  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Roseville",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72305",
      "recalling_firm": "LifeHealth, LLC",
      "address_1": "2656 Patton Rd",
      "address_2": "N/A",
      "postal_code": "55113-1136",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: Nationwide   (AL, AR, CA, CO, FL, IL, IN, IA, KS, KY, MD, MA,MI, MN, MS, MO, NE, NJ, NY,NC, OH, OK, PA, TN, TX, VA)    OUS:  ALBANIA, BELGIUM, BOLIVIA, CANADA, CHILE, CHINA, DENMARK. DOMINICAN REPUBLIC, ECUADOR, EGYPT, ENGLAND, ESTONIA, PHILIPPINES, FRANCE, GERMANY, GREECE, ICELAND, INDONESIA, IRELAND, ITALY, JAMAICA, JAPAN, MOROCCO, NETHERLANDS, NORWAY, PERU, PORTUGAL, SOUTH AFRICA, SOUTH KOREA, SPAIN, TAIWAN, TUNISIA, TURKEY, UNITED ARAB EMIRATES, VIETNAM.",
      "recall_number": "Z-0160-2016",
      "product_description": "BG cartridges for IRMA TRUPOINT Blood Gas Analyzer, PN 048103.     BG Cartridge are single use, disposable, in vitro diagnostics intended for point of care (POC), professional, use with the IRMA TRUpoint¿ Blood Analysis System.",
      "product_quantity": "1735 boxes of 25. (43375 Cartridges)",
      "reason_for_recall": "May report values outside the published performance specifications for pH.  After a certain point in shelf life, the pH sensor may give results that are falsely elevated. The discrepancy is greater as shelf life progresses and can be seen on the aqueous controls before the blood.",
      "recall_initiation_date": "20150916",
      "center_classification_date": "20151023",
      "termination_date": "20160321",
      "report_date": "20151104",
      "code_info": "EOSCU  EOTUZ  EOVMW  EOYUL  EPANW  EPCED  EPDVM  EPFOJ  EPIXX  EPKOJ  EPNXA"
    }
  ]
}