      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66570",
      "recalling_firm": "Intuitive Surgical, Inc.",
      "address_1": "1266 Kifer Rd Bldg 100",
      "address_2": "N/A",
      "postal_code": "94086-5304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally tAustralia, Austria, Belgium, Brazil, *Bulgaria, Canada, Chile, China, Czech  Republic, Denmark, Finland, France, Germany, Great Britain, India, Israel, Italy, Mexico, Monaco, Netherlands, Norway, Pakistan, Panama, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, South Korea, Sweden, Switzerland, Taiwan, and Turkey.",
      "recall_number": "Z-0146-2014",
      "product_description": "da Vinci Vision System Cart, a component of the  da Vinci Si Surgical System IS3000;  Y1903 Illuminator (Illuminator + Lamp Module);   Y1902 Replacement Lamp Modules.    Intuitive Surgical, Sunnyvale, CA.    As a component of the IS3000 da Vinci Si Surgical System, the 3DHD Vision System provides a high resolution image for the surgeon (at the 3D viewer) and the patient-side assistant (at the touchscreen). The 3DHD endoscope assembly (endoscope and camera head) can be used manually (that is, handheld, as with a traditional endoscopy cart) or can be mounted on the camera arm to assist with preoperative exploration. Also integrated in the Vision System Cart is the Illuminator. The illuminator has a single light source attached to the endoscopy assembly by the light guide cable. It provides illumination inside the body for vision of the surgical field. The Illuminator monitors the life left on the lamp module and notifies the customer when replacement is recommended.",
      "product_quantity": "Standalone: 161 units; With Vision Starter Kit: 240 units; With IS3000 Vision Cart: 283 units",
      "reason_for_recall": "Intuitive has identified that the metalized coating on the IS3000 HD lamp module may be incompatible with the illuminator control board, leading to an erroneous red Preventive Maintenance Message that shows up on monitors.",
      "recall_initiation_date": "20131021",
      "center_classification_date": "20131104",
      "termination_date": "20140904",
      "report_date": "20131113",
      "code_info": "Model # 380990  Y1903 (PN951183-05) Illuminators shipped with lamp module with coated housing.  Serial numbers:  11K1903D7944  11K1903D7949  11K1903D7950  11K1903D7951  11K1903D7952  11K1903D7953  11K1903D7954  11L1903D7989  11L1903D7990  11L1903D7991  11L1903D7992  11L1903D7993  11L1903D7994  11L1903D7995  11L1903D7996  11L1903D7997  11L1903D7998  11L1903D7999  11L1903D8000  11L1903D8001  11L1903D8002  11L1903D8003  11L1903D8004  11L1903D8005  11L1903D8006  11L1903D8007  11L1903D8008  11L1903D8009  11L1903D8010  11L1903D8011  11L1903D8012  11L1903D8013  11L1903D8014  11L1903D8015  11L1903D8016  11L1903D8017  11L1903D8018  11L1903D8019  11L1903D8020  11L1903D8021  11L1903D8022  11L1903D8023  11L1903D8024  11L1903D8025  11L1903D8026  11L1903D8027  11L1903D8028  11L1903D8029  11L1903D8030  