  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-21",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Bloomington",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99875",
      "recalling_firm": "Cook Incorporated",
      "address_1": "750 N Daniels Way",
      "address_2": "",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "",
      "distribution_pattern": "US States: FL, IL, NM, TX, WA. Canada, France, Germany, Italy, Spain, Switzerland.",
      "recall_number": "Z-0163-2027",
      "product_description": "Williams Cystoscopic Injection Needles, used for cystoscopic injection into the urethra, bladder neck, and bladder wall, Reference Part Number 090001-S26, Order Number (GPN) G16112",
      "product_quantity": "97 units",
      "reason_for_recall": "The female Luer lock adapter component in the affected device lots may not have been correctly seated due to misalignment and inadequate torquing during manufacturing. Potential events that may occur as a result of this issue include increased procedural time and inadequate treatment.",
      "recall_initiation_date": "20260916",
      "center_classification_date": "20261015",
      "report_date": "20261021",
      "code_info": "UDI-DI: 00827002161122; Lot Numbers: 16613606, 16951548, 16951548X",
      "more_code_info": ""
    }
  ]
}