  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Morrisville",
      "address_1": "9401 Globe Center Drive",
      "reason_for_recall": "Distortion (noise interference) of the liquid level capacitance can result in air aspiration instead of liquid aspiration. This might cause a false signal (air detection) with possible dispense errors resulting in erroneous results.",
      "address_2": "Suite 140",
      "product_quantity": "2 units",
      "code_info": "Part Number: 10641100, SN: 1407006311.",
      "center_classification_date": "20141029",
      "distribution_pattern": "US Distribution to IL.",
      "state": "NC",
      "product_description": "Freedom EVO-2 100 Base Unit, with Air LiHa Arms    The Freedom EVO is an open automation platform product for general laboratory use. It is intended for routine laboratory tasks, such as general purpose pipetting and general purpose liquid handling and robotic processes.",
      "report_date": "20141105",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Tecan US, Inc.",
      "recall_number": "Z-0149-2015",
      "initial_firm_notification": "Telephone",
      "product_type": "Devices",
      "event_id": "69369",
      "termination_date": "20170710",
      "more_code_info": "",
      "recall_initiation_date": "20140926",
      "postal_code": "27560",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}