  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Boston",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90968",
      "recalling_firm": "Rebiscan, Inc.",
      "address_1": "100 Cambridge St Fl 14",
      "address_2": "N/A",
      "postal_code": "02114-2509",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of New Zealand, Belgium.",
      "recall_number": "Z-0168-2023",
      "product_description": "Blinq Pediatric Vision Scanner-Pediatric Vision Scanner-intended for automated detection of misalignment of the visual axes using polarized light.  Model Number BQ830",
      "product_quantity": "63 units",
      "reason_for_recall": "Unable to charge the battery pack- battery packs will allow the batteries to be charged if the voltage of any cell is below 2.5V. When this occurs, the battery  pack will not accept the charge, could vent and possibly catch fire",
      "recall_initiation_date": "20221003",
      "center_classification_date": "20221101",
      "report_date": "20221109",
      "code_info": "UDI: 008648160004  Serial Numbers: B-00001-V thru B-00105-V B-00021 B-00022 B-00036 B-00052 B-00054 thru B-00060 B-00062 thru B-00065 B-00070 B-00071 B-00075 thru B-00077 B-00079 B-00082 thru B-00098 B-00100 thru B-00104 B-00106 B-00109 B-00111 thru B-00113"
    }
  ]
}