11L1903D8031  11L1903D8032  11L1903D8033  11L1903D8034  11L1903D8035  11L1903D8036  11L1903D8037  11L1903D8038  11L1903D8039  11L1903D8040  11L1903D8041  11L1903D8042  11L1903D8043  11L1903D8044  11L1903D8045  11L1903D8046  11L1903D8047  11L1903D8048  11L1903D8049  11L1903D8050  11L1903D8051  11L1903D8052  11M1903D8078  11M1903D8079  11M1903D8080  11M1903D8081  11M1903D8082  11M1903D8083  11M1903D8084  11M1903D8085  11M1903D8086  11M1903D8087  11M1903D8088  11M1903D8089  11M1903D8090  11M1903D8091  11M1903D8092  11M1903D8093  11M1903D8094  11M1903D8095  11M1903D8096  11M1903D8097  11M1903D8098  11M1903D8099  11M1903D8100  11M1903D8101  11M1903D8102  11M1903D8103  11M1903D8104  11M1903D8105  11M1903D8106  11M1903D8107  11M1903D8108  11M1903D8109  11M1903D8127  11M1903D8128  11M1903D8129  11M1903D8130  11M1903D8131  11M1903D8132  11M1903D8133  11M1903D8134  11M1903D8135  11M1903D8136  11M1903D8137  11M1903D8138  11M1903D8139  11M1903D8140  11M1903D8141  11M1903D8142  11M1903D8143  11M1903D8144  11M1903D8145  11M1903D8146  11M1903D8147  11M1903D8148  11M1903D8149  11M1903D8150  11M1903D8151  11M1903D8152  11M1903D8153  11M1903D8154  11M1903D8155  11M1903D8156  11M1903D8157  11M1903D8158  12A1903D8292  12A1903D8293  12A1903D8294  12A1903D8295  12A1903D8296  12A1903D8297  12A1903D8298  12A1903D8299  12A1903D8300  12A1903D8301  12A1903D8302  12A1903D8303  12A1903D8304  12A1903D8305  12A1903D8306  12A1903D8307  12A1903D8308  12A1903D8309  12A1903D8310  12A1903D8311  12A1903D8312  12A1903D8313  12A1903D8314  12A1903D8315  12A1903D8316  12A1903D8317  12A1903D8318  12A1903D8319  12A1903D8320  12A1903D8321  12A1903D8322  12A1903D8323  12A1903D8346  12A1903D8347  12A1903D8348  12A1903D8349  12A1903D8350  12A1903D8351  12A1903D8352  12A1903D8353  12A1903D8354  12A1903D8355  12A1903D8356  12A1903D8357  12A1903D8358  12A1903D8359  12A1903D8360  12A1903D8361  12B1903D8418  12B1903D8419  12B1903D8420  12B1903D8421  12B1903D8422  12B1903D8423  12B1903D8424  12B1903D8425  12B1903D8426  12B1903D8427  12B1903D8428  12B1903D8429  12B1903D8430  12B1903D8431  12B1903D8432  12B1903D8433  12C1903D8434  12C1903D8435  12C1903D8436  12C1903D8437  12C1903D8438  12C1903D8439  12C1903D8440  12C1903D8441  12C1903D8442  12C1903D8443  12C1903D8444  12C1903D8445  12C1903D8446  12C1903D8447  12C1903D8448  12C1903D8449  12C1903D8478  12C1903D8479  12C1903D8480  12C1903D8481  12C1903D8482  12C1903D8483  12C1903D8484  12C1903D8485  12C1903D8486  12C1903D8487  12C1903D8488  12C1903D8489  12C1903D8490  12C1903D8491  12C1903D8492  12C1903D8493  12C1903D8519  12C1903D8520  12C1903D8521  12C1903D8522  12C1903D8523  12C1903D8524  12C1903D8525  12C1903D8526  12C1903D8527  12C1903D8528  12C1903D8529  12C1903D8530  12C1903D8531  12C1903D8532  12C1903D8533  12C1903D8534  12D1903D8573  12D1903D8574  12D1903D8575  12D1903D8576  12D1903D8577  12D1903D8578  12D1903D8579  12D1903D8580  12D1903D8581  12D1903D8582  12D1903D8583  12D1903D8584  12D1903D8585  12D1903D8586  12D1903D8587  12D1903D8588  12D1903D8589  