  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Skaneateles Falls",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72228",
      "recalling_firm": "Welch Allyn Inc",
      "address_1": "4341 State Street Rd",
      "address_2": "N/A",
      "postal_code": "13153-5300",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide, Puerto Rico and foreign distribution to the following countries:  Australia, Bahamas, Bahrain, Belgium, Bermuda, Cambodia, Canada, Czech Republic, China, Cyprus, Denmark, Egypt, France, Germany, Great Britain, Greece, Guam, Guatemala, Iceland, India, Israel, Japan, Kenya, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Panama, Peru, Philippines, Poland, Russia, Saudi Arabia, Serbia, Singapore, South Africa, Spain, Sri Lanka, Thailand, Turkey, United Arab Emirates.",
      "recall_number": "Z-0169-2016",
      "product_description": "Welch Allyn KleenSpec Corded Illumination System, Vaginal Speculum Lighting System, Model Number 78x Series Illuminator, REF 901070, Catalog Numbers 78800, 73410, 73412, 73414, 73416, 78810, 78812, 78814 and 78816, Rx ONLY",
      "product_quantity": "Domestic: 179,401 units, Foreign: 15,063 units",
      "reason_for_recall": "The firm became aware that the KleenSpec¿ 788 Corded Illuminator has the potential to be incorrectly connected to a commonly available AC power cord instead of the illuminator low voltage transformer power cord which is supplied with the illuminator.  This may result in a shattered lamp bulb and a potential electric shock hazard.",
      "recall_initiation_date": "20151002",
      "center_classification_date": "20151027",
      "termination_date": "20180523",
      "report_date": "20151104",
      "code_info": "all lots"
    }
  ]
}