      "status": "Terminated",
      "city": "Roseville",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72305",
      "recalling_firm": "LifeHealth, LLC",
      "address_1": "2656 Patton Rd",
      "address_2": "N/A",
      "postal_code": "55113-1136",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: Nationwide   (AL, AR, CA, CO, FL, IL, IN, IA, KS, KY, MD, MA,MI, MN, MS, MO, NE, NJ, NY,NC, OH, OK, PA, TN, TX, VA)    OUS:  ALBANIA, BELGIUM, BOLIVIA, CANADA, CHILE, CHINA, DENMARK. DOMINICAN REPUBLIC, ECUADOR, EGYPT, ENGLAND, ESTONIA, PHILIPPINES, FRANCE, GERMANY, GREECE, ICELAND, INDONESIA, IRELAND, ITALY, JAMAICA, JAPAN, MOROCCO, NETHERLANDS, NORWAY, PERU, PORTUGAL, SOUTH AFRICA, SOUTH KOREA, SPAIN, TAIWAN, TUNISIA, TURKEY, UNITED ARAB EMIRATES, VIETNAM.",
      "recall_number": "Z-0159-2016",
      "product_description": "CC cartridges for IRMA TRUPOINT Blood Gas Analyzer, PN 039903.  The CC Cartridge are single use, disposable, in vitro diagnostics intended for point of care (POC), professional, use with the IRMA TRUpoint¿ Blood Analysis System.",
      "product_quantity": "4659 boxes of 25. (116475 cartridges)",
      "reason_for_recall": "May report values outside the published performance specifications for pH.  After a certain point in shelf life, the pH sensor may give results that are falsely elevated. The discrepancy is greater as shelf life progresses and can be seen on the aqueous controls before the blood.",
      "recall_initiation_date": "20150916",
      "center_classification_date": "20151023",
      "termination_date": "20160321",
      "report_date": "20151104",
      "code_info": "ASBEZ  ASCXE  ASEMB  ASHWQ  ASJMY  ASLGI  ASMXR  ASOQO  ASRXC  ASTQR  ASWZI"
    }
  ]
}