12D1903D8590  12D1903D8591  12D1903D8592  12D1903D8593  12D1903D8594  12D1903D8595  12D1903D8596  12D1903D8597  12D1903D8598  12D1903D8599  12D1903D8600  12D1903D8601  12D1903D8602  12D1903D8603  12D1903D8604  12D1903D8681  12D1903D8682  12D1903D8683  12D1903D8684  12D1903D8685  12D1903D8686  12D1903D8687  12D1903D8688  12D1903D8689  12D1903D8690  12D1903D8691  12D1903D8692  12D1903D8693  12D1903D8694  12D1903D8695  12D1903D8696  12E1903D8729  12E1903D8730  12E1903D8731  12E1903D8732  12E1903D8733  12E1903D8734  12E1903D8735  12E1903D8736  12E1903D8737  12E1903D8738  12E1903D8739  12E1903D8740  12E1903D8741  12E1903D8742  12E1903D8743  12E1903D8744.    Y1902 (PN 950093-05) Replacement lamp module with coated housing;  Serial numbers:  1HJD9411  1HJD9412  1HJD9415  1HJD9421  1HJD9422  1HJD9424  1HJD9465  1HJD9466  1HJD9467  1HJD9469  1HJD9470  1HJD9472  1HJD9473  1HJD9474  1HJD9475  1HJD9476  1HJD9477  1HJD9478  1HJD9483  1HJD9498  1KJD1554  1KJD1555  1KJD1556  1KJD1557  1KJD1558  1KJD1559  1KJD1560  1KJD1561  1KJD1562  1KJD1563  1KJD1564  1KJD1565  1KJD1566  1KJD1579  1KJD1580  1KJD1581  1KJD1582  1KJD1583  1KJD1584  1KJD1586  1KJD1588  1KJD1589  1KJD1590  1KJD1591  1KJD1592  1KJD1593  1KJD1594  1KJD1595  1KJD1596  1KJD1597  1KJD1599  1KJD1600  1KJD1601  1KJD1602  1KJD1603  1KJD1604  1KJD1605  1KJD1606  1KJD1607  1KJD1609  1KJD1626  1KJD1627  1LJD1831  1LJD1832  1LJD1833  1LJD1834  1LJD1836  1LJD1837  1LJD1839  1LJD1840  1LJD1841  1LJD1842  1LJD1845  1LJD1846  1LJD1847  1LJD1850  1LJD1852  1LJD1853  1LJD1854  1LJD1855  1LJD1856  1LJD1858  1LJD1859  1LJD1861  1LJD1862  1LJD1863  1LJD1864  1LJD1865  1LJD1866  1LJD1867  1LJD1868  1LJD1869  1LJD1872  1LJD1873  1LJD1874  1LJD1875  1LJD1876  1LJD1877  1LJD1878  1LJD1879  1LJD1881  1LJD1882  1LJD1883  1LJD1884  1LJD1886  1LJD1887  1LJD1888  1LJD1889  1LJD1891  1LJD1892  1LJD1893  1LJD1894  1LJD1895  1LJD1897  1LJD1898  1LJD1899  1LJD1901  1LJD1903  1LJD1904  1LJD1905  1LJD1906  1LJD1907  1LJD1908  1LJD1909  1LJD1911  1LJD1912  1LJD1916  1LJD1917  1LJD1918  1LJD1919  1LJD1921  1LJD1922  1LJD1923  1LJD1924  1LJD1925  1LJD1926  1LJD1927  1LJD1928  1LJD1929  1LJD1930  1LJD1931  1LJD1932  1LJD1933  1LJD1935  1LJD1936  1LJD1937  1LJD1938  1LJD1939  1LJD1941  1LJD1942  1LJD1943  1LJD1944  1LJD1945  1LJD1946  1LJD1947  1LJD1948  1LJD1949  1LJD1950  1LJD1952  1LJD1953  1LJD1954  1LJD1955  1LJD1956  1LJD1957  1LJD1958  1LJD1959  1LJD1960  1LJD1961  1LJD1962  1LJD1963  1LJD1964  1LJD1966  1LJD1967  1LJD1968  1LJD1969  1LJD1970  1LJD1971  1LJD1972  1LJD1973  1LJD1974  1LJD1975  1LJD1976  1LJD1977  1LJD1978  1LJD1979  1LJD1980  1MJD2854  1MJD2855  1MJD2856  1MJD2858  1MJD2859  1MJD2862  1MJD2863  1MJD2864  1MJD2865  1MJD2866  1MJD2867  1MJD2868  1MJD2869  1MJD2870  1MJD2871  1MJD2872  1MJD2873  1MJD2874  1MJD2876  1MJD2877  1MJD2878  1MJD2879  1MJD2880  1MJD2881  1MJD2882  1MJD2884  1MJD2885  1MJD2886  1MJD2888  1MJD2889  1MJD2891  1MJD2893  1MJD2896  1MJD2908  1MJD2909  1MJD2910  1MJD2911  1MJD2912  1MJD2913  1MJD2914  