  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Allendale",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63272",
      "recalling_firm": "Stryker Spine",
      "address_1": "2 Pearl Ct",
      "address_2": "N/A",
      "postal_code": "07401-1611",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide distribution.",
      "recall_number": "Z-0163-2013",
      "product_description": "Aviator Drill Bit (10 mm, 12 mm, 14 mm, 16 mm,   Manufactured by: Stryker Spine SAS Z.1 Marticot -33610- Cestas France  +33 (0) 5.57.97.06.30  http://www.stryker.com  Distributed in the USA by Stryker Spine  2 Pearl Ct., Allendale NJ 07401-1677 USA  +1-201-750-8000    Drill bits are reusable instruments, sterilized prior to use, used in conjunction with an Aviator Drill Guide to provide a pathway for screw insertion. The aviator Anterior Cervical Plating (ACP) System is intended for use as an aid in cervical spinal fusion and is intended for unilateral fixation.",
      "product_quantity": "6154 units",
      "reason_for_recall": "Aviator drill bits are 10 mm longer than the Reflex Hybrid drill bits and are not interchangeable which could lead to over-drilling by as much as 10 mm.",
      "recall_initiation_date": "20120705",
      "center_classification_date": "20121029",
      "termination_date": "20130128",
      "report_date": "20121107",
      "code_info": "E207315  Catalog number 48770610, 48770612, 48770616  All lots",
      "more_code_info": ""
    }
  ]
}