  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Lake Forest",
      "address_1": "600 N Field Dr",
      "reason_for_recall": "A software malfunction of the infuser when used with a MedNet Custom Drug Library (CDL).  When a new CDL is downloaded and installed on the infuser the infuser can become inoperable and show a blank screen on the User Interface.",
      "address_2": "",
      "product_quantity": "47,152 devices",
      "code_info": "All Plum 360 Infusers with software versions 15.10.00.010/.011/.012 and 15.11.00.017 when  used with an ICU Medical MedNet CDL",
      "center_classification_date": "20191022",
      "distribution_pattern": "US nationwide distribution, and  Canada",
      "state": "IL",
      "product_description": "Plum 360 Infusion System With MedNet/ Smart Card Plug And Play Module, Model # 30010.    Infusion pump.",
      "report_date": "20191030",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "ICU Medical Inc",
      "recall_number": "Z-0171-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "83480",
      "termination_date": "20210427",
      "more_code_info": "",
      "recall_initiation_date": "20180801",
      "postal_code": "60045-4835",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}