232774GA, SAH07232775GA, SAH07423118GA, SAH07423123GA, SAH08253621GA, SAH08443949GA, SAH04520483GA, SAH05240887GA, SAH05471271GA, SAH05260904GA, SAH05260909GA, SAH05260911GA, SAH05260917GA, SAH07012438GA, SAH06171776GA, SAH06171777GA, SAH08504081GA, SAH04410209GA, SAH07022488GA, SAH07062528GA, SAH08343713GA, SAH08353725GA, SAH08353726GA, SAH08353727GA, SAH08353730GA, SAH08353732GA, SAH08353733GA, SAH07493253GA, SAH06241893GA, SAH06241901GA, SAH06241909GA, SAH06241910GA, SAH08383826GA, SAH08383829GA, SAH09024165GA, SAH09024178GA, SAH09024179GA, SAH09024184GA, SAH06382077GA, SAH06422164GA, SAH06422166GA, SAH06422167GA, SAH06422170GA, SAH06422172GA, SAH06432193GA, SAH06462277GA, SAH06462286GA, SAH06462289GA, SAH05070598GA, SAH05080682GA, SAH05110705GA, SAH05110707GA, SAH05110711GA, SAH05110713GA, SAH07052501GA, SAH08514100GA, SAH04470317GA, SAH04470319GA, SAH04520479GA, SAH05160828GA, SAH05351031GA, SAH05501375GA, SAH06021451GA, SAH06091605GA, SAH06091612GA, SAH06191856GA, SAH06191872GA, SAH06191873GA, SAH06452252GA, SAH07012436GA, SAH07012437GA, SAH07012440GA, SAH07012441GA, SAH07022452GA, SAH07022453GA, SAH07022459GA, SAH07022460GA, SAH07022474GA, SAH07022475GA, SAH07022477GA, SAH07022478GA, SAH07022483GA, SAH07022484GA, SAH07022489GA, SAH07022490GA, SAH07152668GA, SAH07403023GA, SAH07403024GA, SAH07403025GA, SAH07403029GA, SAH07403031GA, SAH07423066GA, SAH07493256GA, SAH07493261GA, SAH08343715GA, SAH08383828GA, SAH06412127GA, SAH05160831GA, SAH07403044GA, SAH08474001GA, SAH06382080GA, SAH06081563GA, SAH06081593GA, SAH06091590GA, SAH05491344GA, SAH06292005GA, SAH06292010GA, SAH05010519GA, SAH06442224GA, SAH06442228GA, SAH07012442GA, SAH07012448GA, SAH07282816GA, SAH07282819GA, SAH06442231GA, SAH05010525GA, SAH05040589GA, SAH05040594GA, SAH05441166GA, SAH07052509GA, SAH07493262GA, SAH08313696GA, SAH04510423GA, SAH04510435GA, SAH04510455GA, SAH05010540GA, SAH05010556GA, SAH05140742GA, SAH05160827GA, SAH05290965GA, SAH05501382GA, SAH05501384GA, SAH05521415GA, SAH05521420GA, SAH06091604GA, SAH06091626GA, SAH06091629GA, SAH06291985GA, SAH06292007GA, SAH06402105GA, SAH06402107GA, SAH06402110GA, SAH06402113GA, SAH06422175GA, SAH06422179GA, SAH06502361GA, SAH07142631GA, SAH07152644GA, SAH07152671GA, SAH07202704GA, SAH07202709GA, SAH07202710GA, SAH07292822GA, SAH07292837GA, SAH07292838GA, SAH07493248GA, SAH08373805GA, SAH08423896GA, SAH08443962GA, SAH08473974GA, SAH08473988GA, SAH08504071GA, SAH08514107GA, SAH07312887GA, SAH07342918GA, SAH07342921GA, SAH07342925GA, SAH07342933GA, SAH07342939GA, SAH07342940GA, SAH07342941GA, SAH08213568GA, SAH05080657GA, SAH05080661GA, SAH08494031GA, SAH08494032GA, SAH07022462GA, SAH07022467GA, SAH07022468GA, SAH04510444GA, SAH04510445GA, SAH04510452GA, SAH05521406GA, SAH08473989GA, SAH08494054GA, SAH05010513GA, SAH07493241GA, SAH08373812GA, SAH06241907GA, SAH06241908GA, SAH06241913GA, SAH08504083GA, SAH08524130GA, SAH08524131GA, SAH09014145GA, SAH07212758GA, SAH07212759GA, SAH07212761GA, SAH07292848GA, SAH07423076GA, SAH07483205GA, SAH06332017GA, SAH06332025GA, SAH06332026GA, SAH07493299GA, SAH05260906GA, SAH05260912GA, SAH05351027GA, SAH05351028GA, SAH05351042GA, SAH05351043GA, SAH05431144GA, SAH08494049GA, SAH05320987GA, SAH07403048GA, SAH04430237GA, SAH08343706GA, SAH08293655GA, SAH08443960GA, SAH06462298GA, SAH06462308GA, SAH08473981GA, SAH08494063GA, SAH07022454GA, SAH07022457GA, SAH07022463GA, SAH07022469GA, SAH08303680GA, SAH04420223GA, SAH05421122GA, SAH06101649GA, SAH06181815GA, SAH06181837GA, SAH06181838GA, SAH06181841GA, SAH06181842GA, SAH06181846GA, SAH04290012GA, SAH04290013GA, SAH04290023GA, SAH07372984GA, SAH07372985GA, SAH07342923GA, SAH07342930GA, SAH05220861GA, SAH06121718GA, SAH08303666GA, SAH07022491GA, SAH07052523GA, SAH07062532GA, SAH07463170GA, SAH07463180GA, SAH07483187GA, SAH07443154GA, SAH08343717GA, SAH05481306GA, SAH04490365GA, SAH05040586GA, SAH07423098GA, SAH07423099GA, SAH07423104GA, SAH07463173GA, SAH07403003GA, SAH07403020GA, SAH07463165GA, SAH05240878GA, SAH05280939GA, SAH05290970GA, SAH07503320GA, SAH06291990GA, SAH08133399GA, SAH08373814GA, SAH08373820GA, SAH08373824GA, SAH06081566GA, SAH06241959GA, SAH07372949GA, SAH07403039GA, SAH07423070GA, SAH07423081GA, SAH07493235GA, SAH07493236GA, SAH07493237GA, SAH07493240GA, SAH07493242GA, SAH07493288GA, SAH07493289GA, SAH07493290GA, SAH07493291GA, SAH07493292GA, SAH07493293GA, SAH07493294GA, SAH04400171GA, SAH04400173GA, SAH08494006GA, SAH08494009GA, SAH09024194GA, SAH06241916GA, SAH06241920GA, SAH06051504GA, SAH07022476GA, SAH07372962GA, SAH08033379GA, SAH06342063GA, SAH07483216GA, SAH07372942GA, SAH04360067GA, SAH06292013GA, SAH04450268GA, SAH04450269GA, SAH06241921GA, SAH07022465GA, SAH05391090GA, SAH06402108GA, SAH05471236GA, SAH05471275GA, SAH05240885GA, SAH06011428GA, SAH07483203GA, SAH06051519GA, SAH06061522GA, SAH06061523GA, SAH06522396GA, SAH06522420GA, SAH05391096GA, SAH05391099GA, SAH05391103GA, SAH05471277GA, SAH06462292GA, SAH06482324GA, SAH06482325GA, SAH06482327GA, SAH06482339GA, SAH06482340GA, SAH06482341GA, SAH05080625GA, SAH05080627GA, SAH05080628GA, SAH05080629GA, SAH05080631GA, SAH05080632GA, SAH05080633GA, SAH05080634GA, SAH06171783GA, SAH04430239GA, SAH06522400GA, SAH07022485GA, SAH05110697GA, SAH05471242GA, SAH06121702GA, SAH06241933GA, SAH07142627GA, SAH07202711GA, SAH07022481GA, SAH07372977GA, SAH07312896GA, SAH07423079GA, SAH04410206GA, SAH04410207GA, SAH04410208GA, SAH07342913GA, SAH07342926GA, SAH07342931GA, SAH05411108GA, SAH05431132GA, SAH05431133GA, SAH05431134GA, SAH05431136GA, SAH05431138GA, SAH05431139GA, SAH05471280GA, SAH06041485GA, SAH06041486GA, SAH07372952GA, SAH07372957GA, SAH07372978GA, SAH071426 14GA, SAH07142637GA, SAH07142640GA, SAH06432197GA, SAH06432201GA, SAH06291979GA, SAH06291980GA, SAH06291983GA, SAH06291989GA, SAH07052551GA, SAH07102587GA, SAH07102588GA, SAH07102591GA, SAH05351045GA, SAH05381047GA, SAH05381048GA, SAH05381049GA, SAH05381050GA, SAH05381051GA, SAH05381052GA, SAH05381053GA, SAH05381054GA, SAH05381059GA, SAH05381062GA, SAH05381064GA, SAH05511402GA, SAH05511403GA, SAH05511404GA, SAH07493279GA, SAH07142632GA, SAH08443924GA, SAH08443947GA, SAH08443953GA, SAH08023356GA, SAH08173490GA, SAH08173494GA, SAH08213575GA, SAH08213577GA, SAH08213580GA, SAH08213581GA, SAH08213582GA, SAH08213583GA, SAH06432190GA, SAH08343708GA, SAH08363755GA, SAH08363756GA, SAH08363759GA, SAH08363762GA, SAH08363764GA, SAH08373770GA, SAH08373802GA, SAH08383844GA, SAH08403845GA, SAH08403846GA, SAH08403847GA, SAH08403848GA, SAH08403852GA, SAH08403853GA, SAH08403854GA, SAH08403856GA, SAH08403859GA, SAH08403860GA, SAH08403861GA, SAH08403863GA, SAH08443936GA, SAH08443939GA, SAH06021457GA, SAH06021461GA, SAH06021462GA, SAH06021463GA, SAH06051518GA, SAH07212736GA, SAH07212752GA, SAH08494005GA, SAH08494011GA, SAH08494012GA, SAH08494013GA, SAH08494014GA, SAH08524135GA, SAH08524136GA, SAH08524137GA, SAH08524138GA, SAH08524139GA, SAH08524140GA, SAH08524141GA, SAH08524142GA, SAH07403055GA, SAH04400168GA, SAH05421119GA, SAH05421124GA, SAH05040588GA, SAH04470328GA, SAH05040585GA, SAH05070614GA, SAH07062542GA, SAH07062544GA, SAH07062550GA, SAH07102598GA, SAH07372993GA, SAH07463172GA, SAH08443926GA, SAH08303673GA, SAH08403882GA, SAH08473980GA, SAH04520477GA, SAH08213556GA, SAH08213558GA, SAH08213563GA, SAH08213564GA, SAH08403870GA, SAH09024175GA, SAH09024204GA, SAH07282810GA, SAH07292829GA, SAH04520476GA, SAH05220853GA, SAH06091583GA, SAH06101663GA, SAH06101665GA, SAH06101666GA, SAH06101667GA, SAH06101668GA, SAH06101669GA, SAH06101670GA, SAH06101671GA, SAH06101672GA, SAH06101673GA, SAH06101674GA, SAH06101675GA, SAH06101676GA, SAH06101677GA, SAH06101678GA, SAH06101679GA, SAH06101682GA, SAH06101683GA, SAH06141746GA, SAH06141755GA, SAH06171794GA, SAH06171795GA, SAH06171799GA, SAH06171801GA, SAH06171802GA, SAH06171803GA, SAH06171806GA, SAH06171807GA, SAH06171808GA, SAH06171809GA, SAH06181817GA, SAH06181818GA, SAH06191863GA, SAH06462280GA, SAH06522431GA, SAH08023369GA, SAH08023374GA, SAH08203533GA, SAH08443963GA, SAH08473984GA, SAH06021436GA, SAH06021437GA, SAH06021439GA, SAH06021440GA, SAH06442229GA, SAH06442233GA, SAH08473969GA, SAH04350040GA, SAH04350054GA, SAH05150782GA, SAH05150788GA, SAH05150789GA, SAH05290974GA, SAH06101680GA, SAH06121710GA, SAH07372966GA, SAH07372973GA, SAH07372974GA, SAH07372975GA, SAH07372980GA, SAH07372983GA, SAH07463157GA, SAH08373795GA, SAH08373801GA, SAH08373804GA, SAH08373806GA, SAH08373808GA, SAH08373809GA, SAH08373813GA, SAH08373817GA, SAH08373819GA, SAH08403850GA, SAH08403851GA, SAH08403855GA, SAH08403857GA, SAH08403858GA, SAH08403864GA, SAH06382081GA, SAH05521416GA, SAH05521417GA, SAH05110696GA, SAH05150804GA, SAH05160813GA, SAH05220859GA, SAH05280929GA, SAH05280931GA, SAH05391086GA, SAH05421123GA, SAH05441181GA, SAH05481309GA, SAH05491315GA, SAH05511400GA, SAH05511401GA, SAH05521425GA, SAH06121706GA, SAH06121722GA, SAH06171810GA, SAH06171813GA, SAH06332015GA, SAH06332019GA, SAH06422171GA, SAH06522384GA, SAH06522418GA, SAH07282803GA, SAH07403054GA, SAH07443127GA, SAH08283644GA, SAH08303672GA, SAH08373803GA, SAH08403866GA, SAH04490373GA, SAH06101694GA, SAH05010528GA, SAH08023341GA, SAH08023345GA, SAH08023351GA, SAH08403880GA, SAH04520475GA, SAH05070599GA, SAH05080635GA, SAH05080637GA, SAH07072579GA, SAH07072581GA, SAH07072582GA, SAH07443148GA, SAH07463178GA, SAH07463179GA, SAH07463181GA, SAH07483190GA, SAH07483192GA, SAH05290967GA, SAH05290968GA, SAH05290978GA, SAH05381066GA, SAH05381070GA, SAH08153473GA, SAH05381055GA, SAH05381056GA, SAH05381057GA, SAH05381058GA, SAH05381060GA, SAH05381061GA, SAH05381063GA, SAH05010553GA, SAH06101653GA, SAH06171779GA, SAH06241891GA, SAH06522413GA, SAH07072567GA, SAH07072573GA, SAH07072574GA, SAH07072576GA, SAH07072577GA, SAH07072580GA, SAH08383825GA, SAH07072556GA, SAH07072557GA, SAH07072559GA, SAH07072562GA, SAH07072566GA, SAH08494019GA, SAH08494023GA, SAH08494024GA, SAH04360071GA, SAH04400162GA, SAH04400163GA, SAH04400164GA, SAH04400165GA, SAH04490374GA, SAH04520481GA, SAH04520482GA, SAH05220848GA, SAH05491317GA, SAH05491319GA, SAH05501381GA, SAH05521408GA, SAH05521410GA, SAH05521411GA, SAH05521412GA, SAH05521414GA, SAH05521418GA, SAH05521419GA, SAH05521421GA, SAH05521423GA, SAH05521424GA, SAH06121730GA, SAH06241924GA, SAH06412144GA, SAH06412145GA, SAH06412146GA, SAH06412147GA, SAH06412150GA, SAH06412151GA, SAH06412152GA, SAH06412153GA, SAH06422154GA, SAH06422155GA, SAH06422156GA, SAH06422157GA, SAH06422158GA, SAH06422159GA, SAH06422160GA, SAH07052511GA, SAH04410191GA, SAH04510449GA, SAH04520480GA, SAH05471278GA, SAH05491343GA, SAH04400160GA, SAH07052512GA, SAH05381071GA, SAH07052507GA, SAH05010546GA, SAH06241877GA, SAH06241912GA, SAH08383831GA, SAH08383832GA, SAH08383833GA, SAH08383834GA, SAH08383839GA, SAH08383840GA, SAH08383842GA, SAH08443948GA, SAH06121715GA, SAH06121732GA, SAH06121733GA, SAH08373777GA, SAH06061527GA, SAH06061529GA, SAH06061531GA, SAH06061533GA, SAH06061535GA, SAH05080672GA, SAH05491338GA, SAH05511399GA, SAH07463166GA, SAH07463169GA, SAH08213554GA, SAH08494025GA, SAH08494027GA, SAH08494034GA, SAH08494048GA, SAH07342932GA, SAH07342934GA, SAH07342935GA, SAH07342938GA, SAH06181834GA, SAH07372982GA, SAH07372987GA, SAH07373000GA, SAH08153468GA, SAH04500406GA, SAH05381073GA, SAH05080673GA, SAH08153466GA, SAH08153469GA, SAH07202730GA, SAH08443958GA, SAH05220837GA, SAH05240867GA, SAH05240868GA, SAH07022472GA, SAH07022479GA, SAH07292843GA, SAH05220836GA, SAH05240889GA, SAH05250898GA, SAH05280946GA, SAH05290963GA, SAH05290972GA, SAH05351035GA, SAH05351039GA, SAH05351040GA, SAH05351044GA, SAH05391101GA, SAH05451208GA, SAH08373787GA, SAH08373788GA, SAH05491339GA, SAH05010557GA, SAH07493249GA, SAH07493254GA, SAH04490382GA, SAH05431135GA, SAH07072569GA, SAH07072570GA, SAH05411107GA, SAH08473971GA, SAH05080678GA, SAH05110712GA, SAH06522430GA, SAH07142622GA, SAH07142628GA, SAH07142635GA, SAH07202707GA, SAH07202708GA, SAH07202715GA, SAH07202716GA, SAH07372981GA, SAH07483214GA, SAH07493280GA, SAH07493285GA, SAH07493286GA, SAH07493287GA, SAH08133406GA, SAH08133407GA, SAH08133408GA, SAH08133410GA, SAH08133411GA, SAH08133415GA, SAH08133417GA, SAH08133418GA, SAH08133419GA, SAH08133421GA, SAH08253588GA, SAH08253591GA, SAH08363754GA, SAH08373800GA, SAH08423895GA, SAH08423908GA, SAH08423912GA, SAH08443950GA, SAH08473966GA, SAH08473967GA, SAH08473976GA, SAH08473977GA, SAH08473978GA, SAH07282802GA, SAH07403036GA, SAH07403038GA, SAH08153465GA, SAH08203526GA, SAH04350051GA, SAH04350053GA, SAH04380091GA, SAH04390158GA, SAH04500410GA, SAH05010524GA, SAH05150780GA, SAH08033392GA, SAH08023368GA, SAH08033381GA, SAH08033388GA, SAH08143444GA, SAH08283626GA, SAH06121698GA, SAH06121703GA, SAH06171778GA, SAH06171780GA, SAH06171788GA, SAH06171789GA, SAH06171805GA, SAH07232784GA, SAH08313700GA, SAH08313701GA, SAH06482315GA, SAH06482323GA, SAH05481297GA, SAH07062524GA, SAH08033378GA, SAH04450285GA, SAH04460284GA, SAH04460286GA, SAH04460287GA, SAH04460288GA, SAH04460289GA, SAH04460290GA, SAH04460291GA, SAH04460292GA, SAH04460293GA, SAH04460294GA, SAH04460295GA, SAH04470309GA, SAH05220862GA, SAH06332029GA, SAH06342051GA, SAH08153445GA, SAH08153446GA, SAH08153447GA, SAH08153448GA, SAH08153449GA, SAH08153451GA, SAH08153452GA, SAH08153453GA, SAH08153454GA, SAH05431159GA, SAH06241892GA, SAH09024166GA, SAH05501386GA, SAH05511390GA, SAH05511391GA, SAH07463158GA, SAH04470324GA, SAH06021447GA, SAH06021456GA, SAH06181853GA, SAH06462281GA, SAH09014154GA, SAH09024160GA, SAH08474003GA, SAH06241899GA, SAH06241905GA, SAH06462278GA, SAH06462293GA, SAH07052506GA, SAH06011429GA, SAH07282792GA, SAH07282798GA, SAH07282801GA, SAH04510442GA, SAH05240876GA, SAH05240877GA, SAH05250903GA, SAH05290982GA, SAH05471261GA, SAH05481292GA, SAH06101655GA, SAH06121723GA, SAH07302862GA, SAH07302865GA, SAH07302867GA, SAH07302872GA, SAH07302873GA, SAH07302875GA, SAH07302876GA, SAH07302880GA, SAH07423091GA, SAH07423100GA, SAH07443125GA, SAH08023358GA, SAH08023360GA, SAH08023362GA, SAH08203517GA, SAH08203537GA, SAH08253600GA, SAH08253606GA, SAH08253610GA, SAH08253613GA, SAH08253615GA, SAH08253618GA, SAH08253619GA, SAH08253623GA, SAH08253624GA, SAH08283625GA, SAH08283627GA, SAH08283630GA, SAH08283633GA, SAH08283636GA, SAH08283637GA, SAH08283640GA, SAH08283642GA, SAH08293649GA, SAH08293651GA, SAH08293652GA, SAH08293653GA, SAH08293664GA, SAH08313690GA, SAH08313692GA, SAH08373776GA, SAH08373791GA, SAH08383830GA, SAH08403871GA, SAH04470332GA, SAH04520463GA, SAH05110698GA, SAH05110703GA, SAH05160808GA, SAH05260905GA, SAH05320991GA, SAH05501389GA, SAH06241876GA, SAH06241880GA, SAH06522382GA, SAH07423082GA, SAH07443146GA, SAH07463162GA, SAH07463168GA, SAH07483191GA, SAH07483194GA, SAH07483225GA, SAH07483227GA, SAH08494056GA, SAH08494059GA, SAH04340033GA, SAH04350039GA, SAH04350041GA, SAH04350042GA, SAH04350047GA, SAH04350050GA, SAH04360062GA, SAH04360072GA, SAH04360074GA, SAH04360075GA, SAH04360077GA, SAH04360078GA, SAH04360079GA, SAH04360080GA, SAH06432199GA, SAH04340036GA, SAH04480341GA, SAH06442216GA, SAH06442217GA, SAH06442219GA, SAH06452246GA, SAH06462275GA, SAH07443149GA, SAH08473965GA, SAH07493245GA, SAH07493246GA, SAH04520464GA, SAH04520465GA, SAH04530488GA, SAH05070596GA, SAH05070601GA, SAH05070609GA, SAH06041470GA, SAH07152642GA, SAH07152646GA, SAH07152648GA, SAH07152650GA, SAH05501383GA, SAH08473998GA, SAH07342919GA, SAH07312891GA, SAH07342912GA, SAH07342917GA, SAH07342922GA, SAH07342924GA, SAH07342928GA, SAH06241904GA, SAH05491349GA, SAH05491354GA, SAH05320986GA, SAH07493301GA, SAH07493303GA, SAH07503317GA, SAH05080646GA, SAH08474002GA, SAH05431137GA, SAH05431149GA, SAH05441177GA, SAH05441178GA, SAH05441182GA, SAH06041478GA, SAH08303675GA, SAH08303676GA, SAH05220856GA, SAH05240886GA, SAH05240890GA, SAH05240892GA, SAH07142636GA, SAH07202703GA, SAH07202714GA, SAH08293658GA, SAH04530492GA, SAH06021454GA, SAH06021458GA, SAH08423910GA, SAH08423915GA, SAH08423916GA, SAH08423917GA, SAH08423918GA, SAH08423920GA, SAH08443925GA, SAH06241951GA, SAH06241971GA, SAH08343724GA, SAH07282820GA, SAH04450270GA, SAH06041475GA, SAH06041479GA, SAH05280950GA, SAH06342057GA, SAH06291987GA, SAH04450273GA, SAH04450274GA, SAH04450277GA, SAH04450278GA, SAH05010549GA, SAH05431158GA, SAH05431162GA, SAH05431163GA, SAH05441176GA, SAH05441180GA, SAH05441183GA, SAH06241955GA, SAH06292002GA, SAH06292008GA, SAH07172692GA, SAH07172693GA, SAH08303674GA, SAH08443952GA, SAH08443954GA, SAH08443955GA, SAH06482314GA, SAH06181832GA, SAH07403028GA, SAH05220854GA, SAH05220855GA, SAH05220858GA, SAH05220864GA, SAH05240872GA, SAH05290964GA, SAH05431148GA, SAH08033383GA, SAH08033384GA, SAH06522389GA, SAH05080647GA, SAH05080648GA, SAH05080649GA, SAH05280938GA, SAH05280944GA, SAH05280945GA, SAH05280953GA, SAH05321005GA, SAH05321007GA, SAH05321008GA, SAH05321009GA, SAH05321010GA, SAH05321011GA, SAH05321013GA, SAH05321014GA, SAH07142616GA, SAH04530490GA, SAH06241918GA, SAH08173486GA, SAH08213576GA, SAH05080645GA, SAH05080650GA, SAH05150784GA, SAH07062527GA, SAH06241883GA, SAH05471281GA, SAH05481314GA, SAH06402114GA, SAH06412124GA, SAH06412128GA, SAH06412130GA, SAH06412133GA, SAH06412136GA, SAH06432211GA, SAH07403056GA, SAH07403062GA, SAH08143433GA, SAH05080638GA, SAH09024182GA, SAH09024183GA, SAH07493265GA, SAH05290979GA, SAH08423897GA, SAH05471264GA, SAH06121714GA, SAH06121721GA, SAH06181847GA, SAH06191858GA, SAH06422174GA, SAH06522417GA, SAH08524125GA, SAH09024177GA, SAH06071538GA, SAH06241954GA, SAH06292000GA, SAH06342060GA, SAH06422177GA, SAH06442234GA, SAH06482326GA, SAH07072558GA, SAH07072563GA, SAH07102610GA, SAH07312903GA, SAH07312907GA, SAH07493309GA, SAH07493314GA, SAH08143436GA, SAH08143442GA, SAH08173475GA, SAH08173484GA, SAH09024188GA, SAH09024189GA, SAH05140725GA, SAH05140726GA, SAH06241923GA, SAH04420222GA, SAH04420234GA, SAH04490367GA, SAH04490372GA, SAH04490375GA, SAH04490378GA, SAH04490381GA, SAH07312886GA, SAH05411111GA, SAH09014146GA, SAH09014147GA, SAH09014148GA, SAH09014149GA, SAH09014150GA, SAH09014151GA, SAH09014152GA, SAH09014153GA, SAH09024156GA, SAH04450271GA, SAH04450280GA, SAH04450281GA, SAH04450282GA, SAH04450283GA, SAH07212760GA, SAH05110702GA, SAH07282793GA, SAH07282794GA, SAH07282796GA, SAH07282800GA, SAH06081576GA, SAH05010522GA, SAH06101657GA, SAH06101662GA, SAH05240865GA, SAH05240866GA, SAH05250901GA, SAH04430251GA, SAH04480343GA, SAH05511394GA, SAH05511396GA, SAH08203508GA, SAH08203509GA, SAH08203510GA, SAH08203511GA, SAH08203512GA, SAH08203513GA, SAH08203514GA, SAH08203518GA, SAH08203519GA, SAH08203522GA, SAH08203523GA, SAH08203524GA, SAH08203535GA, SAH08203540GA, SAH09024169GA, SAH09024173GA, SAH09024174GA, SAH05080630GA, SAH05080639GA, SAH05150785GA, SAH04480336GA, SAH04530501GA, SAH04530506GA, SAH04530507GA, SAH07423115GA, SAH08443934GA, SAH05010533GA, SAH05010541GA, SAH05010550GA, SAH05010554GA, SAH05010561GA, SAH08373810GA, SAH08373816GA, SAH06051520GA, SAH06051521GA, SAH06442235GA, SAH06442236GA, SAH06442238GA, SAH06442239GA, SAH06442240GA, SAH06442241GA, SAH06442242GA, SAH06442243GA, SAH06452244GA, SAH06452245GA, SAH06452248GA, SAH06462279GA, SAH08373779GA, SAH08373818GA, SAH07292847GA, SAH05240870GA, SAH05080626GA, SAH06382067GA, SAH06382073GA, SAH06382074GA, SAH06382075GA, SAH06382076GA, SAH06382078GA, SAH06382082GA, SAH06382083GA, SAH07282814GA, SAH07403001GA, SAH07443144GA, SAH07463174GA, SAH08023339GA, SAH08023342GA, SAH08023350GA, SAH08023373GA, SAH07102589GA, SAH07102590GA, SAH07102594GA, SAH05040590GA, SAH06432212GA, SAH08494045GA, SAH08494046GA, SAH08494047GA, SAH06091595GA, SAH06091596GA, SAH06091597GA, SAH06091598GA, SAH06091599GA, SAH06091601GA, SAH06091607GA, SAH07372972GA, SAH07493251GA, SAH07493252GA, SAH06522376GA, SAH06522381GA, SAH06522394GA, SAH08253616GA, SAH07202713GA, SAH08373807GA, SAH08373821GA, SAH04350043GA, SAH04420220GA, SAH04420226GA, SAH04480340GA, SAH04480342GA, SAH05010539GA, SAH05040573GA, SAH05040577GA, SAH05040579GA, SAH05040582GA, SAH05040593GA, SAH05070615GA, SAH05070620GA, SAH05070622GA, SAH05140721GA, SAH05140723GA, SAH05421118GA, SAH06432189GA, SAH06432195GA, SAH07292842GA, SAH07292850GA, SAH08213562GA, SAH08213570GA, SAH08494010GA, SAH04400167GA, SAH05280934GA, SAH06071536GA, SAH05391087GA, SAH06081572GA, SAH06071542GA, SAH06071552GA, SAH06081560GA, SAH06081580GA, SAH06081575GA, SAH06081579GA, SAH06101656GA, SAH06101661GA, SAH06121700GA, SAH06141748GA, SAH06332039GA, SAH07142612GA, SAH07292844GA, SAH07403015GA, SAH06101648GA, SAH06121696GA, SAH06121712GA, SAH06171797GA, SAH08423905GA, SAH08423906GA, SAH09024164GA, SAH06041480GA, SAH06041483GA, SAH06462304GA, SAH07493271GA, SAH08033391GA, SAH08303677GA, SAH08373797GA, SAH06502359GA, SAH08443937GA, SAH08443942GA, SAH08443946GA, SAH04420225GA, SAH04420228GA, SAH04420230GA, SAH06522401GA, SAH07062534GA, SAH07062537GA, SAH07062541GA, SAH05150786GA, SAH05220841GA, SAH05471257GA, SAH06291997GA, SAH06291998GA, SAH06292009GA, SAH08343716GA, SAH08343718GA, SAH08443927GA, SAH08443930GA, SAH08443935GA, SAH08443941GA, SAH07142617GA, SAH05471251GA, SAH05471253GA, SAH05471254GA, SAH08283628GA, SAH08283629GA, SAH04500420GA, SAH06181844GA, SAH06181849GA, SAH06181851GA, SAH07212735GA, SAH07212743GA, SAH07212744GA, SAH07212745GA, SAH07212746GA, SAH07212748GA, SAH07342929GA, SAH08504066GA, SAH08504074GA, SAH08514102GA, SAH08514103GA, SAH08514104GA, SAH08514110GA, SAH08514111GA, SAH08524124GA, SAH04530500GA, SAH05040567GA, SAH05040570GA, SAH05040583GA, SAH05040587GA, SAH05040592GA, SAH05070607GA, SAH05080660GA, SAH05471260GA, SAH05471263GA, SAH05471266GA, SAH05501379GA, SAH06011430GA, SAH06011431GA, SAH06051494GA, SAH06051496GA, SAH06051497GA, SAH06051498GA, SAH06051499GA, SAH06241965GA, SAH06342061GA, SAH06342062GA, SAH06412142GA, SAH06422163GA, SAH07172682GA, SAH07423075GA, SAH07423080GA, SAH07423105GA, SAH07423107GA, SAH07483215GA, SAH07483217GA, SAH07483221GA, SAH07483223GA, SAH07483228GA, SAH07483231GA, SAH07483232GA, SAH07483233GA, SAH07493278GA, SAH08023372GA, SAH08253593GA, SAH08373811GA, SAH08473970GA, SAH04430240GA, SAH05160815GA, SAH05160817GA, SAH05160820GA, SAH05160821GA, SAH05160822GA, SAH06041481GA, SAH06061528GA, SAH06081569GA, SAH04430236GA, SAH04430243GA, SAH04430253GA, SAH05080655GA, SAH05080659GA, SAH05280952GA, SAH04430257GA, SAH07142641GA, SAH07172700GA, SAH04470311GA, SAH04480357GA, SAH04500413GA, SAH04500416GA, SAH04530491GA, SAH05260919GA, SAH05320985GA, SAH05320990GA, SAH05351024GA, SAH05411110GA, SAH05421128GA, SAH05421129GA, SAH05451215GA, SAH05451216GA, SAH05451222GA, SAH05461230GA, SAH05461234GA, SAH06041476GA, SAH06051513GA, SAH06091615GA, SAH06121704GA, SAH06151766GA, SAH06241874GA, SAH06241878GA, SAH06402121GA, SAH06412129GA, SAH06412132GA, SAH06432202GA, SAH06432209GA, SAH06522427GA, SAH06522428GA, SAH06522429GA, SAH09024159GA, SAH07072578GA, SAH04500421GA, SAH06171784GA, SAH06171811GA, SAH06181839GA, SAH06181852GA, SAH08213552GA, SAH06021443GA, SAH06021444GA, SAH06041466GA, SAH06041467GA, SAH06041468GA, SAH06041469GA, SAH07423111GA, SAH07483224GA, SAH07443150GA, SAH08023321GA, SAH08023322GA, SAH08023323GA, SAH08023324GA, SAH08023325GA, SAH08023326GA, SAH08023328GA, SAH08023329GA, SAH08023334GA, SAH04350055GA, SAH04410199GA, SAH04410200GA, SAH04510427GA, SAH04510431GA, SAH04510433GA, SAH04510457GA, SAH04530496GA, SAH04530502GA, SAH05080671GA, SAH05160811GA, SAH05240869GA, SAH05240871GA, SAH05320994GA, SAH05391100GA, SAH05411114GA, SAH05411115GA, SAH05471273GA, SAH05471274GA, SAH06071540GA, SAH06071541GA, SAH06071543GA, SAH06091603GA, SAH06091608GA, SAH06121695GA, SAH06121699GA, SAH06191867GA, SAH06342054GA, SAH06342055GA, SAH06432186GA, SAH06432187GA, SAH07232789GA, SAH07232790GA, SAH07312901GA, SAH07463183GA, SAH07483189GA, SAH08023370GA, SAH08173480GA, SAH08173481GA, SAH04470312GA, SAH04470313GA, SAH04500409GA, SAH04510426GA, SAH08133400GA, SAH08133403GA, SAH04480345GA, SAH06522414GA, SAH05280936GA, SAH06291999GA, SAH04530504GA, SAH06091606GA, SAH06422180GA, SAH06422165GA, SAH06462276GA, SAH04490368GA, SAH08504068GA, SAH08504075GA, SAH08504076GA, SAH08504077GA, SAH08504078GA, SAH08504084GA, SAH08514086GA, SAH08514090GA, SAH08514091GA, SAH08514092GA, SAH08514093GA, SAH08303667GA, SAH08443945GA, SAH08514097GA, SAH08514098GA, SAH08514099GA, SAH06452268GA, SAH06452269GA, SAH06452270GA, SAH06452271GA, SAH06452272GA, SAH06452273GA, SAH08023337GA, SAH08023354GA, SAH08253614GA, SAH08293656GA, SAH08173488GA, SAH08173489GA, SAH08173504GA, SAH05451204GA, SAH05451214GA, SAH05451218GA, SAH05451219GA, SAH06241967GA, SAH06452259GA, SAH06452263GA, SAH06452264GA, SAH04400170GA, SAH04400175GA, SAH04400177GA, SAH04400178GA, SAH04430249GA, SAH04430255GA, SAH04430256GA, SAH04480353GA, SAH05290976GA, SAH05431155GA, SAH05431156GA, SAH05437361GA, SAH06382086GA, SAH06382087GA, SAH06382088GA, SAH06422182GA, SAH06432206GA, SAH07102584GA, SAH08023353GA, SAH08504067GA, SAH08504072GA, SAH09024180GA, SAH09024181GA, SAH09024196GA, SAH09024197GA, SAH09024200GA, SAH09024201GA, SAH09024203GA, SAH07312902GA, SAH07312905GA, SAH07312911GA, SAH08343710GA, SAH06522373GA, SAH06522374GA, SAH06522377GA, SAH04350044GA, SAH04350048GA, SAH04420232GA, SAH04450272GA, SAH04530485GA, SAH04530486GA, SAH05010536GA, SAH05010543GA, SAH05010547GA, SAH05010563GA, SAH05010564GA, SAH05290957GA, SAH05290960GA, SAH05321012GA, SAH05351037GA, SAH05391084GA, SAH05391091GA, SAH05391094GA, SAH05391097GA, SAH05391098GA, SAH05431143GA, SAH05431150GA, SAH06091594GA, SAH06091600GA, SAH06091610GA, SAH06121701GA, SAH06121707GA, SAH06121709GA, SAH06121728GA, SAH06141734GA, SAH06141759GA, SAH06171786GA, SAH06171796GA, SAH06442230GA, SAH06522372GA, SAH06522375GA, SAH06522380GA, SAH06522402GA, SAH06522411GA, SAH07052510GA, SAH07072555GA, SAH07102609GA, SAH07152649GA, SAH07372947GA, SAH07372989GA, SAH07372990GA, SAH08303671GA, SAH08423890GA, SAH08473987GA, SAH08473993GA, SAH08484004GA, SAH08494030GA, SAH08504082GA, SAH08514085GA, SAH08514094GA, SAH08514096GA, SAH08514112GA, SAH08514113GA, SAH08514114GA, SAH08253609GA, SAH08303665GA, SAH08303682GA, SAH08313698GA, SAH08313699GA, SAH08313704GA, SAH07062526GA, SAH08143432GA, SAH06061534GA, SAH06071539GA, SAH06081554GA, SAH06081562GA, SAH06081570GA, SAH06091582GA, SAH06091589GA, SAH06091631GA, SAH06522383GA, SAH06522390GA, SAH06522415GA, SAH07403053GA, SAH07403063GA, SAH04350056GA, SAH04360076GA, SAH07022487GA, SAH07052515GA, SAH05070608GA, SAH05070613GA, SAH05070618GA, SAH06452262GA, SAH06452266GA, SAH07111313GA, SAH07152660GA, SAH04470325GA, SAH04470327GA, SAH04490360GA, SAH04490384GA, SAH06191866GA, SAH08203541GA, SAH08253585GA, SAH08253586GA, SAH08253587GA, SAH08253590GA, SAH08253592GA, SAH07372964GA, SAH07403037GA, SAH07403041GA, SAH07152647GA, SAH07172694GA, SAH07172695GA, SAH07172696GA, SAH07172697GA, SAH07172698GA, SAH07463164GA, SAH08173497GA, SAH08343714GA, SAH08343722GA, SAH06452258GA, SAH06452261GA, SAH06462299GA, SAH06462309GA, SAH06462311GA, SAH06502347GA, SAH08494028GA, SAH06081571GA, SAH06091623GA, SAH05321003GA, SAH08403875GA, SAH08423907GA, SAH08423911GA, SAH08423913GA, SAH08423922GA, SAH04500395GA, SAH04500397GA, SAH04500398GA, SAH04500400GA, SAH04500401GA, SAH04500402GA, SAH04500407GA, SAH04500411GA, SAH04500417GA, SAH04500419GA, SAH05240891GA, SAH07423117GA, SAH08023327GA, SAH08343705GA, SAH08343712GA, SAH08383837GA, SAH08403867GA, SAH08403868GA, SAH08403869GA, SAH08403872GA, SAH08403873GA, SAH08403874GA, SAH08403877GA, SAH08403878GA, SAH08403879GA, SAH08403881GA, SAH08403883GA, SAH08403884GA, SAH08494026GA, SAH04500415GA, SAH04510459GA, SAH04520473GA, SAH04520478GA, SAH06412126GA, SAH06181845GA, SAH06241934GA, SAH06241940GA, SAH06241952GA, SAH07052508GA, SAH07282808GA, SAH04350058GA, SAH04350060GA, SAH05421121GA, SAH05351025GA, SAH06081564GA, SAH06091630GA, SAH06101659GA, SAH06101660GA, SAH06121713GA, SAH06121725GA, SAH06291981GA, SAH06382070GA, SAH06522398GA, SAH07022482GA, SAH07062553GA, SAH07102602GA, SAH07102603GA, SAH07102604GA, SAH07102607GA, SAH07282804GA, SAH07292851GA, SAH07312882GA, SAH07493313GA, SAH05110699GA, SAH06101693GA, SAH06051501GA, SAH06051503GA, SAH08303683GA, SAH07342927GA, SAH04430248GA, SAH06522386GA, SAH06522391GA, SAH07312897GA, SAH07312909GA, SAH04410195GA, SAH05521422GA, SAH06402111GA, SAH07142619GA, SAH07142620GA, SAH07282817GA, SAH07403019GA, SAH07443131GA, SAH07443132GA, SAH07443133GA, SAH07443134GA, SAH07443136GA, SAH07443138GA, SAH07443139GA, SAH04420224GA, SAH04420227GA, SAH05280951GA, SAH05351019GA, SAH05431164GA, SAH08023344GA, SAH08133420GA, SAH05461225GA, SAH05461226GA, SAH05461227GA, SAH05461228GA, SAH05461229GA, SAH05461231GA, SAH05461232GA, SAH05461233GA, SAH05471239GA, SAH06061530GA, SAH08203529GA, SAH05240881GA, SAH05411105GA, SAH08213557GA, SAH06452256GA, SAH05140728GA, SAH05280940GA, SAH06041491GA, SAH06241889GA, SAH06291977GA, SAH04500412GA, SAH04510430GA, SAH07072571GA, SAH062422176GA, SAH06422176GA, SAH06011426GA, SAH07142615GA, SAH07142621GA, SAH07282795GA, SAH07302866GA, SAH07302868GA, SAH07302869GA, SAH07302870GA, SAH07302871GA, SAH05391095GA, SAH05391102GA, SAH06091624GA, SAH06091625GA, SAH07423119GA, SAH07493267GA, SAH07493238GA, SAH07493243GA, SAH06241960GA, SAH07463155GA, SAH07463161GA, SAH07463167GA, SAH07463171GA, SAH07463175GA, SAH08033389GA, SAH08033393GA, SAH05280942GA, SAH06181825GA, SAH06181829GA, SAH06181831GA, SAH07423094GA, SAH07423096GA, SAH07022480GA, SAH07022486GA, SAH07152658GA, SAH07152666GA, SAH07172686GA, SAH07212756GA, SAH05451209GA, SAH08023363GA, SAH08033377GA, SAH06151765GA, SAH08213567GA, SAH04290002GA, SAH04290003GA, SAH07312904GA, SAH04360059GA, SAH04460306GA, SAH04460307GA, SAH04530493GA, SAH04530494GA, SAH05451210GA, SAH05501385GA, SAH06432204GA, SAH06432205GA, SAH06432210GA, SAH06452247GA, SAH06452249GA, SAH06452251GA, SAH06452253GA, SAH06462287GA, SAH06462295GA, SAH06522395GA, SAH06522397GA, SAH06522407GA, SAH06522408GA, SAH06522409GA, SAH06522410GA, SAH08353734GA, SAH08363746GA, SAH08363750GA, SAH08363751GA, SAH08363752GA, SAH08373792GA, SAH07052496GA, SAH07232785GA, SAH05411106GA, SAH04290009GA, SAH07423106GA, SAH07423120GA, SAH08153467GA, SAH08153470GA, SAH08373773GA, SAH08403862GA, SAH04470326GA, SAH06241957GA, SAH06462305GA, SAH06482329GA, SAH06292012GA, SAH06332036GA, SAH06342044GA, SAH06342046GA, SAH06342047GA, SAH06342052GA, SAH06342053GA, SAH07212762GA, SAH07312899GA, SAH07483220GA, SAH05290969GA, SAH05391092GA, SAH04340035GA, SAH04350052GA, SAH06412125GA, SAH04500403GA, SAH07172683GA, SAH04290004GA, SAH04290005GA, SAH04290007GA, SAH04290010GA, SAH04290026GA, SAH07312898GA, SAH07312900GA, SAH08343723GA, SAH08343709GA, SAH04470316GA, SAH05280949GA, SAH05220847GA, SAH04510443GA, SAH04510446GA, SAH04510447GA, SAH04510450GA, SAH04510451GA, SAH04510453GA, SAH04510454GA, SAH08293648GA, SAH07052516GA, SAH06462301GA, SAH08203543GA, SAH08283635GA, SAH07493312GA, SAH07443141GA, SAH08293660GA, SAH08153460GA, SAH05140724GA, SAH07232769GA, SAH07232773GA, SAH06101654GA, SAH07423122GA, SAH07423124GA, SAH08023365GA, SAH08143429GA, SAH08153462GA, SAH08153464GA, SAH07443130GA, SAH05491361GA, SAH05491362GA, SAH06412140GA, SAH05391088GA, SAH08443932GA, SAH05040581GA, SAH05150777GA, SAH05150779GA, SAH05220838GA, SAH05220840GA, SAH05220845GA, SAH05140719GA, SAH06452255GA, SAH06091587GA, SAH05260915GA, SAH06241898GA, SAH05481295GA, SAH05481300GA, SAH05481301GA, SAH06482330GA, SAH08173500GA, SAH08173502GA, SAH04430259GA, SAH08423885GA, SAH08423887GA, SAH08494044GA, SAH07443128GA, SAH06332035GA, SAH07062549GA, SAH06522416GA, SAH07302864GA, SAH07152654GA, SAH07072572GA, SAH05110700GA, SAH07443126GA , SAH04450264GA, SAH07102611GA, SAH07423116GA, SAH09024198GA, SAH05080652GA, SAH05080653GA, SAH05070610GA, SAH08203520GA, SAH07372979GA, SAH04470330GA, SAH04500394GA, SAH06241896GA, SAH06241902GA, SAH06432213GA, SAH06522399GA, SAH07052521GA, SAH09024185GA, SAH07503319GA, SAH08033387GA, SAH06171800GA, SAH06191855GA, SAH05080644GA, SAH05471245GA, SAH05040574GA, SAH08303670GA, SAH05481296GA, SAH06241881GA, SAH05080683GA, SAH05110708GA, SAH05110709GA, SAH08173503GA, SAH05471249GA, SAH05280947GA, SAH06422173GA, SAH06522392GA, SAH07022458GA, SAH07152670GA, SAH07292834GA, SAH08143438GA, SAH08173476, SAH08173476GA, SAH04490369GA, SAH06422183GA, SAH06502348GA, SAH08133404GA, SAH07443137GA, SAH08443928GA, SAH08023340GA, SAH08023346GA, SAH08023348GA, SAH08203507GA, SAH04520468GA, SAH05010552GA, SAH05471243GA, SAH06121716GA, SAH06502344GA, SAH06502345GA, SAH06502349GA, SAH06502350GA, SAH06502351GA, SAH07062546GA, SAH07142613GA, SAH07202718GA, SAH07503315GA, SAH08033376GA, SAH05040580GA, SAH04460296GA, SAH04480352GA, SAH04480354GA, SAH04480358GA, SAH07493247GA, SAH07493250GA, SAH08313687GA, SAH06292003GA, SAH04480344GA, SAH05080681GA, SAH06081559GA, SAH07463160GA, SAH08293645GA, SAH08373765GA, SAH08373768GA, SAH08494050GA, SAH08494064GA, SAH07463182GA, SAH06482334GA, SAH08423909GA, SAH08473975GA, SAH07403004GA, SAH09024207GA, SAH08514105GA, SAH08514106GA, SAH07012449GA, SAH05320995GA, SAH05471268GA, SAH05471270GA, SAH05391104GA, SAH06522403GA, SAH04430244GA, SAH08373790GA, SAH05491331GA, SAH06522388GA, SAH06452265GA, SAH08423921GA, SAH05110704GA, SAH04410188GA, SAH08153450GA, SAH05280925GA, SAH06071551GA, SAH05110690GA, SAH05280937GA, SAH05220849GA, SAH07403009GA, SAH06121719GA, SAH07292840GA, SAH05070611GA, SAH05411112GA, SAH07022466GA, SAH07403006GA, SAH05160814GA, SAH07423065GA, SAH06432200GA, SAH05421125GA, SAH05160816GA, SAH06241968GA, SAH04410203GA, SAH05431131GA, SAH08343711GA, SAH04460305GA, SAH05471238GA, SAH08473986GA, SAH05160833GA, SAH5160832GA, SAH07012444GA, SAH06101664GA, SAH07493272GA, SAH07423072GA, SAH05320988GA, SAH07372961GA, SAH05220860GA, SAH05441168GA, SAH06241947GA, SAH06291994GA, SAH08213550GA, SAH05290955GA, SAH08504065GA, SAH07152662GA, SAH07172684GA, SAH07172684GS, SAH07202732GA, SAH05351036GA, SAH05411113GA, SAH06291996GA, SAH07052539GA, SAH07423086GA, SAH07302863GA, SAH04480351GA, SAH06241961GA, SAH05320992GA, SAH08293662GA, SAH08303678GA, SAH08313686GA, SAH06452260GA, SAH06141735GA, SAH07312910GA, SAH07172685GA, SAH06241964GA, SAH06292011GA, SAH07403026GA, SAH06241973GA, SAH08293646GA, SAH06011434GA, SAH07312883GA, SAH07292833GA, SAH07202706GA, SAH07423114GA, SAH07423110GA, SAH08173498GA, SAH07403008GA, SAH08173493GA, SAH08373786GA, SAH08443931GA, SAH07423083GA, SAH08504069GA, SAH07403010GA, SAH04430254GA, SAH08423919GA, SAH07423084GA, SAH07423068GA, SAH07423073GA, SAH07423074GA, SAH05451212GA, SAH06291975GA, SAH07292835GA, SAH07403058GA, SAH06241919GA, SAH07072564GA, SAH06071549GA, SAH06101644GA, SAH06291974GA, SAH08473991GA, SAH08473997GA, SAH06081561GA, SAH08213551GA, SAH08494061GA, SAH06482331GA, SAH06482343GA, SAH08494055GA, SAH06502360GA, SAH07423101GA, SAH06363221GA, SAH06462283GA, SAH06482333GA, SAH08203536GA, SAH05240879GA, SAH06181836GA, SAH09024195GA, SAH07403045GA, SAH07403011GA, SAH06412149GA, SAH06241972GA, SAH04340037GA, SAH04520462GA, SAH045220462GA, SAH06141741GA, SAH07152645GA, SAH0715643GA, SAH08504080GA, SAH07282797GA, SAH07463163GA, SAH08133413GA, SAH04510439GA, SAH04410187GA, SAH06432208GA, SAH05140736GA, SAH07423112GA, SAH05290962GA, SAH05240884GA, SAH06141760GA, SAH08373822GA, SAH06502358GA, SAH05451203GA, SAH08443959GA, SAH06382071GA, SAH06041488GA, SAH07062533GA, SAH07142618GA, SAH07202712GA, SAH07202719GA, SAH07202720GA, SAH08213579GA, SAH06171787GA, SAH04530489GA, SAH06051495GA, SAH09024187GA, SAH06462274GA, SAH05521409GA, SAH06171804GA, SAH06482337GA, SAH05080676GA, SAH06241903GA, SAH08173491GA, SAH08504079GA, SAH08153474GA, SAH04470331GA, SAH07372950GA, SAH08133398GA, SAH08033362GA, SAH08033382GA, SAH04480334GA, SAH08213571GA, SAH07072575GA, SAH06332028GA, SAH07403060GA, SAH06422168GA, SAH07062536GA, SAH07052513GA, SAH07102605GA, SAH07102606GA, SAH08033380GA, sah05010558ga, SAH07052547GA, SAH08373778GA, SAH08494033GA, SAH05010548GA, SAH08023355GA, SAH05140734GA, SAH04480339GA, SAH05290973GA, SAH06071544GA, SAH06522385GA, SAH07012447GA, SAH07022471GA, SAH06081557GA, SAH05080663GA, SAH0508633GA, SAH05160830GA, SAH07403051GA, SAH07022456GA, SAH08283631GA, SAH08373823GA, SAH052408826GA, sah05292295ga, SAH07312906GA, SAH06091602, SAH0702565GA, SAH05320996GA, SAH08514095GA, SAH08514101GA, SAH05481290GA, SAH05160805GA.  UPDATE added SAH05080669GA, SAH05080651GA.  5. Product Code: SE2, UDI Number (Serial Numbers): 00840682109260SE217350041SA (SE217350041SA), 00840682109260SE213410181GA (SE213410181GA);    6. Product Code: SEE, UDI Number (Serial Numbers): Not Applicable (SEE11361693GA), Not Applicable (SEE11381821GA), Not Applicable (SEE11381824GA), Not Applicable (SEE12420169GA), Not Applicable (SEE15173206GA), Not Applicable (SEE15173214GA), Not Applicable (SEE09530903GA), Not Applicable (SEE10430627GA), Not Applicable (SEE16324000SA), Not Applicable (SEE16324032SA), Not Applicable (SEE12182218GA), Not Applicable (SEE12041973GA), Not Applicable (SEE12062000GA), Not Applicable (SEE12312406GA), Not Applicable (SEE12322413GA), Not Applicable (SEE12322420GA), Not Applicable (SEE12322422GA), Not Applicable (SEE12322423GA), Not Applicable (SEE12322425GA), Not Applicable (SEE12322428GA), Not Applicable (SEE12322429GA), Not Applicable (SEE12322430GA), Not Applicable (SEE12322431GA), Not Applicable (SEE12322435GA), Not Applicable (SEE12322436GA), Not Applicable (SEE12322437GA), Not Applicable (SEE12332441GA), Not Applicable (SEE12332444GA), Not Applicable (SEE12400093GA), Not Applicable (SEE12400095GA), Not Applicable (SEE13220834GA), Not Applicable (SEE15032878GA), Not Applicable (SEE15032895GA), Not Applicable (SEE15042900GA), Not Applicable (SEE15042902GA), Not Applicable (SEE15042903GA), Not Applicable (SEE15042904GA), Not Applicable (SEE15042905GA), Not Applicable (SEE15042907GA), Not Applicable (SEE15446336SA), Not Applicable (SEE15456714SA), Not Applicable (SEE15456725SA), Not Applicable (SEE15456729SA), Not Applicable (SEE15477106SA), Not Applicable (SEE15477108SA), Not Applicable (SEE15477109SA), Not Applicable (SEE15477113SA), Not Applicable (SEE15477115SA), Not Applicable (SEE16380005SA), Not Applicable (SEE16380009SA), Not Applicable (SEE16380010SA), Not Applicable (SEE16380011SA), Not Applicable (SEE16380012SA), Not Applicable (SEE16380013SA), Not Applicable (SEE16380016SA), Not Applicable (SEE16380017SA), Not Applicable (SEE16380018SA), Not Applicable (SEE16380023SA), 00840682109260SEE18500020SA (SEE18500020SA), 00840682109260SEE18500022SA (SEE18500022SA), 00840682109260SEE18500033SA (SEE18500033SA), 00840682109260SEE18500042SA (SEE18500042SA), Not Applicable (SEE11361705GA), Not A pplicable (SEE11361711GA), Not Applicable (SEE11361714GA), Not Applicable (SEE11361720GA), Not Applicable (SEE11361739GA), 00840682109260SEE19120019SA (SEE19120019SA), 00840682109260SEE19120028SA (SEE19120028SA), 00840682109260SEE19150002SA (SEE19150002SA), 00840682109260SEE19160001SA (SEE19160001SA), 00840682109260SEE19160003SA (SEE19160003SA), 00840682109260SEE19160004SA (SEE19160004SA), 00840682109260SEE19160005SA (SEE19160005SA), 00840682109260SEE19160006SA (SEE19160006SA), 00840682109260SEE19160007SA (SEE19160007SA), 00840682109260SEE19160008SA (SEE19160008SA), 00840682109260SEE19160009SA (SEE19160009SA), 00840682109260SEE19160011SA (SEE19160011SA), 00840682109260SEE19160012SA (SEE19160012SA), 00840682109260SEE19160013SA (SEE19160013SA), Not Applicable (SEE15291745SA), Not Applicable (SEE15291767SA), Not Applicable (SEE15291799SA), Not Applicable (SEE12172193GA), Not Applicable (SEE11151313GA), Not Applicable (SEE11161324GA), Not Applicable (SEE16430004SA), Not Applicable (SEE16430015SA), Not Applicable (SEE16430017SA), Not Applicable (SEE16430018SA), Not Applicable (SEE16430019SA), Not Applicable (SEE16430022SA), Not Applicable (SEE16430026SA), Not Applicable (SEE16430027SA), Not Applicable (SEE16430029SA), 00840682109260SEE18290001SA (SEE18290001SA), 00840682109260SEE18040045SA (SEE18040045SA), 00840682109260SEE18040050SA (SEE18040050SA), Not Applicable (SEE12122151GA), Not Applicable (SEE12122153GA), Not Applicable (SEE12122155GA), Not Applicable (SEE12122158GA), Not Applicable (SEE14151862GA), Not Applicable (SEE14151883GA), Not Applicable (SEE14161910GA), Not Applicable (SEE14171917GA), Not Applicable (SEE14222043GA), Not Applicable (SEE14222044GA), Not Applicable (SEE14222049GA), Not Applicable (SEE14222060GA), Not Applicable (SEE14222062GA), Not Applicable (SEE14222065GA), Not Applicable (SEE14222074GA), Not Applicable (SEE14222077GA), Not Applicable (SEE14452624GA), Not Applicable (SEE15113176GA), Not Applicable (SEE15113182GA), Not Applicable (SEE09100090GA), Not Applicable (SEE09140180GA), Not Applicable (SEE09140182GA), Not Applicable (SEE09140185GA), Not Applicable (SEE09140189GA), Not Applicable (SEE09140193GA), Not Applicable (SEE09140196GA), Not Applicable (SEE11171359GA), Not Applicable (SEE11171381GA), Not Applicable (SEE11181393GA), Not Applicable (SEE11181396GA), Not Applicable (SEE11191425GA), Not Applicable (SEE15113151GA), Not Applicable (SEE16420072SA), Not Applicable (SEE16430023SA), Not Applicable (SEE11361706GA), Not Applicable (SEE11361737GA), Not Applicable (SEE15446244SA), Not Applicable (SEE15446256SA), Not Applicable (SEE10530847GA), 00840682109260SEE18040016SA (SEE18040016SA), 00840682109260SEE18040042SA (SEE18040042SA), 00840682109260SEE18040044SA (SEE18040044SA), 00840682109260SEE18040046SA (SEE18040046SA), Not Applicable (SEE12282348GA), Not Applicable (SEE12282351GA), Not Applicable (SEE12390077GA), Not Applicable (SEE12390079GA), Not Applicable (SEE12390084GA), Not Applicable (SEE15456727SA), Not Applicable (SEE11361709GA), Not Applicable (SEE11361724GA), Not Applicable (SEE11361730GA), Not Applicable (SEE13451325GA), Not Applicable (SEE13451326GA), Not Applicable (SEE13451327GA), Not Applicable (SEE11161339GA), Not Applicable (SEE11161341GA), Not Applicable (SEE11161344GA), Not Applicable (SEE11161346GA), Not Applicable (SEE11201467GA), Not Applicable (SEE11231531GA), Not Applicable (SEE11231537GA), Not Applicable (SEE11231540GA), Not Applicable (SEE15312100SA), Not Applicable (SEE15261403SA), Not Applicable (SEE14171915GA), Not Applicable (SEE14171926GA), Not Applicable (SEE14171927GA), Not Applicable (SEE12510304GA), Not Applicable (SEE12510316GA), Not Applicable (SEE12520340GA), Not Applicable (SEE12520345GA), Not Applicable (SEE12520348GA), Not Applicable (SEE13000390GA), Not Applicable (SEE13000391GA), Not Applicable (SEE11181394GA), Not Applicable (SEE12182215GA), Not Applicable (SEE12182223GA), Not Applicable (SEE12292367GA), Not Applicable (SEE13150668GA), Not Applicable (SEE09040063GA), Not Applicable (SEE16221554SA), Not Applicable (SEE16221560SA), Not Applicable (SEE16221585SA), Not Applicable (SEE16221587SA), Not Applicable (SEE16221588SA), Not Applicable (SEE16221591SA), Not Applicable (SEE16221592SA), Not Applicable (SEE12410122GA), Not Applicable (SEE12410130GA), Not Applicable (SEE12410143GA), Not Applicable (SEE12420150GA), Not Applicable (SEE12172201GA), Not Applicable (SEE12172204GA), 00840682109260SEE19220024SA (SEE19220024SA), 00840682109260SEE19220035SA (SEE19220035SA), 00840682109260SEE18370005SA (SEE18370005SA), Not Applicable (SEE11361710GA), Not Applicable (SEE11361718GA), Not Applicable (SEE14272163GA), Not Applicable (SEE14272165GA), Not Applicable (SEE14272169GA), Not Applicable (SEE14272176GA), Not Applicable (SEE14272202GA), Not Applicable (SEE14272204GA), Not Applicable (SEE14272217GA), Not Applicable (SEE14292230GA), Not Applicable (SEE14292234GA), Not Applicable (SEE14292235GA), Not Applicable (SEE14292236GA), Not Applicable (SEE15363354SA), Not Applicable (SEE15363355SA), Not Applicable (SEE13240862GA), Not Applicable (SEE13351012GA), Not Applicable (SEE13451329GA), Not Applicable (SEE13451332GA), 00840682109260SEE19100053SA (SEE19100053SA), Not Applicable (SEE14151815GA), Not Applicable (SEE09100130GA), Not Applicable (SEE09230279GA), Not Applicable (SEE10420570GA), Not Applicable (SEE10460712GA), Not Applicable (SEE12430189GA), Not Applicable (SEE12430195GA), Not Applicable (SEE12430197GA), Not Applicable (SEE12430211GA), Not Applicable (SEE12430219GA), Not Applicable (SEE09530899GA), 00840682109260SEE17450001SA (SEE17450001SA), 00840682109260SEE17450002SA (SEE17450002SA), Not Applicable (SEE10410502GA), Not Applicable (SEE10420551GA), Not Applicable (SEE11371755GA), Not Applicable (SEE11371763GA), Not Applicable (SEE15415110SA), Not Applicable (SEE15415111SA), Not Applicable (SEE15415114SA), Not Applicable (SEE15415115SA), Not Applicable (SEE15415116SA), Not Applicable (SEE15415117SA), Not Applicable (SEE15415164SA), Not Applicable (SEE11201470GA), Not Applicable (SEE16262162SA), Not Applicable (SEE16272328SA), Not Applicable (SEE13391130GA), Not Applicable (SEE15113121GA), Not Applicable (SEE15113134GA), Not Applicable (SEE14462659GA), Not Applicable (SEE14462669GA), Not Applicable (SEE14462670GA), Not Applicable (SEE14462671GA), Not Applicable (SEE10270274GA), Not Applicable (SEE10400374GA), Not Applicable (SEE10400442GA), Not Applicable (SEE10400443GA), Not Applicable (SEE10400459GA), Not Applicable (SEE10410480GA), Not Applicable (SEE10410483GA), Not Applicable (SEE10410507GA), Not Applicable (SEE10410511GA), Not Applicable (SEE15415093SA), 00840682109260SEE18290012SA (SEE18290012SA), 00840682109260SEE17180002SA (SEE17180002SA), 00840682109260SEE17350011SA (SEE17350011SA), Not Applicable (SEE11071010GA), Not Applicable (SEE11071016GA), Not Applicable (SEE12092067GA), Not Applicable (SEE12242267GA), Not Applicable (SEE12252297GA), Not Applicable (SEE12252306GA), Not Applicable (SEE12370056GA), Not Applicable (SEE13190793GA), Not Applicable (SEE13190794GA), Not Applicable (SEE13190795GA), Not Applicable (SEE13190796GA), Not Applicable (SEE13190797GA), Not Applicable (SEE13190799GA), Not Applicable (SEE13190800GA), Not Applicable (SEE13190801GA), Not Applicable (SEE13190802GA), Not Applicable (SEE15528205SA), Not Applicable (SEE15528213SA), Not Applicable (SEE15528214SA), Not Applicable (SEE15528215SA), Not Applicable (SEE15528216SA), Not Applicable (SEE16303219SA), 00840682109260SEE18040049SA (SEE18040049SA), 00840682109260SEE18290002SA (SEE18290002SA), 00840682109260SEE18290007SA (SEE18290007SA), 00840682109260SEE18290011SA (SEE18290011SA), 00840682109260SEE18290016SA (SEE18290016SA), 00840682109260SEE18390002SA (SEE18390002SA), 00840682109260SEE18390005SA (SEE18390005SA), 00840682109260SEE18390014SA (SEE18390014SA), 00840682109260SEE18390020SA (SEE18390020SA), 00840682109260SEE18400002SA (SEE18400002SA), 00840682109260SEE18400004SA (SEE18400004SA), 00840682109260SEE18400006SA (SEE18400006SA), 00840682109260SEE18400009SA (SEE18400009SA), 00840682109260SEE18400010SA (SEE18400010SA), 00840682109260SEE18400011SA (SEE18400011SA), 00840682109260SEE18400014SA (SEE18400014SA), 00840682109260SEE18400020SA (SEE18400020SA), 00840682109260SEE18470023SA (SEE18470023SA), 00840682109260SEE18510015SA (SEE18510015SA), 00840682109260SEE18510016SA (SEE18510016SA), 00840682109260SEE18510019SA (SEE18510019SA), 00840682109260SEE19110024SA (SEE19110024SA), Not Applicable (SEE15093065GA), Not Applicable (SEE15093069GA), Not Applicable (SEE15113127GA), Not Applicable (SEE15113133GA), Not Applicable (SEE11081200GA), Not Applicable (SEE11081202GA), Not Applicable (SEE11101214GA), Not Applicable (SEE11101223GA), Not Applicable (SEE11111227GA), Not Applicable (SEE11111229GA), Not Applicable (SEE11111230GA), Not Applicable (SEE11111235GA), Not Applicable (SEE11111242GA), Not Applicable (SEE11111243GA), Not Applicable (SEE12430212GA), Not Applicable (SEE12430214GA), Not Applicable (SEE13010401GA), Not Applicable (SEE13401145GA), Not Applicable (SEE13401174GA), Not Applicable (SEE13401176GA), Not Applicable (SEE13401180GA), Not Applicable (SEE13401185GA), Not Applicable (SEE14031603GA), Not Applicable (SEE15312074SA), Not Applicable (SEE15312075SA), Not Applicable (SEE15312076SA), Not Applicable (SEE15312077SA), Not Applicable (SEE15312078SA), Not Applicable (SEE15312079SA), Not Applicable (SEE15312081SA), Not Applicable (SEE15312082SA), Not Applicable (SEE15312083SA), Not Applicable (SEE15312084SA), Not Applicable (SEE15312085SA), Not Applicable (SEE15312086SA), 00840682109260SEE17130002SA (SEE17130002SA), 00840682109260SEE17130004SA (SEE17130004SA), 00840682109260SEE17130007SA (SEE17130007SA), Not Applicable (SEE10410509GA), Not Applicable (SEE14402490GA), 00840682109260SEE17260003SA (SEE17260003SA), 00840682109260SEE17260004SA (SEE17260004SA), Not Applicable (SEE12322438GA), Not Applicable (SEE16324006SA), Not Applicable (SEE14101787GA), Not Applicable (SEE14111804GA), Not Applicable (SEE10450677GA), Not Applicable (SEE10450683GA), Not Applicable (SEE15113116GA), Not Applicable (SEE15113130GA), Not Applicable (SEE15113136GA), Not Applicable (SEE15240991SA), Not Applicable (SEE09380531GA), Not Applicable (SEE09390548GA), Not Applicable (SEE16028367SA), 00840682109260SEE18400019SA (SEE18400019SA), Not Applicable (SEE12340005GA), Not Applicable (SEE12340013GA), Not Applicable (SEE12360037GA), Not Applicable (SEE12360040GA), Not Applicable (SEE15352668SA), Not Applicable (SEE15352678SA), Not Applicable (SEE15538297SA), Not Applicable (SEE16028320SA), Not Applicable (SEE16028321SA), 00840682109260SEE19220021SA (SEE19220021SA), Not Applicable (SEE09530855GA), Not Applicable (SEE09530857GA), Not Applicable (SEE09530860GA), Not Applicable (SEE11201442GA), Not Applicable (SEE11201443GA), Not Applicable (SEE11201454GA), Not Applicable (SEE11231544GA), Not Applicable (SEE15446199SA), Not Applicable (SEE15446213SA), 00840682109260SEE17420002SA (SEE17420002SA), 00840682109260SEE17500019SA (SEE17500019SA), 00840682109260SEE19130011SA (SEE19130011SA), Not Applicable (SEE13040415GA), Not Applicable (SEE12092062GA), Not Applicable (SEE12092064GA), Not Applicable (SEE12092065GA), Not Applicable (SEE12430183GA), Not Applicable (SEE12430193GA), Not Applicable (SEE12430204GA), Not Applicable (SEE15220797SA), Not Applicable (SEE15220798SA), Not Applicable (SEE13351022GA), Not Applicable (SEE13351036GA), Not Applicable (SEE13351042GA), Not Applicable (SEE13351046GA), Not Applicable (SEE13351049GA), 00840682109260SEE18030014SA (SEE18030014SA), 00840682109260SEE18030015SA (SEE18030015SA), 00840682109260SEE18030016SA (SEE18030016SA), 00840682109260SEE18030018SA (SEE18030018SA), 00840682109260SEE18030020SA (SEE18030020SA), 00840682109260SEE18410023SA (SEE18410023SA), Not Applicable (SEE10121006GA), Not Applicable (SEE10121024GA), Not Applicable (SEE10130076GA), Not Applicable (SEE11201451GA), Not Applicable (SEE11371792GA), Not Applicable (SEE11371812GA), Not Applicable (SEE12370059GA), Not Applicable (SEE12370064GA), Not Applicable (SEE12370067GA), Not Applicable (SEE13371087GA), Not Applicable (SEE13371090GA), Not Applicable (SEE13371103GA), Not Applicable (SEE12021906GA), Not Applicable (SEE12021910GA), Not Applicable (SEE14151823GA), Not Applicable (SEE14151833GA), Not Applicable (SEE14482688GA), Not Applicable (SEE15052909GA), Not Applicable (SEE15052937GA), Not Applicable (SEE16089178SA), Not Applicable (SEE15120553SA), Not Applicable (SEE15120554SA), Not Applicable (SEE15120555SA), Not Applicable (SEE15120557SA), Not Applicable (SEE15210700SA), Not Applicable (SEE15210701SA), Not Applicable (SEE13110516GA), Not Applicable (SEE09250308GA), Not Applicable (SEE09250309GA), Not Applicable (SEE09250315GA), Not Applicable (SEE09250326GA), Not Applicable (SEE10120011GA), Not Applicable (SEE10120018GA), Not Applicable (SEE10121018GA), Not Applicable (SEE10260239GA), Not Applicable (SEE10270256GA), Not Applicable (SEE14031589GA), Not Applicable (SEE14031608GA), Not Applicable (SEE14031610GA), Not Applicable (SEE14031611GA), Not Applicable (SEE14031616GA), Not Applicable (SEE14031627GA), Not Applicable (SEE15528071SA), 00840682109260SEE17020023SA (SEE17020023SA), 00840682109260SEE17020029SA (SEE17020029SA), 00840682109260SEE19170006SA (SEE19170006SA), 00840682109260SEE19170007SA (SEE19170007SA), 00840682109260SEE19170009SA (SEE19170009SA), 00840682109260SEE19170012SA (SEE19170012SA), 00840682109260SEE19170015SA (SEE19170015SA), 00840682109260SEE19170019SA (SEE19170019SA), 00840682109260SEE19170022SA (SEE19170022SA), 00840682109260SEE19170024SA (SEE19170024SA), 00840682109260SEE19170038SA (SEE19170038SA), 00840682109260SEE19170039SA (SEE19170039SA), 00840682109260SEE19170040SA (SEE19170040SA), 00840682109260SEE19170042SA (SEE19170042SA), Not Applicable (SEE09530914GA), Not Applicable (SEE13120573GA), Not Applicable (SEE13120576GA), 00840682109260SEE19220001SA (SEE19220001SA), 00840682109260SEE19220003SA (SEE19220003SA), Not Applicable (SEE11361742GA), Not Applicable (SEE13270894GA), Not Applicable (SEE13270899GA), Not Applicable (SEE13270901GA), Not Applicable (SEE13280913GA), Not Applicable (SEE13280918GA), 00840682109260SEE17220002SA (SEE17220002SA), 00840682109260SEE17220004SA (SEE17220004SA), 00840682109260SEE17220005SA (SEE17220005SA), Not Applicable (SEE09020029GA), Not Applicable (SEE09340440GA), Not Applicable (SEE11371766GA), Not Applicable (SEE11371779GA), Not Applicable (SEE12510306GA), Not Applicable (SEE12510307GA), Not Applicable (SEE12510309GA), Not Applicable (SEE12510314GA), Not Applicable (SEE12510315GA), Not Applicable (SEE12510318GA), Not Applicable (SEE12510319GA), Not Applicable (SEE12510320GA), Not Applicable (SEE12510326GA), Not Applicable (SEE12510327GA), Not Applicable (SEE12510328GA), Not Applicable (SEE12510329GA), Not Applicable (SEE12510330GA), Not Applicable (SEE12510331GA), Not Applicable (SEE12510333GA), Not Applicable (SEE12520339GA), Not Applicable (SEE12520343GA), Not Applicable (SEE13000384GA), Not Applicable (SEE13000388GA), Not Applicable (SEE13050442GA), Not Applicable (SEE14031670GA), Not Applicable (SEE14031676GA), Not Applicable (SEE14031680GA), Not Applicable (SEE14031681GA), Not Applicable (SEE14031682GA), Not Applicable (SEE14031688GA), Not Applicable (SEE15093066GA), Not Applicable (SEE15093068GA), Not Applicable (SEE15103103GA), Not Applicable (SEE15528212SA), Not Applicable (SEE09510793GA), Not Applicable (SEE09510802GA), Not Applicable (SEE09510808GA), Not Applicable (SEE09510819GA), 00840682109260SEE19180010SA (SEE19180010SA), Not Applicable (SEE10420534GA), Not Applicable (SEE10420563GA), Not Applicable (SEE10420574GA), Not Applicable (SEE10430651GA), Not Applicable (SEE10450684GA), Not Applicable (SEE10460696GA), Not Applicable (SEE10470784GA), Not Applicable (SEE11071084GA), Not Applicable (SEE11071144GA), Not Applicable (SEE11071155GA), Not Applicable (SEE11081190GA), Not Applicable (SEE09100083GA), Not Applicable (SEE09100132GA), Not Applicable (SEE09350463GA), Not Applicable (SEE09360471GA), Not Applicable (SEE11060979GA), Not Applicable (SEE11060998GA), Not Applicable (SEE11061004GA), Not Applicable (SEE11161329GA), Not Applicable (SEE11371749GA), Not Applicable (SEE16068746SA), Not Applicable (SEE16241835SA), 00840682109260SEE18110002SA (SEE18110002SA), Not Applicable (SEE11071096GA), Not Applicable (SEE11081187GA), Not Applicable (SEE11101216GA), Not Applicable (SEE11101225GA), Not Applicable (SEE11141250GA), Not Applicable (SEE11171356GA), Not Applicable (SEE12430181GA), Not Applicable (SEE13040427GA), Not Applicable (SEE13110494GA), 00840682109260SEE18290008SA (SEE18290008SA), 00840682109260SEE18290014SA (SEE18290014SA), Not Applicable (SEE15352670SA), Not Applicable (SEE10130039GA), Not Applicable (SEE10130044GA), Not Applicable (SEE10130046GA), Not Applicable (SEE10130048GA), Not Applicable (SEE10130050GA), Not Applicable (SEE11151311GA), Not Applicable (SEE15312141SA), 00840682109260SEE17340004SA (SEE17340004SA), 00840682109260SEE17340006SA (SEE17340006SA), Not Applicable (SEE14482697GA), Not Applicable (SEE09470676GA), Not Applicable (SEE09470685GA), Not Applicable (SEE09490758GA), Not Applicable (SEE10530857GA), Not Applicable (SEE11060992GA), Not Applicable (SEE11061000GA), Not Applicable (SEE11061003GA), Not Applicable (SEE11071088GA), Not Applicable (SEE11071136GA), Not Applicable (SEE11141289GA), Not Applicable (SEE13371093GA), Not Applicable (SEE13371101GA), Not Applicable (SEE15012847GA), 00840682109260SEE17400002SA (SEE17400002SA), 00840682109260SEE17400004SA (SEE17400004SA), Not Applicable (SEE09430620GA), Not Applicable (SEE10250134GA), Not Applicable (SEE10390349GA), Not Applicable (SEE10390355GA), Not Applicable (SEE13411219GA), Not Applicable (SEE13411223GA), Not Applicable (SEE11331598GA), Not Applicable (SEE11361704GA), 00840682109260SEE18160023SA (SEE18160023SA), 00840682109260SEE18160032SA (SEE18160032SA), 00840682109260SEE19120018SA (SEE19120018SA), 00840682109260SEE19160048SA (SEE19160048SA), 00840682109260SEE19170032SA (SEE19170032SA), 00840682109260SEE19170033SA (SEE19170033SA), 00840682109260SEE17260002SA (SEE17260002SA), Not Applicable (SEE11511877GA), 00840682109260SEE17130030SA (SEE17130030SA), 00840682109260SEE17130031SA (SEE17130031SA), 00840682109260SEE17130033SA (SEE17130033SA), 00840682109260SEE17130034SA (SEE17130034SA), 00840682109260SEE19180001SA (SEE19180001SA), 00840682109260SEE19190006SA (SEE19190006SA), Not Applicable (SEE11341614GA), Not Applicable (SEE11341616GA), Not Applicable (SEE11341622GA), Not Applicable (SEE11341627GA), Not Applicable (SEE11341629GA), Not Applicable (SEE11341631GA), Not Applicable (SEE11341633GA), Not Applicable (SEE11341638GA), Not Applicable (SEE11361670GA), Not Applicable (SEE14462657GA), 00840682109260SEE17190003SA (SEE17190003SA), 00840682109260SEE17190004SA (SEE17190004SA), 00840682109260SEE17190008SA (SEE17190008SA), 00840682109260SEE17190009SA (SEE17190009SA), 00840682109260SEE17190010SA (SEE17190010SA), 00840682109260SEE17390008SA (SEE17390008SA), Not Applicable (SEE10400387GA), Not Applicable (SEE10400404GA), Not Applicable (SEE10400432GA), Not Applicable (SEE10400434GA), Not Applicable (SEE11141253GA), Not Applicable (SEE14512831GA), Not Applicable (SEE14512834GA), Not Applicable (SEE10100973GA), Not Applicable (SEE10100976GA), Not Applicable (SEE10100977GA), Not Applicable (SEE10120989GA), Not Applicable (SEE11361708GA), Not Applicable (SEE12092043GA), Not Applicable (SEE12252301GA), Not Applicable (SEE13401179GA), Not Applicable (SEE14151808GA), Not Applicable (SEE14151809GA), Not Applicable (SEE15113168GA), Not Applicable (SEE15113169GA), Not Applicable (SEE15113172GA), Not Applicable (SEE15113184GA), Not Applicable (SEE15113185GA), Not Applicable (SEE15456724SA), Not Applicable (SEE12041951GA), Not Applicable (SEE12041966GA), Not Applicable (SEE14221988GA), Not Applicable (SEE14221993GA), Not Applicable (SEE14222000GA), Not Applicable (SEE15352698SA), Not Applicable (SEE15352789SA), Not Applicable (SEE15352791SA), Not Applicable (SEE15352794SA), Not Applicable (SEE15352844SA), Not Applicable (SEE15352845SA), 00840682109260SEE17190019SA (SEE17190019SA), 00840682109260SEE17190020SA (SEE17190020SA), Not Applicable (SEE09150220GA), Not Applicable (SEE09190239GA), Not Applicable (SEE09190243GA), Not Applicable (SEE09410584GA), Not Applicable (SEE09460646GA), Not Applicable (SEE09530861GA), Not Applicable (SEE10130043GA), Not Applicable (SEE10130055GA), Not Applicable (SEE10130073GA), Not Applicable (SEE10390341GA), Not Applicable (SEE10460718GA), Not Applicable (SEE10460735GA), Not Applicable (SEE12292365GA), Not Applicable (SEE12420155GA), Not Applicable (SEE14031605GA), Not Applicable (SEE14031607GA), Not Applicable (SEE14031613GA), Not Applicable (SEE14031614GA), Not Applicable (SEE14031620GA), Not Applicable (SEE14031622GA), Not Applicable (SEE14031628GA), Not Applicable (SEE14041733GA), Not Applicable (SEE16028319SA), Not Applicable (SEE15052908GA), Not Applicable (SEE15103112GA), Not Applicable (SEE15113145GA), Not Applicable (SEE15113146GA), Not Applicable (SEE15113152GA), 00840682109260SEE17330012SA (SEE17330012SA), 00840682109260SEE17330014SA (SEE17330014SA), 00840682109260SEE17330015SA (SEE17330015SA), 00840682109260SEE17330016SA (SEE17330016SA), Not Applicable (SEE11191434GA), Not Applicable (SEE11211473GA), Not Applicable (SEE11211484GA), Not Applicable (SEE11211495GA), Not Applicable (SEE11211506GA), Not Applicable (SEE11241551GA), Not Applicable (SEE12011897GA), Not Applicable (SEE12011900GA), Not Applicable (SEE16119845SA), Not Applicable (SEE16119846SA), 00840682109260SEE18320014SA (SEE18320014SA), 00840682109260SEE18320016SA (SEE18320016SA), 00840682109260SEE18320029SA (SEE18320029SA), 00840682109260SEE18320032SA (SEE18320032SA), 00840682109260SEE18500013SA (SEE18500013SA), 00840682109260SEE18500018SA (SEE18500018SA), 00840682109260SEE18500043SA (SEE18500043SA), Not Applicable (SEE11060977GA), Not Applicable (SEE11071073GA), Not Applicable (SEE11071081GA), Not Applicable (SEE11071085GA), Not Applicable (SEE11081159GA), Not Applicable (SEE11081160GA), Not Applicable (SEE11081171GA), Not Applicable (SEE11081176GA), Not Applicable (SEE11211481GA), Not Applicable (SEE11211491GA), Not Applicable (SEE11321570GA), Not Applicable (SEE11321575GA), Not Applicable (SEE11361716GA), Not Applicable (SEE15497524SA), Not Applicable (SEE15507661SA), Not Applicable (SEE13180776GA), Not Applicable (SEE14302280GA), Not Applicable (SEE14312288GA), Not Applicable (SEE14312290GA), Not Applicable (SEE14312291GA), Not Applicable (SEE14312292GA), Not Applicable (SEE14312296GA), Not Applicable (SEE13180769GA), Not Applicable (SEE13180783GA), Not Applicable (SEE13180787GA), Not Applicable (SEE13180788GA), Not Applicable (SEE13220823GA), Not Applicable (SEE13220824GA), Not Applicable (SEE13220825GA), Not Applicable (SEE13220830GA), Not Applicable (SEE13220835GA), 00840682109260SEE17520002SA (SEE17520002SA), 00840682109260SEE18100026SA (SEE18100026SA), 00840682109260SEE18100037SA (SEE18100037SA), 00840682109260SEE18100040SA (SEE18100040SA), 00840682109260SEE18100041SA (SEE18100041SA), 00840682109260SEE18100042SA (SEE18100042SA), 00840682109260SEE18100043SA (SEE18100043SA), 00840682109260SEE18100044SA (SEE18100044SA), Not Applicable (SEE09510829GA), Not Applicable (SEE12122127GA), Not Applicable (SEE12122129GA), Not Applicable (SEE14221992GA), Not Applicable (SEE14221996GA), Not Applicable (SEE14221997GA), Not Applicable (SEE14221999GA), Not Applicable (SEE14222001GA), Not Applicable (SEE14242083GA), Not Applicable (SEE14242090GA), Not Applicable (SEE14242091GA), Not Applicable (SEE14242095GA), Not Applicable (SEE14242099GA), Not Applicable (SEE14242102GA), Not Applicable (SEE14242106GA), Not Applicable (SEE14242112GA), Not Applicable (SEE14242115GA), Not Applicable (SEE14242116GA), Not Applicable (SEE14242120GA), Not Applicable (SEE14422527GA), Not Applicable (SEE14422532GA), Not Applicable (SEE15113137GA), 00840682109260SEE17310002SA (SEE17310002SA), 00840682109260SEE17380004SA (SEE17380004SA), 00840682109260SEE17380006SA (SEE17380006SA), 00840682109260SEE17380010SA (SEE17380010SA), Not Applicable (SEE15352669SA), Not Applicable (SEE15352786SA), Not Applicable (SEE15352787SA), Not Applicable (SEE15352788SA), Not Applicable (SEE13431297GA), Not Applicable (SEE16420047SA), Not Applicable (SEE16430014SA), Not Applicable (SEE16430031SA), Not Applicable (SEE10430604GA), Not Applicable (SEE10460744GA), Not Applicable (SEE11040899GA), Not Applicable (SEE11040900GA), Not Applicable (SEE11040901GA), Not Applicable (SEE13451315GA), Not Applicable (SEE13451336GA), Not Applicable (SEE13481388GA), Not Applicable (SEE13481406GA), Not Applicable (SEE13481408GA), Not Applicable (SEE16099455SA), Not Applicable (SEE16099456SA), Not Applicable (SEE16099459SA), Not Applicable (SEE15446223SA), Not Applicable (SEE15446225SA), Not Applicable (SEE15446241SA), Not Applicable (SEE15446242SA), Not Applicable (SEE15446243SA), Not Applicable (SEE12400096GA), Not Applicable (SEE12400098GA), Not Applicable (SEE12400102GA), Not Applicable (SEE12400108GA), Not Applicable (SEE12400111GA), Not Applicable (SEE12400113GA), Not Applicable (SEE12400115GA), Not Applicable (SEE14412499GA), Not Applicable (SEE14412505GA), Not Applicable (SEE14412512GA), Not Applicable (SEE14422514GA), Not Applicable (SEE14422522GA), Not Applicable (SEE14422523GA), Not Applicable (SEE16068806SA), 00840682109260SEE19110004SA (SEE19110004SA), 00840682109260SEE19120020SA (SEE19120020SA), 00840682109260SEE19120029SA (SEE19120029SA), Not Applicable (SEE11071148GA), Not Applicable (SEE12162175GA), Not Applicable (SEE11141256GA), Not Applicable (SEE11141261GA), Not Applicable (SEE11141262GA), Not Applicable (SEE11141267GA), Not Applicable (SEE11141275GA), Not Applicable (SEE11141291GA), Not Applicable (SEE11141292GA), Not Applicable (SEE12430213GA), Not Applicable (SEE12430217GA), Not Applicable (SEE12440239GA), Not Applicable (SEE13150670GA), Not Applicable (SEE14432578GA), Not Applicable (SEE14432599GA), Not Applicable (SEE14432600GA), Not Applicable (SEE14432601GA), Not Applicable (SEE14432607GA), Not Applicable (SEE14432608GA), Not Applicable (SEE14432609GA), Not Applicable (SEE14432614GA), Not Applicable (SEE15052942GA), Not Applicable (SEE15052947GA), Not Applicable (SEE15062971GA), Not Applicable (SEE15113173GA), Not Applicable (SEE16334181SA), 00840682109260SEE19110001SA (SEE19110001SA), Not Applicable (SEE10250144GA), Not Applicable (SEE10250148GA), Not Applicable (SEE09020034GA), Not Applicable (SEE13501467GA), Not Applicable (SEE13501474GA), 00840682109260SEE19220008SA (SEE19220008SA), 00840682109260SEE19220010SA (SEE19220010SA), 00840682109260SEE19220011SA (SEE19220011SA), 00840682109260SEE19220012SA (SEE19220012SA), 00840682109260SEE19220013SA (SEE19220013SA), 00840682109260SEE19220014SA (SEE19220014SA), 00840682109260SEE19220015SA (SEE19220015SA), 00840682109260SEE19220016SA (SEE19220016SA), 00840682109260SEE19220019SA (SEE19220019SA), 00840682109260SEE19220026SA (SEE19220026SA), Not Applicable (SEE09100106GA), Not Applicable (SEE11321586GA), Not Applicable (SEE11371805GA), Not Applicable (SEE11511886GA), Not Applicable (SEE13160711GA), Not Applicable (SEE13160712GA), Not Applicable (SEE13160713GA), Not Applicable (SEE13160715GA), Not Applicable (SEE13160716GA), Not Applicable (SEE13160717GA), Not Applicable (SEE13160720GA), Not Applicable (SEE16078920SA), Not Applicable (SEE16078926SA), Not Applicable (SEE16089175SA), Not Applicable (SEE16089176SA), Not Applicable (SEE13150684GA), Not Applicable (SEE13150690GA), 00840682109260SEE18170001SA (SEE18170001SA), 00840682109260SEE18170020SA (SEE18170020SA), Not Applicable (SEE14362381GA), Not Applicable (SEE16078927SA), Not Applicable (SEE16078929SA), Not Applicable (SEE16089172SA), Not Applicable (SEE16089180SA), Not Applicable (SEE16180925SA), Not Applicable (SEE16201177SA), Not Applicable (SEE16201179SA), Not Applicable (SEE16201180SA), Not Applicable (SEE16201182SA), Not Applicable (SEE16201367SA), Not Applicable (SEE16201368SA), Not Applicable (SEE16201408SA), Not Applicable (SEE16440006SA), 00840682109260SEE19100002SA (SEE19100002SA), Not Applicable (SEE16211409SA), Not Applicable (SEE16211510SA), Not Applicable (SEE09330408GA), Not Applicable (SEE11331595GA), Not Applicable (SEE11331599GA), Not Applicable (SEE15153190GA), Not Applicable (SEE15153191GA), Not Applicable (SEE15153192GA), Not Applicable (SEE15153193GA), Not Applicable (SEE15153195GA), Not Applicable (SEE14171929GA), Not Applicable (SEE14171930GA), Not Applicable (SEE14171931GA), Not Applicable (SEE14171932GA), Not Applicable (SEE14171934GA), Not Applicable (SEE14191935GA), Not Applicable (SEE14191936GA), Not Applicable (SEE14191937GA), Not Applicable (SEE14191938GA), Not Applicable (SEE14191939GA), Not Applicable (SEE14191944GA), Not Applicable (SEE14191946GA), Not Applicable (SEE15352793SA), Not Applicable (SEE16150568SA), Not Applicable (SEE16160726SA), Not Applicable (SEE16160727SA), Not Applicable (SEE11521892GA), 00840682109260SEE17340007SA (SEE17340007SA), 00840682109260SEE17340011SA (SEE17340011SA), 00840682109260SEE17340015SA (SEE17340015SA), 00840682109260SEE17340019SA (SEE17340019SA), Not Applicable (SEE09330410GA), Not Applicable (SEE12252307GA), Not Applicable (SEE09530881GA), Not Applicable (SEE09530887GA), Not Applicable (SEE10250142GA), Not Applicable (SEE14021504GA), Not Applicable (SEE16380025SA), Not Applicable (SEE16380026SA), Not Applicable (SEE16380027SA), Not Applicable (SEE16380028SA), Not Applicable (SEE16380029SA), Not Applicable (SEE16380030SA), Not Applicable (SEE16380032SA), Not Applicable (SEE16380033SA), 00840682109260SEE17120001SA (SEE17120001SA), 00840682109260SEE17120002SA (SEE17120002SA), 00840682109260SEE18190003SA (SEE18190003SA), 00840682109260SEE18190004SA (SEE18190004SA), 00840682109260SEE18190009SA (SEE18190009SA), 00840682109260SEE18190010SA (SEE18190010SA), 00840682109260SEE18190012SA (SEE18190012SA), 00840682109260SEE18190013SA (SEE18190013SA), 00840682109260SEE18190016SA (SEE18190016SA), 00840682109260SEE18190020SA (SEE18190020SA), Not Applicable (SEE10450656GA), Not Applicable (SEE10460694GA), Not Applicable (SEE10250105GA), Not Applicable (SEE12510311GA), Not Applicable (SEE12510317GA), Not Applicable (SEE13050434GA), Not Applicable (SEE11371753GA), Not Applicable (SEE12302388GA), Not Applicable (SEE12322415GA), Not Applicable (SEE12322432GA), Not Applicable (SEE12332445GA), Not Applicable (SEE14151839GA), Not Applicable (SEE14171923GA), Not Applicable (SEE14171924GA), Not Applicable (SEE16109589SA), Not Applicable (SEE16109593SA), Not Applicable (SEE16119897SA), Not Applicable (SEE16119899SA), Not Applicable (SEE16119904SA), Not Applicable (SEE16119905SA), Not Applicable (SEE16119906SA), Not Applicable (SEE16119907SA), Not Applicable (SEE16119909SA), Not Applicable (SEE16119912SA), Not Applicable (SEE16119913SA), Not Applicable (SEE16119915SA), Not Applicable (SEE16119916SA), Not Applicable (SEE16119917SA), Not Applicable (SEE16119918SA), Not Applicable (SEE16119920SA), Not Applicable (SEE16119922SA), Not Applicable (SEE16119924SA), Not Applicable (SEE16119926SA), Not Applicable (SEE16119927SA), Not Applicable (SEE16119928SA), Not Applicable (SEE16430036SA), Not Applicable (SEE16430061SA), Not Applicable (SEE16430064SA), Not Applicable (SEE16430065SA), 00840682109260SEE17090012SA (SEE17090012SA), 00840682109260SEE17190021SA (SEE17190021SA), 00840682109260SEE19160047SA (SEE19160047SA), 00840682109260SEE19160051SA (SEE19160051SA), Not Applicable (SEE15477182SA), 00840682109260SEE18060001SA (SEE18060001SA), 00840682109260SEE18060007SA (SEE18060007SA), Not Applicable (SEE09140186GA), Not Applicable (SEE15012842GA), Not Applicable (SEE15012852GA), Not Applicable (SEE15022864GA), Not Applicable (SEE14492746GA), Not Applicable (SEE15012851GA), Not Applicable (SEE15456722SA), Not Applicable (SEE15466919SA), Not Applicable (SEE15466926SA), Not Applicable (SEE15466928SA), Not Applicable (SEE15466932SA), Not Applicable (SEE13220851GA), Not Applicable (SEE13280905GA), 00840682109260SEE17250002SA (SEE17250002SA), 00840682109260SEE17250006SA (SEE17250006SA), 00840682109260SEE17250007SA (SEE17250007SA), 00840682109260SEE17250008SA (SEE17250008SA), 00840682109260SEE17250009SA (SEE17250009SA), Not Applicable (SEE10420583GA), Not Applicable (SEE10420588GA), Not Applicable (SEE10460697GA), Not Applicable (SEE10460698GA), Not Applicable (SEE10460714GA), Not Applicable (SEE11071090GA), Not Applicable (SEE11181400GA), Not Applicab le (SEE11181403GA), Not Applicable (SEE11201445GA), Not Applicable (SEE11211478GA), Not Applicable (SEE14242098GA), Not Applicable (SEE14422534GA), Not Applicable (SEE14422536GA), Not Applicable (SEE14422537GA), Not Applicable (SEE14422538GA), Not Applicable (SEE14422540GA), Not Applicable (SEE14422545GA), Not Applicable (SEE14422546GA), Not Applicable (SEE14422548GA), Not Applicable (SEE14422549GA), Not Applicable (SEE14422550GA), Not Applicable (SEE14422553GA), Not Applicable (SEE14422555GA), Not Applicable (SEE14422556GA), Not Applicable (SEE14422558GA), Not Applicable (SEE14422560GA), Not Applicable (SEE14422561GA), Not Applicable (SEE14422564GA), Not Applicable (SEE14422566GA), Not Applicable (SEE14422567GA), Not Applicable (SEE14422568GA), Not Applicable (SEE14422570GA), Not Applicable (SEE14422571GA), Not Applicable (SEE14422572GA), Not Applicable (SEE14422573GA), Not Applicable (SEE14462645GA), Not Applicable (SEE15241006SA), Not Applicable (SEE15251241SA), Not Applicable (SEE16191063SA), 00840682109260SEE17170004SA (SEE17170004SA), 00840682109260SEE19090001SA (SEE19090001SA), 00840682109260SEE19090003SA (SEE19090003SA), 00840682109260SEE19090018SA (SEE19090018SA), Not Applicable (SEE13150667GA), Not Applicable (SEE15193297GA), Not Applicable (SEE15193302GA), Not Applicable (SEE16038555SA), Not Applicable (SEE16058656SA), Not Applicable (SEE16058657SA), Not Applicable (SEE16058658SA), Not Applicable (SEE16058659SA), Not Applicable (SEE16058665SA), Not Applicable (SEE16058666SA), Not Applicable (SEE16058667SA), Not Applicable (SEE16058668SA), Not Applicable (SEE16058670SA), Not Applicable (SEE16058671SA), Not Applicable (SEE16058672SA), Not Applicable (SEE16058673SA), Not Applicable (SEE16058674SA), Not Applicable (SEE16058675SA), Not Applicable (SEE16058676SA), Not Applicable (SEE16058677SA), Not Applicable (SEE16058678SA), Not Applicable (SEE16058679SA), Not Applicable (SEE16058680SA), Not Applicable (SEE16058699SA), Not Applicable (SEE16130306SA), Not Applicable (SEE16130320SA), Not Applicable (SEE16130322SA), 00840682109260SEE17090017SA (SEE17090017SA), Not Applicable (SEE12021920GA), Not Applicable (SEE12021921GA), Not Applicable (SEE12051988GA), Not Applicable (SEE14382469GA), Not Applicable (SEE14402489GA), Not Applicable (SEE14402491GA), Not Applicable (SEE14412495GA), Not Applicable (SEE14412496GA), Not Applicable (SEE14412497GA), Not Applicable (SEE14412502GA), Not Applicable (SEE14412509GA), Not Applicable (SEE14412510GA), Not Applicable (SEE14412513GA), Not Applicable (SEE16252013SA), Not Applicable (SEE16252019SA), Not Applicable (SEE16252025SA), 00840682109260SEE17110005SA (SEE17110005SA), 00840682109260SEE18100021SA (SEE18100021SA), 00840682109260SEE18100022SA (SEE18100022SA), 00840682109260SEE18100023SA (SEE18100023SA), 00840682109260SEE18100024SA (SEE18100024SA), 00840682109260SEE18100033SA (SEE18100033SA), Not Applicable (SEE12490247GA), Not Applicable (SEE12490257GA), Not Applicable (SEE13100477GA), Not Applicable (SEE13100484GA), Not Applicable (SEE13100489GA), Not Applicable (SEE13100490GA), Not Applicable (SEE13110495GA), Not Applicable (SEE13110496GA), Not Applicable (SEE13110503GA), 00840682109260SEE17070006SA (SEE17070006SA), 00840682109260SEE17070012SA (SEE17070012SA), 00840682109260SEE17070013SA (SEE17070013SA), 00840682109260SEE17070014SA (SEE17070014SA), 00840682109260SEE17070015SA (SEE17070015SA), 00840682109260SEE17070016SA (SEE17070016SA), 00840682109260SEE17070017SA (SEE17070017SA), 00840682109260SEE17070018SA (SEE17070018SA), 00840682109260SEE17070021SA (SEE17070021SA), 00840682109260SEE17070022SA (SEE17070022SA), 00840682109260SEE17070023SA (SEE17070023SA), 00840682109260SEE17070024SA (SEE17070024SA), 00840682109260SEE17070025SA (SEE17070025SA), 00840682109260SEE17070026SA (SEE17070026SA), 00840682109260SEE17350001SA (SEE17350001SA), 00840682109260SEE17350002SA (SEE17350002SA), 00840682109260SEE17350003SA (SEE17350003SA), Not Applicable (SEE12340012GA), Not Applicable (SEE12340016GA), Not Applicable (SEE12340019GA), Not Applicable (SEE12350021GA), Not Applicable (SEE12350024GA), Not Applicable (SEE12360033GA), Not Applicable (SEE12360035GA), Not Applicable (SEE12360036GA), Not Applicable (SEE14432598GA), Not Applicable (SEE14432602GA), Not Applicable (SEE15052917GA), Not Applicable (SEE15052923GA), Not Applicable (SEE15052926GA), Not Applicable (SEE15052932GA), 00840682109260SEE18170005SA (SEE18170005SA), 00840682109260SEE18170025SA (SEE18170025SA), Not Applicable (SEE10390290GA), Not Applicable (SEE10400451GA), Not Applicable (SEE13140657GA), Not Applicable (SEE13150661GA), Not Applicable (SEE13160714GA), Not Applicable (SEE13160718GA), Not Applicable (SEE13160719GA), Not Applicable (SEE13180742GA), Not Applicable (SEE13361074GA), Not Applicable (SEE13371086GA), Not Applicable (SEE13371100GA), Not Applicable (SEE15113139GA), 00840682109260SEE18370001SA (SEE18370001SA), Not Applicable (SEE14161890GA), Not Applicable (SEE14161891GA), Not Applicable (SEE14161892GA), Not Applicable (SEE14161895GA), Not Applicable (SEE14161900GA), Not Applicable (SEE14161901GA), Not Applicable (SEE14161912GA), Not Applicable (SEE10470804GA), Not Applicable (SEE10470805GA), Not Applicable (SEE10470806GA), Not Applicable (SEE10530817GA), Not Applicable (SEE10530825GA), Not Applicable (SEE09020017GA), Not Applicable (SEE09040052GA), Not Applicable (SEE09040053GA), Not Applicable (SEE09040076GA), Not Applicable (SEE15261393SA), Not Applicable (SEE15261394SA), Not Applicable (SEE15261395SA), Not Applicable (SEE15261396SA), Not Applicable (SEE15261397SA), Not Applicable (SEE15261398SA), Not Applicable (SEE15261399SA), Not Applicable (SEE15261401SA), Not Applicable (SEE15261402SA), Not Applicable (SEE15261404SA), Not Applicable (SEE15261405SA), Not Applicable (SEE15261406SA), Not Applicable (SEE15261407SA), Not Applicable (SEE15261511SA), Not Applicable (SEE16364410SA), Not Applicable (SEE16364416SA), 00840682109260SEE18380022SA (SEE18380022SA), Not Applicable (SEE10120001GA), Not Applicable (SEE10120010GA), Not Applicable (SEE10121019GA), Not Applicable (SEE13371120GA), Not Applicable (SEE12490249GA), Not Applicable (SEE12490252GA), Not Applicable (SEE12490258GA), Not Applicable (SEE09430602GA), Not Applicable (SEE15052933GA), Not Applicable (SEE15052935GA), Not Applicable (SEE15052936GA), Not Applicable (SEE15052941GA), Not Applicable (SEE15053021GA), Not Applicable (SEE15062993GA), Not Applicable (SEE15063003GA), Not Applicable (SEE15063005GA), Not Applicable (SEE15063008GA), Not Applicable (SEE15063010GA), Not Applicable (SEE15063015GA), Not Applicable (SEE15063017GA), Not Applicable (SEE15063019GA), Not Applicable (SEE15063020GA), Not Applicable (SEE15063021GA), 00840682109260SEE19220018SA (SEE19220018SA), Not Applicable (SEE11050920GA), Not Applicable (SEE11050932GA), Not Applicable (SEE11050938GA), Not Applicable (SEE11050943GA), Not Applicable (SEE11071082GA), 00840682109260SEE18290017SA (SEE18290017SA), Not Applicable (SEE09100116GA), Not Applicable (SEE12092076GA), Not Applicable (SEE12092077GA), Not Applicable (SEE12092078GA), Not Applicable (SEE12092079GA), Not Applicable (SEE12092080GA), Not Applicable (SEE12102086GA), Not Applicable (SEE12102087GA), Not Applicable (SEE12102088GA), Not Applicable (SEE12102090GA), Not Applicable (SEE12102093GA), Not Applicable (SEE12102095GA), Not Applicable (SEE12102097GA), Not Applicable (SEE12102098GA), Not Applicable (SEE13120606GA), Not Applicable (SEE13130615GA), Not Applicable (SEE13130626GA), Not Applicable (SEE13130628GA), Not Applicable (SEE09100091GA), Not Applicable (SEE09100096GA), Not Applicable (SEE09100114GA), Not Applicable (SEE09100118GA), Not Applicable (SEE09100119GA), Not Applicable (SEE09100120GA), Not Applicable (SEE09100128GA), Not Applicable (SEE09470657GA), Not Applicable (SEE14161884GA), Not Applicable (SEE14161885GA), Not Applicable (SEE14161887GA), Not Applicable (SEE14161888GA), Not Applicable (SEE14161889GA), Not Applicable (SEE14161896GA), Not Applicable (SEE14161899GA), Not Applicable (SEE14161907GA), Not Applicable (SEE14161909GA), Not Applicable (SEE14161913GA), Not Applicable (SEE14161914GA), Not Applicable (SEE15477110SA), Not Applicable (SEE15477112SA), Not Applicable (SEE15415185SA), Not Applicable (SEE15415188SA), Not Applicable (SEE15415195SA), Not Applicable (SEE15446335SA), Not Applicable (SEE15456723SA), Not Applicable (SEE10430629GA), Not Applicable (SEE10450666GA), Not Applicable (SEE10450672GA), Not Applicable (SEE10450681GA), Not Applicable (SEE10450689GA), Not Applicable (SEE10530824GA), Not Applicable (SEE10530836GA), Not Applicable (SEE10530837GA), Not Applicable (SEE10530838GA), Not Applicable (SEE10530839GA), Not Applicable (SEE10530843GA), Not Applicable (SEE10530844GA), Not Applicable (SEE10530845GA), Not Applicable (SEE10530846GA), Not Applicable (SEE10530848GA), Not Applicable (SEE11050933GA), Not Applicable (SEE11050934GA), Not Applicable (SEE15173207GA), Not Applicable (SEE15446123SA), Not Applicable (SEE09100147GA), Not Applicable (SEE10120032GA), Not Applicable (SEE10430603GA), Not Applicable (SEE10450671GA), Not Applicable (SEE10460729GA), Not Applicable (SEE10460753GA), Not Applicable (SEE10510813GA), Not Applicable (SEE10510814GA), Not Applicable (SEE11050941GA), Not Applicable (SEE11071036GA), Not Applicable (SEE11081186GA), Not Applicable (SEE11161320GA), Not Applicable (SEE11191432GA), Not Applicable (SEE13280923GA), Not Applicable (SEE13280927GA), Not Applicable (SEE13280932GA), Not Applicable (SEE13280941GA), Not Applicable (SEE13280946GA), Not Applicable (SEE14272161GA), Not Applicable (SEE14272200GA), Not Applicable (SEE14272203GA), Not Applicable (SEE14272205GA), Not Applicable (SEE14272207GA), Not Applicable (SEE14272219GA), Not Applicable (SEE14472675GA), Not Applicable (SEE16360016SA), Not Applicable (SEE10410499GA), Not Applicable (SEE10420540GA), Not Applicable (SEE11141288GA), Not Applicable (SEE12340009GA), Not Applicable (SEE12340011GA), Not Applicable (SEE12342451GA), Not Applicable (SEE12350029GA), Not Applicable (SEE12390085GA), Not Applicable (SEE12390087GA), Not Applicable (SEE12390088GA), Not Applicable (SEE12390089GA), Not Applicable (SEE12390090GA), Not Applicable (SEE13050439GA), Not Applicable (SEE13060445GA), Not Applicable (SEE13190798GA), Not Applicable (SEE13290951GA), Not Applicable (SEE13290953GA), Not Applicable (SEE13290955GA), Not Applicable (SEE13290958GA), Not Applicable (SEE15353099SA), Not Applicable (SEE15353100SA), Not Applicable (SEE15353119SA), Not Applicable (SEE16150524SA), Not Applicable (SEE16221590SA), Not Applicable (SEE11071037GA), Not Applicable (SEE11071043GA), Not Applicable (SEE13020407GA), Not Applicable (SEE14362385GA), Not Applicable (SEE10260222GA), Not Applicable (SEE10270251GA), Not Applicable (SEE13040419GA), 00840682109260SEE17060005SA (SEE17060005SA), Not Applicable (SEE12172188GA), Not Applicable (SEE16390001SA), Not Applicable (SEE16390005SA), Not Applicable (SEE09340450GA), Not Applicable (SEE09360491GA), Not Applicable (SEE09390537GA), Not Applicable (SEE12430185GA), Not Applicable (SEE12430187GA), Not Applicable (SEE12430203GA), Not Applicable (SEE12490241GA), Not Applicable (SEE12490242GA), Not Applicable (SEE12490253GA), Not Applicable (SEE12490255GA), Not Applicable (SEE12490284GA), Not Applicable (SEE12490286GA), Not Applicable (SEE12490291GA), Not Applicable (SEE12490292GA), Not Applicable (SEE12490297GA), Not Applicable (SEE13371095GA), Not Applicable (SEE13371097GA), Not Applicable (SEE13371098GA), Not Applicable (SEE13371104GA), Not Applicable (SEE13371110GA), Not Applicable (SEE13371111GA), Not Applicable (SEE13371117GA), Not Applicable (SEE13371118GA), 00840682109260SEE17180003SA (SEE17180003SA), Not Applicable (SEE12510301GA), Not Applicable (SEE13000387GA), Not Applicable (SEE13220849GA), Not Applicable (SEE13270880GA), Not Applicable (SEE13270884GA), Not Applicable (SEE13270891GA), Not Applicable (SEE13270895GA), Not Applicable (SEE13270900GA), Not Applicable (SEE13270902GA), Not Applicable (SEE13280906GA), Not Applicable (SEE09230255GA), Not Applicable (SEE09230259GA), Not Applicable (SEE09370499GA), Not Applicable (SEE09530854GA), Not Applicable (SEE10390289GA), 00840682109260SEE17120003SA (SEE17120003SA), Not Applicable (SEE14111800GA), Not Applicable (SEE10250160GA), Not Applicable (SEE10270242GA), Not Applicable (SEE10430628GA), Not Applicable (SEE10430638GA), Not Applicable (SEE10460765GA), Not Applicable (SEE10460771GA), Not Applicable (SEE10460774GA), Not Applicable (SEE10470777GA), Not Applicable (SEE14031587GA), 00840682109260SEE18510002SA (SEE18510002SA), 00840682109260SEE19130007SA (SEE19130007SA), 00840682109260SEE19150001SA (SEE19150001SA), Not Applicable (SEE10250140GA), Not Applicable (SEE10250167GA), Not Applicable (SEE10270248GA), Not Applicable (SEE10270259GA), Not Applicable (SEE10270261GA), Not Applicable (SEE09100104GA), Not Applicable (SEE10260236GA), Not Applicable (SEE10260238GA), Not Applicable (SEE10430614GA), Not Applicable (SEE10430630GA), Not Applicable (SEE10430645GA), Not Applicable (SEE10430653GA), Not Applicable (SEE11361703GA), Not Applicable (SEE12092046GA), Not Applicable (SEE12172198GA), Not Applicable (SEE12172200GA), Not Applicable (SEE12182226GA), Not Applicable (SEE12182227GA), Not Applicable (SEE12182230GA), Not Applicable (SEE12182232GA), Not Applicable (SEE12430190GA), Not Applicable (SEE13150660GA), Not Applicable (SEE13220832GA), Not Applicable (SEE13220839GA), Not Applicable (SEE15062974GA), Not Applicable (SEE15062976GA), Not Applicable (SEE15062987GA), Not Applicable (SEE15062990GA), Not Applicable (SEE15062994GA), Not Applicable (SEE15062995GA), Not Applicable (SEE15062996GA), Not Applicable (SEE15062998GA), Not Applicable (SEE15063011GA), Not Applicable (SEE15113132GA), Not Applicable (SEE15281638SA), Not Applicable (SEE15281639SA), Not Applicable (SEE15281640SA), Not Applicable (SEE15312099SA), Not Applicable (SEE15446220SA), Not Applicable (SEE15446221SA), Not Applicable (SEE15446222SA), Not Applicable (SEE15446224SA), Not Applicable (SEE15446230SA), Not Applicable (SEE15446235SA), Not Applicable (SEE15446236SA), Not Applicable (SEE15446238SA), Not Applicable (SEE15446239SA), Not Applicable (SEE16119850SA), Not Applicable (SEE16201406SA), Not Applicable (SEE16201407SA), 00840682109260SEE17460013SA (SEE17460013SA), 00840682109260SEE17460016SA (SEE17460016SA), 00840682109260SEE17460021SA (SEE17460021SA), 00840682109260SEE17460022SA (SEE17460022SA), Not Applicable (SEE13270886GA), Not Applicable (SEE13270887GA), Not Applicable (SEE13270892GA), Not Applicable (SEE13270893GA), Not Applicable (SEE16241792SA), Not Applicable (SEE16241796SA), Not Applicable (SEE09330409GA), Not Applicable (SEE09330411GA), Not Applicable (SEE09230273GA), Not Applicable (SEE09230284GA), Not Applicable (SEE09230286GA), Not Applicable (SEE09240295GA), Not Applicable (SEE09250317GA), Not Applicable (SEE09380515GA), Not Applicable (SEE09380517GA), Not Applicable (SEE09380522GA), Not Applicable (SEE09380525GA), Not Applicable (SEE10420558GA), Not Applicable (SEE10420564GA), Not Applicable (SEE10420577GA), Not Applicable (SEE10420585GA), Not Applicable (SEE10420595GA), Not Applicable (SEE10430620GA), Not Applicable (SEE10450660GA), Not Applicable (SEE10460692GA), Not Applicable (SEE11201447GA), Not Applicable (SEE11201455GA), Not Applicable (SEE11201456GA), Not Applicable (SEE11201464GA), Not Applicable (SEE15093059GA), Not Applicable (SEE15103080GA), Not Applicable (SEE15103086GA), Not