  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Edison",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66353",
      "recalling_firm": "ITC-Nexus Dx",
      "address_1": "23 Nevsky St",
      "address_2": "N/A",
      "postal_code": "08820-2425",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "Nationwide Distribution including Arizona andTexas.    .",
      "recall_number": "Z-0137-2014",
      "product_description": "Surgicutt fully automated incision making device for bleeding time determination.  ITC 8 Olsen Avenue, Edison, NJ  08820.  732.548.5700.  www.itcmed.com.  A subsidiary of Thoratec Corporation.    The device is a spring loaded cam driven device that provides standardized incisions of 5 mm length and 1 mm depth. This fully automated instrument is used along with blotting paper for bleeding time determinations.",
      "product_quantity": "6 boxes (Each box contains 50 Surgicutt Devices plus 55 sheets of Blotting Paper).",
      "reason_for_recall": "ITC has determined that the blade in their Surgicutt with Blotting Paper (Product Code: SUB50I; Lot: AK301M) devices may be loose.",
      "recall_initiation_date": "20130822",
      "center_classification_date": "20131104",
      "termination_date": "20140609",
      "report_date": "20131113",
      "code_info": "Product Code: SUB50I; Lot No. AK301M. Lot Expiration Date 8/31/2017"
    }
  ]
}