  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waltham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66339",
      "recalling_firm": "Fresenius Medical Care Holdings, Inc.",
      "address_1": "920 Winter St",
      "address_2": "N/A",
      "postal_code": "02451-1521",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "USA Nationwide Distribution",
      "recall_number": "Z-0148-2014",
      "product_description": "Fresenius Liberty¿ Cycler  Part Number: RTLR180111    Product Usage:  The Liberty cycler is a portable device designed to perform acute and chronic peritoneal dialysis.",
      "product_quantity": "16,600",
      "reason_for_recall": "Fluid leaking may be observed inside the pump compartment of the Liberty Cycler  and  may result in dialysate becoming contaminated, potentially causing peritonitis",
      "recall_initiation_date": "20131014",
      "center_classification_date": "20131104",
      "termination_date": "20210518",
      "report_date": "20131113",
      "code_info": "All serial numbers"
    }
  ]
}