Applicable (SEE15103087GA), Not Applicable (SEE15103093GA), Not Applicable (SEE15103095GA), 00840682109260SEE17460007SA (SEE17460007SA), 00840682109260SEE18420007SA (SEE18420007SA), 00840682109260SEE18420013SA (SEE18420013SA), 00840682109260SEE18420015SA (SEE18420015SA), 00840682109260SEE18420016SA (SEE18420016SA), 00840682109260SEE18420020SA (SEE18420020SA), 00840682109260SEE18420021SA (SEE18420021SA), Not Applicable (SEE09190240GA), Not Applicable (SEE10390299GA), Not Applicable (SEE13431285GA), Not Applicable (SEE10400422GA), Not Applicable (SEE10410468GA), Not Applicable (SEE10410481GA), Not Applicable (SEE10410513GA), Not Applicable (SEE10420533GA), Not Applicable (SEE10420553GA), Not Applicable (SEE14312307GA), Not Applicable (SEE15251188SA), Not Applicable (SEE16028317SA), Not Applicable (SEE13120595GA), Not Applicable (SEE13120597GA), Not Applicable (SEE13120600GA), Not Applicable (SEE13120608GA), Not Applicable (SEE13130617GA), Not Applicable (SEE13130619GA), Not Applicable (SEE13130624GA), Not Applicable (SEE13130627GA), Not Applicable (SEE13130629GA), Not Applicable (SEE13130632GA), Not Applicable (SEE13130633GA), Not Applicable (SEE13130635GA), Not Applicable (SEE13130645GA), Not Applicable (SEE13130648GA), Not Applicable (SEE15113119GA), Not Applicable (SEE15113122GA), 00840682109260SEE17490015SA (SEE17490015SA), 00840682109260SEE17490016SA (SEE17490016SA), 00840682109260SEE17490037SA (SEE17490037SA), 00840682109260SEE18470018SA (SEE18470018SA), 00840682109260SEE18470019SA (SEE18470019SA), 00840682109260SEE18470024SA (SEE18470024SA), 00840682109260SEE18500035SA (SEE18500035SA), 00840682109260SEE18500037SA (SEE18500037SA), Not Applicable (SEE13240854GA), Not Applicable (SEE13240859GA), Not Applicable (SEE13240861GA), Not Applicable (SEE13240864GA), Not Applicable (SEE13240865GA), Not Applicable (SEE13240870GA), Not Applicable (SEE12102100GA), Not Applicable (SEE09100092GA), Not Applicable (SEE09100100GA), Not Applicable (SEE12112119GA), Not Applicable (SEE12322416GA), Not Applicable (SEE12322419GA), Not Applicable (SEE15261547SA), Not Applicable (SEE15261548SA), Not Applicable (SEE15261549SA), Not Applicable (SEE15261550SA), Not Applicable (SEE15261551SA), Not Applicable (SEE15261552SA), Not Applicable (SEE15261553SA), Not Applicable (SEE15261555SA), Not Applicable (SEE15261556SA), Not Applicable (SEE15261557SA), Not Applicable (SEE15261558SA), Not Applicable (SEE16360021SA), Not Applicable (SEE16370002SA), Not Applicable (SEE16370004SA), Not Applicable (SEE10100983GA), Not Applicable (SEE10120008GA), Not Applicable (SEE10120991GA), Not Applicable (SEE10120992GA), Not Applicable (SEE10390319GA), Not Applicable (SEE10390328GA), Not Applicable (SEE13010406GA), Not Applicable (SEE15352667SA), 00840682109260SEE17070019SA (SEE17070019SA), 00840682109260SEE17070028SA (SEE17070028SA), 00840682109260SEE18450010SA (SEE18450010SA), 00840682109260SEE18450011SA (SEE18450011SA), 00840682109260SEE18450013SA (SEE18450013SA), Not Applicable (SEE11040917GA), Not Applicable (SEE11050921GA), Not Applicable (SEE11050923GA), Not Applicable (SEE11050924GA), Not Applicable (SEE11050925GA), Not Applicable (SEE11050926GA), Not Applicable (SEE11050928GA), Not Applicable (SEE16221555SA), Not Applicable (SEE12490245GA), Not Applicable (SEE16078863SA), Not Applicable (SEE12490248GA), Not Applicable (SEE12490251GA), Not Applicable (SEE12490300GA), Not Applicable (SEE13371088GA), Not Applicable (SEE13371094GA), Not Applicable (SEE13371099GA), Not Applicable (SEE13371102GA), Not Applicable (SEE13371106GA), Not Applicable (SEE13371107GA), Not Applicable (SEE13371114GA), Not Applicable (SEE13371116GA), Not Applicable (SEE15173227GA), 00840682109260SEE18240052SA (SEE18240052SA), 00840682109260SEE19120016SA (SEE19120016SA), 00840682109260SEE19120026SA (SEE19120026SA), 00840682109260SEE19120027SA (SEE19120027SA), 00840682109260SEE19170034SA (SEE19170034SA), Not Applicable (SEE13110530GA), Not Applicable (SEE13110532GA), Not Applicable (SEE13110534GA), Not Applicable (SEE13110535GA), Not Applicable (SEE13110541GA), Not Applicable (SEE13110543GA), Not Applicable (SEE13110545GA), Not Applicable (SEE16370003SA), Not Applicable (SEE16370005SA), 00840682109260SEE19180005SA (SEE19180005SA), 00840682109260SEE19190002SA (SEE19190002SA), Not Applicable (SEE11071080GA), Not Applicable (SEE11071135GA), Not Applicable (SEE11081175GA), Not Applicable (SEE11081196GA), Not Applicable (SEE12520357GA), 00840682109260SEE18040018SA (SEE18040018SA), 00840682109260SEE18040041SA (SEE18040041SA), 00840682109260SEE18040043SA (SEE18040043SA), 00840682109260SEE18040048SA (SEE18040048SA), Not Applicable (SEE14512822GA), Not Applicable (SEE14512826GA), Not Applicable (SEE14512829GA), Not Applicable (SEE15291766SA), Not Applicable (SEE16360012SA), Not Applicable (SEE16360013SA), Not Applicable (SEE11111244GA), Not Applicable (SEE11211501GA), 00840682109260SEE18320020SA (SEE18320020SA), Not Applicable (SEE09340449GA), Not Applicable (SEE09340453GA), Not Applicable (SEE09350462GA), Not Applicable (SEE14031625GA), Not Applicable (SEE14031626GA), Not Applicable (SEE14031629GA), Not Applicable (SEE14031637GA), Not Applicable (SEE14031648GA), Not Applicable (SEE14031651GA), Not Applicable (SEE14031652GA), Not Applicable (SEE14031653GA), Not Applicable (SEE14031654GA), Not Applicable (SEE14031656GA), Not Applicable (SEE09100133GA), Not Applicable (SEE10390314GA), Not Applicable (SEE10390340GA), Not Applicable (SEE10400426GA), Not Applicable (SEE10410486GA), Not Applicable (SEE12242279GA), Not Applicable (SEE12322408GA), Not Applicable (SEE13110509GA), Not Applicable (SEE13110512GA), Not Applicable (SEE13110517GA), Not Applicable (SEE13110525GA), Not Applicable (SEE13110526GA), Not Applicable (SEE13381123GA), Not Applicable (SEE13391137GA), Not Applicable (SEE13391138GA), Not Applicable (SEE13391139GA), Not Applicable (SEE13391140GA), Not Applicable (SEE15093075GA), Not Applicable (SEE15103076GA), Not Applicable (SEE15103078GA), Not Applicable (SEE15103079GA), Not Applicable (SEE15103081GA), Not Applicable (SEE15103082GA), Not Applicable (SEE15103083GA), Not Applicable (SEE15103088GA), Not Applicable (SEE15103089GA), Not Applicable (SEE15103090GA), Not Applicable (SEE15103091GA), Not Applicable (SEE15103092GA), Not Applicable (SEE15456710SA), 00840682109260SEE18100020SA (SEE18100020SA), 00840682109260SEE18160051SA (SEE18160051SA), 00840682109260SEE18160052SA (SEE18160052SA), 00840682109260SEE18160054SA (SEE18160054SA), 00840682109260SEE18160057SA (SEE18160057SA), 00840682109260SEE18160059SA (SEE18160059SA), 00840682109260SEE18510012SA (SEE18510012SA), Not Applicable (SEE11361745GA), Not Applicable (SEE11371764GA), Not Applicable (SEE16109588SA), Not Applicable (SEE16109592SA), Not Applicable (SEE16109595SA), Not Applicable (SEE16109699SA), Not Applicable (SEE16109704SA), 00840682109260SEE17080010SA (SEE17080010SA), Not Applicable (SEE12162171GA), Not Applicable (SEE14502781GA), Not Applicable (SEE14502782GA), Not Applicable (SEE14502783GA), Not Applicable (SEE14502784GA), Not Applicable (SEE11361727GA), Not Applicable (SEE11521889GA), Not Applicable (SEE11521891GA), Not Applicable (SEE12031931GA), Not Applicable (SEE12031933GA), Not Applicable (SEE12031937GA), Not Applicable (SEE10400370GA), Not Applicable (SEE10400373GA), Not Applicable (SEE10400435GA), Not Applicable (SEE10400441GA), Not Applicable (SEE14502788GA), Not Applicable (SEE14502792GA), Not Applicable (SEE14502794GA), Not Applicable (SEE16354345SA), Not Applicable (SEE12410121GA), Not Applicable (SEE12410123GA), Not Applicable (SEE12410124GA), Not Applicable (SEE12410125GA), Not Applicable (SEE12410127GA), Not Applicable (SEE12410128GA), Not Applicable (SEE12410129GA), Not Applicable (SEE12410145GA), Not Applicable (SEE12430202GA), Not Applicable (SEE12430208GA), Not Applicable (SEE12430210GA), Not Applicable (SEE13411189GA), Not Applicable (SEE13411190GA), Not Applicable (SEE13411191GA), Not Applicable (SEE13411192GA), Not Applicable (SEE13411193GA), Not Applicable (SEE15063004GA), 00840682109260SEE18080007SA (SEE18080007SA), Not Applicable (SEE09140191GA), Not Applicable (SEE09230262GA), Not Applicable (SEE10120017GA), Not Applicable (SEE11071039GA), Not Applicable (SEE11191436GA), Not Applicable (SEE11211480GA), Not Applicable (SEE11241547GA), Not Applicable (SEE13501458GA), Not Applicable (SEE13501465GA), Not Applicable (SEE13501469GA), Not Applicable (SEE14031644GA), Not Applicable (SEE14031671GA), Not Applicable (SEE14211978GA), Not Applicable (SEE14222026GA), Not Applicable (SEE14222032GA), Not Applicable (SEE14222034GA), Not Applicable (SEE14322312GA), Not Applicable (SEE14372420GA), Not Applicable (SEE14372432GA), Not Applicable (SEE14372437GA), Not Applicable (SEE14372441GA), Not Applicable (SEE14372442GA), Not Applicable (SEE14382456GA), Not Applicable (SEE14382457GA), Not Applicable (SEE14382465GA), Not Applicable (SEE14382467GA), Not Applicable (SEE14382471GA), Not Applicable (SEE15062982GA), Not Applicable (SEE15062985GA), Not Applicable (SEE15093024GA), Not Applicable (SEE15093025GA), Not Applicable (SEE15093026GA), Not Applicable (SEE15093027GA), Not Applicable (SEE15093028GA), Not Applicable (SEE15093030GA), Not Applicable (SEE15093031GA), Not Applicable (SEE15093033GA), Not Applicable (SEE15456728SA), Not Applicable (SEE15456730SA), Not Applicable (SEE15497530SA), Not Applicable (SEE15497579SA), Not Applicable (SEE15497580SA), Not Applicable (SEE15497581SA), Not Applicable (SEE15497583SA), Not Applicable (SEE15497584SA), Not Applicable (SEE15497585SA), Not Applicable (SEE15497586SA), Not Applicable (SEE15497587SA), Not Applicable (SEE15497588SA), Not Applicable (SEE10470787GA), Not Applicable (SEE12082030GA), Not Applicable (SEE12082035GA), Not Applicable (SEE12092048GA), Not Applicable (SEE16058669SA), Not Applicable (SEE11181388GA), Not Applicable (SEE11181390GA), Not Applicable (SEE11181398GA), Not Applicable (SEE11201444GA), Not Applicable (SEE11201453GA), Not Applicable (SEE11201463GA), Not Applicable (SEE11231541GA), Not Applicable (SEE11231542GA), Not Applicable (SEE15261392SA), Not Applicable (SEE15261408SA), Not Applicable (SEE16380008SA), 00840682109260SEE17490004SA (SEE17490004SA), 00840682109260SEE18500014SA (SEE18500014SA), Not Applicable (SEE09430609GA), Not Applicable (SEE09430611GA), Not Applicable (SEE09430613GA), Not Applicable (SEE09430616GA), Not Applicable (SEE09430618GA), Not Applicable (SEE09430621GA), Not Applicable (SEE11191433GA), Not Applicable (SEE11211494GA), Not Applicable (SEE10410495GA), Not Applicable (SEE12430200GA), Not Applicable (SEE15332311SA), Not Applicable (SEE15332312SA), Not Applicable (SEE15332313SA), Not Applicable (SEE15332314SA), Not Applicable (SEE15332315SA), Not Applicable (SEE15332339SA), Not Applicable (SEE15332341SA), Not Applicable (SEE11321574GA), Not Applicable (SEE12302383GA), Not Applicable (SEE12312398GA), Not Applicable (SEE12312401GA), Not Applicable (SEE12312404GA), Not Applicable (SEE15415094SA), Not Applicable (SEE15415193SA), Not Applicable (SEE15415194SA), Not Applicable (SEE12340007GA), Not Applicable (SEE12342449GA), Not Applicable (SEE12342452GA), Not Applicable (SEE16364405SA), Not Applicable (SEE09410573GA), Not Applicable (SEE09470670GA), Not Applicable (SEE09480740GA), Not Applicable (SEE10400401GA), 00840682109260SEE17150006SA (SEE17150006SA), 00840682109260SEE18240015SA (SEE18240015SA), 00840682109260SEE18240036SA (SEE18240036SA), 00840682109260SEE18240037SA (SEE18240037SA), 00840682109260SEE18240044SA (SEE18240044SA), 00840682109260SEE18240045SA (SEE18240045SA), 00840682109260SEE18240047SA (SEE18240047SA), 00840682109260SEE18240048SA (SEE18240048SA), Not Applicable (SEE11171382GA), Not Applicable (SEE16109578SA), Not Applicable (SEE16109591SA), Not Applicable (SEE16119767SA), Not Applicable (SEE16119923SA), Not Applicable (SEE16120076SA), Not Applicable (SEE16120077SA), Not Applicable (SEE16130304SA), Not Applicable (SEE16150547SA), Not Applicable (SEE16160735SA), Not Applicable (SEE16160737SA), 00840682109260SEE18170002SA (SEE18170002SA), 00840682109260SEE18170004SA (SEE18170004SA), 00840682109260SEE18170007SA (SEE18170007SA), 00840682109260SEE18170010SA (SEE18170010SA), 00840682109260SEE18170011SA (SEE18170011SA), 00840682109260SEE18170014SA (SEE18170014SA), 00840682109260SEE18170015SA (SEE18170015SA), 00840682109260SEE18170017SA (SEE18170017SA), 00840682109260SEE18170018SA (SEE18170018SA), 00840682109260SEE18170019SA (SEE18170019SA), 00840682109260SEE18170021SA (SEE18170021SA), 00840682109260SEE18170022SA (SEE18170022SA), 00840682109260SEE18170029SA (SEE18170029SA), 00840682109260SEE18170030SA (SEE18170030SA), 00840682109260SEE18170031SA (SEE18170031SA), 00840682109260SEE18170032SA (SEE18170032SA), 00840682109260SEE18170033SA (SEE18170033SA), 00840682109260SEE18280005SA (SEE18280005SA), 00840682109260SEE19180004SA (SEE19180004SA), 00840682109260SEE19180006SA (SEE19180006SA), 00840682109260SEE19180009SA (SEE19180009SA), 00840682109260SEE19180011SA (SEE19180011SA), 00840682109260SEE19180012SA (SEE19180012SA), 00840682109260SEE19190001SA (SEE19190001SA), 00840682109260SEE19190005SA (SEE19190005SA), 00840682109260SEE19190007SA (SEE19190007SA), 00840682109260SEE19190009SA (SEE19190009SA), 00840682109260SEE19190010SA (SEE19190010SA), Not Applicable (SEE09140177GA), Not Applicable (SEE09190242GA), Not Applicable (SEE14422515GA), Not Applicable (SEE14422516GA), Not Applicable (SEE14422517GA), Not Applicable (SEE14422518GA), Not Applicable (SEE14422520GA), Not Applicable (SEE14422521GA), Not Applicable (SEE16160736SA), 00840682109260SEE17490001SA (SEE17490001SA), 00840682109260SEE17490002SA (SEE17490002SA), 00840682109260SEE17490003SA (SEE17490003SA), 00840682109260SEE17490006SA (SEE17490006SA), 00840682109260SEE17490007SA (SEE17490007SA), 00840682109260SEE17490009SA (SEE17490009SA), 00840682109260SEE17490010SA (SEE17490010SA), 00840682109260SEE17490012SA (SEE17490012SA), Not Applicable (SEE11030881GA), Not Applicable (SEE11030883GA), Not Applicable (SEE11030888GA), Not Applicable (SEE13351010GA), Not Applicable (SEE13351023GA), Not Applicable (SEE14372403GA), Not Applicable (SEE10530841GA), Not Applicable (SEE12490256GA), Not Applicable (SEE12490260GA), Not Applicable (SEE12490273GA), Not Applicable (SEE12490274GA), Not Applicable (SEE12490278GA), Not Applicable (SEE13120572GA), Not Applicable (SEE13120574GA), Not Applicable (SEE13120579GA), Not Applicable (SEE13120581GA), Not Applicable (SEE13120582GA), Not Applicable (SEE13120583GA), Not Applicable (SEE13120584GA), Not Applicable (SEE13120599GA), Not Applicable (SEE13120604GA), Not Applicable (SEE14031720GA), Not Applicable (SEE12490295GA), Not Applicable (SEE13401181GA), Not Applicable (SEE13401182GA), Not Applicable (SEE15332309SA), Not Applicable (SEE15332310SA), Not Applicable (SEE09390542GA), Not Applicable (SEE09390547GA), Not Applicable (SEE09460636GA), Not Applicable (SEE09460649GA), Not Applicable (SEE11151312GA), Not Applicable (SEE15113166GA), Not Applicable (SEE15153197GA), Not Applicable (SEE16354346SA), Not Applicable (SEE16360011SA), Not Applicable (SEE16360015SA), 00840682109260SEE17110007SA (SEE17110007SA), Not Applicable (SEE12122137GA), Not Applicable (SEE12122138GA), Not Applicable (SEE12122146GA), Not Applicable (SEE15113138GA), 00840682109260SEE17360010SA (SEE17360010SA), 00840682109260SEE18500030SA (SEE18500030SA), Not Applicable (SEE09360482GA), Not Applicable (SEE14201952GA), Not Applicable (SEE14201953GA), Not Applicable (SEE14201958GA), Not Applicable (SEE14201959GA), Not Applicable (SEE14201960GA), Not Applicable (SEE14201961GA), 00840682109260SEE17150017SA (SEE17150017SA), Not Applicable (SEE11331596GA), 00840682109260SEE18450004SA (SEE18450004SA), Not Applicable (SEE14151816GA), Not Applicable (SEE14151817GA), Not Applicable (SEE14151818GA), Not Applicable (SEE13290964GA), Not Applicable (SEE13290971GA), Not Applicable (SEE15301928SA), Not Applicable (SEE11030890GA), Not Applicable (SEE09250311GA), Not Applicable (SEE09250313GA), Not Applicable (SEE09250314GA), Not Applicable (SEE09300351GA), Not Applicable (SEE09300353GA), Not Applicable (SEE09330416GA), Not Applicable (SEE09330423GA), Not Applicable (SEE09330425GA), Not Applicable (SEE13351050GA), Not Applicable (SEE14502809GA), Not Applicable (SEE14502810GA), Not Applicable (SEE14502811GA), Not Applicable (SEE14502812GA), Not Applicable (SEE14502813GA), Not Applicable (SEE14502814GA), Not Applicable (SEE14502815GA), Not Applicable (SEE14502816GA), Not Applicable (SEE14502817GA), Not Applicable (SEE14502818GA), Not Applicable (SEE15240990SA), Not Applicable (SEE16028369SA), 00840682109260SEE17430016SA (SEE17430016SA), Not Applicable (SEE11071142GA), Not Applicable (SEE12420174GA), Not Applicable (SEE14151837GA), Not Applicable (SEE14151867GA), 00840682109260SEE19220004SA (SEE19220004SA), Not Applicable (SEE09450629GA), Not Applicable (SEE09470673GA), Not Applicable (SEE15405016SA), Not Applicable (SEE15405019SA), 00840682109260SEE17130001SA (SEE17130001SA), 00840682109260SEE17130005SA (SEE17130005SA), 00840682109260SEE17130010SA (SEE17130010 SA), 00840682109260SEE17130012SA (SEE17130012SA), 00840682109260SEE18060024SA (SEE18060024SA), 00840682109260SEE18330001SA (SEE18330001SA), 00840682109260SEE18330006SA (SEE18330006SA), Not Applicable (SEE11371752GA), Not Applicable (SEE11371759GA), Not Applicable (SEE11371797GA), Not Applicable (SEE11371804GA), Not Applicable (SEE12162179GA), Not Applicable (SEE14242096GA), 00840682109260SEE18410009SA (SEE18410009SA), 00840682109260SEE18410024SA (SEE18410024SA), Not Applicable (SEE13140651GA), Not Applicable (SEE13290963GA), Not Applicable (SEE13290985GA), Not Applicable (SEE11511862GA), Not Applicable (SEE11511863GA), Not Applicable (SEE11211477GA), Not Applicable (SEE12021903GA), Not Applicable (SEE12021905GA), Not Applicable (SEE12021912GA), Not Applicable (SEE12021915GA), Not Applicable (SEE12021916GA), Not Applicable (SEE12021917GA), Not Applicable (SEE12021919GA), Not Applicable (SEE12092041GA), Not Applicable (SEE12102105GA), Not Applicable (SEE12102106GA), Not Applicable (SEE14031609GA), 00840682109260SEE17130015SA (SEE17130015SA), 00840682109260SEE17130016SA (SEE17130016SA), 00840682109260SEE19130013SA (SEE19130013SA), 00840682109260SEE19130015SA (SEE19130015SA), 00840682109260SEE19130017SA (SEE19130017SA), 00840682109260SEE18320005SA (SEE18320005SA), 00840682109260SEE18330005SA (SEE18330005SA), Not Applicable (SEE12021901GA), Not Applicable (SEE12410126GA), Not Applicable (SEE12410136GA), Not Applicable (SEE12410141GA), Not Applicable (SEE15312120SA), Not Applicable (SEE15312131SA), Not Applicable (SEE16380015SA), Not Applicable (SEE16380019SA), Not Applicable (SEE09340455GA), Not Applicable (SEE09340456GA), Not Applicable (SEE09350461GA), Not Applicable (SEE09350468GA), Not Applicable (SEE13501455GA), Not Applicable (SEE13501471GA), Not Applicable (SEE13501482GA), 00840682109260SEE18370013SA (SEE18370013SA), 00840682109260SEE18380013SA (SEE18380013SA), Not Applicable (SEE09430605GA), Not Applicable (SEE11141259GA), Not Applicable (SEE14031663GA), Not Applicable (SEE15093034GA), Not Applicable (SEE15093039GA), Not Applicable (SEE11050973GA), Not Applicable (SEE11060953GA), Not Applicable (SEE11060955GA), Not Applicable (SEE11060956GA), Not Applicable (SEE11060959GA), Not Applicable (SEE11060964GA), Not Applicable (SEE11071045GA), Not Applicable (SEE11071145GA), Not Applicable (SEE11081168GA), Not Applicable (SEE13040417GA), Not Applicable (SEE13040428GA), Not Applicable (SEE15312121SA), Not Applicable (SEE12302368GA), Not Applicable (SEE12302369GA), Not Applicable (SEE12302370GA), Not Applicable (SEE12302371GA), Not Applicable (SEE12302372GA), Not Applicable (SEE12302374GA), Not Applicable (SEE12302375GA), Not Applicable (SEE12302378GA), Not Applicable (SEE12302385GA), Not Applicable (SEE09340431GA), Not Applicable (SEE09340433GA), Not Applicable (SEE09340435GA), Not Applicable (SEE11341645GA), Not Applicable (SEE11361698GA), Not Applicable (SEE11361740GA), Not Applicable (SEE09470663GA), Not Applicable (SEE11341653GA), Not Applicable (SEE13411204GA), Not Applicable (SEE13411206GA), Not Applicable (SEE13411208GA), Not Applicable (SEE09330428GA), Not Applicable (SEE12092070GA), Not Applicable (SEE12092071GA), Not Applicable (SEE12102099GA), Not Applicable (SEE12102102GA), Not Applicable (SEE12102103GA), Not Applicable (SEE09480738GA), Not Applicable (SEE09480746GA), Not Applicable (SEE09480747GA), 00840682109260SEE19220017SA (SEE19220017SA), Not Applicable (SEE15022865GA), Not Applicable (SEE16241847SA), Not Applicable (SEE16241856SA), 00840682109260SEE18330037SA (SEE18330037SA), Not Applicable (SEE15517888SA), Not Applicable (SEE15518040SA), Not Applicable (SEE15518041SA), Not Applicable (SEE15518042SA), Not Applicable (SEE15518045SA), Not Applicable (SEE12520335GA), Not Applicable (SEE13000386GA), Not Applicable (SEE10530849GA), Not Applicable (SEE10530859GA), Not Applicable (SEE10530864GA), Not Applicable (SEE10530871GA), Not Applicable (SEE11030880GA), Not Applicable (SEE11040898GA), Not Applicable (SEE10530827GA), Not Applicable (SEE10530831GA), Not Applicable (SEE10530832GA), Not Applicable (SEE10530833GA), Not Applicable (SEE10530834GA), Not Applicable (SEE10530842GA), Not Applicable (SEE12390082GA), Not Applicable (SEE13451333GA), 00840682109260SEE19100032SA (SEE19100032SA), Not Applicable (SEE09100089GA), Not Applicable (SEE09100131GA), Not Applicable (SEE09340441GA), Not Applicable (SEE09410576GA), Not Applicable (SEE10120021GA), Not Applicable (SEE10130074GA), Not Applicable (SEE10260230GA), Not Applicable (SEE10310283GA), Not Applicable (SEE10390325GA), Not Applicable (SEE10390326GA), Not Applicable (SEE10390337GA), Not Applicable (SEE10390353GA), Not Applicable (SEE10390359GA), Not Applicable (SEE10390360GA), Not Applicable (SEE11081173GA), Not Applicable (SEE11081174GA), Not Applicable (SEE11081179GA), Not Applicable (SEE11081182GA), Not Applicable (SEE11081199GA), Not Applicable (SEE11101222GA), Not Applicable (SEE11151315GA), Not Applicable (SEE11171358GA), Not Applicable (SEE11171361GA), Not Applicable (SEE11211512GA), Not Applicable (SEE14312301GA), Not Applicable (SEE14372424GA), Not Applicable (SEE14372439GA), Not Applicable (SEE14382460GA), Not Applicable (SEE14382462GA), Not Applicable (SEE14502785GA), Not Applicable (SEE14502786GA), Not Applicable (SEE14502789GA), Not Applicable (SEE14502790GA), Not Applicable (SEE14502799GA), Not Applicable (SEE14502800GA), Not Applicable (SEE14502802GA), Not Applicable (SEE14502803GA), Not Applicable (SEE14502808GA), Not Applicable (SEE15113158GA), Not Applicable (SEE15113160GA), Not Applicable (SEE15113162GA), Not Applicable (SEE15113164GA), Not Applicable (SEE15113165GA), Not Applicable (SEE15113178GA), Not Applicable (SEE15113179GA), Not Applicable (SEE15220769SA), Not Applicable (SEE15220770SA), Not Applicable (SEE15220773SA), Not Applicable (SEE15220774SA), Not Applicable (SEE15220776SA), Not Applicable (SEE15220778SA), Not Applicable (SEE15220780SA), Not Applicable (SEE15220782SA), Not Applicable (SEE15220783SA), Not Applicable (SEE15220784SA), Not Applicable (SEE15220785SA), Not Applicable (SEE15220791SA), Not Applicable (SEE15261531SA), Not Applicable (SEE15261538SA), Not Applicable (SEE15261539SA), Not Applicable (SEE16099500SA), Not Applicable (SEE16109587SA), Not Applicable (SEE16109701SA), Not Applicable (SEE16109705SA), Not Applicable (SEE16119914SA), Not Applicable (SEE16150555SA), 00840682109260SEE17230046SA (SEE17230046SA), 00840682109260SEE17230047SA (SEE17230047SA), 00840682109260SEE17230048SA (SEE17230048SA), 00840682109260SEE17230049SA (SEE17230049SA), 00840682109260SEE17230050SA (SEE17230050SA), 00840682109260SEE17230051SA (SEE17230051SA), 00840682109260SEE17230053SA (SEE17230053SA), 00840682109260SEE17230054SA (SEE17230054SA), 00840682109260SEE17230064SA (SEE17230064SA), 00840682109260SEE17230065SA (SEE17230065SA), 00840682109260SEE17230066SA (SEE17230066SA), 00840682109260SEE17230067SA (SEE17230067SA), 00840682109260SEE17230068SA (SEE17230068SA), 00840682109260SEE17230069SA (SEE17230069SA), 00840682109260SEE17230070SA (SEE17230070SA), 00840682109260SEE17230071SA (SEE17230071SA), 00840682109260SEE17230072SA (SEE17230072SA), 00840682109260SEE17230073SA (SEE17230073SA), 00840682109260SEE17430046SA (SEE17430046SA), 00840682109260SEE18010002SA (SEE18010002SA), 00840682109260SEE18010017SA (SEE18010017SA), 00840682109260SEE18010018SA (SEE18010018SA), 00840682109260SEE18010019SA (SEE18010019SA), 00840682109260SEE18010021SA (SEE18010021SA), 00840682109260SEE18010022SA (SEE18010022SA), 00840682109260SEE18170003SA (SEE18170003SA), 00840682109260SEE18190026SA (SEE18190026SA), 00840682109260SEE18190030SA (SEE18190030SA), 00840682109260SEE18190032SA (SEE18190032SA), 00840682109260SEE18190037SA (SEE18190037SA), 00840682109260SEE18190038SA (SEE18190038SA), Not Applicable (SEE09020011GA), Not Applicable (SEE09020033GA), Not Applicable (SEE09020038GA), Not Applicable (SEE09430604GA), Not Applicable (SEE11060991GA), Not Applicable (SEE11071017GA), Not Applicable (SEE11071033GA), Not Applicable (SEE11071050GA), Not Applicable (SEE11171371GA), 00840682109260SEE18170028SA (SEE18170028SA), 00840682109260SEE18190011SA (SEE18190011SA), 00840682109260SEE18190017SA (SEE18190017SA), 00840682109260SEE18190024SA (SEE18190024SA), 00840682109260SEE18250031SA (SEE18250031SA), 00840682109260SEE18260001SA (SEE18260001SA), Not Applicable (SEE09430622GA), Not Applicable (SEE14502787GA), Not Applicable (SEE14502796GA), Not Applicable (SEE14502801GA), Not Applicable (SEE14502805GA), Not Applicable (SEE14502807GA), Not Applicable (SEE09340430GA), Not Applicable (SEE09340434GA), Not Applicable (SEE09340437GA), Not Applicable (SEE09100129GA), Not Applicable (SEE10390297GA), Not Applicable (SEE11511882GA), 00840682109260SEE17250001SA (SEE17250001SA), 00840682109260SEE17250003SA (SEE17250003SA), 00840682109260SEE17250004SA (SEE17250004SA), 00840682109260SEE17250005SA (SEE17250005SA), 00840682109260SEE17250010SA (SEE17250010SA), 00840682109260SEE17250011SA (SEE17250011SA), 00840682109260SEE17250012SA (SEE17250012SA), 00840682109260SEE17250013SA (SEE17250013SA), 00840682109260SEE17250014SA (SEE17250014SA), 00840682109260SEE17250015SA (SEE17250015SA), 00840682109260SEE17250016SA (SEE17250016SA), 00840682109260SEE17250017SA (SEE17250017SA), 00840682109260SEE17250018SA (SEE17250018SA), 00840682109260SEE17250019SA (SEE17250019SA), Not Applicable (SEE14322314GA), Not Applicable (SEE15220768SA), Not Applicable (SEE15220771SA), Not Applicable (SEE15220772SA), Not Applicable (SEE15220777SA), Not Applicable (SEE15220779SA), Not Applicable (SEE15220781SA), Not Applicable (SEE10130064GA), Not Applicable (SEE10130075GA), Not Applicable (SEE12082029GA), Not Applicable (SEE12082036GA), Not Applicable (SEE12262327GA), Not Applicable (SEE12262347GA), Not Applicable (SEE14021569GA), Not Applicable (SEE13431288GA), Not Applicable (SEE13431289GA), Not Applicable (SEE10430646GA), Not Applicable (SEE13401183GA), Not Applicable (SEE09530849GA), Not Applicable (SEE09530858GA), Not Applicable (SEE09530875GA), Not Applicable (SEE12340003GA), Not Applicable (SEE12370058GA), Not Applicable (SEE12370066GA), Not Applicable (SEE09530874GA), Not Applicable (SEE09530900GA), Not Applicable (SEE10090957GA), Not Applicable (SEE10121000GA), Not Applicable (SEE10121002GA), Not Applicable (SEE10121013GA), Not Applicable (SEE10121022GA), Not Applicable (SEE14222006GA), Not Applicable (SEE14222009GA), Not Applicable (SEE14222019GA), Not Applicable (SEE09480723GA), Not Applicable (SEE09480730GA), Not Applicable (SEE09480742GA), Not Applicable (SEE09480745GA), Not Applicable (SEE09490752GA), Not Applicable (SEE09490754GA), Not Applicable (SEE10410490GA), Not Applicable (SEE13040420GA), Not Applicable (SEE15153194GA), 00840682109260SEE16078894SA (SEE16078894SA), 00840682109260SEE16078896SA (SEE16078896SA), 00840682109260SEE16078897SA (SEE16078897SA), 00840682109260SEE18040020SA (SEE18040020SA), 00840682109260SEE18040021SA (SEE18040021SA), Not Applicable (SEE13501486GA), Not Applicable (SEE13501487GA), Not Applicable (SEE13501489GA), Not Applicable (SEE13501490GA), Not Applicable (SEE13501494GA), Not Applicable (SEE13501495GA), Not Applicable (SEE13501496GA), Not Applicable (SEE13501497GA), Not Applicable (SEE13501501GA), Not Applicable (SEE15507702SA), 00840682109260SEE11511869GA (SEE11511869GA), 00840682109260SEE11511879GA (SEE11511879GA), 00840682109260SEE11521888GA (SEE11521888GA), Not Applicable (SEE12340001GA), Not Applicable (SEE12340008GA), 00840682109260SEE12340015GA (SEE12340015GA), Not Applicable (SEE12342446GA), 00840682109260SEE12490262GA (SEE12490262GA), 00840682109260SEE12520360GA (SEE12520360GA), 00840682109260SEE12520363GA (SEE12520363GA), 00840682109260SEE12520364GA (SEE12520364GA), Not Applicable (SEE14432613GA), Not Applicable (SEE14492742GA), 00840682109260SEE17430020SA (SEE17430020SA), Not Applicable (SEE11081165GA), 00840682109260SEE15230837SA (SEE15230837SA), 00840682109260SEE17350005SA (SEE17350005SA), 00840682109260SEE17350006SA (SEE17350006SA), 00840682109260SEE17350007SA (SEE17350007SA), 00840682109260SEE17150001SA (SEE17150001SA), 00840682109260SEE17150003SA (SEE17150003SA), 00840682109260SEE17150007SA (SEE17150007SA), 00840682109260SEE18050005SA (SEE18050005SA), 00840682109260SEE18050006SA (SEE18050006SA), 00840682109260SEE18050007SA (SEE18050007SA), 00840682109260SEE18050008SA (SEE18050008SA), 00840682109260SEE18050009SA (SEE18050009SA), 00840682109260SEE18050010SA (SEE18050010SA), 00840682109260SEE18050011SA (SEE18050011SA), Not Applicable (SEE10120029GA), Not Applicable (SEE11371757GA), Not Applicable (SEE11371800GA), Not Applicable (SEE13491442GA), Not Applicable (SEE13491443GA), Not Applicable (SEE13491444GA), Not Applicable (SEE13491446GA), Not Applicable (SEE13491450GA), Not Applicable (SEE13491451GA), Not Applicable (SEE10120023GA), 00840682109260SEE17510001SA (SEE17510001SA), 00840682109260SEE17510010SA (SEE17510010SA), Not Applicable (SEE10121011GA), Not Applicable (SEE10130065GA), Not Applicable (SEE15425793SA), Not Applicable (SEE15425796SA), Not Applicable (SEE15425798SA), Not Applicable (SEE15425799SA), Not Applicable (SEE15425800SA), Not Applicable (SEE09100148GA), Not Applicable (SEE09430607GA), Not Applicable (SEE10400365GA), Not Applicable (SEE10400375GA), Not Applicable (SEE10400385GA), Not Applicable (SEE10400397GA), Not Applicable (SEE12490282GA), Not Applicable (SEE13431259GA), Not Applicable (SEE13431260GA), Not Applicable (SEE13431268GA), Not Applicable (SEE13180784GA), Not Applicable (SEE09470674GA), Not Applicable (SEE12322421GA), Not Applicable (SEE14031672GA), Not Applicable (SEE15113123GA), Not Applicable (SEE15113141GA), Not Applicable (SEE15113156GA), Not Applicable (SEE13110527GA), Not Applicable (SEE13110547GA), Not Applicable (SEE13110551GA), Not Applicable (SEE13110552GA), Not Applicable (SEE13110553GA), Not Applicable (SEE13110555GA), Not Applicable (SEE13110557GA), Not Applicable (SEE13110558GA), Not Applicable (SEE13110559GA), Not Applicable (SEE13110560GA), Not Applicable (SEE13110564GA), Not Applicable (SEE13451317GA), Not Applicable (SEE13451817GA), Not Applicable (SEE13481390GA), Not Applicable (SEE13481391GA), Not Applicable (SEE13481392GA), Not Applicable (SEE13481395GA), Not Applicable (SEE13481396GA), 00840682109260SEE17510005SA (SEE17510005SA), Not Applicable (SEE13491439GA), Not Applicable (SEE13491441GA), Not Applicable (SEE15113120GA), Not Applicable (SEE15113142GA), Not Applicable (SEE15113144GA), Not Applicable (SEE15113148GA), 00840682109260SEE17150018SA (SEE17150018SA), Not Applicable (SEE13361073GA), Not Applicable (SEE13371081GA), Not Applicable (SEE13371091GA), Not Applicable (SEE14362357GA), Not Applicable (SEE14452621GA), Not Applicable (SEE15466917SA), Not Applicable (SEE15466918SA), Not Applicable (SEE15466923SA), Not Applicable (SEE15466924SA), Not Applicable (SEE15466925SA), Not Applicable (SEE15466934SA), Not Applicable (SEE15466935SA), Not Applicable (SEE09170221GA), 00840682109260SEE17130008SA (SEE17130008SA), Not Applicable (SEE14151851GA), Not Applicable (SEE13411198GA), Not Applicable (SEE13411212GA), Not Applicable (SEE13411214GA), 00840682109260SEE19090026SA (SEE19090026SA), 00840682109260SEE19090027SA (SEE19090027SA), Not Applicable (SEE13290960GA), Not Applicable (SEE15183254GA), Not Applicable (SEE15183255GA), Not Applicable (SEE14211983GA), Not Applicable (SEE14252123GA), Not Applicable (SEE14252124GA), Not Applicable (SEE14252125GA), Not Applicable (SEE14252128GA), Not Applicable (SEE14252130GA), Not Applicable (SEE14252131GA), Not Applicable (SEE14252133GA), Not Applicable (SEE14252135GA), Not Applicable (SEE14362353GA), Not Applicable (SEE09510798GA), Not Applicable (SEE10250081GA), Not Applicable (SEE10250091GA), Not Applicable (SEE10250095GA), Not Applicable (SEE10250104GA), Not Applicable (SEE10250110GA), Not Applicable (SEE10250122GA), Not Applicable (SEE10390309GA), Not Applicable (SEE11361725GA), Not Applicable (SEE14512832GA), Not Applicable (SEE14512837GA), Not Applicable (SEE14512838GA), Not Applicable (SEE15012839GA), Not Applicable (SEE15012840GA), Not Applicable (SEE15193279GA), Not Applicable (SEE12510310GA), 00840682109260SEE18370003SA (SEE18370003SA), 00840682109260SEE18370004SA (SEE18370004SA), Not Applicable (SEE09530848GA), Not Applicable (SEE09530852GA), Not Applicable (SEE09020028GA), Not Applicable (SEE10120993GA), Not Applicable (SEE10121005GA), Not Applicable (SEE10121007GA), Not Applicable (SEE10121008GA), Not Applicable (SEE11361713GA), Not Applicable (SEE11361721GA), Not Applicable (SEE11361723GA), Not Applicable (SEE12252316GA), Not Applicable (SEE12252322GA), Not Applicable (SEE12302386GA), Not Applicable (SEE12312397GA), Not Applicable (SEE12312399GA), Not Applicable (SEE12312405GA), Not Applicable (SEE13381124GA), Not Applicable (SEE13381128GA), 01008406821092601116101121SEE16410010SA (SEE13401141GA), Not Applicable (SEE13401154GA), Not Applicable (SEE13401158GA), Not Applicable (SEE13401165GA), Not Applicable (SEE13401168GA), Not Applicable (SEE13401169GA), Not Applicable (SEE13401170GA), Not Applicable (SEE13401171GA), Not Applicable (SEE13501472GA), Not Applicable (SEE14031594GA), Not Applicable (SEE14031604GA), Not Applicable (SEE14031606GA), Not Applicable (SEE14222018GA), Not Applicable (SEE15042906GA), Not Applicable (SEE15052910GA), Not Applicable (SEE15052911GA), Not Applicable (SEE15052924GA), Not Applicable (SEE15052931GA), 00840682109260SEE18160041SA (SEE18160041SA), 00840682109260SEE18160043SA (SEE18160043SA), 00840682109260SEE18160047SA (SEE18160047SA), 00840682109260SEE18160053SA (SEE18160053SA), 00840682109260SEE18160060SA (SEE18160060SA), 00840682109260SEE19120022SA (SEE19120022SA), 00840682109260SEE19120023SA (SEE19120023SA), 00840682109260SEE19120025SA (SEE19120025SA), 00840682109260SEE19120031SA (SEE19120031SA), 00840682109260SEE19120032SA (SEE19120032SA), 00840682109260SEE19120033SA (SEE19120033SA), 00840682109260SEE19120034SA (SEE19120034SA), 00840682109260SEE19130009SA (SEE19130009SA), 00840682109260SEE19130020SA (SEE19130020SA), 00840682109260SEE19150008SA (SEE19150008SA), Not Applicable (SEE15517774SA), Not Applicable (SEE15517821SA), Not Applicable (SEE15517824SA), Not Applicable (SEE11511881GA), Not Applicable (SEE14021519GA), 00840682109260SEE19220020SA (SEE19220020SA), 00840682109260SEE19220030SA (SEE19220030SA), 00840682109260SEE19220031SA (SEE19220031SA), 00840682109260SEE19220032SA (SEE19220032SA), 00840682109260SEE19220040SA (SEE19220040SA), Not Applicable (SEE14502791GA), Not Applicable (SEE14502793GA), Not Applicable (SEE14502797GA), Not Applicable (SEE14502798GA), Not Applicable (SEE14512819GA), Not Applicable (SEE14512823GA), Not Applicable (SEE14512827GA), Not Applicable (SEE16119768SA), Not Applicable (SEE13110502GA), Not Applicable (SEE13110524GA), Not Applicable (SEE14472676GA), Not Applicable (SEE14482685GA), Not Applicable (SEE14482687GA), Not Applicable (SEE14482691GA), Not Applicable (SEE14482692GA), Not Applicable (SEE14482693GA), Not Applicable (SEE14482694GA), Not Applicable (SEE14482696GA), Not Applicable (SEE14482699GA), Not Applicable (SEE14362379GA), Not Applicable (SEE14362380GA), Not Applicable (SEE14362383GA), Not Applicable (SEE14362386GA), Not Applicable (SEE14372388GA), Not Applicable (SEE14372392GA), Not Applicable (SEE14372393GA), Not Applicable (SEE14372394GA), 00840682109260SEE18060005SA (SEE18060005SA), 00840682109260SEE18060006SA (SEE18060006SA), 00840682109260SEE19130016SA (SEE19130016SA), Not Applicable (SEE09250307GA), Not Applicable (SEE15446332SA), Not Applicable (SEE12092056GA), Not Applicable (SEE13180762GA), Not Applicable (SEE09190226GA), Not Applicable (SEE11171374GA), Not Applicable (SEE11081195GA), 00840682109260SEE17150004SA (SEE17150004SA), Not Applicable (SEE16252018SA), Not Applicable (SEE10450670GA), Not Applicable (SEE13401161GA), Not Applicable (SEE12302389GA), Not Applicable (SEE12302390GA), Not Applicable (SEE15251303SA), Not Applicable (SEE12322434GA), Not Applicable (SEE15352700SA), Not Applicable (SEE10420589GA), Not Applicable (SEE12430223GA), Not Applicable (SEE12430225GA), Not Applicable (SEE12440234GA), Not Applicable (SEE12440237GA), Not Applicable (SEE12490261GA), Not Applicable (SEE13371113GA), Not Applicable (SEE14372438GA), Not Applicable (SEE10270244GA), Not Applicable (SEE16303409GA), Not Applicable (SEE10250168GA), 00840682109260SEE18110004SA (SEE18110004SA), Not Applicable (SEE14242082GA), Not Applicable (SEE14242084GA), Not Applicable (SEE14242093GA), Not Applicable (SEE10120006GA), Not Applicable (SEE10120012GA), Not Applicable (SEE10120016GA), Not Applicable (SEE10120034GA), Not Applicable (SEE10121003GA), Not Applicable (SEE14021559GA), Not Applicable (SEE14021574GA), Not Applicable (SEE14021575GA), Not Applicable (SEE14272145GA), Not Applicable (SEE14302274GA), Not Applicable (SEE14312295GA), Not Applicable (SEE14322320GA), Not Applicable (SEE14322335GA), Not Applicable (SEE16038425SA), Not Applicable (SEE14432579GA), 00840682109260SEE17080011SA (SEE17080011SA), Not Applicable (SEE12520337GA), Not Applicable (SEE12520352GA), Not Applicable (SEE12520356GA), Not Applicable (SEE12520361GA), Not Applicable (SEE12520362GA), Not Applicable (SEE14362382GA), Not Applicable (SEE14362387GA), 00840682109260SEE18470010SA (SEE18470010SA), 00840682109260SEE18470011SA (SEE18470011SA), 00840682109260SEE18470012SA (SEE18470012SA), 00840682109260SEE18470013SA (SEE18470013SA), 00840682109260SEE18470014SA (SEE18470014SA), 00840682109260SEE18470017SA (SEE18470017SA), 00840682109260SEE18470021SA (SEE18470021SA), 00840682109260SEE18470025SA (SEE18470025SA), 00840682109260SEE18470026SA (SEE18470026SA), 00840682109260SEE18470027SA (SEE18470027SA), Not Applicable (SEE12520355GA), Not Applicable (SEE14151813GA), Not Applicable (SEE14151881GA), 00840682109260SEE17480011SA (SEE17480011SA), Not Applicable (SEE16430003SA), Not Applicable (SEE14021514GA), Not Applicable (SEE14161893GA), Not Applicable (SEE15415171SA), Not Applicable (SEE14492729GA), Not Applicable (SEE14492743GA), 00840682109260SEE18040040SA (SEE18040040SA), Not Applicable (SEE13411215GA), Not Applicable (SEE10120019GA), 00840682109260SEE17050012SA (SEE17050012SA), 00840682109260SEE17050013SA (SEE17050013SA), 00840682109260SEE17050014SA (SEE17050014SA), Not Applicable (SEE15173226GA), Not Applicable (SEE09100150GA), Not Applicable (SEE11071007GA), Not Applicable (SEE11141258GA), Not Applicable (SEE11141266GA), Not Applicable (SEE15173228GA), Not Applicable (SEE15173229GA), Not Applicable (SEE15173241GA), Not Applicable (SEE15173242GA), Not Applicable (SEE15173246GA), Not Applicable (SEE16221553SA), Not Applicable (SEE16221557SA), 00840682109260SEE19100023SA (SEE19100023SA), 00840682109260SEE19100024SA (SEE19100024SA), 00840682109260SEE19100029SA (SEE19100029SA), 00840682109260SEE19100031SA (SEE19100031SA), 00840682109260SEE19100034SA (SEE19100034SA), 00840682109260SEE19100063SA (SEE19100063SA), 00840682109260SEE19120011SA (SEE19120011SA), 00840682109260SEE19120013SA (SEE19120013SA), 00840682109260SEE17250026SA (SEE17250026SA), 00840682109260SEE17250027SA (SEE17250027SA), 00840682109260SEE17250028SA (SEE17250028SA), 00840682109260SEE17270016SA (SEE17270016SA), Not Applicable (SEE13150696GA), Not Applicable (SEE13150698GA), Not Applicable (SEE13150699GA), Not Applicable (SEE13150700GA), Not Applicable (SEE15352677SA), Not Applicable (SEE16089278SA), Not Applicable (SEE16262252SA), Not Applicable (SEE13361077GA), Not Applicable (SEE13491435GA), Not Applicable (SEE13491438GA), Not Applicable (SEE14101777GA), Not Applicable (SEE14111799GA), Not Applicable (SEE14111805GA), Not Applicable (SEE12520338GA), Not Applicable (SEE12520344GA), 00840682109260SEE17040001SA (SEE17040001SA), 00840682109260SEE17180016SA (SEE17180016SA), Not Applicable (SEE11050976GA), Not Applicable (SEE10100972GA), Not Applicable (SEE11071012GA), Not Applicable (SEE11071022GA), Not Applicable (SEE11071095GA), Not Applicable (SEE14302273GA), Not Applicable (SEE11141272GA), Not Applicable (SEE10420557GA), Not Applicable (SEE10430606GA), Not Applicable (SEE10450659GA), Not Applicable (SEE10450676GA), Not Applicable (SEE14362350GA), Not Applicable (SEE14362351GA), Not Applicable (SEE14362374GA), Not Applicable (SEE15517823SA), Not Applicable (SEE13180767GA), Not Applicable (SEE15173212GA), Not Applicable (SEE12021914GA), Not Applicable (SEE12062005GA), Not Applicable (SEE12082024GA), 00840682109260SEE17100006SA (SEE17100006SA), 00840682109260SEE18510017SA (SEE18510017SA), 00840682109260SEE18510020SA (SEE18510020SA), 00840682109260SEE18510025SA (SEE18510025SA), 00840682109260SEE18510026SA (SEE18510026SA), 00840682109260SEE18510027SA (SEE18510027SA), 00840682109260SEE18510029SA (SEE18510029SA), 00840682109260SEE19010005SA (SEE19010005SA), 00840682109260SEE19010006SA (SEE19010006SA), 00840682109260SEE19010011SA (SEE19010011SA), 00840682109260SEE19010012SA (SEE19010012SA), 00840682109260SEE19010013SA (SEE19010013SA), 00840682109260SEE19010014SA (SEE19010014SA), Not Applicable (SEE13280904GA), Not Applicable (SEE14312299GA), Not Applicable (SEE14312304GA), Not Applicable (SEE14312305GA), Not Applicable (SEE14312306GA), Not Applicable (SEE14422551GA), Not Applicable (SEE11381815GA), 00840682109260SEE18420014SA (SEE18420014SA), 00840682109260SEE18420025SA (SEE18420025SA), 00840682109260SEE18420040SA (SEE18420040SA), 00840682109260SEE18420050SA (SEE18420050SA), Not Applicable (SEE14482690GA), Not Applicable (SEE11181397GA), Not Applicable (SEE11191424GA), Not Applicable (SEE11191426GA), Not Applicable (SEE11191438GA), 00840682109260SEE19190004SA (SEE19190004SA), Not Applicable (SEE16201405SA), Not Applicable (SEE16211410SA), Not Applicable (SEE16211412SA), Not Applicable (SEE16211420SA), Not Applicable (SEE16211507SA), Not Applicable (SEE16211508SA), Not Applicable (SEE16211509SA), Not Applicable (SEE16211511SA), Not Applicable (SEE16211512SA), 00840682109260SEE09510805GA (SEE09510805GA), 00840682109260SEE09510825GA (SEE09510825GA), 00840682109260SEE11081206GA (SEE11081206GA), 00840682109260SEE11081208GA (SEE11081208GA), 00840682109260SEE12041948GA (SEE12041948GA), 00840682109260SEE12092074GA (SEE12092074GA), 00840682109260SEE12122144GA (SEE12122144GA), 00840682109260SEE12162174GA (SEE12162174GA), 00840682109260SEE12162176GA (SEE12162176GA), 00840682109260SEE12172189GA (SEE12172189GA), 00840682109260SEE12172191GA (SEE12172191GA), 00840682109260SEE12172199GA (SEE12172199GA), 00840682109260SEE12182207GA (SEE12182207GA), 00840682109260SEE12182208GA (SEE12182208GA), 00840682109260SEE12182209GA (SEE12182209GA), 00840682109260SEE12182213GA (SEE12182213GA), 00840682109260SEE12182216GA (SEE12182216GA), 00840682109260SEE12182222GA (SEE12182222GA), 00840682109260SEE12182228GA (SEE12182228GA), 00840682109260SEE12182229GA (SEE12182229GA), 00840682109260SEE13501457GA (SEE13501457GA), 00840682109260SEE13501480GA (SEE13501480GA), 00840682109260SEE14041739GA (SEE14041739GA), 00840682109260SEE14041743GA (SEE14041743GA), Not Applicable (SEE15456708SA), Not Applicable (SEE15477181SA), 00840682109260SEE17520003SA (SEE17520003SA), Not Applicable (SEE14312297GA), Not Applicable (SEE14312300GA), Not Applicable (SEE14312302GA), Not Applicable (SEE14312303GA), Not Applicable (SEE16119851GA), Not Applicable (SEE16119851SA), Not Applicable (SEE10460691GA), Not Applicable (SEE10460702GA), Not Applicable (SEE10470786GA), Not Applicable (SEE10470795GA), Not Applicable (SEE13180757GA), Not Applicable (SEE13180759GA), Not Applicable (SEE12182214GA), Not Applicable (SEE12182220GA), 00840682109260SEE18010020SA (SEE18010020SA), Not Applicable (SEE09230263GA), Not Applicable (SEE09240291GA), Not Applicable (SEE14021511GA), Not Applicable (SEE14021530GA), Not Applicable (SEE14161897GA), Not Applicable (SEE14161898GA), Not Applicable (SEE14161902GA), Not Applicable (SEE14161903GA), Not Applicable (SEE14161904GA), Not Applicable (SEE14161906GA), Not Applicable (SEE14161908GA), Not Applicable (SEE13110538GA), Not Applicable (SEE13110542GA), Not Applicable (SEE13120580GA), Not Applicable (SEE13120587GA), Not Applicable (SEE13120589GA), Not Applicable (SEE13120590GA), Not Applicable (SEE13120603GA), Not Applicable (SEE16109700SA), Not Applicable (SEE16120021SA), Not Applicable (SEE16221556SA), 00840682109260SEE17090014SA (SEE17090014SA), 00840682109260SEE17090015SA (SEE17090015SA), Not Applicable (SEE15281648SA), Not Applicable (SEE14482682GA), Not Applicable (SEE14482686GA), Not Applicable (SEE10270241GA), Not Applicable (SEE10310281GA), Not Applicable (SEE14151835GA), Not Applicable (SEE14151840GA), Not Applicable (SEE14151845GA), Not Applicable (SEE14151846GA), Not Applicable (SEE14151852GA), Not Applicable (SEE14151869GA), Not Applicable (SEE14171916GA), Not Applicable (SEE15456716SA), Not Applicable (SEE15497518SA), Not Applicable (SEE15497519SA), Not Applicable (SEE15497527SA), Not Applicable (SEE15497582SA), Not Applicable (SEE15507660SA), Not Applicable (SEE16430049SA), Not Applicable (SEE16430057SA), Not Applicable (SEE16440001SA), Not Applicable (SEE16440005SA), 00840682109260SEE17080012SA (SEE17080012SA), 00840682109260SEE17090011SA (SEE17090011SA), 00840682109260SEE17090013SA (SEE17090013SA), 00840682109260SEE17090016SA (SEE17090016SA), 00840682109260SEE17090018SA (SEE17090018SA), 00840682109260SEE17090019SA (SEE17090019SA), 00840682109260SEE17190022SA (SEE17190022SA), 00840682109260SEE17090024SA (SEE17090024SA), Not Applicable (SEE11371746GA), 00840682109260SEE11371768GA (SEE11371768GA), Not Applicable (SEE11371770GA), Not Applicable (SEE11371771GA), Not Applicable (SEE11371772GA), Not Applicable (SEE11371773GA), Not Applicable (SEE11371776GA), Not Applicable (SEE11371777GA), Not Applicable (SEE11371778GA), Not Applicable (SEE11371780GA), Not Applicable (SEE11371783GA), Not Applicable (SEE12051993GA), Not Applicable (SEE12051994GA), Not Applicable (SEE12051995GA), Not Applicable (SEE15173213GA), 00840682109260SEE19110010SA (SEE19110010SA), 00840682109260SEE19110012SA (SEE19110012SA), 00840682109260SEE19110013SA (SEE19110013SA), 00840682109260SEE19110014SA (SEE19110014SA), 00840682109260SEE19110015SA (SEE19110015SA), 00840682109260SEE19110020SA (SEE19110020SA), 00840682109260SEE19110021SA (SEE19110021SA), Not Applicable (SEE10250126GA), Not Applicable (SEE10250151GA), Not Applicable (SEE10250153GA), Not Applicable (SEE10250154GA), Not Applicable (SEE15373617SA), Not Applicable (SEE14492744GA), Not Applicable (SEE10460747GA), Not Applicable (SEE10460752GA), Not Applicable (SEE11030887GA), 00840682109260SEE18170024SA (SEE18170024SA), Not Applicable (SEE10530872GA), Not Applicable (SEE110308789GA), Not Applicable (SEE11030879GA), Not Applicable (SEE15517782SA), Not Applicable (SEE09020009GA), Not Applicable (SEE09020013GA), Not Applicable (SEE09020023GA), Not Applicable (SEE09020024GA), Not Applicable (SEE09020040GA), Not Applicable (SEE09020042GA), Not Applicable (SEE09040045GA), Not Applicable (SEE09040051GA), Not Applicable (SEE09040054GA), Not Applicable (SEE09040055GA), Not Applicable (SEE09040058GA), Not Applicable (SEE09040060GA), Not Applicable (SEE09040061GA), Not Applicable (SEE09040064GA), Not Applicable (SEE09040069GA), Not Applicable (SEE09040071GA), Not Applicable (SEE09040072GA), Not Applicable (SEE09040075GA), Not Applicable (SEE09040077GA), Not Applicable (SEE09100087GA), Not Applicable (SEE09100095GA), Not Applicable (SEE09100105GA), Not Applicable (SEE09100108GA), Not Applicable (SEE09100109GA), Not Applicable (SEE09100134GA), Not Applicable (SEE09100139GA), Not Applicable (SEE09100140GA), Not Applicable (SEE09100152GA), Not Applicable (SEE09140183GA), Not Applicable (SEE09190244GA), Not Applicable (SEE09230253GA), Not Applicable (SEE09230266GA), Not Applicable (SEE09250310GA), Not Applicable (SEE09370494GA), Not Applicable (SEE09370502GA), Not Applicable (SEE09410552GA), Not Applicable (SEE09410561GA), Not Applicable (SEE09410563GA), Not Applicable (SEE09410564GA), Not Applicable (SEE09410565GA), Not Applicable (SEE09410566GA), Not Applicable (SEE09410567GA), Not Applicable (SEE09410574GA), Not Applicable (SEE09410577GA), Not Applicable (SEE09410578GA), Not Applicable (SEE09410579GA), Not Applicable (SEE09410587GA), Not Applicable (SEE10420525GA), Not Applicable (SEE10420535GA), Not Applicable (SEE10420547GA), Not Applicable (SEE10420597GA), Not Applicable (SEE10430607GA), Not Applicable (SEE10430609GA), Not Applicable (SEE10430611GA), Not Applicable (SEE10430617GA), Not Applicable (SEE10430621GA), Not Applicable (SEE10430635GA), Not Applicable (SEE10430642GA), Not Applicable (SEE10430644GA), Not Applicable (SEE10450682GA), Not Applicable (SEE104606 93GA), Not Applicable (SEE10460700GA), Not Applicable (SEE10460728GA), Not Applicable (SEE10460758GA), Not Applicable (SEE10460775GA), Not Applicable (SEE10470776GA), Not Applicable (SEE10470779GA), Not Applicable (SEE10470780GA), Not Applicable (SEE10470781GA), Not Applicable (SEE10470797GA), Not Applicable (SEE12530365GA), Not Applicable (SEE13270881GA), Not Applicable (SEE13491440GA), Not Applicable (SEE14031597GA), Not Applicable (SEE14031598GA), Not Applicable (SEE14031601GA), Not Applicable (SEE14031602GA), Not Applicable (SEE14031693GA), Not Applicable (SEE14031694GA), Not Applicable (SEE14031699GA), Not Applicable (SEE14031701GA), Not Applicable (SEE14031709GA), Not Applicable (SEE14031711GA), Not Applicable (SEE14031713GA), Not Applicable (SEE14041762GA), Not Applicable (SEE14151868GA), Not Applicable (SEE14151871GA), Not Applicable (SEE16119903SA), Not Applicable (SEE16430033SA), Not Applicable (SEE16430048SA), 00840682109260SEE17430043SA (SEE17430043SA), 00840682109260SEE17430044SA (SEE17430044SA), 00840682109260SEE17490005SA (SEE17490005SA), 00840682109260SEE17490008SA (SEE17490008SA), 00840682109260SEE17490011SA (SEE17490011SA), 00840682109260SEE17490013SA (SEE17490013SA), 00840682109260SEE18040053SA (SEE18040053SA), 00840682109260SEE18290005SA (SEE18290005SA), 00840682109260SEE18320008SA (SEE18320008SA), Not Applicable (SEE11141295GA), Not Applicable (SEE14101784GA), Not Applicable (SEE14222028GA), Not Applicable (SEE14222037GA), Not Applicable (SEE15394376SA), Not Applicable (SEE16068748SA), Not Applicable (SEE16241826SA), Not Applicable (SEE16292869SA), 00840682109260SEE17430045SA (SEE17430045SA), Not Applicable (SEE16390014SA), Not Applicable (SEE11201448GA), Not Applicable (SEE14322313GA), Not Applicable (SEE14322316GA), Not Applicable (SEE09310360GA), Not Applicable (SEE09310378GA), Not Applicable (SEE14422525GA), Not Applicable (SEE14422529GA), Not Applicable (SEE15394730SA), Not Applicable (SEE15394732SA), Not Applicable (SEE15394741SA), Not Applicable (SEE15405018SA), 00840682109260SEE18420054SA (SEE18420054SA), Not Applicable (SEE10400383GA), Not Applicable (SEE10400396GA), Not Applicable (SEE10400429GA), Not Applicable (SEE10400454GA), Not Applicable (SEE10400458GA), Not Applicable (SEE10420569GA), Not Applicable (SEE11231543GA), 00840682109260SEE14041744GA (SEE14041744GA), Not Applicable (SEE14041753GA), Not Applicable (SEE14292241GA), Not Applicable (SEE14302265GA), Not Applicable (SEE14322324GA), Not Applicable (SEE14322328GA), Not Applicable (SEE14322337GA), 00840682109260SEE19240018SA (SEE19240018SA), Not Applicable (SEE09340446GA), Not Applicable (SEE12302384GA), Not Applicable (SEE12340020GA), Not Applicable (SEE13110493GA), Not Applicable (SEE13110540GA), Not Applicable (SEE13120602GA), Not Applicable (SEE13140658GA), Not Applicable (SEE14031612GA), Not Applicable (SEE09020012GA), Not Applicable (SEE09020031GA), Not Applicable (SEE10130042GA), Not Applicable (SEE10250094GA), Not Applicable (SEE10420522GA), Not Applicable (SEE10460764GA), Not Applicable (SEE11030886GA), Not Applicable (SEE11071014GA), Not Applicable (SEE11071042GA), Not Applicable (SEE11071149GA), Not Applicable (SEE11081183GA), Not Applicable (SEE11081184GA), Not Applicable (SEE11111236GA), Not Applicable (SEE11111238GA), Not Applicable (SEE11111246GA), Not Applicable (SEE11141278GA), Not Applicable (SEE11141286GA), Not Applicable (SEE11201462GA), Not Applicable (SEE11211474GA), Not Applicable (SEE11211479GA), Not Applicable (SEE11211485GA), Not Applicable (SEE11211490GA), Not Applicable (SEE11211499GA), Not Applicable (SEE11211502GA), Not Applicable (SEE11211522GA), Not Applicable (SEE11231530GA), Not Applicable (SEE11231545GA), Not Applicable (SEE11241548GA), Not Applicable (SEE11341658GA), Not Applicable (SEE11341662GA), Not Applicable (SEE11371750GA), Not Applicable (SEE11371791GA), Not Applicable (SEE11371811GA), Not Applicable (SEE11511873GA), Not Applicable (SEE11511880GA), Not Applicable (SEE12011896GA), Not Applicable (SEE12092042GA), Not Applicable (SEE12162172GA), Not Applicable (SEE12182205GA), Not Applicable (SEE12262334GA), Not Applicable (SEE12262342GA), Not Applicable (SEE12262346GA), Not Applicable (SEE12322427GA), Not Applicable (SEE12390074GA), Not Applicable (SEE12420173GA), Not Applicable (SEE12510313GA), Not Applicable (SEE13110497GA), Not Applicable (SEE13110499GA), Not Applicable (SEE13110501GA), Not Applicable (SEE13110506GA), Not Applicable (SEE13110518GA), Not Applicable (SEE13110522GA), Not Applicable (SEE13110523GA), Not Applicable (SEE13150659GA), Not Applicable (SEE13361078GA), Not Applicable (SEE13411205GA), Not Applicable (SEE14031675GA), Not Applicable (SEE14031700GA), Not Applicable (SEE14041749GA), Not Applicable (SEE14041752GA), Not Applicable (SEE14041754GA), Not Applicable (SEE14041755GA), Not Applicable (SEE14041764GA), Not Applicable (SEE14191950GA), Not Applicable (SEE14201951GA), Not Applicable (SEE14201954GA), Not Applicable (SEE14201955GA), Not Applicable (SEE14201957GA), Not Applicable (SEE14211965GA), Not Applicable (SEE14211966GA), Not Applicable (SEE14211967GA), Not Applicable (SEE14211968GA), Not Applicable (SEE14211969GA), Not Applicable (SEE14211970GA), Not Applicable (SEE14211971GA), Not Applicable (SEE14211973GA), Not Applicable (SEE14211974GA), Not Applicable (SEE14211975GA), Not Applicable (SEE14211977GA), Not Applicable (SEE14272146GA), Not Applicable (SEE14272164GA), Not Applicable (SEE14272168GA), Not Applicable (SEE14272170GA), Not Applicable (SEE14272175GA), Not Applicable (SEE14272224GA), Not Applicable (SEE14332343GA), Not Applicable (SEE14362369GA), Not Applicable (SEE15032882GA), Not Applicable (SEE15062980GA), Not Applicable (SEE15062986GA), Not Applicable (SEE15210703SA), Not Applicable (SEE15251288SA), Not Applicable (SEE15251289SA), Not Applicable (SEE15251290SA), Not Applicable (SEE15251291SA), Not Applicable (SEE16272332SA), Not Applicable (SEE16293173SA), Not Applicable (SEE16293179SA), Not Applicable (SEE16303212SA), Not Applicable (SEE16303220SA), Not Applicable (SEE16334183SA), Not Applicable (SEE16360018SA), Not Applicable (SEE16360019SA), Not Applicable (SEE16360022SA), Not Applicable (SEE16364417SA), 00840682109260SEE17070001SA (SEE17070001SA), 00840682109260SEE17070003SA (SEE17070003SA), 00840682109260SEE17260007SA (SEE17260007SA), 00840682109260SEE18320004SA (SEE18320004SA), 00840682109260SEE18370006SA (SEE18370006SA), 00840682109260SEE18390003SA (SEE18390003SA), 00840682109260SEE18420041SA (SEE18420041SA), 00840682109260SEE18420045SA (SEE18420045SA), 00840682109260SEE18470016SA (SEE18470016SA), 00840682109260SEE18470020SA (SEE18470020SA), 00840682109260SEE18510022SA (SEE18510022SA), Not Applicable (SEE12282354GA), Not Applicable (SEE12282356GA), Not Applicable (SEE13290984GA), Not Applicable (SEE13311004GA), Not Applicable (SEE13351011GA), Not Applicable (SEE13351013GA), Not Applicable (SEE13351016GA), Not Applicable (SEE13351017GA), Not Applicable (SEE13351024GA), Not Applicable (SEE13351027GA), Not Applicable (SEE13351028GA), Not Applicable (SEE13351029GA), Not Applicable (SEE13351032GA), Not Applicable (SEE13491428GA), Not Applicable (SEE11341655GA), Not Applicable (SEE11341659GA), Not Applicable (SEE11341661GA), Not Applicable (SEE11371758GA), Not Applicable (SEE11371793GA), Not Applicable (SEE12420159GA), Not Applicable (SEE12490250GA), Not Applicable (SEE14031630GA), Not Applicable (SEE16272270SA), Not Applicable (SEE16272272SA), 00840682109260SEE17490020SA (SEE17490020SA), 00840682109260SEE18290009SA (SEE18290009SA), Not Applicable (SEE16068832SA), Not Applicable (SEE09530904GA), Not Applicable (SEE09530838GA), Not Applicable (SEE09530850GA), Not Applicable (SEE09530908GA), Not Applicable (SEE10120007GA), Not Applicable (SEE10390296GA), Not Applicable (SEE10390312GA), Not Applicable (SEE10390313GA), Not Applicable (SEE13040418GA), Not Applicable (SEE16028368SA), Not Applicable (SEE16119908SA), Not Applicable (SEE16450015SA), Not Applicable (SEE16450016SA), 00840682109260SEE17270011SA (SEE17270011SA), Not Applicable (SEE15301890SA), Not Applicable (SEE16089268SA), Not Applicable (SEE16089277SA), Not Applicable (SEE16089279SA), Not Applicable (SEE09250318GA), Not Applicable (SEE11151316GA), Not Applicable (SEE11161319GA), Not Applicable (SEE11161321GA), Not Applicable (SEE11371762GA), Not Applicable (SEE12122130GA), Not Applicable (SEE12410134GA), Not Applicable (SEE12490259GA), Not Applicable (SEE12520354GA), Not Applicable (SEE13190804GA), Not Applicable (SEE13190806GA), Not Applicable (SEE14031679GA), Not Applicable (SEE14222042GA), Not Applicable (SEE09530839GA), Not Applicable (SEE09530878GA), Not Applicable (SEE09530882GA), Not Applicable (SEE09530906GA), Not Applicable (SEE10120999GA), Not Applicable (SEE10460738GA), Not Applicable (SEE11371760GA), Not Applicable (SEE11371761GA), Not Applicable (SEE14021584GA), Not Applicable (SEE14031662GA), Not Applicable (SEE14031668GA), Not Applicable (SEE14031673GA), 00840682109260SEE18410022SA (SEE18410022SA), Not Applicable (SEE14452634GA), Not Applicable (SEE14452636GA), Not Applicable (SEE14452639GA), Not Applicable (SEE14452641GA), Not Applicable (SEE16390003SA), Not Applicable (SEE16390010SA), Not Applicable (SEE15113118GA), Not Applicable (SEE15113140GA), Not Applicable (SEE09450623GA), Not Applicable (SEE09450627GA), Not Applicable (SEE09460632GA), Not Applicable (SEE09460633GA), Not Applicable (SEE09460634GA), Not Applicable (SEE09460635GA), Not Applicable (SEE09460637GA), Not Applicable (SEE15466921SA), Not Applicable (SEE09330421GA), Not Applicable (SEE10120020GA), Not Applicable (SEE10530829GA), Not Applicable (SEE15093067GA), Not Applicable (SEE15103096GA), Not Applicable (SEE15103100GA), Not Applicable (SEE15103102GA), Not Applicable (SEE15103104GA), Not Applicable (SEE15103108GA), Not Applicable (SEE15103110GA), Not Applicable (SEE10420559GA), Not Applicable (SEE10420600GA), Not Applicable (SEE10430608GA), Not Applicable (SEE10430612GA), Not Applicable (SEE10430631GA), Not Applicable (SEE10430632GA), Not Applicable (SEE10430648GA), Not Applicable (SEE10450674GA), Not Applicable (SEE10450688GA), Not Applicable (SEE11060990GA), Not Applicable (SEE11071046GA), Not Applicable (SEE11081170GA), Not Applicable (SEE11081194GA), Not Applicable (SEE11081201GA), Not Applicable (SEE11081209GA), Not Applicable (SEE11101220GA), Not Applicable (SEE11101221GA), Not Applicable (SEE12062011GA), Not Applicable (SEE13411218GA), Not Applicable (SEE13411220GA), Not Applicable (SEE13411221GA), Not Applicable (SEE13411222GA), 00840682109260SEE18500027SA (SEE18500027SA), 00840682109260SEE19120014SA (SEE19120014SA), Not Applicable (SEE12282358GA), Not Applicable (SEE13501481GA), Not Applicable (SEE13501484GA), Not Applicable (SEE13501488GA), Not Applicable (SEE13501491GA), Not Applicable (SEE13501492GA), Not Applicable (SEE13501499GA), Not Applicable (SEE13501500GA), Not Applicable (SEE14372419GA), Not Applicable (SEE15425780SA), Not Applicable (SEE15425797SA), Not Applicable (SEE16241824SA), Not Applicable (SEE16241842SA), Not Applicable (SEE16360023SA), Not Applicable (SEE16410003SA), Not Applicable (SEE09100098GA), Not Applicable (SEE12282353GA), Not Applicable (SEE09340452GA), Not Applicable (SEE15394277SA), Not Applicable (SEE14031657GA), Not Applicable (SEE14031658GA), Not Applicable (SEE13180735GA), Not Applicable (SEE11371807GA), Not Applicable (SEE12031927GA), Not Applicable (SEE12172196GA), Not Applicable (SEE13491430GA), Not Applicable (SEE13270874GA), Not Applicable (SEE09490750GA), Not Applicable (SEE09490759GA), Not Applicable (SEE12031941GA), Not Applicable (SEE11141257GA), Not Applicable (SEE09480741GA), Not Applicable (SEE09130170GA), Not Applicable (SEE09150207GA), Not Applicable (SEE09240299GA), Not Applicable (SEE09310367GA), Not Applicable (SEE09310389GA), Not Applicable (SEE09310390GA), Not Applicable (SEE09310391GA), Not Applicable (SEE09310392GA), Not Applicable (SEE09310393GA), Not Applicable (SEE09310394GA), Not Applicable (SEE09310396GA), Not Applicable (SEE09310397GA), Not Applicable (SEE09310400GA), Not Applicable (SEE09310401GA), Not Applicable (SEE09310402GA), Not Applicable (SEE09310403GA), Not Applicable (SEE09310404GA), Not Applicable (SEE09310405GA), Not Applicable (SEE09310406GA), Not Applicable (SEE10470778GA), Not Applicable (SEE10470798GA), Not Applicable (SEE13180765GA), Not Applicable (SEE13180774GA), Not Applicable (SEE14031705GA), 00840682109260SEE17270001SA (SEE17270001SA), 00840682109260SEE17270014SA (SEE17270014SA), 00840682109260SEE19010002SA (SEE19010002SA), 00840682109260SEE19010004SA (SEE19010004SA), 00840682109260SEE19010009SA (SEE19010009SA), 00840682109260SEE19090002SA (SEE19090002SA), 00840682109260SEE19090024SA (SEE19090024SA), Not Applicable (SEE14472680GA), Not Applicable (SEE15446228SA), Not Applicable (SEE15446229SA), Not Applicable (SEE15446231SA), Not Applicable (SEE15446232SA), Not Applicable (SEE15446233SA), Not Applicable (SEE15517773SA), Not Applicable (SEE11181406GA), Not Applicable (SEE11181408GA), Not Applicable (SEE11181409GA), Not Applicable (SEE11181411GA), Not Applicable (SEE11181416GA), Not Applicable (SEE11181418GA), Not Applicable (SEE11181419GA), Not Applicable (SEE11181421GA), Not Applicable (SEE16089173SA), 00840682109260SEE17300005SA (SEE17300005SA), 00840682109260SEE17300007SA (SEE17300007SA), 00840682109260SEE17300009SA (SEE17300009SA), 00840682109260SEE18190007SA (SEE18190007SA), 00840682109260SEE18190008SA (SEE18190008SA), 00840682109260SEE18190014SA (SEE18190014SA), 00840682109260SEE18190015SA (SEE18190015SA), 00840682109260SEE18190018SA (SEE18190018SA), 00840682109260SEE18190022SA (SEE18190022SA), 00840682109260SEE18190025SA (SEE18190025SA), 00840682109260SEE18190027SA (SEE18190027SA), 00840682109260SEE18190034SA (SEE18190034SA), 00840682109260SEE17100003SA (SEE17100003SA), 00840682109260SEE18080004SA (SEE18080004SA), 00840682109260SEE18100001SA (SEE18100001SA), 00840682109260SEE18100003SA (SEE18100003SA), 00840682109260SEE18100004SA (SEE18100004SA), 00840682109260SEE18100007SA (SEE18100007SA), 00840682109260SEE18100008SA (SEE18100008SA), 00840682109260SEE18100011SA (SEE18100011SA), 00840682109260SEE18100013SA (SEE18100013SA), 00840682109260SEE18100015SA (SEE18100015SA), 00840682109260SEE18100016SA (SEE18100016SA), 00840682109260SEE18100017SA (SEE18100017SA), 00840682109260SEE18100018SA (SEE18100018SA), 00840682109260SEE18100019SA (SEE18100019SA), Not Applicable (SEE11151309GA), Not Applicable (SEE13451313GA), Not Applicable (SEE14021524GA), Not Applicable (SEE14021527GA), Not Applicable (SEE14021528GA), Not Applicable (SEE14021529GA), Not Applicable (SEE14021533GA), Not Applicable (SEE14021534GA), Not Applicable (SEE14021536GA), Not Applicable (SEE14021539GA), Not Applicable (SEE14021540GA), Not Applicable (SEE14021541GA), Not Applicable (SEE14021543GA), Not Applicable (SEE14021546GA), Not Applicable (SEE14021547GA), Not Applicable (SEE14021548GA), Not Applicable (SEE14021550GA), Not Applicable (SEE14021557GA), Not Applicable (SEE14021560GA), Not Applicable (SEE14021565GA), Not Applicable (SEE14021567GA), Not Applicable (SEE14021572GA), Not Applicable (SEE14242089GA), Not Applicable (SEE16430043SA), Not Applicable (SEE16440002SA), Not Applicable (SEE12031932GA), Not Applicable (SEE12031936GA), Not Applicable (SEE12031938GA), Not Applicable (SEE12031940GA), Not Applicable (SEE14031722GA), Not Applicable (SEE15113170GA), Not Applicable (SEE15113181GA), 00840682109260SEE18370007SA (SEE18370007SA), 00840682109260SEE18410032SA (SEE18410032SA), 00840682109260SEE18410033SA (SEE18410033SA), 00840682109260SEE18420001SA (SEE18420001SA), 00840682109260SEE18420002SA (SEE18420002SA), 00840682109260SEE18420003SA (SEE18420003SA), 00840682109260SEE18420005SA (SEE18420005SA), 00840682109260SEE18420019SA (SEE18420019SA), 00840682109260SEE18420046SA (SEE18420046SA), Not Applicable (SEE09040047GA), Not Applicable (SEE09100115GA), Not Applicable (SEE09410582GA), Not Applicable (SEE10120003GA), Not Applicable (SEE10120994GA), Not Applicable (SEE10121001GA), Not Applicable (SEE10130071GA), Not Applicable (SEE14021505GA), Not Applicable (SEE14021513GA), Not Applicable (SEE14021522GA), Not Applicable (SEE14242094GA), Not Applicable (SEE14242107GA), Not Applicable (SEE14242117GA), Not Applicable (SEE14242118GA), Not Applicable (SEE14242119GA), 00840682109260SEE17480006SA (SEE17480006SA), Not Applicable (SEE10450680GA), Not Applicable (SEE15373462SA), Not Applicable (SEE09490756GA), Not Applicable (SEE09490760GA), Not Applicable (SEE09510768GA), Not Applicable (SEE09510782GA), Not Applicable (SEE12490275GA), 00840682109260SEE18160066SA (SEE18160066SA), Not Applicable (SEE12430184GA), Not Applicable (SEE10460731GA), Not Applicable (SEE16119898SA), Not Applicable (SEE10400427GA), 00840682109260SEE11381827GA (SEE11381827GA), Not Applicable (SEE15153187GA), Not Applicable (SEE13431264GA), Not Applicable (SEE11071040GA), 00840682109260SEE15251298SA (SEE15251298SA), 00840682109260SEE15251300SA (SEE15251300SA), 00840682109260SEE15251302SA (SEE15251302SA), 00840682109260SEE15261409SA (SEE15261409SA), 00840682109260SEE15261532SA (SEE15261532SA), 00840682109260SEE15261533SA (SEE15261533SA), Not Applicable (SEE15261534SA), 00840682109260SEE15261540SA (SEE15261540SA), 00840682109260SEE15261541SA (SEE15261541SA), 00840682109260SEE15261542SA (SEE15261542SA), 00840682109260SEE15261545SA (SEE15261545SA), 00840682109260SEE15261546SA (SEE15261546SA), Not Applicable (SEE12041946GA), Not Applicable (SEE14412494GA), Not Applicable (SEE14412500GA), Not Applicable (SEE14412508GA), Not Applicable (SEE14412511GA), Not Applicable (SEE11171377GA), Not Applicable (SEE10530860GA), Not Applicable (SEE10530873GA), Not Applicable (SEE11030882GA), Not Applicable (SEE11030891GA), Not Applicable (SEE12322417GA), Not Applicable (SEE12322424GA), Not Applicable (SEE12322439GA), Not Applicable (SEE10460709GA), Not Applicable (SEE12490246GA), Not Applicable (SEE12182217GA), Not Applicable (SEE14512828GA), Not Applicable (SEE15518043SA), 00840682109260SEE18360001SA (SEE18360001SA), 00840682109260SEE18360009SA (SEE18360009SA), 00840682109260SEE18360011SA (SEE18360011SA), 00840682109260SEE18360013SA (SEE18360013SA), 00840682109260SEE18360014SA (SEE18360014SA), 00840682109260SEE18360017SA (SEE18360017SA), Not Applicable (SEE11030884GA), Not Applicable (SEE11030885GA), Not Applicable (SEE11030889GA), Not Applicable (SEE11030892GA), Not Applicable (SEE11030893GA), Not Applicable (SEE11030894GA), Not Applicable (SEE11030895GA), Not Applicable (SEE12122148GA), Not Applicable (SEE12122150GA), Not Applicable (SEE13381122GA), Not Applicable (SEE13381125GA), Not Applicable (SEE13391134GA), Not Applicable (SEE09460651GA), Not Applicable (SEE12172194GA), Not Applicable (SEE09470656GA), Not Applicable (SEE09470680GA), Not Applicable (SEE09470681GA), Not Applicable (SEE09470682GA), Not Applicable (SEE09470684GA), Not Applicable (SEE09470688GA), Not Applicable (SEE09470689GA), Not Applicable (SEE09470690GA), Not Applicable (SEE09480692GA), Not Applicable (SEE09480693GA), Not Applicable (SEE09480694GA), Not Applicable (SEE09480695GA), Not Applicable (SEE09480721GA), Not Applicable (SEE09480722GA), Not Applicable (SEE09480724GA), Not Applicable (SEE09480725GA), Not Applicable (SEE09480726GA), Not Applicable (SEE09480727GA), Not Applicable (SEE09480731GA), Not Applicable (SEE15332399SA), Not Applicable (SEE15332400SA), 00840682109260SEE17190012SA (SEE17190012SA), 00840682109260SEE17190017SA (SEE17190017SA), 00840682109260SEE18330004SA (SEE18330004SA), 00840682109260SEE18470008SA (SEE18470008SA), 00840682109260SEE18470009SA (SEE18470009SA), 00840682109260SEE18470015SA (SEE18470015SA), 00840682109260SEE18470022SA (SEE18470022SA), Not Applicable (SEE10390318GA), Not Applicable (SEE11211503GA), Not Applicable (SEE11361684GA), Not Applicable (SEE11361690GA), Not Applicable (SEE12262341GA), Not Applicable (SEE13311006GA), Not Applicable (SEE14272162GA), Not Applicable (SEE14272171GA), Not Applicable (SEE14272208GA), Not Applicable (SEE14272209GA), Not Applicable (SEE14272211GA), Not Applicable (SEE14272215GA), Not Applicable (SEE14272216GA), Not Applicable (SEE14272220GA), Not Applicable (SEE14272221GA), Not Applicable (SEE14272222GA), Not Applicable (SEE14272223GA), Not Applicable (SEE14272225GA), Not Applicable (SEE14272226GA), Not Applicable (SEE14272227GA), Not Applicable (SEE14272228GA), Not Applicable (SEE14272229GA), Not Applicable (SEE14472679GA), Not Applicable (SEE15383769SA), 00840682109260SEE19010007SA (SEE19010007SA), 00840682109260SEE19220039SA (SEE19220039SA), Not Applicable (SEE13431270GA), Not Applicable (SEE13441307GA), Not Applicable (SEE14222014GA), Not Applicable (SEE14222021GA), Not Applicable (SEE14222022GA), Not Applicable (SEE14222023GA), Not Applicable (SEE14222024GA), Not Applicable (SEE14222025GA), Not Applicable (SEE14222058GA), Not Applicable (SEE14222066GA), Not Applicable (SEE14222069GA), Not Applicable (SEE14222071GA), 00840682109260SEE17090020SA (SEE17090020SA), Not Applicable (SEE09020020GA), Not Applicable (SEE11341663GA), 00840682109260SEE19100010SA (SEE19100010SA), 00840682109260SEE19100027SA (SEE19100027SA), 00840682109260SEE19100058SA (SEE19100058SA), 00840682109260SEE19190008SA (SEE19190008SA), 00840682109260SEE18420042SA (SEE18420042SA), 00840682109260SEE18430002SA (SEE18430002SA), 00840682109260SEE18430003SA (SEE18430003SA), 00840682109260SEE18430006SA (SEE18430006SA), 00840682109260SEE18430007SA (SEE18430007SA), 00840682109260SEE18430008SA (SEE18430008SA), 00840682109260SEE18430009SA (SEE18430009SA), 00840682109260SEE18450076SA (SEE18450076SA), 00840682109260SEE18400007SA (SEE18400007SA), 00840682109260SEE18420009SA (SEE18420009SA), 00840682109260SEE18420011SA (SEE18420011SA), 00840682109260SEE18420022SA (SEE18420022SA), 00840682109260SEE18420023SA (SEE18420023SA), 00840682109260SEE18420027SA (SEE18420027SA), 00840682109260SEE18420028SA (SEE18420028SA), 00840682109260SEE18420030SA (SEE18420030SA), 00840682109260SEE18420034SA (SEE18420034SA), 00840682109260SEE18420035SA (SEE18420035SA), 00840682109260SEE18420037SA (SEE18420037SA), 00840682109260SEE18420038SA (SEE18420038SA), 00840682109260SEE18420039SA (SEE18420039SA), 00840682109260SEE18420043SA (SEE18420043SA), 00840682109260SEE18420048SA (SEE18420048SA), 00840682109260SEE18420049SA (SEE18420049SA), 00840682109260SEE18420051SA (SEE18420051SA), 00840682109260SEE18420052SA (SEE18420052SA), Not Applicable (SEE13280949GA), Not Applicable (SEE13351044GA), Not Applicable (SEE13351051GA), Not Applicable (SEE13361055GA), Not Applicable (SEE13361056GA), Not Applicable (SEE13361059GA), Not Applicable (SEE13361060GA), Not Applicable (SEE13361063GA), Not Applicable (SEE13361064GA), Not Applicable (SEE13361068GA), Not Applicable (SEE13361071GA), Not Applicable (SEE13361072GA), 00840682109260SEE17110004SA (SEE17110004SA), 00840682109260SEE17180001SA (SEE17180001SA), Not Applicable (SEE12122135GA), Not Applicable (SEE12122141GA), Not Applicable (SEE12122159GA), Not Applicable (SEE12122161GA), Not Applicable (SEE12122166GA), 00840682109260SEE17250023SA (SEE17250023SA), 00840682109260SEE17250025SA (SEE17250025SA), Not Applicable (SEE15456712SA), Not Applicable (SEE15466922SA), Not Applicable (SEE15466927SA), 00840682109260SEE17210017SA (SEE17210017SA), Not Applicable (SEE09130159GA), Not Applicable (SEE09260338GA), Not Applicable (SEE09260339GA), Not Applicable (SEE10400424GA), Not Applicable (SEE10400433GA), Not Applicable (SEE10400452GA), Not Applicable (SEE10400456GA), Not Applicable (SEE10460773GA), Not Applicable (SEE10470793GA), Not Applicable (SEE14502795GA), Not Applicable (SEE10390307GA), Not Applicable (SEE10390317GA), Not Applicable (SEE10390320GA), Not Applicable (SEE10390330GA), Not Applicable (SEE10390331GA), Not Applicable (SEE10390342GA), Not Applicable (SEE16180907SA), Not Applicable (SEE16180908SA), Not Applicable (SEE16180911SA), Not Applicable (SEE16180912SA), Not Applicable (SEE16180913SA), Not Applicable (SEE14031586GA), Not Applicable (SEE14031590GA), Not Applicable (SEE14292231GA), Not Applicable (SEE14292232GA), Not Applicable (SEE14292233GA), Not Applicable (SEE14292239GA), Not Applicable (SEE14292240GA), Not Applicable (SEE14292244GA), Not Applicable (SEE14302257GA), Not Applicable (SEE14302258GA), Not Applicable (SEE15052928GA), Not Applicable (SEE15052930GA), Not Applicable (SEE15052938GA), Not Applicable (SEE15052948GA), Not Applicable (SEE15052949GA), Not Applicable (SEE15052951GA), Not Applicable (SEE15052952GA), Not Applicable (SEE15062963GA), Not Applicable (SEE15062966GA), Not Applicable (SEE15062967GA), Not Applicable (SEE15062968GA), Not Applicable (SEE15062969GA), Not Applicable (SEE13110500GA), 00840682109260SEE18170023SA (SEE18170023SA), 00840682109260SEE19220023SA (SEE19220023SA), Not Applicable (SEE10270260GA), Not Applicable (SEE09350458GA), Not Applicable (SEE10250089GA), Not Applicable (SEE10260202GA), Not Applicable (SEE10260203GA), Not Applicable (SEE10260214GA), Not Applicable (SEE10260215GA), Not Applicable (SEE10260220GA), Not Applicable (SEE10260223GA), Not Applicable (SEE10260225GA), Not Applicable (SEE10260226GA), Not Applicable (SEE10260227GA), Not Applicable (SEE10260228GA), Not Applicable (SEE10260231GA), Not Applicable (SEE16282654SA), Not Applicable (SEE16344281SA), Not Applicable (SEE16344328SA), Not Applicable (SEE16354344SA), Not Applicable (SEE16364411SA), Not Applicable (SEE16364412SA), Not Applicable (SEE16364413SA), Not Applicable (SEE16364414SA), Not Applicable (SEE16364415SA), Not Applicable (SEE16420064SA), Not Applicable (SEE11341665GA), Not Applicable (SEE13401175GA), Not Applicable (SEE16221589SA), Not Applicable (SEE15467028SA), Not Applicable (SEE09020015GA), Not Applicable (SEE09040046GA), Not Applicable (SEE16400001SA), Not Applicable (SEE16410005SA), Not Applicable (SEE16410006SA), Not Applicable (SEE16410012SA), Not Applicable (SEE15240989SA), 00840682109260SEE18160045SA (SEE18160045SA), Not Applicable (SEE14302267GA), Not Applicable (SEE14302276GA), 00840682109260SEE19170026SA (SEE19170026SA), Not Applicable (SEE16120020SA), Not Applicable (SEE12350026GA), Not Applicable (SEE10250124GA), Not Applicable (SEE16241829SA), Not Applicable (SEE11231529GA), Not Applicable (SEE14222033GA), Not Applicable (SEE14222038GA), Not Applicable (SEE14222048GA), Not Applicable (SEE14222052GA), Not Applicable (SEE14222053GA), Not Applicable (SEE14222070GA), 00840682109260SEE18250012SA (SEE18250012SA), 00840682109260SEE18250013SA (SEE18250013SA), 00840682109260SEE18250014SA (SEE18250014SA), Not Applicable (SEE09340432GA), Not Applicable (SEE09340438GA), Not Applicable (SEE09340442GA), Not Applicable (SEE09340443GA), Not Applicable (SEE14312294GA), Not Applicable (SEE15093063GA), Not Applicable (SEE15093064GA), Not Applicable (SEE15093071GA), Not Applicable (SEE15093072GA), Not Applicable (SEE12092068GA), Not Applicable (SEE12092069GA), Not Applicable (SEE12092073GA), Not Applicable (SEE12092075GA), Not Applicable (SEE12092081GA), Not Applicable (SEE12092084GA), Not Applicable (SEE10270275GA), Not Applicable (SEE11040907GA), Not Applicable (SEE14322321GA), Not Applicable (SEE14322323GA), Not Applicable (SEE14322329GA), Not Applicable (SEE14332344GA), Not Applicable (SEE14362348GA), Not Applicable (SEE14362349GA), Not Applicable (SEE14362354GA), Not Applicable (SEE14362355GA), Not Applicable (SEE14362356GA), Not Applicable (SEE14462654GA), Not Applicable (SEE14462655GA), Not Applicable (SEE15193276GA), Not Applicable (SEE16380022SA), Not Applicable (SEE14332338GA), Not Applicable (SEE14332340GA), Not Applicable (SEE15183257GA), 00840682109260SEE18160063SA (SEE18160063SA), Not Applicable (SEE13361057GA), Not Applicable (SEE13361061GA), Not Applicable (SEE12262344GA), Not Applicable (SEE09100123GA), Not Applicable (SEE09250333GA), Not Applicable (SEE10260229GA), Not Applicable (SEE10270249GA), Not Applicable (SEE10270252GA), Not Applicable (SEE10310278GA), Not Applicable (SEE10310284GA), Not Applicable (SEE13361080GA), Not Applicable (SEE14482719GA), Not Applicable (SEE14512835GA), Not Applicable (SEE14512836GA), Not Applicable (SEE15120556SA), Not Applicable (SEE15120558SA), Not Applicable (SEE15210699SA), Not Applicable (SEE15210702SA), Not Applicable (SEE15210705SA), Not Applicable (SEE09330422GA), Not Applicable (SEE11201469GA), Not Applicable (SEE13010403GA), Not Applicable (SEE14021520GA), Not Applicable (SEE14021523GA), Not Applicable (SEE14021525GA), Not Applicable (SEE14021526GA), Not Applicable (SEE14021532GA), Not Applicable (SEE14021535GA), Not Applicable (SEE14021552GA), Not Applicable (SEE14021577GA), Not Applicable (SEE14021582GA), Not Applicable (SEE14031690GA), Not Applicable (SEE15113125GA), Not Applicable (SEE15113129GA), Not Applicable (SEE15113135GA), Not Applicable (SEE16078864SA), Not Applicable (SEE16078865SA), Not Applicable (SEE12122142GA), Not Applicable (SEE12122167GA), Not Applicable (SEE14362384GA), Not Applicable (SEE11301560GA), Not Applicable (SEE11371790GA), Not Applicable (SEE09250306GA), Not Applicable (SEE14452633GA), Not Applicable (SEE15052914GA), Not Applicable (SEE14372389GA), Not Applicable (SEE11071011GA), Not Applicable (SEE11071047GA), Not Applicable (SEE13130616GA), Not Applicable (SEE13130623GA), Not Applicable (SEE13130630GA), Not Applicable (SEE10260219GA), 00840682109260SEE19100039SA (SEE19100039SA), 00840682109260SEE19100040SA (SEE19100040SA), 00840682109260SEE17480004SA (SEE17480004SA), 00840682109260SEE17480005SA (SEE17480005SA), 00840682109260SEE17480007SA (SEE17480007SA), 00840682109260SEE17480009SA (SEE17480009SA), 00840682109260SEE17480013SA (SEE17480013SA), Not Applicable (SEE16272597SA), Not Applicable (SEE09370495GA), Not Applicable (SEE09390535GA), Not Applicable (SEE09460641GA), Not Applicable (SEE11111248GA), Not Applicable (SEE13130649GA), Not Applicable (SEE15193286GA), Not Applicable (SEE15193287GA), Not Applicable (SEE15193288GA), Not Applicable (SEE15193289GA), Not Applicable (SEE15193290GA), Not Applicable (SEE15193291GA), Not Applicable (SEE15193292GA), Not Applicable (SEE15193295GA), Not Applicable (SEE15193305GA), Not Applicable (SEE15394426SA), Not Applicable (SEE15518046SA), Not Applicable (SEE15518047SA), Not Applicable (SEE15528072SA), Not Applicable (SEE15528204SA), Not Applicable (SEE15528206SA), Not Applicable (SEE15528207SA), Not Applicable (SEE11341636GA), 00840682109260SEE19220007SA (SEE19220007SA), Not Applicable (SEE10100969GA), Not Applicable (SEE10100979GA), Not Applicable (SEE10120988GA), Not Applicable (SEE13130636GA), Not Applicable (SEE10270253GA), Not Applicable (SEE13150694GA), Not Applicable (SEE13150697GA), Not Applicable (SEE13220822GA), Not Applicable (SEE13220837GA), Not Applicable (SEE14041737GA), 00840682109260SEE18240059SA (SEE18240059SA), Not Applicable (SEE09390539GA), Not Applicable (SEE09390541GA), Not Applicable (SEE09390543GA), Not Applicable (SEE09390545GA), Not Applicable (SEE09400551GA), 00840682109260SEE17320001SA (SEE17320001SA), 00840682109260SEE17320002SA (SEE17320002SA), 00840682109260SEE17320003SA (SEE17320003SA), 00840682109260SEE17320004SA (SEE17320004SA), 00840682109260SEE17320005SA (SEE17320005SA), 00840682109260SEE17320006SA (SEE17320006SA), Not Applicable (SEE14041740GA), Not Applicable (SEE09410583GA), 00840682109260SEE17090026SA (SEE17090026SA), 00840682109260SEE17090027SA (SEE17090027SA), 00840682109260SEE17360007SA (SEE17360007SA), 00840682109260SEE17460010SA (SEE17460010SA), 00840682109260SEE18160065SA (SEE18160065SA), 00840682109260SEE19240028SA (SEE19240028SA), 00840682109260SEE18190029SA (SEE18190029SA), 00840682109260SEE17020022SA (SEE17020022SA), 00840682109260SEE17400005SA (SEE17400005SA), Not Applicable (SEE14242087GA), Not Applicable (SEE09510828GA), Not Applicable (SEE11071152GA), Not Applicable (SEE11081157GA), Not Applicable (SEE11081164GA), Not Applicable (SEE16430034SA), Not Applicable (SEE09150218GA), Not Applicable (SEE10450667GA), Not Applicable (SEE10460711GA), Not Applicable (SEE11071029GA), Not Applicable (SEE12092083GA), Not Applicable (SEE09390536GA), Not Applicable (SEE10250121GA), Not Applicable (SEE10250139GA), Not Applicable (SEE10250169GA), Not Applicable (SEE10250172GA), Not Applicable (SEE15415177SA), Not Applicable (SEE15425781SA), Not Applicable (SEE15425791SA), 00840682109260SEE17240098SA (SEE17240098SA), 00840682109260SEE17240120SA (SEE17240120SA), 00840682109260SEE17240122SA (SEE17240122SA), 00840682109260SEE17500015SA (SEE17500015SA), 00840682109260SEE17500017SA (SEE17500017SA), 00840682109260SEE17500018SA (SEE17500018SA), 00840682109260SEE18100038SA (SEE18100038SA), 0 0840682109260SEE18100039SA (SEE18100039SA), 00840682109260SEE18170009SA (SEE18170009SA), 00840682109260SEE18170013SA (SEE18170013SA), 00840682109260SEE19170001SA (SEE19170001SA), 00840682109260SEE19170003SA (SEE19170003SA), 00840682109260SEE19170010SA (SEE19170010SA), 00840682109260SEE19170016SA (SEE19170016SA), 00840682109260SEE19170021SA (SEE19170021SA), 00840682109260SEE19170023SA (SEE19170023SA), Not Applicable (SEE11181399GA), Not Applicable (SEE11181414GA), Not Applicable (SEE12340018GA), 00840682109260SEE18270001SA (SEE18270001SA), Not Applicable (SEE09360490GA), Not Applicable (SEE12410133GA), 00840682109260SEE18260002SA (SEE18260002SA), 00840682109260SEE18260003SA (SEE18260003SA), 00840682109260SEE18260004SA (SEE18260004SA), 00840682109260SEE18260005SA (SEE18260005SA), 00840682109260SEE18260006SA (SEE18260006SA), 00840682109260SEE18260008SA (SEE18260008SA), 00840682109260SEE18260009SA (SEE18260009SA), 00840682109260SEE18260010SA (SEE18260010SA), 00840682109260SEE18260011SA (SEE18260011SA), 00840682109260SEE18260012SA (SEE18260012SA), Not Applicable (SEE11081198GA), 00840682109260SEE17130011SA (SEE17130011SA), 00840682109260SEE17130013SA (SEE17130013SA), 00840682109260SEE17210015SA (SEE17210015SA), 00840682109260SEE17210019SA (SEE17210019SA), 00840682109260SEE17260001SA (SEE17260001SA), 00840682109260SEE17380016SA (SEE17380016SA), 00840682109260SEE18100027SA (SEE18100027SA), 00840682109260SEE18100028SA (SEE18100028SA), 00840682109260SEE18280007SA (SEE18280007SA), 00840682109260SEE18290003SA (SEE18290003SA), 00840682109260SEE18350001SA (SEE18350001SA), 00840682109260SEE18350003SA (SEE18350003SA), 00840682109260SEE17180013SA (SEE17180013SA), 00840682109260SEE17180015SA (SEE17180015SA), 00840682109260SEE17180017SA (SEE17180017SA), 00840682109260SEE17200002SA (SEE17200002SA), 00840682109260SEE17410003SA (SEE17410003SA), 00840682109260SEE17460001SA (SEE17460001SA), 00840682109260SEE17460012SA (SEE17460012SA), Not Applicable (SEE09310365GA), Not Applicable (SEE09310374GA), Not Applicable (SEE09310381GA), Not Applicable (SEE09340448GA), Not Applicable (SEE09340454GA), Not Applicable (SEE09360472GA), Not Applicable (SEE09360487GA), Not Applicable (SEE09360489GA), Not Applicable (SEE09410581GA), Not Applicable (SEE09410590GA), Not Applicable (SEE09430608GA), Not Applicable (SEE09430615GA), Not Applicable (SEE09430617GA), Not Applicable (SEE09430619GA), Not Applicable (SEE09510818GA), Not Applicable (SEE09530843GA), Not Applicable (SEE13421227GA), Not Applicable (SEE13421228GA), Not Applicable (SEE13421231GA), Not Applicable (SEE13421233GA), Not Applicable (SEE13421234GA), Not Applicable (SEE13421237GA), Not Applicable (SEE13421242GA), Not Applicable (SEE13421244GA), Not Applicable (SEE12410131GA), Not Applicable (SEE09020032GA), Not Applicable (SEE09040048GA), Not Applicable (SEE09040050GA), Not Applicable (SEE09420595GA), Not Applicable (SEE10390310GA), Not Applicable (SEE15383635SA), Not Applicable (SEE15383636SA), Not Applicable (SEE15383637SA), Not Applicable (SEE15517822SA), 00840682109260SEE18030012SA (SEE18030012SA), 00840682109260SEE18040014SA (SEE18040014SA), 00840682109260SEE18040025SA (SEE18040025SA), 00840682109260SEE18500009SA (SEE18500009SA), 00840682109260SEE18510008SA (SEE18510008SA), 00840682109260SEE18510011SA (SEE18510011SA), 00840682109260SEE17440004SA (SEE17440004SA), 00840682109260SEE17440006SA (SEE17440006SA), 00840682109260SEE17440008SA (SEE17440008SA), Not Applicable (SEE10120028GA), Not Applicable (SEE10120033GA), Not Applicable (SEE10130036GA), Not Applicable (SEE10400384GA), Not Applicable (SEE10400437GA), Not Applicable (SEE10400453GA), Not Applicable (SEE10460725GA), Not Applicable (SEE12410146GA), Not Applicable (SEE15113157GA), Not Applicable (SEE15467027SA), Not Applicable (SEE15517777SA), 00840682109260SEE17020024SA (SEE17020024SA), 00840682109260SEE17020027SA (SEE17020027SA), Not Applicable (SEE12112111GA), Not Applicable (SEE14272160GA), Not Applicable (SEE15193303GA), 00840682109260SEE17460003SA (SEE17460003SA), 00840682109260SEE18100009SA (SEE18100009SA), 