  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Newark",
      "state": "DE",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77381",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "500 Gbc Dr Ms 514",
      "address_2": "PO BOX 6101",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) Internationally to Austria, Australia, Belgium, Canada,  France, Germany, Great Britain, Italy, Japan, Netherlands, Portugal, Slovakia, Spain, Switzerland and Republic of Korea South",
      "recall_number": "Z-0174-2018",
      "product_description": "Dimension Vista Protein 1 Control, PROT 1 CON L, Siemens Material Number (SMN) 10445919,  Product Code KC715; IVD",
      "product_quantity": "N/A",
      "reason_for_recall": "Complaint of high amount of ceruloplasmin patient results below the reference range (< 20 mg/dL [0.2 g/L].   PROT1 CAL and corresponding PROT1 CON L, M, and H are inaccurately assigned for ceruloplasmin (CER).",
      "recall_initiation_date": "20170523",
      "center_classification_date": "20171128",
      "termination_date": "20180330",
      "report_date": "20171206",
      "code_info": "Lot numbers: 6BQL01, 6BQL01A, 6BQL01B, 6BQL01C, 6BQL01D (exp. date  2017-06-13); 6JQL01A, 6JQL01B, 6JQL01C (exp. date 2018-01-31). ---  Lot # (UDI):  6BQL01 (008427680063336BQL0120170613),   6BQL01A (008427680063406BQM01A20170617),   6BQL01B (008427680063406BQM01B20170617),   6BQL01C (008427680063406BQM01C20170617),   6BQL01D (008427680063406BQM01D20170617),   6JQL01A (008427680063336JQL01A20180131),   6JQL01B (008427680063336JQL01B20180131),   6JQL01C (008427680063336JQL01C20180131)."
    }
  ]
}