  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Arvada",
      "state": "CO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63361",
      "recalling_firm": "Sorin Group USA, Inc.",
      "address_1": "14401 W 65th Way",
      "address_2": "N/A",
      "postal_code": "80004",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution in the states of: AL, AR, CA, CO, CT, FL, IL, MI, MN, MO, MS, NC, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA and WI.",
      "recall_number": "Z-0157-2013",
      "product_description": "Gish 3/8 x 3/8\" Straight Connector with Luer fitting, Item Numbers: EC2160B, EC2160S, Lot Numbers: 1210100010, 1215100026, 1217000138, Exp. Dates: 4/16/17, 5/31/15, 7/31/15.    Product Usage:  The connectors are indicated for use in connecting flexible tubing used in extracorporeal blood or fluid circuits. The connectors are distributed as sterile stand alones and bulk non-sterile.",
      "product_quantity": "1,640 units",
      "reason_for_recall": "Due to a molding irregularity, certain lots of connectors have the potential to leak fluid or entrain air at the interface of the luer fitting and mating cap.  Entrained air could result in patient injury.",
      "recall_initiation_date": "20120914",
      "center_classification_date": "20121025",
      "termination_date": "20130423",
      "report_date": "20121031",
      "code_info": "Item Numbers: EC2160B, EC2160S, Lot Numbers: 1210100010, 1215100026, 1217000138, Exp. Dates: 4/16/17, 5/31/15, 7/31/15."
    }
  ]
}