00840682109260SEE18100010SA (SEE18100010SA), 00840682109260SEE19130018SA (SEE19130018SA), 00840682109260SEE19150006SA (SEE19150006SA), Not Applicable (SEE11101215GA), 00840682109260SEE18060017SA (SEE18060017SA), Not Applicable (SEE10530858GA), Not Applicable (SEE10530868GA), Not Applicable (SEE14021551GA), Not Applicable (SEE12122147GA), Not Applicable (SEE12122154GA), Not Applicable (SEE12122156GA), Not Applicable (SEE15012858GA), 00840682109260SEE17250022SA (SEE17250022SA), 00840682109260SEE17250024SA (SEE17250024SA), 00840682109260SEE17390009SA (SEE17390009SA), Not Applicable (SEE11381819GA), Not Applicable (SEE12360032GA), Not Applicable (SEE09100102GA), Not Applicable (SEE13110508GA), Not Applicable (SEE15052916GA), Not Applicable (SEE16130324SA), 00840682109260SEE17480016SA (SEE17480016SA), 00840682109260SEE19100022SA (SEE19100022SA), Not Applicable (SEE12312400GA), Not Applicable (SEE09190231GA), Not Applicable (SEE09190241GA), Not Applicable (SEE09510767GA), Not Applicable (SEE09510770GA), Not Applicable (SEE09510780GA), Not Applicable (SEE11201449GA), Not Applicable (SEE12232252GA), Not Applicable (SEE12232254GA), Not Applicable (SEE12232255GA), Not Applicable (SEE12232256GA), Not Applicable (SEE12232262GA), Not Applicable (SEE12242275GA), Not Applicable (SEE12242277GA), Not Applicable (SEE12242278GA), Not Applicable (SEE12242280GA), Not Applicable (SEE12242283GA), Not Applicable (SEE12242285GA), Not Applicable (SEE12252296GA), Not Applicable (SEE12252298GA), Not Applicable (SEE12252299GA), Not Applicable (SEE12252300GA), Not Applicable (SEE12252302GA), Not Applicable (SEE12252303GA), Not Applicable (SEE12252305GA), Not Applicable (SEE12252308GA), Not Applicable (SEE12252309GA), Not Applicable (SEE12252310GA), Not Applicable (SEE12252312GA), Not Applicable (SEE12252313GA), Not Applicable (SEE12252314GA), Not Applicable (SEE12252315GA), Not Applicable (SEE12322418GA), Not Applicable (SEE13491447GA), Not Applicable (SEE13491448GA), Not Applicable (SEE15113124GA), Not Applicable (SEE15113150GA), Not Applicable (SEE16420061SA), Not Applicable (SEE16420063SA), Not Applicable (SEE16420066SA), Not Applicable (SEE16420067SA), Not Applicable (SEE16420068SA), Not Applicable (SEE16420070SA), Not Applicable (SEE16420073SA), Not Applicable (SEE16420076SA), Not Applicable (SEE16420078SA), Not Applicable (SEE16430042SA), 00840682109260SEE17150009SA (SEE17150009SA), 00840682109260SEE18240013SA (SEE18240013SA), 00840682109260SEE18240034SA (SEE18240034SA), 00840682109260SEE18240035SA (SEE18240035SA), 00840682109260SEE18240039SA (SEE18240039SA), 00840682109260SEE18240040SA (SEE18240040SA), 00840682109260SEE18240041SA (SEE18240041SA), 00840682109260SEE18240043SA (SEE18240043SA), 00840682109260SEE18240046SA (SEE18240046SA), 00840682109260SEE18240049SA (SEE18240049SA), 00840682109260SEE18240051SA (SEE18240051SA), Not Applicable (SEE14432594GA), Not Applicable (SEE13280935GA), Not Applicable (SEE13280943GA), Not Applicable (SEE13290954GA), Not Applicable (SEE13290956GA), Not Applicable (SEE13290957GA), Not Applicable (SEE13290959GA), Not Applicable (SEE13290974GA), Not Applicable (SEE15261400SA), Not Applicable (SEE12490270GA), Not Applicable (SEE12490279GA), Not Applicable (SEE16252014SA), Not Applicable (SEE16282652SA), Not Applicable (SEE16390002SA), Not Applicable (SEE16390012SA), Not Applicable (SEE14452640GA), Not Applicable (SEE14452642GA), Not Applicable (SEE12092051GA), Not Applicable (SEE13040426GA), Not Applicable (SEE16028376SA), 00840682109260SEE17330013SA (SEE17330013SA), 00840682109260SEE17070005SA (SEE17070005SA), 00840682109260SEE17070008SA (SEE17070008SA), 00840682109260SEE17070009SA (SEE17070009SA), 00840682109260SEE17070011SA (SEE17070011SA), Not Applicable (SEE10400372GA), Not Applicable (SEE10410467GA), Not Applicable (SEE10430650GA), Not Applicable (SEE10450663GA), Not Applicable (SEE11141287GA), Not Applicable (SEE11371809GA), Not Applicable (SEE12122165GA), Not Applicable (SEE13371092GA), Not Applicable (SEE14021566GA), Not Applicable (SEE14372411GA), Not Applicable (SEE14382466GA), Not Applicable (SEE14382472GA), Not Applicable (SEE15173211GA), Not Applicable (SEE15291869SA), Not Applicable (SEE15342483SA), Not Applicable (SEE16380021SA), 00840682109260SEE17190018SA (SEE17190018SA), 00840682109260SEE17460008SA (SEE17460008SA), 00840682109260SEE17460009SA (SEE17460009SA), 00840682109260SEE19170027SA (SEE19170027SA), Not Applicable (SEE10090955GA), Not Applicable (SEE10460759GA), Not Applicable (SEE10090946GA), Not Applicable (SEE15032892GA), Not Applicable (SEE11361707GA), Not Applicable (SEE13381129GA), Not Applicable (SEE13391131GA), Not Applicable (SEE14462662GA), Not Applicable (SEE14462663GA), Not Applicable (SEE16364408SA), Not Applicable (SEE16364409SA), Not Applicable (SEE10080931GA), Not Applicable (SEE10090951GA), Not Applicable (SEE16252022SA), Not Applicable (SEE16262133SA), Not Applicable (SEE16262136SA), Not Applicable (SEE16262147SA), Not Applicable (SEE16262148SA), Not Applicable (SEE16262150SA), Not Applicable (SEE16262163SA), Not Applicable (SEE16262165SA), Not Applicable (SEE16262180SA), Not Applicable (SEE16262204SA), Not Applicable (SEE16272331SA), Not Applicable (SEE16272419SA), Not Applicable (SEE16272585SA), Not Applicable (SEE16272586SA), Not Applicable (SEE16272587SA), Not Applicable (SEE16272599SA), Not Applicable (SEE16272600SA), Not Applicable (SEE16272618SA), Not Applicable (SEE16272620SA), 00840682109260SEE17190005SA (SEE17190005SA), 00840682109260SEE17190006SA (SEE17190006SA), 00840682109260SEE17190007SA (SEE17190007SA), 00840682109260SEE17190013SA (SEE17190013SA), 00840682109260SEE17190014SA (SEE17190014SA), 00840682109260SEE17190015SA (SEE17190015SA), 00840682109260SEE17190016SA (SEE17190016SA), Not Applicable (SEE15113154GA), Not Applicable (SEE14432596GA), Not Applicable (SEE14432597GA), 00840682109260SEE19120012SA (SEE19120012SA), Not Applicable (SEE13110528GA), Not Applicable (SEE13110550GA), Not Applicable (SEE13110554GA), Not Applicable (SEE13110556GA), Not Applicable (SEE13110561GA), Not Applicable (SEE13110563GA), 00840682109260SEE19100025SA (SEE19100025SA), 00840682109260SEE19100065SA (SEE19100065SA), Not Applicable (SEE13431274GA), Not Applicable (SEE13451312GA), Not Applicable (SEE13451320GA), Not Applicable (SEE13451331GA), Not Applicable (SEE11391837GA), Not Applicable (SEE12021909GA), Not Applicable (SEE12031926GA), Not Applicable (SEE12031930GA), Not Applicable (SEE12031935GA), Not Applicable (SEE12061997GA), Not Applicable (SEE13491424GA), Not Applicable (SEE14021553GA), Not Applicable (SEE14021564GA), 00840682109260SEE17220006SA (SEE17220006SA), 00840682109260SEE17380001SA (SEE17380001SA), 00840682109260SEE17380002SA (SEE17380002SA), 00840682109260SEE17380013SA (SEE17380013SA), Not Applicable (SEE09470668GA), Not Applicable (SEE09470672GA), Not Applicable (SEE10260195GA), Not Applicable (SEE11071032GA), Not Applicable (SEE11071060GA), Not Applicable (SEE11321562GA), Not Applicable (SEE11321563GA), Not Applicable (SEE11321568GA), Not Applicable (SEE11321569GA), Not Applicable (SEE11321571GA), Not Applicable (SEE11321572GA), Not Applicable (SEE11321584GA), Not Applicable (SEE11321585GA), Not Applicable (SEE11331601GA), Not Applicable (SEE13280907GA), Not Applicable (SEE13280908GA), Not Applicable (SEE13280909GA), Not Applicable (SEE13280910GA), Not Applicable (SEE13280911GA), Not Applicable (SEE13280914GA), Not Applicable (SEE13280915GA), Not Applicable (SEE13280916GA), Not Applicable (SEE13280920GA), Not Applicable (SEE13501460GA), Not Applicable (SEE13501463GA), Not Applicable (SEE13501475GA), Not Applicable (SEE13501479GA), Not Applicable (SEE16150531SA), Not Applicable (SEE16180899SA), Not Applicable (SEE16324003SA), Not Applicable (SEE16324005SA), Not Applicable (SEE10420548GA), Not Applicable (SEE10450657GA), Not Applicable (SEE10460710GA), Not Applicable (SEE11151304GA), 00840682109260SEE17490036SA (SEE17490036SA), Not Applicable (SEE11181395GA), Not Applicable (SEE11191435GA), Not Applicable (SEE14211982GA), Not Applicable (SEE14211986GA), Not Applicable (SEE15507710SA), Not Applicable (SEE15507711SA), Not Applicable (SEE16079148SA), 00840682109260SEE18070004SA (SEE18070004SA), 00840682109260SEE18080008SA (SEE18080008SA), 00840682109260SEE18080011SA (SEE18080011SA), 00840682109260SEE18080014SA (SEE18080014SA), 00840682109260SEE18510004SA (SEE18510004SA), Not Applicable (SEE13000385GA), Not Applicable (SEE12430191GA), Not Applicable (SEE12430192GA), Not Applicable (SEE12430221GA), Not Applicable (SEE12430222GA), Not Applicable (SEE12430226GA), Not Applicable (SEE12430228GA), Not Applicable (SEE12440236GA), Not Applicable (SEE12440240GA), 00840682109260SEE19110002SA (SEE19110002SA), 00840682109260SEE19110005SA (SEE19110005SA), 00840682109260SEE19110008SA (SEE19110008SA), 00840682109260SEE19110009SA (SEE19110009SA), 00840682109260SEE19110011SA (SEE19110011SA), 00840682109260SEE19110016SA (SEE19110016SA), 00840682109260SEE19110017SA (SEE19110017SA), 00840682109260SEE19110018SA (SEE19110018SA), 00840682109260SEE19110027SA (SEE19110027SA), 00840682109260SEE19110030SA (SEE19110030SA), 00840682109260SEE19110031SA (SEE19110031SA), 00840682109260SEE19110032SA (SEE19110032SA), 00840682109260SEE19110033SA (SEE19110033SA), 00840682109260SEE19110034SA (SEE19110034SA), 00840682109260SEE19110035SA (SEE19110035SA), 00840682109260SEE19110036SA (SEE19110036SA), 00840682109260SEE19110038SA (SEE19110038SA), 00840682109260SEE19110042SA (SEE19110042SA), 00840682109260SEE19120007SA (SEE19120007SA), 00840682109260SEE19120008SA (SEE19120008SA), 00840682109260SEE19120009SA (SEE19120009SA), 00840682109260SEE19120010SA (SEE19120010SA), Not Applicable (SEE10420567GA), Not Applicable (SEE10420590GA), Not Applicable (SEE10420596GA), Not Applicable (SEE10450669GA), Not Applicable (SEE11161340GA), Not Applicable (SEE11161342GA), Not Applicable (SEE11161347GA), Not Applicable (SEE11171386GA), Not Applicable (SEE11171387GA), Not Applicable (SEE11191440GA), Not Applicable (SEE16334208SA), Not Applicable (SEE12400119GA), Not Applicable (SEE12410137GA), Not Applicable (SEE12410139GA), Not Applicable (SEE16390017SA), Not Applicable (SEE14432583GA), Not Applicable (SEE14432586GA), Not Applicable (SEE14432589GA), 00840682109260SEE19170008SA (SEE19170008SA), Not Applicable (SEE14041747GA), Not Applicable (SEE14041748GA), Not Applicable (SEE14041751GA), Not Applicable (SEE14041757GA), Not Applicable (SEE14041758GA), Not Applicable (SEE14041760GA), Not Applicable (SEE14041761GA), Not Applicable (SEE14041763GA), Not Applicable (SEE14041765GA), Not Applicable (SEE15312127SA), Not Applicable (SEE15301929SA), Not Applicable (SEE15301930SA), Not Applicable (SEE15301931SA), Not Applicable (SEE15301932SA), Not Applicable (SEE15301933SA), Not Applicable (SEE15301948SA), Not Applicable (SEE14151836GA), Not Applicable (SEE14151838GA), Not Applicable (SEE14151844GA), Not Applicable (SEE14151860GA), Not Applicable (SEE14151861GA), Not Applicable (SEE14222010GA), Not Applicable (SEE13270879GA), Not Applicable (SEE14031677GA), Not Applicable (SEE14151842GA), Not Applicable (SEE14151849GA), Not Applicable (SEE15062979GA), Not Applicable (SEE15153196GA), Not Applicable (SEE16420080SA), Not Applicable (SEE16430010SA), Not Applicable (SEE16430011SA), Not Applicable (SEE16430041SA), Not Applicable (SEE16430044SA), Not Applicable (SEE16430046SA), 00840682109260SEE17060001SA (SEE17060001SA), 00840682109260SEE17060002SA (SEE17060002SA), 00840682109260SEE17060004SA (SEE17060004SA), 00840682109260SEE17380003SA (SEE17380003SA), 00840682109260SEE17380005SA (SEE17380005SA), 00840682109260SEE17380007SA (SEE17380007SA), 00840682109260SEE17380009SA (SEE17380009SA), 00840682109260SEE17380011SA (SEE17380011SA), 00840682109260SEE17380014SA (SEE17380014SA), 00840682109260SEE17380015SA (SEE17380015SA), 00840682109260SEE17380017SA (SEE17380017SA), 00840682109260SEE18190035SA (SEE18190035SA), 00840682109260SEE18450005SA (SEE18450005SA), 00840682109260SEE18450006SA (SEE18450006SA), 00840682109260SEE18450008SA (SEE18450008SA), 00840682109260SEE19100001SA (SEE19100001SA), 00840682109260SEE19100028SA (SEE19100028SA), Not Applicable (SEE09040078GA), Not Applicable (SEE11111237GA), Not Applicable (SEE16334082SA), Not Applicable (SEE16334083SA), Not Applicable (SEE16334084SA), Not Applicable (SEE16334088SA), Not Applicable (SEE16334159SA), Not Applicable (SEE16334160SA), Not Applicable (SEE16334161SA), Not Applicable (SEE16334163SA), Not Applicable (SEE16334164SA), Not Applicable (SEE16334165SA), Not Applicable (SEE16334166SA), Not Applicable (SEE16334167SA), Not Applicable (SEE16334168SA), Not Applicable (SEE16334170SA), Not Applicable (SEE16334171SA), Not Applicable (SEE16334172SA), Not Applicable (SEE16334173SA), Not Applicable (SEE16334174SA), Not Applicable (SEE16334175SA), Not Applicable (SEE16334176SA), Not Applicable (SEE16334177SA), Not Applicable (SEE16334178SA), Not Applicable (SEE16334179SA), Not Applicable (SEE16334180SA), Not Applicable (SEE16334184SA), Not Applicable (SEE16334186SA), Not Applicable (SEE16334187SA), Not Applicable (SEE16334188SA), Not Applicable (SEE16334199SA), Not Applicable (SEE16334200SA), Not Applicable (SEE16334201SA), Not Applicable (SEE16334202SA), Not Applicable (SEE16334203SA), Not Applicable (SEE16334204SA), Not Applicable (SEE16334205SA), Not Applicable (SEE16334206SA), Not Applicable (SEE16334207SA), Not Applicable (SEE16334209SA), Not Applicable (SEE16334210SA), Not Applicable (SEE11171367GA), Not Applicable (SEE10250118GA), Not Applicable (SEE10250184GA), Not Applicable (SEE10250187GA), Not Applicable (SEE13290962GA), Not Applicable (SEE10460733GA), Not Applicable (SEE10460737GA), Not Applicable (SEE10460755GA), Not Applicable (SEE10460760GA), Not Applicable (SEE11050931GA), Not Applicable (SEE10420586GA), Not Applicable (SEE10430616GA), Not Applicable (SEE10460708GA), Not Applicable (SEE10460723GA), Not Applicable (SEE16383412GA), Not Applicable (SEE13110498GA), Not Applicable (SEE13110504GA), Not Applicable (SEE13110507GA), Not Applicable (SEE13120588GA), Not Applicable (SEE11050922GA), Not Applicable (SEE13190803GA), Not Applicable (SEE13190808GA), Not Applicable (SEE13190809GA), Not Applicable (SEE14021531GA), Not Applicable (SEE14432580GA), Not Applicable (SEE14432584GA), Not Applicable (SEE14432585GA), Not Applicable (SEE14432592GA), Not Applicable (SEE16241816SA), 00840682109260SEE18510023SA (SEE18510023SA), Not Applicable (SEE15183247GA), Not Applicable (SEE15183249GA), 00840682109260SEE18220001SA (SEE18220001SA), 00840682109260SEE18220002SA (SEE18220002SA), 00840682109260SEE18220003SA (SEE18220003SA), 00840682109260SEE18220004SA (SEE18220004SA), 00840682109260SEE18220006SA (SEE18220006SA), 00840682109260SEE18220007SA (SEE18220007SA), 00840682109260SEE18220011SA (SEE18220011SA), 00840682109260SEE18220012SA (SEE18220012SA), 00840682109260SEE18220013SA (SEE18220013SA), Not Applicable (SEE14422547GA), Not Applicable (SEE14422557GA), Not Applicable (SEE14422562GA), Not Applicable (SEE14422565GA), Not Applicable (SEE14292243GA), Not Applicable (SEE16390021SA), Not Applicable (SEE16420062SA), Not Applicable (SEE16420069SA), Not Applicable (SEE16420071SA), Not Applicable (SEE16420079SA), Not Applicable (SEE16430012SA), Not Applicable (SEE16430020SA), Not Applicable (SEE16430040SA), Not Applicable (SEE16430047SA), Not Applicable (SEE10390324GA), Not Applicable (SEE10390347GA), Not Applicable (SEE11081158GA), Not Applicable (SEE11081166GA), Not Applicable (SEE11081181GA), Not Applicable (SEE12082020GA), Not Applicable (SEE12082022GA), Not Applicable (SEE13431257GA), Not Applicable (SEE13431262GA), Not Applicable (SEE13431279GA), Not Applicable (SEE13431286GA), Not Applicable (SEE13431292GA), Not Applicable (SEE13401156GA), Not Applicable (SEE09100111GA), Not Applicable (SEE13491426GA), Not Applicable (SEE13491427GA), Not Applicable (SEE13491431GA), Not Applicable (SEE13491432GA), Not Applicable (SEE15405088SA), Not Applicable (SEE11071049GA), Not Applicable (SEE13130639GA), Not Applicable (SEE13130640GA), Not Applicable (SEE13130642GA), Not Applicable (SEE13150689GA), Not Applicable (SEE13150695GA), Not Applicable (SEE10250159GA), Not Applicable (SEE10260197GA), Not Applicable (SEE12410135GA), Not Applicable (SEE12410140GA), Not Applicable (SEE12490268GA), Not Applicable (SEE15517889SA), 00840682109260SEE19120017SA (SEE19120017SA), 00840682109260SEE19120021SA (SEE19120021SA), 00840682109260SEE19120030SA (SEE19120030SA), 00840682109260SEE19170025SA (SEE19170025SA), 00840682109260SEE19170035SA (SEE19170035SA), 00840682109260SEE19170036SA (SEE19170036SA), 00840682109260SEE19170037SA (SEE19170037SA), 00840682109260SEE19170041SA (SEE19170041SA), Not Applicable (SEE13120591GA), Not Applicable (SEE16390019SA), Not Applicable (SEE11061002GA), Not Applicable (SEE13120594GA), Not Applicable (SEE11071058GA), Not Applicable (SEE11071063GA), Not Applicable (SEE11071138GA), Not Applicable (SEE13180758GA), Not Applicable (SEE13180760GA), Not Applicable (SEE13180770GA), Not Applicable (SEE13180771GA), Not Applicable (SEE13270888GA), Not Applicable (SEE13431269GA), Not Applicable (SEE14452623GA), Not Applicable (SEE14452635GA), Not Applicable (SEE10400444GA), Not Applicable (SEE10400460GA), Not Applicable (SEE11081203GA), Not Applicable (SEE11081205GA), Not Applicable (SEE11101213GA), Not Applicable (SEE11101217GA), Not Applicable (SEE12182219GA), Not Applicable (SEE12232253GA), Not Applicable (SEE12232257GA), Not Applicable (SEE12232258GA), Not Applicable (SEE12232259GA), Not Applicable (SEE12232261GA), Not Applicable (SEE12232263GA), Not Applicable (SEE12232265GA), Not Applicable (SEE12232266GA), Not Applicable (SEE12242268GA), Not Applicable (SEE12242269GA), Not Applicable (SEE12242270GA), Not Applicable (SEE12242276GA), Not Applicable (SEE12242281GA), Not Applicable (SEE12242286GA), Not Applicable (SEE12242288GA), Not Applicable (SEE12242291GA), Not Applicable (SEE12242293GA), Not Applicable (SEE12430229GA), Not Applicable (SEE12440231GA), Not Applicable (SEE12440238GA), Not Applicable (SEE12510308GA), Not Applicable (SEE12520346GA), Not Applicable (SEE12520359GA), Not Applicable (SEE12530366GA), Not Applicable (SEE12530369GA), Not Applicable (SEE12530372GA), Not Applicable (SEE13000376GA), Not Applicable (SEE13000377GA), Not Applicable (SEE13000378GA), Not Applicable (SEE13000393GA), Not Applicable (SEE13150683GA), Not Applicable (SEE13150685GA), Not Applicable (SEE13431271GA), Not Applicable (SEE13431273GA), Not Applicable (SEE13431290GA), Not Applicable (SEE13431294GA), Not Applicable (SEE13431299GA), Not Applicable (SEE13431303GA), Not Applicable (SEE13431304GA), Not Applicable (SEE13441305GA), Not Applicable (SEE13451337GA), Not Applicable (SEE13451341GA), Not Applicable (SEE13451350GA), Not Applicable (SEE14021570GA), Not Applicable (SEE14021580GA), Not Applicable (SEE14031698GA), Not Applicable (SEE14031707GA), Not Applicable (SEE14031710GA), Not Applicable (SEE14031716GA), Not Applicable (SEE14041738GA), Not Applicable (SEE14041742GA), Not Applicable (SEE14372390GA), Not Applicable (SEE14372391GA), Not Applicable (SEE14372395GA), Not Applicable (SEE14372396GA), Not Applicable (SEE14372397GA), Not Applicable (SEE14372398GA), Not Applicable (SEE14372399GA), Not Applicable (SEE14372400GA), Not Applicable (SEE14372401GA), Not Applicable (SEE14372406GA), Not Applicable (SEE14372407GA), Not Applicable (SEE14372440GA), Not Applicable (SEE15052925GA), Not Applicable (SEE15052944GA), Not Applicable (SEE15052945GA), Not Applicable (SEE15062970GA), Not Applicable (SEE15062973GA), Not Applicable (SEE15062975GA), Not Applicable (SEE15062981GA), Not Applicable (SEE15063014GA), Not Applicable (SEE15063022GA), Not Applicable (SEE15073023GA), Not Applicable (SEE15241002SA), Not Applicable (SEE15241003SA), Not Applicable (SEE15241005SA), Not Applicable (SEE15241007SA), Not Applicable (SEE15373437SA), Not Applicable (SEE16038440SA), Not Applicable (SEE16038441SA), Not Applicable (SEE16038442SA), Not Applicable (SEE16344283SA), Not Applicable (SEE16344323SA), 00840682109260SEE17070004SA (SEE17070004SA), 00840682109260SEE17070010SA (SEE17070010SA), 00840682109260SEE17200001SA (SEE17200001SA), 00840682109260SEE17200003SA (SEE17200003SA), 00840682109260SEE18080001SA (SEE18080001SA), 00840682109260SEE18080012SA (SEE18080012SA), 00840682109260SEE18240011SA (SEE18240011SA), 00840682109260SEE18240019SA (SEE18240019SA), 00840682109260SEE18510018SA (SEE18510018SA), Not Applicable (SEE13431291GA), Not Applicable (SEE10260232GA), Not Applicable (SEE10260237GA), Not Applicable (SEE10310286GA), Not Applicable (SEE16293172SA), Not Applicable (SEE16293175SA), Not Applicable (SEE16293176SA), Not Applicable (SEE16293181SA), 00840682109260SEE18240056SA (SEE18240056SA), Not Applicable (SEE12340002GA), Not Applicable (SEE14452617GA), Not Applicable (SEE14452620GA), Not Applicable (SEE16241825SA), Not Applicable (SEE16241830SA), Not Applicable (SEE16241843SA), 00840682109260SEE19190003SA (SEE19190003SA), Not Applicable (SEE15052915GA), Not Applicable (SEE15052929GA), Not Applicable (SEE15093032GA), Not Applicable (SEE14372434GA), Not Applicable (SEE10460742GA), Not Applicable (SEE15210698SA), Not Applicable (SEE16430060SA), 00840682109260SEE17480001SA (SEE17480001SA), 00840682109260SEE18240058SA (SEE18240058SA), Not Applicable (SEE11331590GA), Not Applicable (SEE13270878GA), Not Applicable (SEE15517770SA), Not Applicable (SEE15517775SA), Not Applicable (SEE15301891SA), 00840682109260SEE18070005SA (SEE18070005SA), Not Applicable (SEE09530847GA), Not Applicable (SEE09530853GA), Not Applicable (SEE09530867GA), Not Applicable (SEE09530869GA), Not Applicable (SEE13280936GA), Not Applicable (SEE13280948GA), Not Applicable (SEE13501454GA), Not Applicable (SEE13501468GA), Not Applicable (SEE13501477GA), Not Applicable (SEE14312293GA), Not Applicable (SEE16380007SA), 00840682109260SEE18390008SA (SEE18390008SA), 00840682109260SEE19220009SA (SEE19220009SA), 00840682109260SEE19180007SA (SEE19180007SA), Not Applicable (SEE12410149GA), Not Applicable (SEE11071024GA), Not Applicable (SEE14151847GA), Not Applicable (SEE11211492GA), Not Applicable (SEE11211504GA), Not Applicable (SEE11241550GA), Not Applicable (SEE16191062SA), Not Applicable (SEE09140176GA), Not Applicable (SEE09140184GA), Not Applicable (SEE12182210GA), Not Applicable (SEE13240853GA), Not Applicable (SEE13240855GA), Not Applicable (SEE14362358GA), Not Applicable (SEE14362375GA), Not Applicable (SEE10120022GA), Not Applicable (SEE10250116GA), Not Applicable (SEE10250133GA), Not Applicable (SEE10250143GA), Not Applicable (SEE11040916GA), Not Applicable (SEE11050939GA), Not Applicable (SEE13110549GA), 00840682109260SEE17070020SA (SEE17070020SA), Not Applicable (SEE10121020GA), Not Applicable (SEE10121028GA), Not Applicable (SEE11201472GA), Not Applicable (SEE09340444GA), Not Applicable (SEE09510832GA), Not Applicable (SEE09530896GA), Not Applicable (SEE09530907GA), Not Applicable (SEE09530911GA), Not Applicable (SEE10450675GA), Not Applicable (SEE11081193GA), Not Applicable (SEE11371785GA), Not Applicable (SEE11511866GA), Not Applicable (SEE13280924GA), Not Applicable (SEE13290981GA), Not Applicable (SEE13311001GA), Not Applicable (SEE14432587GA), Not Applicable (SEE14462656GA), Not Applicable (SEE14462665GA), Not Applicable (SEE14462666GA), Not Applicable (SEE15113153GA), Not Applicable (SEE15113175GA), Not Applicable (SEE15113180GA), 00840682109260SEE18370010SA (SEE18370010SA), 00840682109260SEE18410016SA (SEE18410016SA), 00840682109260SEE19090021SA (SEE19090021SA), 00840682109260SEE19090023SA (SEE19090023SA), 01008406821092601119061121SEE19240027SA (SEE19240027SA), Not Applicable (SEE11071035GA), Not Applicable (SEE11071056GA), Not Applicable (SEE11071086GA), Not Applicable (SEE11071092GA), Not Applicable (SEE11071094GA), Not Applicable (SEE11151298GA), Not Applicable (SEE13220843GA), Not Applicable (SEE16390023SA), Not Applicable (SEE16420077SA), Not Applicable (SEE16430038SA), 00840682109260SEE19240021SA (SEE19240021SA), 00840682109260SEE19240029SA (SEE19240029SA), Not Applicable (SEE10420575GA), Not Applicable (SEE11361685GA), Not Applicable (SEE12390080GA), Not Applicable (SEE12390081GA), Not Applicable (SEE09390534GA), Not Applicable (SEE09020036GA), Not Applicable (SEE09190228GA), Not Applicable (SEE09240302GA), Not Applicable (SEE09240303GA), Not Applicable (SEE09250304GA), Not Applicable (SEE09250305GA), Not Applicable (SEE09250316GA), Not Applicable (SEE09250320GA), Not Applicable (SEE09250321GA), Not Applicable (SEE09250323GA), Not Applicable (SEE09250325GA), Not Applicable (SEE09250327GA), Not Applicable (SEE09250329GA), Not Applicable (SEE09250330GA), Not Applicable (SEE09250331GA), Not Applicable (SEE09250332GA), Not Applicable (SEE09260334GA), Not Applicable (SEE09260335GA), Not Applicable (SEE09260336GA), Not Applicable (SEE09260342GA), Not Applicable (SEE09260343GA), Not Applicable (SEE09260344GA), Not Applicable (SEE09260345GA), Not Applicable (SEE09300346GA), Not Applicable (SEE09300348GA), Not Applicable (SEE09300349GA), Not Applicable (SEE09300350GA), Not Applicable (SEE09300354GA), Not Applicable (SEE09300355GA), Not Applicable (SEE10260234GA), Not Applicable (SEE10430649GA), Not Applicable (SEE10450686GA), Not Applicable (SEE11141279GA), Not Applicable (SEE11141282GA), Not Applicable (SEE11341625GA), Not Applicable (SEE11361701GA), Not Applicable (SEE12530371GA), Not Applicable (SEE12530373GA), Not Applicable (SEE13050431GA), Not Applicable (SEE13050432GA), Not Applicable (SEE13050433GA), Not Applicable (SEE13050435GA), Not Applicable (SEE13050436GA), Not Applicable (SEE13050437GA), Not Applicable (SEE13050441GA), Not Applicable (SEE13060444GA), Not Applicable (SEE13060447GA), Not Applicable (SEE13060448GA), Not Applicable (SEE13060449GA), Not Applicable (SEE13060451GA), Not Applicable (SEE13060452GA), Not Applicable (SEE13060453GA), Not Applicable (SEE13060454GA), Not Applicable (SEE13180741GA), Not Applicable (SEE13180772GA), Not Applicable (SEE13451344GA), Not Applicable (SEE14382453GA), Not Applicable (SEE14382458GA), Not Applicable (SEE14382461GA), Not Applicable (SEE14382464GA), Not Applicable (SEE14412498GA), Not Applicable (SEE14452629GA), 00840682109260SEE17210006SA (SEE17210006SA), 00840682109260SEE17210007SA (SEE17210007SA), 00840682109260SEE17210009SA (SEE17210009SA), 00840682109260SEE17210010SA (SEE17210010SA), 00840682109260SEE17220001SA (SEE17220001SA), 00840682109260SEE17220003SA (SEE17220003SA), 00840682109260SEE18100029SA (SEE18100029SA), 00840682109260SEE19150003SA (SEE19150003SA), 00840682109260SEE19160010SA (SEE19160010SA), Not Applicable (SEE14422543GA), 00840682109260SEE17130003SA (SEE17130003SA), 00840682109260SEE17130006SA (SEE17130006SA), Not Applicable (SEE13140652GA), Not Applicable (SEE13180739GA), Not Applicable (SEE16450005SA), Not Applicable (SEE16450006SA), Not Applicable (SEE11071051GA), Not Applicable (SEE11071097GA), Not Applicable (SEE11071154GA), Not Applicable (SEE12122149GA), Not Applicable (SEE09330417GA), Not Applicable (SEE09530836GA), Not Applicable (SEE09530873GA), Not Applicable (SEE10250102GA), Not Applicable (SEE10390311GA), Not Applicable (SEE10530851GA), Not Applicable (SEE11161333GA), Not Applicable (SEE11171355GA), Not Applicable (SEE11171357GA), Not Applicable (SEE11171362GA), Not Applicable (SEE11211486GA), Not Applicable (SEE11211518GA), Not Applicable (SEE11321587GA), Not Applicable (SEE11341630GA), Not Applicable (SEE11341635GA), Not Applicable (SEE11351667GA), Not Applicable (SEE12312402GA), Not Applicable (SEE13351014GA), Not Applicable (SEE13351025GA), Not Applicable (SEE13351030GA), Not Applicable (SEE13351053GA), Not Applicable (SEE13361066GA), Not Applicable (SEE13361070GA), Not Applicable (SEE13391135GA), Not Applicable (SEE13411197GA), Not Applicable (SEE14021503GA), Not Applicable (SEE14031689GA), Not Applicable (SEE14031719GA), Not Applicable (SEE14041726GA), Not Applicable (SEE15093042GA), Not Applicable (SEE15093043GA), Not Applicable (SEE15173225GA), Not Applicable (SEE15261581SA), 00840682109260SEE17180004SA (SEE17180004SA), 00840682109260SEE17230074SA (SEE17230074SA), 00840682109260SEE19090025SA (SEE19090025SA), Not Applicable (SEE12021902GA), Not Applicable (SEE13220850GA), Not Applicable (SEE13270873GA), Not Applicable (SEE14151848GA), Not Applicable (SEE15352537SA), Not Applicable (SEE15352538SA), Not Applicable (SEE15352546SA), Not Applicable (SEE15352553SA), Not Applicable (SEE15352554SA), Not Applicable (SEE15352559SA), Not Applicable (SEE15352560SA),  Not Applicable (SEE15352563SA), Not Applicable (SEE15352564SA), Not Applicable (SEE14452622GA), Not Applicable (SEE14452632GA), 00840682109260SEE19100030SA (SEE19100030SA), 00840682109260SEE19100033SA (SEE19100033SA), Not Applicable (SEE10260198GA), Not Applicable (SEE10400363GA), Not Applicable (SEE10400369GA), Not Applicable (SEE10400381GA), Not Applicable (SEE10400382GA), Not Applicable (SEE11141255GA), Not Applicable (SEE11361732GA), Not Applicable (SEE11361735GA), Not Applicable (SEE13150663GA), Not Applicable (SEE13150666GA), Not Applicable (SEE13150669GA), 00840682109260SEE19110023SA (SEE19110023SA), 00840682109260SEE19110025SA (SEE19110025SA), 00840682109260SEE19110028SA (SEE19110028SA), 00840682109260SEE19110041SA (SEE19110041SA), 00840682109260SEE09510834GA (SEE09510834GA), 00840682109260SEE09530844GA (SEE09530844GA), 00840682109260SEE10100975GA (SEE10100975GA), 00840682109260SEE10100978GA (SEE10100978GA), 00840682109260SEE10120002GA (SEE10120002GA), Not Applicable (SEE10120004GA), 00840682109260SEE10120005GA (SEE10120005GA), 00840682109260SEE10120013GA (SEE10120013GA), 00840682109260SEE10120014GA (SEE10120014GA), 00840682109260SEE10120015GA (SEE10120015GA), 00840682109260SEE10120024GA (SEE10120024GA), 00840682109260SEE10120995GA (SEE10120995GA), 00840682109260SEE10120997GA (SEE10120997GA), 00840682109260SEE10121010GA (SEE10121010GA), 00840682109260SEE10121012GA (SEE10121012GA), 00840682109260SEE10121016GA (SEE10121016GA), 00840682109260SEE10130038GA (SEE10130038GA), 00840682109260SEE10430605GA (SEE10430605GA), 00840682109260SEE17060007SA (SEE17060007SA), 00840682109260SEE18350004SA (SEE18350004SA), 00840682109260SEE18350006SA (SEE18350006SA), 00840682109260SEE19110007SA (SEE19110007SA), 00840682109260SEE19110043SA (SEE19110043SA), 00840682109260SEE19110044SA (SEE19110044SA), Not Applicable (SEE10400457GA), Not Applicable (SEE10410465GA), Not Applicable (SEE10410466GA), Not Applicable (SEE10410496GA), Not Applicable (SEE10410497GA), Not Applicable (SEE10410500GA), Not Applicable (SEE10410501GA), Not Applicable (SEE10410510GA), Not Applicable (SEE10410512GA), Not Applicable (SEE11321581GA), Not Applicable (SEE11321582GA), Not Applicable (SEE12092053GA), Not Applicable (SEE12092060GA), Not Applicable (SEE12092061GA), Not Applicable (SEE12092063GA), Not Applicable (SEE13411194GA), Not Applicable (SEE13411196GA), Not Applicable (SEE13411199GA), Not Applicable (SEE13411200GA), Not Applicable (SEE13411201GA), Not Applicable (SEE13411210GA), Not Applicable (SEE13411213GA), Not Applicable (SEE13411216GA), Not Applicable (SEE13411217GA), Not Applicable (SEE14151811GA), Not Applicable (SEE14151814GA), Not Applicable (SEE14151825GA), Not Applicable (SEE14151831GA), Not Applicable (SEE14151850GA), 00840682109260SEE18240038SA (SEE18240038SA), 00840682109260SEE18240042SA (SEE18240042SA), Not Applicable (SEE13150671GA), Not Applicable (SEE10410469GA), Not Applicable (SEE10410482GA), 00840682109260SEE17100002SA (SEE17100002SA), Not Applicable (SEE16380024SA), Not Applicable (SEE10530854GA), Not Applicable (SEE10530856GA), Not Applicable (SEE10530861GA), Not Applicable (SEE10530869GA), Not Applicable (SEE13220844GA), Not Applicable (SEE09250324GA), Not Applicable (SEE09530893GA), Not Applicable (SEE09530912GA), Not Applicable (SEE13401142GA), 00840682109260SEE18160046SA (SEE18160046SA), 00840682109260SEE18160055SA (SEE18160055SA), Not Applicable (SEE11060997GA), Not Applicable (SEE11071005GA), Not Applicable (SEE11071006GA), Not Applicable (SEE10100982GA), Not Applicable (SEE10260216GA), Not Applicable (SEE10270246GA), Not Applicable (SEE11171383GA), Not Applicable (SEE12282357GA), Not Applicable (SEE15220796SA), Not Applicable (SEE15230839SA), 00840682109260SEE16520003SA (SEE16520003SA), 00840682109260SEE17450003SA (SEE17450003SA), 00840682109260SEE17450004SA (SEE17450004SA), 00840682109260SEE17450005SA (SEE17450005SA), 00840682109260SEE18100002SA (SEE18100002SA), 00840682109260SEE18100012SA (SEE18100012SA), Not Applicable (SEE11391832GA), Not Applicable (SEE11391838GA), Not Applicable (SEE11391839GA), 00840682109260SEE17270007SA (SEE17270007SA), Not Applicable (SEE09020027GA), Not Applicable (SEE15022868GA), Not Applicable (SEE15415183SA), 00840682109260SEE18050001SA (SEE18050001SA), 00840682109260SEE18050002SA (SEE18050002SA), Not Applicable (SEE12322414GA), Not Applicable (SEE13421253GA), Not Applicable (SEE13431275GA), Not Applicable (SEE13431301GA), Not Applicable (SEE09100138GA), Not Applicable (SEE09130171GA), Not Applicable (SEE09140187GA), Not Applicable (SEE09140190GA), Not Applicable (SEE09150200GA), Not Applicable (SEE09150202GA), Not Applicable (SEE09150205GA), Not Applicable (SEE09480743GA), Not Applicable (SEE10121021GA), Not Applicable (SEE10390335GA), Not Applicable (SEE10390336GA), Not Applicable (SEE10390338GA), Not Applicable (SEE10390339GA), Not Applicable (SEE10390345GA), Not Applicable (SEE10390346GA), Not Applicable (SEE10390351GA), Not Applicable (SEE10390352GA), Not Applicable (SEE10390354GA), Not Applicable (SEE10390357GA), Not Applicable (SEE10390358GA), Not Applicable (SEE10400371GA), Not Applicable (SEE10430623GA), Not Applicable (SEE10430652GA), Not Applicable (SEE10460705GA), Not Applicable (SEE11050972GA), Not Applicable (SEE11361672GA), Not Applicable (SEE12342453GA), Not Applicable (SEE14322315GA), Not Applicable (SEE14322318GA), Not Applicable (SEE14322327GA), Not Applicable (SEE14322332GA), Not Applicable (SEE14482683GA), Not Applicable (SEE14482702GA), Not Applicable (SEE14492722GA), Not Applicable (SEE14492724GA), Not Applicable (SEE14492725GA), Not Applicable (SEE14492726GA), Not Applicable (SEE14492727GA), Not Applicable (SEE14492728GA), Not Applicable (SEE14492730GA), Not Applicable (SEE14492731GA), Not Applicable (SEE14492745GA), Not Applicable (SEE14492747GA), Not Applicable (SEE14492748GA), Not Applicable (SEE14492749GA), Not Applicable (SEE14492750GA), Not Applicable (SEE14492751GA), Not Applicable (SEE14492752GA), Not Applicable (SEE14492753GA), Not Applicable (SEE14492754GA), Not Applicable (SEE14492755GA), Not Applicable (SEE14492756GA), Not Applicable (SEE14492757GA), Not Applicable (SEE14492758GA), Not Applicable (SEE14492759GA), Not Applicable (SEE14492760GA), Not Applicable (SEE14502762GA), Not Applicable (SEE14502763GA), Not Applicable (SEE14502764GA), Not Applicable (SEE14502765GA), Not Applicable (SEE14502766GA), Not Applicable (SEE14502767GA), Not Applicable (SEE14502768GA), Not Applicable (SEE14502769GA), Not Applicable (SEE14502770GA), Not Applicable (SEE14502771GA), Not Applicable (SEE14502772GA), Not Applicable (SEE14502773GA), Not Applicable (SEE14502774GA), Not Applicable (SEE14502775GA), Not Applicable (SEE14502776GA), Not Applicable (SEE14502777GA), Not Applicable (SEE14502778GA), Not Applicable (SEE14502779GA), Not Applicable (SEE14502780GA), 00840682109260SEE18160028SA (SEE18160028SA), Not Applicable (SEE09410560GA), Not Applicable (SEE09510769GA), Not Applicable (SEE09510776GA), Not Applicable (SEE09510783GA), Not Applicable (SEE09510801GA), Not Applicable (SEE09530863GA), Not Applicable (SEE11071034GA), Not Applicable (SEE11141273GA), Not Applicable (SEE11211514GA), Not Applicable (SEE11391833GA), Not Applicable (SEE11391834GA), Not Applicable (SEE11391836GA), Not Applicable (SEE12021907GA), Not Applicable (SEE12092058GA), Not Applicable (SEE12112121GA), Not Applicable (SEE12112122GA), Not Applicable (SEE12242289GA), Not Applicable (SEE12390071GA), Not Applicable (SEE12420168GA), Not Applicable (SEE13010405GA), Not Applicable (SEE13240857GA), Not Applicable (SEE13401153GA), Not Applicable (SEE13401157GA), Not Applicable (SEE13501453GA), Not Applicable (SEE13501470GA), Not Applicable (SEE14151812GA), Not Applicable (SEE14151841GA), Not Applicable (SEE14151863GA), Not Applicable (SEE14222039GA), Not Applicable (SEE14222050GA), Not Applicable (SEE14222057GA), Not Applicable (SEE14222059GA), Not Applicable (SEE14222067GA), Not Applicable (SEE14222072GA), Not Applicable (SEE14222073GA), Not Applicable (SEE14222075GA), Not Applicable (SEE15022862GA), 00840682109260SEE15032870GA (SEE15032870GA), 00840682109260SEE15032872GA (SEE15032872GA), 00840682109260SEE15032873GA (SEE15032873GA), 00840682109260SEE15032874GA (SEE15032874GA), 00840682109260SEE15032875GA (SEE15032875GA), Not Applicable (SEE15032877GA), Not Applicable (SEE15032881GA), 00840682109260SEE15032885GA (SEE15032885GA), 00840682109260SEE15032888GA (SEE15032888GA), 00840682109260SEE15032890GA (SEE15032890GA), Not Applicable (SEE15032893GA), 00840682109260SEE15301888SA (SEE15301888SA), Not Applicable (SEE15435919SA), Not Applicable (SEE15435920SA), Not Applicable (SEE15435921SA), Not Applicable (SEE15435922SA), Not Applicable (SEE15435927SA), Not Applicable (SEE15435928SA), Not Applicable (SEE15435933SA), Not Applicable (SEE15435934SA), Not Applicable (SEE16324001SA), Not Applicable (SEE16324026SA), Not Applicable (SEE16324027SA), Not Applicable (SEE16324031SA), Not Applicable (SEE16324033SA), Not Applicable (SEE16370006SA), 00840682109260SEE17270010SA (SEE17270010SA), 00840682109260SEE17340005SA (SEE17340005SA), 00840682109260SEE17340009SA (SEE17340009SA), 00840682109260SEE17340010SA (SEE17340010SA), 00840682109260SEE17340012SA (SEE17340012SA), 00840682109260SEE17340013SA (SEE17340013SA), 00840682109260SEE17340014SA (SEE17340014SA), 00840682109260SEE17460004SA (SEE17460004SA), 00840682109260SEE17510002SA (SEE17510002SA), 00840682109260SEE18030010SA (SEE18030010SA), 00840682109260SEE18040013SA (SEE18040013SA), Not Applicable (SEE12420167GA), Not Applicable (SEE10250083GA), Not Applicable (SEE10250085GA), Not Applicable (SEE10250098GA), Not Applicable (SEE10250100GA), Not Applicable (SEE10250101GA), Not Applicable (SEE10250111GA), Not Applicable (SEE10250145GA), Not Applicable (SEE10250156GA), Not Applicable (SEE10260233GA), Not Applicable (SEE15332401SA), Not Applicable (SEE15342480SA), Not Applicable (SEE15342482SA), Not Applicable (SEE15446117SA), Not Applicable (SEE15446119SA), Not Applicable (SEE15446121SA), Not Applicable (SEE15446124SA), Not Applicable (SEE15446169SA), Not Applicable (SEE15446170SA), Not Applicable (SEE15446201SA), Not Applicable (SEE15446203SA), Not Applicable (SEE15446204SA), Not Applicable (SEE15446215SA), Not Applicable (SEE15446216SA), 00840682109260SEE18190005SA (SEE18190005SA), 00840682109260SEE18190006SA (SEE18190006SA), 00840682109260SEE18190021SA (SEE18190021SA), 00840682109260SEE18240012SA (SEE18240012SA), 00840682109260SEE18240014SA (SEE18240014SA), 00840682109260SEE18240050SA (SEE18240050SA), Not Applicable (SEE13150686GA), Not Applicable (SEE09390540GA), Not Applicable (SEE09410554GA), Not Applicable (SEE09410556GA), Not Applicable (SEE09410557GA), Not Applicable (SEE09410558GA), Not Applicable (SEE09410559GA), Not Applicable (SEE09410575GA), Not Applicable (SEE14171933GA), Not Applicable (SEE14191941GA), Not Applicable (SEE14191942GA), Not Applicable (SEE14191943GA), Not Applicable (SEE14191945GA), Not Applicable (SEE14191947GA), Not Applicable (SEE14191948GA), Not Applicable (SEE14191949GA), Not Applicable (SEE09310372GA), Not Applicable (SEE09310379GA), Not Applicable (SEE11091212GA), Not Applicable (SEE11101218GA), Not Applicable (SEE11211475GA), Not Applicable (SEE11211508GA), Not Applicable (SEE14462667GA), Not Applicable (SEE14462668GA), 00840682109260SEE18160022SA (SEE18160022SA), 00840682109260SEE18160030SA (SEE18160030SA), Not Applicable (SEE09100084GA), Not Applicable (SEE09230282GA), Not Applicable (SEE10250125GA), Not Applicable (SEE10270257GA), Not Applicable (SEE09510812GA), Not Applicable (SEE12061996GA), Not Applicable (SEE14272172GA), Not Applicable (SEE14272173GA), Not Applicable (SEE14272213GA), Not Applicable (SEE14272214GA), Not Applicable (SEE10390348GA), Not Applicable (SEE10390350GA), Not Applicable (SEE10390361GA), Not Applicable (SEE10400364GA), Not Applicable (SEE10400366GA), Not Applicable (SEE11341606GA), Not Applicable (SEE11341609GA), Not Applicable (SEE11341612GA), Not Applicable (SEE11341615GA), Not Applicable (SEE11341620GA), Not Applicable (SEE11341660GA), Not Applicable (SEE11361676GA), Not Applicable (SEE11371765GA), Not Applicable (SEE12021908GA), Not Applicable (SEE12021911GA), Not Applicable (SEE12021922GA), Not Applicable (SEE12021923GA), Not Applicable (SEE12021924GA), Not Applicable (SEE12051989GA), Not Applicable (SEE16211513SA), Not Applicable (SEE11511883GA), Not Applicable (SEE11521890GA), Not Applicable (SEE13290980GA), Not Applicable (SEE13300991GA), Not Applicable (SEE13311002GA), Not Applicable (SEE13311005GA), Not Applicable (SEE13311008GA), Not Applicable (SEE09480748GA), Not Applicable (SEE13180733GA), Not Applicable (SEE13180756GA), Not Applicable (SEE13240852GA), Not Applicable (SEE13240858GA), Not Applicable (SEE13240860GA), Not Applicable (SEE13240866GA), Not Applicable (SEE13240869GA), Not Applicable (SEE14362365GA), Not Applicable (SEE12011898GA), Not Applicable (SEE16344325SA), 00840682109260SEE17050010SA (SEE17050010SA), Not Applicable (SEE16450012SA), Not Applicable (SEE16450014SA), Not Applicable (SEE09230265GA), Not Applicable (SEE09230268GA), Not Applicable (SEE09230281GA), Not Applicable (SEE09530837GA), Not Applicable (SEE09530859GA), Not Applicable (SEE11161338GA), Not Applicable (SEE11231532GA), Not Applicable (SEE11231536GA), Not Applicable (SEE10390298GA), Not Applicable (SEE13130620GA), Not Applicable (SEE13130622GA), Not Applicable (SEE12302379GA), Not Applicable (SEE12302391GA), Not Applicable (SEE12302396GA), Not Applicable (SEE15093035GA), Not Applicable (SEE15093036GA), Not Applicable (SEE15093041GA), Not Applicable (SEE15093044GA), 00840682109260SEE19090012SA (SEE19090012SA), Not Applicable (SEE13220842GA), Not Applicable (SEE13220845GA), Not Applicable (SEE13270875GA), Not Applicable (SEE13270876GA), Not Applicable (SEE13270882GA), Not Applicable (SEE13270885GA), Not Applicable (SEE11171369GA), Not Applicable (SEE11181410GA), Not Applicable (SEE11181413GA), Not Applicable (SEE11181415GA), Not Applicable (SEE11171365GA), Not Applicable (SEE09510787GA), Not Applicable (SEE11181401GA), Not Applicable (SEE13270896GA), Not Applicable (SEE13270897GA), Not Applicable (SEE15062999GA), Not Applicable (SEE15063001GA), Not Applicable (SEE15063007GA), Not Applicable (SEE12031929GA), Not Applicable (SEE12031934GA), Not Applicable (SEE09360486GA), Not Applicable (SEE09360488GA), Not Applicable (SEE12031928GA), Not Applicable (SEE12041975GA), Not Applicable (SEE12041976GA), Not Applicable (SEE12041977GA), Not Applicable (SEE12051978GA), Not Applicable (SEE12051979GA), Not Applicable (SEE12051980GA), Not Applicable (SEE12051981GA), Not Applicable (SEE12051982GA), Not Applicable (SEE12051983GA), Not Applicable (SEE12051984GA), Not Applicable (SEE12051985GA), Not Applicable (SEE14322308GA), Not Applicable (SEE14322310GA), Not Applicable (SEE14322317GA), Not Applicable (SEE15446200SA), Not Applicable (SEE15446214SA), 00840682109260SEE12092054GA (SEE12092054GA), 00840682109260SEE14081771GA (SEE14081771GA), 00840682109260SEE14081775GA (SEE14081775GA), 00840682109260SEE15032889GA (SEE15032889GA), Not Applicable (SEE16410002SA), Not Applicable (SEE16410004SA), Not Applicable (SEE16410009SA), Not Applicable (SEE16410014SA), Not Applicable (SEE16410016SA), Not Applicable (SEE10250084GA), Not Applicable (SEE10250099GA), Not Applicable (SEE10250136GA), Not Applicable (SEE15518044SA), Not Applicable (SEE15528074SA), Not Applicable (SEE11081188GA), Not Applicable (SEE15183261GA), 00840682109260SEE18320010SA (SEE18320010SA), 00840682109260SEE18320011SA (SEE18320011SA), Not Applicable (SEE12370048GA), Not Applicable (SEE12370057GA), Not Applicable (SEE09190238GA), Not Applicable (SEE09530840GA), Not Applicable (SEE09530865GA), Not Applicable (SEE12430206GA), Not Applicable (SEE12430209GA), Not Applicable (SEE16099454SA), Not Applicable (SEE12520336GA), Not Applicable (SEE12520351GA), Not Applicable (SEE12520353GA), Not Applicable (SEE13020413GA), Not Applicable (SEE13280922GA), Not Applicable (SEE13280925GA), Not Applicable (SEE13280926GA), Not Applicable (SEE13280928GA), Not Applicable (SEE13280930GA), Not Applicable (SEE13280931GA), Not Applicable (SEE16130303SA), Not Applicable (SEE16130305SA), Not Applicable (SEE16140328SA), Not Applicable (SEE16140330SA), Not Applicable (SEE16150532SA), Not Applicable (SEE16150540SA), Not Applicable (SEE16150541SA), Not Applicable (SEE16150542SA), Not Applicable (SEE16150543SA), Not Applicable (SEE16150544SA), Not Applicable (SEE16180898SA), Not Applicable (SEE16334182SA), Not Applicable (SEE16334185SA), Not Applicable (SEE14382454GA), Not Applicable (SEE16344326SA), Not Applicable (SEE15477175SA), Not Applicable (SEE15477184SA), Not Applicable (SEE15497498SA), Not Applicable (SEE15497500SA), Not Applicable (SEE15497501SA), Not Applicable (SEE15497510SA), Not Applicable (SEE15497511SA), Not Applicable (SEE15497512SA), Not Applicable (SEE15497513SA), Not Applicable (SEE15497514SA), Not Applicable (SEE15497515SA), Not Applicable (SEE15497516SA), Not Applicable (SEE15497517SA), Not Applicable (SEE15497520SA), Not Applicable (SEE15497521SA), Not Applicable (SEE15497522SA), Not Applicable (SEE15497523SA), Not Applicable (SEE15497525SA), Not Applicable (SEE15497526SA), Not Applicable (SEE15497528SA), Not Applicable (SEE15497529SA), Not Applicable (SEE15497531SA), Not Applicable (SEE15497532SA), Not Applicable (SEE15497533SA), Not Applicable (SEE15497534SA), Not Applicable (SEE15497535SA), Not Applicable (SEE15528073SA), Not Applicable (SEE10530852GA), Not Applicable (SEE10530870GA), Not Applicable (SEE11071055GA), Not Applicable (SEE11071147GA), Not Applicable (SEE11371748GA), Not Applicable (SEE11371756GA), Not Applicable (SEE11371795GA), Not Applicable (SEE15241004SA), 00840682109260SEE19110003SA (SEE19110003SA), 00840682109260SEE19110022SA (SEE19110022SA), 00840682109260SEE19110026SA (SEE19110026SA), 00840682109260SEE19110029SA (SEE19110029SA), 00840682109260SEE19110045SA (SEE19110045SA), 00840682109260SEE19110046SA (SEE19110046SA), Not Applicable (SEE11321564GA), Not Applicable (SEE11321567GA), Not Applicable (SEE11321573GA), Not Applicable (SEE16430024SA), Not Applicable (SEE13431295GA), Not Applicable (SEE11321579GA), Not Applicable (SEE10390323GA), Not Applicable (SEE11371788GA), Not Applicable (SEE11151301GA), Not Applicable (SEE11151302GA), Not Applicable (SEE11151305GA), Not Applicable (SEE11161327GA), Not Applicable (SEE12430188GA), Not Applicable (SEE14031684GA), Not Applicable (SEE14462644GA), Not Applicable (SEE14462646GA), Not Applicable (SEE14462649GA), Not Applicable (SEE14462652GA), Not Applicable (SEE15062989GA), Not Applicable (SEE15062991GA), Not Applicable (SEE15062997GA), Not Applicable (SEE15063009GA), Not Applicable (SEE15063012GA), Not Applicable (SEE15063016GA), Not Applicable (SEE16120171SA), Not Applicable (SEE09190232GA), Not Applicable (SEE09190235GA), Not Applicable (SEE09510830GA), Not Applicable (SEE09530842GA), Not Applicable (SEE11081156GA), Not Applicable (SEE14031664GA), Not Applicable (SEE15032883GA), Not Applicable (SEE15032899GA), 00840682109260SEE17410004SA (SEE17410004SA), Not Applicable (SEE13010395GA), Not Applicable (SEE10270254GA), Not Applicable (SEE12061999GA), Not Applicable (SEE12062001GA), Not Applicable (SEE12062010GA), Not Applicable (SEE12082018GA), Not Applicable (SEE12082019GA), Not Applicable (SEE12082021GA), Not Applicable (SEE12082023GA), Not Applicable (SEE12092055GA), Not Applicable (SEE12092059GA), Not Applicable (SEE12102101GA), Not Applicable (SEE13280903GA), Not Applicable (SEE16420074SA), Not Applicable (SEE13190792GA), Not Applicable (SEE14302281GA), Not Applicable (SEE15322281SA), Not Applicable (SEE15322282SA), Not Applicable (SEE15322283SA), Not Applicable (SEE15322284SA), Not Applicable (SEE15322285SA), Not Applicable (SEE15332291SA), Not Applicable (SEE15332292SA), Not Applicable (SEE15332293SA), Not Applicable (SEE15332294SA), Not Applicable (SEE15332295SA), 00840682109260SEE18420044SA (SEE18420044SA), 00840682109260SEE18420047SA (SEE18420047SA), 00840682109260SEE18420053SA (SEE18420053SA), 00840682109260SEE18500011SA (SEE18500011SA), 00840682109260SEE18500034SA (SEE18500034SA), 00840682109260SEE18510014SA (SEE18510014SA), Not Applicable (SEE12102092GA), Not Applicable (SEE13451308GA), Not Applicable (SEE13451322GA), Not Applicable (SEE12332443GA), 00840682109260SEE17510009SA (SEE17510009SA), 00840682109260SEE17400003SA (SEE17400003SA), 00840682109260SEE18080002SA (SEE18080002SA), 00840682109260SEE18080015SA (SEE18080015SA), Not Applicable (SEE10530867GA), Not Applicable (SEE11161317GA), Not Applicable (SEE11171385GA), Not Applicable (SEE13010404GA), Not Applicable (SEE13120613GA), Not Applicable (SEE13270877GA), 00840682109260SEE17500024SA (SEE17500024SA), Not Applicable (SEE13451335GA), Not Applicable (SEE13451345GA), Not Applicable (SEE13140653GA), Not Applicable (SEE13180722GA), Not Applicable (SEE13180743GA), Not Applicable (SEE13180744GA), Not Applicable (SEE13180745GA), Not Applicable (SEE13180746GA), Not Applicable (SEE13180747GA), Not Applicable (SEE13180748GA), Not Applicable (SEE13180749GA), Not Applicable (SEE13180750GA), Not Applicable (SEE13180754GA), Not Applicable (SEE13180755GA), Not Applicable (SEE13190811GA), Not Applicable (SEE15517819SA), Not Applicable (SEE15517820SA), 00840682109260SEE17480010SA (SEE17480010SA), 00840682109260SEE18060009SA (SEE18060009SA), 00840682109260SEE18060010SA (SEE18060010SA), 00840682109260SEE18060012SA (SEE18060012SA), 00840682109260SEE18060013SA (SEE18060013SA), 00840682109260SEE18060014SA (SEE18060014SA), 00840682109260SEE18060016SA (SEE18060016SA), 00840682109260SEE18060018SA (SEE18060018SA), 00840682109260SEE18080003SA (SEE18080003SA), 00840682109260SEE18080005SA (SEE18080005SA), 00840682109260SEE18080006SA (SEE18080006SA), 00840682109260SEE18080009SA (SEE18080009SA), 00840682109260SEE18080010SA (SEE18080010SA), 00840682109260SEE18080013SA (SEE18080013SA), 00840682109260SEE18350002SA (SEE18350002SA), 00840682109260SEE18400017SA (SEE18400017SA), Not Applicable (SEE14151821GA), Not Applicable (SEE14482689GA), Not Applicable (SEE14482701GA), Not Applicable (SEE14211981GA), Not Applicable (SEE14211987GA), Not Applicable (SEE15342446SA), Not Applicable (SEE09250319GA), Not Applicable (SEE09390544GA), Not Applicable (SEE11141277GA), Not Applicable (SEE11141280GA), Not Applicable (SEE11141285GA), Not Applicable (SEE11161326GA), Not Applicable (SEE11371754GA), Not Applicable (SEE11371796GA), Not Applicable (SEE11371801GA), Not Applicable (SEE11371802GA), Not Applicable (SEE11371803GA), Not Applicable (SEE12011893GA), Not Applicable (SEE12011899GA), Not Applicable (SEE16430062SA), 00840682109260SEE17090021SA (SEE17090021SA), 00840682109260SEE17090022SA (SEE17090022SA), 00840682109260SEE17090023SA (SEE17090023SA), 00840682109260SEE17090025SA (SEE17090025SA), 00840682109260SEE17090028SA (SEE17090028SA), Not Applicable (SEE14041730GA), Not Applicable (SEE09460654GA), Not Applicable (SEE09470671GA), Not Applicable (SEE13110566GA), Not Applicable (SEE14272143GA), Not Applicable (SEE14292237GA), Not Applicable (SEE15062988GA), Not Applicable (SEE15063018GA), Not Applicable (SEE15332306SA), Not Applicable (SEE15332308SA), Not Applicable (SEE15373422SA), Not Applicable (SEE16150492SA), Not Applicable (SEE16150497SA), Not Applicable (SEE16150498SA), Not Applicable (SEE16150500SA), Not Applicable (SEE16150501SA), Not Applicable (SEE16150503SA), Not Applicable (SEE16150525SA), Not Applicable (SEE16150526SA), Not Applicable (SEE16150527SA), Not Applicable (SEE16150529SA), Not Applicable (SEE16150530SA), Not Applicable (SEE16150533SA), Not Applicable (SEE16160734SA), 00840682109260SEE17250020SA (SEE17250020SA), 00840682109260SEE17270002SA (SEE17270002SA), 00840682109260SEE17270003SA (SEE17270003SA), 00840682109260SEE17270004SA (SEE17270004SA), 00840682109260SEE17270006SA (SEE17270006SA), 00840682109260SEE17270009SA (SEE17270009SA), 00840682109260SEE17270012SA (SEE17270012SA), 00840682109260SEE17490022SA (SEE17490022SA), 00840682109260SEE17490023SA (SEE17490023SA), 00840682109260SEE17490027SA (SEE17490027SA), 00840682109260SEE17490030SA (SEE17490030SA), 00840682109260SEE17490032SA (SEE17490032SA), 00840682109260SEE18040022SA (SEE18040022SA), 00840682109260SEE18040024SA (SEE18040024SA), Not Applicable (SEE11050974GA), Not Applicable (SEE11060947GA), Not Applicable (SEE12021913GA), 00840682109260SEE18500003SA (SEE18500003SA), 00840682109260SEE18500016SA (SEE18500016SA), 00840682109260SEE18500019SA (SEE18500019SA), Not Applicable (SEE16130321SA), Not Applicable (SEE16140436SA), Not Applicable (SEE16191058SA), Not Applicable (SEE16201178SA), Not Applicable (SEE16201181SA), Not Applicable (SEE16272598SA), Not Applicable (SEE13481389GA), Not Applicable (SEE13481393GA), Not Applicable (SEE13481394GA), Not Applicable (SEE13481397GA), Not Applicable (SEE13481398GA), Not Applicable (SEE13481399GA), Not Applicable (SEE10430633GA), 00840682109260SEE18270002SA (SEE18270002SA), 00840682109260SEE18270003SA (SEE18270003SA), Not Applicable (SEE16078924SA), Not Applicable (SEE16089177SA), 00840682109260SEE17500014SA (SEE17500014SA), 00840682109260SEE17500020SA (SEE17500020SA), 00840682109260SEE17500023SA (SEE17500023SA), 00840682109260SEE17510004SA (SEE17510004SA), 00840682109260SEE17510006SA (SEE17510006SA), Not Applicable (SEE14031708GA), Not Applicable (SEE14031715GA), Not Applicable (SEE14031723GA), Not Applicable (SEE12041947GA), Not Applicable (SEE12041952GA), Not Applicable (SEE12041954GA), Not Applicable (SEE12041955GA), Not Applicable (SEE14031592GA), Not Applicable (SEE12112109GA), Not Applicable (SEE12122157GA), Not Applicable (SEE12122162GA), Not Applicable (SEE12162170GA), Not Applicable (SEE13050438GA), Not Applicable (SEE13050443GA), Not Applicable (SEE13451310GA), Not Applicable (SEE13451318GA), Not Applicable (SEE09100103GA), 00840682109260SEE19240020SA (SEE19240020SA), Not Applicable (SEE09140194GA), Not Applicable (SEE09150204GA), Not Applicable (SEE09150210GA), Not Applicable (SEE10460748GA), Not Applicable (SEE10460751GA), Not Applicable (SEE15394716SA), Not Applicable (SEE15394717SA), Not Applicable (SEE15394718SA), Not Applicable (SEE15394719SA), Not Applicable (SEE15394742SA), Not Applicable (SEE15394743SA), Not Applicable (SEE09170223GA), Not Applicable (SEE09230252GA), Not Applicable (SEE09230257GA), Not Applicable (SEE09230261GA), Not Applicable (SEE09240294GA), Not Applicable (SEE09240296GA), Not Applicable (SEE09510796GA), Not Applicable (SEE09510804GA), Not Applicable (SEE10120027GA), Not Applicable (SEE10121025GA), Not Applicable (SEE10130045GA), Not Applicable (SEE10130069GA), Not Applicable (SEE11060948GA), Not Applicable (SEE11060949GA), Not Applicable (SEE11060951GA), Not Applicable (SEE11060954GA), Not Applicable (SEE11060980GA), Not Applicable (SEE11060981GA), Not Applicable (SEE11060982GA), Not Applicable (SEE11060983GA), Not Applicable (SEE11060984GA), Not Applicable (SEE11060985GA), Not Applicable (SEE11060986GA), Not Applicable (SEE11060987GA), Not Applicable (SEE11060989GA), Not Applicable (SEE11060994GA), Not Applicable (SEE11511872GA), Not Applicable (SEE11511876GA), Not Applicable (SEE11511885GA), Not Applicable (SEE11511887GA), Not Applicable (SEE12122132GA), Not Applicable (SEE12182211GA), Not Applicable (SEE12182224GA), Not Applicable (SEE12312403GA), Not Applicable (SEE13290982GA), Not Applicable (SEE14021554GA), Not Applicable (SEE14101786GA), Not Applicable (SEE14101793GA), Not Applicable (SEE14101794GA), Not Applicable (SEE14101795GA), Not Applicable (SEE14111802GA), Not Applicable (SEE14111803GA), Not Applicable (SEE14111806GA), Not Applicable (SEE14222003GA), Not Applicable (SEE15032891GA), Not Applicable (SEE15507658SA), Not Applicable (SEE15507662SA), 00840682109260SEE16241789SA (SEE16241789SA), 00840682109260SEE16241790SA (SEE16241790SA), 00840682109260SEE16241791SA (SEE16241791SA), 00840682109260SEE16241793SA (SEE16241793SA), 00840682109260SEE16241794SA (SEE16241794SA), 00840682109260SEE16241795SA (SEE16241795SA), 00840682109260SEE16241797SA (SEE16241797SA), Not Applicable (SEE16344282SA), Not Applicable (SEE16430028SA), 00840682109260SEE18060008SA (SEE18060008SA), 00840682109260SEE18060011SA (SEE18060011SA), 00840682109260SEE18060015SA (SEE18060015SA), 00840682109260SEE18160021SA (SEE18160021SA), 00840682109260SEE18160031SA (SEE18160031SA), 00840682109260SEE18160048SA (SEE18160048SA), 00840682109260SEE18160058SA (SEE18160058SA), 00840682109260SEE18160064SA (SEE18160064SA), 00840682109260SEE18160068SA (SEE18160068SA), 00840682109260SEE18160069SA (SEE18160069SA), Not Applicable (SEE09310398GA), Not Applicable (SEE09360492GA), Not Applicable (SEE09370498GA), Not Applicable (SEE09370501GA), Not Applicable (SEE09370503GA), Not Applicable (SEE09370504GA), Not Applicable (SEE09370506GA), Not Applicable (SEE09370509GA), Not Applicable (SEE09380519GA), Not Applicable (SEE09380528GA), Not Applicable (SEE09380529GA), Not Applicable (SEE09390538GA), Not Applicable (SEE10400461GA), Not Applicable (SEE10410473GA), Not Applicable (SEE10410487GA), Not Applicable (SEE10410489GA), Not Applicable (SEE10410491GA), Not Applicable (SEE10410494GA), Not Applicable (SEE10410514GA), Not Applicable (SEE10420536GA), Not Applicable (SEE10420537GA), Not Applicable (SEE10420538GA), Not Applicable (SEE10420550GA), Not Applicable (SEE12041953GA), Not Applicable (SEE12082037GA), Not Applicable (SEE12262330GA), Not Applicable (SEE12262331GA), Not Applicable (SEE12262332GA), Not Applicable (SEE12262333GA), Not Applicable (SEE12262335GA), Not Applicable (SEE12262336GA), Not Applicable (SEE12262337GA), Not Applicable (SEE12262343GA), Not Applicable (SEE12062002GA), Not Applicable (SEE13351019GA), Not Applicable (SEE14031661GA), 00840682109260SEE18460012SA (SEE18460012SA), Not Applicable (SEE11071059GA), Not Applicable (SEE13180753GA), Not Applicable (SEE13220847GA), Not Applicable (SEE13220848GA), Not Applicable (SEE13270872GA), Not Applicable (SEE14272201GA), Not Applicable (SEE16272271SA), Not Applicable (SEE16440004SA), 00840682109260SEE17430024SA (SEE17430024SA), 00840682109260SEE17220007SA (SEE17220007SA), Not Applicable (SEE10450678GA), Not Applicable (SEE16252024SA), Not Applicable (SEE16252037SA), Not Applicable (SEE16262084SA), Not Applicable (SEE10250127GA), Not Applicable (SEE13040422GA), Not Applicable (SEE13100457GA), Not Applicable (SEE13100458GA), Not Applicable (SEE13100459GA), Not Applicable (SEE13100460GA), Not Applicable (SEE13100461GA), Not Applicable (SEE13100462GA), Not Applicable (SEE13100463GA), Not Applicable (SEE13100464GA), Not Applicable (SEE13100465GA), Not Applicable (SEE13100466GA), Not Applicable (SEE13100479GA), Not Applicable (SEE13100480GA), Not Applicable (SEE13100481GA), Not Applicable (SEE13100482GA), Not Applicable (SEE13100483GA), Not Applicable (SEE13100485GA), Not Applicable (SEE13100486GA), Not Applicable (SEE13100487GA), Not Applicable (SEE13100488GA), Not Applicable (SEE13110491GA), Not Applicable (SEE13110492GA), Not Applicable (SEE14302282GA), Not Applicable (SEE15517776SA), Not Applicable (SEE15517818SA), Not Applicable (SEE10410470GA), Not Applicable (SEE10410485GA), 00840682109260SEE18030009SA (SEE18030009SA), 00840682109260SEE19110006SA (SEE19110006SA), 00840682109260SEE17350004SA (SEE17350004SA), Not Applicable (SEE14021579GA), Not Applicable (SEE14021581GA), Not Applicable (SEE14021583GA), Not Applicable (SEE14021585GA), Not Applicable (SEE14031588GA), Not Applicable (SEE14031591GA), 00840682109260SEE18100036SA (SEE18100036SA), Not Applicable (SEE11511865GA), Not Applicable (SEE11511868GA), Not Applicable (SEE11511870GA), Not Applicable (SEE11511878GA), Not Applicable (SEE09130172GA), Not Applicable (SEE09150206GA), Not Applicable (SEE10420593GA), Not Applicable (SEE10460715GA), Not Applicable (SEE12082017GA), Not Applicable (SEE15446202SA), Not Applicable (SEE15446234SA), Not Applicable (SEE15446245SA), Not Applicable (SEE15446255SA), Not Applicable (SEE15446260SA), Not Applicable (SEE15446262SA), Not Applicable (SEE15446331SA), Not Applicable (SEE15528218SA), 00840682109260SEE17150010SA (SEE17150010SA), 00840682109260SEE18500029SA (SEE18500029SA), 00840682109260SEE18500032SA (SEE18500032SA), 00840682109260SEE18500039SA (SEE18500039SA), 00840682109260SEE18500040SA (SEE18500040SA), 00840682109260SEE18500045SA (SEE18500045SA), Not Applicable (SEE16180910SA), Not Applicable (SEE16262145SA), Not Applicable (SEE12430207GA), 00840682109260SEE17290003SA (SEE17290003SA), 00840682109260SEE17290004SA (SEE17290004SA), 00840682109260SEE17290005SA (SEE17290005SA), 00840682109260SEE17290006S A (SEE17290006SA), 00840682109260SEE17290007SA (SEE17290007SA), 00840682109260SEE17300001SA (SEE17300001SA), 00840682109260SEE17300002SA (SEE17300002SA), 00840682109260SEE17300003SA (SEE17300003SA), 00840682109260SEE17300008SA (SEE17300008SA), 00840682109260SEE17300010SA (SEE17300010SA), 00840682109260SEE17300011SA (SEE17300011SA), 00840682109260SEE17300012SA (SEE17300012SA), 00840682109260SEE17300013SA (SEE17300013SA), 00840682109260SEE17300014SA (SEE17300014SA), Not Applicable (SEE13371108GA), 00840682109260SEE17490021SA (SEE17490021SA), Not Applicable (SEE13180777GA), Not Applicable (SEE14322309GA), 00840682109260SEE17430015SA (SEE17430015SA), 00840682109260SEE17430017SA (SEE17430017SA), 00840682109260SEE17430018SA (SEE17430018SA), 00840682109260SEE17430019SA (SEE17430019SA), 00840682109260SEE17430021SA (SEE17430021SA), 00840682109260SEE17430025SA (SEE17430025SA), 00840682109260SEE17430026SA (SEE17430026SA), 00840682109260SEE17430027SA (SEE17430027SA), 00840682109260SEE17430028SA (SEE17430028SA), 00840682109260SEE17430031SA (SEE17430031SA), 00840682109260SEE17430032SA (SEE17430032SA), 00840682109260SEE17430033SA (SEE17430033SA), 00840682109260SEE17430034SA (SEE17430034SA), 00840682109260SEE17430036SA (SEE17430036SA), 00840682109260SEE17430037SA (SEE17430037SA), 00840682109260SEE17430039SA (SEE17430039SA), 00840682109260SEE18160044SA (SEE18160044SA), 00840682109260SEE18160050SA (SEE18160050SA), 00840682109260SEE18160070SA (SEE18160070SA), Not Applicable (SEE12410138GA), Not Applicable (SEE12410144GA), Not Applicable (SEE12410147GA), Not Applicable (SEE12410148GA), Not Applicable (SEE09140181GA), Not Applicable (SEE09250322GA), Not Applicable (SEE09250328GA), Not Applicable (SEE09260337GA), Not Applicable (SEE16282785SA), Not Applicable (SEE16292871SA), 00840682109260SEE18160061SA (SEE18160061SA), 00840682109260SEE19180003SA (SEE19180003SA), Not Applicable (SEE09510764GA), Not Applicable (SEE09510803GA), Not Applicable (SEE10400403GA), Not Applicable (SEE10400440GA), Not Applicable (SEE10400455GA), Not Applicable (SEE10400462GA), Not Applicable (SEE10400463GA), Not Applicable (SEE10410472GA), Not Applicable (SEE10410479GA), Not Applicable (SEE10410492GA), Not Applicable (SEE10510811GA), Not Applicable (SEE11171376GA), Not Applicable (SEE11181391GA), Not Applicable (SEE11211509GA), Not Applicable (SEE11231525GA), Not Applicable (SEE12061998GA), Not Applicable (SEE15062984GA), Not Applicable (SEE15063013GA), 00840682109260SEE17160001SA (SEE17160001SA), 00840682109260SEE17160002SA (SEE17160002SA), 00840682109260SEE17440015SA (SEE17440015SA), 00840682109260SEE17440016SA (SEE17440016SA), 00840682109260SEE17440017SA (SEE17440017SA), 00840682109260SEE17440018SA (SEE17440018SA), 00840682109260SEE17440019SA (SEE17440019SA), Not Applicable (SEE10390356GA), 00840682109260SEE18450018SA (SEE18450018SA), 00840682109260SEE18450020SA (SEE18450020SA), 00840682109260SEE18450061SA (SEE18450061SA), 00840682109260SEE18450069SA (SEE18450069SA), 00840682109260SEE18450073SA (SEE18450073SA), 00840682109260SEE18460001SA (SEE18460001SA), 00840682109260SEE18460002SA (SEE18460002SA), Not Applicable (SEE09100099GA), Not Applicable (SEE09100112GA), Not Applicable (SEE09100137GA), Not Applicable (SEE09530846GA), Not Applicable (SEE10430626GA), 00840682109260SEE18370008SA (SEE18370008SA), 00840682109260SEE19160002SA (SEE19160002SA), 00840682109260SEE19170004SA (SEE19170004SA), Not Applicable (SEE11060963GA), Not Applicable (SEE11071146GA), 00840682109260SEE17380012SA (SEE17380012SA), Not Applicable (SEE10100980GA), 00840682109260SEE17050011SA (SEE17050011SA), Not Applicable (SEE15517769SA), Not Applicable (SEE14512825GA), Not Applicable (SEE11361743GA), Not Applicable (SEE11371775GA), Not Applicable (SEE11371806GA), Not Applicable (SEE11371813GA), Not Applicable (SEE11381816GA), Not Applicable (SEE11381817GA), Not Applicable (SEE11381818GA), Not Applicable (SEE11381820GA), Not Applicable (SEE11381822GA), Not Applicable (SEE11381823GA), Not Applicable (SEE11381825GA), Not Applicable (SEE11381828GA), Not Applicable (SEE11381830GA), Not Applicable (SEE11381831GA), Not Applicable (SEE12122163GA), Not Applicable (SEE13120586GA), Not Applicable (SEE13351037GA), Not Applicable (SEE13351038GA), Not Applicable (SEE13351039GA), Not Applicable (SEE13351041GA), Not Applicable (SEE13351047GA), Not Applicable (SEE13351052GA), Not Applicable (SEE14101790GA), Not Applicable (SEE14222027GA), Not Applicable (SEE14222047GA), Not Applicable (SEE14222054GA), Not Applicable (SEE14222055GA), Not Applicable (SEE16180909SA), Not Applicable (SEE16180924SA), Not Applicable (SEE16190968SA), Not Applicable (SEE16190970SA), Not Applicable (SEE16364421SA), 00840682109260SEE17150019SA (SEE17150019SA), 00840682109260SEE18330039SA (SEE18330039SA), 00840682109260SEE18330040SA (SEE18330040SA), 00840682109260SEE18370009SA (SEE18370009SA), 00840682109260SEE18370014SA (SEE18370014SA), 00840682109260SEE19100005SA (SEE19100005SA), 00840682109260SEE19100006SA (SEE19100006SA), 00840682109260SEE19100008SA (SEE19100008SA), 00840682109260SEE19100021SA (SEE19100021SA), Not Applicable (SEE16130323SA), Not Applicable (SEE16140329SA), Not Applicable (SEE09040066GA), Not Applicable (SEE10470796GA), Not Applicable (SEE13481403GA), Not Applicable (SEE13481410GA), Not Applicable (SEE13481413GA), Not Applicable (SEE13481419GA), 00840682109260SEE19120006SA (SEE19120006SA), Not Applicable (SEE10410503GA), Not Applicable (SEE14021507GA), Not Applicable (SEE14021509GA), Not Applicable (SEE14021515GA), Not Applicable (SEE14021516GA), Not Applicable (SEE15052934GA), 00840682109260SEE18510007SA (SEE18510007SA), Not Applicable (SEE13501456GA), Not Applicable (SEE13501483GA), Not Applicable (SEE13501485GA), Not Applicable (SEE13501493GA), Not Applicable (SEE13501498GA), Not Applicable (SEE09410553GA), Not Applicable (SEE12430216GA), Not Applicable (SEE12430220GA), Not Applicable (SEE15507704SA), Not Applicable (SEE14362367GA), Not Applicable (SEE14362372GA), Not Applicable (SEE14362373GA), Not Applicable (SEE14362377GA), Not Applicable (SEE13290969GA), Not Applicable (SEE13290972GA), Not Applicable (SEE13290975GA), Not Applicable (SEE14031600GA), Not Applicable (SEE16334089SA), 00840682109260SEE19100007SA (SEE19100007SA), 00840682109260SEE19100043SA (SEE19100043SA), 00840682109260SEE19100044SA (SEE19100044SA), 00840682109260SEE19100055SA (SEE19100055SA), 00840682109260SEE19100056SA (SEE19100056SA), 00840682109260SEE18030011SA (SEE18030011SA), Not Applicable (SEE09470669GA), Not Applicable (SEE10400439GA), Not Applicable (SEE10410498GA), Not Applicable (SEE10420520GA), Not Applicable (SEE10420532GA), Not Applicable (SEE10420562GA), Not Applicable (SEE10420580GA), Not Applicable (SEE10420582GA), Not Applicable (SEE10420602GA), Not Applicable (SEE10430641GA), Not Applicable (SEE10460722GA), Not Applicable (SEE10460726GA), Not Applicable (SEE10460739GA), Not Applicable (SEE10460763GA), Not Applicable (SEE10470799GA), Not Applicable (SEE10530815GA), Not Applicable (SEE10530816GA), Not Applicable (SEE10530818GA), Not Applicable (SEE10530819GA), Not Applicable (SEE10530820GA), Not Applicable (SEE10530821GA), Not Applicable (SEE10530822GA), Not Applicable (SEE10530823GA), Not Applicable (SEE10530826GA), Not Applicable (SEE10530830GA), Not Applicable (SEE10530835GA), Not Applicable (SEE12011895GA), Not Applicable (SEE15394284SA), Not Applicable (SEE15467026SA), 00840682109260SEE17440010SA (SEE17440010SA), 00840682109260SEE17440012SA (SEE17440012SA), 00840682109260SEE19170031SA (SEE19170031SA), 00840682109260SEE19180008SA (SEE19180008SA), 00840682109260SEE19220005SA (SEE19220005SA), Not Applicable (SEE09530862GA), Not Applicable (SEE13190805GA), Not Applicable (SEE15507701SA), Not Applicable (SEE12031925GA), Not Applicable (SEE14432610GA), Not Applicable (SEE16140432SA), Not Applicable (SEE16140434SA), Not Applicable (SEE12430194GA), Not Applicable (SEE12430198GA), Not Applicable (SEE12490271GA), Not Applicable (SEE16440007SA), Not Applicable (SEE10130040GA), Not Applicable (SEE10130049GA), Not Applicable (SEE10130051GA), Not Applicable (SEE10130056GA), Not Applicable (SEE10130059GA), Not Applicable (SEE10130061GA), Not Applicable (SEE10130062GA), Not Applicable (SEE10130063GA), Not Applicable (SEE10130067GA), Not Applicable (SEE10130068GA), 00840682109260SEE18100031SA (SEE18100031SA), 00840682109260SEE18160040SA (SEE18160040SA), 00840682109260SEE12390073GA (SEE12390073GA), Not Applicable (SEE13290965GA), Not Applicable (SEE13290966GA), Not Applicable (SEE13290968GA), Not Applicable (SEE13290976GA), Not Applicable (SEE13290979GA), Not Applicable (SEE13290987GA), Not Applicable (SEE13290988GA), Not Applicable (SEE13290989GA), Not Applicable (SEE13290990GA), Not Applicable (SEE13311003GA), Not Applicable (SEE13311007GA), Not Applicable (SEE13311009GA), Not Applicable (SEE13421224GA), Not Applicable (SEE13421225GA), Not Applicable (SEE13421226GA), Not Applicable (SEE13421229GA), Not Applicable (SEE13421230GA), Not Applicable (SEE13421236GA), Not Applicable (SEE13421240GA), Not Applicable (SEE13421241GA), Not Applicable (SEE13421243GA), Not Applicable (SEE13421251GA), Not Applicable (SEE13431261GA), Not Applicable (SEE13431263GA), Not Applicable (SEE13431265GA), Not Applicable (SEE13431267GA), Not Applicable (SEE13431277GA), Not Applicable (SEE13431278GA), Not Applicable (SEE13431280GA), Not Applicable (SEE13431281GA), Not Applicable (SEE13431282GA), Not Applicable (SEE13431284GA), Not Applicable (SEE14031704GA), 00840682109260SEE14031724GA (SEE14031724GA), 00840682109260SEE14031725GA (SEE14031725GA), Not Applicable (SEE14041735GA), Not Applicable (SEE15093037GA), Not Applicable (SEE15093040GA), Not Applicable (SEE15093050GA), 00840682109260SEE17250021SA (SEE17250021SA), 00840682109260SEE18510006SA (SEE18510006SA), Not Applicable (SEE09230256GA), Not Applicable (SEE09230258GA), Not Applicable (SEE09240297GA), Not Applicable (SEE09240301GA), Not Applicable (SEE09250312GA), Not Applicable (SEE11381826GA), Not Applicable (SEE15383876SA), Not Applicable (SEE09380516GA), Not Applicable (SEE09380523GA), Not Applicable (SEE09380527GA), Not Applicable (SEE09510790GA), Not Applicable (SEE15446254SA), Not Applicable (SEE15093074GA), Not Applicable (SEE15103077GA), Not Applicable (SEE15103094GA), Not Applicable (SEE16344327SA), Not Applicable (SEE10390329GA), 00840682109260SEE18500001SA (SEE18500001SA), 00840682109260SEE18500010SA (SEE18500010SA), 00840682109260SEE18500012SA (SEE18500012SA), 00840682109260SEE18500017SA (SEE18500017SA), 00840682109260SEE18500021SA (SEE18500021SA), Not Applicable (SEE14452631GA), Not Applicable (SEE12510302GA), Not Applicable (SEE15467025SA), Not Applicable (SEE16109691SA), Not Applicable (SEE16109693SA), Not Applicable (SEE16109694SA), Not Applicable (SEE16109695SA), Not Applicable (SEE14272136GA), Not Applicable (SEE14272140GA), Not Applicable (SEE14272144GA), Not Applicable (SEE11351668GA), Not Applicable (SEE11361674GA), Not Applicable (SEE11361677GA), Not Applicable (SEE11361679GA), Not Applicable (SEE11361691GA), Not Applicable (SEE11361694GA), Not Applicable (SEE12252311GA), 00840682109260SEE18060003SA (SEE18060003SA), Not Applicable (SEE09510817GA), Not Applicable (SEE09510824GA), Not Applicable (SEE10121026GA), Not Applicable (SEE10121027GA), Not Applicable (SEE10121029GA), Not Applicable (SEE10130070GA), Not Applicable (SEE10390305GA), Not Applicable (SEE10390306GA), Not Applicable (SEE10390334GA), Not Applicable (SEE15093062GA), Not Applicable (SEE15507659SA), Not Applicable (SEE15507663SA), Not Applicable (SEE16150528SA), Not Applicable (SEE16150545SA), Not Applicable (SEE16150554SA), Not Applicable (SEE16150556SA), Not Applicable (SEE16150557SA), Not Applicable (SEE16150569SA), 00840682109260SEE19220006SA (SEE19220006SA), Not Applicable (SEE14101782GA), Not Applicable (SEE11341641GA), Not Applicable (SEE11341644GA), Not Applicable (SEE11341647GA), Not Applicable (SEE11341648GA), Not Applicable (SEE11341649GA), Not Applicable (SEE11341650GA), Not Applicable (SEE11341652GA), Not Applicable (SEE11361715GA), Not Applicable (SEE11361717GA), Not Applicable (SEE11361719GA), Not Applicable (SEE11361728GA), Not Applicable (SEE11361731GA), Not Applicable (SEE13220828GA), Not Applicable (SEE13220829GA), Not Applicable (SEE13220831GA), Not Applicable (SEE13220833GA), Not Applicable (SEE13220836GA), Not Applicable (SEE14362352GA), Not Applicable (SEE14362359GA), Not Applicable (SEE14362360GA), Not Applicable (SEE14362361GA), Not Applicable (SEE14362362GA), Not Applicable (SEE14362363GA), Not Applicable (SEE14362364GA), Not Applicable (SEE14362368GA), Not Applicable (SEE14432590GA), Not Applicable (SEE15012845GA), Not Applicable (SEE16252023SA), Not Applicable (SEE16292877SA), Not Applicable (SEE16303213SA), 00840682109260SEE17010001SA (SEE17010001SA), 00840682109260SEE17010002SA (SEE17010002SA), 00840682109260SEE17010003SA (SEE17010003SA), 00840682109260SEE17010004SA (SEE17010004SA), 00840682109260SEE17010005SA (SEE17010005SA), 00840682109260SEE17010006SA (SEE17010006SA), 00840682109260SEE17010007SA (SEE17010007SA), 00840682109260SEE17010008SA (SEE17010008SA), 00840682109260SEE17010009SA (SEE17010009SA), 00840682109260SEE17020025SA (SEE17020025SA), 00840682109260SEE17260006SA (SEE17260006SA), 00840682109260SEE18500023SA (SEE18500023SA), 00840682109260SEE18500025SA (SEE18500025SA), 00840682109260SEE18500026SA (SEE18500026SA), 00840682109260SEE18500028SA (SEE18500028SA), 00840682109260SEE19160044SA (SEE19160044SA), 00840682109260SEE19160045SA (SEE19160045SA), Not Applicable (SEE16360020SA), Not Applicable (SEE12282350GA), Not Applicable (SEE12282360GA), Not Applicable (SEE12292361GA), Not Applicable (SEE12292364GA), Not Applicable (SEE15352545SA), Not Applicable (SEE12510332GA), Not Applicable (SEE12510334GA), Not Applicable (SEE12520341GA), Not Applicable (SEE12520350GA), Not Applicable (SEE12530370GA), Not Applicable (SEE13000383GA), Not Applicable (SEE13010398GA), Not Applicable (SEE13020411GA), Not Applicable (SEE13020412GA), Not Applicable (SEE13020414GA), Not Applicable (SEE16252015SA), Not Applicable (SEE13110511GA), Not Applicable (SEE13110519GA), Not Applicable (SEE15405017SA), Not Applicable (SEE15405060SA), Not Applicable (SEE16068747SA), Not Applicable (SEE09510820GA), Not Applicable (SEE12082033GA), Not Applicable (SEE13481409GA), Not Applicable (SEE13481411GA), Not Applicable (SEE13481412GA), Not Applicable (SEE13481414GA), Not Applicable (SEE13481420GA), 00840682109260SEE17220008SA (SEE17220008SA), 00840682109260SEE17220009SA (SEE17220009SA), 00840682109260SEE17220011SA (SEE17220011SA), 00840682109260SEE17110001SA (SEE17110001SA), 00840682109260SEE19170030SA (SEE19170030SA), Not Applicable (SEE10450662GA), Not Applicable (SEE10460703GA), Not Applicable (SEE13000380GA), Not Applicable (SEE13010396GA), Not Applicable (SEE13010397GA), Not Applicable (SEE10460727GA), Not Applicable (SEE10460746GA), Not Applicable (SEE10460767GA), Not Applicable (SEE11040902GA), Not Applicable (SEE11040904GA), Not Applicable (SEE11040906GA), Not Applicable (SEE11040912GA), Not Applicable (SEE11040913GA), Not Applicable (SEE11040918GA), Not Applicable (SEE12390086GA), Not Applicable (SEE12400100GA), Not Applicable (SEE14021512GA), Not Applicable (SEE14021518GA), Not Applicable (SEE13130621GA), Not Applicable (SEE13130638GA), Not Applicable (SEE11071031GA), Not Applicable (SEE11071099GA), Not Applicable (SEE12162183GA), Not Applicable (SEE12510303GA), Not Applicable (SEE12520349GA), Not Applicable (SEE13110510GA), Not Applicable (SEE13110529GA), Not Applicable (SEE13110531GA), Not Applicable (SEE13110533GA), Not Applicable (SEE13110536GA), Not Applicable (SEE13110537GA), Not Applicable (SEE13110539GA), Not Applicable (SEE13110544GA), Not Applicable (SEE13110546GA), Not Applicable (SEE13110562GA), Not Applicable (SEE13110565GA), Not Applicable (SEE13120577GA), Not Applicable (SEE13120578GA), Not Applicable (SEE14422528GA), Not Applicable (SEE14422530GA), Not Applicable (SEE14422531GA), Not Applicable (SEE14422533GA), Not Applicable (SEE16282651SA), Not Applicable (SEE16282653SA), Not Applicable (SEE16282788SA), Not Applicable (SEE16282790SA), Not Applicable (SEE16282791SA), 00840682109260SEE17130009SA (SEE17130009SA), 00840682109260SEE17130014SA (SEE17130014SA), 00840682109260SEE17130017SA (SEE17130017SA), 00840682109260SEE17130018SA (SEE17130018SA), 00840682109260SEE17130019SA (SEE17130019SA), 00840682109260SEE17430029SA (SEE17430029SA), 00840682109260SEE18400005SA (SEE18400005SA), 00840682109260SEE18400008SA (SEE18400008SA), 00840682109260SEE18400012SA (SEE18400012SA), 00840682109260SEE18400013SA (SEE18400013SA), 00840682109260SEE18400015SA (SEE18400015SA), 00840682109260SEE18400016SA (SEE18400016SA), Not Applicable (SEE09040079GA), Not Applicable (SEE09240300GA), Not Applicable (SEE09260341GA), Not Applicable (SEE09300357GA), Not Applicable (SEE09310358GA), Not Applicable (SEE09310362GA), Not Applicable (SEE09310364GA), Not Applicable (SEE09310366GA), Not Applicable (SEE09310369GA), Not Applicable (SEE09310370GA), Not Applicable (SEE09310373GA), Not Applicable (SEE09310377GA), Not Applicable (SEE09350466GA), Not Applicable (SEE09410569GA), Not Applicable (SEE09410572GA), Not Applicable (SEE09430601GA), Not Applicable (SEE09530884GA), Not Applicable (SEE10250082GA), Not Applicable (SEE10250088GA), Not Applicable (SEE10250092GA), Not Applicable (SEE10250096GA), Not Applicable (SEE11071015GA), Not Applicable (SEE11071025GA), Not Applicable (SEE11071027GA), Not Applicable (SEE11071028GA), Not Applicable (SEE11081210GA), Not Applicable (SEE11171363GA), Not Applicable (SEE11181404GA), Not Applicable (SEE11191429GA), Not Applicable (SEE11371782GA), Not Applicable (SEE15312135SA), Not Applicable (SEE15517771SA), Not Applicable (SEE16068741SA), Not Applicable (SEE16068742SA), Not Applicable (SEE16068743SA), Not Applicable (SEE16068745SA), Not Applicable (SEE16068749SA), Not Applicable (SEE16252026SA), Not Applicable (SEE16262132SA), Not Applicable (SEE16262134SA), Not Applicable (SEE16262137SA), Not Applicable (SEE16262138SA), Not Applicable (SEE16262140SA), Not Applicable (SEE16262142SA), Not Applicable (SEE16262146SA), Not Applicable (SEE16262149SA), Not Applicable (SEE16262151SA), Not Applicable (SEE16262161SA), Not Applicable (SEE16293180SA), Not Applicable (SEE16313375SA), Not Applicable (SEE16313376SA), Not Applicable (SEE16313377SA), Not Applicable (SEE16313379SA), Not Applicable (SEE16313380SA), Not Applicable (SEE16383418GA), Not Applicable (SEE16390013SA), 00840682109260SEE17330004SA (SEE17330004SA), 00840682109260SEE17330005SA (SEE17330005SA), 00840682109260SEE17330006SA (SEE17330006SA), 00840682109260SEE17330007SA (SEE17330007SA), 00840682109260SEE17330008SA (SEE17330008SA), 00840682109260SEE17330009SA (SEE17330009SA), 00840682109260SEE17330010SA (SEE17330010SA), 00840682109260SEE17360004SA (SEE17360004SA), 00840682109260SEE17360005SA (SEE17360005SA), 00840682109260SEE18050003SA (SEE18050003SA), 00840682109260SEE18050004SA (SEE18050004SA), 00840682109260SEE18250007SA (SEE18250007SA), 00840682109260SEE18250008SA (SEE18250008SA), 00840682109260SEE18250010SA (SEE18250010SA), 00840682109260SEE18250011SA (SEE18250011SA), 00840682109260SEE18250015SA (SEE18250015SA), 00840682109260SEE18250016SA (SEE18250016SA), 00840682109260SEE18250017SA (SEE18250017SA), 00840682109260SEE18250018SA (SEE18250018SA), 00840682109260SEE18250019SA (SEE18250019SA), 00840682109260SEE18250020SA (SEE18250020SA), 00840682109260SEE18250021SA (SEE18250021SA), 00840682109260SEE18250023SA (SEE18250023SA), 00840682109260SEE18250024SA (SEE18250024SA), 00840682109260SEE18250027SA (SEE18250027SA), 00840682109260SEE18250029SA (SEE18250029SA), 00840682109260SEE18250030SA (SEE18250030SA), 00840682109260SEE18250032SA (SEE18250032SA), 00840682109260SEE18250033SA (SEE18250033SA), 00840682109260SEE18250034SA (SEE18250034SA), Not Applicable (SEE15477111SA), Not Applicable (SEE15477183SA), 00840682109260SEE17180011SA (SEE17180011SA), 00840682109260SEE17180012SA (SEE17180012SA), 00840682109260SEE17180014SA (SEE17180014SA), Not Applicable (SEE13180785GA), Not Applicable (SEE13180786GA), Not Applicable (SEE09360483GA), Not Applicable (SEE09370505GA), Not Applicable (SEE09370507GA), Not Applicable (SEE12082016GA), Not Applicable (SEE13361058GA), Not Applicable (SEE13361065GA), Not Applicable (SEE14502804GA), Not Applicable (SEE14502806GA), Not Applicable (SEE11111228GA), Not Applicable (SEE11111233GA), Not Applicable (SEE11111239GA), Not Applicable (SEE11111247GA), Not Applicable (SEE15153186GA), Not Applicable (SEE15153188GA), Not Applicable (SEE15153198GA), Not Applicable (SEE15153201GA), Not Applicable (SEE15153202GA), Not Applicable (SEE15153205GA), Not Applicable (SEE16360025SA), 00840682109260SEE17080015SA (SEE17080015SA), 00840682109260SEE18450023SA (SEE18450023SA), 00840682109260SEE18450070SA (SEE18450070SA), 00840682109260SEE18450071SA (SEE18450071SA), Not Applicable (SEE16410007SA), Not Applicable (SEE16410015SA), 00840682109260SEE17010010SA (SEE17010010SA), Not Applicable (SEE12242287GA), Not Applicable (SEE12252317GA), Not Applicable (SEE12252318GA), Not Applicable (SEE12262326GA), Not Applicable (SEE12262345GA), Not Applicable (SEE13110515GA), Not Applicable (SEE13401143GA), Not Applicable (SEE13401150GA), Not Applicable (SEE14242092GA), Not Applicable (SEE14242100GA), Not Applicable (SEE14242108GA), Not Applicable (SEE14242113GA), Not Applicable (SEE14242114GA), Not Applicable (SEE15012848GA), Not Applicable (SEE15251299SA), Not Applicable (SEE15251301SA), Not Applicable (SEE16180900SA), 00840682109260SEE17270005SA (SEE17270005SA), 00840682109260SEE17270013SA (SEE17270013SA), 00840682109260SEE17270015SA (SEE17270015SA), 00840682109260SEE17410005SA (SEE17410005SA), 00840682109260SEE18040012SA (SEE18040012SA), 00840682109260SEE18100006SA (SEE18100006SA), 00840682109260SEE18100014SA (SEE18100014SA), 00840682109260SEE19090007SA (SEE19090007SA), Not Applicable (SEE10390291GA), Not Applicable (SEE10390343GA), Not Applicable (SEE10400392GA), Not Applicable (SEE10400406GA), Not Applicable (SEE10400421GA), Not Applicable (SEE11181389GA), Not Applicable (SEE11191441GA), Not Applicable (SEE15093070GA), Not Applicable (SEE15093073GA), 00840682109260SEE18040008SA (SEE18040008SA), 00840682109260SEE18040009SA (SEE18040009SA), 00840682109260SEE18040010SA (SEE18040010SA), 00840682109260SEE18040017SA (SEE18040017SA), 00840682109260SEE18040023SA (SEE18040023SA), 00840682109260SEE18040027SA (SEE18040027SA), 00840682109260SEE18040047SA (SEE18040047SA), Not Applicable (SEE15415184SA), Not Applicable (SEE15415186SA), Not Applicable (SEE15425741SA), Not Applicable (SEE15425782SA), Not Applicable (SEE09300352GA), Not Applicable (SEE09300356GA), Not Applicable (SEE10120026GA), Not Applicable (SEE10121009GA), Not Applicable (SEE10470803GA), Not Applicable (SEE10510810GA), Not Applicable (SEE11050940GA), Not Applicable (SEE11050944GA), Not Applicable (SEE11060996GA), Not Applicable (SEE11511867GA), Not Applicable (SEE11511874GA), Not Applicable (SEE13351020GA), Not Applicable (SEE13351021GA), Not Applicable (SEE13351033GA), Not Applicable (SEE13351034GA), Not Applicable (SEE13411195GA), Not Applicable (SEE13411207GA), Not Applicable (SEE14031666GA), Not Applicable (SEE14031692GA), Not Applicable (SEE14372405GA), Not Applicable (SEE14372410GA), Not Applicable (SEE14372414GA), Not Applicable (SEE14372415GA), Not Applicable (SEE14372422GA), Not Applicable (SEE14382459GA), Not Applicable (SEE14472673GA), Not Applicable (SEE15301887SA), Not Applicable (SEE15301889SA), Not Applicable (SEE15394375SA), Not Applicable (SEE16221692SA), Not Applicable (SEE16221693SA), Not Applicable (SEE16221697SA), Not Applicable (SEE16221698SA), Not Applicable (SEE16221699SA), 00840682109260SEE16520005SA (SEE16520005SA), 00840682109260SEE17350009SA (SEE17350009SA), 00840682109260SEE19240025SA (SEE19240025SA), 00840682109260SEE19240026SA (SEE19240026SA), Not Applicable (SEE09150203GA), Not Applicable (SEE09230287GA), Not Applicable (SEE09240290GA), Not Applicable (SEE12092085GA), Not Applicable (SEE12182206GA), Not Applicable (SEE09020035GA), Not Applicable (SEE09260340GA), Not Applicable (SEE10430639GA), Not Applicable (SEE10460706GA), Not Applicable (SEE12182231GA), Not Applicable (SEE14041732GA), Not Applicable (SEE15425786SA), Not Applicable (SEE15425787SA), Not Applicable (SEE15425788SA), Not Applicable (SEE15425789SA), Not Applicable (SEE15425790SA), Not Applicable (SEE15425792SA), Not Applicable (SEE15425794SA), Not Applicable (SEE15425795SA), Not Applicable (SEE15425801SA), Not Applicable (SEE09100086GA), Not Applicable (SEE09100110GA), Not Applicable (SEE09190227GA), Not Applicable (SEE09190229GA), Not Applicable (SEE09370510GA), Not Applicable (SEE09380518GA), Not Applicable (SEE09380526GA), Not Applicable (SEE09380530GA), Not Applicable (SEE09430599GA), Not Applicable (SEE09430606GA), Not Applicable (SEE09460650GA), Not Applicable (SEE09510811GA), Not Applicable (SEE09510822GA), Not Applicable (SEE09510833GA), Not Applicable (SEE09530866GA), Not Applicable (SEE10080925GA), Not Applicable (SEE10080927GA), Not Applicable (SEE10080928GA), Not Applicable (SEE10090948GA), Not Applicable (SEE10250106GA), Not Applicable (SEE10270272GA), Not Applicable (SEE10270273GA), Not Applicable (SEE10400376GA), Not Applicable (SEE10400377GA), Not Applicable (SEE10400399GA), Not Applicable (SEE10400416GA), Not Applicable (SEE10400417GA), Not Applicable (SEE10420560GA), Not Applicable (SEE11040905GA), Not Applicable (SEE11040908GA), Not Applicable (SEE11040909GA), Not Applicable (SEE11040910GA), Not Applicable (SEE11040914GA), Not Applicable (SEE11060957GA), Not Applicable (SEE11060962GA), 00840682109260SEE11171353GA (SEE11171353GA), Not Applicable (SEE11171368GA), Not Applicable (SEE11171372GA), Not Applicable (SEE11181407GA), Not Applicable (SEE11321578GA), Not Applicable (SEE11321583GA), Not Applicable (SEE12041949GA), Not Applicable (SEE12041958GA), Not Applicable (SEE12041959GA), Not Applicable (SEE12041960GA), Not Applicable (SEE12041961GA), Not Applicable (SEE12041962GA), Not Applicable (SEE12041963GA), Not Applicable (SEE12041964GA), Not Applicable (SEE12041965GA), Not Applicable (SEE12041967GA), Not Applicable (SEE12041968GA), Not Applicable (SEE12041969GA), Not Applicable (SEE12041970GA), Not Applicable (SEE12041971GA), Not Applicable (SEE12041972GA), Not Applicable (SEE12041974GA), Not Applicable (SEE12390083GA), Not Applicable (SEE12490277GA), Not Applicable (SEE12520347GA), Not Applicable (SEE12530367GA), Not Applicable (SEE12530368GA), Not Applicable (SEE13050430GA), Not Applicable (SEE13180728GA), Not Applicable (SEE13180729GA), Not Applicable (SEE13280939GA), Not Applicable (SEE13290961GA), Not Applicable (SEE13290970GA), Not Applicable (SEE13290978GA), Not Applicable (SEE13451346GA), Not Applicable (SEE13451349GA), Not Applicable (SEE14031667GA), Not Applicable (SEE14031669GA), Not Applicable (SEE14412503GA), Not Applicable (SEE14412506GA), Not Applicable (SEE14422524GA), Not Applicable (SEE14422526GA), Not Applicable (SEE14422539GA), Not Applicable (SEE14422552GA), Not Applicable (SEE14482695GA), Not Applicable (SEE15063000GA), Not Applicable (SEE15113143GA), Not Applicable (SEE15517772SA), Not Applicable (SEE15517817SA), Not Applicable (SEE16078930SA), Not Applicable (SEE16078932SA), Not Applicable (SEE16109579SA), Not Applicable (SEE16109580SA), Not Applicable (SEE16109583SA), Not Applicable (SEE16109584SA), Not Applicable (SEE16109594SA), Not Applicable (SEE16233391GA), Not Applicable (SEE16303405GA), 00840682109260SEE17460006SA (SEE17460006SA), 00840682109260SEE17460014SA (SEE17460014SA), 00840682109260SEE17460018SA (SEE17460018SA), 00840682109260SEE17490017SA (SEE17490017SA), 00840682109260SEE17490019SA (SEE17490019SA), 00840682109260SEE17490024SA (SEE17490024SA), 00840682109260SEE17490025SA (SEE17490025SA), 00840682109260SEE17490026SA (SEE17490026SA), 00840682109260SEE17490028SA (SEE17490028SA), 00840682109260SEE17490029SA (SEE17490029SA), 00840682109260SEE17490031SA (SEE17490031SA), 00840682109260SEE17490033SA (SEE17490033SA), 00840682109260SEE17490034SA (SEE17490034SA), 00840682109260SEE17490035SA (SEE17490035SA), 00840682109260SEE18160001SA (SEE18160001SA), 00840682109260SEE18160024SA (SEE18160024SA), 00840682109260SEE18160027SA (SEE18160027SA), 00840682109260SEE18330045SA (SEE18330045SA), 00840682109260SEE18330046SA (SEE18330046SA), 00840682109260SEE18330047SA (SEE18330047SA), 00840682109260SEE18330048SA (SEE18330048SA), 00840682109260SEE18370011SA (SEE18370011SA), 00840682109260SEE18370012SA (SEE18370012SA), 00840682109260SEE18450084SA (SEE18450084SA), 00840682109260SEE19090009SA (SEE19090009SA), 00840682109260SEE19090022SA (SEE19090022SA), 00840682109260SEE19130008SA (SEE19130008SA), 00840682109260SEE19170020SA (SEE19170020SA), 00840682109260SEE19370038SA (SEE19370038SA), 00840682109260SEE19370058SA (SEE19370058SA), 00840682109260SEE19370059SA (SEE19370059SA), 00840682109260SEE11060999GA (SEE11060999GA), 00840682109260SEE11071008GA (SEE11071008GA), 00840682109260SEE11071019GA (SEE11071019GA), 00840682109260SEE11071023GA (SEE11071023GA), 00840682109260SEE13481415GA (SEE13481415GA), 00840682109260SEE13491425GA (SEE13491425GA), 00840682109260SEE16058655SA (SEE16058655SA), 00840682109260SEE16058660SA (SEE16058660SA), Not Applicable (SEE13361069GA), Not Applicable (SEE13361075GA), Not Applicable (SEE13371083GA), Not Applicable (SEE13371085GA), Not Applicable (SEE15022859GA), Not Applicable (SEE16303224SA), Not Applicable (SEE16313378SA), Not Applicable (SEE16450002SA), Not Applicable (SEE16450004SA), Not Applicable (SEE16450009SA), Not Applicable (SEE12292362GA), Not Applicable (SEE12302373GA), Not Applicable (SEE10460740GA), Not Applicable (SEE10460761GA), Not Applicable (SEE10460766GA), Not Applicable (SEE10460772GA), Not Applicable (SEE10510809GA), Not Applicable (SEE10510812GA), Not Applicable (SEE11371774GA), Not Applicable (SEE12021904GA), Not Applicable (SEE12021918GA), Not Applicable (SEE12062013GA), Not Applicable (SEE14222051GA), Not Applicable (SEE14222063GA), Not Applicable (SEE14222064GA), Not Applicable (SEE14242104GA), Not Applicable (SEE14242105GA), 00840682109260SEE18240018SA (SEE18240018SA), Not Applicable (SEE09350457GA), Not Applicable (SEE09350460GA), Not Applicable (SEE09350467GA), Not Applicable (SEE09360473GA), Not Applicable (SEE11050975GA), Not Applicable (SEE11060958GA), Not Applicable (SEE11060978GA), Not Applicable (SEE14432588GA), Not Applicable (SEE14432591GA), Not Applicable (SEE14432603GA), Not Applicable (SEE14432604GA), Not Applicable (SEE14432605GA), Not Applicable (SEE14432606GA), Not Applicable (SEE15093047GA), Not Applicable (SEE15093052GA), Not Applicable (SEE15352795SA), Not Applicable (SEE15352796SA), Not Applicable (SEE15352846SA), 00840682109260SEE16520002SA (SEE16520002SA), 00840682109260SEE16520004SA (SEE16520004SA), Not Applicable (SEE10410506GA), Not Applicable (SEE10460762GA), Not Applicable (SEE10460769GA), 00840682109260SEE18500038SA (SEE18500038SA), Not Applicable (SEE12302376GA), Not Applicable (SEE12302381GA), Not Applicable (SEE12302392GA), Not Applicable (SEE12302393GA), Not Applicable (SEE12302394GA), Not Applicable (SEE14452615GA), Not Applicable (SEE14452616GA), Not Applicable (SEE14452630GA), Not Applicable (SEE14452638GA), Not Applicable (SEE15012854GA), Not Applicable (SEE15012855GA), Not Applicable (SEE11191437GA), Not Applicable (SEE11201459GA), Not Applicable (SEE11211489GA), Not Applicable (SEE11211505GA), Not Applicable (SEE11211511GA), Not Applicable (SEE11211513GA), Not Applicable (SEE11211517GA), Not Applicable (SEE11211523GA), Not Applicable (SEE11361681GA), Not Applicable (SEE11361686GA), Not Applicable (SEE11361692GA), Not Applicable (SEE12410142GA), Not Applicable (SEE11071030GA), Not Applicable (SEE11071052GA), Not Applicable (SEE11141263GA), Not Applicable (SEE11141265GA), Not Applicable (SEE11141276GA), Not Applicable (SEE11141284GA), Not Applicable (SEE15281641SA), Not Applicable (SEE09310368GA), Not Applicable (SEE09310371GA), Not Applicable (SEE09310375GA), Not Applicable (SEE09310382GA), Not Applicable (SEE09310384GA), Not Applicable (SEE09310387GA), Not Applicable (SEE09330412GA), Not Applicable (SEE09330413GA), Not Applicable (SEE09330427GA), Not Applicable (SEE14472674GA), Not Applicable (SEE09530845GA), Not Applicable (SEE13180751GA), Not Applicable (SEE13180752GA), Not Applicable (SEE13180761GA), Not Applicable (SEE13180764GA), Not Applicable (SEE13180773GA), Not Applicable (SEE13180779GA), Not Applicable (SEE13180782GA), Not Applicable (SEE14031721GA),  Not Applicable (SEE14081767GA), Not Applicable (SEE14081768GA), Not Applicable (SEE14081772GA), Not Applicable (SEE14081773GA), Not Applicable (SEE14222040GA), Not Applicable (SEE14222056GA), Not Applicable (SEE14242079GA), Not Applicable (SEE14242080GA), Not Applicable (SEE14242085GA), Not Applicable (SEE14242086GA), Not Applicable (SEE14242088GA), Not Applicable (SEE14242109GA), Not Applicable (SEE14242111GA), Not Applicable (SEE11151308GA), Not Applicable (SEE11161325GA), Not Applicable (SEE13130625GA), Not Applicable (SEE13130641GA), Not Applicable (SEE13130643GA), Not Applicable (SEE13130644GA), Not Applicable (SEE13130647GA), Not Applicable (SEE13220826GA), 00840682109260SEE18420018SA (SEE18420018SA), Not Applicable (SEE14452618GA), Not Applicable (SEE14452625GA), 00840682109260SEE17390003SA (SEE17390003SA), 00840682109260SEE18450012SA (SEE18450012SA), 00840682109260SEE18450015SA (SEE18450015SA), Not Applicable (SEE09100085GA), Not Applicable (SEE09100126GA), Not Applicable (SEE09130160GA), Not Applicable (SEE09130165GA), Not Applicable (SEE11181417GA), Not Applicable (SEE13421232GA), Not Applicable (SEE13421235GA), Not Applicable (SEE13421238GA), Not Applicable (SEE13421246GA), Not Applicable (SEE13421247GA), Not Applicable (SEE14222029GA), Not Applicable (SEE14222068GA), Not Applicable (SEE14222076GA), Not Applicable (SEE15373463SA), Not Applicable (SEE15373464SA), Not Applicable (SEE15373465SA), Not Applicable (SEE15373466SA), Not Applicable (SEE15373467SA), Not Applicable (SEE15373531SA), Not Applicable (SEE15373532SA), 00840682109260SEE17460011SA (SEE17460011SA), 00840682109260SEE18360002SA (SEE18360002SA), 00840682109260SEE18360003SA (SEE18360003SA), 00840682109260SEE18360008SA (SEE18360008SA), Not Applicable (SEE14031617GA), Not Applicable (SEE14031623GA), Not Applicable (SEE14031631GA), Not Applicable (SEE14031634GA), Not Applicable (SEE14031642GA), Not Applicable (SEE15456718SA), Not Applicable (SEE15456720SA), Not Applicable (SEE15487295SA), Not Applicable (SEE13140654GA), Not Applicable (SEE13140656GA), Not Applicable (SEE13150664GA), Not Applicable (SEE13180724GA), Not Applicable (SEE16364406SA), Not Applicable (SEE16364407SA), 00840682109260SEE17440005SA (SEE17440005SA), 00840682109260SEE17440009SA (SEE17440009SA), 00840682109260SEE17440011SA (SEE17440011SA), 00840682109260SEE17440013SA (SEE17440013SA), 00840682109260SEE17520001SA (SEE17520001SA), Not Applicable (SEE09100097GA), Not Applicable (SEE09100117GA), Not Applicable (SEE09130158GA), Not Applicable (SEE09190230GA), Not Applicable (SEE09380532GA), Not Applicable (SEE09470686GA), Not Applicable (SEE09470687GA), Not Applicable (SEE09480737GA), Not Applicable (SEE09510763GA), Not Applicable (SEE09510765GA), Not Applicable (SEE09510766GA), Not Applicable (SEE09510772GA), Not Applicable (SEE09510773GA), Not Applicable (SEE09510777GA), Not Applicable (SEE09510778GA), Not Applicable (SEE09510788GA), Not Applicable (SEE09510795GA), Not Applicable (SEE09510806GA), Not Applicable (SEE09510807GA), Not Applicable (SEE09510809GA), Not Applicable (SEE10130053GA), Not Applicable (SEE10250090GA), Not Applicable (SEE10250093GA), Not Applicable (SEE10250108GA), Not Applicable (SEE10250112GA), Not Applicable (SEE10250120GA), Not Applicable (SEE10250161GA), Not Applicable (SEE10390293GA), Not Applicable (SEE10390294GA), Not Applicable (SEE10390308GA), Not Applicable (SEE10390316GA), Not Applicable (SEE10390321GA), Not Applicable (SEE10390322GA), Not Applicable (SEE10420573GA), Not Applicable (SEE10460756GA), Not Applicable (SEE11071054GA), Not Applicable (SEE11071064GA), Not Applicable (SEE11071132GA), Not Applicable (SEE11101224GA), Not Applicable (SEE11161318GA), Not Applicable (SEE11161323GA), Not Applicable (SEE11161331GA), Not Applicable (SEE11211498GA), Not Applicable (SEE11341602GA), Not Applicable (SEE11341604GA), Not Applicable (SEE11341605GA), Not Applicable (SEE11341607GA), Not Applicable (SEE11341608GA), Not Applicable (SEE11341610GA), Not Applicable (SEE11341611GA), Not Applicable (SEE11341613GA), Not Applicable (SEE11341617GA), Not Applicable (SEE11341618GA), Not Applicable (SEE11341621GA), Not Applicable (SEE11341623GA), Not Applicable (SEE11341637GA), Not Applicable (SEE11341639GA), Not Applicable (SEE11341654GA), Not Applicable (SEE11361689GA), Not Applicable (SEE11371786GA), Not Applicable (SEE11371799GA), Not Applicable (SEE11381829GA), Not Applicable (SEE12031943GA), Not Applicable (SEE12031944GA), Not Applicable (SEE12182221GA), Not Applicable (SEE12510325GA), Not Applicable (SEE13280929GA), Not Applicable (SEE13280933GA), Not Applicable (SEE13280940GA), Not Applicable (SEE13290952GA), Not Applicable (SEE13351015GA), Not Applicable (SEE13351026GA), Not Applicable (SEE13401152GA), Not Applicable (SEE13401166GA), Not Applicable (SEE13401167GA), Not Applicable (SEE13401177GA), Not Applicable (SEE14021576GA), Not Applicable (SEE14322336GA), Not Applicable (SEE14362366GA), Not Applicable (SEE14362370GA), Not Applicable (SEE14362371GA), Not Applicable (SEE14362376GA), Not Applicable (SEE14402473GA), Not Applicable (SEE14402474GA), Not Applicable (SEE14402475GA), Not Applicable (SEE14402476GA), Not Applicable (SEE14402477GA), Not Applicable (SEE14402478GA), Not Applicable (SEE14402479GA), Not Applicable (SEE14402480GA), Not Applicable (SEE14402482GA), Not Applicable (SEE14402483GA), Not Applicable (SEE14402487GA), Not Applicable (SEE14482684GA), Not Applicable (SEE14482700GA), Not Applicable (SEE14482703GA), Not Applicable (SEE14482705GA), Not Applicable (SEE14482707GA), Not Applicable (SEE14482708GA), Not Applicable (SEE14482709GA), Not Applicable (SEE14482710GA), Not Applicable (SEE14482711GA), Not Applicable (SEE14482712GA), Not Applicable (SEE14482713GA), Not Applicable (SEE14482714GA), Not Applicable (SEE14482716GA), Not Applicable (SEE14482721GA), Not Applicable (SEE15022861GA), Not Applicable (SEE15022867GA), Not Applicable (SEE15042901GA), Not Applicable (SEE16099458SA), Not Applicable (SEE16160725SA), Not Applicable (SEE16191065SA), Not Applicable (SEE16262164SA), Not Applicable (SEE16360014SA), Not Applicable (SEE16360017SA), Not Applicable (SEE16364418SA), Not Applicable (SEE16364420SA), Not Applicable (SEE16390024SA), Not Applicable (SEE16410011SA), Not Applicable (SEE16430032SA), Not Applicable (SEE16430035SA), Not Applicable (SEE16430037SA), Not Applicable (SEE16430052SA), 00840682109260SEE17020021SA (SEE17020021SA), 00840682109260SEE17310006SA (SEE17310006SA), 00840682109260SEE17310007SA (SEE17310007SA), 00840682109260SEE17310008SA (SEE17310008SA), 00840682109260SEE17310009SA (SEE17310009SA), 00840682109260SEE17310011SA (SEE17310011SA), 00840682109260SEE17310013SA (SEE17310013SA), 00840682109260SEE17310014SA (SEE17310014SA), 00840682109260SEE17310016SA (SEE17310016SA), 00840682109260SEE17310017SA (SEE17310017SA), 00840682109260SEE17340016SA (SEE17340016SA), 00840682109260SEE17380008SA (SEE17380008SA), 00840682109260SEE17410001SA (SEE17410001SA), 00840682109260SEE17410002SA (SEE17410002SA), 00840682109260SEE17430023SA (SEE17430023SA), 00840682109260SEE17430035SA (SEE17430035SA), 00840682109260SEE18170012SA (SEE18170012SA), 00840682109260SEE18450002SA (SEE18450002SA), 00840682109260SEE18450007SA (SEE18450007SA), 00840682109260SEE18450014SA (SEE18450014SA), 00840682109260SEE19100003SA (SEE19100003SA), 00840682109260SEE19100004SA (SEE19100004SA), 00840682109260SEE19100009SA (SEE19100009SA), 00840682109260SEE19100035SA (SEE19100035SA), 00840682109260SEE19220029SA (SEE19220029SA), 00840682109260SEE19220033SA (SEE19220033SA), Not Applicable (SEE13120593GA), Not Applicable (SEE13120607GA), Not Applicable (SEE13120609GA), Not Applicable (SEE13120610GA), Not Applicable (SEE13120611GA), Not Applicable (SEE13120614GA), Not Applicable (SEE16380031SA), Not Applicable (SEE16068744SA), Not Applicable (SEE10420517GA), Not Applicable (SEE10420519GA), Not Applicable (SEE09410586GA), Not Applicable (SEE11050930GA), Not Applicable (SEE11201452GA), Not Applicable (SEE11201465GA), Not Applicable (SEE11201468GA), Not Applicable (SEE11201471GA), Not Applicable (SEE14031596GA), Not Applicable (SEE14031691GA), Not Applicable (SEE14031695GA), Not Applicable (SEE14031696GA), Not Applicable (SEE14031697GA), Not Applicable (SEE14221998GA), Not Applicable (SEE15261530SA), Not Applicable (SEE15261537SA), Not Applicable (SEE16038553SA), Not Applicable (SEE16038554SA), Not Applicable (SEE16038556SA), Not Applicable (SEE16038557SA), Not Applicable (SEE16390018SA), Not Applicable (SEE16390020SA), 00840682109260SEE18100005SA (SEE18100005SA), 00840682109260SEE18360016SA (SEE18360016SA), Not Applicable (SEE11321576GA), Not Applicable (SEE11321580GA), Not Applicable (SEE11331591GA), Not Applicable (SEE11331593GA), Not Applicable (SEE11331597GA), Not Applicable (SEE09410568GA), Not Applicable (SEE09410571GA), Not Applicable (SEE09410580GA), Not Applicable (SEE14031641GA), Not Applicable (SEE14031674GA), Not Applicable (SEE14031683GA), Not Applicable (SEE14031686GA), Not Applicable (SEE14031703GA), Not Applicable (SEE16221558SA), Not Applicable (SEE16221559SA), Not Applicable (SEE16430005SA), Not Applicable (SEE16430007SA), Not Applicable (SEE16430021SA), 00840682109260SEE17480012SA (SEE17480012SA), 00840682109260SEE17480014SA (SEE17480014SA), 00840682109260SEE19130019SA (SEE19130019SA), 00840682109260SEE19150004SA (SEE19150004SA), 00840682109260SEE19150005SA (SEE19150005SA), 00840682109260SEE19170028SA (SEE19170028SA), 00840682109260SEE19170029SA (SEE19170029SA), Not Applicable (SEE10420531GA), Not Applicable (SEE12082026GA), Not Applicable (SEE12082027GA), Not Applicable (SEE12082038GA), Not Applicable (SEE12082039GA), Not Applicable (SEE12082040GA), Not Applicable (SEE12092050GA), Not Applicable (SEE12520342GA), 00840682109260SEE18330044SA (SEE18330044SA), 00840682109260SEE18330050SA (SEE18330050SA), Not Applicable (SEE11171360GA), Not Applicable (SEE11211515GA), Not Applicable (SEE15446240SA), Not Applicable (SEE15446334SA), Not Applicable (SEE16109582SA), Not Applicable (SEE16119901SA), Not Applicable (SEE16119921SA), Not Applicable (SEE11071026GA), Not Applicable (SEE09310380GA), Not Applicable (SEE09310383GA), Not Applicable (SEE10420598GA), Not Applicable (SEE10430640GA), Not Applicable (SEE10450658GA), Not Applicable (SEE10460704GA), Not Applicable (SEE10460707GA), Not Applicable (SEE10460716GA), Not Applicable (SEE10460730GA), Not Applicable (SEE10460736GA), Not Applicable (SEE10460754GA), Not Applicable (SEE11071020GA), Not Applicable (SEE11071021GA), Not Applicable (SEE11071057GA), Not Applicable (SEE12182233GA), Not Applicable (SEE14272218GA), Not Applicable (SEE15251286SA), Not Applicable (SEE15251287SA), Not Applicable (SEE09110156GA), Not Applicable (SEE09170222GA), Not Applicable (SEE09190233GA), Not Applicable (SEE14031632GA), Not Applicable (SEE14031635GA), Not Applicable (SEE14031636GA), Not Applicable (SEE14031638GA), Not Applicable (SEE14031639GA), Not Applicable (SEE14031643GA), Not Applicable (SEE14031647GA), Not Applicable (SEE14031649GA), Not Applicable (SEE14031650GA), Not Applicable (SEE14422542GA), Not Applicable (SEE15193268GA), Not Applicable (SEE15193271GA), Not Applicable (SEE13220827GA), 00840682109260SEE17430030SA (SEE17430030SA), Not Applicable (SEE12430215GA), Not Applicable (SEE12430218GA), Not Applicable (SEE16383417GA), Not Applicable (SEE14272174GA), Not Applicable (SEE14272199GA), Not Applicable (SEE13371084GA), Not Applicable (SEE09020043GA), Not Applicable (SEE09100094GA), Not Applicable (SEE09100101GA), Not Applicable (SEE09100124GA), Not Applicable (SEE09100125GA), Not Applicable (SEE14322319GA), Not Applicable (SEE14322326GA), Not Applicable (SEE14322330GA), Not Applicable (SEE14322331GA), Not Applicable (SEE14322333GA), Not Applicable (SEE14322334GA), 00840682109260SEE15415112SA (SEE15415112SA), 00840682109260SEE15415113SA (SEE15415113SA), 00840682109260SEE15415118SA (SEE15415118SA), 00840682109260SEE18170008SA (SEE18170008SA), Not Applicable (SEE10460695GA), 00840682109260SEE17500016SA (SEE17500016SA), Not Applicable (SEE09340429GA), Not Applicable (SEE09340439GA), Not Applicable (SEE09340445GA), Not Applicable (SEE09340447GA), Not Applicable (SEE11201457GA), Not Applicable (SEE11201466GA), Not Applicable (SEE12370044GA), Not Applicable (SEE12370046GA), Not Applicable (SEE12370047GA), Not Applicable (SEE12370050GA), Not Applicable (SEE12370051GA), Not Applicable (SEE12370052GA), Not Applicable (SEE12370054GA), Not Applicable (SEE12370055GA), Not Applicable (SEE14322322GA), Not Applicable (SEE15173209GA), Not Applicable (SEE15173210GA), Not Applicable (SEE15173216GA), Not Applicable (SEE15173217GA), Not Applicable (SEE15173220GA), Not Applicable (SEE15173221GA), Not Applicable (SEE15173222GA), Not Applicable (SEE15173223GA), Not Applicable (SEE15173224GA), Not Applicable (SEE16109692SA), SEE16410010SA (SEE16410010SA), 00840682109260SEE19090005SA (SEE19090005SA), 00840682109260SEE19090006SA (SEE19090006SA), 00840682109260SEE19090034SA (SEE19090034SA), Not Applicable (SEE16241827SA), Not Applicable (SEE16241828SA), Not Applicable (SEE16241831SA), Not Applicable (SEE16241832SA), Not Applicable (SEE16241833SA), Not Applicable (SEE16241845SA), Not Applicable (SEE16241848SA), Not Applicable (SEE16241849SA), Not Applicable (SEE16241850SA), Not Applicable (SEE16241853SA), Not Applicable (SEE16241854SA), Not Applicable (SEE16241857SA), Not Applicable (SEE16241858SA), Not Applicable (SEE16241859SA), Not Applicable (SEE16251979SA), Not Applicable (SEE16251981SA), Not Applicable (SEE16251982SA), Not Applicable (SEE16251983SA), Not Applicable (SEE16251984SA), Not Applicable (SEE16252017SA), Not Applicable (SEE11151306GA), Not Applicable (SEE11151307GA), Not Applicable (SEE11151314GA), Not Applicable (SEE11161322GA), Not Applicable (SEE15466929SA), Not Applicable (SEE15466930SA), Not Applicable (SEE15466931SA), Not Applicable (SEE15466933SA), Not Applicable (SEE15466936SA), 00840682109260SEE17100001SA (SEE17100001SA), Not Applicable (SEE13431276GA), Not Applicable (SEE13431302GA), Not Applicable (SEE13441306GA), Not Applicable (SEE13451368GA), Not Applicable (SEE13451375GA), Not Applicable (SEE14041727GA), 00840682109260SEE19100050SA (SEE19100050SA), Not Applicable (SEE09530888GA), Not Applicable (SEE10400386GA), Not Applicable (SEE10400393GA), Not Applicable (SEE10400394GA), Not Applicable (SEE10400395GA), Not Applicable (SEE10400398GA), Not Applicable (SEE10400405GA), Not Applicable (SEE10400407GA), Not Applicable (SEE10400408GA), Not Applicable (SEE10400410GA), Not Applicable (SEE10400411GA), Not Applicable (SEE10400412GA), Not Applicable (SEE10400413GA), Not Applicable (SEE10400418GA), Not Applicable (SEE10400423GA), Not Applicable (SEE10400431GA), Not Applicable (SEE11141254GA), Not Applicable (SEE11201460GA), Not Applicable (SEE11211483GA), Not Applicable (SEE11211487GA), Not Applicable (SEE11211488GA), Not Applicable (SEE11211493GA), Not Applicable (SEE11211496GA), Not Applicable (SEE11211510GA), Not Applicable (SEE11211521GA), Not Applicable (SEE11231526GA), Not Applicable (SEE11231533GA), Not Applicable (SEE11231534GA), Not Applicable (SEE11231535GA), Not Applicable (SEE11231538GA), Not Applicable (SEE11231539GA), Not Applicable (SEE11231546GA), Not Applicable (SEE11371781GA), Not Applicable (SEE11371784GA), Not Applicable (SEE12420151GA), Not Applicable (SEE12420152GA), Not Applicable (SEE12420153GA), Not Applicable (SEE12420154GA), Not Applicable (SEE12420156GA), Not Applicable (SEE12420157GA), Not Applicable (SEE12420166GA), Not Applicable (SEE12430177GA), Not Applicable (SEE13431258GA), Not Applicable (SEE13431266GA), Not Applicable (SEE14021555GA), Not Applicable (SEE14021558GA), Not Applicable (SEE14021561GA), Not Applicable (SEE14021568GA), Not Applicable (SEE14021573GA), Not Applicable (SEE14101780GA), Not Applicable (SEE14101781GA), Not Applicable (SEE14101785GA), Not Applicable (SEE14101791GA), Not Applicable (SEE14101797GA), Not Applicable (SEE15281659SA), Not Applicable (SEE15446261SA), Not Applicable (SEE15446333SA), Not Applicable (SEE12340004GA), Not Applicable (SEE12342447GA), Not Applicable (SEE12342450GA), Not Applicable (SEE12342454GA), Not Applicable (SEE09510831GA), Not Applicable (SEE09530856GA), Not Applicable (SEE09530864GA), Not Applicable (SEE09530880GA), Not Applicable (SEE09530886GA), Not Applicable (SEE09530889GA), Not Applicable (SEE09530905GA), Not Applicable (SEE09530910GA), Not Applicable (SEE10260196GA), Not Applicable (SEE11381814GA), Not Applicable (SEE09450624GA), Not Applicable (SEE09450625GA), Not Applicable (SEE09450626GA), Not Applicable (SEE09450628GA), Not Applicable (SEE09450630GA), Not Applicable (SEE09510816GA), Not Applicable (SEE12011894GA), Not Applicable (SEE13451367GA), Not Applicable (SEE13451376GA), Not Applicable (SEE14372413GA), Not Applicable (SEE14041746GA), 00840682109260SEE18070001SA (SEE18070001SA), Not Applicable (SEE11071038GA), Not Applicable (SEE11071041GA), Not Applicable (SEE11071044GA), Not Applicable (SEE11151297GA), 00840682109260SEE18320007SA (SEE18320007SA), 00840682109260SEE19100061SA (SEE19100061SA), Not Applicable (SEE11060952GA), Not Applicable (SEE11060988GA), Not Applicable (SEE14512830GA), Not Applicable (SEE14512833GA), Not Applicable (SEE15012841GA), Not Applicable (SEE15012843GA), Not Applicable (SEE15012846GA), Not Applicable (SEE15012849GA), Not Applicable (SEE15012853GA), Not Applicable (SEE15022860GA), Not Applicable (SEE15022863GA), Not Applicable (SEE15022866GA), 00840682109260SEE17390001SA (SEE17390001SA), 00840682109260SEE17390002SA (SEE17390002SA), 00840682109260SEE17390004SA (SEE17390004SA), 00840682109260SEE17390005SA (SEE17390005SA), 00840682109260SEE17390006SA (SEE17390006SA), 00840682109260SEE17390007SA (SEE17390007SA), 00840682109260SEE18160049SA (SEE18160049SA), 00840682109260SEE18160071SA (SEE18160071SA), 00840682109260SEE18170016SA (SEE18170016SA), 00840682109260SEE18190033SA (SEE18190033SA), 00840682109260SEE18330007SA (SEE18330007SA), 00840682109260SEE18330014SA (SEE18330014SA), 00840682109260SEE18330015SA (SEE18330015SA), 00840682109260SEE19220038SA (SEE19220038SA), Not Applicable (SEE09140178GA), Not Applicable (SEE09150219GA), Not Applicable (SEE09510821GA), Not Applicable (SEE11071072GA), Not Applicable (SEE11071077GA), Not Applicable (SEE11071079GA), Not Applicable (SEE11071133GA), Not Applicable (SEE11071151GA), Not Applicable (SEE11081169GA), Not Applicable (SEE11081172GA), Not Applicable (SEE11081211GA), Not Applicable (SEE11141293GA), Not Applicable (SEE12262328GA), Not Applicable (SEE12262329GA), Not Applicable (SEE12302377GA), Not Applicable (SEE12370049GA), Not Applicable (SEE12370061GA), Not Applicable (SEE12370063GA), Not Applicable (SEE12370068GA), Not Applicable (SEE12400101GA), Not Applicable (SEE12400103GA), Not Applicable (SEE12400104GA), Not Applicable (SEE12400105GA), Not Applicable (SEE12400107GA), Not Applicable (SEE12400109GA), Not Applicable (SEE12400110GA), Not Applicable (SEE12400112GA), Not Applicable (SEE12400114GA), Not Applicable (SEE12400116GA), Not Applicable (SEE12400117GA), Not Applicable (SEE12400118GA), Not Applicable (SEE13020410GA), Not Applicable (SEE13060450GA), Not Applicable (SEE13150693GA), Not Applicable (SEE14021537GA), Not Applicable (SEE15352790SA), Not Applicable (SEE15352792SA), Not Applicable (SEE16282707SA), Not Applicable (SEE16360026SA), Not Applicable (SEE16390008SA), Not Applicable (SEE16410008SA), 00840682109260SEE17100005SA (SEE17100005SA), 00840682109260SEE17370002SA (SEE17370002SA), 00840682109260SEE18170026SA (SEE18170026SA), 00840682109260SEE18450075SA (SEE18450075SA), 00840682109260SEE19010008SA (SEE19010008SA), 00840682109260SEE19130010SA (SEE19130010SA), Not Applicable (SEE13150662GA), Not Applicable (SEE13150682GA), Not Applicable (SEE13371082GA), Not Applicable (SEE13371105GA), 00840682109260SEE19090010SA (SEE19090010SA), 00840682109260SEE19090011SA (SEE19090011SA), 00840682109260SEE19090016SA (SEE19090016SA), 00840682109260SEE19090017SA (SEE19090017SA), 00840682109260SEE19090020SA (SEE19090020SA), 00840682109260SEE19090033SA (SEE19090033SA), Not Applicable (SEE14031618GA), Not Applicable (SEE15332340SA), Not Applicable (SEE15332342SA), Not Applicable (SEE11081167GA), Not Applicable (SEE13180763GA), Not Applicable (SEE09100135GA), Not Applicable (SEE15093046GA), Not Applicable (SEE14472672GA), Not Applicable (SEE09510789GA), Not Applicable (SEE12490263GA), Not Applicable (SEE15477107SA), Not Applicable (SEE10430625GA), Not Applicable (SEE10430634GA), Not Applicable (SEE10460741GA), Not Applicable (SEE15394729SA), Not Applicable (SEE09510785GA), Not Applicable (SEE11060993GA), Not Applicable (SEE11061001GA), Not Applicable (SEE12322407GA), Not Applicable (SEE12322410GA), Not Applicable (SEE15261536SA), Not Applicable (SEE16140435SA), Not Applicable (SEE16150494SA), Not Applicable (SEE16150496SA), Not Applicable (SEE16150499SA), Not Applicable (SEE16150502SA), Not Applicable (SEE16272269SA), 00840682109260SEE18320026SA (SEE18320026SA), 00840682109260SEE18320027SA (SEE18320027SA), 00840682109260SEE18320031SA (SEE18320031SA), Not Applicable (SEE16241823SA), Not Applicable (SEE12292363GA), Not Applicable (SEE12390072GA), Not Applicable (SEE12390075GA), Not Applicable (SEE15173208GA), Not Applicable (SEE15173215GA), Not Applicable (SEE16420041SA), Not Applicable (SEE16420042SA), Not Applicable (SEE16420043SA), Not Applicable (SEE16420044SA), Not Applicable (SEE16420045SA), Not Applicable (SEE16420046SA), Not Applicable (SEE16420048SA), Not Applicable (SEE16420049SA), Not Applicable (SEE16420050SA), 00840682109260SEE18240016SA (SEE18240016SA), 00840682109260SEE18320002SA (SEE18320002SA), 00840682109260SEE18320012SA (SEE18320012SA), 00840682109260SEE18320015SA (SEE18320015SA), 00840682109260SEE18320018SA (SEE18320018SA), 00840682109260SEE18320019SA (SEE18320019SA), 00840682109260SEE18320022SA (SEE18320022SA), 00840682109260SEE18320023SA (SEE18320023SA), 00840682109260SEE18330009SA (SEE18330009SA), 00840682109260SEE18330016SA (SEE18330016SA), Not Applicable (SEE10120030GA), Not Applicable (SEE14272147GA), Not Applicable (SEE14272149GA), Not Applicable (SEE14272150GA), Not Applicable (SEE14272151GA), Not Applicable (SEE14272152GA), Not Applicable (SEE14272154GA), Not Applicable (SEE14272155GA), Not Applicable (SEE16272588SA), Not Applicable (SEE16272594SA), Not Applicable (SEE16272595SA), Not Applicable (SEE09190236GA), Not Applicable (SEE09190246GA), Not Applicable (SEE13220846GA), Not Applicable (SEE16120170SA), Not Applicable (SEE14422541GA), Not Applicable (SEE13451309GA), Not Applicable (SEE09460639GA), Not Applicable (SEE09460640GA), Not Applicable (SEE09460643GA), Not Applicable (SEE09460644GA), Not Applicable (SEE09470665GA), Not Applicable (SEE12172195GA), Not Applicable (SEE12530374GA), Not Applicable (SEE13000375GA), Not Applicable (SEE13000379GA), Not Applicable (SEE13000381GA), Not Applicable (SEE13000382GA), Not Applicable (SEE14432574GA), Not Applicable (SEE11211497GA), Not Applicable (SEE14151834GA), Not Applicable (SEE14151866GA), Not Applicable (SEE12340017GA), Not Applicable (SEE12400091GA), Not Applicable (SEE12400092GA), 00840682109260SEE17180005SA (SEE17180005SA), Not Applicable (SEE14171919GA), Not Applicable (SEE12490264GA), Not Applicable (SEE12490269GA), Not Applicable (SEE12490272GA), Not Applicable (SEE12490280GA), Not Applicable (SEE16109590SA), Not Applicable (SEE16119910SA), Not Applicable (SEE16430008SA), Not Applicable (SEE11111240GA), Not Applicable (SEE09470677GA), Not Applicable (SEE09480744GA), Not Applicable (SEE09490755GA), Not Applicable (SEE15103105GA), Not Applicable (SEE15103113GA), Not Applicable (SEE15103115GA), Not Applicable (SEE14242081GA), 00840682109260SEE19240043SA (SEE19240043SA), Not Applicable (SEE10270245GA), Not Applicable (SEE14031678GA), Not Applicable (SEE14031685GA), Not Applicable (SEE14031702GA), 00840682109260SEE18190036SA (SEE18190036SA), 00840682109260SEE18190040SA (SEE18190040SA), Not Applicable (SEE11141260GA), Not Applicable (SEE10420579GA), Not Applicable (SEE10420576GA), Not Applicable (SEE10460719GA), Not Applicable (SEE10460734GA), Not Applicable (SEE16334162SA), Not Applicable (SEE09100127GA), Not Applicable (SEE09230254GA), Not Applicable (SEE09240293GA), Not Applicable (SEE10460724GA), Not Applicable (SEE14031621GA), Not Applicable (SEE10460690GA), Not Applicable (SEE14211979GA), Not Applicable (SEE15032876GA), Not Applicable (SEE09310376GA), Not Applicable (SEE10270263GA), Not Applicable (SEE10400419GA), Not Applicable (SEE10400438GA), Not Applicable (SEE10400446GA), Not Applicable (SEE10410477GA), Not Applicable (SEE10420561GA), Not Applicable (SEE10470789GA), Not Applicable (SEE11071018GA), Not Applicable (SEE11111231GA), Not Applicable (SEE11111241GA), Not Applicable (SEE11511875GA), Not Applicable (SEE12112110GA), Not Applicable (SEE12112116GA), Not Applicable (SEE13240868GA), Not Applicable (SEE13290973GA), Not Applicable (SEE15113128GA), Not Applicable (SEE15113131GA), Not Applicable (SEE15183260GA), Not Applicable (SEE16119844SA), Not Applicable (SEE16282862SA), Not Applicable (SEE15173236GA), Not Applicable (SEE15173243GA), Not Applicable (SEE14512821GA), Not Applicable (SEE09490761GA), Not Applicable (SEE11211520GA), Not Applicable (SEE11231527GA), Not Applicable (SEE13381121GA), Not Applicable (SEE13391133GA), Not Applicable (SEE13391136GA), Not Applicable (SEE16211411SA), Not Applicable (SEE09510792GA), Not Applicable (SEE09510823GA), Not Applicable (SEE12162178GA), Not Applicable (SEE12162180GA), Not Applicable (SEE12162184GA), Not Applicable (SEE13381126GA), Not Applicable (SEE13381127GA), Not Applicable (SEE13391132GA), Not Applicable (SEE13280938GA), Not Applicable (SEE13280942GA), Not Applicable (SEE13280944GA), Not Applicable (SEE13280950GA), Not Applicable (SEE09460648GA), Not Applicable (SEE11161337GA), Not Applicable (SEE10420566GA), Not Applicable (SEE10420599GA), Not Applicable (SEE10430610GA), Not Applicable (SEE10430622GA), Not Applicable (SEE10430647GA), Not Applicable (SEE10450664GA), Not Applicable (SEE11371769GA), Not Applicable (SEE12400094GA), Not Applicable (SEE13451316GA), Not Applicable (SEE13451319GA), Not Applicable (SEE13451330GA), Not Applicable (SEE13451338GA), Not Applicable (SEE13451339GA), Not Applicable (SEE13451340GA), Not Applicable (SEE13451343GA), Not Applicable (SEE13451351GA), Not Applicable (SEE13451352GA), Not Applicable (SEE13451354GA), Not Applicable (SEE13451355GA), Not Applicable (SEE13451356GA), Not Applicable (SEE13451357GA), Not Applicable (SEE13451358GA), Not Applicable (SEE13451359GA), Not Applicable (SEE13451360GA), Not Applicable (SEE13451361GA), Not Applicable (SEE13451362GA), Not Applicable (SEE13451363GA), Not Applicable (SEE13451364GA), Not Applicable (SEE13451365GA), Not Applicable (SEE13451366GA), Not Applicable (SEE14472681GA), 00840682109260SEE19100045SA (SEE19100045SA), 00840682109260SEE19100059SA (SEE19100059SA), Not Applicable (SEE09100093GA), Not Applicable (SEE09100121GA), Not Applicable (SEE12420161GA), Not Applicable (SEE12490288GA), Not Applicable (SEE12490289GA), Not Applicable (SEE12490290GA), Not Applicable (SEE12490293GA), Not Applicable (SEE12490294GA), Not Applicable (SEE12490298GA), Not Applicable (SEE14021506GA), Not Applicable (SEE10390315GA), Not Applicable (SEE10390327GA), Not Applicable (SEE10390332GA), Not Applicable (SEE10390333GA), Not Applicable (SEE12322409GA), Not Applicable (SEE13451328GA), Not Applicable (SEE13451347GA), Not Applicable (SEE13491449GA), Not Applicable (SEE13501462GA), Not Applicable (SEE13501473GA), Not Applicable (SEE13501476GA), Not Applicable (SEE14041756GA), Not Applicable (SEE15032898GA), Not Applicable (SEE15507700SA), Not Applicable (SEE15507703SA), Not Applicable (SEE16078898SA), Not Applicable (SEE12122133GA), Not Applicable (SEE09130162GA), Not Applicable (SEE09130166GA), Not Applicable (SEE09460652GA), Not Applicable (SEE09470683GA), Not Applicable (SEE09480729GA), Not Applicable (SEE09530870GA), Not Applicable (SEE10130060GA), Not Applicable (SEE10250079GA), Not Applicable (SEE10250080GA), Not Applicable (SEE10250097GA), Not Applicable (SEE10250109GA), Not Applicable (SEE10250113GA), Not Applicable (SEE10250114GA), Not Applicable (SEE10250115GA), Not Applicable (SEE10250117GA), Not Applicable (SEE10260194GA), Not Applicable (SEE10260205GA), Not Applicable (SEE10260211GA), Not Applicable (SEE10260235GA), Not Applicable (SEE10270240GA), Not Applicable (SEE10270243GA), Not Applicable (SEE10270247GA), Not Applicable (SEE10310279GA), Not Applicable (SEE10410464GA), Not Applicable (SEE11071065GA), Not Applicable (SEE11171366GA), Not Applicable (SEE11171370GA), Not Applicable (SEE11171375GA), Not Applicable (SEE11171378GA), Not Applicable (SEE11191427GA), Not Applicable (SEE11191430GA), Not Applicable (SEE11341651GA), Not Applicable (SEE11361733GA), Not Applicable (SEE11361736GA), Not Applicable (SEE11361738GA), Not Applicable (SEE11361741GA), Not Applicable (SEE11361744GA), Not Applicable (SEE12122136GA), Not Applicable (SEE12122145GA), Not Applicable (SEE12400106GA), Not Applicable (SEE12420172GA), Not Applicable (SEE13110513GA), Not Applicable (SEE13110514GA), Not Applicable (SEE13110520GA), Not Applicable (SEE13110521GA), Not Applicable (SEE13280937GA), Not Applicable (SEE13421239GA), Not Applicable (SEE13421245GA), Not Applicable (SEE13421248GA), Not Applicable (SEE13421249GA), Not Applicable (SEE13421250GA), Not Applicable (SEE13421252GA), Not Applicable (SEE13501459GA), Not Applicable (SEE13501461GA), Not Applicable (SEE13501466GA), Not Applicable (SEE14041731GA), Not Applicable (SEE14211972GA), Not Applicable (SEE14222036GA), Not Applicable (SEE14292245GA), Not Applicable (SEE14302255GA), Not Applicable (SEE14302259GA), Not Applicable (SEE14302261GA), Not Applicable (SEE14302262GA), Not Applicable (SEE14302264GA), Not Applicable (SEE14302266GA), Not Applicable (SEE14302268GA), Not Applicable (SEE14302269GA), Not Applicable (SEE14302270GA), Not Applicable (SEE14302271GA), Not Applicable (SEE14302272GA), Not Applicable (SEE14302275GA), Not Applicable (SEE14302277GA), Not Applicable (SEE14302278GA), Not Applicable (SEE14302279GA), Not Applicable (SEE14302283GA), Not Applicable (SEE14302284GA), Not Applicable (SEE14302285GA), Not Applicable (SEE14302286GA), Not Applicable (SEE14302287GA), Not Applicable (SEE14372409GA), Not Applicable (SEE14372425GA), Not Applicable (SEE14372427GA), Not Applicable (SEE14372429GA), Not Applicable (SEE14372436GA), Not Applicable (SEE14382463GA), Not Applicable (SEE14382470GA), Not Applicable (SEE14402481GA), Not Applicable (SEE14402484GA), Not Applicable (SEE14402485GA), Not Applicable (SEE14402486GA), Not Applicable (SEE14402488GA), Not Applicable (SEE14402492GA), Not Applicable (SEE14432575GA), Not Applicable (SEE15032880GA), Not Applicable (SEE15220795SA), Not Applicable (SEE15261579SA), Not Applicable (SEE15383875SA), Not Applicable (SEE16109585SA), Not Applicable (SEE16150493SA), Not Applicable (SEE16324002SA), 00840682109260SEE19090019SA (SEE19090019SA), 00840682109260SEE19130006SA (SEE19130006SA), Not Applicable (SEE12062003GA), Not Applicable (SEE16140433SA), 00840682109260SEE15103097GA (SEE15103097GA), 00840682109260SEE15103098GA (SEE15103098GA), 00840682109260SEE15103099GA (SEE15103099GA), 00840682109260SEE15103101GA (SEE15103101GA), 00840682109260SEE15103106GA (SEE15103106GA), 00840682109260SEE15103107GA (SEE15103107GA), 00840682109260SEE15103111GA (SEE15103111GA), 00840682109260SEE15103114GA (SEE15103114GA), Not Applicable (SEE16191064SA), 00840682109260SEE19130012SA (SEE19130012SA), 00840682109260SEE19130014SA (SEE19130014SA), Not Applicable (SEE12322433GA), Not Applicable (SEE12282352GA), Not Applicable (SEE15446226SA), Not Applicable (SEE16450008SA), Not Applicable (SEE16450010SA), Not Applicable (SEE12350027GA), Not Applicable (SEE15332307SA), 00840682109260SEE18330018SA (SEE18330018SA), Not Applicable (SEE09100107GA), Not Applicable (SEE13150674GA), Not Applicable (SEE14031615GA), Not Applicable (SEE14031640GA), Not Applicable (SEE11050945GA), Not Applicable (SEE10420549GA), Not Applicable (SEE14372402GA), Not Applicable (SEE10400445GA), Not Applicable (SEE15312080SA), Not Applicable (SEE11181405GA), Not Applicable (SEE14151810GA), Not Applicable (SEE15032896GA), Not Applicable (SEE10460701GA), Not Applicable (SEE11071087GA), 00840682109260SEE18160029SA (SEE18160029SA), Not Applicable (SEE10250130GA), Not Applicable (SEE09510794GA), Not Applicable (SEE09510814GA), 00840682109260SEE19220027SA (SEE19220027SA), Not Applicable (SEE09490757GA), Not Applicable (SEE11191431GA), Not Applicab le (SEE12092072GA), Not Applicable (SEE09040062GA), Not Applicable (SEE09040068GA), Not Applicable (SEE12340014GA), Not Applicable (SEE12342455GA), Not Applicable (SEE12350022GA), Not Applicable (SEE12350025GA), Not Applicable (SEE12350028GA), Not Applicable (SEE12350030GA), Not Applicable (SEE13290977GA), Not Applicable (SEE14272139GA), Not Applicable (SEE15183248GA), Not Applicable (SEE15183252GA), 00840682109260SEE18320030SA (SEE18320030SA), Not Applicable (SEE11040915GA), Not Applicable (SEE10420565GA), Not Applicable (SEE16364419SA), Not Applicable (SEE10250107GA), Not Applicable (SEE10250129GA), Not Applicable (SEE10260200GA), Not Applicable (SEE10260217GA), Not Applicable (SEE10260218GA), Not Applicable (SEE10260221GA), Not Applicable (SEE15332398SA), Not Applicable (SEE16221696SA), 00840682109260SEE17510011SA (SEE17510011SA), 00840682109260SEE18510001SA (SEE18510001SA), 00840682109260SEE18510010SA (SEE18510010SA), Not Applicable (SEE16109690SA), Not Applicable (SEE16109702SA), Not Applicable (SEE14312289GA), Not Applicable (SEE13240867GA), Not Applicable (SEE13240871GA), Not Applicable (SEE14412501GA), Not Applicable (SEE12242292GA), Not Applicable (SEE12252320GA), Not Applicable (SEE12252323GA), Not Applicable (SEE12252325GA), Not Applicable (SEE15173238GA), 00840682109260SEE17210014SA (SEE17210014SA), 00840682109260SEE17290001SA (SEE17290001SA), 00840682109260SEE17290002SA (SEE17290002SA), Not Applicable (SEE12122134GA), Not Applicable (SEE10530850GA), Not Applicable (SEE12390078GA), Not Applicable (SEE11050929GA), Not Applicable (SEE11201446GA), Not Applicable (SEE11211500GA), Not Applicable (SEE09370496GA), Not Applicable (SEE09370500GA), Not Applicable (SEE09370508GA), Not Applicable (SEE14372423GA), Not Applicable (SEE14372426GA), Not Applicable (SEE14372428GA), Not Applicable (SEE14372430GA), Not Applicable (SEE14372433GA), Not Applicable (SEE14372435GA), Not Applicable (SEE14422535GA), Not Applicable (SEE14422569GA), Not Applicable (SEE14432611GA), Not Applicable (SEE14432612GA), Not Applicable (SEE11060995GA), Not Applicable (SEE11141274GA), Not Applicable (SEE11141281GA), Not Applicable (SEE11371751GA), Not Applicable (SEE14482698GA), Not Applicable (SEE10460713GA), Not Applicable (SEE14031717GA), Not Applicable (SEE19110019SA), Not Applicable (SEE19110040SA), Not Applicable (SEE16333410GA), 00840682109260SEE18200013SA (SEE18200013SA), Not Applicable (SEE13290967GA), Not Applicable (SEE15413316GA), 00840682109260SEE17150002SA (SEE17150002SA), Not Applicable (SEE13290983GA), Not Applicable (SEE13290986GA), Not Applicable (SEE13351018GA), Not Applicable (SEE14462651GA), Not Applicable (SEE14462658GA), Not Applicable (SEE16303407GA), Not Applicable (SEE13180775GA), Not Applicable (SEE13180780GA), Not Applicable (SEE13180781GA), Not Applicable (SEE15193282GA), Not Applicable (SEE16420075SA), 00840682109260SEE17210013SA (SEE17210013SA), 00840682109260SEE17310005SA (SEE17310005SA), 00840682109260SEE18500015SA (SEE18500015SA), 00840682109260SEE18160038SA (SEE18160038SA), 00840682109260SEE18160039SA (SEE18160039SA), 00840682109260SEE19170002SA (SEE19170002SA), Not Applicable (SEE15153203GA), Not Applicable (SEE15193272GA), 00840682109260SEE18510028SA (SEE18510028SA), Not Applicable (SEE12122140GA), 00840682109260SEE18160026SA (SEE18160026SA), 00840682109260SEE18160037SA (SEE18160037SA), Not Applicable (SEE14222002GA), 00840682109260SEE17070002SA (SEE17070002SA), 00840682109260SEE18460004SA (SEE18460004SA), 00840682109260SEE18460011SA (SEE18460011SA), Not Applicable (SEE14021562GA), Not Applicable (SEE10420523GA), Not Applicable (SEE12112108GA), Not Applicable (SEE11341656GA), Not Applicable (SEE11341656SA), Not Applicable (SEE10470800GA), Not Applicable (SEE11071134GA), Not Applicable (SEE11371787GA), Not Applicable (SEE11191428GA), Not Applicable (SEE09100122GA), Not Applicable (SEE09190237GA), Not Applicable (SEE12370060GA), Not Applicable (SEE12400097GA), Not Applicable (SEE14151819GA), Not Applicable (SEE10130047GA), Not Applicable (SEE12490243GA), Not Applicable (SEE14021544GA), Not Applicable (SEE12360034GA), Not Applicable (SEE13270898GA), Not Applicable (SEE13280912GA), Not Applicable (SEE16272628SA), Not Applicable (SEE16282787SA), Not Applicable (SEE16282789SA), Not Applicable (SEE15423320GA), Not Applicable (SEE14101783GA), Not Applicable (SEE15183256GA), 00840682109260SEE17150005SA (SEE17150005SA), Not Applicable (SEE09100143GA), Not Applicable (SEE13351031GA), Not Applicable (SEE12430201GA), 00840682109260SEE17210008SA (SEE17210008SA), Not Applicable (SEE13451334GA), 00840682109260SEE17410006SA (SEE17410006SA), Not Applicable (SEE15423321GA), Not Applicable (SEE15423319GA), Not Applicable (SEE15423323GA), Not Applicable (SEE14021545GA), Not Applicable (SEE14021549GA), Not Applicable (SEE14222020GA), Not Applicable (SEE16430050SA), Not Applicable (SEE13140655GA), 00840682109260SEE18160025SA (SEE18160025SA), 00840682109260SEE18160033SA (SEE18160033SA), Not Applicable (SEE15373309GA), Not Applicable (SEE16123366GA), Not Applicable (SEE16420065SA), Not Applicable (SEE14452627GA), Not Applicable (SEE16183379GA), Not Applicable (SEE09530895GA), Not Applicable (SEE09530913GA), Not Applicable (SEE14242103GA), Not Applicable (SEE14252134GA), 00840682109260SEE18160034SA (SEE18160034SA), 00840682109260SEE18220008SA (SEE18220008SA), 00840682109260SEE18220009SA (SEE18220009SA), Not Applicable (SEE14161894GA), Not Applicable (SEE16390004SA), 00840682109260SEE18510005SA (SEE18510005SA), Not Applicable (SEE12092052GA), 00840682109260SEE17160004SA (SEE17160004SA), Not Applicable (SEE13000394GA), Not Applicable (SEE12162182GA), Not Applicable (SEE10420578GA), 00840682109260SEE19220022SA (SEE19220022SA), Not Applicable (SEE15012850GA), Not Applicable (SEE12430205GA), Not Applicable (SEE12490287GA), Not Applicable (SEE13481407GA), Not Applicable (SEE13501464GA), Not Applicable (SEE10400400GA), Not Applicable (SEE16173377GA), Not Applicable (SEE16183378GA), Not Applicable (SEE14502761GA), Not Applicable (SEE13371096GA), Not Applicable (SEE15423328GA), Not Applicable (SEE15423329GA), Not Applicable (SEE15423330GA), Not Applicable (SEE16123367GA), Not Applicable (SEE16173375GA), Not Applicable (SEE09330414GA), Not Applicable (SEE16450003SA), Not Applicable (SEE16173370GA), Not Applicable (SEE15193281GA), Not Applicable (SEE09310361GA), Not Applicable (SEE09310386GA), Not Applicable (SEE09310388GA), Not Applicable (SEE10410471GA), Not Applicable (SEE12082032GA), Not Applicable (SEE12082034GA), Not Applicable (SEE16440003SA), Not Applicable (SEE09140179GA), Not Applicable (SEE13481418GA), Not Applicable (SEE14031619GA), Not Applicable (SEE14302263GA), Not Applicable (SEE14472677GA), Not Applicable (SEE15193277GA), Not Applicable (SEE15193283GA), 00840682109260SEE17060003SA (SEE17060003SA), 00840682109260SEE17400006SA (SEE17400006SA), 00840682109260SEE17500022SA (SEE17500022SA), 00840682109260SEE18190031SA (SEE18190031SA), 00840682109260SEE18380034SA (SEE18380034SA), Not Applicable (SEE16173372GA), 00840682109260SEE18070002SA (SEE18070002SA), 00840682109260SEE18190039SA (SEE18190039SA), 00840682109260SEE18200008SA (SEE18200008SA), 00840682109260SEE18200009SA (SEE18200009SA), 00840682109260SEE18200010SA (SEE18200010SA), 00840682109260SEE18200011SA (SEE18200011SA), 00840682109260SEE18200012SA (SEE18200012SA), Not Applicable (SEE10250174GA), Not Applicable (SEE09370511GA), Not Applicable (SEE12370045GA), Not Applicable (SEE12360038GA), Not Applicable (SEE11361682GA), Not Applicable (SEE11391835GA), Not Applicable (SEE13180790GA), Not Applicable (SEE09190225GA), Not Applicable (SEE11361673GA), Not Applicable (SEE11151300GA), Not Applicable (SEE11511871GA), Not Applicable (SEE12122128GA), Not Applicable (SEE12370062GA), Not Applicable (SEE12182234GA), Not Applicable (SEE13361076GA), Not Applicable (SEE13451369GA), Not Applicable (SEE13451374GA), Not Applicable (SEE14151822GA), Not Applicable (SEE14322325GA), Not Applicable (SEE14332341GA), Not Applicable (SEE14422559GA), Not Applicable (SEE15312143SA), Not Applicable (SEE15473332GA), Not Applicable (SEE15473334GA), Not Applicable (SEE16173369GA), Not Applicable (SEE16450001SA), 00840682109260SEE17110006SA (SEE17110006SA), Not Applicable (SEE18390015SA), 00840682109260SEE18420017SA (SEE18420017SA), Not Applicable (SEE11111245GA), Not Applicable (SEE12192236GA), Not Applicable (SEE13270890GA), Not Applicable (SEE14021542GA), 00840682109260SEE17420003SA (SEE17420003SA), Not Applicable (SEE09340436GA), Not Applicable (SEE10260199GA), Not Applicable (SEE10270267GA), Not Applicable (SEE10420601GA), Not Applicable (SEE10430613GA), Not Applicable (SEE10430643GA), Not Applicable (SEE10460721GA), Not Applicable (SEE10460770GA), Not Applicable (SEE11231528GA), Not Applicable (SEE13020408GA), 00840682109260SEE18170027SA (SEE18170027SA), Not Applicable (SEE12420165GA), Not Applicable (SEE14101779GA), Not Applicable (SEE11371767GA), Not Applicable (SEE14482704GA), Not Applicable (SEE10120998GA), Not Applicable (SEE10130072GA), Not Applicable (SEE12252321GA), Not Applicable (SEE12162181GA), Not Applicable (SEE12282359GA), Not Applicable (SEE13361067GA), Not Applicable (SEE14432576GA), Not Applicable (SEE14432577GA), Not Applicable (SEE14452628GA), Not Applicable (SEE15052946GA), 00840682109260SEE17020030SA (SEE17020030SA), Not Applicable (SEE11201450GA), Not Applicable (SEE11361726GA), Not Applicable (SEE12340006GA), Not Applicable (SEE12342448GA), Not Applicable (SEE13401151GA), Not Applicable (SEE13401172GA), 00840682109260SEE18320017SA (SEE18320017SA), 00840682109260SEE18030019SA (SEE18030019SA), Not Applicable (SEE15113183GA), Not Applicable (SEE13401163GA), Not Applicable (SEE13401178GA), Not Applicable (SEE09460631GA), Not Applicable (SEE10530840GA), Not Applicable (SEE13190791GA), Not Applicable (SEE13451311GA), Not Applicable (SEE13451323GA), Not Applicable (SEE13451324GA), Not Applicable (SEE14382468GA), 00840682109260SEE17450007SA (SEE17450007SA), 00840682109260SEE19120024SA (SEE19120024SA), Not Applicable (SEE12370070GA), Not Applicable (SEE13270883GA), 00840682109260SEE17350010SA (SEE17350010SA), Not Applicable (SEE12302395GA), Not Applicable (SEE09470660GA), Not Applicable (SEE09470662GA), Not Applicable (SEE09470675GA), Not Applicable (SEE09470679GA), Not Applicable (SEE12302380GA), Not Applicable (SEE10430619GA), Not Applicable (SEE10430624GA), Not Applicable (SEE10430636GA), Not Applicable (SEE14031624GA), Not Applicable (SEE14031633GA), Not Applicable (SEE14081769GA), Not Applicable (SEE14081774GA), 00840682109260SEE17080001SA (SEE17080001SA), 00840682109260SEE18280004SA (SEE18280004SA), 00840682109260SEE18290006SA (SEE18290006SA), 00840682109260SEE18290015SA (SEE18290015SA), Not Applicable (SEE16053354GA), Not Applicable (SEE16053355GA), Not Applicable (SEE15473339GA), Not Applicable (SEE09530898GA), Not Applicable (SEE15063006GA), Not Applicable (SEE12490276GA), Not Applicable (SEE15052940GA), Not Applicable (SEE11161328GA), Not Applicable (SEE12440232GA), Not Applicable (SEE09510813GA), Not Applicable (SEE11081177GA), Not Applicable (SEE11081197GA), Not Applicable (SEE15173219GA), Not Applicable (SEE15193301GA), Not Applicable (SEE16292870SA), Not Applicable (SEE10400388GA), Not Applicable (SEE10400390GA), Not Applicable (SEE10430654GA), Not Applicable (SEE11050919GA), Not Applicable (SEE11050935GA), Not Applicable (SEE11050936GA), Not Applicable (SEE11050937GA), Not Applicable (SEE11071071GA), Not Applicable (SEE11151303GA), Not Applicable (SEE11151310GA), Not Applicable (SEE12420170GA), Not Applicable (SEE12510305GA), Not Applicable (SEE13431272GA), Not Applicable (SEE13431300GA), Not Applicable (SEE12420160GA), Not Applicable (SEE12031942GA), 00840682109260SEE18320009SA (SEE18320009SA), Not Applicable (SEE09330418GA), Not Applicable (SEE14412507GA), Not Applicable (SEE14452619GA), Not Applicable (SEE14462650GA), Not Applicable (SEE15193284GA), 00840682109260SEE18330036SA (SEE18330036SA), 00840682109260SEE18330038SA (SEE18330038SA), 00840682109260SEE18330041SA (SEE18330041SA), 00840682109260SEE18330042SA (SEE18330042SA), 00840682109260SEE18330043SA (SEE18330043SA), Not Applicable (SEE15052950GA), Not Applicable (SEE15062992GA), Not Applicable (SEE16324029SA), Not Applicable (SEE14422519GA), Not Applicable (SEE11071137GA), Not Applicable (SEE11081185GA), Not Applicable (SEE11081178GA), Not Applicable (SEE12092045GA), Not Applicable (SEE12092047GA), Not Applicable (SEE12092049GA), Not Applicable (SEE12420162GA), Not Applicable (SEE12490283GA), Not Applicable (SEE13150665GA), Not Applicable (SEE13150675GA), Not Applicable (SEE14482706GA), Not Applicable (SEE14482715GA), Not Applicable (SEE14482717GA), Not Applicable (SEE14482718GA), Not Applicable (SEE14482720GA), 01008406821092601118020921SEE18060023SA (SEE18060023SA), Not Applicable (SEE13431287GA), Not Applicable (SEE11081191GA), Not Applicable (SEE11050927GA), Not Applicable (SEE10250103GA), Not Applicable (SEE11030897GA), Not Applicable (SEE11071061GA), Not Applicable (SEE13240856GA), Not Applicable (SEE15032884GA), Not Applicable (SEE15032887GA), Not Applicable (SEE15113174GA), Not Applicable (SEE14462647GA), Not Applicable (SEE14462653GA), Not Applicable (SEE14472678GA), Not Applicable (SEE10090947GA), Not Applicable (SEE09470664GA), Not Applicable (SEE09530897GA), Not Applicable (SEE10400436GA), Not Applicable (SEE11341640GA), Not Applicable (SEE13451314GA), Not Applicable (SEE13451348GA), Not Applicable (SEE13451370GA), Not Applicable (SEE13451371GA), Not Applicable (SEE13451372GA), Not Applicable (SEE13451373GA), Not Applicable (SEE14031595GA), Not Applicable (SEE14101788GA), Not Applicable (SEE14101789GA), Not Applicable (SEE14111801GA), Not Applicable (SEE14161905GA), Not Applicable (SEE16313483SA), Not Applicable (SEE16313484SA), Not Applicable (SEE16430006SA), 00840682109260SEE18220014SA (SEE18220014SA), 00840682109260SEE18220015SA (SEE18220015SA), Not Applicable (SEE13110505GA), Not Applicable (SEE13000389GA), Not Applicable (SEE09310363GA), Not Applicable (SEE11171380GA), Not Applicable (SEE12322411GA), Not Applicable (SEE13150676GA), Not Applicable (SEE13150673GA), Not Applicable (SEE13150691GA), Not Applicable (SEE14031655GA), Not Applicable (SEE14031665GA), Not Applicable (SEE12172197GA), Not Applicable (SEE16430001SA), Not Applicable (SEE16430025SA), Not Applicable (SEE14302256GA), Not Applicable (SEE12092082GA), Not Applicable (SEE09460653GA), Not Applicable (SEE13481405GA), Not Applicable (SEE11361669GA), Not Applicable (SEE12051992GA), Not Applicable (SEE13050440GA), Not Applicable (SEE13180738GA), Not Applicable (SEE11111232GA), Not Applicable (SEE09470655GA), Not Applicable (SEE11361700GA), Not Applicable (SEE11161332GA), Not Applicable (SEE10400379GA), Not Applicable (SEE12082028GA), Not Applicable (SEE13180740GA), Not Applicable (SEE09510771GA), Not Applicable (SEE10120009GA), Not Applicable (SEE10120031GA), Not Applicable (SEE10121017GA), Not Applicable (SEE11371808GA), Not Applicable (SEE12332440GA), Not Applicable (SEE13180731GA), Not Applicable (SEE13180732GA), Not Applicable (SEE13180736GA), Not Applicable (SEE10410488GA), Not Applicable (SEE10460750GA), Not Applicable (SEE10470785GA), Not Applicable (SEE13150672GA), Not Applicable (SEE13180737GA), Not Applicable (SEE12430199GA), Not Applicable (SEE12430182GA), Not Applicable (SEE13150679GA), Not Applicable (SEE11331592GA), Not Applicable (SEE10460745GA), Not Applicable (SEE13180730GA), Not Applicable (SEE13150677GA), Not Applicable (SEE13180721GA), Not Applicable (SEE13180727GA), Not Applicable (SEE13180734GA), Not Applicable (SEE13100478GA), Not Applicable (SEE16233398GA), Not Applicable (SEE16383414GA), Not Applicable (SEE16390022SA), Not Applicable (SEE18400001SA), Not Applicable (SEE09230283GA), Not Applicable (SEE09390533GA), Not Applicable (SEE09460645GA), Not Applicable (SEE10410504GA), Not Applicable (SEE10420581GA), Not Applicable (SEE10430615GA), Not Applicable (SEE10430618GA), Not Applicable (SEE10460717GA), Not Applicable (SEE10460732GA), Not Applicable (SEE10470782GA), Not Applicable (SEE10470783GA), Not Applicable (SEE10470792GA), Not Applicable (SEE10470794GA), Not Applicable (SEE11341664GA), Not Applicable (SEE11361734GA), Not Applicable (SEE12122152GA), Not Applicable (SEE12122160GA), Not Applicable (SEE12122164GA), Not Applicable (SEE13040424GA), Not Applicable (SEE13110548GA), Not Applicable (SEE13180789GA), Not Applicable (SEE13401173GA), Not Applicable (SEE14031593GA), 00840682109260SEE18320003SA (SEE18320003SA), Not Applicable (SEE18420008SA), Not Applicable (SEE15383313GA), Not Applicable (SEE15413318GA), 00840682109260SEE17340017SA (SEE17340017SA), 00840682109260SEE18190028SA (SEE18190028SA), 00840682109260SEE18320006SA (SEE18320006SA), 00840682109260SEE18380025SA (SEE18380025SA), 00840682109260SEE18380026SA (SEE18380026SA), 00840682109260SEE18380028SA (SEE18380028SA), 00840682109260SEE18380030SA (SEE18380030SA), Not Applicable (SEE15062983GA), Not Applicable (SEE15173234GA), Not Applicable (SEE15183263GA), Not Applicable (SEE15193293GA), Not Applicable (SEE15312130SA), Not Applicable (SEE16252020SA), 00840682109260SEE16520001SA (SEE16520001SA), 00840682109260SEE17360008SA (SEE17360008SA), 00840682109260SEE17460020SA (SEE17460020SA), Not Applicable (SEE12252319GA), Not Applicable (SEE12252324GA), Not Applicable (SEE14031718GA), Not Applicable (SEE12262340GA), Not Applicable (SEE14021538GA), Not Applicable (SEE14021556GA), Not Applicable (SEE11321589GA), Not Applicable (SEE13371089GA), Not Applicable (SEE16313485SA), Not Applicable (SEE15373307GA), Not Applicable (SEE15312128SA), Not Applicable (SEE15373308GA), 00840682109260SEE17210018SA (SEE17210018SA), Not Applicable (SEE15093038GA), Not Applicable (SEE15373310GA), Not Applicable (SEE13361062GA), Not Applicable (SEE15193270GA), Not Applicable (SEE15423322GA), Not Applicable (SEE15032886GA), Not Applicable (SEE16430039SA), Not Applicable (SEE09370512GA), Not Applicable (SEE11361712GA), 00840682109260SEE17460002SA (SEE17460002SA), 00840682109260SEE17460015SA (SEE17460015SA), 00840682109260SEE17460019SA (SEE17460019SA), Not Applicable (SEE18450022SA), 00840682109260SEE17480008SA (SEE17480008SA), Not Applicable (SEE09530877GA), Not Applicable (SEE11301554GA), Not Applicable (SEE12182225GA), Not Applicable (SEE14432593GA), Not Applicable (SEE10120990GA), 00840682109260SEE18410026SA (SEE18410026SA), Not Applicable (SEE11181422GA), Not Applicable (SEE10250166GA), Not Applicable (SEE16053344GA), 00840682109260SEE18160056SA (SEE18160056SA), 00840682109260SEE18290010SA (SEE18290010SA), 00840682109260SEE18290013SA (SEE18290013SA), 00840682109260SEE18410034SA (SEE18410034SA), 00840682109260SEE18420031SA (SEE18420031SA), 00840682109260SEE17510003SA (SEE17510003SA), Not Applicable (SEE12242295GA), Not Applicable (SEE14171928GA), 01008406821092601118101621SEE18420033SA (SEE18420033SA), Not Applicable (SEE11351666GA), 00840682109260SEE17330002SA (SEE17330002SA), Not Applicable (SEE15193296GA), Not Applicable (SEE12172185GA), Not Applicable (SEE11331600GA), Not Applicable (SEE11361699GA), Not Applicable (SEE15473333GA), Not Applicable (SEE11200793GA), Not Applicable (SEE14412493GA), Not Applicable (SEE10130035GA), Not Applicable (SEE15473338GA), Not Applicable (SEE09510827GA), 00840682109260SEE17270008SA (SEE17270008SA), Not Applicable (SEE16043341GA), Not Applicable (SEE12112113GA), Not Applicable (SEE14101778GA), Not Applicable (SEE16370001SA), Not Applicable (SEE10400380GA), Not Applicable (SEE09230280GA), Not Applicable (SEE14031687GA), 00840682109260SEE17240117SA (SEE17240117SA), Not Applicable (SEE11161345GA), Not Applicable (SEE14222041GA), Not Applicable (SEE13220841GA), Not Applicable (SEE13371112GA), Not Applicable (SEE10400391GA), Not Applicable (SEE09140173GA), Not Applicable (SEE15012857GA), Not Applicable (SEE15373306GA), Not Applicable (SEE09330426GA), Not Applicable (SEE093304426GA), Not Applicable (SEE10100970GA), Not Applicable (SEE14211985GA), Not Applicable (SEE13270889GA), Not Applicable (SEE10450668GA), Not Applicable (SEE11301557GA), Not Applicable (SEE15032869GA), Not Applicable (SEE09530841GA), Not Applicable (SEE090100088GA), Not Applicable (SEE09100088GA), Not Applicable (SEE09130167GA), Not Applicable (SEE10530853GA), Not Applicable (SEE10130037GA), Not Applicable (SEE11211476GA), Not Applicable (SEE09390549GA), Not Applicable (SEE09390550GA), Not Applicable (SEE14512824GA), Not Applicable (SEE15423324GA), Not Applicable (SEE15473336GA), 00840682109260SEE18100030SA (SEE18100030SA), 00840682109260SEE1810003SA (SEE1810003SA), Not Applicable (SEE14021510GA), Not Applicable (SEE14021521GA), Not Applicable (SEE09420596GA), 00840682109260SEE18320021SA (SEE18320021SA), Not Applicable (SEE16252012SA), Not Applicable (SEE16053358GA), 00840682109260SEE18410018SA (SEE18410018SA), Not Applicable (SEE09330415GA), Not Applicable (SEE15113159GA), 00840682109260SEE18420010SA (SEE18420010SA), 00840682109260SEE18420024SA (SEE18420024SA), Not Applicable (SEE09130164GA), Not Applicable (SEE11341624GA), Not Applicable (SEE12520358GA), Not Applicable (SEE12520385GA), Not Applicable (SEE14362378GA), 00840682109260SEE18290004SA (SEE18290004SA), Not Applicable (SEE16430063SA), Not Applicable (SEE11341657GA), Not Applicable (SEE09230270GA), Not Applicable (SEE13371119GA), 00840682109260SEE17430041SA (SEE17430041SA), 00840682109260SEE18240010SA (SEE18240010SA), Not Applicable (SEE10450673GA), Not Applicable (SEE14272142GA), Not Applicable (SEE12430186GA), Not Applicable (SEE12490267GA), Not Applicable (SEE12242284GA), Not Applicable (SEE11361696GA), Not Applicable (SEE11301556GA), Not Applicable (SEE12172187GA), 00840682109260SEE18410017SA (SEE18410017SA), 00840682109260SEE18410019SA (SEE18410019SA), Not Applicable (SEE11141264GA), Not Applicable (SEE15193300GA), 00840682109260SEE18410025SA (SEE18410025SA), 00840682109260SEE18410036SA (SEE18410036SA), 00840682109260SEE18420026SA (SEE18420026SA), 00840682109260SEE18420032SA (SEE18420032SA), Not Applicable (SEE11361695GA), Not Applicable (SEE15153204GA), Not Applicable (SEE09350459GA), Not Applicable (SEE09350464GA), Not Applicable (SEE09350465GA), Not Applicable (SEE09380514GA), Not Applicable (SEE13280917GA), 01008406821092601116102521SEE16430009SA (SEE16430009SA), 01008406821092601116102521SEE16430013SA (SEE16430013SA), 01008406821092601116102721SEE16430058SA (SEE16430058SA), 00840682109260SEE18420012SA (SEE18420012SA), Not Applicable (SEE11171364GA), Not Applicable (SEE09330420GA), Not Applicable (SEE09510800GA), Not Applicable (SEE18390001SA), Not Applicable (SEE10310285GA), Not Applicable (SEE11060961GA), Not Applicable (SEE14372404GA), Not Applicable (SEE16252016SA), Not Applicable (SEE09360474GA), Not Applicable (SEE12420158GA), Not Applicable (SEE12490265GA), Not Applicable (SEE15153200GA), Not Applicable (SEE15193280GA), Not Applicable (SEE15423327GA), 00840682109260SEE17210016SA (SEE17210016SA), Not Applicable (SEE120620007GA), Not Applicable (SEE09330407GA), Not Applicable (SEE12102096GA), Not Applicable (SEE12370069GA), Not Applicable (SEE14492723GA), Not Applicable (SEE12322412GA), Not Applicable (SEE10400428GA), Not Applicable (SEE10420554GA), 00840682109260SEE17390010SA (SEE17390010SA), Not Applicable (SEE10270250GA), Not Applicable (SEE16430059SA), Not Applicable (SEE13371109GA), Not Applicable (SEE11211507GA), Not Applicable (SEE14222004GA), Not Applicable (SEE12370065GA), Not Applicable (SEE10310280GA), 00840682109260SEE17460005SA (SEE17460005SA), Not Applicable (SEE16173376GA), 00840682109260SEE17310012SA (SEE17310012SA), Not Applicable (SEE11241549GA), Not Applicable (SEE12430230GA), Not Applicable (SEE11191423GA), Not Applicable (SEE10400409GA), 00840682109260SEE18410015SA (SEE18410015SA), 00840682109260SEE18450016SA (SEE18450016SA), Not Applicable (SEE15183264GA), Not Applicable (SEE09470659GA), Not Applicable (SEE09470667GA), Not Applicable (SEE15193278GA), 00840682109260SEE18380033SA (SEE18380033SA), 00840682109260SEE18380035SA (SEE18380035SA), Not Applicable (SEE12282349GA), 00840682109260SEE17190011SA (SEE17190011SA), 00840682109260SEE17040002SA (SEE17040002SA), 00840682109260SEE18390009SA (SEE18390009SA), 00840682109260SEE18390011SA (SEE18390011SA), Not Applicable (SEE16272583SA), Not Applicable (SEE09330424GA), Not Applicable (SEE10080926GA), Not Applicable (SEE11361680GA), Not Applicable (SEE09380524GA), Not Applicable (SEE15052912GA), 00840682109260SEE18220010SA (SEE18220010SA), 00840682109260SEE17430047SA (SEE17430047SA), Not Applicable (SEE16241851SA), 00840682109260SEE17480003SA (SEE17480003SA), 00840682109260SEE18040052SA (SEE18040052SA), 00840682109260SEE18040054SA (SEE18040054SA), Not Applicable (SEE10450679GA), Not Applicable (SEE12242290GA), 00840682109260SEE18380039SA (SEE18380039SA), Not Applicable (SEE15113117GA), Not Applicable (SEE12302387GA), Not Applicable (SEE12192239GA), Not Applicable (SEE13481401GA), Not Applicable (SEE14332345GA), 01008406821092601117080121SEE17310001SA (SEE17310001SA), 00840682109260SEE17370001SA (SEE17370001SA), Not Applicable (SEE16053356GA), Not Applicable (SEE11060946GA), Not Applicable (SEE11201461GA), Not Applicable (SEE12051990GA), Not Applicable (SEE14041745GA), Not Applicable (SEE14041750GA), 00840682109260SEE17260005SA (SEE17260005SA), 00840682109260SEE18380043SA (SEE18380043SA), 00840682109260SEE18410021SA (SEE18410021SA), Not Applicable (SEE16410017SA), Not Applicable (SEE16183387GA), Not Applicable (SEE16313482SA), Not Applicable (SEE15113171GA), 00840682109260SEE18240053SA (SEE18240053SA), 00840682109260SEE18240054SA (SEE18240054SA), 00840682109260SEE18240057SA (SEE18240057SA), Not Applicable (SEE13190807GA), Not Applicable (SEE13240863GA), Not Applicable (SEE15153199GA), Not Applicable (SEE16272420SA), Not Applicable (SEE14151873GA), Not Applicable (SEE10250123GA), 00840682109260SEE18380036SA (SEE18380036SA), 00840682109260SEE18390006SA (SEE18390006SA), 00840682109260SEE18390010SA (SEE18390010SA), Not Applicable (SEE15173231GA), Not Applicable (SEE15173244GA), Not Applicable (SEE18380040SA), Not Applicable (SEE18390004SA), 00840682109260SEE18010003SA (SEE18010003SA), Not Applicable (SEE13180778GA), Not Applicable (SEE14031660GA), 00840682109260SEE18330008SA (SEE18330008SA), Not Applicable (SEE09110155GA), Not Applicable (SEE14031706GA), Not Applicable (SEE15173245GA), 00840682109260SEE16410013SA (SEE16410013SA), Not Applicable (SEE14041729GA), Not Applicable (SEE12430227GA), Not Applicable (SEE13120571GA), Not Applicable (SEE13120585GA), Not Applicable (SEE14222011GA), Not Applicable (SEE14222013GA), 00840682109260SEE17100004SA (SEE17100004SA), Not Applicable (SEE15052927GA), Not Applicable (SEE15413317GA), Not Applicable (SEE16183382GA), 00840682109260SEE17460017SA (SEE17460017SA), Not Applicable (SEE14161886GA), 00840682109260SEE17310010SA (SEE17310010SA), 00840682109260SEE18320013SA (SEE18320013SA), 00840682109260SEE17130032SA (SEE17130032SA), Not Applicable (SEE12031945GA), Not Applicable (SEE16241852SA), 00840682109260SEE18380038SA (SEE18380038SA), Not Applicable (SEE11030896GA), Not Applicable (SEE10090945GA), Not Applicable (SEE10090958GA), Not Applicable (SEE10090968GA), Not Applicable (SEE11111226GA), Not Applicable (SEE14432581GA), Not Applicable (SEE15052943GA), Not Applicable (SEE12490254GA), Not Applicable (SEE12490281GA), Not Applicable (SEE12490285GA), Not Applicable (SEE14101792GA), Not Applicable (SEE14101796GA), Not Applicable (SEE14151872GA), Not Applicable (SEE14171921GA), Not Applicable (SEE12162173GA), Not Applicable (SEE18380041SA), Not Applicable (SEE18380042SA), 00840682109260SEE18030013SA (SEE18030013SA), Not Applicable (SEE18380044SA), Not Applicable (SEE16173373GA), Not Applicable (SEE09100080GA), Not Applicable (SEE13150678GA), Not Applicable (SEE15103109GA), Not Applicable (SEE16241846SA), Not Applicable (SEE16390007SA), Not Applicable (SEE16430051SA), 00840682109260SEE18260007SA (SEE18260007SA), 00840682109260SEE18420004SA (SEE18420004SA), Not Applicable (SEE15032894GA), Not Applicable (SEE15032897GA), 00840682109260SEE18500002SA (SEE18500002SA), 00840682109260SEE18500024SA (SEE18500024SA), 00840682109260SEE18500031SA (SEE18500031SA), 00840682109260SEE18500041SA (SEE18500041SA), Not Applicable (SEE09510791GA), Not Applicable (SEE15062978GA), Not Applicable (SEE15062977GA), Not Applicable (SEE15193273GA), Not Applicable (SEE15193294GA), Not Applicable (SEE16390011SA), Not Applicable (SEE18410035SA), Not Applicable (SEE15052939GA), Not Applicable (SEE15113177GA), Not Applicable (SEE09430614GA), Not Applicable (SEE12490299GA), Not Applicable (SEE14332339GA), Not Applicable (SEE13280947GA), Not Applicable (SEE15173232GA), Not Applicable (SEE11071100GA), Not Applicable (SEE12510312GA), Not Applicable (SEE13361079GA), Not Applicable (SEE15173232SA), 00840682109260SEE17430022SA (SEE17430022SA), 00840682109260SEE18360024SA (SEE18360024SA), Not Applicable (SEE13491429GA), Not Applicable (SEE13491433GA), Not Applicable (SEE13491434GA), Not Applicable (SEE13491445GA), 00840682109260SEE18330003SA (SEE18330003SA), 00840682109260SEE18220005SA (SEE18220005SA), Not Applicable (SEE16390006SA), 00840682109260SEE18420006SA (SEE18420006SA), Not Applicable (SEE11101219GA), Not Applicable (SEE11341643GA), Not Applicable (SEE09110154GA), Not Applicable (SEE10420591GA), Not Applicable (SEE10090961GA), Not Applicable (SEE12332442GA), Not Applicable (SEE16173374GA), Not Applicable (SEE16272593SA), Not Applicable (SEE16272596SA), 00840682109260SEE17500013SA (SEE17500013SA), Not Applicable (SEE10420539GA), Not Applicable (SEE13180766GA), 00840682109260SEE17360006SA (SEE17360006SA), 00840682109260SEE17430038SA (SEE17430038SA), 00840682109260SEE17430042SA (SEE17430042SA), 00840682109260SEE18450019SA (SEE18450019SA), Not Applicable (SEE12192238GA), 00840682109260SEE18320028SA (SEE18320028SA), Not Applicable (SEE16053359GA), 00840682109260SEE17310003SA (SEE17310003SA), 00840682109260SEE18240020SA (SEE18240020SA), Not Applicable (SEE12420163GA), Not Applicable (SEE09310399GA), Not Applicable (SEE13010402GA), Not Applicable (SEE09300347GA), Not Applicable (SEE09530885GA), Not Applicable (SEE09380513GA), Not Applicable (SEE11060950GA), Not Applicable (SEE11081161GA), 00840682109260SEE18320001SA (SEE18320001SA), Not Applicable (SEE13150692GA), Not Applicable (SEE10390295GA), Not Applicable (SEE09530879GA), Not Applicable (SEE09530894GA), Not Applicable (SEE10120025GA), Not Applicable (SEE10130041GA), Not Applicable (SEE10130054GA), Not Applicable (SEE10130058GA), Not Applicable (SEE12062009GA), Not Applicable (SEE13120612GA), Not Applicable (SEE16272627SA), Not Applicable (SEE11060960GA), Not Applicable (SEE09510784GA), 00840682109260SEE17440020SA (SEE17440020SA), 00840682109260SEE18100034SA (SEE18100034SA), Not Applicable (SEE11111249GA), Not Applicable (SEE11321577GA), Not Applicable (SEE13060446GA), Not Applicable (SEE10420518GA), Not Applicable (SEE10450665GA), Not Applicable (SEE15473335GA), Not Applicable (SEE12192235GA), 00840682109260SEE18240055SA (SEE18240055SA), 00840682109260SEE18450062SA (SEE18450062SA), Not Applicable (SEE13120592GA), 00840682109260SEE17500021SA (SEE17500021SA), Not Applicable (SEE09530902GA), Not Applicable (SEE15183258GA), 00840682109260SEE17050009SA (SEE17050009SA), Not Applicable (SEE15423331GA), Not Applicable (SEE15183251GA), Not Applicable (SEE15183250GA), Not Applicable (SEE16123364GA), Not Applicable (SEE10310288GA), Not Applicable (SEE12192237GA), Not Applicable (SEE11081180GA), Not Applicable (SEE09530909GA), Not Applicable (SEE09230285GA), Not Applicable (SEE11181420GA), 00840682109260SEE18060004SA (SEE18060004SA), 00840682109260SEE18070003SA (SEE18070003SA), 00840682109260SEE18250026SA (SEE18250026SA), Not Applicable (SEE15312142SA), Not Applicable (SEE10130052GA), Not Applicable (SEE16262152SA), Not Applicable (SEE12192240GA), Not Applicable (SEE12112120GA), Not Applicable (SEE13040421GA), Not Applicable (SEE10270258GA), Not Applicable (SEE13010399GA), Not Applicable (SEE16183384GA), Not Applicable (SEE16173368GA), Not Applicable (SEE10090964GA), Not Applicable (SEE12041956GA), Not Applicable (SEE10400402GA), Not Applicable (SEE10410508GA), Not Applicable (SEE10420521GA), Not Applicable (SEE12051986GA), Not Applicable (SEE15423325GA), Not Applicable (SEE09430603GA), Not Applicable (SEE10310287GA), Not Applicable (SEE11301555GA), Not Applicable (SEE14021508GA), Not Applicable (SEE13481416GA), Not Applicable (SEE11301553GA), Not Applicable (SEE10410505GA), Not Applicable (SEE12262338GA), Not Applicable (SEE12262339GA), Not Applicable (SEE13150687GA), Not Applic able (SEE09360476GA), Not Applicable (SEE10390292GA), 00840682109260SEE14462664GA (SEE14462664GA), Not Applicable (SEE15113147GA), Not Applicable (SEE15113155GA), Not Applicable (SEE09480749GA), Not Applicable (SEE09480739GA), Not Applicable (SEE09490753GA), Not Applicable (SEE09490751GA), Not Applicable (SEE16053348GA), Not Applicable (SEE16053361GA), Not Applicable (SEE16053362GA), Not Applicable (SEE16233393GA), Not Applicable (SEE16233395GA), Not Applicable (SEE16233397GA), Not Applicable (SEE16233399GA), Not Applicable (SEE16344324SA), Not Applicable (SEE17160003SA), Not Applicable (SEE14432582GA), Not Applicable (SEE16213389GA), 00840682109260SEE17300015SA (SEE17300015SA), 00840682109260SEE17300016SA (SEE17300016SA), 00840682109260SEE17310004SA (SEE17310004SA), 00840682109260SEE18200014SA (SEE18200014SA), 00840682109260SEE18510003SA (SEE18510003SA), 00840682109260SEE18510013SA (SEE18510013SA), Not Applicable (SEE14432595GA), Not Applicable (SEE14332342GA), 00840682109260SEE17020028SA (SEE17020028SA), 00840682109260SEE17070027SA (SEE17070027SA), Not Applicable (SEE11161335GA), Not Applicable (SEE14332346GA), Not Applicable (SEE15113161GA), 00840682109260SEE14222015GA (SEE14222015GA), Not Applicable (SEE15193274GA), Not Applicable (SEE15193275GA), Not Applicable (SEE16183381GA), Not Applicable (SEE11071062GA), Not Applicable (SEE11071069GA), Not Applicable (SEE11071091GA), Not Applicable (SEE11071093GA), Not Applicable (SEE11081163GA), Not Applicable (SEE11141294GA), Not Applicable (SEE13451342GA), 00840682109260SEE17450006SA (SEE17450006SA), 00840682109260SEE17110003SA (SEE17110003SA), 00840682109260SEE18380024SA (SEE18380024SA), 00840682109260SEE18380027SA (SEE18380027SA), 00840682109260SEE18380031SA (SEE18380031SA), 00840682109260SEE17150011SA (SEE17150011SA), Not Applicable (SEE11040911GA), Not Applicable (SEE13411211GA), Not Applicable (SEE10250078GA), Not Applicable (SEE13220838GA), Not Applicable (SEE16170870SA), Not Applicable (SEE16123363GA), Not Applicable (SEE10400389GA), Not Applicable (SEE12360039GA), Not Applicable (SEE12082031GA), Not Applicable (SEE16293174SA), Not Applicable (SEE16043343GA), Not Applicable (SEE10090965GA), Not Applicable (SEE10260201GA), Not Applicable (SEE10400414GA), Not Applicable (SEE10400415GA), Not Applicable (SEE13481387GA), Not Applicable (SEE14031599GA), Not Applicable (SEE12282355GA), Not Applicable (SEE13351043GA), Not Applicable (SEE13351045GA), 00840682109260SEE18170006SA (SEE18170006SA), Not Applicable (SEE16053346GA), Not Applicable (SEE11341634GA), Not Applicable (SEE13351048GA), Not Applicable (SEE15173230GA), Not Applicable (SEE15062972GA), 00840682109260SEE17220010SA (SEE17220010SA), Not Applicable (SEE15183262GA), Not Applicable (SEE15062965GA), Not Applicable (SEE16233396GA), Not Applicable (SEE16123365GA), Not Applicable (SEE16383413GA), 01008406821092601116102521SEE16430045SA (SEE16430045SA), 00840682109260SEE17050008SA (SEE17050008SA), 00840682109260SEE17230052SA (SEE17230052SA), Not Applicable (SEE14031659GA), Not Applicable (SEE14041759GA), Not Applicable (SEE15063002GA), Not Applicable (SEE14021517GA), Not Applicable (SEE13120598GA), 00840682109260SEE17490018SA (SEE17490018SA), Not Applicable (SEE10460743GA), Not Applicable (SEE10460699GA), Not Applicable (SEE12370053GA), Not Applicable (SEE09510835GA), Not Applicable (SEE11191439GA), Not Applicable (SEE15193299GA), Not Applicable (SEE15261529SA), Not Applicable (SEE15261535SA), Not Applicable (SEE15093054GA), Not Applicable (SEE15093055GA), Not Applicable (SEE15093056GA), Not Applicable (SEE15093060GA), Not Applicable (SEE15093061GA), Not Applicable (SEE13501478GA), Not Applicable (SEE16324028SA), Not Applicable (SEE16324030SA), 00840682109260SEE18190001SA (SEE18190001SA), 00840682109260SEE18190002SA (SEE18190002SA), 00840682109260SEE18190019SA (SEE18190019SA), 00840682109260SEE18190023SA (SEE18190023SA), Not Applicable (SEE09100149GA), Not Applicable (SEE10250131GA), Not Applicable (SEE10250171GA), Not Applicable (SEE10250190GA), Not Applicable (SEE10390344GA), Not Applicable (SEE10400367GA), Not Applicable (SEE10400368GA), Not Applicable (SEE12340010GA), Not Applicable (SEE12350023GA), Not Applicable (SEE16303406GA), Not Applicable (SEE16303408GA), Not Applicable (SEE16333411GA), Not Applicable (SEE16383415GA), Not Applicable (SEE16383416GA), Not Applicable (SEE16450011SA), 00840682109260SEE17330001SA (SEE17330001SA), Not Applicable (SEE10100971GA), Not Applicable (SEE10100981GA), Not Applicable (SEE10450661GA), Not Applicable (SEE11161343GA), Not Applicable (SEE12410132GA), Not Applicable (SEE09480720GA), Not Applicable (SEE09510786GA), Not Applicable (SEE09510799GA), Not Applicable (SEE09510826GA), Not Applicable (SEE10250188GA), Not Applicable (SEE10260210GA), Not Applicable (SEE11181402GA), Not Applicable (SEE12102089GA), Not Applicable (SEE12102091GA), Not Applicable (SEE12122131GA), Not Applicable (SEE12122139GA), Not Applicable (SEE12430178GA), Not Applicable (SEE15173237GA), Not Applicable (SEE15373312GA), Not Applicable (SEE16233392GA), 00840682109260SEE16233394GA (SEE16233394GA), 00840682109260SEE18430004SA (SEE18430004SA), 00840682109260SEE18430005SA (SEE18430005SA), 00840682109260SEE18450017SA (SEE18450017SA), Not Applicable (SEE10400420GA), Not Applicable (SEE09360480GA), Not Applicable (SEE10130066GA), Not Applicable (SEE11201458GA), Not Applicable (SEE11361722GA), Not Applicable (SEE13190810GA), 00840682109260SEE17340008SA (SEE17340008SA), Not Applicable (SEE18400003SA), 00840682109260SEE18400018SA (SEE18400018SA), 00840682109260SEE13481417GA (SEE13481417GA), 00840682109260SEE18380016SA (SEE18380016SA), Not Applicable (SEE09460642GA), Not Applicable (SEE16241855SA), Not Applicable (SEE16183385GA), 00840682109260SEE18380012SA (SEE18380012SA), 00840682109260SEE17510008SA (SEE17510008SA), 00840682109260SEE17430040SA (SEE17430040SA), Not Applicable (SEE09360485GA), Not Applicable (SEE12322426GA), 00840682109260SEE17480002SA (SEE17480002SA), Not Applicable (SEE10530862GA), Not Applicable (SEE10420516GA), Not Applicable (SEE10420546GA), Not Applicable (SEE10420552GA), Not Applicable (SEE10420555GA), Not Applicable (SEE13371115GA), 00840682109260SEE17350008SA (SEE17350008SA), Not Applicable (SEE09310395GA), Not Applicable (SEE15173240GA), 00840682109260SEE18160035SA (SEE18160035SA), Not Applicable (SEE14222045GA), Not Applicable (SEE14252122GA), 00840682109260SEE17150012SA (SEE17150012SA), Not Applicable (SEE12232251GA), Not Applicable (SEE15413314GA), Not Applicable (SEE11081162GA), Not Applicable (SEE10090950GA), 00840682109260SEE17240119SA (SEE17240119SA), 00840682109260SEE18380014SA (SEE18380014SA), Not Applicable (SEE12440235GA), Not Applicable (SEE12430196GA), Not Applicable (SEE14021502GA), Not Applicable (SEE15062964GA), Not Applicable (SEE16173371GA), 00840682109260SEE17200004SA (SEE17200004SA), 00840682109260SEE18010001SA (SEE18010001SA), 00840682109260SEE18110001SA (SEE18110001SA), 00840682109260SEE18110003SA (SEE18110003SA), Not Applicable (SEE09140175GA), Not Applicable (SEE09150198GA), Not Applicable (SEE09510774GA), Not Applicable (SEE11071068GA), Not Applicable (SEE12041950GA), 00840682109260SEE17480015SA (SEE17480015SA), Not Applicable (SEE10530863GA), Not Applicable (SEE16313481SA), 00840682109260SEE17330011SA (SEE17330011SA), 00840682109260SEE18380015SA (SEE18380015SA), Not Applicable (SEE18380021SA), Not Applicable (SEE15183253GA), Not Applicable (SEE09360477GA), 00840682109260SEE17440007SA (SEE17440007SA), 00840682109260SEE17150008SA (SEE17150008SA), 00840682109260SEE18040019SA (SEE18040019SA), 00840682109260SEE18040026SA (SEE18040026SA), Not Applicable (SEE18390013SA), Not Applicable (SEE18390016SA), 00840682109260SEE17510007SA (SEE17510007SA), Not Applicable (SEE16183383GA), Not Applicable (SEE11361675GA), Not Applicable (SEE16251980SA), 00840682109260SEE17440014SA (SEE17440014SA), Not Applicable (SEE13180723GA), Not Applicable (SEE14422554GA), Not Applicable (SEE15093058GA), Not Applicable (SEE13280921GA), Not Applicable (SEE13481404GA), Not Applicable (SEE14512820GA), Not Applicable (SEE15093051GA), Not Applicable (SEE16053357GA), Not Applicable (SEE16183386GA), 01008406821092601118101121SEE18410020SA (SEE18410020SA), Not Applicable (SEE10090967GA), Not Applicable (SEE13351054GA), Not Applicable (SEE14292238GA), Not Applicable (SEE14292246GA), Not Applicable (SEE14302260GA), Not Applicable (SEE09410585GA), Not Applicable (SEE15173218GA), Not Applicable (SEE10250170GA), Not Applicable (SEE10250128GA), Not Applicable (SEE10250164GA), Not Applicable (SEE11071143GA), Not Applicable (SEE13351035GA), Not Applicable (SEE10250180GA), Not Applicable (SEE16410001SA), Not Applicable (SEE16053353GA), Not Applicable (SEE10250152GA), Not Applicable (SEE13431296GA), Not Applicable (SEE15423326GA), Not Applicable (SEE09020025GA), Not Applicable (SEE09530892GA), Not Applicable (SEE11071013GA), Not Applicable (SEE11361678GA), Not Applicable (SEE11361683GA), Not Applicable (SEE12062006GA), Not Applicable (SEE16380006SA), 00840682109260SEE17080013SA (SEE17080013SA), 00840682109260SEE17150013SA (SEE17150013SA), 00840682109260SEE17300006SA (SEE17300006SA), 00840682109260SEE18390017SA (SEE18390017SA), 00840682109260SEE18390018SA (SEE18390018SA), 00840682109260SEE18390019SA (SEE18390019SA), Not Applicable (SEE14021563GA), Not Applicable (SEE09510810GA), Not Applicable (SEE12112114GA), Not Applicable (SEE14222035GA), Not Applicable (SEE14382455GA), 00840682109260SEE17390011SA (SEE17390011SA), Not Applicable (SEE16390009SA), Not Applicable (SEE15113163GA), Not Applicable (SEE15113167GA), Not Applicable (SEE14452626GA), Not Applicable (SEE10130057GA), Not Applicable (SEE12041957GA), Not Applicable (SEE13411203GA), Not Applicable (SEE14292242GA), 00840682109260SEE18360015SA (SEE18360015SA), Not Applicable (SEE12112117GA), Not Applicable (SEE11181392GA), Not Applicable (SEE14462661GA), Not Applicable (SEE12400120GA), Not Applicable (SEE13180768GA), Not Applicable (SEE14272206GA), Not Applicable (SEE15193298GA), Not Applicable (SEE15183265GA), 00840682109260SEE18330019SA (SEE18330019SA), Not Applicable (SEE10420594GA), Not Applicable (SEE13431298GA), 00840682109260SEE14312298GA (SEE14312298GA), Not Applicable (SEE16183380GA), 01008406821092601118090721SEE18360019SA (SEE18360019SA), Not Applicable (SEE10410484GA), 00840682109260SEE18360018SA (SEE18360018SA), 00840682109260SEE18500004SA (SEE18500004SA), 00840682109260SEE18360010SA (SEE18360010SA), Not Applicable (SEE13130631GA), Not Applicable (SEE15113149GA), Not Applicable (SEE15173233GA), Not Applicable (SEE15173235GA), Not Applicable (SEE15193269GA), Not Applicable (SEE16043342GA), 00840682109260SEE18030008SA (SEE18030008SA), 00840682109260SEE18030017SA (SEE18030017SA), 01008406821092601118110821SEE18450074SA (SEE18450074SA), Not Applicable (SEE11371794GA), Not Applicable (SEE10250146GA), 01008406821092601118090721SEE18360022SA (SEE18360022SA), Not Applicable (SEE09470666GA), Not Applicable (SEE10090966GA), Not Applicable (SEE13000392GA), Not Applicable (SEE15193267GA), 00840682109260SEE17020026SA (SEE17020026SA), Not Applicable (SEE14041736GA), Not Applicable (SEE09040056GA), 00840682109260SEE17400001SA (SEE17400001SA), Not Applicable (SEE09020018GA), Not Applicable (SEE09190224GA), Not Applicable (SEE10400430GA), Not Applicable (SEE09240298GA), 00840682109260SEE18360007SA (SEE18360007SA), 01008406821092601116052521SEE16213388GA (SEE16213388GA), Not Applicable (SEE12062008GA), Not Applicable (SEE09530891GA), Not Applicable (SEE13431283GA), Not Applicable (SEE13481400GA), 00840682109260SEE17300004SA (SEE17300004SA), Not Applicable (SEE09460647GA), 00840682109260SEE17060006SA (SEE17060006SA), 00840682109260SEE18360012SA (SEE18360012SA), Not Applicable (SEE11071089GA), Not Applicable (SEE11071066GA), Not Applicable (SEE09020016GA), Not Applicable (SEE09040067GA), Not Applicable (SEE09230250GA), Not Applicable (SEE9040073GA), Not Applicable (SEE10090963GA), Not Applicable (SEE10120996GA), Not Applicable (SEE10121004GA), Not Applicable (SEE10121015GA), Not Applicable (SEE10420584GA), Not Applicable (SEE11371747GA), Not Applicable (SEE12062012GA), Not Applicable (SEE12062014GA), Not Applicable (SEE12092066GA), Not Applicable (SEE12420171GA), Not Applicable (SEE14452637GA), Not Applicable (SEE14462643GA), Not Applicable (SEE14462660GA), 00840682109260SEE17170007SA (SEE17170007SA), Not Applicable (SEE14462648GA), Not Applicable (SEE11071083GA), 00840682109260SEE18360023SA (SEE18360023SA), Not Applicable (SEE15193304GA), Not Applicable (SEE09330419GA), Not Applicable (SEE09510775GA), 0100840682109260SEE18450064SA (SEE18450064SA), Not Applicable (SEE10250186GA), Not Applicable (SEE10420592GA), Not Applicable (SEE15093057GA), 00840682109260SEE18280006SA (SEE18280006SA), 00840682109260SEE18450066SA (SEE18450066SA), Not Applicable (SEE25093057GA), Not Applicable (SEE09230275C), Not Applicable (SEE09230275GA), Not Applicable (SEE10090954GA), 01008406821092601118090721SEE18360021SA (SEE18360021SA), Not Applicable (SEE09020039GA), Not Applicable (SEE09040059GA), Not Applicable (SEE15093029GA), Not Applicable (SEE15093045GA), Not Applicable (SEE15093048GA), Not Applicable (SEE15093049GA), Not Applicable (SEE15093053GA), 00840682109260SEE18100025SA (SEE18100025SA), 00840682109260SEE18420029SA (SEE18420029SA), Not Applicable (SEE13020409GA), Not Applicable (SEE09100081GA), Not Applicable (SEE13411202GA), Not Applicable (SEE13411209GA), Not Applicable (SEE13431293GA), Not Applicable (SEE10260191GA), Not Applicable (SEE10430655GA), 00840682109260SEE18100032SA (SEE18100032SA).  UPDATE. added No applicable UDI (SEE12122168GA), Not Applicable (SEE09100113GA), Not Applicable (SEE08510008GA).   7. Product Code: SPV, No UDI Number; Serial Number: SPV18300540SA."
    }
  ]
}