  "results": [
    {
      "status": "Terminated",
      "city": "Dayton",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72011",
      "recalling_firm": "Innovative Sterlization Technologies LLC",
      "address_1": "7625 Paragon Rd Ste A",
      "address_2": "N/A",
      "postal_code": "45459-4063",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Visit",
      "distribution_pattern": "US Nationwide Distribution in the states of IL, IN, LA, OH, PA, TX & WV.",
      "recall_number": "Z-0153-2016",
      "product_description": "OneTray Sealed Sterilization Containers    Product Usage:  A rigid sterilization container used to steam sterilize medical devices",
      "product_quantity": "M2104 - 25 units; M2408 - 84 units",
      "reason_for_recall": "The wrong sterilization time was on the label. The label lists Steam Gravity 10 - 34 minutes at  270 degrees F instead of Steam Gravity 17 - 34 minutes at 270 degrees F.",
      "recall_initiation_date": "20140203",
      "center_classification_date": "20151021",
      "termination_date": "20160204",
      "report_date": "20151028",
      "code_info": "M2104 Lot # 044305, 044379, 044380 & 044418.    M2408 Lot # 044296, 022304, 044381, 044420, 044421 & 044422."
    }
  ]
}