  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Acton",
      "state": "MA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90987",
      "recalling_firm": "Insulet Corporation",
      "address_1": "100 Nagog Park",
      "address_2": "N/A",
      "postal_code": "01720-3440",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution: US nationwide, Canada, Europe, Australia, and the Middle East.",
      "recall_number": "Z-0154-2023",
      "product_description": "18239: ASM Omnipod Dash PDM, insulin delivery system.    PT-000010: Assembly, Dash Final PDM U100, mg/dL  PT-000011: Assembly, Dash Final PDM U100, mmol/L  PT-000030: Assembly, DASH PDM, Canada    Remote Controller (Personal Diabetes Manager or PDM) associated with Insulin Pump (Pod)      *The part numbers above may be incorporated  into higher level assemblies such as DASH Starter  Kits.",
      "product_quantity": "248,288 US; 167,802 OUS",
      "reason_for_recall": "The firm has become aware of PDM battery issues, including battery swelling, fluid leaking from the battery, and extreme overheating, which may pose a fire hazard. The firm has received a complaint alleging fire as a result of thermal energy from a DASH PDM.",
      "recall_initiation_date": "20221017",
      "center_classification_date": "20221115",
      "termination_date": "20240917",
      "report_date": "20221123",
      "code_info": "Part Number/UDI-DI 18239 : 10385082000009 PT 000030: 10385082000122 PT 000010: 10385082000139 PT 000011: 10385082000146  All Serial Numbers and Lots"
    }
  ]
}