  "results": [
    {
      "status": "Ongoing",
      "city": "Somerville",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90890",
      "recalling_firm": "Ethicon, Inc.",
      "address_1": "Us Highway 22 West",
      "address_2": "N/A",
      "postal_code": "08876",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "International distribution of the country of China.",
      "recall_number": "Z-0150-2023",
      "product_description": "STRATAFIX SPIRAL PGA-PCL UNI UD 20CM USP2-0(M3) S/A SH. indicated for use in soft tissue approximation where the use of absorbable sutures is appropriate.  Product No.: SXMD1B405-12",
      "product_quantity": "2,280 eaches",
      "reason_for_recall": "Packages may contain a PS-1 cutting edge needle and not the intended SH taper point needle may result in more tissue trauma than is anticipated for taper point needles, which may necessitate repair",
      "recall_initiation_date": "20220920",
      "center_classification_date": "20221031",
      "report_date": "20221109",
      "code_info": "UDI-DI: 10705031228061 Lot/Serial No.: B903HPU Expiry Date: 08/21/2026"
    }
  ]
}