      "status": "Terminated",
      "city": "Gainesville",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78111",
      "recalling_firm": "Exactech, Inc.",
      "address_1": "2320 NW 66th Ct",
      "address_2": "N/A",
      "postal_code": "32653-1630",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution was made to AL, FL, IN, ME, MN, NJ, NY, OH, OK, VA,  and Puerto Rico.  Foreign distribution was made to Argentina, Australia, Brazil, China, Colombia, France, India, Italy, Luxembourg, The Netherlands, Spain, Switzerland, Tunisia, and United Kingdom.",
      "recall_number": "Z-0136-2018",
      "product_description": "Optetrak Offset Tibial Tray and Screws, Size 3F/3T, Alpha, Ref 208-04-33, Sterile, RX.",
      "product_quantity": "683 devices",
      "reason_for_recall": "There is a design incompatibility between certain sizes of the offset spine stiffener screw and the constrained condylar tibial insert when used in conjunction with an offset tibial tray.",
      "recall_initiation_date": "20170919",
      "center_classification_date": "20171120",
      "termination_date": "20210621",
      "report_date": "20171129",
      "code_info": "All lot numbers"
    }
  ]
}