{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86462",
      "recalling_firm": "Smith & Nephew, Inc.",
      "address_1": "150 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1031",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide  Foreign:  AE, AT, AU, BE, BM, BR, CA, CH, CL, CN, CO, CY,  CZ, DE, DK, DO, EC, EE, ES, FI, FR, GB, GE, GR, HK,  HU, IE, IL, IN, IT, JP,KR, LV,MY, NL, NO, PA,PE, PL,  PR, PT, RU, SE, SG, SI, TH, TR, TW, UA,  & ZA",
      "recall_number": "Z-0165-2021",
      "product_description": "OSTEORAPTOR CRVD 2.3 SA UB COBRD BLUE- Absorbable Suture anchor  Product Number: 72203290",
      "product_quantity": "263,874 (total)",
      "reason_for_recall": "Product packaging improper or incomplete seal of the pouch surrounding the sterile product may result in a breach of the sterile barrier created by the pouch",
      "recall_initiation_date": "20200909",
      "center_classification_date": "20201021",
      "report_date": "20201028",
      "code_info": "Batch Numbers:  2012935  2013910  2016229  2017812  2018635  2019990  2021454  2022268  2024711  2025147  2027400  2029394  2030945  2037184  2038460  2038795  2039253  2039444  2045859  2045858  2045860  2045861  2048368  2049884  2048367"
    }
  ]
}