  "results": [
    {
      "status": "Ongoing",
      "city": "Sanford",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93018",
      "recalling_firm": "Omega Medical Imaging, Inc.",
      "address_1": "3400 Saint Johns Pkwy Ste 1020",
      "address_2": "N/A",
      "postal_code": "32771-6769",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the country of Pakistan.",
      "recall_number": "Z-0156-2024",
      "product_description": "Omega Medical Imaging Elevating Monitor Suspension, Part Number 1000-0085",
      "product_quantity": "52 units",
      "reason_for_recall": "An actuator separated from the pivot mechanism on a lower monitor boom system column and became detached and was restrained by the installed safety cable at one location.",
      "recall_initiation_date": "20230228",
      "center_classification_date": "20231024",
      "report_date": "20231101",
      "code_info": "all units manufactured between 10-22-2012 and 07-09-2019"
    }
  ]
}