  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Gainesville",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78111",
      "recalling_firm": "Exactech, Inc.",
      "address_1": "2320 NW 66th Ct",
      "address_2": "N/A",
      "postal_code": "32653-1630",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution was made to AL, FL, IN, ME, MN, NJ, NY, OH, OK, VA,  and Puerto Rico.  Foreign distribution was made to Argentina, Australia, Brazil, China, Colombia, France, India, Italy, Luxembourg, The Netherlands, Spain, Switzerland, Tunisia, and United Kingdom.",
      "recall_number": "Z-0156-2018",
      "product_description": "Optetrak Constrained Condylar Offset Screw, Size MM, Ref 208-04-80, Sterile, RX.",
      "product_quantity": "520 devices",
      "reason_for_recall": "There is a design incompatibility between certain sizes of the offset spine stiffener screw and the constrained condylar tibial insert when used in conjunction with an offset tibial tray.",
      "recall_initiation_date": "20170919",
      "center_classification_date": "20171120",
      "termination_date": "20210621",
      "report_date": "20171129",
      "code_info": "All lot numbers"
    }
  ]
}