      "status": "Terminated",
      "city": "Billings",
      "state": "MT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71968",
      "recalling_firm": "Windstone Medical Packaging, Inc.",
      "address_1": "1602 4th Ave N",
      "address_2": "N/A",
      "postal_code": "59101-1521",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution to California, Florida, Georgia, Illinois, Kansas, Louisiana, Michigan, New Jersey, Nevada, Texas, Virginia, and Wyoming.",
      "recall_number": "Z-0151-2017",
      "product_description": "Bayview Abdominoplasty Pack, Kit part number 009079-1  convenience custom kits used for general surgery in hospital operating room",
      "product_quantity": "90 kits",
      "reason_for_recall": "The Medtronic COVER LIGHT HANDLE DISPOSABLE GLOVES, Part # 13140, of the custom AMS kit were recalled because gloves may contain splits or holes compromising sterility.",
      "recall_initiation_date": "20150422",
      "center_classification_date": "20161020",
      "termination_date": "20161025",
      "report_date": "20161026",
      "code_info": "Lot Numbers/Expiration Dates:    57718 6/18/2016  60804 5/16/2015  62210 5/7/2015  62802 6/6/2015  66875 11/20/2015  72825 7/29/2016  73912 7/5/2016"
    }
  ]
}