  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77456",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution to MN",
      "recall_number": "Z-0168-2018",
      "product_description": "Syngo.plaza, Picture Archiving and Communications System(PACS), Model 10863171      Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific work flows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.",
      "product_quantity": "2 units",
      "reason_for_recall": "Siemens is releasing a non-medical software application LTA Incomplete Archive Check Tool V1.0.  It is intended to eliminate an issue that occurs during syngo.plaza de-archiving from Dicom LTA.  It has been defined that the number of de-archived images is less than the count of the archived images for the series. The issues were discovered through Siemens Global Complaint Handling System.",
      "recall_initiation_date": "20170522",
      "center_classification_date": "20171121",
      "termination_date": "20180702",
      "report_date": "20171129",
      "code_info": "non-medical software application Check Tool V1 .0 (L TA Incomplete Archive Check Tool V1 .0)"
    }
  ]
}