1MJD2915  1MJD2916  1MJD2917  1MJD2919  1MJD2920  1MJD2921  1MJD2922  1MJD2923  1MJD2924  1MJD2925  1MJD2926  1MJD2927  1MJD2928  1MJD3005  1MJD3007  1MJD3008  1MJD3009  1MJD3011  1MJD3012  1MJD3014  1MJD3016  1MJD3017  1MJD3021  1MJD3022  1MJD3023  1MJD3025  1MJD3027  1MJD3028  1MJD3029  1MJD3030  1MJD3031  1MJD3033  1MJD3034  1MJD3035  1MJD3037  1MJD3042  2AJD3454  2AJD3455  2AJD3457  2AJD3459  2AJD3460  2AJD3461  2AJD3462  2AJD3463  2AJD3465  2AJD3466  2AJD3467  2AJD3468  2AJD3469  2AJD3471  2AJD3472  2AJD3476  2AJD3477  2AJD3480  2AJD3481  2AJD3484  2AJD3485  2AJD3489  2AJD3491  2AJD3492  2AJD3493  2AJD3494  2AJD3495  2AJD3496  2AJD3497  2AJD3498  2AJD3500  2AJD3501  2AJD3502  2AJD3503  2AJD3504  2AJD3505  2AJD3506  2AJD3507  2AJD3508  2AJD3509  2AJD3510  2AJD3511  2AJD3512  2AJD3513  2AJD3518  2AJD3524  2AJD3526  2AJD3528  2AJD3533  2AJD3534  2AJD3535  2AJD3536  2AJD3537  2AJD3538  2AJD3539  2AJD3541  2AJD3542  2AJD3543  2AJD3544  2AJD3545  2AJD3546  2AJD3547  2AJD3548  2AJD3549  2AJD3550  2AJD3551  2AJD3552  2AJD3558  2AJD3559  2AJD3561  2AJD3562  2AJD3563  2AJD3564  2AJD3566  2AJD3567  2AJD3568  2AJD3569  2AJD3570  2AJD3572  2AJD3573  2AJD3575  2AJD3576  2AJD3578  2AJD3579  2AJD3580  2AJD3581  2AJD3582  2AJD3583  2AJD3584  2AJD3588  2AJD3589  2AJD3609  2AJD3612  2AJD3613  2AJD3614  2AJD3615  2BJD4398  2BJD4399  2BJD4400  2BJD4401  2BJD4402  2BJD4404  2BJD4406  2BJD4408  2BJD4409  2BJD4412  2BJD4413  2BJD4415  2BJD4416  2BJD4425  2BJD4426  2BJD4428  2BJD4431  2BJD4432  2BJD4433  2BJD4434  2BJD4435  2BJD4436  2BJD4437  2BJD4438  2BJD4439  2BJD4440  2BJD4441  2BJD4443  2BJD4444  2BJD4445  2BJD4446  2BJD4448  2BJD4449  2BJD4450  2BJD4451  2BJD4452  2BJD4453  2BJD4454  2BJD4455  2BJD4456  2BJD4457  2BJD4473  2BJD4474  2BJD4475  2BJD4476  2BJD4477  2BJD4490  2BJD4538  2BJD4539  2BJD4540  2BJD4541  2BJD4542  2BJD4543  2BJD4544  2BJD4545  2BJD4546  2BJD4547  2BJD4548  2BJD4549  2BJD4550  2BJD4551  2BJD4552  2BJD4553  2BJD4554  2BJD4555  2BJD4557  2BJD4579  2BJD4580  2BJD4581  2BJD4582  2BJD4583  2BJD4584  2BJD4585  2BJD4586  2BJD4587  2BJD4588  2BJD4589  2BJD4590  2BJD4591  2BJD4592  2BJD4593  2BJD4594  2BJD4595  2BJD4596  2BJD4597  2BJD4619  2BJD4620  2BJD4622  2BJD4623  2BJD4624  2BJD4625  2BJD4627  2BJD4628  2BJD4629  2BJD4630  2BJD4631  2BJD4632  2BJD4633  2BJD4634  2BJD4635  2BJD4677  2BJD4678  2BJD4679  2BJD4680  2BJD4681  2BJD4682  2BJD4683  2BJD4686  2BJD4687  2BJD4688  2BJD4689  2BJD4690  2BJD4691  2BJD4718  2BJD4719  2BJD4722  2BJD4723  2BJD4750  2BJD4752  2BJD4753  2BJD4754  2BJD4755  2CJD4897  2CJD4898  2CJD4899  2CJD4902  2CJD4903  2CJD4906  2CJD4909  2CJD4910  2CJD4911  2CJD4912  2CJD4913  2CJD4948  2CJD4950  2CJD4951  2CJD4953  2CJD4954  2CJD4955  2CJD4957  2CJD4958  2CJD4959  2CJD4960  2CJD4961  2CJD4963  2CJD4964  2CJD4975  2CJD4976  2CJD4979  2CJD4980  2CJD4981  2CJD4982  2CJD4983  2CJD4984  2CJD4985  2CJD4987  2CJD4988  2CJD4989  2CJD4990  2CJD4991  2CJD4992  2CJD5038  2CJD5051  2CJD5700  2CJD5743  2CJD5755  2CJD5760  2CJD5762."
    }
  ]
}