{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Walsall",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88732",
      "recalling_firm": "Xstrahl  Limited",
      "address_1": "Unit 2",
      "address_2": "Maybrook Industrial Estate Maybro; Maybrook Road",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of AZ, CA, FL, KY, NV, OR, PA, TN, and TX.  The countries of Austria, Germany, and Switzerland.",
      "recall_number": "Z-0174-2022",
      "product_description": "Concerto User Interface Software provided with the following systems:    (1) Xstrahl 100 Electronic Brachytherapy - 100kV Superficial X-Ray Therapy System;  (2) Xstrahl 150 Electronic Brachytherapy - 150kV Superficial X-Ray Therapy System;  (3) Xstrahl 200 - 225kV Superficial / Orthovoltage X-Ray Therapy System;  (4) Xstrahl 300 - 300kV Orthovoltage X-Ray Therapy System;  (5) Xstrahl X 80 Photoelectric Therapy System.",
      "product_quantity": "25 systems in the U.S.",
      "reason_for_recall": "If a saved treatment plan with 2 opposing beams is edited prior to approval, then Beam 2 is not updated with the changed parameters upon selecting save, resulting in error messages during the treatment and possible mis-treatment.",
      "recall_initiation_date": "20210426",
      "center_classification_date": "20211026",
      "termination_date": "20240126",
      "report_date": "20211103",
      "code_info": "V2.0, V2.1, and V2.2 when 2 opposing beam treatment plans are used."
    }
  ]
}