  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Santa Clara",
      "address_1": "4551 Great America Pkwy",
      "reason_for_recall": "The diluent tubing used on the instrument appears cloudy or to have a powder/film on the inner and outer surfaces which may result in falsely elevated platelet results.",
      "address_2": "",
      "product_quantity": "22 analyzers with tubing",
      "code_info": "Tubing lot #2018-10-16; Instrument serial numbers 007689, 007690, 007691, 007692, 007705, 007706, 007707, 007708, 007709, 007728, 007694, 007695, 007697, 007766, 007767, 007768, 007765, 007710, 007699, 007700, 007701, and 007703.",
      "center_classification_date": "20191022",
      "distribution_pattern": "Distribution was made to GA, LA, MA, NC, NY, SC, and TX.    There was no military/government distribution.      Foreign distribution was made to Australia, Germany, Greece, Honduras, Indonesia, Italy, and Serbia.",
      "state": "CA",
      "product_description": "Emerald Diluent Tubing, List number 09H50-01, accompanying CELL-DYN Emerald instrument, List number 09H39-01.  The firm name on the label is Abbott Laboratories, Abbott Park, IL.",
      "report_date": "20191030",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Abbott Laboratories",
      "recall_number": "Z-0173-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "83793",
      "termination_date": "20201007",
      "more_code_info": "",
      "recall_initiation_date": "20190509",
      "postal_code": "95054-1208",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}